8/8/2019 Megha Deep
1/28
MALAVIYA NATIONAL INSTITUE OF TECHNOLOGY,
JAIPUR
SEMINAR ON-
BETA RAY SPECTROMETER
SUPERVISED BY: SUBMITTED BY:
Deptt. of physics Megha Gupta
2009PPH118IIIrd Sem.
8/8/2019 Megha Deep
2/28
Introduction & Properties of beta-rays
Beta ray are positively and negatively charge particle. it
also consists electron and positron.
Beta ray has ionizing nature.
Energy spectrum of beta rays is continuous
They have typically energies in the range of a few
hundred keV to several MeV.
Beta ray are also corpuscular in nature.
8/8/2019 Megha Deep
3/28
BEAT-DECAY
Beta-decay is types of radio active decay, in which betaparticle (electron, positron) is emitted.
The emitted beta particles have a continuous energyspectrum range in (0 to maximum).
It has definate energy limit at point E0, is the max. energy ,it
is also called End-point energy.
Energy continuous spectrum has definate height. it dependson nucleus emitting -particle.
8/8/2019 Megha Deep
4/28
Continuous Energy spectrum ofbeta rays
8/8/2019 Megha Deep
5/28
TYPES OF -DECAY
1. decay (electron emission)-neutron transformed in protonwith emission of eand antineutrino .
n p + e + e
A remains same Z increase by 1
2.+ decay (positron emission)-proton transformed in neutronwith emission of e+ and neutrino.
p n + e+
+ e
A remains same Z decrease by -1
8/8/2019 Megha Deep
6/28
decay and + decay
8/8/2019 Megha Deep
7/28
decay
On a fundamental level, an up quark is convertedinto a down quark, emitting W boson that thendecays into e- andantineutrino.
http://en.wikipedia.org/wiki/File:Beta_Negative_Decay.svg8/8/2019 Megha Deep
8/28
-ray spectrometer
Electron and positron emitted during beta-decay internalconversion electron emitted during the de-excitation ofnuclear levels.
This instrument carry out the energy analysis of electron iscalled ray spectrometer .
8/8/2019 Megha Deep
9/28
Types of -ray spectrometer
1.Magnetic spectrometer
2.Electrostatic spectrometer
3.Scintillation spectrometer
4. Semiconductor spectrometer
8/8/2019 Megha Deep
10/28
8/8/2019 Megha Deep
11/28
Magnetic spectrometer
Momentumof e- detected in spectrometer. it will be exhibit fullwidth at half the height of line.
Ratio of R=(B)/(B)(%) is called resolution(resolving
power)of spectrometer. It denoted as percentage.
It is constant for spectrometer ,if electron follow a fixedgeometry.
Fraction of e- detected & counted by a detector ,this fraction iscalled transmission(t).
8/8/2019 Megha Deep
12/28
Electron conversion spectra in amagnetic spectrometer
8/8/2019 Megha Deep
13/28
Resolution-Resolution of a spectrometer is measure of itsability to resolve two peaks that are close together inenergy.
Resolution=(FHWM)x100/peak channel no. =(E)x100/E
Where,E-full width of peak at half of maximum.
E-based line voltage at the centroid of photo-peak.
Calibration-energy of photo peak of Cs137
corrosponding channel no .
8/8/2019 Megha Deep
14/28
Transverse spectrometer
Velocity of e- perpendicular direction to magnetic field.
Two types-
(1)Semicircular spectrometer(2)Double focusing spectrometer
Semicircular spectrometer produces lower resolution andtransmission.
Double focusing spectrometer produces best resolution .01%to .01%.
8/8/2019 Megha Deep
15/28
Longitudinal spectrometer Velocity of e- parallel direction to magnetic field.
In which Magnetic field obtained by solenoid coils.it is alsocalled long-lens spectrometer. it has R=1%,t=4%.
Disadvantage- Use coil to obtained magnetic field.
Source &detector placed in magnetic field.
To overcome these difficulties, a short lens spectrometer was
developed
8/8/2019 Megha Deep
16/28
Electrostatic spectrometer
Electrostatic spectrometer is modify form of magneticspectrometer.
electrostatic spectrometer mostly used at lower
energies(energy limit 1.5Mev)
8/8/2019 Megha Deep
17/28
Scintillation spectrometer
Electron from an external source normally pass through protective
covering and light reflector.
Detector does not opaque secondary bremsstrahulung photon.
Full energy peak corresponding to energy absorption of incidentof electron energy.
scintillator with low atomic are generally prefer for electron
spectroscopy.
8/8/2019 Megha Deep
18/28
split crystal-spectrometer in which -ray source wassandwiched between a pair of anthracene phosphors.
Scintillator for detection of beta ray spectra Beta source
Collimator
anthrance Scintillator
Photomultiplier
tube
8/8/2019 Megha Deep
19/28
Scintillation spectrometer
Scintillation spectrometer ,in high efficiency causes cannotdistinguish between two coincide pulses.
one proper beta pulse.
Other due to electron conversion.
Advantage-
high deflection efficiency.
Simple structure
Disadvantageit have poor resolution & non linear responsecompared to magnetic or electro magnetic spectrometer.
8/8/2019 Megha Deep
20/28
Electron conversion spectra forplastic scintillator & Cs(TI)
8/8/2019 Megha Deep
21/28
Semiconductor spectrometer Semi-conductor radiation detector used for study of beta ray
and measured electron conversion spectra.
These spectrometer are commonly used at low energy and highenergy detection.
Magnetic -ray spectrometer are preferable to semiconductordetector high energy resolution.
Advantage (i)cheaper
(ii)Si(Li) detector used to measure internal conversioncoefficient for.
(iii)used for large angle geometry & total spectrum can berecorded.
8/8/2019 Megha Deep
22/28
Semiconductor spectrometer
Disadvantage two problem arise for detection of electron.back scattering of e- from the surface of the crystal give to
low energy with conversion lines.
window counter used to so that loss of energy in detector
window kept to minimum.
8/8/2019 Megha Deep
23/28
Electron conversion spectra withSi(Li) detector
8/8/2019 Megha Deep
24/28
Comparison
Ideal spectrometer produces higher transmission and lowerresolution.
Electrostatic spectrometer is modify form of magneticspectrometer. it is also used at lower energies.
Semiconductor spectrometer used to both low energy and highenergy electron detection.
Magnetic beta ray spectrometer are preferable to semiconductordetector for high energy resolution.
scintillation spectrometer have poor resolution compared tomagnetic and electrostatic spectrometer, so it is the best
spectrometer for beta rays is mostly used in practical lab.
O i
8/8/2019 Megha Deep
25/28
ObservationS.No. Channel number Count rate per 10
seconds1 1.0 1514
2 1.1 10903 1.2 8734 1.3 767
5 1.4 9286 1.5 4824
7 1.6 198828 1.7 283529 1.8 13586
10 1.9 186611 2.0 112
12 2.1 5113 2.2 42
14 2.3 3915 2.4 39
8/8/2019 Megha Deep
26/28
Complete spectra of Sr-90
Channel number
8/8/2019 Megha Deep
27/28
Reference
R.M.Singru
Glenn.F.Knoll
Jadish verma
8/8/2019 Megha Deep
28/28
THANKS