IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
2 of 38
IST(02) 029 A4 모(101025)
TABLE OF CONTENTS
Table of contents 2
Information of test laboratory, Power Supply System Used, 3
Measurement Uncertainty, Product Information
Equipment Under Test and test setup 4
Summary 5
Test Conditions and Data - Emissions
Conducted Emissions 0.15 MHz - 30 MHz Applicable
Test Conditions / Data and Plots 6-9
Radiated Emissions 30 MHz - 5000 MHz Applicable
Test Conditions / Data and Plots 10-11
Harmonic Current And Voltage Fluctuations
Harmonic Current Not applicable
Test Conditions / Data and Plots
Voltage Flicker Applicable
Test Conditions / Data and Plots 12-14
Test Conditions and Data - Immunity
Electro-Static Discharge Applicable
Test Conditions / Data and Plots 15-17
Radiated Susceptibility Applicable
Test Conditions / Data and Plots 18-19
Electrical Fast Transient (Burst) Applicable
Test Conditions / Data and Plots 20-21
Surge Transient Applicable
Test Conditions / Data and Plots 22-23
Conducted Susceptibility Applicable
Test Conditions / Data and Plots 24-25
Power Frequency Magnetic Not applicable
Test Conditions / Data and Plots
Voltage Dips, interruptions Applicable
Test Conditions / Data and Plots 26-27
Appendix A. The Test Setup Photos 28-33
B. The Photos of ESD Measurement 34-35
C. The EUT Photos 36-38
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
3 of 38
IST(02) 029 A4 모(101025)
INFORMATION OF TEST LABORATORY
IST Co., Ltd.
400-19, Singal-dong, Giheung-gu, Yongin-si,
Kyonggi-Do, 446-599, Korea
TEL : +82 31 326 6700 FAX : +82 31 326 6797 KOLAS Testing No. : 118 RRA Designation No. : KR0018 FCC Registration No. : 400603 FCC(DoC) Registration No. : 801060 VCCI Member No. : 1739
POWER SUPPLY SYSTEM USED
Power supply system AC 230 V, 50 Hz
(Refer to the product information)
Measurement Uncertainty
The measurement uncertainty was evaluated for all test items listed in
this report. Also it, “ The evaluation and treatment of uncertainty “,
is described in IST Quality Manual according to ISO17025 Guide. The data
and results described in this report are true and include evaluated
uncertainty. It may cause some deviation of uncertainty by change component
or process of the test for similar products.
PRODUCT INFORMATION
Video Input 16 Channel Composite
Video Outputs CCTV Monitor(NTSC/PAL), CRT(VGA) Monitor 16 CH Looping Composite, 3 Spot CCTV Monitor(NTSC/PAL)
Monitor Display NTSC : 480 fps, PAL : 400 fps Monitor Screen split 1/4/9/16/PIP Recording Compression H.264 Recording Storage media Internal 2 x HDD(Option), USB Recording Audio 4 Channel, G.726 Compression Camera Control Pan/Tilt, Zoom, Focus, Iris Back-up DVD +/- RW, USB
Power(Adapter) Input : AC(100~240) V, (50~60) Hz, 1.5 A Output : DC12 V, 5.0 A
Size 400mm x 88 mm x 400 mm
-EMC suppression device is not used during the test.
-Find information in user manual.
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
4 of 38
IST(02) 029 A4 모(101025)
Equipment Under Test
EUT Type :
Table-Top. Floor-Standing. Table-Top and Floor-Standing (Combination).
Operation – mode of the E.U.T. :
The equipment under test was operated during the measurement under following
conditions :
Standby. Operational Condition : Recording during Network communication(ping) mode
Configuration of the equipment under test :
Following peripheral devices and interface cables were connected during the
measurement :
Equipment Type Brand Serial No. Adapter(EUT) PA-1061-0 LITE-ON TECHNOLOGY CORPORATION N/A LCD Monitor M2294DJ LG Electronics 902KCYQ4R914Notebook PC CP10 Arima Computer NC2CCHS4J1LCAdapter(Notebook) ADP-50HH DELTA KJW0249000842CCD Camera CNB-GN258 CNB CN560241278Adapter(Camera) DSA-04215-12 1 42 Dee Van Electronic Co.,Ltd. 60308032 Mouse(Serial) M-MD14-2 Logitech Inc. N/A USB Memory US01M1 EO Electron N/A Track ball mouse SPR-8695TU SEJIN ELECTRON INC. 7AFZ007772 Connecting Interface Cables :
- Shielded VGA(D-Sub) cable(with two ferrite core): 1.8 m - Shielded VIDEO IN(BNC) cable(without ferrite core): 3.0 m - Shielded VIDEO OUT(BNC) cable(without ferrite core): 3.0 m - Shielded TV/SPOT OUT(BNC) cable(without ferrite core): 3.0 m - Unshielded RS485 cable(without ferrite core): 5.0 m - Unshielded SENSOR/ALARM cable(without ferrite core): 3.0 m - Unshielded LAN(RJ-45) cable(without ferrite core): 3.0 m - Unshielded Mouse’s RS-232 cable(without ferrite core): 1.6 m - Unshielded Track ball mouse USB cable(without ferrite core): 1.0 m
The drawing of general test setup :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
5 of 38
IST(02) 029 A4 모(101025)
SUMMARY
Emissions
Conducted Emissions
The requirements are MET Not MET
Minimum limit margin 3.25 dB at 23.130 MHz
Maximum limit exceeding
Remarks : Limits are kept with more than 3 dB margin
Radiated Emissions
The requirements are MET Not MET
Minimum limit margin 3.54 dB at 126.03 MHz
Maximum limit exceeding
Remarks : Limits are kept with more than 3 dB margin
Generic Test
Harmonic Current/Voltage Flicker
The requirements are MET Not MET
Immunity
Electrostatic Discharge MET Not MET
Radiated Electromagnetic Fields MET Not MET
Electrical Fast Transient (Burst) MET Not MET
Surge Transient MET Not MET
Conducted Susceptibility MET Not MET
Power Frequency Magnetic MET Not MET
Voltage Dips, interruption MET Not MET
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
6 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Conducted Emissions
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
ESCI Test Receiver Rohde & Schwarz May 21, 2011 100374
ESH2-Z5 LISN Rohde & Schwarz May 21, 2011 842966/007
ESH3-Z2 Pulse Limiter Rohde & Schwarz May 21, 2011 357.8810.52
ISN T800 ISN TESEQ May 21, 2011 26107
Environmental Conditions
Temperature (18.5 ± 0.2)
Humidity (38 ± 0.1) % R.H.
Atmosphere pressure 1011 hPa
Test Program See the operation mode on page 4.
Test Area Conducted Room #2
Test Date May 02, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
7 of 38
IST(02) 029 A4 모(101025)
Conducted Emissions Live
Freq. [MHz]
Measurement [dB]
Limit [dB]
InsertionLoss
CableLoss
Result [dB]
Margin [dB]
Q-peak Average Q-peak Average [dB] [dBuV] Q-peak Average Q-peak Average
0.150 44.48 30.40 66.00 56.00 0.15 0.01 44.64 30.56 21.36 25.44
0.186 41.01 24.30 64.21 54.21 0.14 0.01 41.16 24.45 23.05 29.76
0.214 36.12 18.74 63.05 53.05 0.14 0.01 36.27 18.89 26.78 34.16
0.610 29.11 18.54 56.00 46.00 0.17 0.02 29.30 18.73 26.70 27.27
1.098 34.30 22.54 56.00 46.00 0.19 0.04 34.53 22.77 21.47 23.23
24.214 35.56 34.03 60.00 50.00 1.05 0.20 36.81 35.28 23.19 14.72
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
8 of 38
IST(02) 029 A4 모(101025)
Conducted Emissions Neutral
Freq. [MHz]
Measurement [dB]
Limit [dB]
InsertionLoss
CableLoss
Result [dB]
Margin [dB]
Q-peak Average Q-peak Average [dB] [dBuV] Q-peak Average Q-peak Average
0.150 43.89 30.71 66.00 56.00 0.15 0.01 44.05 30.87 21.95 25.13
0.182 41.74 24.69 64.39 54.39 0.15 0.01 41.90 24.85 22.49 29.54
0.210 38.85 21.28 63.21 53.21 0.15 0.01 39.01 21.44 24.19 31.76
0.626 33.31 24.26 56.00 46.00 0.18 0.02 33.51 24.46 22.49 21.54
1.114 32.05 19.51 56.00 46.00 0.21 0.04 32.30 19.76 23.70 26.24
24.214 34.98 33.35 60.00 50.00 0.88 0.20 36.06 34.43 23.94 15.57
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
9 of 38
IST(02) 029 A4 모(101025)
Conducted Emissions
Freq. [MHz]
Measurement [dB]
Limit [dB]
InsertionLoss
CableLoss
Result [dB]
Margin [dB]
Q-peak Average Q-peak Average [dB] [dB] Q-peak Average Q-peak Average
5.238 44.51 41.32 74.00 64.00 9.67 0.09 54.27 51.08 19.73 12.92
7.926 49.71 46.50 74.00 64.00 9.59 0.11 59.41 56.20 14.59 7.80
13.358 53.56 49.80 74.00 64.00 9.67 0.14 63.38 59.62 10.62 4.38
17.694 54.76 49.90 74.00 64.00 9.81 0.17 64.74 59.88 9.26 4.12
23.130 58.00 50.50 74.00 64.00 10.05 0.20 68.25 60.75 5.75 3.25
Note : Telecommunication ports
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
10 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Radiated Emissions
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
ESCS30 Test Receiver Rohde & Schwarz Sep. 17, 2011 100171
VULB 9160 Antenna Schwarzbeck Aug. 06, 2011 3048
E7405A Spectrum Analyzer Schwarzbeck Sep. 17, 2011 US41160290
BBHA 9120D Horn Antenna H.P. Jun. 11, 2012 9120D-826
8348A Preamplifier Max-Full Antenna Corp
Sep. 16, 2011 3307A02865
MFT-120S Turn-Table AFS42-01001800-28-10P-42 N/A -
MITEQ Inc. Amplifier Schwarzbeck Feb. 14, 2012 1565819
Environmental Conditions
Temperature (14.4 ± 0.5)
Humidity (37 ± 0.8) % R.H.
Atmosphere pressure 1013 hPa
Test Program See the operation mode on page 4.
Test Area Open Test Site #2 (10 m)
Full-Anechoic Room (3 m) (ETL LABORATORY)
Test Date May 3, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
11 of 38
IST(02) 029 A4 모(101025)
Radiated Emissions
(Disturbance Radiation)
Freq. [MHz]
Reading [dBuV]
AntennaFactor[dB/m]
CableLoss [dB]
Polar.[H/V]
Limit [dBuV/m]
Result [dBuV/m]
Margin [dB]
30.970 11.50 12.03 0.96 V 30.00 24.49 5.51
37.760 12.30 12.22 1.05 V 30.00 25.57 4.43
63.950 11.50 11.17 1.32 V 30.00 23.99 6.01
126.030 13.00 11.62 1.84 V 30.00 26.46 3.54
150.280 11.20 12.62 1.99 V 30.00 25.81 4.19
175.500 11.10 11.45 2.16 V 30.00 24.71 5.29
500.451 11.00 17.46 3.82 H 37.00 32.28 4.72
750.703 5.50 21.60 4.59 V 37.00 31.69 5.31
Note :
Radiated Emissions
(Disturbance Radiation)
Freq. [MHz]
Reading [dBuV]
c.f [dB(1/m)]
Polar.[H/V]
Limit [dBuV/m]
Result [dBuV/m]
Margin [dB]
Peak Detector
1076.76 43.60 -3.30 H 70.00 46.90 23.10
1125.24 43.50 -3.30 V 70.00 40.20 29.80
1500.96 52.00 -2.50 V 70.00 49.50 20.50
1864.56 42.90 -2.00 V 70.00 40.90 29.10
2123.21 43.70 -0.20 V 70.00 43.50 26.50
Average Detector
1076.76 38.40 -3.30 H 50.00 41.70 8.30
1125.24 36.50 -3.30 V 50.00 33.20 16.80
1500.96 39.80 -2.50 V 50.00 37.30 12.70
1864.56 37.00 -2.00 V 50.00 35.00 15.00
2123.21 36.90 -0.20 V 50.00 36.70 13.30
Note : Limits above 1 GHz(3 m method)
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
12 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Harmonic Current/Voltage Flicker
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacturer Due Calibration Serial No.
PM6000 Power Analyzer Voltech Jan. 31, 2012 100006700213
IEC Standard 555 Impedance Network Voltech Jan. 28, 2012 1B0318251
4500L AC Power Source California May 21, 2011 HK51408
Environmental Conditions
Temperature (22 ± 0.2)
Humidity (39 ± 0.1) % R.H.
Atmosphere pressure 1013 hPa
Test Program See the operation mode on page 4.
Test Area EMI Test Room
Test Date May 4, 2011
Note : Harmonic test is omitted because rated power is less than 75 watts.
(See clause 7 of EN61000-3-2)
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
13 of 38
IST(02) 029 A4 모(101025)
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
14 of 38
IST(02) 029 A4 모(101025)
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
15 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Electrostatic Discharge
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
ESS-2000 ESD Simulator NoiseKen Jul. 18, 2011 ESS0442747
TC-815P ESD Tip NoiseKen Jul. 18, 2011 ESS0442694
Environmental Conditions
Temperature (21.5 ± 0.2)
Humidity (37.4 ± 0.1) % R.H.
Atmosphere pressure 1011 hPa
Test Program See the operation mode on page 4.
Test Area EMS Test Room
Test Date May 6, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
16 of 38
IST(02) 029 A4 모(101025)
Electrostatic Discharge(ESD)
[Applicable]
Test Specification:
Discharge Voltage (Air): 2 kV 6 kV 15 kV
4 kV 8 kV kV
Discharge Voltage : 2 kV 6 kV kV
(Contact) 4 kV 8 kV
Discharge Impedance: 330 Ω / 150 pF 150 Ω / 150 pF
Discharge Repetition Rate : 1 s
Number of Discharges: 10 times at all locations (Air discharge)
10 times at all locations (Contact discharge)
Kind of Discharges: Air discharge Contact discharge
Direct Indirect (HCP/VCP)
Polarity: Positive Negative
Location of Discharge: Each location on the surface accessible by hand
See drawing in page 33-34
Result :
No degradation of function Met Criterion A
Distortion of function Met Criterion B
Error of function Met Criterion C
Loss of function Unrecoverable Failure
Note: During the performed test there was no change of operation mode.
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
17 of 38
IST(02) 029 A4 모(101025)
Electrostatic Discharge (ESD)
List of ESD Points Discharge
Voltage(V) Test Point
Test Result(s)
±2 kV ±4 kV ±6 kV ±8 kV
HCP(C) A A A -
VCP(C) A A A -
Front CD-ROM(A) A A - A
Front Surface(C) A A A -
Upper Surface(C) A A A -
Rear Surface/Screw/Ports(C) A A A -
Side Surface/Screw(C) A A A -
(A): Air Discharge (C): Contact Discharge
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
18 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Radiated Susceptibility
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
SML03 Signal Generator Rohde & Schwarz May 21, 2011 102212
NRVD Power Meter Rohde & Schwarz May 21, 2011 101242
NRV-Z51 Power Sensor Rohde & Schwarz May 21, 2011 100330
NRV-Z51 Power Sensor Rohde & Schwarz May 21, 2011 100331
HI-6005 Field sensor ETS Aug. 08, 2011 00052474
AR200W Amplifier Amplifier Research - 14124
LPDA-0803 Log-periodic Ant. TDK RF - 130232
AT4002A Microwave Horn Ant. TDK RF - 312314
Environmental Conditions
Temperature (20 ± 0.2)
Humidity (40 ± 0.1) % R.H.
Atmosphere pressure 1013 hPa
Test Program See the operation mode on page 4.
Test Area Full anechoic chamber
Test Date May 8, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
19 of 38
IST(02) 029 A4 모(101025)
Radiated Susceptibility
[Applicable]
Test Specification
Frequency Range: 27 MHz – 500 MHz 900 MHz
9 kHz - 27 MHz 80 MHz – 2000 MHz
Field Strength: 1 V/m 3 V/m
10 V/m V/m
Distance Antenna – EUT: 1 m 3 m
Modulation: AM : 80 %
FM : kHz dev.
Sine wave: 1 kHz
Unmodulated
Pulse : 1 Hz
Duty Cycle: 50 %
Step: < 0.0015 decades/s
1 % / 3 s log 1 % / 3 s
Polarization of Antenna: Horizontal Vertical
Position Front Left
Right Rear
Result
No degradation of function Met Criterion A
Distortion of function Met Criterion B
Error of function Met Criterion C
Loss of function Unrecoverable Failure
Note : During the performed test there was no change of operation mode.
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
20 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Electrical Fast Transient
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
PHV41.2 Burst Generator Heafely May 21, 2011 083057-18
IP4A Capacitive Clamp Heafely May 21, 2011 083-078-32
Environmental Conditions
Temperature (21.5 ± 0.2)
Humidity (37.4 ± 0.1) % R.H.
Atmosphere pressure 1011 hPa
Test Program See the operation mode on page 4.
Test Area EMS Test Room
Test Date May 6, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
21 of 38
IST(02) 029 A4 모(101025)
Electrical Fast Transient (Burst)
[Applicable]
Test Specification
Pulse Amplitude - AC Power Port: 0.5 kV 1.0 kV
2.0 kV
Pulse Amplitude - DC Power Port: 1.0 kV 2.0 kV
4.0 kV
Pulse Amplitude - Signal/Data 0.5 kV 1.0 kV
Non control Port: 2.0 kV 0.25 kV
Pulse Amplitude - Process: 0.5 kV 1.0 kV
Measurement & Control Port 2.0 kV
Burst Frequency: 2.5 kHz 5.0 kHz
Duration per application : 1 (+0.2/-0) (min)
Time of Coupling: 60 s 120 s
Coupling Method: Coupling/decoupling network Coupling clamp
Polarity: Positive Negative
Location of Coupling:
name of lines: AC Power cable(Adapter) length : 0.5 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
name of lines: BNC(VIDEO IN, VIDEO/TV/SPOT OUT) cable length : 3.0 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
name of lines: LAN, RS485, SENSOR/ALARM cable length : 5.0 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
Result :
No degradation of function Met Criterion A
Distortion of function Met Criterion B
Error of function Met Criterion C
Loss of function Unrecoverable Failure
Note: Color camera’s input picture was flickerd during the test, but it was
recovered by itself after test.
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
22 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Surge Transient
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
PSURGE 4.1 Surge Generator Heafely May 21, 2011 082876-05
IP6.2 COUPLING NETWORK Heafely - 082876-02
Environmental Conditions
Temperature (21.1 ± 0.2)
Humidity (38.5 ± 0.1) % R.H.
Atmosphere pressure 1012 hPa
Test Program See the operation mode on page 4.
Test Area EMS Test Room
Test Date May 7, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
23 of 38
IST(02) 029 A4 모(101025)
Surge Transient
[Applicable]
Test Specification
Pulse amplitude
– Power line :
0.5 kV 1.0 kV : Line to line
0.5 kV 1.0 kV 2.0 kV : Line to ground
Pulse amplitude
– Signal line :
0.5 kV
2.0 kV
1.0 kV
Source impedance : 18 µF : Line to line
Via a 40 series resistor 10 + 9 µF: Line to ground
Number of surges : 3 surges/angle 20 surges/polarity
Angle 0° 90°
180° 270°
Repetition rate : 10 s 60 s
Polarity : Positive Negative
Location of Coupling:
name of lines: AC Power cable(Adapter) length : 1.0 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
name of lines: BNC(VIDEO IN) cable length : 30.0 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
name of lines: RS485(TX +/-) cable length : 30.0 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
Result :
No degradation of function Met Criterion A
Distortion of function Met Criterion B
Error of function Met Criterion C
Loss of function Unrecoverable Failure
Note : During the performed test there was no change of operation mode.
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
24 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Conducted Susceptibility
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
SMT 02 Signal Generator Rohde & Schwarz Oct. 12, 2011 833414/019
250LM1 Amplifier Amplifier Research - 14229
URV5 AC Milivoltmeter Rohde & Schwarz Oct. 12, 2011 836671/003
URV-Z2 INSERTION UNIT 50 Rohde & Schwarz Oct. 13, 2011 DE105101
URV-Z2 INSERTION UNIT 50 Rohde & Schwarz Oct. 13, 2011 DE105102
50FH-006-50 Attenuator JFW May 21, 2011 -
FCC-150-50 (50-150) Adapter FCC May 21, 2011 05039
FCC-150-50 (50-150) Adapter FCC May 21, 2011 05040
FCC-801-M3-50 CDN FCC Aug. 18, 2011 28
FCC-203I EM Clamp FCC Jan. 31, 2012 2
FCC-205ADC Decoupling Clamp FCC Jan. 31, 2012 1
Environmental Conditions
Temperature (22 ± 0.2)
Humidity (38 ± 0.1) % R.H.
Atmosphere pressure 1012 hPa
Test Program See the operation mode on page 4.
Test Area Full anechoic chamber
Test Date May 9, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
25 of 38
IST(02) 029 A4 모(101025)
Conducted Susceptibility
[Applicable]
Test Specification
Frequency – range : 0.01 MHz – 150 MHz 80 MHZ – 230 MHz
0.15 MHz – 100 MHz
Voltage level (EMF) : 1 V 10 V
3 V
Modulation : AM: 80 %, 1 kHz, sinusoidal Pulse: 1 Hz(0.5 s ON : 0.5 s OFF)
FM : Unmodulated
Frequency step : 0.015 oct/s Log 1 %
Dwell Time : 3 s 2 s 1 s
Location of Coupling:
name of lines: AC Power cable(Adapter) length : 0.3 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
name of lines: BNC(VIDEO IN, VIDEO/TV/SPOT OUT) cable length : 3.0 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
name of lines: LAN, RS485, SENSOR/ALARM cable length : 5.0 m
type of lines shielded unshielded
status of lines: passive active
kind of transmission: analog digital
Result :
No degradation of function Met Criterion A (1 V, 3 V)
Distortion of function Met Criterion B (10 V)
Error of function Met Criterion C
Loss of function Unrecoverable Failure
Note : For power line test, color camera’s input picture was flickerd at voltage
level 10 V, but the EUT operated as normal after the test.
For voltage level 1 V and 3 V test, the EUT operated normally.
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
26 of 38
IST(02) 029 A4 모(101025)
TEST CONDITIONS AND DATA
Voltage Dips, Interruptions
[Applicable]
Test Equipment Used
The test equipment used is calibrated in regular for every year.
Model Name Description Manufacture Due Calibration Serial No.
PLINE 1610 Dips Generator Haefely May 21, 2011 083690-19
Environmental Conditions
Temperature (21.1 ± 0.2)
Humidity (38.5 ± 0.1) % R.H.
Atmosphere pressure 1012 hPa
Test Program See the operation mode on page 4.
Test Area EMS Test Room
Test Date May 7, 2011
Note :
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
27 of 38
IST(02) 029 A4 모(101025)
Dips/Interruptions
[Applicable]
Test Specification
Normal Voltage/frequency: 100 Vac/50 Hz 230 Vac/50 Hz
Duration of reduction: 0.5 P 1 P 5 P 10 P
(Voltage reduction 30 %)
Duration of reduction: 0.5 P 1 P 5 P 10 P
(Voltage reduction 60 %)
Duration of reduction: 0.5 P 1 P 5 P
(Voltage reduction 100 %)
Number of reductions at each duration : 3
Interval between reductions : 10 s
Phase: 0˚ 45˚ 90˚ 135˚
180˚ 225˚ 270˚ 315˚
Result : 100 Vac/50 Hz
Test Dips(30 %) Dips(60 %) Dips(100 %)
Variation (Unom -15 %)
0.5/1/5/10 period
0.5/1/5/10 period
0.5/1/5 period
Criteria A A A A B B B B C C C C
Result : 230 Vac/50 Hz
Test Dips(30 %) Dips(60 %) Dips(100 %)
Variation (Unom +10 %)
0.5/1/5/10 period
0.5/1/5/10 period
0.5/1/5 period
Criteria A A A A B B B B C C C C
Note: During the performed test there was no change of operation mode.
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
28 of 38
IST(02) 029 A4 모(101025)
Appendix A. Test Setup Photos
Conducted Emissions
Conducted Emissions at telecommunication ports
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
29 of 38
IST(02) 029 A4 모(101025)
Appendix A. Test Setup Photos
Radiated Emissions – Front View
Radiated Emissions – Rear View
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
30 of 38
IST(02) 029 A4 모(101025)
Appendix A. Test Setup Photos
Radiated Emissions above 1 GHz
H&F
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
31 of 38
IST(02) 029 A4 모(101025)
Appendix A. Test Setup Photos
Electro-Static Discharge
Radiated Susceptibility
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
32 of 38
IST(02) 029 A4 모(101025)
Appendix A. Test Setup Photos
Electrical Fast Transient
Surge Transient
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
33 of 38
IST(02) 029 A4 모(101025)
Appendix A. Test Setup Photos
Conducted Susceptibility
Voltage Dips
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
34 of 38
IST(02) 029 A4 모(101025)
Discharge (Air)
Appendix B. The Photos of ESD Measurement Discharge (Contact)
ESD Point
C
A
C
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
35 of 38
IST(02) 029 A4 모(101025)
Discharge (Air)
Appendix B. The Photos of ESD Measurement Discharge (Contact)
ESD Point
C
C
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
36 of 38
IST(02) 029 A4 모(101025)
Appendix C. The EUT Photos
Front View
Rear View
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
37 of 38
IST(02) 029 A4 모(101025)
Appendix C. The EUT Photos
Inner View
IST Co., Ltd. TEST REPORT NO. : 11-IST-0272
38 of 38
IST(02) 029 A4 모(101025)
Appendix C. The EUT Photos
Adapter View
Adapter label View