nice NeOumltech Medical Systems Pvt Ltd
Presentation Flow
COMPANY OVERVIEW
CLINICAL ASPECTS OF nice 2000 SERIES INFANT RADIANT WARMER
PRODUCT OVERVIEW
TROUBLESHOOTING
Company INFO
Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive
for growth and delighted customers
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
Why Infant Radiant Warmers are Used
Assist Thermoregulation
Body warming by providing Far IR radiant heat
Ensure baby is stabilized after birth
INFANT RADIANT WARMERS
MODES OF HEAT LOSS
Conduction Heat Loss Direct heat transfer from skin to object
Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature
Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum
(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator
Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation
Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Presentation Flow
COMPANY OVERVIEW
CLINICAL ASPECTS OF nice 2000 SERIES INFANT RADIANT WARMER
PRODUCT OVERVIEW
TROUBLESHOOTING
Company INFO
Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive
for growth and delighted customers
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
Why Infant Radiant Warmers are Used
Assist Thermoregulation
Body warming by providing Far IR radiant heat
Ensure baby is stabilized after birth
INFANT RADIANT WARMERS
MODES OF HEAT LOSS
Conduction Heat Loss Direct heat transfer from skin to object
Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature
Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum
(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator
Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation
Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Company INFO
Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive
for growth and delighted customers
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
Why Infant Radiant Warmers are Used
Assist Thermoregulation
Body warming by providing Far IR radiant heat
Ensure baby is stabilized after birth
INFANT RADIANT WARMERS
MODES OF HEAT LOSS
Conduction Heat Loss Direct heat transfer from skin to object
Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature
Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum
(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator
Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation
Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
Why Infant Radiant Warmers are Used
Assist Thermoregulation
Body warming by providing Far IR radiant heat
Ensure baby is stabilized after birth
INFANT RADIANT WARMERS
MODES OF HEAT LOSS
Conduction Heat Loss Direct heat transfer from skin to object
Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature
Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum
(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator
Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation
Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
Why Infant Radiant Warmers are Used
Assist Thermoregulation
Body warming by providing Far IR radiant heat
Ensure baby is stabilized after birth
INFANT RADIANT WARMERS
MODES OF HEAT LOSS
Conduction Heat Loss Direct heat transfer from skin to object
Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature
Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum
(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator
Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation
Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Why Infant Radiant Warmers are Used
Assist Thermoregulation
Body warming by providing Far IR radiant heat
Ensure baby is stabilized after birth
INFANT RADIANT WARMERS
MODES OF HEAT LOSS
Conduction Heat Loss Direct heat transfer from skin to object
Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature
Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum
(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator
Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation
Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
MODES OF HEAT LOSS
Conduction Heat Loss Direct heat transfer from skin to object
Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature
Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum
(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator
Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation
Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Electromagnetic Spectrum
Infrared Spectrum
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Effects of Far IR in Infant body
Far Infrared Rays are invisible waves of energy
They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Types of heaters Quartz
A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping
Far IR Infrared
Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Ceramic Heaters
Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc
Tubular Heater
A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material
Types of heaters(Contrsquodhelliphellip)
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Benefits of Heaterhellip
QUARTZ
bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack
CERAM
IC
bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)
TUBULAR
bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating
Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
A range of Baby friendly equipments
6 Layer security Introducing nice NeOumltech 2000 series
warmers
Designed to perfection for your baby
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
nice 2000 series Infant Radiant Warmers
nice 2000 B Infant Radiant Warmer Servo skin and Manual Mode
Sleek and Economic
Warmer with bed
nice 2000 CFL Infant Radiant Warmer
Servo skin and Manual Mode
LEDCFL Phototherapy
Warmer with bed
nice 2000 Infant Radiant Warmer
Servo skin and Manual Mode
Standalone
Height Adjustable
Without Bed
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
bullMicro controller based electronicsbull6 Layer securitybullServo Skin and Manual mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullShort wave heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode
Features
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
How to operate
Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray Procedure
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Control Panel Operation
1 Skin Temperature Display 10 Manual Mode Indicator
2 Alarm Indicator 11 Skin Probe
3 Manual Timer Indicator 12 Power failure Indicator
4 Mute ON Indicator 13 Skin Mode Indicator
5 Sensor Failure Indicator 14 Increase Key
6 Set TemperatureManualAPGAR Timer Display 15 Mute Key
7 Decrease Key 16 Heater ON Indicator
8 Set Key 17 Skin Temperature Low Indicator
9 Mode Key 18 Skin Temperature High Indicator
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Operating Modes
Manual Mode Operation
1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output
1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)
1048729 of Heater output Setting ndash Press increase ()Decrease () switch to adjust the of Heater Output in 10 increments
1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Skin Mode Operation
Note The skin temperature probe must be properly attached before choosing Skin mode operation
1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode
1048729How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC
APGAR Timer The APGARManual timer display indicates the APGAR timer select in the APGAR Mode
Operating Modes(contrsquodhellip)
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console
Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)
Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)
Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)
Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect
Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power
Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Heater Module
Parabolic design
Even Distribution of heat over bed surface
Extensive safety
measures(6 Layer safety for thermal
cut- off)
Far IR heating ensures
stimulation of cells for growth
The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Bed Platform Operation
bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Adjust the height of the rod to adjust the position of the bed (Up or down)
bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Side Panel Operation
bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
X-ray Procedure
bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed
bullThe heater housing rotates to the left or right for X-ray procedures
bullThe X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed
bull An X-ray cassette can be placed on the tray and slid into the cavity beneath the bed without disturbing the patient
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Comparison of nice 2000 Infant Radiant Warmersnice 2000 B nice 2000 CFL nice 2000
Skin Mode Y Y Y
Manual Mode Y Y Y
X Ray Facility Y Y N
Bed Y Y N
Pivoting Warmer ModuleY Y Y
Apgar Timer N N N
Height Adjustment N N Y
Utilities optional optional optional
Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
How nice Infant Radiant Warmers fare over competitors
nice 2000 series warmers
User-friendly easy to access all parameters
Long life quartz heater has a long life compared to other heaters
6 Layer safety for the protection of infant from burning
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
6 Layer Safety Features
Alarms if infant temperature greater or less than Set Temperature
Thermal cut off if infant temperature exceeds 39 degC
Heater output reduces to 30 after 15 mins in Manual mode
Heater output reduces to 30 if electronics senses no variable feedback
Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby
Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Cleaning Procedure
1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions
2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing
3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Preventive Maintenance
1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer
Sl No Problem Cause Remedy
1 Power failure LED is ON with audio alarm
Power FailureCheck the unit is Plugged in to Main supply
Check the mains are switched ON
Else Contact nice Neotech
2System Failure LED is ON with audio alarm
System Failure Contact nice Neotech
General System Failure
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy
1 No Display on the Control PanelCheck AC Supply Voltage
Check if Power Cord is disconnected
Check Fuse in the Rear side of Pillar
Contact nice Neotech
2 No Heater OutputCheck Link cord connected between Control Module toHeater Module
Set Temperature may be less than Actual Temperature
Check Heater Connection
Contact nice Neotech
3 Equipment Working But Observation Lamp amp Heater output not working
Check Link cord connected between Control Module toHeater Module
Check Fuse in the Rear side of Heater Module
Check Transformer Voltage - 12V AC
Replace Lamp
4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature
Check Mute ON
Contact nice Neotech
5 Patient Temperature Display shows ldquoS-Ordquo
with Continuous Alarm(S-O Sensor Open)
Check if skin Probe is Disconnected from theEquipment
If skin probe is defective replace new probe
Contact nice Neotech
7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module
Check Heater Connection
Contact nice Neotech
8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)
Check Environment Temperature above 39deg C
Check Infant Fever
9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set
Temperature
Check Environment Temperature is 1degC above SetTemperature
Contact nice Neotech
10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Temperature
Probe may be wet
Contact nice Neotech
11 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)
Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode
Low Supply Voltage
12 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource
Percentage Heater Output is Set at High in ManualMode
13 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)
Check for Strong Air conditioning draft
14Temperature Display does not display InfantsTemperature accurately
Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly
15Baby skin Temperature not correlate with display Temperature
Check sensitive portion of the Probe is fixed with babyskin properly
Check Probe is fixed to the lower abdomen of baby
16 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech
17 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm
Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)
System Fault Infant Radiant Warmer(Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product IEC 60601-2-21Electrical safety IEC 60601-2-50
IEC 60601-1-4
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594