NCIGTNIH National Center for Image-Guided Therapyhttp://www.ncigt.org
System Test for the Integration ofBrainlab, BioImage Suite and Slicer3
Haiying Liu
Surgical Planning Lab
Brigham & Women's Hospital
Harvard Medical School
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
2
National Center for Image Guided TherapyNIH U41RR019703
NEDO Intelligent Surgical Instruments Project
Acknowledgements
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
3
The Team
• Haiying Liu, BWH• Noby Hata, BWH• Alex Golby, BWH• Junichi Tokuda, BWH• Steve Pieper, Isomics• Xenios Papademetris, Yale • Pratik Patel, Brainlab AG
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
4
Outline
• Testing connectivity between Brainlab and BioImage Suite
• Testing connectivity between Slicer3 and BioImage Suite
• Testing the entire system
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
5
Assertion
• Only the setup of the system in last section has been successful can we start to perform the following testing process.
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
6
Test Connectivity between Brainlab and BioImage Suite
If the connection is good, you will see a list images which are available on Brainlab system. Highlight one and then clicking Transfer Image button will get that image from Brainlab to BioImage. The image will display in the VVLink Viewer window (next slide).
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
7
Test Connectivity between Brainlab and BioImage Suite
The image display window.
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
8
Test Connectivity between Brainlab and BioImage Suite
If the Enable button is clicked and when somebody moves the tracking device with the Brainlab system, BioImage will receive real-time tracking data and the image will be reformatted accordingly (next slide).
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
9
Test Connectivity between Brainlab and BioImage Suite
Image reformats as receiving tracking data from Brainlab
Coordinates update as receiving tracking data from Brainlab
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
10
Test Connectivity between Brainlab and BioImage Suite
Make sure all three buttons are checked
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
11
Test Connectivity between Brainlab and BioImage Suite
Clicking this button will send the current image in the VVLink Viewer to Slicer3 (next slide)
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
12
Test Connectivity between Brainlab and BioImage Suite
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
13
Test Connectivity between Brainlab and BioImage Suite
Clicking this button will send the current coordinates in the VVLink Viewer to Slicer3 (next slide)
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
14
Test Connectivity between Brainlab and BioImage Suite
Go to the Transform module to see the coordinates.
Make sure the node name is BioImageSuite
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
15
Testing the entire System
• To test the entire system, make sure:– Enable button is clicked on Slide 8– All three buttons are checked on Slide 10
• Then we will see Slicer3 receives real-time tracking data as the tracking device with Brainlab is moved (next slide)
© NIH National Center for Image-Guided Therapyhttp://www.ncigt.org
16
Testing the entire System
Values will update on a real-time base