Southeast User Group Meeting
March 5, 2013
DSM – Will Gilliland
FSR – Roger Herzog
Sysmex Service
Tips and Tricks
2/20/2013
2
Session Goal
• To provide an overview of SNCS features
• To provide the Sysmex Hematology customer with information and troubleshooting tips for some morecommon issues with table top instruments
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
Session Objectives
At the end of this session the operator will:• Have a better understanding of the normal operation
of the X-series analyzers.
• Understand some of the causes for common error messages.
• Be able to troubleshoot and resolve some of the common errors.
• Be familiar with SNCS functionality
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
3
SNCS
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
Windows Desktop
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
4
Launch SNCS Status Tool
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
SNCS Status Tool
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
5
Launch SNCS Remote Session
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
Remote Session Password Window
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
6
XE-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
X-Series
• Overall Uptime for X-Series is >95%
• In the event you have an issue, here are common solutions that you can try before calling TAC.
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
7
XE-Series
• WBC or NRBC QC Shift– Possible Cause: Insufficient reagent/sample
delivered to reaction chamber due to partial hydraulic obstruction
– Possible Solution: Clean SRV and while SRV is disassembled flush tubing from reaction chamber to SRV and verify fluid stream is clean.
– Flush tubing and port above reaction chamber into reaction chamber
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-Series
• WBC or NRBC QC shift
• Reaction Chamber– Barb Needs a Diamond Ring
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
8
XE-Series
• WBC or NRBC QC shift
• SRV Structure
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-Series
• NRBC, DIFF or RET QC shift– Possible Cause: Insufficient draining of
reaction chambers thus next sample being diluted out
– Possible Solution: Clean partially occluded t-fittings below master valves 37-40
– Dye port clogged
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
9
XE-Series
• NRBC, DIFF or RET QC shift
• T-Fitting location
To waste chamber 1
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-2100 Example
XE-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
10
XE-5000 Example
XE-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-Series
• IMI Count too long/short– Possible Cause: Plugged aperture or
insufficient vacuum
– Possible Solution: Brush aperture as described in the operator’s manual or flush vacuum system from bellow unit to pneumatic unit vacuum trap
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
11
XE-Series
• Flushing vacuum system– Remove clear plastic cover over the copper
colored bellows unit where vacuum adjustments are made without removing the lower white plastic base.
– Squirt approximately 5-10 ml of DI H2O into hole on top of copper colored bellows unit followed by 5-10 ml of methanol.
– Re-assemble bellows unit and adjust vacuum.
– Drain pneumatic unit vacuum trap.
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-Series
• Flushing Vacuum System
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
12
XE-Series
• IMI Bkgd errors and/or excessive false positive PLT Clumps? Flags
• Possible causes: Air bubbles from poor quick connect connection or possible contamination of reagent or analyzer SIM hydraulics
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-Series
• IMI Bkgd errors and/or excessive false positive PLT Clumps? Flags• Possible solutions:
• Re-seat SIM quick connect connection to eliminate air bubbles
• Perform SIM reagent extended reagent replenish sequence to replace reagent in reagent chamber with reagent from cubitainer, may need to be done multiple times.
• If issue persists, the cubitainer may be the problem.
• If fresh reagent doesn’t resolve issue, analyzer SIM hydraulics may need decontaminated.
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
13
XE-Series
• SIM Reagent Cubitainer Quick Connect
• Customer Bulletin 61-1080 on Customer Resource Center
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-Series
• SIM Quick Connect
• O-ring importance
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
14
XE-Series
• SIM Quick Connect
• Cleanliness
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XE-Series
• SIM Quick Connect
• Handling
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
15
XE-Series
IMI Transducer DetailCopyright © 2013 by Sysmex America, Inc. All
rights reserved..
XE-Series
• MCV shift or trend high or low• Possible Cause: 0.03 Mpa pressure mis-adjusted
• Possible solution: Adjust 0.03 Mpa pressure– 0.03 Mpa pressure range is 0.270-0.0330 Mpa
– Increasing the pressure will decrease the MCV
– Decreasing the pressure will increase the MCV
– Affect is minimal but may help, approximately 0.5-1.5 flwithin that pressure range.
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
16
XE-Series
• WC-2 not draining errors:• Possible Cause: MV 43 partially occluded.
• Possible Solution: Through Maintenance program drain waste chambers 3 times. Best done if side panel is open and chamber can be seen– Fill chamber with Clorox through open port on top
of waste chamber 2.
– Perform another waste chamber drain sequence to get bleach in and through MV 43.
– Let soak awhile and perform drain waste sequence again.Copyright © 2013 by Sysmex America, Inc. All
rights reserved..
XE-Series
• Waste Chamber 2 Not Draining
To top of MV43
To bottom of WC-2
To top of WC-1
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
17
Questions?
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XT-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
18
XT Series
• Waste chamber 2 not draining errors.– MV18 clogged.
– Flush tube from WC2 to MV18.
– Manual waste drain.
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XT-Series
MV 18
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
19
XT-Series
• Waste Chamber 1 drain errors.– Back pressure from waste tubing
– Nipple clogged
– Replace drain tubing from back of instrument to drain
– Fluid connector clogged
– Excessive build up in waste chamber 1
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XT-Series
Pinch Valve for WC1 DrainCopyright © 2013 by Sysmex America, Inc. All
rights reserved..
2/20/2013
20
XT-Series
• Reaction unit overflows/shift on QC– Fluid in tray under manual pipette
– Reaction unit drain port clogged
– MV16 2, 3, 4 fittings clogged
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XT-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
21
XT-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
Questions?
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
22
XS-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XS-Series
• WBC QC shift/abnormal scatter grams– Clean reaction unit with bleach
– Filter 13 clogged
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
23
XS-Series
Reaction Unit Detail
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
• RBC/HGB mixing chamber errors– Filter 13 clogged
– Cleaned chamber with bleach and perform drain function
XS-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
24
XS-Series
RBC Mixing Unit Detail
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XS-Series
• Leaving fluid on caps of tubes– Piercer rinse cup is likely clogged
– Remove, clean rinse cup with 10% bleach, and re-install
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
25
XS-Series
Piercer Rinse Cup Detail
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
XS-Series
• Pressure / Vacuum Issues– Pressure or vacuum is not stable
– Cannot be adjusted within acceptable range
– QC shifts or other data errors
– Replacement of pneumatic pump may be indicated
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
2/20/2013
26
Removing top cover
Removing air pump
XS-Series
Copyright © 2013 by Sysmex America, Inc. All rights reserved..
Sysmex Service Tips
• Questions…