321 Lecture ECG

38
Electrocardiogram: The Basics BMEN 321 October 31, 2006

Transcript of 321 Lecture ECG

Page 1: 321 Lecture ECG

Electrocardiogram: The Basics

BMEN 321

October 31, 2006

Page 2: 321 Lecture ECG

Background Reading

• This lecture summarizes information in Bioelectromagnetism by Malmivuo and Plonsey and can be found on the web at butler.cc.tut.fi/~malmivuo/bem/bembook

• We cover chapters 15 and 19

Page 3: 321 Lecture ECG

The Heart

• The heart consists of four chambers– The right atrium and

right ventricle: responsible for delivery of deoxygenated blood to lungs

– The left atrium and left ventricle: responsible for delivery of oxygenated blood to the body

Page 4: 321 Lecture ECG

The Heart: PhasesThere are two phases of the cardiac cycle

• Systole: The ventricles are full of blood and begin to contract. The mitral and tricucuspid valves close (between atria and ventricles). Blood is ejected through the pulmonic and aortic valves.

• Diastole: Blood flows into the atria and through the open mitral and tricuspid valves into the ventricles.

Page 5: 321 Lecture ECG

ECG

• The ECG records the electrical signal of the heart as the muscle cells depolarize (contract) and repolarize.

• Normally, the SA Node generates the initial electrical impulse and begins the cascade of events that results in a heart beat.

• Recall that cells resting have a negative charge with respect to exterior and depolarization consists of positive ions rushing into the cell

Page 6: 321 Lecture ECG

Cell Depolarization

• Flow of sodium ions into cell during activation

Depol Repol. Restoration ofionic balance

Page 7: 321 Lecture ECG

ECG Leads• In 1908, Willem

Einthoven developed a system capable of recording these small signals and recorded the first ECG.

• The leads were based on the Einthoven triangle associated with the limb leads.

• Leads put heart in the middle of a triangle

Page 8: 321 Lecture ECG

ECG LeadsThe basic values

The lead values

Also note that by KVL: VI + VIII = VII

Page 9: 321 Lecture ECG

ECG: Electric Signal

• Assumptions– Model cardiac source as a dipole producing an

electric heart vector, p.– Model body as an infinite, homogeneous volume

conductor

• The leads will pick up the projection of the electric heart vector, p, along the lead

Page 10: 321 Lecture ECG

Propagating Activation Wavefront

• When the cells are at rest, they have a negative transmembrane voltage – surrounding media is positive

• When the cells depolarize, they switch to a positive transmembrane voltage – surrounding media becomes negative

• This leads to a propagating electric vector (pointing from negative to positive)

Page 11: 321 Lecture ECG

Propagating Activation Wavefront

Page 12: 321 Lecture ECG

Propagating Activation Wavefront

Depol. toward positive electrodePositive Signal

Repol. toward positive electrodeNegative Signal

Depol. away from positive electrodeNegative Signal

Repol. Away from positive electrodePositive Signal

Page 13: 321 Lecture ECG

Propagating Activation Wavefront

• When the activation does not align directly with the lead (or propagate directly toward and electrode), the signal is proportional to component of the activation direction along the lead direction.

Page 14: 321 Lecture ECG

ECG Signal• Heart behaves as a syncytium:

a propagating wave that once initiated continues to propagate uniformly into the region that is still at rest.

• The depolarization wavefront defines a dividing line between activated and resting cells.

• Elsewhere, the signal is zero• Will propagate along conduction

paths – sinus node – AV node – bundle branches – Purkinjie fibers

Page 15: 321 Lecture ECG

ECG Signal• The excitation begins at the

sinus (SA) node and spreads along the atrial walls

• The resultant electric vector is shown in yellow

• Cannot propagate across the boundary between atria and ventricle

• The projections on Leads I, II and III are all positive

Page 16: 321 Lecture ECG

ECG Signal

• Atrioventricular (AV) node located on atria/ventricle boundary and provides conducting path

• Pathway provides a delay to allow ventricles to fill

• Excitation begins with the septum

Page 17: 321 Lecture ECG

ECG Signal

• Depolarization continues to propagate toward the apex of the heart as the signal moves down the bundle branches

• Overall electric vector points toward apex as both left and right ventricles depolarize and begin to contract

Page 18: 321 Lecture ECG

ECG Signal• Depolarization of the right

ventricle reaches the epicardial surface

• Left ventricle wall is thicker and continues to depolarize

• As there is no compensating electric forces on the right, the electric vector reaches maximum size and points left

• Note the atria have repolarized, but signal is not seen

Page 19: 321 Lecture ECG

ECG Signal

• Depolarization front continues to propagate to the back of the left ventricular wall

• Electric vector decreases in size as there is less tissue depolarizing

Page 20: 321 Lecture ECG

ECG Signal

• Depolarization of the ventricles is complete and the electric vector has returned to zero

Page 21: 321 Lecture ECG

ECG Signal• Ventricular repolarization

begins from the outer side of the ventricles with the left being slightly dominant

• Note that this produces an electric vector that is in the same direction as the depolarization traveling in the opposite direction

• Repolarization is diffuse and generates a smaller and longer signal than depolarization

Page 22: 321 Lecture ECG

ECG Signal

• Upon complete repolarization, the heart is ready to go again and we have recorded an ECG trace

Page 23: 321 Lecture ECG

Normal ECG Signal

• P – atrial depolarization

• QRS complex – ventricular depolarization

• T – ventricular repolarization

Page 24: 321 Lecture ECG

Cardiac Cycle

Page 25: 321 Lecture ECG

Augmented Leads• Three additional limb leads are also used:

aVR, aVL, and aVF

• These are unipolar leads

• Each lead uses the average of the average of the other two leads as reference– VR = ΦR – (ΦL + ΦF)/2

Page 26: 321 Lecture ECG

Precordial Leads

• Measure potentials close to the heart, V1- V6

• Unipolar leads

Page 27: 321 Lecture ECG

ECG Information

• The 12 leads allow tracing of electric vector in all three planes of interest

• Not all the leads are independent, but are recorded for redundant information

Page 28: 321 Lecture ECG

ECG Diagnosis

• The trajectory of the electric vector resulting from the propagating activation wavefront can be traced by the ECG and used to diagnose cardiac problems

Page 29: 321 Lecture ECG

Electric Axis of the Heart• This axis changes during cardiac cycle as shown

earlier – generally lies between +30º and -110º in the frontal plane and +30º and -30º in the transverse plane

• Clinically, it is generally taken where the QRS complex has the largest positive deflection

• Note: Often use –aVR

• Deviation to R: increased activity in R vent. – obstruction in lung, pulmonary emboli, some heart disease

• Deviation to L: increased activity in L vent. – hypertension, aortic stenosis, ischemic heart disease

Page 30: 321 Lecture ECG

Cardiac Rhythm: SupraventricularNORMAL SINUS RHYTHMImpuses originate at S-A node at normal rate

                   

   

                                                                         

  SINUS TACHYCARDIAImpuses originate at S-A node at rapid rate

                   

   

                                                                         

  All complexes normal, evenly spacedRate > 100/min

SINUS TACHYCARDIAImpuses originate at S-A node at rapid rate

                   

   

                                                                         

  All complexes normal, rhythm is irregularLongest R-R interval exceeds shirtest > 0.16 s

Page 31: 321 Lecture ECG

Cardiac Rhythm: Supraventricular

ATRIAL FLUTTERImpulses travel in circular course in atria – No interval between T and P

                   

   

                                                                         

  Rapid flutter waves, ventricular response irregular

ATRIAL FIBRILLATIONImpuses have chaotic, random pathways in atria

                   

   

                                                                         

  Baseline irregular, ventricular response irregular

Page 32: 321 Lecture ECG

Cardiac Rhythm: VentricularPREMATURE VENTRICULAR CONTRACTIONA single impulse originates at right ventricle

                   

   

                                                                         

  Time interval between normal R peaks is a multiple of R-R intervals

VENTRICULAR TACHYCARDIAImpulse originate at ventricular pacemaker – odd/wide QRS complex - often due to myocardial infarction

                   

   

                                                                         

  Wide ventricular complexesRate> 120/min

Page 33: 321 Lecture ECG

Cardiac Rhythm: Ventricular

VENTRICULAR FIBRILLATIONChaotic ventricular depolarization – ineffective at pumping blood – death within minutes

                   

   

                                                                         

  Rapid, wide, irregular ventricular complexes

PACER RHYTHMImpulses originate at transvenous pacemaker

                   

   

                                                                         

  Wide ventricular complexes preceded by pacemaker spikeRate is the pacer rhythm

Page 34: 321 Lecture ECG

Activation Sequence DisordersA-V BLOCK, FIRST DEGREEAtrio-ventricular conduction lengthened

                   

   

                                                                         

  P-wave precedes each QRS-complex but PR-interval is > 0.2 s

A-V BLOCK, SECOND DEGREESudden dropped QRS-complex

                    

  

                                                                        

   Intermittently skipped ventricular beat

Page 35: 321 Lecture ECG

Bundle-branch BlockRIGHT BUNDLE-BRANCH BLOCKQRS duration greater than 0.12 sWide S wave in leads I, V5 and V6

                                                                           

               

Page 36: 321 Lecture ECG

Atrial Hypertropy: Enlarged AtriaRIGHT ATRIAL HYPERTROPHYTall, peaked P wave in leads I and II

LEFT ATRIAL HYPERTROPHYWide, notched P wave in lead IIDiphasic P wave in V1

                              

               

                              

               

Page 37: 321 Lecture ECG

Ventricular Hypertropy: Enlarged VentricleLEFT VENTRICULAR HYPERTROPHYLarge S wave in leads V1 and V2

Large R wave in leads V6 and V6

                                                                          

                

Page 38: 321 Lecture ECG

Myocardial Ischemia and Infarction

• Oxygen depletion to heart can cause an oxygen debt in the muscle (ischemia)

• If oxygen supply stops, the heart muscle dies (infarction)

• The infarct area is electrically silent and represents an inward facing electric vector…can locate with ECG