3130 Series Performance Verification User Bulletin (PN ...

64
User Bulletin Applied Biosystems 3130 Series Genetic Analyzers July 2006 SUBJECT: Performance Verification After Upgrade from ABI PRISM ® 3100 Series Genetic Analyzers In This User Bulletin This user bulletin includes the following topics: 1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 2 Before Upgrading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3 Materials and Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 3.1 Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 3.2 Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 4 Experimental Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 5 Data Collection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 6 Data Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 7 Results and Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 7.1 Results of a 3100-Avant-to-3130 Instrument Upgrade . . . . . . . . . 12 7.2 Results of a 3100-to-3130xl Instrument Upgrade . . . . . . . . . . . . . 24 7.3 Results of a 3100-Avant-to-3130xl Instrument Upgrade . . . . . . . 36 7.4 Results of a 3130-to-3130xl Instrument Upgrade . . . . . . . . . . . . . 48 8 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 9 References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Transcript of 3130 Series Performance Verification User Bulletin (PN ...

User BulletinApplied Biosystems 3130 Series Genetic Analyzers

July 2006

SUBJECT: Performance Verification After Upgrade from ABI PRISM® 3100 Series Genetic Analyzers

In This UserBulletin

This user bulletin includes the following topics:

1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Before Upgrading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Materials and Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

3.1 Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43.2 Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

4 Experimental Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Data Collection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Data Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Results and Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

7.1 Results of a 3100-Avant-to-3130 Instrument Upgrade . . . . . . . . .127.2 Results of a 3100-to-3130xl Instrument Upgrade. . . . . . . . . . . . .247.3 Results of a 3100-Avant-to-3130xl Instrument Upgrade . . . . . . .367.4 Results of a 3130-to-3130xl Instrument Upgrade. . . . . . . . . . . . .48

8 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 609 References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

2 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

1 Introduction

InstrumentUpgrade Paths

The following paths are available to upgrade an ABI PRISM® 3100 or 3100-Avant™ Genetic Analyzer to an Applied Biosystems 3130 or 3130xl Analyzer:

InstrumentComponents

Modified

The instrument components modified during the upgrade include:

• Polymer delivery system• Oven door gasket• CCD camera shutter• Detection cell heater• Data Collection Software upgrade to 3130 Series version

3.0 software

For details on the upgraded components, refer to the Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0, Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products User Bulletin (PN 4363787).

PerformanceVerification

StudiesPerformed

Although components of the optical system are not changed during the upgrade, DAB/SWGDAM guidelines require a performance verification after an upgrade. The following performance verification studies ensure that key performance parameters have not changed:

• Sensitivity• Mixture• Precision• Resolution• Concordance

IMPORTANT! Human identification laboratories should perform their own performance verification studies.

Instrument Can be upgraded to

3100 3130xl

3100-Avant 3130

3130xl

3130 3130xl

2 Before Upgrading

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 3

2 Before UpgradingIMPORTANT! Perform the following steps before upgrading an instrument:

• Review the Applied Biosystems 3130 and 3130xl Genetic Analyzer Site Preparation Guide to ensure that the laboratory environment meets specifications for a 3130 or 3130xl instrument.

• Back up all sample data. During the upgrade, the instrument computer is reimaged (all data are erased) before the Data Collection software is installed. It is not necessary to back up spatial calibration and spectral calibration files because these calibrations are performed at the end of the upgrade.

Reagents

4 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

3 Materials and Methods

3.1 ReagentsTo minimize variability between studies, the following setup procedures were used throughout the study:

• For the sensitivity studies, serial dilutions were prepared from the quantified stock control DNA (9947A, 007, and 9948) based on the average concentration determined using the Quantifiler® Human DNA Quantification Kit and the ABI PRISM® 7000 Sequence Detection System with SDS Software v1.0.

• For all studies, multiple PCR product reactions were pooled to provide sufficient quantity and minimize variability.

• For all studies, one manufactured lot of each of the following kits was used:

3.2 InstrumentsThree instruments of each type were used for each study. Figure 1 shows the instrument upgrade paths, instrument numbers, and the performance assessment test numbers performed. Table 1 on page 5 explains the performance assessment tests.

Figure 1 Tests performed on each instrument before and after upgrades

Kit AB Part Number

Quantifiler® Human DNA Quantification Kit 4343895

AmpFlSTR® Identifiler® PCR Amplification Kit 4322288

AmpFlSTR® Profiler Plus® PCR Amplification Kit 4303326

3100-Avant-003

Test 1

Upgrade to3100-Avant-040

3100-Avant-019

3130-003

Test 2

3130-040

3130-019

3130xl-003

Test 3

3130xl-040

3130xl-019

3100-006

Test 4

3100-015

3100-028

3130xl-006

Test 5

3130xl-015

3130xl-028

Upgrade to

Upgrade to

3 Materials and Methods

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 5

Table 1 Performance assessment tests (studies performed in each test: sensitivity, mixture, precision, resolution, and concordance)

Test Instrument Instrument ID Numbers Purpose

1 3100-Avant™ before upgrade

003, 040, and 019 To generate baseline data on the3100-Avant Genetic Analyzer before upgrade.

2 3130:

• After upgrade from 3100-Avantand

• Before upgrade to 3130xl

003, 040, and 019 3100-Avant-to- 3130 upgrade: To generate data using the same studies as Test 1, then to compare the results from Test 1 to the results from Test 2.

3130-to-3130xl upgrade: To generate baseline data on the 3130 Genetic Analyzer before upgrade.

3 3130xl after upgrade from 3130

003, 040, and 019 To generate data using the same studies as Test 1, then to compare the results from Test 1 to the results from Test 3.

4 3100 before upgrade 006, 015, and 028 To generate baseline data on the3100 Genetic Analyzers before upgrade.

5 3130xl after upgrade from 3100

006, 015, and 028 To generate data using the same studies as Test 4, then to compare the results from Test 4 to the results from Test 5.

Instruments

6 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

4 Experimental Setup

Quantitation All DNA samples were quantitated using the Quantifiler® Human DNA Quantification Kit according to the protocols defined in the Quantifiler® Kits User's Guide.

Amplification DNA was amplified:

• By PCR on a GeneAmp® PCR System 9700.

Note: The Aluminum 96-Well GeneAmp® PCR System 9700 Sample Block Module is not supported for use with the AmpFlSTR® chemistries.

• Using the following kits and reaction volumes according to the recommended protocols in the AmpFlSTR® user's manuals:

Sensitivity Study Sensitivity studies were run using three different DNA samples: 007, 9947A, and 9948.

When DNA input quantity is low, stochastic PCR effects may result in heterozygote peak imbalance, possible allele dropout, and low overall relative fluorescence signal.

Before and after upgrade, sensitivity studies were conducted to determine the detection capabilities of the instruments. As expected and reported in Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection v3.0 (PN 4363787), variation in sensitivity from instrument to instrument was observed.

Kit Reaction Volume (µL)

AB Part Number

AmpFlSTR® Identifiler® PCR Amplification Kit

25 4322288

AmpFlSTR® Profiler Plus®

PCR Amplification Kit50 4303326

4 Experimental Setup

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 7

Serial dilutions were made with 0.1 mM Na2EDTA (T10E0.1 buffer). The following DNA input amounts were amplified:

All samples were replicated twice on the plate, and each replicate was injected three times (for a total of six injections per sample).

The sensitivity study evaluated the following before and after upgrade of each instrument:

• The ability to detect a full genotype profile in a range of DNA input amounts, determined by evaluating the number of alleles that fell below the detection threshold

• Average peak heights• Average heterozygote peak height ratios

The GeneMapper® ID Software v3.2 analysis threshold for the sensitivity study was set at 50 RFU (relative fluorescence units).

Peak Height Calculations

Peak heights for all alleles from all samples in a test were used to calculate a single average peak height value. This average peak height value was then compared between tests.

To compare peak heights by chemistry, peak heights for all alleles from all samples in a test run using one chemistry were used to calculate a single average peak height value. This average peak height value was then compared between tests performed using the same chemistry.

Kit DNA Input Amounts Amplified

AmpFlSTR Identifiler PCR Amplification Kit

• 1.5 ng• 1.0 ng• 500 pg

• 250 pg• 125 pg

AmpFlSTR Profiler Plus PCR Amplification Kit

• 3.0 ng• 2.0 ng• 1.0 ng

• 500 pg• 250 pg

Instruments

8 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Mixture Study DNA samples 9947A and 9948 were used to prepare a set of DNA mixtures at the following 9947A:9948 ratios: 0:1, 1:1, 2:1, 4:1, 9:1, and 1:0 (Table 2).

All amplified samples were replicated twice on the plate, and each replicate was injected three times (for a total of six injections per sample). The samples were evaluated for the ability to detect the minor contributor at each ratio.

Total DNA input was 1.5 ng per reaction for the Identifiler® Kit and 3.0 ng per reaction for the Profiler Plus® Kit kit.

The GeneMapper ID Software v3.2 analysis threshold for this study was set at 50 RFU (relative fluorescence units).

Number of Expected Alleles

The number of expected alleles for mixtures was determined from the 1:1 mixture ratio. When the major and minor contributor shared an allele, the shared allele was counted as a single allele.

If an allele from the minor contributor fell in the stutter position of an allele from the major contributor, the allele from the minor contributor was identified only if its peak height exceeded the stutter filter threshold. Typically, minor contributor alleles in the stutter position of a major contributor allele are filtered as stutter in the 1:10 mixture ratio.

Table 2 Mixture composition

Kit Mixture Ratio

Total Female DNA Input (ng)

Total Male DNA Input (ng)

AmpFlSTR Identifiler PCR Amplification Kit

1:0 1.50 0.001:1 0.75 0.754:1 1.20 0.309:1 1.35 0.150:1 0.00 1.50

AmpFlSTR Profiler Plus PCR Amplification Kit

1:0 3.00 0.001:1 1.50 1.504:1 2.40 0.609:1 2.70 0.300:1 0.00 3.00

5 Data Collection

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 9

Precision Study Allelic ladders from the Identifiler® Kit and Profiler Plus® Kit were used to calculate the standard deviation of the size (in bp) for each allele in each allelic ladder. Sixteen replicates of each ladder were loaded on a 96-well plate, and each replicate was injected three times (for a total of 48 injections per sample).

Sizing averages were calculated within a run (that is, 16 samples for the 3100 and 3130xl 16-capillary instruments and four samples for the 3100-Avant and 3130 4-capillary instruments).

Sizing precision between runs of allelic ladder was not calculated during this study.

Resolution Study The TH01 locus of the Identifiler® Kit allelic ladder was examined to evaluate the resolution and detection of alleles differing in size by 1 base pair.

ConcordanceStudy

The genotype results from the sensitivity study were evaluated for concordance.

5 Data CollectionData was collected using the following versions of software:

All instruments used the default Run Module with the appropriate dye set.

Instrument Instrument ID Numbers

Data Collection Version

3100-Avant™ 003 2.0040 1.0019 1.0

3100 006 1.1015 1.1028 2.0

3130 003, 040, and 019 3.03130xl 003, 040, 019,

006, 015, and 0283.0

Instruments

10 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

For more information, see:

• User Bulletin ABI PRISM® 3100/3100-Avant™ Genetic Analyzers Using Data Collection v2.0: Protocols for Processing AmpFlSTR® PCR Amplification Kit PCR Products (PN 4350218, Rev. A)

• User Bulletin Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 (PN 4363787)

• ABI PRISM® 3100 Genetic Analyzer User's Manual (PN 4315834B)

• Applied Biosystems 3130/3130xl Genetic Analyzers: Getting Started Guide (PN 4352715)

6 Data Analysis

Initial DataCompiling and

Analysis

All data were analyzed using GeneMapper® ID Software v3.2 with:

• HID_Advanced Analysis Method • Peak Detection Threshold of 50 RFU

For information on setting up the analysis method, see the GeneMapper® ID Software Version 3.1 and 3.2 Human Identification Analysis Tutorial (P/N 4357520 Rev. A).

Final DataAnalysis

Minitab® Statistical Software was used to:

• Calculate averages • Generate the box and whisker plots

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 11

7 Results and DiscussionElectropherograms in this section were generated using GeneMapper® ID Software Version 3.2.

Data plots in this section were generated using Minitab® Statistical Software and reported in Box-and-Whisker-Plot format (Figure 2).

Figure 2 Box-and-whisker plot elements

Maximum value

Minimum value

75th percentile

Median value

25th percentile

Outliers

where:• Whiskers extend to the minimum and maximum data values.• Median = The middle of the data when it is ordered from smallest to largest value. • 25th percentile (lower quartile) = The median of the lower half of the data. • 75th percentile (upper quartile) = The median of the upper half of the data. • Minimum value = The smallest observed value. • Maximum value = The largest observed value.• Values beyond the whiskers are outliers. Outliers are unusually large or small

values and are represented as dots on the plot.

Results of a 3100-Avant™-to-3130 Instrument Upgrade

12 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.1 Results of a 3100-Avant™-to-3130 Instrument Upgrade

7.1.1 Sensitivity Study

The sensitivity study evaluated the following before and after upgrade of each instrument:

• The ability to detect a full genotype profile in a range of DNA input amounts, determined by evaluating the number of alleles that fell below the detection threshold

• Average peak heights• Average heterozygote peak height ratios

Number of AllelesBelow Detection

Threshold

Table 3 shows the number of alleles that fell below the detection threshold before and after upgrading the 3100-Avant instrument to a 3130 instrument:

• Before upgrade – Sixty four (64) alleles (out of a total of 8725) alleles fell below the detection threshold (Figure 3).

• After upgrade – Sixty four (64) alleles (out of a total of 9915) alleles fell below the detection threshold (Figure 4).

As expected, instrument-to-instrument variation was observed between instruments of the same type before and after upgrade.

Table 3 Sensitivity study – Number of alleles below detection threshold – 3100-Avant™-to-3130 instrument upgrade

Test 1 (before upgrade) Test 2 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

3100-Avant-003 22 3678 3130-003 2 3259

3100-Avant-040 5 3082 3130-040 9 3273

3100-Avant-019 37 1965 3130-019 53 3383

Total 64 8725 Total 64 9915

‡. Data from failed injections not included

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 13

Full genotypes were consistently obtained with 250 pg DNA input amounts for all samples amplified with the Identifiler® and Profiler Plus® kits before and after upgrading the 3100-Avant instrument to a 3130 instrument. All alleles that fell below the detection limit were obtained from the 125 pg amplification.

Alleles that fall below the threshold can be caused by:

• Low DNA input, which results in low overall relative fluorescence

• Stochastic PCR effects, which result in:– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3100-Avant™-to-3130 Instrument Upgrade

14 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 3 Sensitivity study – Identifiler® Kit electropherograms for a 3100-Avant™ instrument before upgrade (Y-scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

Serial dilutions of 9947A DNA

0.125 ng

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 15

Figure 4 Sensitivity study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 3) after 3100-Avant™-to-3130 instrument upgrade (Y scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

Serial dilutions of 9947A DNA

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

0.125 ng

Results of a 3100-Avant™-to-3130 Instrument Upgrade

16 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Peak Height A comparison of peak heights for each DNA input amount before and after instrument upgrade showed little variation in sensitivity (for a description of how peak heights were evaluated, see page 7):

• Overall, a 1.2% increase in peak heights was observed after upgrading the 3100-Avant instruments to 3130 instruments.

• Identifiler® Kit data showed a 4.1% decrease in overall peak height after upgrade (Figure 5).

• Profiler Plus® Kit data showed an 11.2% increase in overall peak height after upgrade (Figure 6).

Figure 5 Sensitivity study – Comparison of Identifiler® Kit peak heights by DNA input amount for all samples before (Test 1) and after (Test 2) 3100-Avant™-to-3130 instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 17

Figure 6 Sensitivity study – Comparison of Profiler Plus® Kit peak heights by DNA input amount for all samples before (Test 1) and after (Test 2) 3100-Avant™-to-3130 instrument upgrade

HeterozygotePeak Height

Ratio

Heterozygote peak height ratios were calculated and plotted for each DNA sample and concentration in the sensitivity study (for a description of how peak heights were evaluated, see page 7).

Heterozygote peak height ratios for all loci in the Identifiler® Kit (Figure 7) and Profiler Plus® Kit (Figure 8) averaged 80% or greater before and after upgrading the 3100-Avant instruments to 3130 instruments.

Heterozygote peak height ratios below the 70% cutoff before and after upgrade occurred mainly with lower DNA input amounts (Figures 9 and 10).

Low heterozygote peak height ratios can be caused by:

• Low DNA input, which results in low overall relative fluorescence

• Stochastic PCR effects, which result in:– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3100-Avant™-to-3130 Instrument Upgrade

18 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 7 Sensitivity study – Comparison of Identifiler® Kit

heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 1) and after (Test 2) 3100-Avant™-to-3130 instrument upgrade

Figure 8 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 1) and after (Test 2) 3100-Avant™-to-3130 instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 19

Figure 9 Sensitivity study – Comparison of Identifiler® Kit heterozygote peak height ratio values by input amount for all samples before (Test 1) and after (Test 2) 3100-Avant™-to-3130 instrument upgrade

Figure 10 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by input amount for all samples before (Test 1) and after (Test 2) 3100-Avant™-to-3130 instrument upgrade

Results of a 3100-Avant™-to-3130 Instrument Upgrade

20 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.1.2 Mixture Study

Mixed DNA profiles were evaluated, and the number of alleles falling below 50 RFU was recorded before and after instrument upgrade.

Number of AllelesBelow Detection

Threshold

For a description of how the number of expected alleles was determined, see page 8.

Table 4 shows the number of alleles that fell below the detection threshold before and after upgrading the 3100-Avant instrument to a 3130 instrument:

• Before upgrade – Fifty seven (57) alleles (out of a total of 5832) fell below the detection threshold (Figure 11).

• After upgrade – Thirty eight (38) alleles (out of a total of 5643) fell below the detection threshold (Figure 12).

The limit of detection of the minor contributor was approximately 9:1 before and after upgrading the 3100-Avant instrument to a 3130 instrument

Table 4 Mixture study – Number of alleles below detection threshold – 3100-Avant™-to-3130 instrument upgrade (data for 1:1 mixture not included)

Test 1 (before upgrade) Test 2 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

3100-Avant-003 12 1944 3130-003 12 1944

3100-Avant-040 33 1944 3130-040 13 1755

3100-Avant-019 12 1944 3130-019 13 1944

Total 57 5832 Total 38 5643

‡. Data from failed injections not included

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 21

Figure 11 Mixture study – Identifiler® Kit electropherograms for a 3100-Avant™ instrument before upgrade (Y-scale is constant in each of the upper panels)

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9 FGA 26

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

Results of a 3100-Avant™-to-3130 Instrument Upgrade

22 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 12 Mixture study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 11) after 3100-Avant™-to-3130 instrument upgrade

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9FGA 26

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 23

7.1.3 Precision Study

The standard deviation in the size value for each allele of the allelic ladder was calculated within a run (four samples per run) before and after upgrading each of the three instruments:

• All Identifiler® Kit and Profiler Plus® Kit runs before and after upgrade yielded standard deviation values <0.15 bp (Figure 13).

• A small overall increase in the standard deviation was observed after upgrade in the Identifiler® Kit data (0.005 bp) and Profiler Plus® Kit data (0.007 bp) (data not shown).

Figure 13 Precision study – Sizing data for Identifiler® Kit and Profiler Plus® Kit allelic ladders before (Test 1) and after (Test 2) 3100-Avant™-to-3130 instrument upgrade

7.1.4 Resolution Study

The TH01 locus of the Identifiler® Kit allelic ladder was examined to evaluate the resolution and detection of alleles differing in size by 1 base pair. The 9.3 and 10 alleles of TH01 were resolved and individually detected for all injections before and after upgrading all instruments.

7.1.5 Concordance Study

Amplification of the 9947A, 9948, and 007 DNA samples produced correct genotypes for all loci for both the Identifiler® Kit and the Profiler Plus® Kit before and after upgrading all instruments.

Results of a 3100-to-3130xl Instrument Upgrade

24 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.2 Results of a 3100-to-3130xl Instrument Upgrade

7.2.1 Sensitivity Study

The sensitivity study evaluated the following before and after upgrade of each instrument:

• The ability to detect a full genotype profile in a range of DNA input amounts, determined by evaluating the number of alleles that fell below the detection threshold

• Average peak heights• Average heterozygote peak height ratios

Number of AllelesBelow Detection

Threshold

Table 5 shows the number of alleles that fell below the detection threshold before and after upgrading the 3100 instrument to a 3130xl instrument:

• Before upgrade – Fifty six (56) alleles (out of a total of 7218) fell below the detection threshold (Figure 14).

• After upgrade – Four (4) alleles (out of a total of 10330) fell below the detection threshold (Figure 15).

As expected, instrument-to-instrument variation was observed between instruments of the same type before and after upgrade.

Table 5 Sensitivity study – Number of alleles below detection threshold – 3100-to-3130xl instrument upgrade

Test 4 (before upgrade) Test 5 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

3100-006 Not included§

3183 3130xl-006 2 3396

3100-015 55 3540 3130xl-015 2 3432

3100-028 1 3678 3130xl-028 0 3502

Total 56 7218 Total 4 10330

‡. Data from failed injections not included§. Instrument 3100-006 injections failed

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 25

Full genotypes were consistently obtained with 250 pg DNA input for all samples amplified with the Identifiler® Kit and Profiler Plus® Kit before and after upgrading the 3100 instrument to a 3130xl instrument. All alleles that fell below the detection limit were obtained from the 125 pg amplification.

Alleles that fall below the threshold can be caused by:• Low DNA input, which results in low overall relative

fluorescence • Stochastic PCR effects, which result in:

– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3100-to-3130xl Instrument Upgrade

26 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 14 Sensitivity study – Identifiler® Kit electropherograms for a 3100 instrument before upgrade (Y-scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

Serial dilutions of 9947A DNA

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

0.125 ng

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 27

Figure 15 Sensitivity study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 14) after 3100-to-3130xl instrument upgrade (Y scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

Serial dilutions of 9947A DNA

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

0.125 ng

Results of a 3100-to-3130xl Instrument Upgrade

28 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Peak Height A comparison of peak heights for each DNA input amount before and after instrument upgrade showed some variation in sensitivity (for a description of how peak heights were evaluated, see page 7):

• Overall, a 8.9% increase in peak heights was observed after upgrading the 3100 instruments to 3130xl instruments.

• Identifiler® Kit data showed a 4.5% increase in overall peak height after upgrade (Figure 16).

• Profiler Plus® Kit data showed a 13.7% increase in overall peak height after upgrade (Figure 17).

Figure 16 Sensitivity study – Comparison of Identifiler® Kit peak heights by input amount for all samples before (Test 4) and after (Test 5) 3100-to-3130xl instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 29

Figure 17 Sensitivity study – Comparison of Profiler Plus® Kit peak heights by input amount for all samples before (Test 4) and after (Test 5) 3100-to-3130xl instrument upgrade

HeterozygotePeak Height

Ratio

Heterozygote peak height ratios were calculated and plotted for each DNA sample and concentration in the sensitivity study (for a description of how peak heights were evaluated, see page 7).

Heterozygote peak height ratios for all loci in Identifiler® Kit (Figure 18) and Profiler Plus® Kit (Figure 19) kits averaged 80% or greater before and after upgrading the 3100 to 3130xl instruments.

Heterozygote peak height ratios below the 70% cutoff before and after upgrade occurred mainly with lower DNA input amounts (Figures 20 and 21).

Low heterozygote peak height ratios can be caused by:

• Low DNA input, which results in low overall relative fluorescence

• Stochastic PCR effects, which result in:– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3100-to-3130xl Instrument Upgrade

30 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 18 Sensitivity study – Comparison of Identifiler® Kit heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 4) and after (Test 5) 3100-to-3130xl instrument upgrade

Figure 19 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 4) and after (Test 5) 3100-to-3130xl instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 31

Figure 20 Sensitivity study – Comparison of Identifiler® Kit heterozygote peak height ratio values by input amount for all samples before (Test 4) and after (Test 5) 3100-to-3130xl instrument upgrade

Figure 21 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by input amount for all samples before (Test 4) and after (Test 5) 3100-to-3130xl instrument upgrade

Results of a 3100-to-3130xl Instrument Upgrade

32 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.2.2 Mixture Study

Mixed DNA profiles were evaluated and the number of alleles falling below 50 RFU was recorded before and after instrument upgrade.

Number of AllelesBelow Detection

Threshold

For a description of how the number of expected alleles was determined, see page 8.

Table 6 shows the number of alleles that fell below the detection threshold before and after upgrading the 3130 to a 3130xl:

• Before upgrade – Twenty one (21) alleles (out of a total of 4536) fell below the detection threshold (Figure 22).

• After upgrade – Forty two (42) alleles (out of a total of 5832) fell below the detection threshold (Figure 23).

The limit of detection of the minor contributor was approximately 9:1 before and after upgrading the 3100 instrument to a 3130xl instrument.

Table 6 Mixture study – Number of alleles below detection threshold – 3100-to-3130xl upgrade (data for 1:1 mixture not included)

Test 4 (before upgrade) Test 5 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

3100-006 8 1620 3130xl-006 14 1944

3100-015 0 972 3130xl-015 14 1944

3100-028 13 1944 3130xl-028 14 1944

Total 21 4536 Total 42 5832

‡. Data from failed injections not included

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 33

Figure 22 Mixture study – Identifiler® Kit electropherograms for a 3100 instrument before upgrade (Y-scale is constant in each of the upper panels)

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9 FGA 26

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

Results of a 3100-to-3130xl Instrument Upgrade

34 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 23 Mixture study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 22) after 3100-to-3130xl instrument upgrade

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9 FGA 26

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 35

7.2.3 Precision Study

The standard deviation in the size value for each allele of the allelic ladder was calculated within a run (sixteen samples per run for the 3100 instrument, 16 samples per run for the 3130xl instrument) before and after upgrading each of the three instruments:

• All Identifiler® Kit and Profiler Plus® Kit runs before and after instrument upgrade yielded standard deviation values <0.15 bp (Figure 24).

• A small overall decrease in the standard deviation was observed after upgrade in the Identifiler® Kit data (0.003 bp) and Profiler Plus® Kit data (0.011 bp) (data not shown).

Figure 24 Precision study – Sizing data for Identifiler® Kit and Profiler Plus® Kit before (Test 4) and after (Test 5) 3100-to-3130xl instrument upgrade

7.2.4 Resolution Study

The TH01 locus of the Identifiler® Kit allelic ladder was examined to evaluate the resolution and detection of alleles differing in size by 1 base pair. The 9.3 and 10 alleles of TH01 were resolved and individually detected for all injections before and after upgrading all instruments.

7.2.5 Concordance StudyAmplification of the 9947A, 9948, and 007 DNA samples produced correct genotypes for all loci for both Identifiler® Kit and Profiler Plus® Kit before and after upgrading.

Results of a 3100-Avant™-to-3130xl Instrument Upgrade

36 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.3 Results of a 3100-Avant™-to-3130xl Instrument Upgrade

For this upgrade path:• Data was collected on an 3100-Avant instrument (Test 1).• The 3100-Avant instrument was upgraded to a 3130 instrument.• The 3130 instrument was upgraded to a 3130xl instrument.• Data was collected on the 3130xl instrument (Test 3).

7.3.1 Sensitivity Study

The sensitivity study evaluated the following before and after upgrade of each instrument:

• The ability to detect a full genotype profile in a range of DNA input amounts, determined by evaluating the number of alleles that fell below the detection threshold

• Average peak heights• Average heterozygote peak height ratios

Number of AllelesBelow Detection

Threshold

Table 7 shows the number of alleles that fell below the detection threshold before and after upgrading the 3100-Avant instrument to a 3130xl instrument:

• Before upgrade – Sixty four (64) alleles (out of a total of 8725) fell below the detection threshold (Figure 25).

• After upgrade – Thirty four (34) alleles (out of a total of 8795) fell below the detection threshold (Figure 26).

As expected, instrument-to-instrument variation was observed between instruments of the same type before and after upgrade.

Table 7 Sensitivity study – Number of alleles below detection threshold – 3100-Avant™-to-3130xl instrument upgrade

Test 1 (before upgrade) Test 3 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50

RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50

RFU)

Total Number of

Alleles‡

3100-Avant-003 22 3678 3130xl-003 34 34133100-Avant-040 5 3082 3130xl-040 0 35903100-Avant-019 37 1965 3130xl-019 0 1792Total 64 8725 Total 34 8795

‡. Data from failed injections not included

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 37

Full genotypes were consistently obtained with 250 pg DNA input amounts for all samples amplified with the Identifiler® and Profiler Plus® kits before and after upgrading the 3100-Avant instrument to a 3130xl instrument. All alleles that fell below the detection limit were obtained from the 125 pg amplification.

Alleles that fall below the threshold can be caused by:• Low DNA input, which results in low overall relative

fluorescence • Stochastic PCR effects, which result in:

– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3100-Avant™-to-3130xl Instrument Upgrade

38 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 25 Sensitivity study – Identifiler® Kit electropherograms for a 3100-Avant™ instrument before upgrade (Y-scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

Serial dilutions of 9947A DNA

0.125 ng

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 39

Figure 26 Sensitivity study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 25) after 3100-Avant™-to-3130xl instrument upgrade (Y-scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

Serial dilutions of 9947A DNA

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

0.125 ng

Results of a 3100-Avant™-to-3130xl Instrument Upgrade

40 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Peak Height A comparison of peak heights for each DNA input amount before and after instrument upgrade showed some variation in sensitivity (for a description of how peak heights were evaluated, see page 7):

• Overall, a 28.7% increase in peak heights was observed after upgrading the 3100-Avant instruments to 3130xl instruments.

• Identifiler® Kit data showed a 29.8% increase in overall peak height after upgrade (Figure 27).

• Profiler Plus® Kit data showed a 27.6% increase in overall peak height after upgrade (Figure 28).

Figure 27 Sensitivity study – Comparison of Identifiler® Kit peak heights by input amount for all samples before (Test 1) and after (Test 3) 3100-Avant™-to-3130xl instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 41

Figure 28 Sensitivity study – Comparison of Profiler Plus® Kit peak heights input amount for all samples before (Test 1) and after (Test 3) 3100-Avant™-to-3130xl instrument upgrade

HeterozygotePeak Height

Ratio

Heterozygote peak height ratios were calculated and plotted for each DNA sample and concentration in the sensitivity study (for a description of how peak heights were evaluated, see page 7).

Heterozygote peak height ratios for all loci in Identifiler® Kit (Figure 29) and Profiler Plus® Kit (Figure 30) kits averaged 80% or greater before and after upgrading the 3100-Avant instruments to 3130xl instruments.

Heterozygote peak height ratios below the 70% cutoff before and after upgrade occurred mainly with lower DNA input amounts (Figures 31 and 32).

Low heterozygote peak height ratios can be caused by:

• Low DNA input, which results in low overall relative fluorescence

• Stochastic PCR effects, which result in:– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3100-Avant™-to-3130xl Instrument Upgrade

42 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 29 Sensitivity study – Comparison of Identifiler® Kit heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 1) and after (Test 3) 3100-Avant™-to-3130xl instrument upgrade

Figure 30 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 1) and after (Test 3) 3100-Avant™-to-3130xl instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 43

Figure 31 Sensitivity study – Comparison of Identifiler® Kit heterozygote peak height ratio values by input amount for all samples before (Test 1) and after (Test 3) 3100-Avant™-to-3130xl instrument upgrade

Figure 32 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by input amount for all samples before (Test 1) and after (Test 3) 3100-Avant™-to-3130xl instrument upgrade

Results of a 3100-Avant™-to-3130xl Instrument Upgrade

44 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.3.2 Mixture Study

Mixed DNA profiles were evaluated and the number of alleles falling below 50 RFU was recorded before and after instrument upgrade.

Number of AllelesBelow Detection

Threshold

For a description of how the number of expected alleles was determined, see page 8.

Table 8 shows the number of alleles that fell below the detection threshold before and after upgrading the 3100-Avant instrument to a 3130xl instrument:

• Before upgrade – Thirty five (35) alleles (out of a total of 5832) fell below the detection threshold (Figure 33).

• After upgrade – Thirty seven (37) alleles (out of a total of 5562) fell below the detection threshold (Figure 34).

The limit of detection of the minor contributor was approximately 9:1 before and after upgrading the 3100-Avant instrument to a 3130xl instrument.

Table 8 Mixture study – Number of alleles below detection threshold – 3100-Avant™-to-3130xl instrument upgrade (data for 1:1 mixture not included)

Test 1 (before upgrade) Test 3 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

3100-Avant-003 12 1944 3130xl-003 0 1782

3100-Avant-040 33 1944 3130xl-040 13 1863

3100-Avant-019 12 1944 3130xl-019 24 1917

Total 35 5832 Total 37 5562

‡. Data from failed injections not included

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 45

Figure 33 Mixture study – Identifiler® Kit electropherograms for a 3100-Avant™ instrument before upgrade (Y-scale is constant in each of the upper panels)

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9 FGA 26

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

Results of a 3100-Avant™-to-3130xl Instrument Upgrade

46 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 34 Mixture study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 33) after 3100-Avant™-to-3130xl instrument upgrade

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9FGA 26

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 47

7.3.3 Precision StudyThe standard deviation in the size value for each allele of the allelic ladder was calculated within a run (four samples per run for the 3100-Avant instrument, 16 samples per run for the 3130xl instrument) before and after upgrading each of the three instruments

• All Identifiler® Kit and Profiler Plus® Kit runs before and after instrument upgrade gave standard deviation values <0.15 bp (Figure 35).

• A small overall increase in the standard deviation was observed after upgrade in the Identifiler® Kit data (0.045 bp) and Profiler Plus® Kit data (0.030 bp) (data not shown).

Figure 35 Precision study – Sizing data for Identifiler® Kit and Profiler Plus® Kit before (Test 1) and after (Test 3) 3100-Avant™-to-3130xl instrument upgrade

7.3.4 Resolution Study

The TH01 locus of the Identifiler® Kit allelic ladder was examined to evaluate the resolution and detection of alleles differing in size by 1 base pair. The 9.3 and 10 alleles of TH01 were resolved and individually detected for all injections before and after upgrading all instruments.

7.3.5 Concordance Study

Amplification of the 9947A, 9948, and 007 DNA samples produced correct genotypes for all loci for both the Identifiler® Kit and the Profiler Plus® Kit before and after upgrading all instruments.

Results of a 3130-to-3130xl Instrument Upgrade

48 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.4 Results of a 3130-to-3130xl Instrument Upgrade

7.4.1 Sensitivity Study

The sensitivity study evaluated the following before and after upgrade of each instrument:

• The ability to detect a full genotype profile in a range of DNA input amounts, determined by evaluating the number of alleles that fell below the detection threshold

• Average peak heights• Average heterozygote peak height ratios

Number of AllelesBelow Detection

Threshold

Table 9 shows the number of alleles that fell below the detection threshold before and after upgrading the 3130 instrument to a 3130xl instrument:

• Before upgrade – Forty one (41) alleles (out of a total of 9851) fell below the detection threshold (Figure 36).

• After upgrade – Thirty four (34) alleles (out of a total of 8795) fell below the detection threshold (Figure 37).

As expected, instrument-to-instrument variation was observed between instruments of the same type before and after upgrade.

Table 9 Sensitivity study – Number of alleles below detection threshold – 3130-to-3130xl instrument upgrade

Test 2 (before upgrade) Test 3 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

3130-003 2 3257 3130xl-003 34 3413

3130-040 9 3264 3130xl-040 0 3590

3130-019 30 3330 3130xl-019 0 1792

Total 41 9851 Total 34 8795

‡. Data from failed injections not included

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 49

Full genotypes were consistently obtained with 250 pg DNA input for all samples amplified with the Identifiler® Kit and Profiler Plus® Kit before and after upgrading the 3130 instrument to a 3130xl instrument. All alleles that fell below the detection limit were obtained from the 125 pg amplification.

Alleles that fall below the threshold can be caused by:• Low DNA input, which results in low overall relative

fluorescence • Stochastic PCR effects, which result in:

– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3130-to-3130xl Instrument Upgrade

50 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 36 Sensitivity study – Identifiler® Kit electropherograms for a 3130 instrument before upgrade (Y-scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

Serial dilutions of 9947A DNA

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

0.125 ng

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 51

Figure 37 Sensitivity study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 36) after 3130-to-3130xl instrument upgrade (Y scale is constant in each of the upper panels)

1.5 ng

1.0 ng

0.5 ng

0.25 ng

0.125 ng

Serial dilutions of 9947A DNA

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

0.125 ng

Results of a 3130-to-3130xl Instrument Upgrade

52 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Peak Height A comparison of peak heights for each DNA input amount before and after instrument upgrade showed some variation in sensitivity (for a description of how peak heights were evaluated, see page 7):

• Overall, a 27.4% increase in peak heights was observed after upgrading the 3130 to 3130xl instruments.

• Identifiler® kit data showed a 32.7% increase in overall peak height after upgrade (Figure 38).

• Profiler Plus® Kit data showed a 19.5% increase in overall peak height after upgrade (Figure 39).

Figure 38 Sensitivity study – Comparison of Identifiler® Kit peak heights by DNA input amount for all samples before (Test 2) and after (Test 3) 3130-to-3130xl instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 53

Figure 39 Sensitivity study – Comparison of Profiler Plus® Kit peak heights by DNA input amount for all samples before (Test 2) and after (Test 3) 3130-to-3130xl instrument upgrade

HeterozygotePeak Height

Ratio

Heterozygote peak height ratios were calculated and plotted for each DNA sample and concentration in the sensitivity study (for a description of how peak heights were evaluated, see page 7). Heterozygote peak height ratios for all loci in Identifiler® Kit (Figure 40) and Profiler Plus® Kit (Figure 41) kits averaged 80% or greater before and after upgrading the 3130 instruments to 3130xl instruments.

Heterozygote peak height ratios below the 70% cutoff before and after upgrade occurred mainly with lower DNA input amounts (Figures 42 and 43).

Low heterozygote peak height ratios can be caused by:

• Low DNA input, which results in low overall relative fluorescence

• Stochastic PCR effects, which result in:– Peak imbalance between the heterozygote alleles – Allele dropout

Results of a 3130-to-3130xl Instrument Upgrade

54 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 40 Sensitivity study – Comparison of Identifiler® Kit heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 2) and after (Test 3) 3130-to-3130xl instrument upgrade

Figure 41 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by locus for all samples and DNA input amounts before (Test 2) and after (Test 3) 3130-to-3130xl instrument upgrade

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 55

Figure 42 Sensitivity study – Comparison of Identifiler® Kit heterozygote peak height ratio values by DNA input amount for all samples before (Test 2) and after (Test 3) 3130-to-3130xl instrument upgrade

Figure 43 Sensitivity study – Comparison of Profiler Plus® Kit heterozygote peak height ratio values by DNA input amount for all samples before (Test 2) and after (Test 3) 3130-to-3130xl instrument upgrade

Results of a 3130-to-3130xl Instrument Upgrade

56 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.4.2 Mixture Study

Mixed DNA profiles were evaluated and the number of alleles falling below 50 RFU was recorded before and after instrument upgrade.

Number of AllelesBelow Detection

Threshold

For a description of how the number of expected alleles was determined, see page 8.

Table 10 shows the number of alleles that fell below the detection threshold before and after upgrading the 3130 to a 3130xl:

• Before upgrade – Forty one (41) alleles (out of a total of 9851) fell below the detection threshold (Figure 44).

• After upgrade – Thirty four (34) alleles (out of a total of 8795) fell below the detection threshold (Figure 45).

The limit of detection of the minor contributor was approximately 9:1 before and after upgrading the 3130 to a 3130xl.

Table 10 Mixture study – Number of alleles below detection threshold – 3130-to-3130xl upgrade (data for 1:1 mixture not included)

Test 2 (before upgrade) Test 3 (after upgrade)

Instruments and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

Upgraded Instruments

and ID Numbers

Number of Alleles Below

Detection Threshold (50 RFU)

Total Number of

Alleles‡

3130-003 2 3257 3130xl-003 34 3413

3130-040 9 3264 3130xl-040 0 3590

3130-019 30 3330 3130xl-019 0 1792

Total 41 9851 Total 34 8795

‡. Data from failed injections not included

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 57

Figure 44 Mixture study – Identifiler® Kit electropherograms for a 3130 instrument before upgrade (Y-scale is constant in each of the upper panels)

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9FGA 26

Results of a 3130-to-3130xl Instrument Upgrade

58 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

Figure 45 Mixture study – Identifiler® Kit electropherograms generated on the same instrument (see Figure 44) after 3130-to-3130xl instrument upgrade

1:0

1:1

4:1

9:1

0:1

Mixtures9947A:9948(female:male)

9:1 mixture – Alleles from the minor contributor highlighted

TPOX 9FGA 26

GeneMapper® ID Software v3.2 Detection Threshold 50 RFU

7 Results and Discussion

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 59

7.4.3 Precision Study

The standard deviation in the size value for each allele of the allelic ladder was calculated within a run (four samples per run for the 3130 instrument, 16 samples per run for the 3130xl instrument) before and after upgrading each of the three instruments:

• All Identifiler® Kit and Profiler Plus® Kit runs before and after instrument upgrade yielded standard deviation values <0.15 bp (Figure 46).

• A small overall increase in the standard deviation was observed after upgrade in the Identifiler® Kit data (0.040 bp) and Profiler Plus® Kit data (0.023 bp) (data not shown).

Figure 46 Precision study – Sizing data for Identifiler® Kit and Profiler Plus® Kit allelic ladders before (Test 2) and after (Test 3) 3130-to-3130xl instrument upgrade

7.4.4 Resolution Study

The TH01 locus of the Identifiler® Kit allelic ladder was examined to evaluate the resolution and detection of alleles differing in size by 1 base pair. The 9.3 and 10 alleles of TH01 were resolved and individually detected for all injections before and after upgrading all instruments.

Results of a 3130-to-3130xl Instrument Upgrade

60 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

7.4.5 Concordance Study

Amplification of the 9947A, 9948, and 007 DNA samples produced correct genotypes for all loci for both the Identifiler® Kit and the Profiler Plus® Kit before and after upgrading all instruments.

8 ConclusionThe performance of the upgraded 3100-Avant™ and 3100 instruments to 3130 and 3130xl instruments respectively is consistent with the expected performance of the previously validated 3100-Avant™ and 3100 platforms.

9 ReferencesScientific Working Group on DNA Analysis Methods. Revised validation guidelines (approved July 2003). Available at: http//www.fbi.gov/hq/lab/fsc/backissu/july2004/standards/2004_03_standards02.htm.

DNA Advisory Board. Quality assurance standards for forensic DNA testing laboratories (approved October 1998). Forensic Science Communications 2000;2(3). Available at: http//www.fbi.gov/programs/lab/fsc/backissu/july2000/codispre.htm.

Applied Biosystems Quantifiler® Kit User’s Manual, PN 4344790, Rev. B, Applied Biosystems, 11/2003.

User Bulletin ABI PRISM® 3100/3100-Avant™ Genetic Analyzers Using Data Collection v2.0: Protocols for Processing AmpFlSTR® PCR Amplification Kit PCR Products, PN 4350218, Rev. A., 12/2003.

User Bulletin Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0, PN 4363787, Rev. A, 02/2005.

ABI PRISM® 3100 Genetic Analyzer User's Manual, PN 4315834, Rev. B., 01/2001.

9 References

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 61

Results of a 3130-to-3130xl Instrument Upgrade

62 Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin

9 References

Applied Biosystems 3130 Series Genetic Analyzers Performance Verification User Bulletin 63

www.appliedbiosystems.com

Worldwide Sales and Support

Applied Biosystems vast distribution and service network, composed of highly trained support and applications personnel, reaches 150 countries on six continents. For sales office locations and technical support, please call our local office or refer to our Web site at www.appliedbiosystems.com.

Applera is committed to providing the world’s leading technology and information for life scientists. Applera Corporation consists of the Applied Biosystems and Celera Genomics businesses.

Headquarters850 Lincoln Centre DriveFoster City, CA 94404 USAPhone: +1 650.638.5800Toll Free (In North America): +1 800.345.5224Fax: +1 650.638.5884

© Copyright 2006, Applied Biosystems. All rights reserved.

Information in this document is subject to change without notice.

NOTICE TO PURCHASER:

PLEASE REFER TO THE ABI PRISM® 3100 Genetic Analyzer USER'S MANUAL FOR LIMITED LABEL LICENSE OR DISCLAIMER INFORMATION.

TRADEMARKS:

AB (Design), Applied Biosystems, ABI PRISM, AmpFlSTR, GeneMapper, Identifiler, ProFiler Plus, and Quantifiler are registered trademarks of Applera Corporation or its subsidiaries in the U.S. and/or certain other countries.

Applera and Avant are trademarks of Applera Corporation or its subsidiaries in the U.S. and/or certain other countries.

All other trademarks are the sole property of their respective owners.

07/2006

Part Number 4376468 Rev. AStock Number 106UB49-01