A Multi Channel, Wireless Telemetric Micro System
-
Upload
lalo-viana -
Category
Documents
-
view
221 -
download
0
Transcript of A Multi Channel, Wireless Telemetric Micro System
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
1/16
IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006 187
A Multichannel, Wireless Telemetric Microsystemfor Small Animal Studies
Chung-Chiun Liu, Edward OConnor, and Kingman P. Strohl
AbstractConventional means of collecting biophysiological pa-rameters in small animals often involve cumbersome direct wiringand/or restraint of the animal. At present, there is no system forvery small animals that can provide multichannel monitoring ofbiopotentials without restraining the animal or small enough insize or light enough in weight for studies with smaller animals. Forlarger animals, such as monkeys or larger rodents, systems havebeen proposed where the transmitter of the system has dimensionssuch as 2 5 2 5 1 3 cm3 and the weight is 9 g; this is fartoo high for smaller animals. Also, the battery life of that systemis relatively short ( 10 h). In this study, a multichannel wirelesstelemetric microsystem for biopotential monitoring in small ani-
mals, such as mice or rats, has been designed, fabricated, and eval-uated. This microsystem has four input channels with one calibra-tion channel. There are 8 channels on the chip, of which five, thefour electroencephalogram (EEG)/electromyogram (EMG) chan-nels, and the calibration channel, are now in use. The system canalso be expanded to more than eight input channels, if desired. Inthat case, a larger ASIC chip and larger circuit substrate might berequired, depending on the type of biopotentials being measured.The amount of ASIC and circuit substrate space consumption islarger for biopotentials such as EEG or EMG than for others suchas temperature or pressure. However, the same clocking-demod-ulation system could be retained up to 128 channels. The mul-tichannel telemetric chip for the present embodiment is approxi-mately 2 2 mm, and the overall size of the microsystem is ap-proximately 1 0 1 0 5 mm, including the enclosure package
and battery, with a total weight of 1 g. The power consumed bythis four-channel version, where two channels are EEG and twoare EMG, is 0.41 mW, and the fabrication process is AMI_ABN.There is a magnetic on/off provision. The microsystem has beenused to monitor EEG, Theta activity, and nuchal EMG in mice withexcellent results. This wireless telemetric microsystem can be ef-fectively used to record multiple biopotentials from freely movingsmall animals. This platform microsystem can be extended to in-clude other physiological parameters, such as temperature, pres-sure, and biological parameters.
Index TermsBiopotentials, electroencephalogram (EEG),electromyogram (EMG), multichannel, small animals, wirelesstelemetry.
I. INTRODUCTION
PHYSIOLOGICAL investigation of various biological is-
sues usually requires the collection of biophysical and
biochemical parameters in animal studies. It is often that small
Manuscript received June 29, 2004; revised August 3, 2005. The associateeditor coordinating the review of this paper and approving it for publicationwas Prof. Ralph Etienne-Cummings.
C.-C. Liuand E. OConnor arewith theDepartment of ChemicalEngineeringand the Electronics Design Center, CaseWestern Reserve University, Cleveland,Ohio 44106 USA (e-mail: [email protected]; [email protected]).
K. P. Strohl is with the School of Medicine, Case Western Reserve Univer-sity and the Louis Stokes VA Hospital, Cleveland, OH 44106 USA (e-mail: [email protected]).
Digital Object Identifier 10.1109/JSEN.2005.860358
animals, such as a mouse or rat, are used. Regardless of the
sensing elements employed for the biophysical and biochem-
ical parameters, conventional techniques to transmit the sensor
outputs to the external environment involve wire connections
and restraint tethering which limit the animal movement and
the recording conditions. It will be desirable if the transmis-
sion interface can be accomplished using a wireless telemetric
approach. This allows the monitoring of various biological
functions of an unrestrained small animal.
One multichannel telemetry system used a sequential conver-
sion of the input signals to a current to control a current-con-trolled oscillator based upon a monolithic chip [2], [3]. This
chip (3 3 mm) was complementary bipolar (BJT), and con-
tained a single set of amplifiers, reference circuits and a current
controlled oscillator (CCO). It lacked clocking provisions and
required the addition of commercial CMOS chips. Other limi-
tations included an inability for providing proper preamplifica-
tion, filtering, or input impedance for weak biological signals,
such as EEG or EMG. The overall package size (not including
the battery) was large, cm, precluding its use in
small animals.
Fryer et al. [4] described a multichannel telemetry system
using a time-sharing sequential multiplexing. However, the size
of the system, cm not including the battery, was largeand designed for signals such as strain gauges or electrocar-
diogram (EKG) rather than far more difficult to detect signals
such as EEG and EMG. Input impedance was 150 K for EKG.
Input-referred noise was 20 uV p-p at 50-Hz bandwidth. The
current drain was 2.5 mA using two 1.35-V mercury cells.
The frequency deviation of the FM transmitter was required to
be trimmed to match the discriminator of a system FM receiver.
No in vivo EKG recordings were shown.
Ruedin et al. [5] described a miniaturized EEG transmitter
with two asymmetric channels that was anchored to the skull of
a small animal with screws. The size was large,
cm. The input impedance was stated to be 2 10e6 ; forEEG and EMG recordings, this was internally shunted to 6.8
10e4 (68 K) . Other versions of EEG transmitters [6], [7] have
four single-ended rather than differential channels and had the
same construction and size disadvantages as that by Ruedin et
al. [5]. A system reported by Borbely et al. [8] had only a single
channel.
An intraperitoneal telemetry device that transmits EEG is
commercially available. However, this unit has only one EEG
channel and is large, cm in size. The single EEG
channel unit employs a pair of built-in silicone-insulated double
helix stainless steel EEG leads [9], and users cannot connect
their own electrodes.
1530-437X/$20.00 2005 IEEE
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
2/16
188 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
TABLE ICOMPARISON OF WIRELESS MICROSYSTEM PERFORMANCE
Commercial implantable transmitters to monitor preterm
labor by measuring pressure changes are available [10]. A
biotelemetry system for EMG and tendon force measurement
in rats has been proposed [11], as are implantable biotelemetry
transmitters for mice, sensing temperature and pressure, and
incorporating ion-selective microelectrodes and biosensors
[12]. However, the existing units in this type of devices are
fairly large (20 8 mm diameter).
A 128-channel EEG monitoring system was described using
time multiplexing with a clock rate of, e.g., 25.6 KHz [13]. This
system utilized large commercial integrated circuit packages
and discrete parts, and operated via a cable rather than telemetry.
However, in such applications, a larger chip using the system
described herein could accommodate more channels and enable
such EEG monitoring to be accomplished by telemetry rather
than cable.
Irazoqui-Pastor et al. [14] described a miniaturized neural or
EEG device, operating at an r.f. carrier frequency of 3.2 GHzusing analog FM. It was inductively powered and a 100-W Ham
radio transmitter driving a large external coil with a passive
impedance matching circuit was required. It was approximately
cm and required a 2.5-cm monopole transmitting
antenna. The power consumption was approximately 5.8 mW.
It appeared to be single channel. The input-referred noise of the
OTA portion of the design was 8 uV, but the crest factor ([15,
p. 299]) was not stated.
Mohseni et et al. [16] described a wireless neural mi-
crosystem with three signal channels and a marker channel
operating with FM in the 88108 FM band. The marker
channel was used for sequencing the channels in the systemsdemultiplexer. The size of the microsystem was 1.8 1.3 cm
without the two 1.5-V batteries. The power consumption was
2.44 mW. The input-referred noise was 7.5-uV RMS but
the crest factor was not stated. The device had not been tested
in vivo. The dc offset of the input signals was over a range of
0.25 V. The ASIC used the AMI ABN process but required
laser trimming, both to set clocking rate and to control the
amount of frequency modulation.
Takeuchi et al. [17] described a hybrid neural device which
used commercial parts. It was a single channel and operated
with an 8090 MHz FM r.f. carrier. It was tested with a 500-uV
test signal. The dimensions were relatively large, 1.5 0.8 cm,
not including any powering system. The power consumptionof this device was 10 mW. It operated for only 30 min with a
silver-oxide battery. It was mounted with an adhesive material
directly on the back of an insect.
DeMichele et al. [18] described a 16-channel inductively-
powered system for neural or EEG signals using an ASIC
of .46 .46 cm, fabricated in the AMI ABN process. The
overall package size with an enclosure or encapsulation was not
specified. It drew 3.8 mA at 4.75 V with a power consumption
of 18 mW. It required gain adjustment. The range of the
transmitter was about 3 ft at 385 MHz, with a 1 antenna
connected to the transmitter. The r.f. bandwidth was specified
as 15 MHz and the device was tested with a signal consisting of
a 100-uV 6-Hz square wave, but it was not tested in vivo. The
stated amplifier input-referred noise was high, e.g., 121-uV
RMS. Switching noise injected by the scanning process was
found to be a significant problem. The amplifiers demonstrated
an operating point shift in the presence of r.f. interference
including the VCO.
Harrison et al. [19] described a neural amplifier built in a1.5- m CMOS process (AMI_ABN), with six amplifiers on a
2.2 2.2 mm chip. The amplifier was designed with MOSFETs
and on-chip capacitors. The supply voltage was 5-V split-supply
and it had a frequency range for use with neural electrodes
from millihertz to 7 KHz. The measured input-referred noise
was 14-uV p-p. The RMS was 2.2 uV; this would lead to a
crest factor of 3.8 [15, p. 299]. The neural amplifier was re-
designed for low-frequency biosignal applications such as EEG
or brain-surface electrodes, to exhibit a bandwidth of below 1 to
30 Hz. It was stated that the input-referred noise voltage for the
EEG version was 1.6-uV RMS. There was a neural waveform
recording but there was no in vivo EEG recording presented inthis study.
Table I summarizes the results of our telemetry system and
those offive others referenced in this paper. The input-referred
noise, power dissipation, detectable signal, and transmission
range, as well as the parameters of telemetry link frequency,
number of data channels, power supply, system clock frequency,
communication scheme, number of external components, total
weight, and package dimensions are presented. Our system
shows the lowest input-referred noise, and our system has the
lowest power dissipation of a complete telemetry system. By
comparison, our system has a small footprint. The telemetry link
frequency of our system is higher than those of [ 16] and [17],
but lower than those of [14] and [18]. Our system has more datachannels with the exception of [18]; in that system, a much larger
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
3/16
LIU et al.: MULTICHANNEL, WIRELESS TELEMETRIC MICROSYSTEM 189
ASIC chip is used. Table I also provides information on the total
weight (where provided by the investigators) and the power
requirements of the systems cited, and our system compared
favorably in these aspects. The communication scheme of our
system is amplitude-modulation-based (onoff keying) while
those of the other complete telemetry systems are FM-based;
this use of an FM-based system is apparently associated withhigher power dissipation. The power supply voltage of our
system also compares well with other systems.
The advance in microfabrication techniques as well as micro-
electronics in recent years provides an opportunity to develop
microsize, wireless telemetric interface suitable for small an-
imal monitoring. Sensors for physiological parameters include
those for common parameters such as EEG, EMG, EKG, pres-
sure, temperature, and also for various chemical and biochem-
ical parameters. Thus, a wireless telemetric interface needs to
be capable of transmitting these types of sensor outputs.
EEG refers to recording graphically the electric activity of
the brain, particularly the cerebral cortex, by means of exter-
nally placed electrodes. The frequency range is 0.5100 Hz.
Theta EEG refers to a recording from the temporal region,
having a frequency range of 47 Hz. Problems associated
with recording EEG are low signal levels and high 5060 Hz
power line interference. EMG refers to recording the electrical
impulses that pass through a muscle as it contracts and relaxes.
EKG refers to recording electrical impulses as they vary during
the cardiac (heart) cycle.
In this study, we had developed a multichannel (four channels
with a calibration channel) wireless telemetric microsystem for
EEG and EMG monitoring for small animals. Specifically, this
microsystem was used for the study of sleep disorders using
mice as the animal model. From the physiological viewpoint,sleep apnea is initiated and sustained by instability in the
respiratory control system [20]. Short-term potentiation of
ventilation (STP), also called ventilatory after-discharge,
is evoked by brief hypoxia, promotes ventilatory stability,
and protects against dysrhythmic breathing or posthypoxic
frequency and ventilatory decline [21]. An absence of STP
promotes the appearance of repetitive apneas, as supported by
studies on patients with obstructive sleep hypo-apnea syndrome
(OSAHS) [22] or congestive heart failure (CHF) patients
with Cheyne-Stokes respiration (CSR) [23]. Hence, ventilatory
instability and periodicity are common to CSR, OSAHS, and
the appearance of periodic breathing at altitude [24]. Centraland obstructive apneas may occur in the same patient over a
night [25]. These studies indicated that posthypoxic behavior
and periodicity are fundamental features in the pathophysiology
of sleep apnea syndromes.
The complexity of understanding the pathophysiology of
sleep disorder would require the investigation of certain organs
such as the brain involved in both central chemosensory and
coordination of chemical and nonchemical reflexes. Such inves-
tigation would need to monitor the EEG and EMG developed
from a small animal model. It would be important that the
animal is not restrained during this study. Thus, it is mean-
ingful to have a multichannel wireless microsystem capable
of monitoring EEG and EMG that can be used in studies withunrestrained small animals.
Fig. 1. Transmitter package (ca.1 2 1 2 0 : 5
cm).
II. DESIGN AND FABRICATION OF THE
INTERFACE MICROSYSTEMIn our laboratory, a low-power 2 2 mm integrated circuit
signal processor chip was developed and applied to a transmitter
package as shown in Fig. 1, that could be anchored to the head
of a small animal such as a rat or mouse. The dimensions of
the transmitter including the enclosure package and battery are
approximately cm. The vertical spacefor the circuitry
in the circuit compartment of the package is about 1.1 mm.
The input signals of the bipolar (differential) EEG or EMG
channels modulated the period of a sub-carrier oscillator and
the time-multiplexed sub-carrier oscillator output was converted
to pulses, which gated a wireless transmitter on and off. The
wireless transmitter was an r.f. oscillator with a tank coil as thetransmitting antenna. The pulsed RF output from the tank coil
or other antenna structure was then picked up by a radio re-
ceiver, which drove a demodulator to reconstruct the individual
input signals and output them to a PC with waveform acquisi-
tion software.
The digital clocking on the chip provided for up to eight mul-
tiplexed channels; four signal channels plus a fifth on-chip cal-
ibration channel. The purpose of the calibration channel was to
produce a reference output the amplitude of which corresponded
to an EEG or EMG input level of 50-uV p-p. It also served as
an error detector in that it indicated by its frequency, waveshape
and output channel that the signal was properly received by the
radio receiver and processed by the demodulator.For animal study chronically implanted brain or muscle elec-
trodes were connected to the transmitter and used to record
spontaneous or evoked brain potentials (EEG) and neck muscle
activity (EMG). The package itself could be anchored to the
skull with cranioplastic cement or fitted with pins for insertion
into a small socket mounted on the animals head.
The details of this multichannel wireless telemetric mi-
crosystem are given in the following sections.
A. Technology
A 1.5- 2-metal 2-poly CMOS process with an NPN op-
tion, which can operate at 3 V, was used to implement an in-tegrated circuit chip providing the main part of the circuitry for
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
4/16
190 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
Fig. 2. Block diagram of the ASIC.
the telemetry transmitter package. This fabrication process al-
lows p-channel and n-channel enhancement mode MOSFETs,
bipolar NPNs and other electronic structures such as resistors,
capacitors and diodes to be integrated. This process, AMI ABN
[16], [18], [19], and [26], is applicable to low-noise analog de-
signs or mixed-signal designs.
B. Process Characteristics
Resistors can be fabricated from n-wells in this process. Poly1and poly2 can be used to form on-chip capacitors. Diodes may
be laid out for the bonding pads for ESD protection. Guard
rings may be laid out for signal isolation between circuit blocks
and for latch up prevention. An NPN option can be used for
BiCMOS digital speed-up purposes and was used for analog
purposes in this application.
C. Circuit Considerations
This CMOS process having an NPN option [26] was used so
that along with the digital clocking circuitry, the bipolar devices
could be used to provide a higher transconductance, a better
matching, lower offsets and lower flicker noise for the analogEEG and EMG amplifiers that were needed for the application.
A block diagram of the ASIC chip is shown in Fig. 2. The cir-
cuit was designed with 4 input differential preamplifiers, four
selectable second-stage amplifiers that could activate sequen-
tially from pulse inputs from the clocking circuitry, and a cur-
rent-controlled subcarrier oscillator. The gain of each selectable
second-stage amplifier was a function of the pulse current of the
select pulse for that channel. The gain of each input differential
preamplifier stage was fixed. Between each input differential
preamplifier stage and the associated selectable second-stage
amplifier was a pair of capacitors (indicated between dashed
lines), of which one coupled the signal and the other provided
roll-off to the band of signal frequencies. As previously indi-cated there was a calibration circuit for producing a reference
signal. There was also CMOS clocking circuitry, a block of
monostable multivibrators with outputs combined by gating to
provide output pulses and synchronization pulses, and a power
toggle-on/toggle-off switch circuit.
The power switch was toggled by an external magnetic sensor
such as a Hall-effect sensor or a reed switch turning the trans-
mitter on and off with a magnet. The Hall-effect sensor was
an SMT having a footprint of mm and was incorpo-
rated onto the circuit substrate of the hybrid package along withthe ASIC and the bare-die BJT r.f. oscillator/transmitter chip.
A reed switch selected from currently-available types would
have required a slightly larger overall package. The Hall-effect
sensor, a pole-independent device with a latched digital output,
worked by producing an output which went high to low as a
small magnet was brought near and went low to high again when
it was withdrawn, thus toggling the chip supply voltage via the
on-ASIC power switch circuit mentioned above. The onoff cir-
cuitry of the ASIC does not have significant static power con-
sumption; the Hall-effect sensor, an Allegro A3212 ELHLT, has
a static power consumption of 15 uW. The manufacturers data
sheet is numbered as Allegro Microsystems 27 622.61G.
The set of four selectable amplifier stages converted the input
EEG or EMG signals into linearly proportional output currents,
which, along with a reference current from the calibration cir-
cuit, were fed into the current controlled subcarrier oscillator.
These currents were fed in sequence so that the channels were
time-division multiplexed. A diagram describing the encoding
scheme is shown in Fig. 3. Each channel was turned on for two
complete cycles of the subcarrier oscillator; a half-cycle on each
end of this period served to provide setup time, and one sub-
carrier oscillator cycle was used for the data measurement of
an input channel or for the calibration signal. Each channels
value was encoded in the duration between RF pulses, i.e., the
subcarrier oscillators period was modulated by the channelsvalue. An increase in input current to the subcarrier oscillator
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
5/16
LIU et al.: MULTICHANNEL, WIRELESS TELEMETRIC MICROSYSTEM 191
Fig. 3. Encoding scheme diagram.
Fig. 4. Timing and reference/calibration signals diagram.
resulted in a shortening of the period. If the subcarrier oscil-
lator is operated at 1015 kHz it will produce a minimum sam-
pling frequency per channel of 625 Hz and a Nyquist frequency
of 312 Hz, which exceeds the 100-Hz bandwidth which is re-
quired for the EEG and EMG signals in this application. The
CMOS timing circuit was a chain of toggle-connected master-
slave D-type flip-flops, some of which drove a 1-of-8 logic de-
coder block. Three more input channels could have been used
with the 1-of-8 decoder arrangement, but the physical space was
not available on the 2 2 mm chip to accommodate the input
amplifiers for the three extra channels. Additional toggle-con-
nected D flip-flops divided down the clock signal from the sub-
carrier oscillator to about 6 Hz, which was used to drive thecalibration circuit and inject a square wave as a reference signal
onto one of the multiplexed channels. A timing and reference
signals diagram is shown in Fig. 4.
Simulated waveforms at circuit nodes are shown in Figs. 57.
Fig. 5 simulates the output of an input channel amplifier-pream-
plifier pair, having an input signal of 70 Hz at 5 000-uV p-p am-
plitude. The second amplifier is gated on and off by the select
pulse input for that channel. In this simulation, the select pulse
is actually a SPICE pulse rather than an SCO clock pulse; the
period of the SCO clock would actually be varying. The output
current is 70 uA p-p. Fig. 6 simulates the select pulse (with pe-
riod held constant for simplification purposes) for the channel.
The amplitude is 3 V, the on time is 162 uS and the frame time is
1296 uS. Fig. 7 simulates the frequency response of the channelfor a differential signal input level of 50 uV p-p.
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
6/16
192 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
Fig. 5. Simulation of output of signal channel stage.
Fig. 6. Simulation of channel amplifier select pulse.
Fig. 7. Simulation of channel amplifier frequency response.
Separate parts of the block diagram are shown in Figs. 811.
They are the input preamplifier, Fig. 8, the subcarrier oscillator
(SCO), Fig. 9, the divide-by-128 section (DFFs 511), Fig. 10,and the monostable multivibrators block, Fig. 11. In the SCO,
the magnitudes of the current sources are controlled by an ex-
ternal current, e.g., the current from a signal channel amplifier
or from the calibration section. A latching circuit is formed fromtwo 2-input CMOS NAND gates. The positive feedback around
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
7/16
LIU et al.: MULTICHANNEL, WIRELESS TELEMETRIC MICROSYSTEM 193
Fig. 8. Signal channel input preamplifier stage diagram.
Fig. 9. SCO block diagram.
Fig. 10. DFFs 511 (divide-by-128) block diagram.
the loop is used to ensure that only one of the MOS transis-
tors is on at a time. The switching points of the comparators
combined with the current source determine the oscillator fre-
quency of the SCO. Two 22-pF timing capacitors were required
for the SCO and they were implemented by very small 0402 or
0201 external chip capacitors, which were mm or
smaller. The divide-by-128 chain of toggle-connected CMOS
DFFs 511 provides a low frequency for the calibration signal.
In the Monostable Multivibrators block, each one-shot consists
of a 2-input CMOS NOR gate, a CMOS inverter, an on-chip ca-
pacitor, and an on-chip resistor. The trigger input of a one-shot
can be longer than the output pulse width.
The power for the chip and transmitter unit was suppliedfrom a CR1025 3-V watch cell which was located in the bat-
tery compartment of the enclosure package, above the circuit
compartment. The battery was located inside of the wireless
transmitter coil, which was wound around the outside of the
enclosure package of the transmitter unit.
The battery was an Li/MnO2 watch cell with a nominal
voltage of 3 V and an average capacity of 30 mAh to 2.0 V.
The volume of the battery is 0.2 cm . It is speci fied by the
manufacturer that if the load is such that the current drain is
64 uA and the operation is 24/7 the time to cutoff voltage (2.0
V) is 467 h. The wireless transmitter drew 40 uA and the Hall
sensor drew 12 uA; the ASIC drew 93 uA. The current drain
of the telemetry hybrid transmitter including ASIC, wireless
transmitter and Hall-effect sensor was 145 uA, and, thus, thissystem could operate for 168 h (7 days, 24/7) or longer. It was
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
8/16
194 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
Fig. 11. Monostable multivibrators block.
found by measurement that the CR1025 dropped below 3 to
2.9 V within a few hours; it then gradually reached 2.83 V
and remained at this level during most of its lifetime. When
it reached the end of its lifetime it then dropped within a few
hours, to 2.0 V. The effective supply voltage of the ASIC andhybrid transmitter was, thus, approximately 2.83 V.
The power consumption of the entire hybrid transmitter was
0.41 mW. This was considerably less than that of other multi-
plexed or wireless designs referred to above.
This device, like most battery-operated devices, was subject
to the effects of battery drop and parameters were expected to
change somewhat during operational life of the battery as a re-
sult. Such changes, as,e.g., changes in channel gains, were small
due to the use of on-chip voltage regulators and the use of cur-
rent mirrors tending to act as current regulators. However, it was
found that the change in signal amplitude was tracked by the
change in calibration waveform amplitude over the range of 3.2to 2.7 V. The amplitude change for both signal output wave-
form and calibration channel output square-wave amplitude was
2.0% per 0.1-V change in battery voltage over the range of 3.0
to 2.7 V. Thus, the amplitude change throughout the operational
life of the battery was small but would be correctible by refer-
encing the demodulators signal-channel gain to the calibration
square-wave amplitude.
The conductive substrate in epi wafers can have currents in-
duced into it by the magnetic field of a (spiral) on-chip inductor,
and this inductor-induced noise through the substrate can affect
other circuits on the same chip [27]. Although our device did not
utilize an on-chip inductor, the integrated circuit chip used was
within the field of an inductor. No degradation of performancewas observed in the miniaturized unit in comparison to larger
prototypes with the coil remote from the chip.
D. Circuit Implementation
The ASIC constituted the processor section of the complete
telemetry link. The only requirements to complete the imple-
mentation of the telemetry function of the transmitter were a
single external bare-die BJT chip, a resistor and a capacitor on
the hybrid substrate, plus a tank circuit. The resistor could have
been included on the ASIC but it was put instead on the cir-
cuit substrate so that it would be physically closer to the wire-
less oscillator section and, thus, tend to better decouple RF fromthe ASIC. The transistor and capacitor were not laid out on the
Fig. 12. Microphoto of ASIC chip (ca. 2.22
2.2 mm).
ASIC because of considerations of space and flexibility in re-
gard to the wireless link, as well as to avoid any problems which
might arise from having an RF generating circuit directly on
the ASIC in close proximity to analog circuitry which was pro-
cessing low-level signals such as EEG and EMG. A micropho-
tograph of the ASIC processor chip of the telemetry system is
shown in Fig. 12.
The block diagram of the ASIC chip was shown previously in
Fig. 2. The input stages utilized dc blocking capacitors of very
small physical size (0402 and 0603) on the hybrid substrate insuch a way that capacitive dc blocking, a standard precaution
with EEG amplifiers, was implemented.
The input impedance of the input preamplifier stages was re-
lated to circuit parameters and represented a tradeoff between
stage current drain, electronic flicker noise and chip size. The
measured input impedance was 670 K and was found to be
adequate for the EEG and EMG signals of the application; how-
ever, in future versions, this impedance may be increased to
3.8 M as indicated below. The in vivo electrode-tissue inter-
face impedances of the animal electrodes used in the in vivo
system testing were relatively low. Information about the an-
imal electrodes used is given later in this paper.
In a test, the result of loading (reducing the input impedance)of standard Grass EEG amplifiers was examined for in vivo
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
9/16
LIU et al.: MULTICHANNEL, WIRELESS TELEMETRIC MICROSYSTEM 195
mouse EMG and EEG at 300 K resistive shunt impedance.
In vivo EMG signals were not influenced by the 300-K shunt
across the input terminals of the Grass amplifier. The in vivo
EEG signals were not significantly attenuated (less than 3 dB
below 5 Hz) by the 300-K shunt. The higher frequency
components were not noticeably attenuated. The equivalent
electrode impedance appeared to increase as the frequencydecreases; however, the signals appeared entirely usable even
at 300 K.
It is of interest that earlier, transistorized commercial and re-
search EEG equipment utilized lower input impedances than the
value mentioned above; in [5], cited above, the input impedance
was stated to be approximately 2 10e6 ; however, for EEG
and EMG recordings, this was stated by the authors to be inter-
nally shunted to only 6.8 10e4 (68 K) .
Noise is considered to be a dominant factor in EEG equip-
ment, particularly in CMOS input amplifiers [14], [16], [18],
and [19] where the electronic (circuit-generated) noise, such as
flicker noise, is often too high. In EEGequipment, both the RMS
and the crest factor [15, p. 299] should be stated in equipmentspecifications but frequently are not.
In this development, BJTs were used for input stages of the
EEG/EMG telemeter because the flicker noise of MOSFETs is
known to be as much as 10 to 1000 times larger [15, p. 123]
unless space-and-power-consuming techniques are used.
In a BJT, the total equivalent input noise is
Also
where ;
; ;
;
' ;
;
;
: :
. The above equation for Eni^2 [15,pp. 142143] is valid for a 1-Hz bandwidth at the frequency f.
Increasing the transistor quiescent collector current increases
the 1/f noise [15, p. 126]. As the collector current drops, elec-
tronic 1/f noise of the stage may be expected to decrease, while
small-signal input impedance rises. The CMRR of long-tailed
pair Q1, Q2 in Fig. 8 increases as emitter resistor R7 is in-
creased. Thus, because collector current drops but CMRR in-
creases as R7 increases, R7 was made as large as space on the
chip permitted. Because R1 and R2 shunt the input impedance
of the stage, they were also both made as large as space per-
mitted. The resulting input impedance was 670 K but simula-
tion showed that the unshunted input impedance was 4 M .
R1 and R2 may, thus, be replaced in future versions by currentsources to increase the channel input impedance. R7 may also
be replaced by a current source to further increase the CMRR
in future versions. The output node of the input preamplifier
stage EXT1, drove filter capacitors C1 and C2, connected to
nodes EXT2 and EXT3 of the following stage. The purpose of
C1 and C2 was to set the high and low frequency roll-offs as
narrow as possible, to pass only the spectrum required, since
the greater the amplifier bandwidth the greater the output noiseand input-referred noise [15, p. 125]. Essentially, C1 of Fig. 8
may be regarded as coupling the signal to the associated se-
lectable second-stage amplifier (not shown in Fig. 8 but indi-
cated in Fig. 2) while C2 may be regarded as providing roll-off
to the band of signal frequencies. These capacitors, indicated
in Fig. 8 as being separated from the amplifier circuitry by a
dashed line, are also indicated (between a pair of dashed lines)
in Fig. 2, and separated from the input differential preamplifier
stage by the first of the dashed lines.
The input-referred noise of the EEG channels, from trans-
mitter channel input to system demodulator output, was found
by measurements and calculation [15, p. 275] to be 0.69 uV
RMS with a crest factor [15, p. 299] of 3.84 at 100-Hz BW,and 0.36-uV RMS with a crest factor of 3.69 at 30-Hz BW.
The input-referred noise of the EMG channels was found to be
0.63-uV RMS with a crest factor of 3.84 at 100-Hz BW.
Thus, the EEG signal processing in this system exhibited less
noise than the systems using CMOS circuitry for signal pream-
plification [14], [16], [18], and [19].
E. Wireless Transmitter
The train of output pulses from the ASIC, which processed
the input EEG and EMG signals and converted the signals to
a pulse format, was applied to the input resistor R1 of an r.f.
oscillator used as a wireless transmitter, as shown in Fig. 13.It was an Armstrong oscillator in which the collector winding
coil L1 and the tank circuit winding coil L2 of the transformer
were combined into an autotransformer single tapped tank coil
L, serving as the transmitting antenna of the telemetry system
and having a tuning capacitor C1 across it. Coupling capacitor
C2 provided feedback to the base of Q1. Capacitor C3 placed the
tap point at r.f. ground. Resistor R1 provided base current and
set the forward bias on Q1 when the output pulses of the ASIC
were present, thereby causing a burst of r.f. oscillation during
each ASIC output pulse. The circuit operated class C during
the intervals when it was gated on by the output pulses of the
ASIC. Because of the flow of r.f. current in L, an r.f. electromag-
netic field was generated by the coil; this r.f. field then served to
propagate the wireless signal. The use of crystal control was not
found necessary. Because the transistor and coupling capacitor
were not incorporated on the ASIC, any r.f generating circuit
which can be onoff keyed, using any carrier frequency or any
antenna type, of sufficiently small footprint, might be substi-
tuted in this system.
F. Demodulator
The train of transmitter pulses from the output of the radio
receiver was applied to the input of the demodulator of the
system, a block diagram of which is shown in Fig. 14. An
LED was turned on by a pulse-sensing circuit to indicate signalreception from the transmitter. These pulses were applied to
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
10/16
196 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
Fig. 13. Wireless transmitter circuit diagram.
a pulse separator circuit which separated the active interval,
or signal channel pulses, from the synchronization, or marker
pulses which were also sent by the transmitter, so that the
demodulator was able to sequence the channels. When the
marker and signal interval pulses were separated, the signal
interval pulses were applied to the clock input of a flip-flop;
a pulse at the beginning of a signal channel interval set the
flip-flop output high and the pulse at the end of the channel
interval reset the flip-flop output low. Thus, a series of pulses
were generated at the output of the flip-flop, whose high levels
corresponded in duration to the signal channel intervals. These
high levels were then used to gate a clock input of a counter
in such a way that when the flip-flop output pulses were high
the counter up-counted, and when the flip-flop pulses went
low, the up-counter stopped counting. At the end of each signal
channel interval of the transmitter, a binary (digital) output was,
thus, generated. This digital output was applied to a D-to-Aconverter, which converted the digital to an analog voltage,
corresponding to the EEG or EMG signal sample, and held this
sample until the next frame of signal channel intervals. The
up-counter was then reset and at the start of the next signal
channel interval the process started again. There were eight
D-to-As, corresponding to eight channels, of which four were
used for signals, one was used for the calibration signal and
three of which were reserved for future ASIC versions. At the
end of each frame of eight channels, the flip-flop mentioned
above was reset by the marker pulses from the pulse-separation
circuit mentioned previously. The output of each D/A was
then passed through a dc amplifier/filter to eliminate the D/Aincrement-noise and calibrate the signal level so that 1 V p-p
output from the demodulator corresponded to 50-uV p-p input
to the transmitter. The output of each amplifier/filter was then
passed through an additional low-pass filter to limit channel
bandwidth to 100 Hz, for the purpose of reducing electronic
flicker noise [15, pp. 19-20]. The amplified and filtered signal
was then passed through a dc block and a buffer to eliminate
dc offset. In addition, because the subcarrier oscillator of the
transmitter was current-to-frequency-modulated, a 1/f converter
was added between the low-pass filter and the dc blocking filter
in the four signal channels. Although eight channels were
sent from the transmitter, only four signal channels and one
calibration channel were used, so that there were five outputamplifier/filter/buffer circuits in the demodulator. Any channel
mismatch on the chip was trimmed out in the demodulator as
the system was calibrated channel-by-channel with a 50-uV
p-p signal being applied to the channel input of the hybrid
transmitter containing the particular ASIC and the channel was
trimmed for a 1.0-V p-p output signal from the corresponding
output of the demodulator unit being calibrated for that specific
transmitter. In the future, the use of a different BiCMOS typeof process in which p-n-ps as well as n-p-ns are available
may allow closer on-chip matching between channels, by
using BJTs to replace less-well-matched MOSFETs in analog
circuitry, thus allowing for less need for trimming.
G. Radio Receiver
The output of the systems oscillator/transmitter was sensed
using a small antenna such as a half-dipole or loop, which fed a
single-conversion receiver, a block diagram of which is shown
in Fig. 15. The receiver incorporated a front end, an i.f. section
and a second detector. A commercial front end having a tuning
range of 50810 MHz in four bands was used in this receiver.
An internal switch was used to select the correct band for re-
ceiving the transmitter. The i.f. bandwidth of this receiver was
adjusted to correspond to the spectral content of the transmitter
pulses while rejecting external interference; the i.f. circuitry was
broadly-tuned and the bandwidth was in the range of 2 MHz.
The receiver gain was limited by the adjustment of the front-end
and IF gains, to suppress interference of extraneous signals and
noise with the transmitter pulses. The second detector circuit in
this receiver produced a pulse output, which corresponded to
the pulse output of the transmitters ASIC chip, which onoff
keyed the transmitter units wireless transmitter circuit. The re-
ceiver second detectors pulse output was fed to the demodu-
lator unit, to be processed in order to recover the transmitterchannel interval and marker signals, and subsequently the EEG
or EMG input signal information. The systems receiver also in-
corporated an audio section, which produced an audible signal
to facilitate tuning the receiver to the transmitter signal. The RF
receiver is not a commercial one and it, therefore, has no manu-
facturer information or data sheet reference; however, the front
end is a Zenith 175-00 014 CATV tuner made by Zenith Elec-
tronics Corporation, Glenview, IL.
H. Miniature Enclosure for Telemetry Transmitter
The miniature enclosure for the telemetry transmitter was
a micromachined box that could be made of alumina ceramicor macor ceramic. A microphotograph of the box is shown in
Fig. 1. A cover slid in low-friction micromachined grooves, and
served to clamp the battery against the internal battery contacts
as well as to hold the battery. The battery shelf also acted as
the top of the circuitry compartment, which was only 1.1 mm
in height. The enclosure box was set on the circuit substrate,
which carried the ASIC chip, a SMT Hall-effect sensor, an R.F.
oscillator chip BJT, and the various chip capacitors and a chip
resistor; the .22- and .001-uF chip capacitors and the chip re-
sistor were 0402 or 0201 size and some of the remaining chip
capacitors were 0603 size. The ASIC chip and the chip BJT were
connected to the circuit substrate with wire bonds. Flip-chip at-
tachment could also have been used. A photograph of the com-plete multichannel wireless telemetry circuitry on the alumina
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
11/16
LIU et al.: MULTICHANNEL, WIRELESS TELEMETRIC MICROSYSTEM 197
Fig. 14. System demodulator block diagram.
Fig. 15. System radio receiver block diagram.
Fig. 16. Microphoto of transmitter hybrid substrate (ca. 12
1 cm).
substrate is shown in Fig. 16. The size of this circuit substrate
was cm and was smaller than the footprint of a U.S.
dime. It may be assembled manually or by automated methods.A transmitting coil of a few turns of wire of approximately #40
TABLE IITABLE OF OFF-CHIP COMPONENTS OF THE TRANSMITTER
gauge was wound around the top of the enclosure, leaving clear-
ance for the sliding cover. The assembly was held together by
a cement. Leads from the electrodes implanted in the animal,
or from a connector on which the box might be mounted, were
brought to the side or bottom terminals. The replacement of
the battery was relatively simple by sliding the top cover of the
miniature enclosure to access the battery.
A list of the off-chip components of the hybrid substrateand enclosure package is shown in Table II. These off-chip
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
12/16
198 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
Fig. 17. B6 Mouse recording (active) made with ASIC system.
Fig. 18. B6 Mouse recording (slow wave) made with ASIC system.
components are annotated as to the components shown in the
schematics included in this paper.
III. RESULTS AND DISCUSSIONS
After the ASIC chip was fabricated and the hybrid trans-
mitter was built, the in vivo performance of this system wasassessed by recording the EEG and EMG waveforms of small
mice. Two C57BL/6J and two A/J mice (Jackson Laboratory,
Bar Harbor, ME) were implanted with stainless steel electrodes
for the recording of the cortical and theta EEGs and of the EMGs
of the nuchal (neck) muscles. A midline incision was made to
expose the skull and neck muscles posterior to the skull. Two
pairs of stainless steel wires 0.21 mm in diameter and stripped
for 0.5 mm at the ends were surgically placed to contact the durafor bipolar theta and cortical EEG recordings. Two other pairs of
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
13/16
LIU et al.: MULTICHANNEL, WIRELESS TELEMETRIC MICROSYSTEM 199
Fig. 19. AJ Mouse recording (REM) made with ASIC system.
Fig. 20. B6 Mouse recording (slow arousal) made with ASIC system.
stainless steel electrodes made by knotting stainless steel wires
and stripping the knotted portion were sutured into the surface
of the neck muscle for bipolar EMG recordings.
The electrodes were connected to the telemetry transmitter
and the animal was placed in a Lucite chamber (10 cm in di-
ameter and 6 cm high) with bedding and food and water. Eachanimal was studied for three days.
A data analysis program was used to view the demodulated
EEG and EMG data via an analog conditioning filter/amplifier
(CWE, Inc.) and a 12-bit Data Acquisition System (National In-
struments PCI-MIO-16E) or 16-bit PCI-6033E DAQ. The Data
Acquisition System (DAS) was used in a LabView environment
on a P-III desktop computer. The resultant data was sampledat 512 Hz and stored on a hard disk. Records were scored for
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
14/16
200 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
TABLE IIITABLE OF SPECIFICATIONS OF THIS SYSTEM
sleep stage, i.e., assignment was made of the kind of waveform
as, e.g., Waking, Slow-Wave, or REM.
For each mouse, several segments of data, each about 5 h
long, were recorded. The data was then analyzed. Example
waveforms telemetered by the telemetry system are shown.
Each recording has visible at the left the labels EEG, Theta,
EMG1, EMG2, CAL. Each recording also has vertical bars
in the center that are labeled to show the signal level corre-
sponding to the height of the bar.
Fig. 17 shows cortical EEG, Theta activity and nuchal EMG
from a B6 mouse in active wakefulness. Fig. 18 shows the EEG,
Theta and EMG from a B6 mouse in slow wave sleep. Fig. 19
shows a set of waveforms from a B6 mouse in slow wave sleep
with arousal from sleep. Fig. 20 shows signals from an AJ
mouse in REM sleep. The calibration signal, a square wave of
56 Hz, is shown on the bottom channel of each recording.
The bandwidth used for the cortical EEG and the EMG was
100 Hz. The bandwidth used for the theta EEG was reduced
by the recording instrumentation to improve the display of the
characteristics of this waveform parameter.
By referring to the waveform recordings, which represented
the entire system as well as the ASIC and hybrid transmitter, itcould be seen that system CMRR was sufficient so that the effect
of 5060 Hz power line interference, often a problem with EEG
systems, was not observed. The system channel crosstalk effects
were visually absent, since clearly evidence of the calibration
square wave was not seen in the signal channels, or vice versa,
and evidence of the EEG waveforms was not seen in the EMG
or calibration channels, or vice versa.
The calibration signal served an important purpose in therecordings in that it indicated at all times during the recordings,
by continuously providing a waveform of known wave-shape,
frequency, and amplitude, that the system was working and that
r.f. transmission and reception were being properly achieved. It
indicated also that demodulation was properly accomplished,
and all channels including signal channels were being displayed
in their proper positions on the recording. In other systems
there has often been no indicator to assure that transmission
was being adequately accomplished or that channels were
being properly demodulated or output in proper sequence,
so that, for example, what might have seemed to be EEG
might have been not a true EEG waveform but possibly only
a system artifact. The calibration channel, thus, served thepurpose of an error detector in the recordings as well as an
amplitude reference.
In the recorded waveforms, the noise levels represented
the resultant of all noise sources, including digital, 5060 Hz
power line interference, and ambient r.f. interference as well as
electronic system noise such as flicker, shot and thermal. There
was good signal integrity as compared with, e.g., recordings
made directly with Grass amplifiers, despite these noise sources
and despite changes in light and temperature as well during
the in vivo recording experiment. The measured EEG channel
electronic noise parameters at 30-Hz BW have been previously
stated in this paper to be 0.36-uV RMS with a crest factorof 3.69, and at 100-Hz BW to be 0.69 uV with a crest factor
of 3.84. This was the overall system noise from the channel
input of the hybrid transmitter to the channel output of the
system demodulator. It is evident that this noise is considerably
less than that of the amplifier alone at 30-Hz BW in [18]
above, which is an all-CMOS design in the same process [26].
In the future, however, the use of another BiCMOS type of
chip process where both polarities of BJTs are available, may
allow still further reduction in noise to be achieved, by using
BJTs to replace noisier MOSFETs in the analog circuitry.
Specifications of the system are indicated in Table III. A com-
parison of this wireless microsystem with other wireless mi-
crosystems is shown in Table I, above. In Table III, the parameter
RF BW is measured on the bench by tuning a calibrated com-
mercial telemetry receiver between the points where the signal
reception from the telemetry transmitter drops off. The commer-
cial receiver has switchable bandpass filters with bandwidths
suitable for television testing purposes, e.g., several megahertz.
IV. CONCLUSION
We have designed and tested a multichannel wireless
telemetry system for up to four biopotential signals plus a
calibration channel for use in the monitoring of small animals
(mouse or rat). It is expandable to more channels. The systemhas a package size of approximately cm (including
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
15/16
LIU et al.: MULTICHANNEL, WIRELESS TELEMETRIC MICROSYSTEM 201
the enclosure package and the battery power source) appropriate
for monitoring small animals (mouse or rat). The footprint of
the complete package including ASIC, BJT transmitter chip,
Hall-effect SMT, and all other parts including enclosure, is
smaller than that of a small coin, such as a U.S. dime. To
achieve specifications of such small size, a monolithic I.C.
chip, i.e., ASIC, was fabricated. The I.C. chip is a low-powersignal processor chip, only 2 2 mm in size. This signal
processor chip amplifies, filters and time division multiplexes
the signals that are in turn transmitted via an RF link, i.e.,
BJT chip and associated parts as described above, contained
within the package, to an external radio receiver. The receiver
drives a demodulator to reconstruct the individual signals for
display or analysis by waveform acquisition software. Finally,
the system incorporates a Hall-effect sensor, i.e., an SMT also
in the package, as mentioned above, providing magnetic onoff
capability, initially for conservation of power, but which also
could be used for interactive procedures.
This development demonstrates the feasibility of recording
of multiple biopotentials using the miniature telemetric systemfor a freely moving small animal. The ASIC chip design used
in the telemetry system is flexible and can accommodate more
channels and both unipolar and bipolar signals, as well as other
physical and biochemical sensor outputs. This forms the tech-
nical foundation for future research in this wireless telemetric
microsystem for small animal study.
ACKNOWLEDGMENT
This study was approved by the IACUC of Louis Stokes, VA,
Research Center, Cleveland, OH, and complied with the Na-tional Institutes of Health Guide for the Care and Use of Lab-
oratory Animals.
REFERENCES
[1] J. Morizio, P. Irazoqui, V. Go, and J. Parmentier, Wireless headstagefor neural prosthetics, in Proc. 2nd Int. IEEE EMBS Conf. Neural En-gineering, Arlington, VA, Mar. 2005.
[2] C. S. Sander, J. W. Knutti, and J. D. Meindl, A monolithic signal pro-cessor for multichannel implantable telemetry, in Proc. IEEE Int. Solid-State Circuits Conf., Dig. Tech. Papers, 1979, pp. 198199.
[3] S. J. Gschwend, J. W. Knutti, H. V. Allen, and J. D. Meindl, A general-
purpose implantable multichannel telemetry system for physiologicalresearch, Bio Telemetry Patient Monitoring, vol. 6, pp. 107117, 1979.
[4] T. B. Fryer, H. Sandler, and B. Datnow, A multichannel implantabletelemetry system, Med. Res. Eng., Mar.Apr. 9, 1969.
[5] P. Ruedin, J. Bisang, P. G. Waser, and A. A. Borbely, Sleep telemetryin the rat: I. A miniaturized FM-AM transmitter for EEG and EMG,
Electroencephalogr. Clin. Neurophysiol., vol. 44, pp. 112114, 1978.[6] R. Moser, M. Daniker, and A. A. Borbely, EEG-telemetry in the rat:
selective recording from5 outof 12 chronically implantedelectrodes, inProc. Biotelemetry II. 2nd Int. Symp., Davos, Basel, Switzerland, 1974,
pp. 182184.[7] W. Kraft and F. Voegeli, 4-Kanal Miniatursender zur Uebertragung
des Elektroencephalogramms von Kleintieren, AGEN-Mitteilungen,vol. 15, pp. 1924, 1973.
[8] A. A. Borbely, I. Baumann, and N. M. Waser, Multi-channel telemetryof physiological parameters (body temperature, ECG,EEG) in the rat.
II. Applicationsin neuropharmacology, Kimmich and Vos Biotelemetry,pp. 381388, 1972.
[9] N. Herold, S. Spray, T. Horn, and S. J. Henriksen, Measurements ofbehavior in the naked mole-rat after intraperitoneal implantation of a
radio-telemetry system, J. Neurosci. Meth., vol. 81, pp. 151158, 1998.[10] Biotelemetry Using Implanted Unit to Monitor Preterm Labor,
NASA Tech Briefs. (1998, Nov. 19). [Online]. Available:
http://www.nasatech.com/NEWS/nasa 1119.html
[11] Biotelemetry Systemfor EMGand Tendon Force Measurementsin Rats,
NASA Sensors 2000!. [Online]. Available: www.datafilter.com/mc/sen-
sors2000biotelemetry.html[12] Implantable Biotelemetry Transmitters for Mice, NASA Sen-
sors 2000!. [Online]. Available: www.datafilter.com/mc/sen-sors2000biotelemetry.html
[13] J.R. Ives, N.R. Mainwaring,L. J.Gruber, G.R. Cosgrove, H.W.Blume,
and D. L. Schomer, 128-channel cable telemetry EEG recording systemfor long-term invasive monitoring, Electroencephalogr. Clin. Neuro-
physiol., vol. 79, pp. 6972, 1991.[14] P. Irazoqui-Pastor, I. Mody, and J. W. Judy, In-vivo EEG recording
using a wireless implantable neural transceiver, in Proc. 1st Int. Conf.Neural Engineering, Mar. 2003, pp. 622625.
[15] C. D. Motchenbacher and F. C. Fitchen, Low-Noise Electronic De-
sign. New York: Wiley, 1973.
[16] P. Mohseni and K. Najafi, A wireless FM multi-channel microsystemfor biomedical neural recording applications, in Proc. Mixed-Signal
Design, Southwest Symp., Feb. 2003, pp. 217222.[17] S. Takeuchi and I. Shimoyama, A radio-telemetry system with a shape
memory alloy microelectrode for neural recording of freely moving in-
sects, IEEE Trans. Biomed. Eng., vol. 5, no. 1, pp. 133137, Jan. 2004.[18] G. A. DeMichele and P. R. Troyk, Integrated multichannel wireless
biotelemetry system, in Proc. 25th Annu. Int. Conf. Engineering inMedicine and Biology Soc., vol. 4, Sep. 2003, pp. 33723375.
[19] R. R. Harrison andC. Charles, A low-powerlow-noiseCMOS amplifierfor neural recording applications, IEEE J. Solid-State Circuits, vol. 38,no. 6, pp. 958965, Jun. 2003.
[20] M. C. Khoo, Determinants of ventilatory instability and variability [InProcess Citation], Respir. Physiol., vol. 122, no. 23, pp. 167182,2000.
[21] M. Younes, The physiologic basis of central apnea and periodicbreathing, Curr. Pulmonol., vol. 10, pp. 265326, 1989.
[22] D. Georgopoulus, E. Giannouli, V. Tsara, R. Argiopulou, D. Patakas,
andN. R. Authonisen, Respiratoryshort-termpoststimulus potentiation(after-discharge) in patients with obstructive sleep apnea, Amer. Rev.Respir. Dis., vol. 146, pp. 12501255, 1995.
[23] M. Ahmod et al., Ventilatory instability in patients with congestiveheart failure and nocturnal Cheyne-Stokes breathing, Sleep, vol. 17, no.6, pp. 527534, 1994.
[24] M. Younes, M. Ostrowski, W. Thompson, N. C. Leslie, andW. Sawchuk,
Chemical control stability in patients with obstructive sleep apnea,Amer. J. Respir. Crit. Care Med., vol. 163, no. 5, pp. 829839, 2001.
[25] F. Han and K. P. Strohl, Inheritance of ventilatory behavior in rodentmodels, Respir. Physiol., vol. 121, no. 23, pp. 247256, 2000.
[26] AMI Semiconductor., Pocatello, ID. [Online]. Available:
www.amis.com
[27] L.L. Pun, An on-chipplanar spiralinductor inducedsubstrateeffectsonradio frequency integrated circuits in CMOS technology, M.S. thesis,Dept. Elect. Electron. Eng., The Hong Kong University of Science and
Technology, Hong Kong, Jan. 1998.
Chung-Chiun Liu is the Wallace R. Persons Pro-fessor of Sensor Technology and Control and aProfessor of chemical engineering at Case WesternReserve University, Cleveland, OH, where he isalso the Director of the Center for Micro and NanoProcessing. His research areas include chemical andbiological sensors and sensor arrays, applicationsof microfabrication to the development of chemicaland biological microsystems, wireless telemetric
interface technology, and microelectrochemical en-ergy sources, including microfuel cells and printablebatteries. He has authored 190 journal publications and holds 12 U.S. patents.
Authorized licensed use limited to: Universidad Nacional Autonoma de Mexico. Downloaded on June 16, 2009 at 13:30 from IEEE Xplore. Restrictions apply.
-
8/8/2019 A Multi Channel, Wireless Telemetric Micro System
16/16
202 IEEE SENSORS JOURNAL, VOL. 6, NO. 1, FEBRUARY 2006
Edward OConnor is a Technical Specialist in elec-tronics at the Center for Micro and Nano Processing,Case Western Reserve University, Cleveland, OH, apost which he has held for the past 31 years. He hasbeen working in the areas of biomedical electronicsand radio frequency telemetry. He has attendedthe Case Institute of Technology where he studiedelectrical engineering and has published his work in
biotelemetry. He holds four U.S. patents.
Kingman P. Strohl received the B.S. in anthro-pology from Yale University, New Haven, CT, in1970, the M.D. degree from Northwestern Uni-versity, Evanston, IL, in 1974, and he completedhis training in internal medicine at the Universityof Kentucky, Lexington, in 1977, and a researchfellowship in respiratory physiology and pulmonarymedicine at Peter Bent Brigham Hospital, Harvard
School of Health, Cambridge, MA, in 1980.Since 1980, he has been with Case Western Re-serve University, Cleveland, OH, and is now a Pro-
fessor of medicine, anatomy, and oncology at the School of Medicine. Thethemes advanced in funded research over this period of time include the me-chanical properties of the upper airway (1981 to 1988), biomarkers of hypoxia(1988 to 1997), and generic features of breathing and sleep (1997 to present),all relevant to clinical disorders of sleep apnea.