Page 1: Ion Chromatography Paper

Ion Chromatography An Instrumental Method

Jordan Sedlock | CHEM 310 | November 25, 2014

Page 2: Ion Chromatography Paper

2

Methods of quantifying ions and polar molecules have been developed and used since the

mid 19th century.1 Ion chromatography (IC) was developed around the 1950s and is used to

separate and quantify ions or ionizable molecules as well as large proteins and small nucleotides

in solution at very low levels.2 It is a form of liquid chromatography and describes efficient

chromatographic peak separation amongst ionic species.2 Ion chromatography has proven to be

very effective at quantifying ionic species in many fields including, but not limited to: food

analysis, water quality treatment, and pharmaceuticals.

Ion chromatography refers to a group of liquid-chromatographic methods that are used to

quantify certain groups of analytes.3 Analytes can include low-molecular-mass acids and bases,

also, inorganic anions and cations.3 Different methods of IC exist and include: ion-exchange,

ion-exclusion, ion-pair, and ion-suppression chromatography.2 Ion-exchange chromatography

(IEC) is one of the most applicable areas of IC, used to separate inorganic anions and cations.3

Ion-exclusion chromatography (also abbreviated IEC) is commonly used to separate weak acids

such as carboxylic acids, weak bases such as ammonia, and carbohydrates as well as other

hydrophilic biomolecules.2 Ion-pair chromatography (IPC) is commonly used to separate

analytes with widely different components as would be the case in mixtures of acidic and basic

analytes as well as compounds that are very difficult to separate, possibly due to presence of

covalent bonds.2 Ion-suppression chromatography (ISC) is generally used to separate polar, weak

acids and weak bases.4

Ion-exchange chromatography (IEC) separation techniques are based on the affinities of

the analytes of interest and the oppositely charged ionic functional groups within the stationary

phase of an ion-exchange column.2 There are two types of exchangers used in IEC, anion and

cation exchangers.5 Anion exchangers contain positively charged functional groups on the

Page 3: Ion Chromatography Paper

3

stationary phase to attract solute anions.5 Cation exchangers contain negatively charged

functional groups on the stationary phase to attract solute cations.5 Of these two types of

exchangers, there are three different classes used for ion-exchange based on particle size. There

are resins, which are used for applications involving small molecules, gels, which are used for

applications involving macromolecules, and inorganic exchangers, which are used for

separations involving extreme chemical conditions such as powerful oxidizing agents or high

temperatures.5 Anion exchanger resins are known to be either strongly basic, containing

quaternary ammonium groups attached to styrene and divinylbenzene copolymers or weakly

basic containing polyalkylamine groups attached to styrene and divinylbenzene copolymers.5

Cation exchanger resins are known to be either strongly acidic, containing sulfonic acid groups

attached to styrene and divinylbenzene copolymers or weakly acidic containing carboxylic acid

groups attached to styrene and divinylbenzene copolymers.5 Gels used in ion-exchange are

commonly made from cellulose and dextran, which are softer materials than polystyrene resins.5

Ion-exclusion chromatography (IEC), another type of IC, is especially useful for

separating organic acids as well as small, neutral molecules and weak bases.6 This technique

involves the separation of molecular species, rather than ions but can be used to separate analytes

by differences in molecular size, shape, or charge.2 This technique also involves what is known

as the Donnan Exclusion mechanism, wherein, certain molecules (depending on size, shape, or

charge) can pass through a semipermeable membrane and onto the stationary phase (or resin

phase) after absorbing into an occluded liquid phase surrounding it and other molecules cannot,

which are therefore, not retained.2 This is useful when separating weak organic acids because

such compounds can easily adsorb onto the resin phase while strong, anionic compounds are

unable.6

Page 4: Ion Chromatography Paper

4

Ion-pair chromatography (IPC) is another type of IC that separates polar and ionic

compounds via a reverse-phase HPLC column rather than an ion-exchange column.5 This

method is commonly detected using UV detectors, because analytes of interest contain

chromaphores, as well as conductivity detectors, because they are universal.7 IPC relies heavily

on the composition of the mobile phase in order to achieve separation.7 The mobile phase

contains an organic solvent as well as an ion-pair reagent.7 The ion-pair reagent contains a large,

ionic molecule whose charge is opposite that of the analyte trying to be analyzed.7 This large

molecule is composed of two different regions, one wherein there is a strongly hydrophobic

segment that interacts with the stationary phase and another region which contains the opposing

charge of the analyte of interest.7 The stationary phase is commonly comprised of

polystyrene/divinylbenzene (PS/DVB).7 The retention mechanism in this separation technique is

not fully understood, however, three theories exist in attempt to explain this: ion pair formation,

dynamic ion exchange, and ion interaction.7 The first theory is similar to reversed-phase

chromatography in that the ion-pair reagent and the analyte of interest form a neutral pair and

then partition between the mobile and stationary phases; retention times can be varied depending

on concentration of organic solvent in the mobile phase.7 The second theory involves the

hydrophobic region of the ion-pair reagent adsorbing to the stationary phase and the analyte of

interest interacting with the ionic region of the ion-pair reagent like it would in ion-exchange

chromatography.7 The ion interaction model involves an electrical double layer that is formed

from the permeation of the stationary phase by the ion-pair reagent which carries with it a

counter-ion; retention is dependent upon factors described in the first two models.7

Page 5: Ion Chromatography Paper

5

Ion-suppression chromatography (ISC), the last type of IC being discussed, is comprised

of two sub-categories: suppressed- ion anion chromatography and suppressed- ion cation

chromatography.5 In ion-suppression chromatography, an unwanted electrolyte is removed

before the analyte of interest undergoes detection via electrical conductivity.5 In the suppressed

anion technique, a mixture of anions is separated using an anion separator column and strong

base as an eluent.5 Anions and eluent are then passed through a suppressor where the strong base,

which possesses high conductivity is converted into H2O, which has low conductivity.5 This

process allows analytes of interest to be detected without interference of eluate.5 In the

suppressed cation technique, a mixture of cations is separated using a cation separator column

and strong acid as an eluent.5 Cations and eluent are then passed through a suppressor where the

strong acid is converted into water, reducing conductivity of the eluate and allowing for accurate

detection of the analytes of interest.5 This technique essentially neutralizes weak acids and bases

which is what causes the “suppression” in ISC.4 ISC is commonly used with reversed-phase

HPLC columns, however, the silica material is not preferred as much as PS/DVB is due to the

small pH range that the silica material possesses (2.5 – 7.5).4 The mobile phase for this technique

contains an organic modifier and a buffer and the retention mechanism is similar to that on a

reversed-phase HPLC column when separating weak acids and bases or neutral molecules.4

Based on the types of samples that are analyzed using IC, sample preparation steps must

be taken in order to ensure sufficient resolution. The most basic rule of sample preparation is that

all samples must be in the aqueous phase as IC is a form of liquid chromatography.8 It is

particularly difficult to quantify compounds that exist at trace levels in matrices with extreme

conditions such as pH or high ionic strength.3 To reduce the amount of an interfering matrix, a

simple solution is to inject a smaller volume onto a column as well as using a selective detection

Page 6: Ion Chromatography Paper

6

method, however, this is problematic because it also reduces the amount of analyte of interest

that is present.3 A second method, which involves handling matrices with extreme pH is to

perform active dialysis (Donnan dialysis), a commonly employed sample cleanup technique

which involves transferring ions of a specific charge across a membrane, for example, with

highly alkaline samples.3 This type of sample cleanup helps reduce peak distortion, baseline

noise, and prolong column life.3 This method can be further refined by adding an electrical

current across the membrane, known as electrodialysis, which only differs from Donnan dialysis

due to the presence of the electrical current.3 Electrodialysis is also helpful when analyzing

alkaline samples which contain trace amounts of inorganic anions as well as neutralizing acids

and bases prior to analysis of trace metals such as magnesium(II) and calcium(II).3

Samples with matrices of high, ionic strength must be prepared in order to account for

matrix effects, which greatly affect resolution and noise in the final chromatogram. A technique

to do this is referred to as matrix elimination IC.3 This technique, simply put, involves using the

dominant ion in the matrix as the eluent in a separation which causes sufficient separation of the

analyte of interest and no retainment of the matrix on the column.3 This method greatly reduces

band broadening in the final chromatogram.3 Another method of reducing matrix effects is to

perform sample cleanup and a matrix eliminating step via solid-phase extraction (SPE).9

Environmental samples, particularly water samples need very little sample preparation in

order to be separated and quantified via IC.9 Drinking water generally only needs to be filtered

through a 0.45 µm filter to remove particulates and is then ready for analysis.9 Wastewater

samples generally only need dilution (to range of 2 – 100 µg/l) and filtering through a 0.45 µm

filter and are then prepared fully.9 Solid samples such as soil can go through very simple aqueous

extractions before being prepared fully.9 Known amounts of sample are mixed with water,

Page 7: Ion Chromatography Paper

7

preferably so as not to introduce extraneous peaks into final chromatogram, but can be mixed

with other solvents such as methanol or weak acids, and filtered through a 0.45 µm filter and are

then ready for analysis.9 Acid digestion in a strongly concentrated acid, such as nitric acid, can

be employed for solid samples such as rock.9 This method is not favorable for IC analysis of

anions because the co-anions that are present from the acid digestion can cause large, distorted

peaks in a final chromatogram.9 This sample preparation method can be employed for analysis of

cyanide, sulfide, and fluoride, however, when coupled with amperometric detection, which

involves applying a voltage between two electrodes positioned in column effluent and measuring

the difference between them.9 Acid digestion is used frequently when using IC to determine

cations; examples include nitrogen, transition metals, and rare earth metals.9

Alkali fusion is an alternative to acid digestion which requires mixing a sample

composed of geological material with an alkaline flux and heated to high temperatures ranging

from 800 – 1100 °C.9 The flux becomes molten which is then cooled and digested in suitable

solution prior to IC analysis.9 This method works well for analyzing fluoride and chloride in

materials of geological origin.9 Another method used for solid samples includes combustion,

wherein, the entire sample is burned in an oxygen flame which leads to conversion of

nonmetallic elements into volatile and gaseous compounds that are collected in an absorbing

solution and then analyzed via IC.9 This method of sample preparation works well for

determining total nitrogen, sulfur, and phosphorous in plant material and different geological,

rocky material.9

After sample preparation, analysis per sample is generally dependent upon which type of

sample is being analyzed, as well as other general parameters such as mobile phase composition,

stationary phase composition, column length, column diameter, flow rate, etc. Analysis is also

Page 8: Ion Chromatography Paper

8

dependent upon which form of IC is used. As previously stated, the four forms of IC are each

used to detect different forms and classes of analytes. Sea water samples can be analyzed for

transition metals in approximately 12 minutes.9 Lake water samples can be analyzed for anions

in approximately 10 minutes.9 Wastewater samples, which are of particular environmental

concern, can be analyzed for anionic and cationic nutrients in approximately 6.5 minutes.10 Weak

acid determination, used in food chemistry, can be performed in approximately 18 minutes.6

Determination of toxic elements such as arsenic, in different types of cereal-based food can be

completed in as little as 7 minutes.11 Separation of proteins from complex biological mixtures,

such as human plasma, can be completed in approximately 20 minutes.12

Ion chromatography analysis is performed on an HPIC instrument that generally costs

approximately $35,000 dollars.13 Thermo Scientific has manufactured an HPIC system designed

to maximize resolution, speed, and sensitivity.13 All chromatography systems consist of the same

basic components: eluent, pump, injection valve, columns, suppressor (which is specific to ion

chromatography), detector, and some sort of data collection system.14 The eluent establishes

basic ionic conditions, stabilizes sample in solution, and promotes progression of sample through

the system.14 Pumps are used to move eluent and samples throughout the system.14 There are

different types of pumps that can be used including serial and parallel, gradient and isocratic.14

Continuous, pulse-free flow is required so as not to introduce noise into the chromatogram.14

Eluent flows from the pump, into the injection site which contains a 2-position valve used to

direct flow of eluent and to introduce the sample into the system.14 Leaving the injection site, the

sample and eluent flow into the column where separation occurs based on the stationary phase

within the column.14 Following the column, the sample will enter a suppressor (selected based

on application).14 There are three types of suppressors: self regenerating suppressor (SRS), Atlas

Page 9: Ion Chromatography Paper

9

Electrolytic Suppressor (AES), and micromembrane suppressor (MMS).14 Suppression is

required so that the ions being analyzed can be detected via one of three different methods:

conductivity, amperometry, or absorbance.14 Readings from the detector appear in the data

collecting system and final chromatograms are produced.14

Many applications have been pursued using instruments and separation techniques of this

nature. IC has proven to be particularly useful in environmental analysis of water samples due to

its ability to detect trace amounts of compounds with high accuracy. Perchlorate, an essential

ingredient in rocket fuel, poses serious threats to human health as it is contaminating drinking

water.15 Perchlorate causes interference of iodide uptake by the thyroid gland and therefore

greatly affects proper thyroid functions.16 An anion exchange column and ion exchange

chromatography can be employed to detect levels of perchlorate in drinking water.16 Separation

of all ions can be completed within 7 minutes.16 It is concluded that this method provides valid

results for determining trace-level perchlorate in drinking water.16

Other methods of determining trace level perchlorate in drinking water have been

employed and compared to IC.17 One such method includes electrospray ionization mass

spectrometry (ESI-MS).18 This technique involves creating ions by applying high voltage to

liquid to create an aerosol.18 It is appropriate to use this method in analysis of permanently

ionized or polar, ionizable compounds.18 Ions are then quantified via mass spectrometric

methods. Unlike IC, ESI-MS requires the employment of a standard addition method to correct

for severe matrix effects in the samples being analyzed and therefore requires more care during

sample preparation.18 ESI-MS has linearity of 10-5 M and a detection limit in reagent water

below 1 µg/l, however, it is almost impossible to detect trace levels of perchlorate in wastewater

samples.18 The IC method has a linear range of 2 – 100 µg/l and a limit of detection of 20 µg/l.16

Page 10: Ion Chromatography Paper

10

High ionic strength of chloride greatly interferes with chromatogram signal, however, this can be

greatly reduced with solid-phase extraction before analysis.16 More training would be required to

perform the ESI-MS method of determining perchlorate in drinking water.

Ion chromatography is a relatively simple technique overall. It is very important in

environmental chemistry and will continue to be as it is so effective at determining trace

elements as well as inorganic anions and cations. Ion chromatography has progressed a long way

from development and has been proven to be a very dependable and accurate analytical

separation technique.

Page 11: Ion Chromatography Paper

11

References

1. Ion Chromatography. http://ionchromatography.co.uk/2011/12/25/the-history-of- ion-

exchange-chromatography/ (accessed Oct 15, 2014). 2. Chromacademy. http://www.chromacademy.com/lms/sco111/theory_of_hplc_ion-

chromatography.pdf (accessed Oct 15, 2014).

3. Haddad, P. R.; Doble, P.; & Macka, M. Developments in sample preparation and separation techniques for the determination of inorganic ions by ion chromatography and

capillary electrophoresis. Journ. Of Chrom. A. 1999, 856, 145-177. 4. Chromedia Analytical Sciences.

http://www.chromedia.org/chromedia?waxtrapp=lcejiGsHqnOxmOlIEcCbCeHhHcC&su

bNav=xododDsHqnOxmOlIEcCbCeHhHcCfC (accessed Oct 15, 2014). 5. Harris, D. C. Chromatograpic Methods and Capillary Electrophoresis. In Quantitative

Chemical Analysis; W. H. Freeman and Company: New York; 2010; 635-646. 6. Milagres, M. P.; Brandao, S. C. C.; Malgalhaes, M. A.; Minim, V. P. R.; & Minim, L. A.

Development and validation of high-perfomance liquid chromatography- ion exclusion

method for detection of lactic acid in milk. Food Chem. 2012, 135, 1078-1082. 7. Dionex. http://www.dionex.com/en-us/webdocs/4696-TN12_LPN0705-01.pdf (accessed

Oct 16, 2014). 8. Koch, W. F. Sample preparation in ion chromatography. Journ. Of Res. 1979, 84, 241-

245.

9. Jackson, P. E. Ion Chromatography in Environmental Analysis. In Encyclopedia of Analytical Chemistry; Meyers, R. A.; Ed.; John Wiley & Sons Ltd: Chichester; 2000;

2779-2801. 10. Karmarkar, S. V. Analysis of wasterwater for anionic and cationic nutrients by ion

chromatography in a single run with sequential flow injection analysis. Journ. Of Chrom.

A. 1999, 850, 303-309. 11. Llorente-Mirandes, T.; Calderon, J.; Centrich, F.; Rubio, R.; & Lopez-Sanchez, J. F. A

need for determination of arsenic species at low levels in cereal-based food and infant cereals. Validation of a method by IC-ICPMS. Food Chem. 2014, 147, 377-385.

12. Gajdoski, M. S.; Kovac, S.; Malatesti, N.; Muller, E.; & Josic, D. Ion-exchange sample

displacement chromatography as a method for fast and simple isolation of low- and high-abundance proteins from complex biological mixtures. Food Technol. Biotechnol. 2014,

52(1), 58-63. 13. Thermo Scientific. http://www.thermoscientific.com/en/product/dionex- ics-5000-

capillary-hpic-system.html (accessed Oct 27, 2014).

14. Dionex. Principles and Troubleshooting Techniques in Ion Chromatography. Dionex Corporation: 2002; Document No. 034461.

15. Jackson, P. E.; Gokhale, S.; Streib, T.; Rohrer, J. S.; & Pohl, C. A. Improved method for the determination of trace perchlorate in ground and drinking waters by ion chromatography. Journ. Of Chem. A. 2000, 888, 151-158.

16. Jiang, S.; Li, Y.; & Sun, B. Determination of trace level perchlorate in Antarctic snow and ice by ion chromatography coupled with tandem mass spectrometry using an

automated sample on-line preconcentration method. Chin. Chem. Lett. 2013, 24, 311-314.

Page 12: Ion Chromatography Paper

12

17. Li, Y.; Whitaker, J. S.; & McCarty, C. L. Analysis of iodinated haloacetic acids in drinking water by reversed-phase liquid chromatography/electrospray ionization/tandem

mass spectrometry with large volume direct aqueous injection. Journ. Of Chrom. A. 2012, 1245, 75-82.

18. Roehl, R.; Slingsby, R.; Avdolovic, N.; & Jackson, P. E. Applications of ion chromatography with electrospray mass spectrometric detection to the determination of environmental contaminants in water. Journ. Of Chrom. A. 2002, 956, 245-254.