1 Next-Generation Performance and More Capabilities corbett LIFE SCIENCE Introducing the.
-
Upload
robyn-carr -
Category
Documents
-
view
216 -
download
2
Transcript of 1 Next-Generation Performance and More Capabilities corbett LIFE SCIENCE Introducing the.
1
Next-Generation Performance and More Capabilities
www.corbettlifescience.com
corbett
LIFE SCIENCE
Introducing the
2
What is the Rotor-Gene 6000?An integrated device that:
Maintains identical well-to-well conditions to monitor micro reaction tubes
Illuminates and collects a wide range of optical signals
Enables exquisite control of thermal conditions
Provides an open platform for all chemistries
Features a very fast data acquisition rate
Capable of the broadest application set:
Real-time analysis (e.g. quantitative amplification)
End-point analysis (e.g. SNP genotyping)
SYBR melt analysis
HRM (high resolution melt) analysis
Concentration analysis
+ future applications…
3
How does the rotary format work?
Samples spin continually during a run 400 RPM (heating or cooling)
High-speed data collection all samples read in one revolution (0.15 sec)
G-force keeps reagent at base of tube removes bubbles & condensation
will not pellet components
Continuous movement means no variation
well-to-well thermal or optical
400 RPM
>10 G
4
Why rotary?
All wells are iso-thermal
All wells are iso-illuminated
Super fast data acquisition rate
Minimal maintenance and calibration
Simple and automated verification testing
Enables the broadest application set
Many advantages over other designs including:
5
3rd Generation Rotary DesignThe Rotor-Gene 6000 is based on proven technology
Rotor-Gene 20002000
Rotor-Gene 60002006Rotor-Gene 3000
2003
6
7
Optics
8
Cross-section of rotary optics
Reaction Chamber
PMT DetectorAssembly
LED Light Source
Assembly
Tubes in Rotor Spin Past Optics
Lens
Detection Filters
Spindle/Motor Assembly
9
Rotary optics 3D animation
LED light source(rotates for each channel)
rotor spins tubes at 400 rpm
lens
filter set(rotates for
each channel)
sensitive PMT (photomultiplier)
detector
10
Rotary Design means:low maintenance & maximum convenience
No optical normalization needed (thus no special reagents either)
No passive reference dye needed (e.g. ROX™)
No lamp replacement or maintenance (lifetime guarantee on lightsource)
No block to clean (because tubes hang in mid air)
No bubbles in reaction (automatically removed by centrifugal rotor)
Tube caps can be labeled (not possible with 96-well systems)
Supports several tube formats (by simply swapping rotors)
11
Light-Emitting Diode (LED) Light Source
LEDs last >100,000 hrs
Low-power, low-temperature device
LEDs have a focussed light-emitting area
Separate LED used for each channel from infra-red (IR) to ultra-violet (UV)
Estimated lifespan ~40 years (assuming 4 runs every working day)
Rotor-Gene LEDs are guaranteed for the life of the instrument!
12
5 Models to suit different needs and budgets:
1. 2-Plex Green/Yellow
2. 2-Plex HRM Green/Yellow + High Resolution Melt channel
3. 5-Plex Green/Yellow/Orange/Red/Crimson
4. 5-Plex HRM Green/Yellow/Orange/Red/Crimson + High Resolution Melt channel
5. 6-Plex Blue/Green/Yellow/Orange/Red/Crimson
Channel Excite/Detect (nm) Example fluorophores detected
Blue 365/460 BiosearchBlue™, Marina Blue®, Bothell Blue®, Alexa Fluor® 350
Green 470/510 FAM™, SYBR® Green 1, Fluorescein, EvaGreen™, Alexa Fluor® 488
Yellow 530/555 JOE™, VIC™, HEX™, TET™, Yakima Yellow®, Cal Flour® Gold 540
Orange 585/610 ROX™, Cy3.5®, Texas Red®, Alexa Fluor® 568, CAL Fluor™ Red 610
Red 625/660 Cy5®, Quasar 670™, LightCycler Red 640®, Alexa Fluor™ 633
Crimson 680/712 LP Quasar705™, LightCycler Red 705®, Alexa Fluor® 680
HRM 460/510 SYTO®9, LC Green®, LC Green™Plus+, EvaGreen™
UV
IR
NOTE: ROX™ normalization is not needed so all channels can be “plexed” for separate reactions
13
Thermal Control
14
Thermal Precision (uniformity) in a Block-based Cycler
± 0.50 ºC (or more) across the 96 well block
(thus > 1.00 ºC variation is typical)
Corner and Edge wells most affected
NOTE: Fluorescence 1/Temp
Localized “hotspots” as Peltier device junctions begin to fail
The Rotor-Gene does not use Peltier devices—because they fail unpredictably and are expensive to repair
NOTERotor-Gene 6000 specifications: Uniformity: ±0.01°C, Resolution: ±0.02°C
15
Centrifugal fan drives air around chamber
Chamber vent seals to contain air
Heating mechanism
Note: holes in the rotor allow
free airflow
Heater elements switch on
16
Cool air in
Centrifugal fan Drives air into chamber
Centrifugal fan drives air around chamber
Chamber vent opens expelling hot air
Cooling mechanism
Heater elements switch off
Note: holes in the rotor allow
free airflow
17
TIME
Equilibration time in Rotor-Gene is 0 sec (for ± 0.01 °C)
Equilibration time in 96 well block is 15 sec (for ± 0.5 °C)
Up to 50% faster run times with better uniformity between samples
96 ºC
Thermal Equilibration: Rotor vs. Block
72 ºC
60 ºC
72.5 Degrees
71.5 Degrees
72 Degrees +/- 0.01
18
Thermal Accuracy
Instrument Calibration andPerformance Verification
19
For Checking and CalibratingThermal Accuracy and Optical Performance
Increasingly, laboratories require routine verification and validation of instrument performance and thermal accuracy
The OTV Kit automates routine verification testing on the Rotor-Gene
The OTV system comprises an OTV Disc, optical insert accessories and a CD
With the kit, verification of instrumentperformance and thermal accuracy can bedone at up to 30 times within the 6 monthexpiry date of the rotor consumable
Automated verification testing in thismanner is unique to the Rotor-Gene
20
OTV Mechanism
Rotor-Gene 3000 Scatter Plate Insert
OTV disc in rotor
The OTV system* uses the chemicalproperties of three different thermochromicliquid crystals (TLCs) as an absolutetemperature reference.
When heated, a TLC changes from opaqueto transparent at a very precise temperature.But because TLCs do not fluoresce, a fluorescencescatter plate is inserted over the optics to enable detection by the Rotor-Gene.
The Rotor-Gene measures the precise temperature transition of each TLC. This reported value is compared to the known calibrated value to verify the instrument is within specification. If not within specification, automatic re-calibration of the Rotor-Gene can be done at the press of a button.
*patent pending
Rotor-Gene 6000 Scatter Plate Insert
21
The HTML report file can be saved, printed, e-mailed or exported to MS Word as a record of the verification test
OTV Report
The report indicates the instrument was within specification when “No Adjustment Required” isstated (as shown here)
Detailed analysis data is also reported
22
Software
More analysis options
Raw data exportFor 3rd party software analysis
Unlimited user software licenseNo additional license fees
All users can copy & run the softwareto analyze files remotely
Upgrades free (by web download)
Standard Curve Quantitation2 Standard Curves Relative QuantitationDelta Delta Ct Relative QuantitationComparative QuantitationRelative Expression Software Tool (REST)LinReg (Assumption-Free Analysis)Melt AnalysisHigh Resolution Melt AnalysisEnd-Point AnalysisAllelic DiscriminationScatter Graph AnalysisConcentration Analysis
Standard Curve Quantitation2 Standard Curves Relative QuantitationDelta Delta Ct Relative QuantitationComparative QuantitationRelative Expression Software Tool (REST)LinReg (Assumption-Free Analysis)Melt AnalysisHigh Resolution Melt AnalysisEnd-Point AnalysisAllelic DiscriminationScatter Graph AnalysisConcentration Analysis
Relative Expression Software Tool (REST)
23
Plug-and-Play Portability
Robust design suits transportability
No optical alignment or calibration needed
Self-configuring USB connection to computer
Easy to carry
Small: 370 mm (14.6") W, 420 mm (16.5") D, 275 mm (10.8") H
Light: 14 kg (31 lbs)
24
Tube Formatsand
Sample Handling
25
Tube Formats
36 0.2 mL PCR tubes Attached flat or domed capsNOTE: optical caps are not required since detection is through the base of the tube
1.
2. 72 0.1 mL tubes- allow small reaction volumes (5–10 µL)- in strips of 4 for ease of use- Frosted cap extensions allow write-on labelling + easy handling
26
Gene-Disc™ Plates
3. Gene-Disc™ 72- 0.1 mL tubes in a rotary “plate” design - Tubes oriented vertically (not angled)- Manual or automated loading- Heat-sealed in seconds
4. Gene-Disc™ 100- For 96 sample workflow + 4 extra controls-30 µL wells in a rotary “plate” design - Manual or automated loading- Wells oriented vertically (not angled)- Heat-sealed in seconds
27
Gene-Disc rotor ready for cycling
The new heat sealer provides a permanent or removable film seal(user selectable switch)
Direct Robotic Setup in Gene-Disc™ Plates
Vertical tube orientation means a robot can set up all reactions directly into Gene-Disc tubes
28
Amplification Performance
29
Replicates (full 72-well rotor)
No ROX normalization
With optional ROX normalization
30
2-fold quantitative discrimination
BCL-2 human gene target (68 bp amplicon) amplified from total genomic DNA template. Semi-log amplification plots shown of normalized fluorescence vs. cycle number with no smoothing applied and without ROX™ normalization. Primer concentration 300 nM, dual-labeled probe 60 nM, 40 cycle amplification completed in 46 min using standard Platimum® qPCR SuperMix-UDG commercial master mix (Invitrogen Corp., Carlsbad, CA).
256,000 copies
500 copies
2-fold discrimination (=1 PCR cycle)
10 separate dilutions in triplicate
No ROX normalization
No data “smoothing”
single-copy gene amplified from whole human genomic template
Fast cycling (40 cycles, 46 min)
Standard commercial master-mix
Low probe (60 nM = ¼ dilution)
31
Concentration Measurement
32
DNA Concentration Measurement
The Rotor-Gene is fully equipped to do DNA concentration measurement using fluorescent dyes
Using a standard run protocol and integrated analysis software, the concentration of unknown samples is determined from a standard curve.
A DNA standard curve with replicates is shown. Curve interpolated using a spline curve fit (Rotor-Gene analysis software). Data was obtained using reagents in the Quant-iT™ PicoGreen® dsDNA Kit (Invitrogen Corp., Carlsbad CA). Standard Rotor-Gene concentration analysis run protocol was used. 10 µL PicoGreen®
(diluted1/200 in 1 TE buffer) was combined with 10 µL of each standard (diluted in 1 TE buffer). Final volume 20 µL.
Concentration pg/µL
Flu
ore
sce
nce
1000
900
800
700
600
500
400
300
200
100
No DNA controls
33
SYBR Melt Analysis and
HRM™ (high resolution melt)
a new application fora new type of instrument
34
Generic dsDNA intercalation dye
Inexpensive & simple
Used for real-time PCR product detection
Used for DNA dissociation (melt) analysis
Widely used
SYBR™ Green I
35
SYBR melt analysis of a DNA fragment
Raw data plot
Fluorescence drops as DNA melts and SYBR is released
Derivative data plot
This “rate” curve peaks at maximum dissociation rate
which is indicative of the Tm (temperature of melting)
36
SYBR™ melts can reflect product size
250 bp fragment
500 bp fragment
Raw data plot:fluorescence vs. temp.
Derivative data plot:dF/dT vs. temp
500 bp fragment
250 bp fragment
37
SYBR™ Green I melts can reflect sequencedetection of alleles
Low primer conc (50 nM)
Single band contains two species (alleles)
High primer conc (900 nM)
Primer-dimer appears as a third species
Allele A
Allele B
Primer Dimer
Allele A
Allele B
38
SYBR™ Green I is toxic to PCR,so concentration used is very low
“Saturation” Hypothesis
SYBR® Green I
Some dye can relocateas melting begins
New dye technology for HRMExamples: EvaGreen, SYTO9
LC Green™ IDye saturation leaves no room for
relocation events during melting
“Saturation” dyes are much less toxic, so concentration usedcan be higher
This may reduce dye relocation events and improve HRM results
Theoretically, unsaturated binding may allow dye relocation during melts, making it less suitable for HRM
39
HRM on the Rotor-Gene 6000
To support HRM an instrument requires:high-intensity + high sensitivity optics
high-speed data capture
very precise temperature control
extreme temperature resolution
To support multiple wells:Superlative thermal and optical well-to-well uniformity
Mutants(T allele)
Wild types(C allele)
Heterozygotes
Mutants(T allele)
Wild types(C allele)
Heterozygotes
Temperature (°C)
Flu
ore
sce
nce
82 83 84 85 86 87 88
•Example SNP genotyping using HRM analysis. ACTN3 (R577X) SNP genotypes (C—T).
•Ten replicates each genotype are shown.
•Fragment pre-amplified using a 40 cycle fast protocol (46 min).
40
HRM workflow on the Rotor-Gene 6000
Autocall genotypesUp to 100 at a
time
Choose preferred tube
0.1 mL tubes
Gene-Disc™ 72
0.2 mL tubes
Gene-Disc™ 100
Run PCR and HRM
41
HRM Applications
Some HRM applications currently under investigation include:
Mutation discovery
Screening for loss of heterozygosity
DNA fingerprinting
SNP genotyping
Characterization of haplotype blocks
DNA methylation analysis
DNA mapping
Species identification
Somatic acquired mutation ratios
HLA compatibility testing
Association (case/control) studies
Alleleic prevalence in a population
Identification of candidate predisposition genes
42
Automation
43
Precision Robotic Pipetting Improves Results
CAS-1200™
Precision Liquid Handling System
Rotor-Gene 6000™
Six Channel Multiplexing System
44
Rotor-Gene™ results
10 µL reaction volume, 18 replicates
Hand pipettingCT std dev 0.12
CAS-1200 robotCT std dev 0.10
45
Rotor-Gene™ results
5 µL reaction volume, 18 replicates
Hand pipettingCT std dev 0.64
CAS-1200 robotCT std dev 0.12
46
Robotic Workflow for Real-Time Analysis
“Extraction-to-Reaction”
47
Summary
Real-Time analysis is an exact science with many variables
Through good design, the rotary format best minimizes these variables
Precise and reproducible data is easier to achieve with the Rotor-Gene
Future applications such as HRM and concentration measurement can only be achieved on the Rotor-Gene, underlining it’s superior design
Choose from more tube formats and data analysis options
Verification testing and calibration is automated by the OTV system
The robust design delivers minimum maintenance, lowest operating costs and maximum convenience
48
colors
49
Sydney AustraliaCorbett Research Pty Ltd14 Hilly StreetMortlake, NSW 2137T 1 800 803 915 (Toll free)T +61 2 9736 1320F +61 2 9736 1364
United KingdomCorbett Research UK LimitedUnit 296 Cambridge Science ParkMilton Road, Cambridge CB4 0WDT +44 (0)1223 424 288F +44 (0)1223 424 144
All slides 2006 Corbett Life Science. All rights reserved
E-mail [email protected]
USACorbett Robotics Inc185 Berry Street, Suite 5200San Francisco, CA 94107T +1 866 380 1166 (Toll free)T +1 415 348 1166F +1 415 348 1177
Brisbane AustraliaCorbett Robotics Pty Ltd42 McKechnie DriveEight Mile Plains, QLD 4113T +61 7 3841 7077F +61 7 3841 6077
Web www.corbettlifescience.com
Offices