7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 1/32
Standing guard over the flame of life
Oxygen Monitoring in Acute Medicine
Dave SainsburyStaff Anaesthetist, Women’s & Children’s Hospital
Adelaide, South Australia
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 2/32
1/25/2011 Oxygen Monitoring 2
n 2nd law of thermodynamics
n Entropy is time’s arrow
Entropy Life
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 3/32
1/25/2011 Oxygen Monitoring 3
CnHnOn
O2
CO2
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 4/32
1/25/2011 Oxygen Monitoring 4
1 5 91 4 9
1 0 09 2
4 0 4 0
2
0
5 0
1 0 0
1 5 0
Tissue Vein
Mitochondria
I
n
s
p
ir
e
d
T
ra
c
h
ea
A
l
ve
o
l
u
s
A
rt
e
r
y
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 5/32
1/25/2011 Oxygen Monitoring 5
Monitoring Oxygen As a Gas
ª Inspired
ª Tracheal
ª
Alveolar
â Oxygen analyser
â Calculated from FIO2
â
Calculated: Alveolar air equationEnd tidal oxygen analyser
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 6/32
1/25/2011 Oxygen Monitoring 6
Ideal Oxygen Analyser
n Rapid response
n Continuous
n Robust
n
Low maintenancen Do no harm
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 7/32
1/25/2011 Oxygen Monitoring 7
Oxygen Analysers
n Polarographic sensor
n Fuel cell
n Paramagnetic analyzer – Robust
– Rapid response time
– Low maintenance
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 8/32
1/25/2011 Oxygen Monitoring 8
Paramagnetic Oxygen Analyser
Pressure transducer
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 9/32
1/25/2011 Oxygen Monitoring 9
Oxygram
Inspired
Expired
30%
20%
10%
0%
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 10/32
1/25/2011 Oxygen Monitoring 10
Alveolar Air Equation
n P A O2 = FIO2 * (Pb- 47) - PaCO2
n P A
O2 = FI
O2 * (Pb
- 47) - Pa
CO2 /R
n P A O2 = FIO2 * (Pb- 47) - PaCO2 /R(1-FIO2(1-R))
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 11/32
1/25/2011 Oxygen Monitoring 11
1 5 91 4 9
1 0 09 2
4 0 4 0
2
0
5 0
1 0 0
1 5 0
Tissue Vein
Mitochondria
I
n
s
p
ir
e
d
T
ra
c
h
ea
A
l
ve
o
l
u
s
A
rt
e
r
y
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 12/32
1/25/2011 Oxygen Monitoring 12
Partial pressure of oxygen (mmHg) O
x y g e n s a t u r a t i o n o f h a e m o g l o b i n ( % )
O x y g e n c o n t e n t o f b l o o d ( m l / 1 0 0 m l )
0
10
20
30
40
50
60
70
80
90
100
5
10
20
15
2010 806040 100 500300 600200 400
Artery
Vein
P50
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 13/32
1/25/2011 Oxygen Monitoring 13
Ideal blood oxygen monitor
n Indicate oxygen tension
n Indicate oxygen supply
– Haemoglobin concentration
– Haemoglobin saturation
– Blood flow
n Rapid response
n Continuous
n Safe
– Non-invasive
n Robust
n Low maintenance
n Stable calibration
n Affordable
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 14/32
1/25/2011 Oxygen Monitoring 14
Monitoring blood oxygen
ª Arterial tension
ª Arterial saturation
â Blood gas analyser
â Trans cutaneous
â Calculated from tensionCo-oximeterPulse oximeter
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 15/32
1/25/2011 Oxygen Monitoring 15
Blood gas analyser
n Polarographic
n Measures tension
n Calculates saturation (functional)
n Problems
– invasive
– intermittent
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 16/32
1/25/2011 Oxygen Monitoring 16
Functional Saturation
OxyHb(OxyHb+reducedHb)
7g OxyHb 1g reducedHb 2g metHb
Functional Saturation = 7/(7+1)
= 87%
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 17/32
1/25/2011 Oxygen Monitoring 17
Transcutaneous gas analyser
n Polarographic
n Measures tension
– persistent pulmonary hypertension
– diaphragmatic hernia
n Problems
– Delicate
– Heated
– Repositioning
– Recalibrating
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 19/32
1/25/2011 Oxygen Monitoring 19
Fractional Saturation
OHb(OHb+reducedHb+COHb +metHb)
7g OxyHb 1g reducedHb 2g MetHb
Fractional Saturation = 7/(7+1+2)
= 70%
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 20/32
1/25/2011 Oxygen Monitoring 20
Oximeters
• 1875 vonVierordt
• 1972 Takuo Aoyagi - "Pulse Oximetry"
• 1980's Nellcor
• 1988 pulse oximetry mandatory in Australia
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 21/32
1/25/2011 Oxygen Monitoring 21
Light absorption
due to tissuesArtery
(pulsatile tissue)
Non-arterial,
(non-pulsatile) tissu
Systole DiastoleLight absorption
due to arterial blood
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 22/32
1/25/2011 Oxygen Monitoring 22
Light wavelength (NM)
6 0 0 6 8 0 7 2 0 7 6 0 8 0 0 8 8 0 9 2 0 9 6 0 1 0 0 06 4 0 8 4 0
E x t i n c t i o n c
o e f f i c i e n t
.0 1
.1
1 .0
1 0
Red
Deoxyhaemoglobin
Oxyhaemoglobin
Infrared
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 23/32
1/25/2011 Oxygen Monitoring 23
O x y
g e n s a t u r a t i o n o
f h a e m o g l o b i n (
% )
0
10
20
30
4050
60
70
80
90
100
1.00.4 2.0 3.0 3.4
R =Red absorption
Infra-red absorption
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 24/32
1/25/2011 Oxygen Monitoring 24
Oximeter errors
n Signal artefact
n Light absorption artefact
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 25/32
1/25/2011 Oxygen Monitoring 25
Signal artefact
n Ambient light
n Poor perfusion
n Movement artefactn If R = 1 pulse oximeter shows 85%
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 26/32
1/25/2011 Oxygen Monitoring 26
Light absorption artefact
n carboxyHb
n metHb
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 27/32
1/25/2011 Oxygen Monitoring 27
Light wavelength (NM)
6 0 0 6 8 0 7 2 0 7 6 0 8 0 0 8 8 0 9 2 0 9 6 0 1 0 0 06 4 0 8 4 0
E x t i n c t i o n c
o e f f i c i e n t
.0 1
.1
1 .0
1 0
Red
Deoxyhaemoglobin
Oxyhaemoglobin
Carboxyhaemoglobin
Infrared
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 28/32
1/25/2011 Oxygen Monitoring 28
Light wavelength (NM)
6 0 0 6 8 0 7 2 0 7 6 0 8 0 0 8 8 0 9 2 0 9 6 0 1 0 0 06 4 0 8 4 0
E x t i n c t i o n c
o e f f i c i e n t
.0 1
.1
1 .0
1 0
Red
Methaemoglobin
Deoxyhaemoglobin
Oxyhaemoglobin
Infrared
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 29/32
1/25/2011 Oxygen Monitoring 29
Light absorption artefact
n carboxyHb
– error showing low functional saturation
– error showing high fractional saturation
n metHb
– error toward 85%
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 30/32
1/25/2011 Oxygen Monitoring 30
Light absorption artefact
n carboxyHb – reads low functional, high fractional
n metHb
– error toward 85%
n fetal Hb
n dyes
n nail polish
n bilirubin
n skin pigment
7/31/2019 Oxygen Pulse Oximetry
http://slidepdf.com/reader/full/oxygen-pulse-oximetry 31/32
1/25/2011 Oxygen Monitoring 31
Monitoring oxygen
ª Tissue
ª Venous tension
ª Venous saturation
â Cytochrome oximeters
â Blood gas analyser
â Optical fiber oximeter