2 THE SIMPLE TOC-DETERMINATION –QuickTOCtrace FAST AND RELIABLE. LAR | PROCESS ANALYSERS AG
A MEASURING SYSTEM FORULTRA PURE WATER .The lowest organic impurities are detected fast and cost-effectively - for opti-mal process control of challenging industries.
In order to ensure the high product quality andto guarantee process security, the fast and re-liable monitoring of ultra pure water is of theutmost importance. This applies in particularto manufacture of pharmaceutical productsand semiconductor technologies.
UPW - Ultra pure water. Highest requirementswith minimal impurities.Ultra pure water is prepared under particularly strin-gent specifications. It is a matter of ensuring puri-ty in respect of all types of contaminants: organic/inorganic compounds, dissolved/solid or volatile/non-volatile components, dissolved gases, reactive/inert substances etc.
Depending on the respective application, require-ments regarding UPW purity are defined in normssuch as ASTM D5127 and SEMI F63, as well aspharma regulations (pharmacopoeia).
UPW In the manufacture of semiconductors.Alongside solar technology and photovoltaics, en-ergy generation and the pharmaceutical industry,UPW is primarily used in the manufacture of semi-conductors. It is used inter alia in the rinsing of wa-fers, the dilution of chemicals and in optical systemsin photolithography. The primary and critical applica-tion of UPW is front end cleaning in themanufactureof integrated circuit boards.
UPW In the pharmaceuticalindustry.Water for injections (WFI) is a demanding applica-tion of UPW in the pharmaceutical industry. In themanufacture of any type of active ingredient andmedicine, the entire production process, includingthe cleaning process, is seamlessly monitored.
The QuickTOCtrace is suitable for the determination of TOC in ultra pure and purified water - especially in the manufacture of semi-conductors and pharmaceutical industries as well as in any process, where ultra pure water is essential.
THE SIMPLE TOC-DETERMINATION – 3FAST AND RELIABLE. QuickTOCtraceLAR | PROCESS ANALYSERS AG
Quality assurance and process security throughcontinuous monitoring.The monitoring of the above mentioned parametersis of great importance with regard to quality assur-ance. In the manufacture of semiconductors, par-ticles and organic/inorganic components can havea significant effect on sensitive photo-lithographicprocesses. Biological growth, often promoted by anincreased TOC content, can contribute to unintend-ed chemical processes.In the pharmaceutical industry, the assurance ofproduct quality is paramount. Depending on the ar-eas of application, WFI is used for injections, highlypurified water (HPW) for the processing of sterilepreparations and purified water (PW) for non-sterilepreparations such as pills, and for cleaning process-es.
Sources of impurities (particles, bacteria, organic(TOC) and anorganic components) are the feed wa-ter and additional water used in the preparation ofultra pure water on the one hand. On the other hand,distribution systems can in turn be a source of con-tamination, as can additives and cleaning process-es. Regular chemical and/or steam-based cleaningprocesses (pharma), as well as ultrafiltration (semi-conductors), ozonisation and general optimising ofduct systems canminimise impurities.The effectiveness of cleaning processes and thequality of the ultra pure water used are subject tomonitoring through TOC analyses and conductivitymeasurements.
TOC determination using UV oxidation.As a cumulative parameter, TOC is a measure of theorganic impurity of water and is thus an importantindicator of its quality. TOC is determined by meansof oxidation of aqueous samples and the quantita-tive measurement of the resulting CO2. In typicalUPW application, organic contamination is <1mg/l
TOC and the sample is enriched with oxygen, whichensures total oxidation through UV light. This is anaffordable method that yields quick results.
Differential conductivity measurement.In ultra pure water systems, electrolytic conductivi-ty and resistance measurement are the most wide-ly used indicators of contamination. In differentialmeasurement, the conductivity of the sample priorto UV oxidation is measured in a first step, and asubsequent measurement is carried out after oxi-dation. Under UV light, organic components are ox-idised into CO2, which increases conductivity. Thischange is used to calculate TOC concentration.
The QuickTOCtrace.Using UV oxidation and differential conductivitymeasurement, LAR’s QuickTOCtrace reliably deter-mines organic concentration in a range of 0.1 to1,000 ppb TOC and yields results within 30 seconds.
AT A GLANCE
• The quality of high-purity and ultra pure watermust be assured.
• Quality assurance and process security requirereliable methods.
• The TOC value is the measure of water sample’sorganic load.
• The QuickTOCtrace is designed to be easy tomaintain and user-friendly, yielding results in afew seconds.
Requirements for pure water in semiconductor and pharmaceutical industriesParameter SemiconductorUPW PharmaWFI Pharma HPW Pharma PWConductivity 0.055 µS/cm <1.1 µS/cm (20 °C, Ph. Eur.) /
<1.3 µS/cm (25 °C, USP)
<1.1 µS/cm (20 °C, Ph. Eur.) <4.3 µS/cm (20 °C, Ph. Eur.) /
<1.3 µS/cm (25 °C, USP)
TOC <1 µg/l (online <10 ppb) <0.5 mg/l (Ph. Eur., USP) <0.5 mg/l (Ph. Eur.) <0.5 mg/l (Ph. Eur., USP)
Bacteria <1 CFU/100 ml <10 CFU/100 ml (Ph. Eur., USP) <10 CFU/100 ml (Ph. Eur.) <100 CFU/100 ml (Ph. Eur., USP)
Ph. Eur. = Europäische Pharmacopoeia USP = US Pharmacopoeia
4 THE SIMPLE TOC-DETERMINATION –QuickTOCtrace FAST AND RELIABLE. LAR | PROCESS ANALYSERS AG
Continuous TOC measurement by means of UVoxidation and differential conductivity measure-ment.The QuickTOCtrace is an online system working onthe basis of the traditional continuous flow processwith conductivity detection. UV oxidation usingthe “direct surface” process guarantees highestprecision thanks to highly sensitive conductivitysensors. With integrated temperature sensors, alldata that is relevant of the measurement of TOCare monitored and processed. The QuickTOCtracecan thus reliably determine the parameter TOC ina measuring range of 0.1 - 1,000 ppb.
The QuickTOCtrace is an easy to use, low-mainte-nance online measuring device that enables quickand reliable monitoring of demanding applications.
It is particularly suited for applications in the phar-maceutical industry (WFI, HPW, PW), as well asultra pure water (UPW).
System Suitability Test - SST.The verification of the TOC result is carried outusing periodical implementation of the systemsuitability test and complies with the standardsprescribed by the US pharmacopoeia, as well asthe European pharma regulation (European phar-macopoeia).QuickTOCtrace’s software enables a quick andsimple initiation of the SST at any time and offersautomatic data evaluation as well as automatic im-plementation of the testing procedure.
Qualification and calibration.Calibration can be carried out by the user at anytime. By connecting the inlets for the calibrationsolution and the diluted water sample, as well asterminating the online measurement, the calibra-tion can be carried out using the software. Statis-tical evaluation of measurements and the respec-tive calibration parameters are displayed on thescreen (plausibility test).
Software and data output.The QuickTOCtrace is controlled via an integratedprocessing unit and an 8” touch screen monitor.Based on the Windows 7 operating system, datacan be embedded and documented as encrypteddate-specific files, so that crude data is not manip-ulated as per pharma guidelines. All saved filescan be accessed using the viewer function and/orexported for further processing.
THE ANALYSER.Especially for challenging applications that require reliable detection withinseconds of lowest impurities in pure water.
TheQuickTOCtrace
is very easy tooperate and requi-
res extremly lowmaintenance.
THE SIMPLE TOC-DETERMINATION – 5FAST AND RELIABLE. QuickTOCtraceLAR | PROCESS ANALYSERS AG
THE PRINCIPLE.Safe and fast measurements.
UV oxidation. formation of free radicals.Oxidation using the “direct surface” procedure -thus without the use of chemical oxidants - hasproven very effective in the measurement of verysmall TOC concentration, commonly found in ultrapure water.When water containing O2 is exposed to UV radia-tion (185 nm, 254 nm), short-lived, highly reactivespecies with a high oxidation potential are formed.Amongst them is for example the OH●
radical,which can oxidise organic compounds to CO2.
The conductivity of ultra pure water is strongly de-pendent on dissolved CO2. UV oxidation producesCO2, which increases the conductivity of the sam-ple. This change is used in differential conductiv-ity measurement to determine the correspondingTOC concentration.
Differential conductivity measurement.Conductivity measurement is based on the meas-urement of ohmic resistance, or its reciprocal val-ue, the electrolytic conductivity value. This con-ductivity is influenced by the value, volubility andparticle density of ions in the aqueous solution. Itis therefore an effective indicator of any impurities.
In differential conductivity measurement, the in-itial conductivity before oxidation, as well as theconductivity after the treatment of the sample, aremeasured. The TOC concentration results fromthe difference obtained from subtracting the first,lower conductivity measurement from the second,higher one.
1) Sample preparation2) UV-Oxidation of orga-
nic carbon to CO2 byHydroxyl-Radicals
3) Determination of theTOC contaminationa) Measurement of
conductivity prior tooxidation (LF 1)
b) Measurement ofconductivity afteroxidation (LF 2)
c) CalculationLF 2 - LF 1 = TOC
Fig.1
MEASUREMENT PRINCIPLEof QuickTOCtrace
CHEMICAL REACTIONS
H2O + hv (185 nm/ 254nm)� OH
●
+ H●
und 2 H●
� H2CO2 + H2O � H2CO3
-� HCO3- + H+
HCO3- + H+ � CO3
2- + H+
UV Oxidation
2
Sample
1
Measurement3
Sample
SST/ Calibration
Sample inletUV-Lamp
UV Reactor
Inlet 2
Inlet 3Drain
2. Conductivitymeasurement
1. Conductivitymeasurement
a b
Software• Process control• Data input and output• Quantitative determination of TOC
c
With the UV-Oxi-dation CO2 is crea-ted which is a partof the carbonate-hydrogencarbonateequilibrium.
TOC-Analysis LAR | PROCESS ANALYSERS AG
The information and the illustrations in this brochure regarding appearance, specifications, service, measure, weight, consumption, maintenance times and so forth are approxi-mate, in no way binding and subject to change. All information is correct at the time of publication. We reserve the right to deviate in construction, design, colour, as well as makechanges in our delivery options.Version QTtrace-1E2218.
QuickTOCtrace AN OVERVIEWOnline TOC measurement – the fastest way to analyse ultra pure water.
The LAR QuickTOCtrace is a measuring system for the continuous online determination of thetotal organic carbon (TOC) in ultra pure water applications such as ultra-pure process waterin the semiconductor manufacture.
Measurement Technique and Sample Preparation
Measurement Method UV oxidation (partial) withdifference-conductivity measurement
Measurement Ranges 0.1 - 1,000 ppb TOC,0.055 - 2.0 µS/cm conductivity
Response Time <30 seconds (T90)
Parameter TOC, conductivity
Calibration Automatic and manually
Sample Streams 1 to 3 (optional)
Dimensions and weight
Housing Stainless Steel IP 20, IP 54 (optional)
Dimensions 300 x 500 x 200 mm (W x H x D)
Weight ca. 7 kg
Electric Specifications
Power Supply 230 / 115 VAC, 50 Hz
Analog Output 4– 20 mA (0– 20 mA optional)
Interface USB 2.0
Equipment Devices and Data Output
8“ touch screen graphic display, backlit
Integrated computer for control
Operating system Windows 7 embedded
Self explanatory software incl. Viewer
Automatic System Suitability Test (SST)
Accessories and Options
Calibration solutions, SST-Kit, IQ/OQ documentation(conform to NIST, USP etc.)
Heat-exchanger for samples with > 50 °C
TECHNICAL DATA
Fast and safe –you can rely on the QuickTOCtrace!
ADVANTAGES& FEATURES
✔ Recognised differential conductivitymeasurement with UV oxidation
✔ Reaction time of <30 seconds
✔ Continuous measurement(every 2 seconds)
✔ Highly effcient oxidation
✔ UV lamp with a long life span
✔ Easy to use
✔ Easy calibration (1 time per year)
✔ Easy system validation with fully auto-matic SST (optional)
✔ Comprehensive data storage
✔ Very low maintenance
THE SIMPLE TOC-DETERMINATION – 7FAST AND RELIABLE. QuickTOCtraceLAR | PROCESS ANALYSERS AG
ALL cLAR?LAR Process Analysers AG: Water is our Element.We do everything for its protection.
COD-ANALYSIS
With our analysers, thechemical oxygen demandis cleanly and safely deter-mined online, without usinghazardous chemicals.
BOD/TOXICITY
We detect the BOD withthe plant‘s own biomassand determine the toxicitywith highly sensitive bacte-ria, fast and reliably.
TOC-ANALYSIS
From complex industrywaster water to pharma-ceutical pure water, ourTOC analysers determinethe parameter quickly andprecisely.
TNb/TP-ANALYSIS
TNb and TP are importantparameters for waste watertreatment. We are the onlyones who offer a combina-tion of these with TOC andCOD in one system.
FURTHER PRODUCTS
LAR offers a specificsolution for nearly all appli-cations. With our protectivehousings, you are always onthe safer side. Learn moreabout our product range atwww.lar.com.
We are the leading manufacturer of wateranalysers for industrial and municipal wastewater treatment, process monitoring, as wellas for pure water analysis. further productsin the areas of environmental technology andindustrial processing complete our productportfolio.
Unique and state of the art.l Ar’s ultra High Temperaturemethodat 1,200°C!LAR formed in 1986, gained prominence throughtheir TOC and COD analysers. LAR is the onlycompany worldwide that, using a high temperaturemethod at 1,200°C, can completely oxidise asample to accurately determine sum parameters.Particularly when measuring the TRUE TOC withdiffering concentrations.
LAR is only satisfied once the customer is.We offer application specific analysers developedby our research and development team. Inaddition, we maintain close contact with ourclients and continually analysis the exact problemareas of every application.Since the availability of our devices is a decidingcriteria, they are constructed in a very user-friendly way. All important areas require little effort
to be accessed and the protective housing offersadditional safety.
After sales. A familiar word to us.Servicing is carried out by our qualified partnersworldwide. Technical support, via telephone ore-mail is available at all times. Additionally, weoffer practically orientated seminars and trainings,operator meetings and workshops, that leave noquestions unanswered.
We always take a closer look.Setting ourselves the highest quality standards,we closely cooperate with our partners to fulfillthe customers expectations throughout the world.Thus, we regularly evaluate our distributors andwhen necessary, introduce measures to improveour collaboration with them.LAR has established its own system forguaranteeing its standards of quality. Not onlydo we fulfill the requirements of the ISO 9001,but we also work continually on improving ourstandards of quality. To enable this, we collectinformation about all applications in our databasethat are subsequently analysed and evaluated.Regular meetings are held to address every issueguaranteeing highest quality standards.
QuickTOCtrace
TOC-ANALYSIS
The information and the illustrations in this brochure regarding appearance, specifications, service, measure, weight, consumption, maintenance times and so forth are approximate, inno way binding and subject to change. All information is correct at the time of publication. We reserve the right to deviate in construction, design, colour, as well as make changes in ourdelivery options.Pictures: ©LAR Process Analysers AG, ©Image Source-Fotolia.com (Cover), ©prescott09-Fotolia.com,©Bigstock/livertoon.Version QTtrace-1E2218.
Neukoellnische Allee 134D-12057 Berlinwww.lar.com
Telefon +49 30 278 958 -43Telefax +49 30 278 958 - 703E-Mail [email protected]
LAR Process AnalysersAG
TÜV certified company
A R E A S O F A P P L I C AT I O N
ENVIRONMENT/MUNICIPAL FACILITIES / INDUSTRY
I N D U S T R I E S
ENVIRONMENTAL MONITORING / WASTE WATER TREATMENT /
POWER / WASTEPROCESSING / AIRPORTS / AUTOMOBILE /
PHARMACEUTICAL / LABORATORY/ CHEMICAL / PETROCHEMICAL/
REFINERIES / COAL AND STEEL / PAPER MANUFACTURE /
BREWERIES / FOOD MANUFACTURE / DRINK MANUFACTURE /
MILK PROCESSING/SEMICONDUCTOR MANUFACTURE
T Y P E S O F WAT E R
GROUNDWATER / SURFACE WATER / DRINKING WATER /
WATER INFLUENT / WATER EFFLUENT / DISCHARGE CONTROL /
INDUSTRIALWASTEWATER / DE-ICINGWATER / PROCESS
WATER / OIL-IN-WATER/COOLINGWATER/ PURE WATER /
BOILER FEED WATER / CONDENSATE RETURN /
HIGH SALTCONCENTRATION/PHARMA HPW / PHARMA WFI /
SEMICONDUCTOR UPW