Count should be 0 (zero) RLU.novaLUM II ATP Detect ion System Without swab in chamber, press RUNnext to EMPTY CHAMBER.

novaLUM® II Monthly Calibration Procedural Flow Chart Refer to Current Charm ® Operator’s Manual for Complete Test Procedure

EMPTY CHAMBER

NEGATIVE CONTROL

POSITIVE CONTROL

RUN

RUN

RUN

PLAN

LOCATION

LIMIT

RESULT

SURFACE

OPERATOR1

OPERATOR2

MISC

NLII CALIB VERFI . . .

EMPTY CHAMBER

0

0

CALIBRATION

JOHN

SAM

TEST 17MAR16 15:54:34

PASS

1 2

EMPTY CHAMBER

NEGATIVE CONTROL

POSITIVE CONTROL

RUN

RUN

RUN

PLAN

LOCATION

LIMIT

RESULT

SURFACE

OPERATOR1

OPERATOR2

MISC

NLII CALIB VERFI . . .

NEGATIVE CONTROL

0

0

CALIBRATION

JOHN

SAM

TEST 17MAR16 15:54:34

PASS

Twist top to act ivate without touching swab t ip. Mix side to side to break white tablet.

Insert swab in chamber of novaLUM II system. Press RUN next to NEGATIVE CONTROL.

Count should be 0 (zero) RLU.

1 2 3

EMPTY CHAMBER

NEGATIVE CONTROL

POSITIVE CONTROL

RUN

RUN

RUN

PLAN

LOCATION

LIMIT

RESULT

SURFACE

OPERATOR1

OPERATOR2

MISC

NLII CALIB VERFI . . .

POSITIVE CONTROL

0

0

CALIBRATION

JOHN

SAM

TEST 17MAR16 15:54:34

PASS

Add 1.0 mL Reagent Water to a clean microtube and add 1 ATP Posit ive Control Tablet. Do not touch the tablet.

Cap and shake vigorously for 10 seconds to disperse tablet. Some undissolved sol ids are acceptable. Let si t for 10 minutes then shake again for 10 seconds.

Withdraw the PocketSwab® Plus swab handle from the body.

Holding the swab horizontal ly, slowly add 20 uL of the posit ive control solut ion to the swab t ip.

Perform PocketSwab Plus swab procedure. Press RUN next to POSITIVE CONTROL.

Results on novaLUM II system should read 10,000 to 40,000 RLU.

1 2 3

4 5 6

EMPTY CHAMBER

NEGATIVE CONTROL

POSITIVE CONTROL