Spectrum Analyzer
GSP-930
USER MANUALREVISION 1.1 APRIL 2012
ISO-9001 CERTIFIED MANUFACTURER
7/27/2019 Gsp 900 Manual Final
2/255
This manual contains proprietary information, which is protected bycopyright. All rights are reserved. No part of this manual may bephotocopied, reproduced or translated to another language withoutprior written consent of Good Will company.
The information in this manual was correct at the time of printing.However, Good Will continues to improve products and reserves therights to change specification, equipment, and maintenanceprocedures at any time without notice.Good Will Instrument Co., Ltd.No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
7/27/2019 Gsp 900 Manual Final
3/255
SAFETY INSTRUCTIONS
1
Table of ContentsSAFETY INSTRUCTIONS .................................................. 3 GETTING STARTED .......................................................... 8
GSP-930 Introduction ............................ 9Accessories .......................................... 11Appearance .......................................... 13First Use Instructions .......................... 24
BASIC OPERATION ........................................................ 36Frequency Settings ............................... 39Span Settings ....................................... 43Amplitude Settings .............................. 46Autoset ................................................ 59Bandwidth/Average Settings ................ 61Sweep .................................................. 66
Trace .................................................... 72Trigger ................................................. 79Marker ................................................. 84Display................................................. 98System Settings ................................. 105Preset ................................................ 109
ADVANCED OPERATION ...... ...... ...... ...... ...... ...... ...... ..... 111Measurement ..................................... 112Limit Line Testing .............................. 159Sequence ........................................... 165Tracking Generator ............................ 170Power Meter ...................................... 173
FILE ............................................................................. 178File Overview ..................................... 178
7/27/2019 Gsp 900 Manual Final
4/255
GSP-930 User Manual
2
REMOTE CONTROL ...................................................... 194Interface Configuration ...................... 195
FAQ .............................................................................. 205APPENDIX .................................................................... 206
Replace the Clock Battery ................... 206Glossary of Acronyms ........................ 207GSP-930 Default Settings ................... 209Menu Tree .......................................... 211 GSP-930 Specifications ...................... 240GSP-930 Dimensions ......................... 249Declaration of Conformity .................. 250
INDEX .......................................................................... 251
7/27/2019 Gsp 900 Manual Final
5/255
SAFETY INSTRUCTIONS
3
SAFETY INSTRUCTIONSThis chapter contains important safetyinstructions that you must follow duringoperation and storage. Read the following beforeany operation to insure your safety and to keepthe instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on theinstrument.
WARNINGWarning: Identifies conditions or practices thatcould result in injury or loss of life.
CAUTIONCaution: Identifies conditions or practices thatcould result in damage to the instrument or toother properties.
DANGER High Voltage
Attention Refer to the Manual
Earth (ground) Terminal
Frame or Chassis Terminal
Do not dispose electronic equipment as unsortedmunicipal waste. Please use a separate collection
facility or contact the supplier from which thisinstrument was purchased.
7/27/2019 Gsp 900 Manual Final
6/255
GSP-930 User Manual
4
Safety Guidelines
GeneralGuideline
CAUTION
Do not place any heavy object on theinstrument.
Avoid severe impact or rough handling thatleads to damaging the instrument.
Do not discharge static electricity to theinstrument.
Use only mating connectors, not bare wires, forthe terminals.
Ensure signals to the RF input do not exceed
+30dBm. Ensure reverse power to the TG output terminal
does not exceed +30dBm.
Do not supply any input signals to the TGoutput.
Do not block the cooling fan opening.
Do not disassemble the instrument unless you
are qualified.(Measurement categories) EN 61010-1:2010 specifies themeasurement categories and their requirements as follows. Theinstrument falls under category II.
Measurement category IV is for measurement performed at thesource of low-voltage installation.
Measurement category III is for measurement performed in thebuilding installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed oncircuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range: 100V~240V
Frequency: 50/60Hz
To avoid electrical shock connect the protectivegrounding conductor of the AC power cord toan earth ground.
7/27/2019 Gsp 900 Manual Final
7/255
SAFETY INSTRUCTIONS
5
Battery
CAUTION
Rating: 10.8V, 6 cell Li-ion battery
Turn off the power and remove the power cordbefore installing or removing the battery.
Cleaning Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of milddetergent and water. Do not spray any liquid.
Do not use chemicals containing harsh materialsuch as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below) Temperature: 5C to 45C
Humidity:
7/27/2019 Gsp 900 Manual Final
8/255
GSP-930 User Manual
6
Disposal Do not dispose this instrument as unsortedmunicipal waste. Please use a separate collectionfacility or contact the supplier from which thisinstrument was purchased. Please make sure
discarded electrical waste is properly recycled toreduce environmental impact.
7/27/2019 Gsp 900 Manual Final
9/255
SAFETY INSTRUCTIONS
7
Power cord for the United Kingdom
When using the instrument in the United Kingdom, make sure thepower cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHEDIMPORTANT: The wires in this lead are coloured in accordance with thefollowing code:Green/ Yellow: EarthBlue: NeutralBrown: Live (Phase)
As the colours of the wires in main leads may not correspond withthe coloured marking identified in your plug/appliance, proceedas follows:
The wire which is coloured Green & Yellow must be connected tothe Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminalwhich is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to theterminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipmentor contact the supplier.
This cable/appliance should be protected by a suitably rated andapproved HBC mains fuse: refer to the rating information on theequipment and/or user instructions for details. As a guide, a cableof 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on theconnection method used.
Any exposed wiring from a cable, plug or connection that isengaged in a live socket is extremely hazardous. If a cable or plug isdeemed hazardous, turn off the mains power and remove the cable,any fuses and fuse assemblies. All hazardous wiring must beimmediately destroyed and replaced in accordance to the abovestandard.
7/27/2019 Gsp 900 Manual Final
10/255
GSP-930 User Manual
8
GETTING STARTEDThis chapter provides a brief overview of theGSP-930, the package contents, instructions forfirst time use and an introduction to the frontpanel, rear panel and GUI.
GSP-930 Introduction ...................................................... 9Main Features .................................................................................................................9Accessories..................................................................... 11Package Contents ....................................................................................................... 12
7/27/2019 Gsp 900 Manual Final
11/255
GETTING STARTED
9
GSP-930 IntroductionThe GSP-930 is the most advanced spectrum analyzer GW Instek
has produced to date. The GSP-930 features a split window displayto view data in spectrum, topographic or spectrographic views.
Main Features
Performance 9kHz~3GHz bandwidth
1Hz resolution
Nominal RBW accuracy of 5% 750kHz
Video bandwidth 1Hz~1MHz (10 steps)
Amplitude measurement range: DANL~30dBm(frequency dependent)
Input attenuation: 0 ~ 50dB
Phase noise: < -88dBc/Hz@1GHz, 10kHz
Features 10%-step increments for RBW bandwidth
Three display modes: Spectrum, Topographicand Spectrographic
Split window display
Built-in EMI filter
Auto Wake-up
Built-in preamplifier
Gate sweep
Marker Frequency counter
Two operating modes: Spectrum and PowerMeter mode
SEM measurement
ACPR measurement
OCBW measurement
7/27/2019 Gsp 900 Manual Final
12/255
GSP-930 User Manual
10
Channel power measurement
Demodulation analyzer
Diverse marker functions and features with
Peak Table Sequence function to automatically perform pre-
programmed sequential operations
Optional battery operation
Interface 8.4 color LCD (800600)
On-screen menu icons
DVI-I video output RS-232 with RTS/CTS hardware flow control
USB 2.0 with support for USB TMC
LAN TCP/IP with LXI support
Optional GPIB/IEEE488 interface
IF output @ 886MHz
Headphone output
REF (reference clock) input/output BNC ports
Alarm/Open collector output BNC port
Trigger/Gate input BNC ports
RF N-type input port
Tracking generator output
DC +7V/500mA output SMB port
7/27/2019 Gsp 900 Manual Final
13/255
GETTING STARTED
11
Accessories
Standard
Accessories
Part number Description
Region dependant User manual
Region dependant Power cord
Options Option number Description
Opt1. Tracking generator
Opt2. Battery (11.1V/5200mAH
Li-ion battery)Opt3. GPIB interface (IEEE 488
bus)
OptionalAccessories
Part number Description
PWS-06 USB Average Power Sensor(up to 6200 MHz;-32 to 20 dBm)
GRA-415 6U Rack mount kit
7/27/2019 Gsp 900 Manual Final
14/255
GSP-930 User Manual
12
Package Contents
Check the contents before using the GSP-930.
Opening the box
Contents(single unit)
Main unit(may include optionalGPIB, TG output)
Quick Start manual
User Manual CD
Power cord x1 (regiondependent)
Optional battery pack
Calibration certificate
7/27/2019 Gsp 900 Manual Final
15/255
GETTING STARTED
13
Appearance
GSP-930 Front Panel
Frequency
Autoset
Amplitude
Span
BW/Avg
Display
Trace
Sweep
Meas
Trigger
Sequence
File
Recall
Save
Peak
Marker->
Marker
System
Mode
Preset
Search
QuickSave
LimitLine
7
4
1
0
8
5
2
9
6
3
/ Enter
BK SP
F4
F3
F2
F1
F6
F5
F7
GSP-930 9 kHz 3 GHzSpectrum Analyzer
LOCAL
RF INPUT50
REV PWR +30dBm
TG OUTPUT50 W W500mA MAX.
DC 7VOUTPUT +30dBm MAX.
MAX.DC 50V MAX.DC 50V
CONTROL FILE
M AR KER S TA TE
LCDdisplay
Mainkeys
Functionkeys
Controlkeys Power key
File keys
State keys
Scroll wheel
Arrow keys
RF inputterminal
DC powersupply
Trackinggeneratoroutput
Numeric,Enter andBK SP keys
USB A,Micro SDport
Marker keys
LCD display 800600 color LCD display. The display shows thesoft keys for the current function, frequency,amplitude and marker information.
Function keysF 1
~ F 7The F1 to F7 function keys directlycorrespond to the soft keys on theright-hand side of display.
Main keysFrequency
Sets the center frequency, startfrequency, stop frequency, centerfrequency step and frequency
offset values.
7/27/2019 Gsp 900 Manual Final
16/255
GSP-930 User Manual
14
Span
Sets the span, with options for fullspan, zero span and last span.
Amplitude
Sets the amplitude reference level,attenuation, pre-amplifiercontrols, scale and other optionsfor attenuation and scale.
Autoset
Automatically searches the peaksignal with maximum amplitudeand displays it with appropriate
horizontal and vertical scales.
Control keysBW/Avg
Sets the resolution bandwidth,video bandwidth, average typeand turns the EMI filter on/off.
Sweep
Sets the sweep time and gate time.
Trace
Sets traces and trace relatedfunctions.
Display
The Display key configures thewindowing mode and basicdisplay properties.
Meas
Accesses measurement options
such as ACPR, OCBW,demodulation measurements,SEM, TOI and other advancedmeasurements.
LimitLine
Sets and tests Pass/Fail limit lines.
Sequence
Access, set and edit program
sequences.
7/27/2019 Gsp 900 Manual Final
17/255
GETTING STARTED
15
Trigger
Sets the triggering modes.
File File
File utilities options
QuickSave
The Quick Save utility allows youto save either the state, trace,screen limit line, correction orsequence with only a single press.
Save
Save the trace, state etc., and saveoptions.
Recall
Recall the trace, state etc., andrecall options.
MarkerMarker
Turns the Markers on/off andconfigures the markers.
Marker->
TheMarker-> key positions themarkers on the trace.
PeakSearch
Finds each maximum andminimum peak. Used with theMarker function.
StatePreset
LOCAL
The Preset key will restore the
spectrum analyzer to the Factoryor User-defined settings.
The Preset key will also return theinstrument back to local controlafter it has been in remote controlmode.
7/27/2019 Gsp 900 Manual Final
18/255
GSP-930 User Manual
16
Mode
TheMode key sets the spectrumanalyzer to either Spectrum orPower Meter mode.
System
The System key shows systeminformation, settings and othersystem related functions.
Power key Turns the instrument on/off.
Scroll wheel Edit values, select listed items.
Arrow keys Increment/decrement values (insteps), select listed items.
RF input terminal
RF INPUT 50W
+30dBm MAX.MAX.DC 50V
RF input port. Accepts RF inputs.
Maximum input: +33dBm
Input impedance: 50
Maximum DC voltage: 50V
N-type: female
DC power supply
500mA MAX.
DC 7VOUTPUT
SMB port supplies power foroptional accessories.
DC +7V
500mA Max.
7/27/2019 Gsp 900 Manual Final
19/255
GETTING STARTED
17
Numeric keypad7 8 9
654
1 2 3
/0
BK SP
Enter
The numerickeypad is used toenter vales and
parameters. It isoften used inconjunction withthe arrow keys andscroll wheel.
TG output port
REV PWR +30dBm
TG OUTPUT 50 W
MAX.DC 50V
The Tracking Generator (TG)
output source. N-type: female
Input impedance: 50
Output power: -50dBm to 0dBm
Maximum reversed power:+30dBm
USB A, Micro SD USB A port, Micro SD port forsaving/recalling settings/files.
7/27/2019 Gsp 900 Manual Final
20/255
7/27/2019 Gsp 900 Manual Final
21/255
GETTING STARTED
19
Battery pack Voltage: 10.8VCapacity: 5200mAH
REF IN BNC female reference input.
REF OUT BNC female reference output:10MHz, 50 impedance
Security Lock
ALARM OUT BNC female open collector Alarmoutput.
TRIG IN/GATE IN BNC female 3.3V CMOS triggerinput/gated sweep input.
Phone 3.5mm stereo headphone jack(wired for mono operation)
USB B USB B Device port. USB 1.1/2.0
LAN RJ-45 10Base-T/100Base-Tx
7/27/2019 Gsp 900 Manual Final
22/255
GSP-930 User Manual
20
Display
Scale andReferencelevel
Attenuationlevel
Dateand time Function
menu
Markerinformation
Triggersettings
Pre-ampsettings
Trace andDetectionsettings
Frequency/Bandwidthsettings
Status icons
Sweepsettings
Traces and
waveforms
LXI icon
Entry /Messagearea
Reference level Displays the reference level. For details, see page46.
Attenuation Displays the vertical scale (attenuation) of theinput signal. For details, see page 47.
Date/Time Displays the date and time. See page 106 for
details.
Markerinformation
Displays marker information. For details see page83.
LXI icon This icon indicates the status of the LXI connection.For details, see page 196.
Function menu Soft menu keys associated with the F1 to F7function keys to the right of the display.
7/27/2019 Gsp 900 Manual Final
23/255
GETTING STARTED
21
Sweep settings Sweep icon that shows the sweepstatus. See page 66 for details.
Trigger settings Trigger icon that shows the triggerstatus. See page 79 details.
Pre-amp settings Pre-amplifier icon that shows thePre-amplifier status. See from page48 for details.
Trace anddetection settings
Trace icon that shows the trace typeand the detection mode used foreach trace. See from page 72 fordetails.
Status Icons Displays the interface status, power source status,and alarm status, etc. See the Status Icon Overviewon page 22 for a list of the status icons.
Frequency/Bandwidthsettings
Displays the Start, Center and Stop frequencies,RBW, VBW, Span and Sweep settings.
Entry/Messagearea
This area is used to show system messages, errorsand input values/parameters.
Trace and
waveforms
Main display showing the input signals, traces
(page 72), limit lines (159) and marker positions(83).
7/27/2019 Gsp 900 Manual Final
24/255
GSP-930 User Manual
22
Status Icon Overview
PreAmp Indicates that the pre amplifier ison.
AC Shown when running on ACpower.
AC Charge Shown when the AC power ischarging the battery.
Alarm Off Alarm buzzer output is currentlyoff.
Alarm On Alarm buzzer output is currentlyon.
Amplitude Offset Indicates that the amplitude-shift is
active. This icon appears whenamplitude-related functions areused:Reference level offsetAmplitude CorrectionInput Z = 75Input Z cal >0
Battery indicator
~
Indicates the battery charge.
BandwidthIndicator
Indicates that the RBW or VBWsettings are in manual mode.
Average Indicates that the Average functionis active.
7/27/2019 Gsp 900 Manual Final
25/255
GETTING STARTED
23
External Lock Indicates that the system is nowlocked and refers to the externalreference input signal
External Trigger External trigger signal is beingused.
Math Trace math is being used.
SequenceIndicator
Shown when a sequence is running.
Sweep Indicator Indicates that the sweep time ismanually set.
Trackinggenerator
Indicates the tracking generator isturned on.
TG Normalization Indicates that the tracking generator
has been normalized.
Wake-up clock Indicates that the wake-up clock isturned on.
USB Indicates that a USB flash drive isinserted into the front panel and isrecognized.
Micro SD Indicates that a micro SD card isinserted into the front panel and isrecognized.
7/27/2019 Gsp 900 Manual Final
26/255
GSP-930 User Manual
24
First Use InstructionsUse the procedures below when first using the GSP-930 to tilt the
stand, insert the battery pack, power up the instrument, setting theinternal clock, the wake-up clock, updating the firmware andrestoring the default settings. Lastly, the Conventions sections willintroduce you to the basic operating conventions used throughoutthe user manual.
Tilting the Stand
Description The GSP-930 has two adjustable rubber feetthat can used to position the instrument intotwo preset orientations.
Upright Position Tuck the feet underthe bottom of theinstrument to stand
the instrumentupright.
Leaning Position Pull the feet back to
have the instrumentleaning back.
7/27/2019 Gsp 900 Manual Final
27/255
GETTING STARTED
25
Inserting the Battery Pack
Description The GSP-930 has an optional battery pack. Thebattery should be inserted before power isconnected to the AC power socket and beforethe unit is turned on.
Steps 1. Ensure the power is off and the AC power isdisconnected.
2. Remove the battery cover.
3. Insert the battery as shown in the diagrambelow.
4. Replace the battery cover.
Display Icon The battery icon is displayed when GSP-930 is running on battery power.
Insertion Diagram
7/27/2019 Gsp 900 Manual Final
28/255
GSP-930 User Manual
26
Power UP
Steps 1. Insert the AC power cord into the powersocket.
2. The power button exterior will be lit blue toindicate that the GSP-930 is in standby mode.
3.
Press the power button to turn the GSP-930 on.
4. The power button will turn orange and theGSP-930 will start to boot up.
Note
It takes approximately 1.5 minutes for the GSP-930
to fully startup.
7/27/2019 Gsp 900 Manual Final
29/255
GETTING STARTED
27
Power Down
Description The GSP-930 has two methods to power down:Normal and Forced Power Down.
The normal power down method will save thesystem state and end any running processes.The state is saved for the next time theinstrument is turned back on.
The forced power down method only does aminimum state save.
Normal PowerDown
Press the power button. The system willautomatically handle the power downprocedure in the following order:
The system state is saved.
Outstanding processes are closed in
sequence.
The LCD backlight is turned off.
The system enters standby mode (thepower key changes from orange toblue).
NoteThe process takes ~10 seconds.
Forced PowerDown
Press and hold the power button for ~4seconds until the system turns off and thepower button turns blue.
NoteThe forced down mode might cause the GSP-930 to perform a longer system check the nexttime it is powered up.
7/27/2019 Gsp 900 Manual Final
30/255
GSP-930 User Manual
28
Setting the Date, Time and Wake-Up Clock
Description The GSP-930 can be setup to power-upautomatically using the Wakeup Clockfunction. This feature is useful to wake-up theinstrument early and eliminate settling time.
System Date Example: Set the System Date to March 1, 2012
1. Press System >Date/Time[F4]>Set
Date[F1]>Year[F1].
2. Press 2012>Enter[F1].
3. PressMonth[F2]>3>Enter[F1].
4. Press Day[F3]>1>Enter[F1].
5. Press Return[F7].
NoteThe System Date will be shown at the top of thedisplay.
System Time Example: Set the System Time to 9.00 AM
1. Press System >Date/Time[F4]>SetTime[F2]>Hour[F1].
2. Press 9>Enter[F1].
3. PressMinute[F2]>0>Enter[F1].
4. Press Second[F3]>0>Enter[F1].
5. Press Return[F7].
7/27/2019 Gsp 900 Manual Final
31/255
GETTING STARTED
29
NoteThe System Time will be shown at the top of thedisplay.
System Wake-UpClock
Example: Set the GSP-930 to wake up at 9.00AM
1. Press System >Date/Time[F4]>Wake-UpClock[F3]>Clock[F1].
2. Press Clock[F1] to choose a clock (1 ~ 7).
3. Press State[F2] to turn the clock on/off.
4. Press Hour[F3]>9>Enter[F1].
5. PressMinute[F4]>0>Enter[F1].
6. Press [F5] and choose Rept. (Repeat) or Single.
7. Press Select Date[F6] and select a day.
8. Press Return[F7] to save the Wake-Up Clocksettings.
NoteThe system time is kept with the CR2032 clockbattery. If the system time/ wake up clock can nolonger be set, please replace the clock battery. Seepage 206.
7/27/2019 Gsp 900 Manual Final
32/255
GSP-930 User Manual
30
Firmware Update
Description The GSP-930 allows the firmware to beupdated by end-users. Before using the GSP-930, please check the GW Instek website or askyour local distributor for the latest firmware.
System version Before updating the firmware, please check thefirmware version.
1. Press System >System Information[F1].
2. The firmware will be listed on the display.
Firmware: T. 1. 0. 0
3. Press any other key to exit out of the SystemInformation screen.
4. To upgrade the firmware, insert the newfirmware onto a USB flash drive or Micro SDcard and put the drive/ card into theappropriate front panel port. The firmwarefiles should be located in a directory namedgsp930.
5. Press System >More 1/2[F7]>Upgrade[F3].
7/27/2019 Gsp 900 Manual Final
33/255
GETTING STARTED
31
6. The spectrum analyzer will automatically findthe firmware on the USB flash drive and startto update the firmware. When finished, the
message Upgrade is finished will be shownat the bottom of the screen followed byRebooting.
Rebooting
7.
The system will automatically restart after therebooting message.
NoteThe upgrade process may take a few minutes.
Restoring Default Settings
Description The factory default settings or user-definedpresets can be easily restored using the Presetkey on the front panel. By default, the factorydefault settings are restored with the Presetkey.
For details on how to configure the presetsettings, please see page 109.
Steps 1. Press Preset .
2. The spectrum analyzer will load the presetsettings.
7/27/2019 Gsp 900 Manual Final
34/255
GSP-930 User Manual
32
Conventions
The following conventions are used throughout the user manual.
Read the conventions below for a basic grasp of how to operate theGSP-930 menu system and front panel keys.
Soft Menu keys The F1 to F7 function keys on the right side ofthe display correspond directly to the soft-menu keys on their left.
Frequency
Autoset
Amplitude
Span
BW/Avg
Display
Trace
Sweep
Meas
Trigger
Sequence
File
Recall
Save
Peak
Marker->
Marker
System
Mode
Preset
Search
QuickSave
LimitLine
7
4
1
0
8
5
2
9
6
3
/ Enter
BKSP
F4
F3
F2
F1
F6
F5
F7
GSP-930 9 kHz 3 GHzSpectrum Analyzer
LOCAL
R FIN P U T50
REVPWR +30dBm
TGOU TP UT 50W W500mAMAX.
D C 7VO UT PU T + 30 dB m M AX.
MAX.DC 50V MAX.DC50V
CONTROL FILE
M AR KE R S TA TE
Soft-menu keys
F1 ~ F7 function keys
Input ParameterValues
Selecting this type of menu key will allow youto enter a new value with the numeric keypador increment/decrement the value using thescroll wheel.
Toggle State
Pressing this menu key will toggle the state.
7/27/2019 Gsp 900 Manual Final
35/255
GETTING STARTED
33
Toggle State &Input Parameter
Pressing this menu key will allow you to togglethe state of the function between Auto andMan(ual) state. When in the Man state, theparameter value can be manually edited. Usethe numeric keypad to enter the new value oruse the scroll wheel to increment/decrementthe current value.
Sub Menu
Pressing this menu key will enter a submenu.
Sub Menu to
select parameter
Pressing this menu key will enter a submenu toselect a parameter.
Active Function
Pressing this type of menu key will activatethat function. The menu key will behighlighted to show it is the active function.
7/27/2019 Gsp 900 Manual Final
36/255
GSP-930 User Manual
34
Parameter input Numerical keypadScroll wheel
7
4
1
0
8
5
2
9
6
3
/ Enter
BK SP
Directional arrow keysBackspace, Enter keys
Parameter values can be entered using thenumeric keypad, the scroll wheel andoccasionally with the arrow keys.
Using the numeric
keypad
When prompted to enter a parameter, use the
number keys (0~9), the decimal key (.) and thesign key (+/-) to enter a value. After a valuehas been entered, the soft-menu keys can beused to select the units.
The value of the parameter is shown at thebottom of the screen as it is edited.
Edited parameter
Span: 1.5
Back Space Use the backspace key to delete the lastcharacter or number entered.
7/27/2019 Gsp 900 Manual Final
37/255
GETTING STARTED
35
Using the scrollwheel
Use the scroll wheel to alter the current value.Clockwise increases the value, anti-clockwisedecreases the value.
Directional arrows Use the directional arrows to select discreteparameters or to alter values by a coarserresolution than the scroll wheel. Left decreasesthe value, right increases the value.
7/27/2019 Gsp 900 Manual Final
38/255
GSP-930 User Manual
36
BASIC OPERATIONFrequency Settings .......................................................... 39Center Frequency ....................................................................................................... 39Start and Stop Frequency .......................................................................................... 40Center Frequency Step .............................................................................................. 41Frequency Offset ........................................................................................................ 42Span Settings ................................................................. 43Span .............................................................................................................................. 43Full Span ...................................................................................................................... 44Zero Span .................................................................................................................... 44Last Span ...................................................................................................................... 45Amplitude Settings ......................................................... 46Reference Level .......................................................................................................... 46
Attenuation .................................................................................................................. 47Scale/Div ..................................................................................................................... 48Scale Type .................................................................................................................... 48
View Scale .................................................................................................................... 49Vertical Scale Units .................................................................................................... 50Reference Level Offset .............................................................................................. 50
Amplitude Correction ................................................................................................ 51Create a Correction Set ............................................................................... 52
Amplitude Correction On/Off .................................................................. 54Delete Correction Set .................................................................................. 55Save Correction Set To Memory ............................................................... 55Recall Correction Set From Memory ........................................................ 56
Input Impedance ........................................................................................................ 56Input Impedance Calibration ................................................................................... 57Using the Built-in Pre-Amplifier ............................................................................. 57Autoset .......................................................................... 59Using Autoset ............................................................................................................. 59Limiting the Autoset Vertical Search Range .......................................................... 60
7/27/2019 Gsp 900 Manual Final
39/255
BASIC OPERATION
37
Limiting the Autoset Horizontal Search Range .....................................................60Bandwidth/Average Settings ........................................... 61Resolution Bandwidth Setting (RBW) .....................................................................61
Video Bandwidth Settings (VBW) ...........................................................................62
VBW/RBW Ratio .......................................................................................................62Average Trace ..............................................................................................................63Average Type ...............................................................................................................65EMI Filter .....................................................................................................................66Sweep ............................................................................ 66Sweep Time ..................................................................................................................66Single Sweep .................................................................................................................67Continuous Sweep ......................................................................................................68
Gated Sweep Overview ..............................................................................................68Using the Gated Sweep Mode ..................................................................................70Trace ............................................................................. 72Selecting a Trace ..........................................................................................................72
Trace Math ...................................................................................................................74Trace Detection Mode ...............................................................................................75Trigger ........................................................................... 79Selecting a Trigger Type .............................................................................................79
Free Run Mode ..............................................................................................79Activate Video Trigger .................................................................................79Activate External Trigger .............................................................................81
Selecting the Trigger Mode .......................................................................................82Set the Trigger Delay Time .......................................................................................83Marker ........................................................................... 84
Activating a Marker.....................................................................................................85
Activate a Normal Marker ...........................................................................85Move Marker Manually ................................................................................86Move Marker to Preset Locations ..............................................................86
Activate Delta Marker ..................................................................................87Move Delta Marker(s)Manually ..................................................................88
Marker Functions ........................................................................................................89Marker Noise .................................................................................................89Frequency Counter .......................................................................................90
Move Marker to Trace ...............................................................................................91Show Markers in Table ..............................................................................................92Peak Search ..................................................................................................................93
7/27/2019 Gsp 900 Manual Final
40/255
GSP-930 User Manual
38
Move Marker to Peak .................................................................................. 93Move Marker and Peak to Center ............................................................. 93Search for Peaks ........................................................................................... 94Peak Configuration ...................................................................................... 95
Peak Table ................................................................................................................... 96Display ........................................................................... 98
Adjusting the LCD Brightness ................................................................................. 98Turning the LCD Backlight Off .............................................................................. 98Setting a Display Line (Reference Level Line) ...................................................... 99Using the Video Out Port ......................................................................................... 99Setting the Display Mode ........................................................................................ 100Split Spectrum View ................................................................................................ 103System Settings ............................................................ 105System Information ................................................................................................. 105Error Messages ......................................................................................................... 105Set the System Language ......................................................................................... 106Set the Date and Time ............................................................................................. 106Using the Wake-Up Clock ...................................................................................... 107
Alarm Output ............................................................................................................ 108Preset ........................................................................... 109Using the Preset Key ............................................................................................... 109Save the User Preset Settings ................................................................................. 109Preset Type Settings ................................................................................................. 110Power on Preset Settings ........................................................................................ 110
7/27/2019 Gsp 900 Manual Final
41/255
BASIC OPERATION
39
Frequency Settings
Center Frequency
Description The center frequency function sets the centerfrequency and centers the display to the centerfrequency.
Operation 1. Press Frequency >Center[F1] and enter the
frequency and unit.
Range:Resolution:Default
0kHz~3GHz1Hz1.5GHz
Display Center frequency
Set Center Frequency
7/27/2019 Gsp 900 Manual Final
42/255
GSP-930 User Manual
40
Start and Stop Frequency
Description The start/stop frequency function will set thestart and stop frequency of the span.
Operation 1. To set the start frequency, press Frequency > Start
Freq[F2] and enter the frequency and unit.
2. To set the stop frequency, press Frequency > Stop
Freq[F3] and enter the frequency and unit.
Range:Resolution:Default Start frequency:Default Stop frequency:
0kHz~3GHz1Hz0Hz3GHz
Display Start Frequency
Start Frequency Stop Frequency
Stop Frequency
NoteThe start and stop frequency can change when thespan settings are used.The stop frequency must be set higher than the
start frequency (for spans 0), otherwise the span
will be automatically set to 100Hz.
7/27/2019 Gsp 900 Manual Final
43/255
BASIC OPERATION
41
Center Frequency Step
Description The CF Step function sets the step size of thecenter frequency when using the arrow keys.
When the arrow keys are used to alter thecenter frequency, each press will move thecenter frequency by the step size specified bythe CF Step function.
In auto mode, the center frequency step size isequal to 10% (1 division) of the span.
Operation 1. Press Frequency > CF Step[F4] and set the CF Step to
Auto or Man.
2. If Man was selected, set the frequency and unitof the center frequency step size.
Manual Range:Auto range:
100Hz~3GHz1/10 of span frequency
Display
CF Step size
7/27/2019 Gsp 900 Manual Final
44/255
GSP-930 User Manual
42
Frequency Offset
Description The Freq Offset function allows you to add anoffset to the Center, Start and Stop frequenciesas well as the marker frequencies. The offsetvalue does not affect displaying the trace on thedisplay.
Operation 1. Press Frequency >Freq Offset[F5] and set the offset
value.
The Center, Start, Stop and Marker frequenciesare updated accordingly.
Offset Range: 0Hz~100GHz
Display
Frequency
Offset
7/27/2019 Gsp 900 Manual Final
45/255
BASIC OPERATION
43
Span Settings
Span
Description The Span function will set the frequency rangeof the sweep. The sweep will be centeredaround the center frequency.
Setting the span will alter the start and stopfrequencies.
Operation 1. Press Span > Span[F1] and enter the span
frequency range and unit.
Range:Resolution:Default Span:
0kHz~3GHz1Hz3GHz
Display Span
Set Span
7/27/2019 Gsp 900 Manual Final
46/255
GSP-930 User Manual
44
Full Span
Description The Full Span function will set the span to thefull frequency range.
This function will set the start and stopfrequencies to 0Hz and 3GHz respectively.
Operation 1. Press Span > Full Span[F2].
Zero Span
Description The Zero Span function will set the frequencyrange of the sweep to 0Hz and fixes the startand stop frequencies to the center frequency.The Zero Span function measures the timedomain characteristics of the input signal at the
center frequency. The horizontal axis isdisplayed in the time domain.
Operation 1. Press Span > Zero Span[F3].
The span changes accordingly.
7/27/2019 Gsp 900 Manual Final
47/255
BASIC OPERATION
45
Display Time domain
0Hz Span
Example: Amplitude modulation
NoteThe measurement functions such as TOI, SEM,CNR, CTB, CSO, ACPR, OCBW, phase, Jitter andNdB are not available with the zero span setting:
Last Span
Description The last span function returns the spectrumanalyzer to the previous span settings.
Operation 1. Press Span > Last Span[F4].
7/27/2019 Gsp 900 Manual Final
48/255
GSP-930 User Manual
46
Amplitude SettingsThe vertical display scale is defined by the reference level
amplitude, attenuation, scale and external gain/loss.
Reference Level
Description The reference level defines the absolute level ofthe amplitude on the top graticule in voltage orpower.
Operation 1. Press Amplitude > Ref Level[F1] and enter the
reference level amplitude and unit.
Range:Units:Resolution:
-120dBm ~ 30dBmdBm, W, V, dBmV, dBuV1dBm
DisplayReference LevelRef Level reading
7/27/2019 Gsp 900 Manual Final
49/255
BASIC OPERATION
47
Attenuation
Description The attenuation of the input signal level can setto automatic (Auto) or manual (Man). When theattenuation is set to Man, the input attenuatorcan be changed manually in 1dB steps.
Operation 1. Press Amplitude >Attenuation[F2] and select Auto
or Man.
2. If Man was selected, enter the attenuation leveland unit.
Range:Units:Resolution:
0dBm ~ 50dBmdBm1dB
Display Attenuation level
7/27/2019 Gsp 900 Manual Final
50/255
GSP-930 User Manual
48
Scale/Div
Description Sets the logarithmic units for the verticaldivisions when the scale is set to Log.
Operation 1. Press Amplitude > Scale/Div[F3] repeatedly to select
the vertical division units.
Unit Range: 10, 5, 2, 1
Display Scale
Note The Scale/Div function is only selectable when thescale is set to Log (logarithmic).
Scale Type
Description Sets the vertical scale in linear or logarithmicunits.
By default the linear scale is set to volts and thelogarithmic scale is set to dBm.
Operation 1. Press Amplitude > Scale Type[F4] and set the vertical
scale to Log or Lin.
NoteIf the unit scale is changed (i.e. dBm volts), thedisplayedvertical scale type will remain in the set
linear or logarithmic setting.
7/27/2019 Gsp 900 Manual Final
51/255
BASIC OPERATION
49
View Scale
Description The Scale function turns the vertical scale
on/off. The value of each graticule division isdisplayed with same units used for the RefLevel settings.
Operation 1. Press Amplitude > Scale[F5] to toggle the Scale on or
off.
2. Press Scale Pos.[F6] to toggle the position of thescale when on.
Scale position: Left, Center, Right
Display
ScaleLeft
The vertical scale is displayed on the left hand sideby default.
7/27/2019 Gsp 900 Manual Final
52/255
GSP-930 User Manual
50
Vertical Scale Units
Description Change the vertical units for both linear orlogarithmic scales.
Operation 1. Press Amplitude >More[F7]>Y Axis[F1] and then
choose the desired units.
2. The units are changed accordingly.
Units: dBm, dBmV, dBuV,Watts, Volts
Reference Level Offset
Description The Reference Level Offset function sets anoffset value to the reference level to compensate
for any loss or gain from an external network ordevice.The offset value does not affect the inputattenuation or the on-screen trace.This setting will change the reference levelreadout, the scale readout and the markerreadout.
Operation 1.
PressAmplitude
>More[F7]>RefLvlOffset[F2] and setthe offset level and unit.
2. To remove the offset level, set the referenceoffset to 0 dB.
Range: 0dB ~ 50dB
Display Icon The AMP icon is displayed at the bottom
of the screen.
7/27/2019 Gsp 900 Manual Final
53/255
BASIC OPERATION
51
Example: Ref: 0dBm
Before reference level offset(offset: 0dB)
Ref: 10dBmReference level
offset: 10dB
After reference level offset (offset: 10dB)
Amplitude Correction
Description Amplitude correction adjusts the frequencyresponse of the spectrum analyzer by alteringthe amplitudes at specified frequencies. Thisallows the spectrum analyzer to compensate forloss or gain from an external network or deviceat certain frequencies.
Range Correction Sets: 5 sets of 30 points
Amplitude: -40dB to +40dBAmplitude Resolution: 0.1dBFrequency: 9kHz to 3GHzFrequency Resolution: 1Hz
7/27/2019 Gsp 900 Manual Final
54/255
GSP-930 User Manual
52
Display Correction points
Original waveform
Correctedoutput
Example: The diagram above shows how amplitude
correction is used to compensate for any losses orgains at specific frequencies.
Create a Correction Set
Description The GSP-930 can create and edit up to 5 sets ofcorrection points. The correction points andassociated values are all tabulated for ease ofuse.
Operation 1. Press Amplitude >More[F7]>Correction[F3]>
Correction Set[F1] and choose a correction set toedit/create.
Correction set: 1~5
7/27/2019 Gsp 900 Manual Final
55/255
BASIC OPERATION
53
2. Press Edit[F3].
The GSP-930 will split into two screens. The top
screen will show the waveform and the bottomscreen will provide an overview of thecorrection points.
Spectrum display
Correction points
Frequency ofselected point
3. Press Point Num[F1] and choose a point numberto edit.
Point Num: 1~30
4. Press Frequency[F2] and choose the frequency of
the selected point.Press Gain Offset[F3] and choose the amplitudeof the select point. The units will be the same asthose used for the vertical scale.
The frequency of the point values displayed inthe correction table on the bottom display.
7/27/2019 Gsp 900 Manual Final
56/255
GSP-930 User Manual
54
Correction Table
5. Repeat steps 2 to 4 for any other correctionpoints.
6. To delete the selected point, press DeletePoint[F6].
7. Press Return[F7]>Save Correction[F5] to save thecorrection set.
NoteNote that the correction points are automaticallysorted by frequency (low high).The correction set must be saved before it can beturned on.
The frequency values displayedin the correctiontable are rounded down for display purposes only.The actual frequency for each point can be seen inthe Frequency soft-key.
Amplitude Correction On/Off
Description Any one of the 5 correction sets can be turnedon.
ActivateCorrection
1. Press Amplitude >More[F7]>Correction[F3]>
Correction Set[F1] and choose a correction set.
Correction Set: 1~5
2.
Press Correction[F2] and toggle correction on.
7/27/2019 Gsp 900 Manual Final
57/255
BASIC OPERATION
55
DeactivateCorrection
1. Press Amplitude >More[F7]>Correction[F3]>
Correction[F2] to turn correction back off.
Delete Correction Set
Operation 1. Press Amplitude >More[F7]>Correction[F3]>
Correction Set[F1] and choose the correction setto delete.
Correction Set: 1~5
2. Press DeleteCorrection[F6].The selected correction set will be deleted.
Save Correction Set To Memory
Operation 1. Press Save > Save To[F1] and choose the save
location.
Location: Register, Local, USB, SD
2. Press Type[F2]> Correction[F5].
3. Press Data Source[F3] and choose a correction.
Correction Set: Correction 1~5
4. Press Save Now[F7].
5. The correction set will be saved to the selectedlocation. For more information on Save andRecall, please see page 170.
7/27/2019 Gsp 900 Manual Final
58/255
GSP-930 User Manual
56
Recall Correction Set From Memory
Operation 1. PressRecall
> Recall[F1] and choose the recalllocation:
Location: Register, Local, USB, SD
2. Press Type[F2]> Correction[F5].
3. Press Destination[F3] and choose a correctionset.
Correction Set: Correction 1~5
4. Press Recall Now[F4].
5. The current correction set will be recalled fromthe selected location. For more information onSave and Recall, please see page 170.
Input Impedance
Description Sets the input impedance to 75 or 50.
Operation 1. Press Amplitude >More[F7]>Input Z[F4] to toggle
the input impedance.
Range: 75, 50
7/27/2019 Gsp 900 Manual Final
59/255
BASIC OPERATION
57
Input Impedance Calibration
Description When an external impedance converter module(optional accessory ADP-101) is used to convertthe impedance of a device from 50 to 75,some external loss can be induced. The Input ZCal function can be used to compensate forthese losses with an offset value.
Note
The Input Z Cal function is only available when the
input impedance is set to 75.
Operation 1. Press Amplitude >More[F7]>Input Z Cal[F5] and set
the impedance offset.
Range:Resolution:
0dB to +10dB1dB
Display Icon The AMP icon is displayed at the bottomof the screen when Input Z Cal0dB.
Using the Built-in Pre-Amplifier
Description The built-in pre-amplifier boosts weak inputsignals, such as EMI testing signals, to levels
that are easy to handle, over the entirefrequency range. The built-in pre-amplifier onthe GSP-930 has a nominal gain of 20dB.
In the Auto setting, the pre-amplifier will beautomatically turned on when the referencelevel is less than -30dBm. When the referencelevel is greater than -30dBm, the pre-amplifieris turned off.
The bypass setting turns the pre-amplifier off.
7/27/2019 Gsp 900 Manual Final
60/255
GSP-930 User Manual
58
Operation 1. Press Amplitude >More[F7]>Preamp[F6] to toggle
the Preamp state.
Range: Auto, Bypass
Display Icon The Pr-amp icon indicates that the preamplifier is on.
Example: Pr-amp icon
NoteWhen the pre-amplifier is on, the attenuatorbecomes fixed at 0dB (i.e. Attenuation = 0dB).
7/27/2019 Gsp 900 Manual Final
61/255
BASIC OPERATION
59
AutosetThe Autoset function searches the peak signals in two stages (full
span & 0Hz - 100MHz limited span), picks the signal peak with themaximum amplitude, and then shows it in the display.
Using Autoset
Operation 1. Press Autoset >Autoset[F1].
Autoset Range Amplitude:Span:
-80dBm ~ +20dBm100Hz ~ 3GHz
Example:
Before Autoset, preset state
After Autoset
7/27/2019 Gsp 900 Manual Final
62/255
GSP-930 User Manual
60
NoteRBW, VBW and sweep settings are reset to Autowhen the Autoset function is used.
Limiting the Autoset Vertical Search Range
Description You can set the amplitude floor so that thesignals lower than the setting will be ignoredby the Autoset search.
Operation 1. Press Autoset >Amp.Floor[F2] and switch the
range from Auto to Man.
2. Enter the amplitude limit and unit for theAutoset search.
Range: 80 to +20dBm
NoteSee page 48 for setting the amplitude units.
Limiting the Autoset Horizontal Search Range
Description You can change the frequency span limit in thedisplay to get a better view of the Autosetresult. By default, the frequency span afterAutoset is set at 3MHz.
Operation 1. Press Autoset > Span[F3] and switch the range
from Auto to Man.
2. Enter the span frequency for the Autosetsearch.
Manual Range: 100Hz to 3GHz
7/27/2019 Gsp 900 Manual Final
63/255
BASIC OPERATION
61
Bandwidth/Average SettingsBW/Avg key sets the resolution bandwidth (RBW), video
bandwidth (VBW) and averaging functions. The resolution, sweeptime, and averaging are in a trade-off relationship, so configurationshould be done with care.
Resolution Bandwidth Setting (RBW)
Description RBW (Resolution Bandwidth) defines the width
of the IF (intermediate frequency) filter that isused to separate signal peaks from one another.The narrower the RBW, the greater thecapability to separate signals at closefrequencies. But it also makes the sweep timelonger under specific frequency spans (thedisplay is updated less frequently).
Operation 1. Press BW/Avg > RBW[F1] and set the RBW to
Auto or Man.
2. Set the resolution bandwidth and unit for Manmode.
Mode:Frequency Range(3dB):
Frequency Range(6dB):
Auto, Man10Hz~3kHz (1-3-10 step)10kHz~1MHz (10% step)
200Hz, 9kHz, 120kHz
Display Icon The BW icon is displayed at the bottom ofthe screen when the RBW is in Man mode.
NoteIf the RBW settings have an asterisk (*), itindicates that the -6dB filters are used.
7/27/2019 Gsp 900 Manual Final
64/255
GSP-930 User Manual
62
Video Bandwidth Settings (VBW)
Description VBW (Video Bandwidth) defines thesmoothness of the trace on the display.Combined with RBW, VBW defines the abilityto sort out the target signal from surroundingnoise or adjacent peaks.
Operation 1. Press BW/Avg > VBW[F2] and set the VBW to
Auto or Man.
2. Set the video bandwidth and unit for Manmode.
Mode:Frequency Range(3dB):
Auto, Man1Hz~1MHz (1-3-10 step)
Display Icon The BW icon is displayed at the bottom of
the screen when the VBW is in Man mode.
VBW/RBW Ratio
Description The VBW/RBW function is used to view theratio between the video bandwidth and theresolution bandwidth.
The VBW/RBW ratio is altered by setting theRBW and or VBW settings, see page 61 & 62respectively.
View VBW/RBWratio
1. Press BW/Avg .
2. The ratio is displayed on the VBW/RBW[F3] softkey.
7/27/2019 Gsp 900 Manual Final
65/255
BASIC OPERATION
63
Display
VBW/RBWRatio
Tip Signals that are masked by the noise floor levelshould have a ratio of less than 1 to smooth thenoise out.
Signals with strong frequency components shoulduse a ratio equal to or greater than 1.
Average Trace
Description The Average function averages the trace for auser-defined number of times before it is
displayed. This feature smoothes the noiselevel, but has the drawback of slowing downthe display update rate.
Operation 1. Press BW/Avg >Average[F4] and toggle Average
on or off.
2. Set the number of averages.
Range:Default:
4 ~ 20020
Display Icon The AVG icon is displayed at the bottom ofthe screen when the Average function ison.
7/27/2019 Gsp 900 Manual Final
66/255
GSP-930 User Manual
64
Example:
Average:Off
Average: On (20)
7/27/2019 Gsp 900 Manual Final
67/255
BASIC OPERATION
65
Average Type
Description The Average Type function determines how theGSP-930 determines the average value.
LOG Average: Averages the trace points on alogarithmic scale.
Volt Average: Averages the amplitudes of thetrace points on a linear voltage scale.
Power Average: Averages the trace points on alogarithmic scale in watts.
Operation 1. Press BW/Avg >Average Type[F5] and choose the
average type.
Range:
Default:
LOG Power, VoltAverage, Power AverageLOG Power
7/27/2019 Gsp 900 Manual Final
68/255
GSP-930 User Manual
66
EMI Filter
Description The built-in EMI filter is used for specificmeasurement situations such as EMI averagedetection, where a higher level of sensitivity isrequired than the standard configuration.When turned on, the RBW is set to -6dB,indicated by an asterisk (*).
When any measurement functions are turned
on (see page 111 for details), the EMI filter isautomatically disabled. Conversely if the EMIfilter is turned on, any measurement functionsare turned off.
Operation 1. Press BW/Avg > EMI Filter[F6] and toggle EMI
filter on or off.
SweepThe GSP-930 has a number of sweep options including setting thesweep time and sweep mode(continuous, single). The GSP-930 alsohas gated sweep modes.
Sweep Time
Description Sweep time defines the length of time thesystem takes to "sweep the current frequencyspan. Note, however, that sweep time andRBW/VBW are in a trade-off. Faster sweeptimes update the display more frequently butmake RBW and VBW wider, reducing thecapability to separate signals at closefrequencies.
7/27/2019 Gsp 900 Manual Final
69/255
BASIC OPERATION
67
Operation 1. Press Sweep > Sweep Time[F1] and toggle the
Sweep time to Auto or Man.
2. Set the sweep time for the Man mode.
Mode:Range:
Resolution:
Auto, Man22ms ~ 1000s (span>0Hz)50us ~ 1000s (span=0Hz)10us
Display Icon The SWT icon will be displayed at the
bottom of the screen when in the sweep isin manual mode.
Single Sweep
Description The single sweep function is used to perform asingle sweep. When Sweep Single is pressed theGSP-930 will perform a single sweep and thenstop.
Operation 1. Press Sweep > Sweep Single[F2] to put the
spectrum analyzer into single sweep mode.
2. Press Sweep Single[F2] again to perform a singlesweep.
Display Icon The Sweep Single icon is displayed on theright-hand side of the screen when thesweep is in single mode.
NoteYou must wait for the single sweep to finish beforepressing the Single Sweep key again.
If a setting is changed whilst the spectrum analyzer
is still sweeping, the single sweep will immediatelystart over.
7/27/2019 Gsp 900 Manual Final
70/255
GSP-930 User Manual
68
Continuous Sweep
Description The GSP-930 has two main sweeping modes:single and continuous. Use the continuousmode to have the sweep constantly updated.
Operation 1. Press Sweep > Sweep Cont[F3] to put the
spectrum analyzer into continuous sweepmode.
Display Icon The Sweep Cont icon is displayed on theright-hand side of the screen when thesweep is in continuous mode.
NoteThe GSP-930 will now continuously sweep unlessthe mode is changed to single sweep mode or ifthe system is waiting for a trigger condition.
Gated Sweep Overview
Description The Gated Sweep mode allows a trigger signalto dictate when the spectrum analyzer cansweep. This mode is useful for characterizingsignals that are pulsed on and off, such as RFburst transmissions or for measuring spuriousnoise levels between transmission bursts.
Overview 1. The trigger signal must be synchronized to theperiod of the input signal (shown as RF burstbelow).
2. The start of the gate time is produced from thepositive or negative edge of the trigger signal +the delay time.
3. The end of the gate time is determined by the
7/27/2019 Gsp 900 Manual Final
71/255
BASIC OPERATION
69
set gate length.
4. The gated sweep should not be positioned ateither end of the transmission.
Delay Length
Gate
Trigger
RF Burst
period
Example: The diagram above demonstrates therelationship between the input trigger, the inputsignal and the position of the gated sweep relativeto the input signal.
NotePlease take into consideration RBW settling time.Setting the delay time too short may not leaveenough time for the RBW filter to resolve.
7/27/2019 Gsp 900 Manual Final
72/255
GSP-930 User Manual
70
Using the Gated Sweep Mode
Connection 1. Connect a trigger signal (3.3v CMOS) to theGATE IN port on the rear panel.
Trigger
Operation 1. Press Sweep > GateDelay[F5] and set the gate
delay time.
2. Press Sweep > Gated Length[F6] and set the gate
time length.
3. Press Sweep > Gated Sweep[F4] and turn the
mode on.
Gate Delay:Gate Length:
0s ~ 1000s10us ~ 1000s
Display Icon The Sweep Gated icon is displayed whenGated Sweep is turned on.
7/27/2019 Gsp 900 Manual Final
73/255
BASIC OPERATION
71
Example: The example below shows the spectrum of an FSKmodulated signal when gated sweep mode is off.
The example below shows the same signal with thegated sweep timed to sweep when only the desiredfrequency is output.
Gated sweep icon
NoteGate Delay and Gate Length must first be setbefore Gated Sweep is turned on.
7/27/2019 Gsp 900 Manual Final
74/255
GSP-930 User Manual
72
TraceThe GSP-930 is able to set the parameters of up to 4 different
trances on the display at once. Each trace is represented by adifferent color and is updated with each sweep.
Selecting a Trace
Description Each trace (1, 2, 3, 4) is represented by adifferent color. When activated, an icon for each
trace color and function is shown to the left ofthe display. When a trace is selected,parameters can be set/edited from the tracemenu.
Trace Color: 1: Yellow2: Pink3: Blue4: Red
Trace Type The type of trace used determines how the tracedata is stored or manipulated before beingdisplayed. The analyzer updates each traceaccording to the type of trace used.
Clear andWrite
The GSP-930 continuouslyupdates the display with each
sweep.
Hold Max/Hold Min
The maximum or minimumpoints are maintained for theselected trace. The trace pointsare updated each sweep if newmaximum or minimum pointsare found.
7/27/2019 Gsp 900 Manual Final
75/255
BASIC OPERATION
73
View View will hold the selectedtrace and stop updating thetrace data for the selected trace.
Pressing View[F4] will displaythe trace data that was clearedusing the Blank[F5] key.
Blank Clears the selected trace fromthe display and stores tracedata. The trace data can berestored by pressing View[F4].
Display Icon
Clear & Write
Hold Max
View
Blank
Trace 1
Trace 2
Trace 3
Trace 4
Operation 1. Press Trace > Trace[F1] and choose the trace
number.
Trace: 1, 2, 3, 4
2. Select the trace type:
Clear & Write[F2]Max Hold[F3]Min Hold[F4]View[F5]Blank[F6]
Note Traces, 2, 3 and 4 are set to Blankby default.
7/27/2019 Gsp 900 Manual Final
76/255
GSP-930 User Manual
74
Trace Math
Description Performs trace math from two traces (TR1, TR2)and stores the result in the currently selectedtrace. It also performs trace shift.
Math functions Power Diff Subtracts the TR1 amplitudedata from the TR2 amplitudedata. The TR1 data TR2 dataare converted to watts. The
result is converted back todBm.
Log Diff Subtracts the TR1 amplitudedata from the TR2 amplitudedata and then adds alogarithmic reference. Both theTR1 and TR2 data is in dBm.The resultant trace of thesubtraction is in dB. When theresult is added to a logarithmicreference the resulting data isin dBm.
LOG Offset Adds a reference to the TR1trance
Operation 1. Press Trace >More[F1]>Trace Math[F1].
2. Press TR1[F1] and select the first trace source:
TR1: Trace 2, Trace 3, Trace 4
3. Press TR2[F2] and select the second tracesource:
TR2: Trace 2, Trace 3, Trace 4
7/27/2019 Gsp 900 Manual Final
77/255
BASIC OPERATION
75
4. Select the trace math function:
PowerDiff[F3]
LogDiff[F4]LogOffset[F5]
5. If LogDiff was selected, set the reference leveland unit.
LogDiff ref range:LogDiff ref units:
-120dBm ~ 30dBmdBm, W
6. If LogOffset was selected, set the offset leveland unit.
LogOffset range: -50dB~+50dB
7. To turn trace math off, press the OFF[F6].
Display Icon The Math icon is displayed when tracemath is turned on.
Trace Detection Mode
Description Each time the spectrum analyzer samples datafor each point on the trace, a number of samples
are usually taken for each point, known as asample bucket. The actual value of each point isdetermined by the detector from the samples ineach bucket.
Each selected trace, (1, 2, 3, 4), can use adifferent detection mode.
7/27/2019 Gsp 900 Manual Final
78/255
GSP-930 User Manual
76
Detection modes Auto Automatically chooses anappropriate mode based on thevalues of all the samples.
Normal While the signal level is constantlyincreasing or decreasing, thepositive peaks are detected.Otherwise, detecting modeswitches between positive peakand negative peaks. Useful forpicking up burst phenomenon
while avoiding excessive noise.
Peak+ Detects positive peak signals byselecting the maximum peak valuefor each point from each bucket.This mode is useful for sinusoidalsignals.
Peak- Detects negative peak signals by
selecting the lowest peak value foreach point from each bucket. Thismode is not recommended foramplitude measurement.
Sample Randomly selects a value from thebucket sample. Useful for noisesignals.
Average Calculates the average of all thesamples in the sample bucket.
7/27/2019 Gsp 900 Manual Final
79/255
BASIC OPERATION
77
Auto DetectorSelection Method
Below is a flow chart diagram showing theDetector selection for the Auto mode.
Is (trace) Average
on?
Are any traces set
to Max Hold?
Are any traces setto Min Hold?
Is Average Type
set to Power?
AverageSample
Peak+
Peak-
Normal
Y
N
Y
N
Y
N
N Y
Is Phase Jitter
measurement on?Sample
Y
N
Are any windows
set to Topographic?Peak+
Y
N
Detector(Auto mode)
Operation 1. Press Trace >More[F7]>Detection[F2].
2. Select the trace detection mode for the selected
trace:
Auto[F1]Normal[F2]Peak+[F3]Peak-[F4]Sample[F5]Average[F6]
7/27/2019 Gsp 900 Manual Final
80/255
GSP-930 User Manual
78
3. The display will return the Trace menu.
Display Icon Normal Average icon
Peak+ icon Peak - icon
Sample icon Average icon
7/27/2019 Gsp 900 Manual Final
81/255
7/27/2019 Gsp 900 Manual Final
82/255
GSP-930 User Manual
80
Parameters Video Edge: Determines the polarity of thevideo trigger.
Positive: The signal voltageexceeds the video level at thetrigger frequency.
Negative: The signal voltage islower than the video level at thetrigger frequency.
Video Level:The trigger voltage level.
TriggerFrequency:
Sets the frequency to starttriggering
Operation 1. Press Trigger >Trigger Condition[F2]>Video[F1]
2. Press Video Edge[F1] and choose the edge.
Range: Positive, Negative
3. Press Video Level[F2] and set the video voltagetrigger level.
Trigger level: (-120dBm to +30dBm)+Ref Level Offset
4. Press Trigger Freq[F3] and choose the frequencyat which the spectrum analyzer will check thetriggering conditions.
Frequency: 0-3GHz+frequency offset
NoteSet the trigger back to Free Run to disable thevideo trigger.
7/27/2019 Gsp 900 Manual Final
83/255
BASIC OPERATION
81
Activate External Trigger
Description The external trigger is used when an externaltrigger signal is input into the rear panel TRIGIN port. The external trigger signal can beconfigured as positive or negative edge.
Trigger: 3.3V, CMOS
Operation 1. Press Trigger >Trigger Condition[F2]>Ext.Edge[F2]
and select the trigger edge:
Pos:Neg:
Positive edgeNegative edge
2. Connect the external trigger signal to the rearpanel TRIG IN port.
Trigger
3. PressAction Now[F5] to activate the externaltrigger.
4. The system will now wait the trigger conditionsto be matched before starting a sweep.
Display Icon The EXT Trigger icon is displayed when theexternal trigger is activated.
NoteThe trigger will revert back to the Free Run mode ifany parameter settings are changed, such as thespan or amplitude settings.
7/27/2019 Gsp 900 Manual Final
84/255
GSP-930 User Manual
82
Selecting the Trigger Mode
Description In free run mode all signals are captured andthe trigger conditions are not used.
Modes Normal: The spectrum analyzer capturesevery signal that meets the triggerconditions.
Single: The spectrum analyzer captures
the first signal that meets thetrigger conditions.
Continuous: The spectrum analyzer capturesthe first signal that meets thetrigger conditions then switches tofree run mode thereafter.
Operation 1.
PressTrigger
> Trigger Mode[F3] to toggle thetrigger mode:
Nor.:Sgl.:Cont.:
NormalSingleContinuous
2. PressAction Now[F5] to manually starttriggering.
7/27/2019 Gsp 900 Manual Final
85/255
BASIC OPERATION
83
Set the Trigger Delay Time
Description Sets the delay time between when the analyzertriggers and when the analyzer begins tocapture the signal.
Delay time range: 1ns to 1ks
Operation 1. Press Trigger >Trigger Delay[F4] and set the
trigger delay time.
Delay range: 0~1000s
7/27/2019 Gsp 900 Manual Final
86/255
GSP-930 User Manual
84
MarkerA Marker shows the frequency and amplitude of a waveform point.
The GSP-930 can activate up to 6 markers or marker pairssimultaneously as well as up to 10 peak markers in the marker table.
The marker table helps editing and viewing multiple markers in asingle display.
A delta marker shows the frequency and amplitude differencesbetween the reference marker.
The GSP-930 can automatically move a marker to various locationsincluding the peak signal, center frequency, and start/stop
frequency. Other marker operations regarding signal peaks areavailable in the Peak Search function.
Activating a Marker from page85
Move Marker Manually from page86
Move Marker to Preset Locations from page86
Activate Delta Marker from page87
Move Delta Marker(s)Manually from page 88
Marker Functions from page89
Move Marker to Trace from page91
Show Markers in Table from page92
Peak Search from page93
Peak Configuration from page95
Peak Table from page96
7/27/2019 Gsp 900 Manual Final
87/255
BASIC OPERATION
85
Activating a Marker
There are two basic marker types, normal markers and delta
markers. Normal markers are used to measure the frequency/timeor amplitude of a point on the trace. Delta markers are used tomeasure the difference between a reference point and a selectedpoint on the trace.
Activate a Normal Marker
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
Marker: 1~6
2. Press [F2] to turn the selected marker on.
3. Press Normal[F3] to set the selected marker tothe Normal type.
4. The display will show the marker on the trace(centered by default) with the markermeasurement at the top of the display.
Marker
Maker ID, Frequency, Amplitude
7/27/2019 Gsp 900 Manual Final
88/255
GSP-930 User Manual
86
Move Marker Manually
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. Use the left/right arrow keys tomove the marker one grid division.
3. Use the scroll wheel to move themarker in fine increments.
4. Alternatively, the numeric keypadcan be used to directly enter thefrequency of the marker position.
7 8 9
654
1 2 3
/0
Move Marker to Preset Locations
Description The Marker-> key is used to move the selected
marker to a number of preset positions.
Functions Mkr>Center:Mkr>Start:Mkr>Stop:Mkr>CF Step:Mkr>Ref Lvl:
Move to center frequency.Move to start frequency.Move to stop frequency.Move to step frequency.Move to reference level
amplitude.
NoteWhen the Marker-> key is used, the span and other
settings may be automatically changed.
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. Press Marker-> and select a marker position:
7/27/2019 Gsp 900 Manual Final
89/255
7/27/2019 Gsp 900 Manual Final
90/255
GSP-930 User Manual
88
Move Delta Marker(s)Manually
Move Delta orReference Marker 1. PressMarker
> Delta[F4]>Move[F2]>Move Ref[F2]to move the reference marker.
2. Press Marker > Delta[F4]>Move[F2]>Move
Delta[F3] to move the Delta marker.
3. Move the selected marker in the same fashionas a normal marker, see page 86
Move Bothreference anddelta marker
1. Press eitherMove Pair Span[F4] orMove PairCenter[F5] to move both markers at the sametime.
Move Pair Span: Sets the frequency spanbetween both markers.The span can be positiveor negative:
1+span1
1-span1
Move Pair Center: Moves both markers atthe same time, keepingthe span between both
markers eventhroughout.
2. Move both markers in the same fashion as anormal marker, see page 86
7/27/2019 Gsp 900 Manual Final
91/255
BASIC OPERATION
89
Marker Functions
Marker Noise
Description The noise marker function calculates theaverage noise level over a bandwidth of 1Hz,referenced from the marker position.
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. Press [F2] to turn the selected marker on.
3. Press Normal[F3] and then position the markerto the desired location.
4. Press Function[F5]>Marker Noise and turnMarker Noise on.
5. The display will show the noise levelmeasurement at the top of the screen indBm/Hz.
Marker ID, Frequency, dBm/Hz
Marker
7/27/2019 Gsp 900 Manual Final
92/255
GSP-930 User Manual
90
Frequency Counter
Description The frequency counter function is used to makeaccurate frequency measurements.
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. Press [F2] to turn the selected marker on.
3. Press Normal[F3] and then position the markerto the desired location.
4. Press Function[F5]>Frequency Counter[F1] andturn the counter function on.
5. Press Resolution[F2] and set the resolution:
Auto: Automatically chooses
the best resolution.Man: Allows the resolution to
be manually set.
Man Range: 1Hz, 10Hz, 100Hz, 1kHz
6. The display will show the frequencymeasurement at the top of the screen at the
selected resolution.
Frequency
Marker
7/27/2019 Gsp 900 Manual Final
93/255
BASIC OPERATION
91
Move Marker to Trace
Description The Marker Trace function moves the selectedmarker to any of the currently active traces.
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. Press [F2] to turn the selected marker on.
3. PressMore 1/2[F7]>Marker Trace[F1] and choosea trace to move the current marker to. Onlyactive traces can be selected.
Auto[F1]Trace1[F2]Trace2[F3]Trace3[F4]Trace4[F5]
4. In the example below, marker 1 is set to Trace1and marker 2 is set to Trace2.
Marker 1, Trace 1
Marker 2, Trace 2
7/27/2019 Gsp 900 Manual Final
94/255
GSP-930 User Manual
92
Show Markers in Table
Description The GSP-930 has a Marker Table function to
show all the active markers and measurementsat once.
Operation 1. Press Marker >More 1/2[F7]>Marker Table[F2] and
turn the marker table on.
2. The display will split into two screens. Thebottom half will show the Marker Table withthe marker ID(normal, reference or delta), trace,x-axis position (frequency/time) and theamplitude of the marker.
Marker Table
7/27/2019 Gsp 900 Manual Final
95/255
BASIC OPERATION
93
Peak Search
Move Marker to Peak
Description ThePeak
Search key is used to find trace peaks.
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. PressPeak
Search >Peak Search[F1]. The marker will
move to the highest signal peak.
3. To continually search for the peak each sweep,
press,Peak
Search >More 1/2[F7]>Peak Track[F1] and
set Peak Track to on.
Move Marker and Peak to Center
Description The Center function moves the marker to thehighest signal peak and moves the centerfrequency to that peak.
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. PressPeak
Search >Mkr>Center[F2].
NoteThe span will not be changed.
7/27/2019 Gsp 900 Manual Final
96/255
GSP-930 User Manual
94
Search for Peaks
Description ThePeak
Search key can be used to search for anumber of different peaks.
Peak Search Next Peak: Searches for next highestpeak visible on the display.
Next Peak Right: Searches for the next peak tothe right of the marker.
Next Peak Left: Searches for the next peak tothe left of the marker.
Min Search: Searches for the lowest peak.
Operation 1. Press Marker > Select Marker[F1] and select a
marker number.
2. PressPeak
Search and select the type of peak you
wish to find.
Example:Next Peak
7/27/2019 Gsp 900 Manual Final
97/255
BASIC OPERATION
95
Example:Next Peak Right
Example:Next Peak Left
Peak Configuration
Description There are two peak search configurationoptions: Peak Excursion and Peak Threshold.
Peak Excursion: Peak Excursion sets theminimum value above thepeak threshold for whichpeaks will be detected.
7/27/2019 Gsp 900 Manual Final
98/255
GSP-930 User Manual
96
Peak Threshold: Peak threshold sets theminimum threshold level forthe analyzer to detect peaks.
Any value above the PeakThreshold + Peak Excursionwill be detected as a peak.
Peak ThresholdPeak Excursion
1 Peakdetection area
Operation 1. PressPeak
Search >More 1/2[F7].
2. Press Peak Excursion[F2] to set the excursionlevel.
3. Press Peak Threshold[F3] to set the peakthreshold.
Peak Excursion: 0~100dB
Peak Threshold: -120dB~+30dB
Peak Table
Description The Peak Table function will display all peaks(up to 10) that meet the peak configurationsettings. The amplitude and frequency for eachpeak is listed.
Operation 1. PressPeak
Search >More 1/2[F7]>Peak Table[F5].
7/27/2019 Gsp 900 Manual Final
99/255
BASIC OPERATION
97
2. Press Peak Sort[F2] and set the sorting type:
Freq:
Amp:
Sort by frequency in
ascending order.Sort by amplitude inascending order.
3. Press Peak Table[F1] to turn the peak table on.
4. The display splits in two. The bottom screenshows the peak table with the peak marker ID,
X-axis position and amplitude.
Peak Table
Peak markers
7/27/2019 Gsp 900 Manual Final
100/255
GSP-930 User Manual
98
DisplayThe Display key configures the basic display settings as well as
setting up the display mode (spectrum, spectrographic,topographic) and the split screen modes.
Adjusting the LCD Brightness
Description The LCD brightness levels can be adjusted tothree pre-set levels.
Operation 1. Press Display > LCD Brightness[F2] to toggle the
display brightness:
Hi:Mid:Lo:
High brightnessMedium brightnessLow brightness
Turning the LCD Backlight Off
Description The LCD backlight can be turned off topreserve power or to prolong the lifetime of theLCD display when not in use.
Operation 1. Press Display > LCD Backlight[F3] and turn the
LCD backlight off.
2. When the backlight is off, press any functionkey to turn the LCD backlight back on.
7/27/2019 Gsp 900 Manual Final
101/255
BASIC OPERATION
99
Setting a Display Line (Reference Level Line)
Description The Display Line function is used to super-impose a reference level line over the traces.
Operation 1. Press Display > Display Line[F4] to turn the
display line on.
2. Set the display line level and unit.
Example:
Display line
Display line set at -50dBm
Using the Video Out Port
Description The GSP-930 has a dedicated DVI terminal to
output the display to an external monitor. Thevideo output is always on.
Output resolution 800 x 600 (fixed)
Operation 1. Connect an external monitor to the rear panelDVI terminal.
7/27/2019 Gsp 900 Manual Final
102/255
GSP-930 User Manual
100
Setting the Display Mode
Description The GSP-930 has three different display modesfor viewing: spectrum, spectrograph andtopographic. It is also possible to view thespectrum with the spectrographic ortopographic views using a split screen.
Spectrum Default display mode.
Spectrogram Useful for viewingfrequency or power inthe time domain.
Topographic Useful for observing thefrequency of events witha trace.
Operation 1.
PressDisplay
>Window Setup[F1] and select thedisplay mode:
Spectrum[F1]:Spectrogram[F3]:Topographic[F4]:Spectrogram+Spectrum[F5]:Topographic+Spectrum[F6]:
NoteThe same trace is used on the top and bottom forthe Spectrogram+Spectrum andTopographic+Spectrum modes.
7/27/2019 Gsp 900 Manual Final
103/255
BASIC OPERATION
101
Example:Spectrogram