U. S. Deparhnent of Commerce National Bureau of Standards Resear ch Paper RP1883 Volume 40, May 1948 Part of the Journal of Research of the National Bureau of Standards Absorption of X-Rays in Air By Frank H. Day and Lauriston S. Taylor Studies have been unde r taken to d ete rmine the ab so rption in air of X-rays produced by voltages from 10 to 200 kv (con stant potential) for variou s ini tial filtration s. A free- air ion ization chamber on tracks is used to obtain ab so rption data over a 150-cm di st ance . No appreciabl e change in qua li ty of X -rays is observed in t his di sta nce becau se of the ail' abso rp ti on. H ence, for s hort di stances, effective ail' absorp tion coefficients fo r h ete rogeneous X- rays of var ious qua li ties have been det e rmin ed. 1. Introduction Th e use, in recent years, of beryllium-window X-ray tubes with their high-intensity outputs of soft X-rays has developed a n eed for air-absorption corr ections where di stance is a factor in X -r ay exposur e m eas ur em ent s. Such, for example, is the case where the inverse-square law is used in co mputin g the expos ur e at a point when the actual measur em ent is of necessity mad e at some other point. Th ere is also need for corr ections to be applied to the fr ee-air type of ionization cham bel' du e to air absorption b etween the limiting dia- phragm of the ch amb er and the front ed ge of the collector plates . Since the exposur e rat e is specifi ed in the phtIle of the limiting diaphragm, air-absorp- tion along the pa th between the diaphragm and the collector plat e result s in the r ea.dings of the cha.mber being too low. This distance in the case of the guarded-field fr ee -air standard for calibra- tions up to 200 kv at the Bur eau of Sta.ndards is 21.1 em, a nd for unfil tered radiations produ ced at, sa.y, 50 kv, the necessary corr ections may be considerable. II. Apparatus and Experimental Procedure A diagram of the experimental arrangement is shown in figure 1, where a mobil e, fr ee-air- guard-plat e ionization chamber is indicated. Th e opening in its lead shie ld is 10 cm in diameter. This opening is adequate to accommodate, at any point along the track, the total area of X-rays determined by the fix ed, limiting diaphragm, the diameter of which is 1.75 cm. The e ff ec tive width Absorption of X-Rays in Air of the collector plat es is 10 .2 cm , and the diameter of the beam with the chamb er in its farth est position from the tube (position B, fig . 1) is 6.5 cm . The intercIectrode separation is a fix ed di stan ce of 12. 7 cm , and this is insufficient spacing to allow for the produ ction and collection of all seco ndar y electrons associated with a primary beam 6.5 cm in diameter. H ence, th ere is a slight lo ss of ionization, which earlier st udi es show does not exceed 101' 2 percent . l No corr ection has b een applied for this loss. Th e char ge co llected in the ioniz at ion chamb er at various points along the track is measured with a capacity co mp ensator ancl string electrometer 2 with an accura cy of ± 1 percent. Th e exposur e of the chamb er is controlled by a t, imer operating a lead shutl er. Th e timer is actuated by an accurately co ntrolled 60-cycle signal from the Radio Section of the Bur eau of Standards. Tungsten- target X-ray tubes with l ow inh er ent filtration are used in conjunction with various types and thicknesses of initial filtration. Th e term "initial filtr ation ," as here used, refers to filters of aluminum, co pper, or lead introdu ced b et ween the X-ray tube and the co lumn of air over which air-absorption measur ements are mad e. The window thickness of the cerium glass in one X -ray tube used is 1.29 mm and of the beryllium in another tube, 1.5 mm . Power supplied to the X-ray tube is furnished by a kenotron rectifi er with a r esistance-capaci- tance filter incorporated to reduce the ripple to I Lauri ston S. Tay lor, George SingC/', and A. L. Ch ar lio Ll , Am. J. Roent. 41, 256 (1939). ' Lauri s ton S. 'r ay lor, R esearch 6, 80i (1931) RP 306 . 393

Transcript of Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately...

Page 1: Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately 1 percent. The difference would not be greater than this in the case of 5O-kv X-rays,

U. S. Deparhnent of Commerce National Bureau of Standards

Research Paper RP1883 Volume 40, May 1948

Part of the Journal of Research of the National Bureau of Standards

Absorption of X-Rays in Air By Frank H . Day and Lauriston S. Taylor

Studies have been under taken to determine the absorption in air of X-rays produced by voltages from 10 to 200 kv (constant potential) for various ini tial filtrations. A free-air ion ization chamber on t r acks is used to obtain absorption data over a 150-cm distance . N o appreciable change in quali ty of X-rays is observed in t his distance because of the ail' absorp tion. H ence, for short distances, effective ail' absorpt ion coefficients fo r heterogeneous X -rays of var ious quali ties have been determined .

1. Introduction

The use, in recent years, of beryllium-window X-ray tubes with their high-intensity outputs of soft X-rays has developed a need for air-absorption corrections where distance is a factor in X -ray exposure measurements. Such, for example, is the case where the inverse-square law is used in computing the exposure at a point wh en the actual measurement is of necessity made at some other point . There is also need for corrections to be applied to the free-air type of ionization cham bel' du e to air absorption between the limiting dia­phragm of the chamber and th e front edge of the collector plates . Since the exposure rate is specifi ed in the phtIle of the limiting diaphragm, air-absorp­tion along the path between the diaphragm and the collector plate results in the r ea.dings of the cha.mber being too low. This distance in the case of the guarded-field free-air standard for calibra­tions up to 200 kv at the Bureau of Sta.ndards is 21.1 em, and for unfiltered radiations produced at, sa.y, 50 kv, the necessary corrections may be considerable.

II. Apparatus and Experimental Procedure

A diagram of the experimental arrangement is shown in figure 1, wh ere a mobile, free-air­guard-plate ionization chamber is indicated. The opening in its lead shield is 10 cm in diameter . This opening is adequate to accommodate, at any point along the track, th e total area of X -rays determined by the fixed, limiting diaphragm, the diameter of which is 1.75 cm. The effective width

Absorption of X-Rays in Air

of the collector plates is 10.2 cm, and the diameter of the beam with the chamber in its farthest position from the tube (position B, fig . 1) is 6.5 cm . The intercIectrode separation is a fixed distance of 12.7 cm, and this is insufficient spacing to allow for the production and collection of all secondary electrons associated with a primary beam 6.5 cm in diameter. H ence, there is a slight loss of ionization, which earli er studies show does not exceed 101' 2 percent.l No correction has been applied for this loss.

The charge collected in the ionization chamber at various points along the track is measured with a capacity compensator ancl string electrom eter 2

with an accuracy of ± 1 percent. The exposure of the chamber is controlled by a t,imer operating a ~-in. lead shutler. The timer is actuated by an accurately controlled 60-cycle signal from the Radio Section of the Bureau of Standard s.

Tungsten-target X -ray tubes with low inherent filtration are used in conjunction with various types and thicknesses of initial filtration. The term "initial filtration," as here used, r efers to filters of aluminum, copper, or lead introduced between the X -ray tube and the column of air over which air-absorption measurements are made. The window thickness of the cerium glass in one X -ray tube used is 1.29 mm and of the beryllium in another tube, 1.5 mm.

Power supplied to the X-ray tube is furnished by a kenotron r ectifier with a resistance-capaci­tance filter incorporated to reduce the ripple to

I Lauriston S. Taylor , George SingC/', and A. L . CharlioLl , Am. J . Roent. 41, 256 (1939) .

' Lauris ton S. 'r aylor, B S~J . R esearch 6, 80i (1931) RP306.

393

Page 2: Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately 1 percent. The difference would not be greater than this in the case of 5O-kv X-rays,

LEAD PROTECTIVE ~USING I 74 CM

I 45 CM

I 150 CM

I I I

! FI XED I POSITION (A) - IONIZAT ION CHAMBER

- '0'''''' '"\ LIMITING

~PHRAGM I LEAD GLASS I GUARD COL LECTOR

""""'1\ ,.,,, '""' 1 . LEAD , PLATE PLATE

r----~~~~~~~-~J-~---~=~~-- - -l SHUT TER- a ~ GUARD PLATE J i r -1 i l . r I :0. .... ;~cc . ,'-::~~:~' ~ -rr ~ U L ~-I !'--LEAD '1 I I 'J i !

FILTERS SHIELD ! i ~- It :

t,~====-_-::-=h::~_-=-_-_-_-L _ -:rJ n r j

.-- A

" II :: TO "-TABLE ._----',i--ELECTROMETER I' ,:

~ Vvr

• F IGURE 1. Experimental arrangement for ail' abs01'ption meaSltrements.

0.05 p ercent per milliampere current drain. The t ube currents used are not in excess of 10 mao

The voltage input to the rectifier units is con­stant to 0.25 percen t or better , th e 208-volt com­m ercial supply being fed tlu'ough a voltage stabilizer. The voltage to the rectifier units is th en varied by induction r egulator control. It is estimated that the absolute direct-current voltage applied to the X-ray tubes is known to an accuracy of 1 or 2 percen t.

The filament voltage of the X -ray tubes is s tabilized by an electronic volt age stabilizer in the ground circuit feeding tlu'ough an insulating transformer. The output of the stabilizer is also manually controlled by an adjustable-ratio au to­transformer to reduce long-period fluctuations.

In view of the above-describ ed stabilizat ion, the variation in quantity of X -ray output over a p eriod of exposure of 10 seconds or more is not grea ter than 0.5 percen t . This is also verified from previous experience on flu ctuations in X -ray out­put of this equipmen t, as determined by the Bureau 's standard chamb er for measuremen ts up to 200 kv.

III . Observations and Calculations

Und er steady opera ting condit ions of the X-ray t ube, the ionization chamber is exposed for a given number of milliampere-seconds at successive points along the track The ratio of the ionization cur­r ents measured a t two points along the track is a m easure of the air absorp tion b etween these points, inasmuch as the total flux passing tlu'ough the

394

limi ting diaphragm is measured by the chamber , excep t for th e par t lost by air absorption .

11{easurements are first made, using the beam in its minimum-filter condition. Fil tel'S of alumi­num, copper , or lead are then interposed in the beam up to the point where the beam is rendered sufficiently homogeneous, so tha t fur ther addit ions of filt ers will no t lower the air-absorption coefficien t by an appreciable amount. This is illustrated (fig. 2) by th e semilogarithmic plots of X -ray

10

0.9

~ ~ ~

~ 6 0 MM AL

~ ~ ~ ---.::::; -,- -.::::.;

~ :::::::------

-------= -- 2.0 MM AL

\~ "'-... ~ -0::: ;----.: ~ r----- 10 MM AL

~

0 8

>-!: U> z '" 0 .7 >-~

>-« 0:

"- 0.6 0

z S'

~ '~ ~ ~ ~ ----- ----- ---- -----"- ---0.5 MM AL

~. ~ ""- ~ ............ r--... -----OA MM AL

'" ~ "" "'::~ K 0.3 MM AL

~ ~ '" ~ "- ""- 0.2 MM AL

~ ~ ~ "" "" "" ~

~ 0.1 MM AL >-

" « a: "- "" ~ ~ ~ 0 .05 MM AL

0 .5

~ ~ 0.02 MM AL

"- MIN IMUM FIL TER

0 . 4 9 25 50 75 100 125 150

AIR ABSORPT ION DISTAN CE ( CENT IMETER S )

FIGURE 2. Effect of aluminum fi ltration on air absorption of 50-kv X -rays produced by beryllium-window X -ray I

tube with inherent filtration 1.5 mm Be.

Journal of Research

Page 3: Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately 1 percent. The difference would not be greater than this in the case of 5O-kv X-rays,

intensity as a function of the length of the ab­sorbing air path for the 50 lev X rays from the beryllium-window tub C'. H ere i t is observed that fi ltrations ill excess of 2 mm of aluminum will not greatly affect the air absorption.3 Similarly, it is found that for any voltage up to 200 lev, filtrations of 6 mm of AI, 0.2 mm of Ou, or 0.1 mm of Pb are sufficien t for the attainment of a prac­tical lower limit to the air absorption.

These studies are made with X rays of 200 kv and less, to voltages as low as transmission through filters will allow. Minimum-filter conditiollP allow observations on X rays as low as 7.5 kv h'om a beryllium-window tube. Even these 7 .5-kv X rays show no great change in quality due to absorption in the air itself over the distance of observation, as indicated by a ncar straigh t-line semilogarithmic plot in figure 3. Since the beam is effectively

1 0

09 \ \

08 \

0 7

~

\ 1\

I 06

\ 05

\ 1\ ,.

t-in z 04 \ '" t-~ " ,. '" '" x 0.3 .... 0

z

\ .\

Q t-U

'" cr .... 0 .2

\ \

\ 0.1

\ 25 50 75 100 125

AIR AS SORPTI ON DIST AN CE (CENT IM E TE R5 )

FIGURE 3. Air absorption of 7.5-kv X -rays produced by beryllium-window tube wilh inherent filtration 1.5 mm

Be.

3 The curve for 6-mm filt rat io,r was not obsen 'ed at this voltage because of tbe ex tremely long exposure which would be necessary. At 100 k v, however, the di fference between the 2-mm and tbe 6-mm curves is approximately 1

percent . The difference would not be greater than t his in the case of 5O-kv X-rays, for the same amoun t of filter would render t he beam more homo­geneous in the latter case.

Absorption of X-Rays in Air

monochromatic, it is permi sible to prepare tables of linear absorption coefficients involving a in crlo term 4 for air under normal atmospheric condi tions, (760 mm of mercury pressure and 22° 0 ), the e to be applied over short distances.

Tables 1, 2, and 3 give effective Ii neal' absorp­tion coefficien ts , J.k (em- I) for ail' , where J.k i used in the formula , l x = l oe-"x, I x being the intensity at a distance X cm beyond the point where the original intensity, 10 , is measured. The X-ray beam is here considered to be parallel.

T ABLE 1. Linear absorption coeffi cients for X-rays of air under atmospheric conditions obtained with beryllium­window X -ray tube

JOO ------- 5. 1 2.0 .8 .3 .3 J20 -----_. 4.8 1. 6 . 7 .3 .3 ]40 ------- 'I. 6 1.3 . 6 .3 .3 J60 ------- 4.3 1.1 .5 .3 .3 ISO .... ____ 4. I 1.0 .5 .3 .3

200 ------- 3.9 0.8 . 4 .3 .3

1 l\1i nimum filtration is 1.5 mm of berylliu m plus 119 ern of ai r.

Data in th ese tables are derived by calculating J.k for ail' between 11 9 em and 269 cm from the t ube target. Errors due to change in quality of the X-rays will be involved in applying these data to determine intensities at great distances 5 from the target, and, also , at distances 9 few centimeters from the target. However, within the range of approximately 1 to 3 m from the target, theOle coefficients may be applied with li ttle error.

For r eady reference, som e air absorption cor­r ection factors have been prepared and are shown

, '1'0 accurate ly specify the absorptiou of a beterogeneous beam, it is neces­sary to employ an ex prcss ion that is the sum of exponential terms involvin g di ffere nt absorption coetllciellts.

5 "Grea t distances" , as herc used, refers to the absorption by a suffici nt quantity of air to reduce the intenSity of the original beam by sevrra l balf, value layers. However, reductions of the order of on ly onc ha lf-value layer, such as observed in this experiment (fi g. 2), show very little qua lity de­pendence.

395

Page 4: Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately 1 percent. The difference would not be greater than this in the case of 5O-kv X-rays,

TABLE 2. Linear absorption coefficients for X rays of air tinder atmospheric conditions obtained with beryllium-window X-ray tube

Absorption coefficient (It )

Voltage Added filtration ~f~IT~n~ 1---------.-------.--------,--------.----·----,--------,---------

0.05 mm Al 0.2 mm Al 0.5 mm Al 1.0 mm Al 2.0 mm Al 4.0 mm Al 6.0 mm AI

kv cm- 1 cm- i cm- i cm-1 cm- i cm- i cm- i cm- i

7.5___ ___ ___ ______ ____ _ ___ ___ ___ _ __ __ _______ 2O.5x1O-' 10 _ _ _ _____ ___ _ __ __ _ ___ _____ _ ____ _ __ _ __ __ __ _ 12.9 9.7xlO-' 8.3xlO-3 15 ____________________________________ , ___ _ 7.8

20 ______________________ ._ ___ __ _ ___ __ ___ _ __ _ 6. 9 5.3 3. 9 2.5x1O-' - -- --- --- ---- ----_.- --.--- ------------- ----- ---- ----30 _ _ _ __ _____ _ ___ ___ _ _ _ _ _ _ _ __ ___ ___ _ _ ___ ____ 6. 1 4.9 3.3 1.8 0. 9xlO-' 0.8x1O-' --- ---------- -- -- ------- --

40 _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 5. 9 4.8 3.0 1.5 . 8 . 6 --- ------ - --- -.- - --- --- ---SO _ _ _ __ _______ __ __ _ ___ ___ _ __ __ ____ __ __ _ ____ 5.8 4.6 2. 8 1.4 . 7 .6 0.5xlO-' 0.4x10-' 60 ____________________________ _______ _ .____ 5. 6 4.5 2. 7 1.3 . 7 . 6 .5 . 4 70 _ _ _ _____ _ _ _ _ _ __ __ _ _ _ __ ____ _ __ ___ __ _ ______ 5.5 4.4 2.6 1.2 . 7 .5 . 5 .4 80 _ _ _ __ _________ ____ _____ _ __ __ _____ ____ ____ 5.3 4. 3 2.5 1.1 . 6 . 5 .5 . 4

90 _____ ___ _________________________________ 5.2 4.2 2.4 1.0 . 6 . 5 . 4 . 4 100________________________________________ 5. J 4.1 2.3 0. 9 .6 . 5 . 4 . 4

1 Minimum filtration is 1.5mm of beryllium plus 119 cm of air.

TABLE 3. Linear absorption coefficients for X rays of air under atmospheric conditions obtained with cerium glass-window X -ray tube

Voltage Minimum filtration 1

Absorption coefficient (It )

Added filtration

0.5 mm AI 1.0 mm Al 2.0 mm Al 4.0 mm Al 0.0 mm Al - ---·-----------·--------1--------1--------1-------1-------1--------------

kv cm-1 cm- 1 cm- 1

20 ______________________________________ _ 30 ______________________________________ _

cm- 1

3.2X lO-' 2. 6 I. 3XIO-3 0.9 X l0-' 0. 8X IO-' - -- -- -. -. -. --- -- --- - --.- .---

40_______________________________________ 2.2 1.1 00_______________________________________ 2. J 1.0 60 ___ _ __ _ _ ______ _ ______ _ __ ____ ___ _____ ___ 2. 0 . 9

70___ _ ____ _ __ _ __ _ _ __ ___ _ _ _ __ __ _ _ _____ __ __ 1. 9 . 9 80___ _ _ _______ _____ ___ __ __ _______ ____ ____ 1. 8 . 8 90___ __ __ _ ___ ___ _ _ _ _ __ _ _ __ _______ _ __ _ __ __ 1. 8 . 8 100______________________________________ 1. 7 . 7 JlO ___ ____ ___ _ __ ____ __ _ ___ ____ __ _____ ____ 1. 6 .7

120___ ____ ____ ____ _ ____ __ _____ __ _ _ __ _ __ __ 1. 5 . 7

1 Minimum filtration is 1.29 mID of cerium glass pIns 119 em of air.

in figures 4,5,6, and 7. For a parallel beam, the air absorption correction factor, C, is taken to be,

In the case of a diverging beam, it may be defined as,

where I~ and I~ are the intensities measured by

396

.8 . 6 - ------ -- -- --- --. ----- -- -- --

. 7 . 6 0.5X IO-3 O. 4X IO-'

.7 . 6 . 5 . 4

. 7 . 5 . 5 . 4

. 6 .5 . 5 .4

. 6 . 5 . 4 . 4

. 6 . 5 . 4 . 4

. 6 . 5 . 4 . 4

. 6 .5 . 4 . 4

r-meter 6 at distances Xo and X + Xo cm, respec­tively, from the tube target.

The correction factor, C, may be used in calcu­lating intensities at various points, where a thimble chamber 6 measurement is made at one point only. Its application in field use is to determine the intensity at one position by ail' ionization measure­ment, calculate by inverse-square law the intensity to be expected at some distance farther from the

6 John A. Victorl'en. Medical physics, p . 1370- 1382 (The Year Rook Pub­lishers, Inc. , C hicago, Ill. , 19H) .

Journal of Research I

Page 5: Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately 1 percent. The difference would not be greater than this in the case of 5O-kv X-rays,

source, if it Were operating alone as a reducing factor; then, multiply this intensity by the air absorption concction factor to determine the true intensity. If the intrnsity is to be determin ed

at some point closer to the source than the ioniza­tion measurement is made, the reciprocal of the air absorption correction factor will be used as the multiplying factor.

o

t--.. .IoD=±=±=±±+±t=±=t=~;:;=t~~~I~~

20 40 6 0 80 100 120 140 160 180 200 AIR AB SO RPTI ON DIS TAN CE ( CENT IMETER S )

FI(i;I)R E 4, A ir-absQrption j'eduction in intensity oj minimum-filter X-ray beam from beryllium-window tube fi ltration 1.5 mm Be.

with inherent

1.00

0. 96

092

0 .8e

<r

~ ~ 0.84 z o ~ u ~ 080

g z o 0 76

i ~ 0.72

<r ;;

0 . 68

0 . 4

0.60 o

'" !:b" ~ ~

y

20

~

~ ~ ~ ~~ §§ S:-

"'" ~ ~ 8: ~ '--~ ~ t': :s I::' s:: l'---

"- "'" ~ ':::' t:--:: t' t--:t---- 10

'\ l"-I" 0_.

t' t'~~t---'\ '" I"- ~~~ t---- r---

,,\ ['..... '" .~",i'--" l'---I" " K '" r-----f..,,>o .. i'---..:

" r'< I'.... "-I'.... I "~o "-

-«- '" '" '" "

40 60 80 100 120 140 160 leo 200 AIR ABSORPT ION DISTAN CE (C ENTIMETERS I

FIG URE 5, Air-absorption reduction in intensity of X-ray beam filtered by 0,5 mm of Al.

(Inherent filtration of X-my tube used is 1.5 !TI1ll Be,)

Absorption of X-Rays in Air

100 ~

09 9 ~~

~ 0.98

;!.

6 0 .97

~ 8 0.96

z o ~ :£ 095

~ ~ 0 .94

0.93

0 .92

FIGU RI!] 6,

~ ~ ,,,", ~ ~ ~~ ,.::::; ~ ~

'" 1":(:':: ~ ~ lJ ~ IB O TO 200 K V

~" " ~ ~ / 160 KV ~ Vl4 0KV

" ~ ~ ~<V I-- t--

"" "" ~~~~ ~ ~ ~ f"'R--.. ~ "" ~ I'.... "- "" .. ~

['..... 1'--"'I'....

r---..

t--- f-I- ~~ K 20 40 60 80 100 120 140 160 18 0 200

AIR ABSORPTION DI STANCE ( CENT IMETER S)

A ir-absorption j'eduction in intensity of Y-my beam filt ered by 0.2 mm oj Cu,

(InhNen t filtl'a tion of X -ray l ube usee! is 1.5 mm Be.)

397

Page 6: Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately 1 percent. The difference would not be greater than this in the case of 5O-kv X-rays,

---------------------------------------------- --

100

~ :---. ~ " ~

" 099

~ '-.. 1 '\ ~~ '-..

0.9 8

" ~0-. 1', I'\. ~ ~ r"

097

0 .96 ~ 0- i'--"- '~~ '-..

'\~ ,'" I '-.

:'\.f"-0 .95

0 .94 I'\. '\. I'\.. '" '\. '" "'''''~ f"- ~~"" 9 1~/'l- "'" " O~" ["'-, "" "l

"- '" "-..1 '"

0.93

0.92

'\. "" I'" '\. I'\.. '" 0.91

'\. '" " 0 .90

f'\ "-.. 0 .89

0 .8e 1

o 20 <40 60 60 100 120 140 160 180 200

AIR A~~"ls:lC>Tlt:lN nl!:' TANCI: (CE NTtt.lE T ERSl

FIGURE 7. Air-absorption reduction in intensity of X -ray beam filtered by 2.0 mm of Al (solid lines).

Dashed lin e indicates reduction produced by 0.1 mm of Pb filter for X·rays of all qualities from 50 to 200 kv. (Inherent fi l tration of X·ray tube used is 1.5 mm Be.)

In cases where the graphs do no t cover the conditions desired, the air absorption correction factor may be calculated, using the absorption coefficients given in tables 1, 2, and 3, and the equation, 0 e-~x. As an example of this calcula­tion, table 3 indicates that J..L for 40 lev with 1.0 mm Al of added filtration is 0.8xl0- 3 em-I. 0 is then defined for arbitrary values of X beyond 119 em from the tube target. It is convenient to arrange the work in tabular form, as follows:

X

cm 25 ___________________ _ 50 ___ ________ _______ _ 100 _________________ _ 150 __________________ _ 200 _____ __ __________ _

398

}J. X

0. 02 . 04 . 08 . 12 . 16

0. 98 .96 . 92 . 89 . 85

IV. Free-Air Chamber Versus Thimble­Chamber Air-Absorption Data

That the air absorption factors determined with the free-air chamber are applicable also to thimble chamber use is indicated by reference to table 4. In this case, the limiting diaphragm is removed and a 25-1' thimble chamber is placed at the midpoint of the collector plates of the free­air chamber, the latter chamber being here used only as a vehicle for transport of the thimble chamber. Data obtained are at 90 kv with no filtration other than l.29 mm of cerium glass, the tube window thickness.

TABLE 4. Comparison of air absorption data for 90-kv '(-rays obtained by thimble i onization chamber with those obtained by fl'ee-air ionization chamber under minimum filter condition (1.29 mm of cerium glass)

Intensity, as I Thimble chamber In tensity as Distance, X, measured by intensity calcu· m easured by of air lated to indicate abso rption 1

thImble air absor ption free-air chamber' only 3

chamber

-em percent percent pecrenl

0 100 JOO 100 25 64.9 95.1 94.9 50 44.9 90.7 91.0 75 32.5 86. 4 87.0

100 24.5 83. 1 83.1

125 18. 9 79.4 SO. 3 , 150 15.0 76.7 77.7

I

1 X is the ionizat ion chamber travel distance, t he zero position of w hich is 119 em from t he X·ray t ube target.

Z This includes inverse-square reduction in intensity superimposed upon air absorption reduction .

3 '1'he observed thimble chamber intensity times (119+Xj119)' y ields an intensity indicating the a ir absorption red uction only.

These data show r eductions in intensity due both to the inverse-square l(l,w and to air absorp­tion. The minimum target-to-thimble-chamber distance used is 119 cm. Multiplication of the thimble chamber readings by (119 + X /119),z where X is the thimble-chamber travel distance, yields data indicating air absorption only. The data obtained in this manner with the thimble chamber agree within a percent with those using the free-air chamber as described in section III .

Journal of Research

Page 7: Absorption of x-rays in air · the difference between the 2-mm and tbe 6-mm curves is approximately 1 percent. The difference would not be greater than this in the case of 5O-kv X-rays,

In the case of soft X-ray , thimble-chamber measurement of the ratio of intensities between two points may be made without appreciable error, because there is no significant selective ab­sorption of the soft components of the X-rays in traversing the amounts of air considered. The calibration of thimble chambers for soft X -rays, however, will be in error, the readings being less than those for hard X -rays, because there is

Absorption of X-Rays in Air

appreciable absorption of the soft rays in the thimble chamber wall. The subject of such cali­brations for soft X-rays will be di cu sed in another paper now in preparation.

The authors thank John H . Baltrukoni for obtaining much of the data reported above.

W ASHING'rON, September 15, 1947.