![Page 1: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/1.jpg)
2D Digital Mammography- An update on vendor-recommended
QC tests
D. R. Jacobson, Ph.D. Asst. Professor of Radiology
Medical College of Wisconsin Milwaukee, WI 53226
![Page 2: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/2.jpg)
We will consider:
1. Benefit of screening mammography 2. Importance of good image quality in
mammography 3.Value of QC in mammography 4.Current QC program requirements
for FFDM 5.Potential changes for the future
![Page 3: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/3.jpg)
0
5
10
15
20
25
30
35
40
45
50
1930 1940 1950 1960 1970 1980 1990 2000 2009
American Cancer Society. Cancer Facts & Figures 2013. Atlanta: American Cancer Society; 2013.
Age-adjusted Cancer Death Rates for Females in the US, 1930-2009 deaths/100,000
Approximately 38 million mammography procedures were performed in 2013
![Page 4: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/4.jpg)
0
5
10
15
20
25
30
35
40
45
50
1930 1940 1950 1960 1970 1980 1990 2000 2009
American Cancer Society. Cancer Facts & Figures 2013. Atlanta: American Cancer Society; 2013.
Age-adjusted Cancer Death Rates for Females in the US, 1930-2009 deaths/100,000
1986 ACS establishes Breast Cancer Awareness Screening Program
1990 ACR Standards for the Performance of Screening Mammography
1992 MQSA
![Page 5: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/5.jpg)
Survival is very good if the cancer is found early
5-year survival rates (%):
98 if cancer is local
84 if regional
24 if metastasized
American Cancer Society. Cancer Facts & Figures 2013. Atlanta: American Cancer Society; 2013.
![Page 6: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/6.jpg)
Why do we do mammography? What is the purpose of mammography?
early and accurate detection of breast cancer
with the minimum required radiation dose
![Page 7: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/7.jpg)
What is mammography image quality? How is breast cancer detected in a mammogram?
![Page 8: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/8.jpg)
How is breast cancer detected in a mammogram? What is mammography image quality?
![Page 9: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/9.jpg)
What is the benefit of QC in mammography?
How does it assure good image quality?
How does it improve breast cancer detection?
![Page 10: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/10.jpg)
Quality control should not be thought of as an end in itself
![Page 11: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/11.jpg)
Pathology to be seen:
high density mass
shape and margin
are important
What is mammography image quality?
![Page 12: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/12.jpg)
Pathology to be seen:
micro-calcifications
morphology and distribution
are important
What is mammography image quality?
![Page 13: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/13.jpg)
Pathology to be seen:
architectural distortion
deviation from normal parenchymal pattern
What is mammography image quality?
![Page 14: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/14.jpg)
Tabar L, Tot T, Dean PB. Breast cancer: the art and science of early detection with mammography.
Stuttgart, Germany, Thieme Verlag, 2005
* 65% stellate, 35% circular
How is breast cancer detected in a mammogram?
64% stellate and circular malignant tumors* without calcifications
17% stellate and circular malignant tumors* with calcifications
19% calcifications only
5 year retrospective study of 866 cancers
![Page 15: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/15.jpg)
Tabar L, Tot T, Dean PB. Breast cancer: the art and science of early detection with mammography.
Stuttgart, Germany, Thieme Verlag, 2005
“…perception of the stellate lesion—while they are small [is]
the number one task of mammography” L. Tabar
How is breast cancer detected in a mammogram?
malignant stellate lesions 93%
benign stellate lesions 7%
![Page 16: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/16.jpg)
Mammography QC- a little history:
• 1985 NEXT (FDA) • 1986 Galkin et al
• 1986 ACS establishes Breast Cancer Awareness Screening Program ACS-ACR collaboration sets standard for quality mammography • 1987- ACR launches voluntary
accreditation program including mandatory QC testing
Wide variability in image
quality and radiation dose
in radiography
![Page 17: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/17.jpg)
A little more history ….
1987: ACR initiates voluntary accreditation 1990: 1st ACR QC manuals 10/27/92: President Bush signs MQSA bill to take effect 10/1/94 (21CFR Part 900) 1992: 2nd ACR manual published 10/23/93: Interim MQSA regulations published 10/28/97: Final MQSA regulations published to take effect 4/28/99 (a few effective 10/28/02) 1999: New ACR manual to work with MQSA
![Page 18: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/18.jpg)
1990
1992
1999
The ACR QC program
![Page 19: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/19.jpg)
• FDA certification and ACR accreditation
Mammography Quality Standards Act, 1992
Sec. 900.11
Requirements for certification.
(a) General. After October 1, 1994,
a certificate issued by FDA is required
for lawful operation of all
mammography facilities subject to
the provisions of this subpart. To
obtain a certificate from FDA, facilities
are required to meet the quality
standards in Sec. 900.12 and to be
accredited by an approved
accreditation body or other entity as
designated by FDA.
![Page 20: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/20.jpg)
21CRF 900.12(e)(6):
Quality Control tests — other modalities.
For systems with image receptor modalities other than screen-film, the quality assurance program shall be substantially the same as the quality assurance program recommended by the image receptor manufacturer, except that the maximum allowable dose shall not exceed the maximum allowable dose for screen-film systems in paragraph (e)(5)(vi) of this section.
![Page 21: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/21.jpg)
21CRF 900.12(e)(5)(vi):
Dosimetry.
The average glandular dose delivered during a single cranio-caudal view of an FDA-accepted phantom simulating a standard breast shall not exceed 3.0 milligray (mGy) (0.3 rad) per exposure. The dose shall be determined with technique factors and conditions used clinically for a standard breast. Standard breast is defined as 4.2 cm compressed breast consisting of 50% glandular and 50% adipose tissue
![Page 22: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/22.jpg)
Other modalities (as of August 2013) are: • Full field digital mammography (FFDM) • General Electric 2000D was first – 01/28/00 • Currently 26 different models are approved
• Digital Breast Tomosynthesis (DBT) • Hologic Selenia Dimensions DBT – 02/11/11
![Page 23: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/23.jpg)
Generic description of selected mammographic
QC tests
![Page 24: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/24.jpg)
Medical physicist QC tests for Screen-Film Systems: ACR and MQSA
![Page 25: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/25.jpg)
Unit Evaluation – flat and parallel compression
![Page 26: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/26.jpg)
Unit Evaluation – flat and parallel compression
Δ ≤ 1 cm
![Page 27: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/27.jpg)
Non-flat and non-parallel compression designs
![Page 28: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/28.jpg)
Light field – x-ray field – detector alignment
![Page 29: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/29.jpg)
light to x-ray: X and Y sum Δ ≤ 2% SID X-ray to detector: Δ ≤ 2% along any side
![Page 30: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/30.jpg)
Paddle at ~ 4.5 cm
Compression paddle to detector alignment
Paddle cannot extend beyond detector by > 1% SID
![Page 31: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/31.jpg)
![Page 32: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/32.jpg)
Focal Spot Performance (System Resolution)
![Page 33: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/33.jpg)
11 lp/mm - length
13 lp/mm - width
![Page 34: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/34.jpg)
AEC performance – thickness compensation – density control
Δ O.D. 2-6 cm ≤ +/- 0.15
![Page 35: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/35.jpg)
Screen uniformity and artifacts
Artifacts are not apparent or significant
![Page 36: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/36.jpg)
Table 1. Frequency of X-ray Unit, AWS, and Printer Tests.
Phantom image quality – MAPP – 4/3/3
![Page 37: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/37.jpg)
kVp accuracy and repeatability accuracy 5% CV ≤ 0.02
![Page 38: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/38.jpg)
Beam quality (HVL) HVL ≥ kVp/100
![Page 39: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/39.jpg)
Table 1. Frequency of X-ray Unit, AWS, and Printer Tests.
Entrance exposure, average glandular dose, radiation output rate
Air kerma and mAs CV ≤ 0.05 Output rate ≥ 7.0 mGy/sec (800 mR/sec ) Dg ≤ 3.0 mGy (300 mrad)
![Page 40: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/40.jpg)
Viewbox luminance and viewing conditions (optional under MQSA)
Viewbox luminance ≥ 3000 cd/m2
Luminance ≤ 50 lux
![Page 41: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/41.jpg)
Masking
Hot light
Masking and “hot light” are required (from equipment requirements)
![Page 42: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/42.jpg)
Currently approved FFDM systems
![Page 43: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/43.jpg)
FFDM and DBT Systems FDA approved, cleared, or accepted the following FFDM and DBT units for use as indicated by date:
Siemens Mammomat Inspiration Prime Edition cleared on 6/27/13
Konica Minolta Xpress Digital Mammography Computed Radiography (CR) System on 12/23/11
Agfa Computed Radiography (CR) Mammography System on 12/22/11
Fuji Aspire Computed Radiography for Mammography (CRM) System on 12/8/11
Giotto Image 3D-3DL Full-Field Digital Mammography (FFDM) System on 10/27/11
Fuji Aspire HD Full-Field Digital Mammography (FFDM) System on 9/1/11
GE Senographe Care Full-Field Digital Mammography (FFDM) System on 10/7/11
Planmed Nuance Excel Full-Field Digital Mammography (FFDM) System on 9/23/11
Planmed Nuance Full-Field Digital Mammography (FFDM) System on 9/23/11
Siemens Mammomat Inspiration Pure Full-Field Digital Mammography (FFDM) System on 8/16/11
Hologic Selenia Encore Full-Field Digital Mammography (FFDM) System on 6/15/11
Philips (Sectra) MicroDose L30 Full-Field Digital Mammography (FFDM) System on 4/28/11
Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System on 2/11/11
Siemens Mammomat Inspiration Full Field Digital Mammography (FFDM) System on 2/11/11
Carestream Directview Computed Radiography (CR) Mammography System on 11/3/10
Hologic Selenia Dimensions 2D Full Field Digital Mammography (FFDM) System on 2/11/09
Hologic Selenia S Full Field Digital Mammography (FFDM) System on 2/11/09
Siemens Mammomat Novation S Full Field Digital Mammography (FFDM) System on 2/11/09
Hologic Selenia Full Field Digital Mammography (FFDM) System with a Tungsten target in 11/2007
Fuji Computed Radiography Mammography Suite (FCRMS) on 07/10/06
GE Senographe Essential Full Field Digital Mammography (FFDM) System on 04/11/06
Siemens Mammomat Novation DR Full Field Digital Mammography (FFDM) System on 08/20/04
GE Senographe DS Full Field Digital Mammography (FFDM) System on 02/19/04
Lorad/Hologic Selenia Full Field Digital Mammography (FFDM) System on 10/2/02
Lorad Digital Breast Imager Full Field Digital Mammography (FFDM) System on 03/15/02
Fischer Imaging SenoScan Full Field Digital Mammography (FFDM) System on 09/25/01
GE Senographe 2000D Full Field Digital Mammography (FFDM) System on 01/28/00
Konica- Minolta Agfa
Giotto Fuji-DR Planmed Philips Carestream Fuji-CR Siemens Lorad Fischer GE
![Page 44: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/44.jpg)
Common QC tests for FFDM Mechanical/safety checks Acquisition monitor checks X-ray beam collimation and alignment (dead space) Compression paddle position, flat and parallel Spatial resolution (detector and system: GE) AEC: thickness response, exposure compensation Artifacts/uniformity (flat-field uniformity) Image quality: 4/3/3 or 5/4/4 kVp accuracy and repeatability HVL Radiation dose Radiation output rate SNR/CNR Reading workstation Film printer
![Page 45: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/45.jpg)
Common QC tests for FFDM – same as S/F Mechanical/safety checks Acquisition monitor checks X-ray beam collimation and alignment (dead space) Compression paddle position, flat and parallel Spatial resolution (detector and system: GE) AEC: thickness response, exposure compensation Artifacts/uniformity (flat-field uniformity) Image quality: 4/3/3 or 5/4/4 kVp accuracy and repeatability HVL Radiation dose Radiation output rate SNR/CNR Reading workstation Film printer
![Page 46: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/46.jpg)
Acquisition Monitor check - SMPTE pattern Acquisition Monitor IQ check - SMPTE pattern
![Page 47: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/47.jpg)
Flat field calibration and test
![Page 48: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/48.jpg)
Collimation testing without film is a challenge
![Page 49: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/49.jpg)
![Page 50: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/50.jpg)
video ~ 1 sec
take snap-
shot from video
![Page 51: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/51.jpg)
![Page 52: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/52.jpg)
Set to “true size”
![Page 53: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/53.jpg)
Table 1. Frequency of X-ray Unit, AWS, and Printer Tests.
Dead space
![Page 54: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/54.jpg)
Collimation – sliding paddle
![Page 55: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/55.jpg)
Table 1. Frequency of X-ray Unit, AWS, and Printer Tests.
System Spatial Resolution
![Page 56: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/56.jpg)
Image quality is still assessed with the MAP phantom
Fibers = 4, 5
Specks = 3, 4
Masses = 3, 4
Radiation dose must be < 300 mrad (3 mGy)
![Page 57: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/57.jpg)
ROI SNR and CNR
![Page 58: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/58.jpg)
Current Manufacturer-required
QC tests for FFDM
![Page 59: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/59.jpg)
GE Senographe 2000D
![Page 60: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/60.jpg)
GE Senographe 2000DS
![Page 61: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/61.jpg)
GE
Senographe Essential
![Page 62: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/62.jpg)
General Electric
![Page 63: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/63.jpg)
Contrast to Noise Ratio
• Contrast is difference between means of ROI 1 and 2.
• Noise is the SD of the ROI 2 (background)
• Contrast is measured between background and largest mass
ROI 2
ROI 1
GE
![Page 64: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/64.jpg)
MTF Measurement
0
0.2
0.4
0.6
0.8
1
0 2 4 6
Unit A
Unit B
MTF @ 2 lp/mm > 0.58
MTF @ 4 lp/mm > 0.25
![Page 65: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/65.jpg)
![Page 66: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/66.jpg)
Contact and mag. mode
Sub-system MTF
![Page 67: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/67.jpg)
Sub-System MTF Test
3.93 4.37
1.0
1.11
1.23
1.37
1.52
1.69
1.88
2.09
2.32
2.58
2.87
3.19
3.54
4.86
2.09
3.93
3
4
1
1
2
2
20
5
8
1
0
11
12
13
14
15
16
17
18
19
3 4
1
2
5
1
8 2
Measure Signal and Noise in ROIs
2.09 and 3.93 lp/mm 5 and 8 lp/mm
Contact Magnification
![Page 68: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/68.jpg)
Fisher Senoscan
![Page 69: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/69.jpg)
Fischer
![Page 70: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/70.jpg)
Hologic/Lorad Selenia
![Page 71: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/71.jpg)
Hologic Selenia Dimensions 2D (3D)
![Page 72: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/72.jpg)
Hologic/Lorad EE- MP performs RT tests
![Page 73: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/73.jpg)
SMPTE pattern
![Page 74: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/74.jpg)
Image quality is still assessed with the MAP phantom
Fibers = 5
Specks = 4
Masses = 4
![Page 75: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/75.jpg)
Siemens
Mammomat
Novation
![Page 76: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/76.jpg)
Siemens
Inspiration
![Page 77: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/77.jpg)
Siemens
![Page 78: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/78.jpg)
Siemens
![Page 79: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/79.jpg)
Contrast to noise ratio
ROI
• Contrast is difference
between means of
ROI 1 and 2.
• Noise is the SD of the
ROI 2 (background)
![Page 80: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/80.jpg)
Image quality is still assessed with the MAP phantom
Fibers = 5
Specks = 4
Masses = 4
![Page 81: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/81.jpg)
Table 1. Frequency of X-ray Unit, AWS, and Printer Tests.
Radiation dose must not exceed 3 mGy Radiation dose should not exceed 2 mGy
![Page 82: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/82.jpg)
Table 1. Frequency of X-ray Unit, AWS, and Printer Tests.
![Page 83: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/83.jpg)
Fuji ClearView
CRm
![Page 84: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/84.jpg)
Fuji CR
![Page 85: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/85.jpg)
At 20 mR, for 25 kVp and Mo/Mo: S= 120 +/- 20% (96-144) Expose – read should be 5-10 minutes
![Page 86: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/86.jpg)
![Page 87: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/87.jpg)
![Page 88: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/88.jpg)
![Page 89: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/89.jpg)
![Page 90: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/90.jpg)
Carestream DirectView
![Page 91: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/91.jpg)
Carestream ROI target = 1000 x log10(mR) + 1000
![Page 92: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/92.jpg)
p.93 Unrestricted Internal Use
© 2011, Carestream Health
QC Manual – Important Points
Calibration = 5 minutes TQT = 5 minutes QC testing = 5 minutes
![Page 93: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/93.jpg)
Sectra
MicroDose
Mammography
![Page 94: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/94.jpg)
Calibration phantom Daily QC phantom
![Page 95: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/95.jpg)
System uses scanning multi-slot geometry X-ray exposure is pulsed: 10-50 msec, 5-30 pulses/sec.
mAs ≈ 1% conventional mAs Assure that exposure and kVp meters will work with this system Place Pb aperture on paddle for HVL
![Page 96: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/96.jpg)
Philips (Sectra)
![Page 97: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/97.jpg)
![Page 98: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/98.jpg)
The average glandular dose (AGD) is calculated by AGD = ESAK ´ g ´ c´ s , where g and c are functions of the breast thickness and HVL and s is a function of breast or PMMA thickness. The g-, c- and s-tables can be interpolated. The entrance surface air kerma (ESAK) should be calculated at the entrance surface of the breast.
Table 11 g-factor as a function of breast thickness and HVL [10]. Table 12 c-factors for average breasts for women 50 to 64 [10]. Table 13 s-factors
Dg ≤ 1 mGy (100mrad)
![Page 99: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/99.jpg)
Contrast-to-noise ratio measured using the Daily QC Phantom
![Page 100: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/100.jpg)
![Page 101: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/101.jpg)
![Page 102: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/102.jpg)
Planmed
Nuance
![Page 103: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/103.jpg)
Planmed
![Page 104: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/104.jpg)
FUJI Aspire
![Page 105: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/105.jpg)
Fuji Aspire HD
![Page 106: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/106.jpg)
Gioto
Image
![Page 107: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/107.jpg)
Giotto
![Page 108: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/108.jpg)
![Page 109: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/109.jpg)
![Page 110: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/110.jpg)
Konica-Minolta
Xpress CR - Contact Mammography Upgrade
![Page 111: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/111.jpg)
Konica Minolta
![Page 112: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/112.jpg)
QC PHANTOMS AND AUTOMATED ANALYSIS PROGRAMS: GE flat field test AOP and SNR IQST (Image Quality Signature Test) Phantom- MTF and CNR
![Page 113: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/113.jpg)
![Page 114: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/114.jpg)
Flat Field Tests AWS Screen Results
Action Level
</= 10%
</= 0.8%
<100
None
</= 40%
Left Chest
Wall
Left
Anterior
Right Chest
Wall
Right
Anterior
18x24 25 4.5 4.9 4.5 4.8 4.5 Pass
1. Flat Field and Image Quality Checks (Weekly technologist test.)
a. Flat Field: 2.5 cm thick uniform PMMA;26 kVp; 200mAs; Mo/Mo. Phantom evaluated on AWS.
Results Status Objects Observed Status
1.82 Pass Fibers 5.5 Pass
0.49 Pass Specks 4 Pass
0 Pass Masses 4 Pass
0 Pass
23.03 Pass
b. Phantom Image Quality (Print, AWS and RWS).
RMI Model156 Phantom: 26 kVp;125 mAs, Mo/Mo.
Print AWS RWS(1) RWS(2)
4.5 5 5 5
4 4 4 4
3.5 4 4 4
RWS(1) and RWS (2) are located in the second floor mammography reading room.
Facility Survey Date
Unit or Room # Report DateSenographe 2000D 2/14/2001
Waukesha Mem Hosp-NewBerlin
MEDICAL PHYSICIST'S FULL FIELD DIGITAL MAMMOGRAPY SURVEY REPORT
SNR Uniformity < 40%
Number of fibers; = or > 4.
Number of speck groups; = or > 3.
Number of masses; = or > 3.
2/7/2001
Force lbs
Breast
Thickness
Indicated
Bad ROI = 0
Test
High Freq Modulation < 0.8
Bad Pixel < 100
Brightness Uniformity < 10%
StatusField Size
![Page 115: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/115.jpg)
![Page 116: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/116.jpg)
![Page 117: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/117.jpg)
![Page 118: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/118.jpg)
DS IQ Signature Test
automates SNR and MTF
![Page 119: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/119.jpg)
![Page 120: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/120.jpg)
QC PHANTOMS AND AUTOMATED ANALYSIS PROGRAMS: Hologic Auto SNR and CNR
![Page 121: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/121.jpg)
Contrast to Noise Ratio
![Page 122: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/122.jpg)
![Page 123: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/123.jpg)
QC PHANTOMS AND AUTOMATED ANALYSIS PROGRAMS: Fuji CR Auto SNR and CNR
![Page 124: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/124.jpg)
![Page 125: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/125.jpg)
2 cm
6 cm
ROI placement for automated CNR and SNR calculation
![Page 126: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/126.jpg)
QC calculation tool - offline
page 237
![Page 127: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/127.jpg)
QC PHANTOMS AND AUTOMATED ANALYSIS PROGRAMS: Fuji DR Fuji One Shot phantom and auto QC program
![Page 128: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/128.jpg)
Fuji 1 Shot Phantom M Plus weekly constancy test tool
![Page 129: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/129.jpg)
![Page 130: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/130.jpg)
![Page 131: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/131.jpg)
![Page 132: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/132.jpg)
![Page 133: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/133.jpg)
![Page 134: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/134.jpg)
QC PHANTOMS AND AUTOMATED ANALYSIS PROGRAMS: Carestream CR Mammo TQT
![Page 135: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/135.jpg)
![Page 136: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/136.jpg)
![Page 137: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/137.jpg)
![Page 138: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/138.jpg)
Potential future changes • ACR FFDM QC program • Universal phantom • Alternative standard (MQSA)
![Page 139: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/139.jpg)
19 cm
13 cm
2 cm x 0.1 cm
31 cm
7 cm
6 fibers 6 speck groups 6 masses
New ACR accreditation phantom (proposed)
![Page 140: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/140.jpg)
Who determines the QC requirements for a new mammography modality? 1. FDA 2. ACR 3. Imaging system manufacturer 4. On-site physicist 5. Image receptor manufacturer
21 C.F.R. 900.12(e)(6)
![Page 141: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/141.jpg)
Which of the following are new modalities- requiring 8 hours of initial training before use? 1. Screen/film and stereo mammography 2. Full-field digital mammography and
digital breast tomosynthesis 3. Stereotactic biopsy 4. Digital stereotactic biopsy 5. Automated breast ultrasound
www.fda.gov/Radiation-EmittingProducts/MammographyQualityStandardsActandProgram /Guidance/PolicyGuidanceHelpSystem/ucm136925.htm www.fda.gov/Radiation-EmittingProducts/MammographyQualityStandardsActandProgram /FacilityCertificationandInspection/ucm243765.htm
![Page 142: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/142.jpg)
The 5 year survival for a minimal breast cancer is 1. Less then 50% 2. 60-70% 3. 70-80% 4. 80-90% 5. Greater than 90%
American Cancer Society. Cancer Facts & Figures 2013. Atlanta: American Cancer Society; 2013
![Page 143: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/143.jpg)
The most reliable indication of breast cancer in a mammogram is 1. Architectural distortion 2. Clustered calcifications 3. Linear and branching calcifications 4. Spiculated mass 5. Skin thickening
Tabar L, Tot T, Dean PB. Breast cancer: the art and science of early detection with mammography. Stuttgart, Germany, Thieme Verlag, 2005
![Page 144: 2D Digital Mammography- - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22637-311436-101907.pdf · 2D Digital Mammography- An update on vendor-recommended QC tests D. R. Jacobson, Ph.D.](https://reader035.fdocuments.in/reader035/viewer/2022070705/5e963aceb996893b5f585b31/html5/thumbnails/144.jpg)
What performance parameter(s) does MQSA specify for FFDM quality assurance? 1. Phantom dose < 3.0 mGy for one CC view 2. Phantom dose < 2.0 mGy for one CC view Image 3. Image quality must exceed S/F IQ 4. Dose can’t exceed max. allowed S/F dose 5. Patient dose ≤ 3.0 mGy for one CC view
21 C.F.R. 900.12(e)(6); 21 C.F.R. 900.12(e)(5)(vi)