1
DICOM WG-02 DICOM WG-02 Advances in X-Ray Angiography Advances in X-Ray Angiography
Projection Imaging and 3DProjection Imaging and 3D
SPIE Medical Imaging2010, San Diego
Authors: Heinz Blendinger Siemens Healthcare Bas Revet Philips Healthcare Francisco Sureda GE Healthcare (Chair DICOM WG-02)
2
IntroductionIntroductionOverview of X-Ray Angiography in DICOM
X-Ray N-Dimensional Presentation X-Ray N-Dimensional Presentation CasesCases
3D Shutter
Volume Subtraction
Stabilized Region in Moving Volume
Trajectory Tracking in Volume
2D-3D Blending
ConclusionConclusion
Presentation OutlinePresentation Outline
3
Overview of X-Ray Angiography in Overview of X-Ray Angiography in DICOMDICOM
2D Projection Images2D Projection Images
X-Ray AcquisitionX-Ray Acquisition
3D Reconstruction3D Reconstruction
Approved in the
Standard
Work in Progress
Supp 94:Radiation Dose
Reporting
Supp 83:Enhanced XA/XRF
Supp 116:X-Ray 3D Storage
Supp 139-LB: Enhanced XA
Informative Annex
N-Dimensional Presentation State
Follow-up of IHE REM Profile
CR-DX Dose Reporting
Supp 140: Presentation State
X-Ray 3D Informative Annex
4
Workflow in X-Ray N-Dimensional Workflow in X-Ray N-Dimensional PresentationPresentation
X-Ray 3D X-Ray 3D Reconstruction Reconstruction
SystemSystem
X-Ray Calibratio
n Procedure
Calibration Data
Proprietary
X-Ray X-Ray Acquisition Acquisition
SystemSystem
3D Visualization 3D Visualization SystemsSystems
X-Ray Acquisitio
n Procedure
Enhanced XA
Storage SOP Class
X-Ray 3D Storage
SOP Class
Reconstruction
Procedure
N-D Presentation
State SOP Class
In progress
Visualization Visualization
5
X-Ray 3D Angiography X-Ray 3D Angiography – Rotational – Rotational AcquisitionAcquisition
Frame #5:X-ray settings 5
Geometry settings 5
Frame #2:X-ray settings 2
Geometry settings 2
Frame #3:X-ray settings 3
Geometry settings 3
Frame #4:X-ray settings 4
Geometry settings 4
Frame #1:X-ray settings 1
Geometry settings 1
Optimized 3D Reconstruction
6
X-Ray 3D Angiography X-Ray 3D Angiography – Presentation – Presentation StateState
Needs for 3D Angiography Presentation– Presentation features common to all 3D modalities (rendering...)– Speficic presentation of X-Ray 3D Angiography:
• Acquisition 3D shutter (e.g. collimation)• Volume Subtraction and voxel shift• Stabilized point in all volumes (e.g. cardiac wall motion, stent stabilized)• Catheter tracking trajectory in one volume• 2D-3D blending presentation (3D conic projection on 2D fluoroscopy)
N-Dimensional Presentation State• Work Item 2008-04-C. Addresses needs of multi-modalities• Led by Working Group 11, participation of Web3D Consortium and other DICOM working groups• Supplement in progress...
7
X-Ray 3D Angiography X-Ray 3D Angiography – Presentation – Presentation StateStateVolume Regions to define 3D shutter
Punch Hole: EV: “ELLIPSE“ or “POLYGON“Plane: Orientation, Position, inside or outsideFrame: EV: “ELLIPSE“ or “POLYGON“ or “BITMAP“
Proposal for Volume Region Sequence:• Region Purpose : DT: “DISPLAY SHUTTER“ or “ROI“• Region Display Mode : DT: “DISPLAY“ or “HIDE“• Region Definition Method : Enumerated Values• ...
Punch Holes Clipping Planes Frames
8
X-Ray 3D Angiography X-Ray 3D Angiography – Presentation – Presentation StateStateVolume Subtraction
Proposal for 3D Mask Subtraction Sequence:
• Mask Operation: DT: “Volume Selection“• Applicable Volume: Pair of Frame Numbers (e.g. N+1, N+k)• Mask Volume: Pair of Frame Numbers (e.g. 1, N)• Mask Voxel Shift: Triple of Float Numbers• Mask Contrast Registration 4x3 matrix• Mask Operation Explanation: Free Text• Mask Selection Mode: DT: SYSTEM or USER• Recommended Viewing Mode: DT: SUB or NAT
Mask Frames Contrast Frames
SubtractedContrast
9
X-Ray 3D Angiography X-Ray 3D Angiography – Presentation – Presentation StateStateStabilized region in moving volume
Applicable to 4D data (i.e. 3D volumes with time)Goal: Stabilize the position and orientation of an object during the dynamic view
The result is an animated sequence to view the volumes stabilized at a given point:
• a) with same volume orientation• b) or changing the orientation of the volume to keep the same orientation of the object of interest
Proposal for Stabilized Region Sequence:// contains as many items as number of volumes
• Applicable Volume: Pair of Frame Numbers• Point: 3D coordinates• View Angle: 3D view angle
10
X-Ray 3D Angiography X-Ray 3D Angiography – Presentation – Presentation StateStateTrajectory Tracking in Volume
During the planning phase of a catheterization intervention
• Define a 3D curve from point A to point B inside an artery (i.e. catheter trajectory)
• Define an animated sequence to view the progress of the curve:
Applicable to 3D dataGoal: View the volume along a pre-defined trajectory of the camera
Example:
a) from the outside of the volume, with same volume orientation
b) from the outside of the volume, changing the orientation of the volume to be perpendicular to the tip of the curve
c) from inside the artery (fly-thru) as virtual endoscopy
AA
BB
11
2D-3D Blending in X-Ray Angiography 2D-3D Blending in X-Ray Angiography (1/2)(1/2)
X-Ray X-Ray AcquisitioAcquisition Systemn System
X-Ray Rotational Acquisition X-Ray 2D
Projection SOP Class
X-Ray 3D Storage
SOP Class
X-Ray 3D Reconstruction
3D Segmentation
Segmentation SOP
Class
X-Ray Acquisitio
n
2D-3D BlendingX-Ray 2D Projectio
n SOP Class
3D View 3D Viewing Settings
Conic Projection2D Presentation3D Presentation
new
2D Presentati
on SOP Class
Same Frame Of Reference
2D View
12
2D-3D Blending in X-Ray Angiography 2D-3D Blending in X-Ray Angiography (2/2)(2/2)
X-Ray X-Ray AcquisitioAcquisition Systemn System
CT Storage SOP Class 3D
Segmentation
Segmentation SOP
Class
X-Ray Acquisitio
n
2D-3D BlendingX-Ray 2D Projectio
n SOP Class
3D View 3D Viewing Settings
Conic Projection2D Presentation3D Presentation
new
2D Presentati
on SOP Class
2D View
3D Image (CT, …)
Registration to Isocenter
System
new
13
ConclusionConclusionEnhanced XA (2D)Enhanced XA (2D)
Supplement 139Supplement 139 in Letter Ballot. Informative (DICOM Part 17)
Will facilitate the adoption of the Enhanced XA SOP Class (Sup 83)
XA Presentation State (2D)XA Presentation State (2D) Supplement 140Supplement 140 in Final Text
X-Ray 3D AngiographyX-Ray 3D Angiography New IOD approved in Standard 2007 (Supplement 116Supplement 116)
Application cases (Informative) - Work In Progress
N-D Presentation StateN-D Presentation State N-D Presentation State - Work In Progress