O P E R A T I N G I N S T R U C T I O N S
FRIGOSTREAM
01639 GB
Legal Notice
Distribution, reproduction and translation of the software and its documentation (or excerpts thereof) are
prohibited without the prior written consent of PHYSIOMED ELEKTROMEDIZIN AG.
PHYSIOMED ELEKTROMEDIZIN AG reserves the right to change the software and associated data as well
as the documentation without notice. All other rights reserved.
PHYSIOMED ELEKTROMEDIZIN AG
Hutweide 10
D-91220 Schnaittach/Laipersdorf (Germany)
Phone: +49 9126 2587-0
Fax: +49 9126 2587-25
E-Mail: [email protected]
Internet: http://www.physiomed.de
FRIGOSTREAM is produced in accordance with the quality requirements set forth in DIN EN ISO
13485:2003 and complies with current safety standards and regulations in accordance with the Medical
Devices Act as set forth in Directive 93/42/EEC on Medical Devices dated 14 June 1993.
The conformity with the specified directive is confirmed by the CE marking on the instrument. The declar-
ation of conformity can be requested under the above mentioned address.
Last updated February 16, 2010.
Table of Contents
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Conventions Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Instrument Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 General Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.4 Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Instrument Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.6 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.6.1 Rheumatology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.6.2 Surgery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6.3 Neurology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6.4 Sports Medicine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.7 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 2 Display and Control Elements . . . . . . . . . . . . . . . . . . . . . . . 6
2.1 Function of the Display and Control Elements . . . . . . . . . . . . . . . . . . 6
2.1.1 Display <1> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1.2 Mode Key <2>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1.3 Start/Stop Key <3> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.4 Plus/Minus Keys <4> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.5 Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 3 Operation of the Instrument . . . . . . . . . . . . . . . . . . . . . . . 8
3.1 Start-up of the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Functional Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Instrument Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 Defrosting the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Chapter 4 Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Appendix A Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
A.1 Customer Service, Maintenance, Service . . . . . . . . . . . . . . . . . . . . 13
A.2 Cleaning and Disinfecting Agent . . . . . . . . . . . . . . . . . . . . . . . . 13
A.3 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
PHYSIOMED ELEKTROMEDIZIN AG Table of Contents
FRIGOSTREAM i
A.4 Service Life and Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
A.5 Electromagnetic Compatibility. . . . . . . . . . . . . . . . . . . . . . . . . . 14
A.6 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
A.7 Manufacturer Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Appendix B Scope of Delivery and Accessories . . . . . . . . . . . . . . . . . . . . 16
B.1 Scope of Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
B.2 Available Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Appendix C Supplementary Documents . . . . . . . . . . . . . . . . . . . . . . . . 17
C.1 Manufacturer’s Recommendation . . . . . . . . . . . . . . . . . . . . . . . . 17
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Table of Contents PHYSIOMED ELEKTROMEDIZIN AG
ii FRIGOSTREAM
Chapter 1 Introduction
By purchasing FRIGOSTREAM you have acquired an extremely versatile, advanced cryotherapeutic instru-
ment. However, it can be utilized optimally only once you are informed about the function of the control
elements. Therefore, read the given instruction manual carefully and make yourself thoroughly familiar
with the instrument before its use.
1.1 Conventions Used
Please note the following typographical conventions in these Operating Instructions:
• Cross references and important terms used for the first time in this document are written in italics.
• Names of menus and symbols on the display are written in bold typeface.
Paragraphs that deserve special attention are highlighted in the following way:
Symbol Type Meaning
Tip Intended to give you some extra hints for more convenient operation
Note Provides background information for better understanding
Important Prevents misunderstandings that might lead to limited operation of the in-
strument or insufficient therapeutical results
Caution Alerts you in cases of possible damage to the instrument or risks of injury
1.2 Instrument Symbols
The symbols affixed to the instrument have the following meaning:
Symbol Meaning
Make sure that the instrument is disposed of in an environmentally sound manner!
It is absolutely necessary to separate old instruments from household waste (see
Service Life and Disposal on page 14) before disposal once their lifetime has ex-
pired!
1.3 General Instructions
FRIGOSTREAM may be used only in dry rooms suitable for medical purposes as set forth in DIN VDE
0100-710, Application Group 1. These are rooms with beds, physiotherapy rooms, massage rooms and
practitioner’s rooms. The instrument complies with the technical requirement of VDE 0750/IEC 60601,
classification in accordance with EU Directive on Medical Devices (93/42/EEC): Class IIa.
PHYSIOMED ELEKTROMEDIZIN AG 1 Introduction
FRIGOSTREAM 1
Caution
The instrument may be operated only and exclusively by medically qualified personnel after
appropriate instruction! Kindly ensure proper usage of the instrument in accordance with the
instruction manual!
The instrument is unsuitable for use in explosive areas and hydrotherapy rooms. A drastic change in tem-
perature must be avoided owing to possible bedewing of the instrument; it must be put into operation
only after a temperature equalisation!
The operation of the instrument in the proximity of (e.g. at a distance of 1 m from) a shortwave or mi-
crowave therapeutic instrument can lead to fluctuations in the original values of the instrument and must,
therefore, be absolutely avoided!
1.4 Instrument Overview
Control Panel
Instrument View
Legend
1 Display 2 Mode Key
3 Start/Stop Key 4 Plus/Minus Keys
1 Introduction PHYSIOMED ELEKTROMEDIZIN AG
2 FRIGOSTREAM
1.5 Instrument Description
FRIGOSTREAM is a cryotherapeutic instrument, which has been applied successfully in different medical
fields. It is primarily foreseen as a therapeutic instrument for application in a clinical and therapeutic envir-
onment.
The cooling circuit of the FRIGOSTREAM is a closed circuit. The instrument has been filled with a pre-
defined volume of protective cooling medium and is hermetically sealed.
The instrument complies with all the applicable safety standards, in particular:
• IEC 60601-1: Electronic Medical Devices, Part 1: General Safety Requirements
1.6 Application
Cryotherapy is used in different medical fields. These fields include:
• rheumatology
• orthopaedics
• surgery
• sports medicine
• neurology
• internal medicine
The treatment of these indications takes place through the application of different cooling media. These
are:
• ice chips
• compresses
• ice packs
• mudpacks
• cooling spray
• nitrogen
• cooling air
1.6.1 Rheumatology
Applications of cryotherapy within rheumatology in connection with massages and kinesitherapy:
In the case of inflammatory diseases:
• arthritides
• spondylitides
• bursitides
• gout
• rheumatoid arthritis in acute episode
In the case of non-inflammatory diseases:
• arthroses
• acute arthrosis humeroscapularis
PHYSIOMED ELEKTROMEDIZIN AG 1 Introduction
FRIGOSTREAM 3
• acute epicondylitis
• myalgias
• lumbar ischialgias
• dorsalgias
• sacralgias
• acute lumbago
• acute cervical syndrome
1.6.2 Surgery
Applications of cryotherapy within accident surgery, orthopaedics, mouth and jaw surgery:
• prophylaxis of swellings in the case of fresh fractures and in the post-operative phase
• sprains
• crushing injuries
• contusions
• pulled muscles
• luxations (all kinds of injuries to the ligament and joint system)
• Sudeck's syndrome
• lymphedema after amputation of the breast
• prevention of reactive effusions
• acute inflammations
1.6.3 Neurology
Applications of cryotherapy within neurology:
• spastic hemipareses and parapareses
• neuralgias (e.g. migraine, trigeminus neuralgia)
• multiple sclerosis
• post-apoplectical conditions
• acute neuritis
1.6.4 Sports Medicine
Applications of cryotherapy within sports medicine:
• fresh sports injuries
• soft tissue injuries
• injuries to the ligament and joint system (e.g. muscle contusions, traumatic bursitis, humeral epicon-
dylitis, cervical spine whiplash injury)
1.7 Contraindications
May not be treated with FRIGOSTREAM in the case of:
• arterial occlusive diseases (especially Fontaine stages III and IV)
• grave functional blood circulatory disorders
• common cryoglobulinaemia
1 Introduction PHYSIOMED ELEKTROMEDIZIN AG
4 FRIGOSTREAM
• cold urticaria
• trophic disturbances and tissue lesions
• Raynaud’s symptomatology
• pregnancy (a treatment during pregnancy may be performed only in consultation with the attending
doctor!)
PHYSIOMED ELEKTROMEDIZIN AG 1 Introduction
FRIGOSTREAM 5
Chapter 2 Display and Control Elements
The FRIGOSTREAM is designed in such a way that it is very easy to operate in spite of manifold functions.
All the display and control elements are integrated in the housing in order that the surface can be cleaned
easily and the instrument is free from dust.
The micro-processor of the instrument monitors the safety-relevant components, prevents maloperation
and tests the instrument every time it is switched on.
2.1 Function of the Display and Control Elements
The individual display and control elements of the FRIGOSTREAM have been presented in the following
sections. The numbers in angle brackets refer to the Instrument Overview on page 2.
2.1.1 Display <1>
Display <1>
In the Display <1> the selectable therapy paramet-
ers are displayed depending upon the current con-
text.
2.1.2 Mode Key <2>
Mode Key <2>
The Mode Key <2> above the Display <1> serves the purpose of selecting
the programmes Flow, Time or Defrost.
2 Display and Control Elements PHYSIOMED ELEKTROMEDIZIN AG
6 FRIGOSTREAM
2.1.3 Start/Stop Key <3>
Start/Stop Key <3>
The Start/Stop Key <3> above the Display <1> serves the purpose of start-
ing and ending the therapy in the individual menus.
2.1.4 Plus/Minus Keys <4>
Plus und Minus Keys <4>
The Plus/Minus Keys <4> below the Display <1> serve the purpose
of setting the parameters in the individual menus.
2.1.5 Power Switch
The Power Switch is on the rear side of the instrument. With the Power Switch, the mains voltage of
the FRIGOSTREAM is switched on and/or off.
PHYSIOMED ELEKTROMEDIZIN AG 2 Display and Control Elements
FRIGOSTREAM 7
Chapter 3 Operation of the Instrument
Caution
The filter attached to the rear side of the instrument must be cleaned and/or washed every
week!
Note
Optimal operating conditions are given if used in air-conditioned rooms and at a humidity of
maximum 50-60%. In the worst case, there can be water formation in connection with a de-
frosting process if there is very high humidity and high outside temperatures.
Important
Increased humidity and temperatures above 30°C may impact the cooling performance of the
instrument. The average permanent operating time may decrease with rising temperature and
humidity. We recommend using FRIGOSTREAM only in air-conditioned rooms.
3.1 Start-up of the Instrument
Before you can perform the first treatment with the FRIGOSTREAM, you must first start up the instrument
properly.
Since the instrument is mounted on rollers ex-factory, it may be operated only on level surfaces.
Caution
While setting up the FRIGOSTREAM, a distance of approx. 10 cm must be observed between
the instrument and the wall, since it can otherwise result in loss of performance. The lateral
distance from the walls must be minimum 20 cm.
Instruments with strong electromagnetic fields (X-ray and/or diathermic apparatuses) can in-
fluence the functioning of the instrument.
How to Start up the Instrument
(1) Make sure that the instrument is switched off (the Power Switch is on Position 0).
(2) Plug the power cord on the rear side of the instrument tightly into the power socket. A safe connec-
tion can be established via a shockproof socket with protective earthing conductor.
(3) Actuate the Power Switch on the rear side of the instrument.
(4) The instrument cools down to the point where the operating temperature is achieved. The message
Loading flashes on the display and the status of the cooling process is displayed by means of a pro-
gress bar.
3 Operation of the Instrument PHYSIOMED ELEKTROMEDIZIN AG
8 FRIGOSTREAM
Progress Display of the Cooling Process
At the end of the cooling phase, the instrument automatically changes over to the menu Flow and is
then ready for operation.
For longer treatments, it is urgently recommended to allow the instrument to cool down for another
10-15 minutes. The further progress of the cooling is no longer visible in the display during this time
span.
3.2 Functional Test
If you are not sure as to whether your FRIGOSTREAM is working properly, you can conduct a self-test.
How to Execute a Functional Test
(1) Switch off the instrument by means of the Power Switch and then switch it on again.
The instrument will conduct an automatic functional test whereby all the functions and original val-
ues will be checked.
The instrument is now ready to operate. You are in the Flow menu.
3.3 Instrument Errors
In case a malfunction is ascertained during the self-test or during operation, the Display <1> shows the
message ERROR LOAD.
In case the message was caused by a temporary disturbance and disappears after a brief period, the in-
strument is immediately ready to operate once again without switching off and on.
However, if the error continues, the message once again appears. In this case, switch the instrument on
and off several times. In doing so, the instrument will conduct a self-test (see Functional Test on page 9).
Ascertain afterwards as to whether the message is still being displayed.
Important
Notify your service partner in case the error recurs even after several self-tests! Under no cir-
cumstances is it allowed to carry out any treatment unless the functional fitness of the instru-
ment is ensured!
PHYSIOMED ELEKTROMEDIZIN AG 3 Operation of the Instrument
FRIGOSTREAM 9
3.4 Defrosting the Instrument
In case there is a perceptible decrease in air flow rate or small ice particles are blown out (as a rule, only
after approx. 4 hours of continuous operation) you ought to defrost the instrument in order that the full
air flow rate can be guaranteed.
How to Defrost the Instrument
(1) Press the Mode key twice in order to reach the Defrost menu.
Defrost Menu
(2) Press the Start/Stop key for approx. 5 seconds. The defrosting process starts.
The message Run flashes on the bottom right of the display. In addition, the current storage temper-
ature is displayed.
The defrosting process is running
(3) The defrosting process runs until a storage temperature of +10°C is achieved.
Thereafter, the FRIGOSTREAM once again cools down automatically to the operating temperature.
3 Operation of the Instrument PHYSIOMED ELEKTROMEDIZIN AG
10 FRIGOSTREAM
Chapter 4 Treatment
The FRIGOSTREAM knows the following menus and operating conditions:
• Flow: Adjust the blower position and perform the treatment
• Time: Adjust the treatment time
• Defrost: Defrost the instrument
How to Perform a Treatment
(1) Make sure that the instrument is ready to operate (see Start-up of the Instrument on page 8). You
are in the Flow menu.
(2) Press the Start/Stop key for starting the treatment.
The message Run flashes on the bottom left of the display.
(3) Set the desired blower position by means of the Plus and Minus keys (nine positions). On the Dis-
play <1> the set blower position is displayed:
Flow Menu
(4) Once again press the Start/Stop key, if you wish to end the treatment.
Note
You can also limit the treatment to a specific period by setting a treatment time (see
How to Set the Treatment Time on page 11). Once the treatment time is over, the
blower switches off.
How to Set the Treatment Time
(1) Press the Mode key, in order to reach the Time menu.
Time Menu
PHYSIOMED ELEKTROMEDIZIN AG 4 Treatment
FRIGOSTREAM 11
(2) Set the desired treatment time by means of the Plus and Minus keys. On the Display <1> the set
set treatment time is displayed.
Continuous treatment with adjusted therapy time
Note
You can interrupt the treatment at any time by pressing the Start/Stop key.
In the Time menu, If you do not change the treatment time, the instrument automatically
changes over to the Flow menu after a few seconds.
4 Treatment PHYSIOMED ELEKTROMEDIZIN AG
12 FRIGOSTREAM
Appendix A Appendix
Cooling Medium
The cooling circuit of the cryotherapeutic instrument FRIGOSTREAM is a closed system. The
instrument has been filled with a predefined volume of protective cooling medium and is her-
metically sealed. Thus, no maintenance of the cooling circuit is required.
A.1 Customer Service, Maintenance, Service
The manufacturer is responsible only for the original security relevant properties of the instrument. The in-
strument must be used generally in conformance with the user manual.
Maintenance measures on the instrument may be carried out only at the PHYSIOMED ELEKTROMEDIZIN
AG authorised locations. If the work is carried out by such locations, a certification should be obtained
from them about the type and extent of the work. possibly with information relating to the modifications
in the rated data or the work range. The certification must also contain the date of execution as well as
company name and signature. Failed components, which influence the safety of the instrument, must be
replaced with original parts. Circuit diagrams, parts lists and service instructions shall be made available to
qualified technical personnel of the user on request.
We recommend carrying out regular servicing of the instrument and accessories. The defined safety-rel-
evant inspections are given in our manufacturer's recommendations (see Manufacturer’s Recommenda-
tion on page 17).
A.2 Cleaning and Disinfecting Agent
Clean the instrument and accessories regularly with customary chlorine-free and bleach-free cleaning
agents (e.g. a disinfecting agent based on aldehyde). In doing so, make absolutely sure the instrument is
previously switched off!
Clean the instrument with a soft cloth. Make sure that no liquid penetrates into the instrument during the
cleaning!
We recommend that you clean the hand-piece of the tube likewise with a customary disinfecting agent,
e.g. Dismozon®
pur.
Caution
If set up in extremely dirty rooms, the condensers and cooling units must be cleaned of dust
and dirt approx. every 6 months!
Caution
The filter attached to the rear side of the instrument must be cleaned and/or washed every
week!
PHYSIOMED ELEKTROMEDIZIN AG Appendix
FRIGOSTREAM 13
A.3 Transport
In order to avoid all kinds of transportation losses, the shipment of a FRIGOSTREAM must always be un-
dertaken in the original packing envisaged for it. The instrument must not be transported in recumbent
position. However, if this should be the case, the instrument ought not to be started up for another half
an hour.
A.4 Service Life and Disposal
Due to legislation, the service life of this medical product has a limit of sieben years.
The instrument has to be disposed in compliance with the legal obligations. Therefore, pay attention to
the applicable regulations for environment protection.
Please contribute to environment protection by not putting used instruments into your domestic waste.
A.5 Electromagnetic Compatibility
Declaration in accordance with Standard EN 60601-1-2: Electromagnetic Compatibil-
ity
Electronic medical instruments are subject to special precautionary measures in respect of
electromagnetic compatibility.
Portable and mobile HF communication instruments can influence electronic medical instru-
ments.
Appendix PHYSIOMED ELEKTROMEDIZIN AG
14 FRIGOSTREAM
A.6 Technical Data
Protection Class VDE 0750 / IEC 60601 1, Type B
CE Marking in accordance with Directive 93/42/EEC on Medical Devices
Class in accordance with Directive on
Medical Devices (93/42/EEC)
IIa
Mains Supply 220...240 V or 100...120 V
Mains Frequency 50 / 60 Hz
Mains Fuse 2 x T16 A at 220...240 V or 2 x T15 A at 100...120 V
Power Consumption 1,800 VA
Air Flow Rate 1,500 l/min
Power Drain 9 Positions
Length of Treatment Tube 170 cm
Optimum Operating Temperature 15...28°C
Admissible Operating Temperature 10...40°C
Maximum Humidity 65%
Dimensions (W x H x D) 36.0 x 93.0 x 48.0 cm
Weight 53 kg
A.7 Manufacturer Information
Manufacturer
mecoTec GmbH
Mannheimer Straße 14
D-75179 Pforzheim (Germany)
For queries, kindly contact the distributor:
PHYSIOMED ELEKTROMEDIZIN AG
Hutweide 10
D-91220 Schnaittach/Laipersdorf (Germany)
PHYSIOMED ELEKTROMEDIZIN AG Appendix
FRIGOSTREAM 15
Appendix B Scope of Delivery and Accessories
Important
For safety reasons, the instrument is to be used exclusively with original accessories. The use
of other manufacturers' accessories is at the user's risk!
B.1 Scope of Delivery
FRIGOSTREAM is be delivered with the following accessories:
Item No. Identification Quantity
01116 Cryotherapy instrument FRIGOSTREAM 1
00556 Nozzle attachment 5 mm 1
00557 Nozzle attachment 10 mm 1
00558 Nozzle attachment 15 mm 1
00553 Slotted nozzle 1
00554 Tube 1
01639 Operating Instructions (English) 1
B.2 Available Accessories
For FRIGOSTREAM the following accessories are included in the delivery:
Item No. Identification
00556 Nozzle attachment 5 mm
00557 Nozzle attachment 10 mm
00558 Nozzle attachment 15 mm
00553 Slotted nozzle
00298 Supporting arm FRIGOSTREAM incl. fixation (laser version)
00554 Tube
00549 Tube laser version
Scope of Delivery and Accessories PHYSIOMED ELEKTROMEDIZIN AG
16 FRIGOSTREAM
Appendix C Supplementary Documents
C.1 Manufacturer’s Recommendation
PHYSIOMED ELEKTROMEDIZIN AG Supplementary Documents
FRIGOSTREAM 17
Index
A
accessories 16
included 16
ambient temperature 15
application 3
arterial occlusions
arterial 4
arthritides 3
arthritis
rheumatoid 3
arthroses 3
arthrosis humeroscapularis
ccute 3
available
accessories 16
B
bursitides 3
C
CE marking 15
cervical syndrome
acute 3
circulatory disorders
functional 4
cleaning 13
cold urticaria 4
contraindications 4
control elements
function 6
control elements 6
contusions 4
cooling 8
crushing injuries 4
cryoglobulinaemia 4
customer service 13
D
Defrost 10
defrosting 10
dimensions 15
disinfecting agent 13
Display 6
display elements 6
function 6
disposal 14
disturbances
trophic 4
dorsalgias 3
E
electromagnetic
compatibility 14
EMV 14
epicondylitis
acute 3
error Codes 9
F
functional test 9
functions 3
G
gout 3
H
hemipareses 4
I
indications
neurology 4
rheumatology 3
sports medicine 4
surgery 4
inflammations
acute 4
instructions
general 1
instrument description 3
instrument errors 9
instrument overview 2
instrument symbols 1
L
load impedance 15
lumbago
acute 3
lumbar ischialgias 3
luxations 4
lymphedema 4
M
maintenance 13
manufacturer’s recommendation 17
medical fields 3
mode key 6
modulation 15
multiple sclerosis 4
myalgias 3
Index PHYSIOMED ELEKTROMEDIZIN AG
18 FRIGOSTREAM
N
neuritis
acute 4
neurology 4
O
operating conditions 11
operation 8
original packing 14
output frequency 15
output voltage 15
P
parapareses 4 , 4
plus/minus keys 7
power switch 7
pregnancy 4
protection class 15
pulled muscles 4
R
Raynaud’s symptomatology 4
reactive effusions 4
S
sacralgias 3
service 13
service life 14
soft tissue injuries 4
spondylitides 3
sports injuries 4
sports medicine 4
sprains 4
start-up 8
start/stop key 7
storage temperature 10
Sudeck’s syndrome 4
surgery 4
T
technical data 15
tissue lesions 4
transport 14
treatment 11
W
wave form 15
weight 15
PHYSIOMED ELEKTROMEDIZIN AG Index
FRIGOSTREAM 19
Top Related