MAINTENANCESCHEDULEMAITAI,eV5,REGENCHILLER
MODEL/SERIALNUMBER: 6360T11SP20C/1A17500482POWER: 120VAC,13.1AINSTALLDATE: 1NOV17FILTERMODEL: PENTEKP5,Spunpolypropylenefilter,5umratingCLEANING&WATERFILTERCHANGEPROCEDUREShutoffalllasersandchillers.Placebucketandcakepanunderfilterassemblyinback.Drainwaterintobucketbyopeningpetcockvalve.Openfillcaptoventdraining.Openfilterhousingwithfilterhousingwrench.Drainwaterfromfilterhousingandcleanifnecessary.Replacefilterwithanewone.Replaceandtightenfilterhousing.DONOTOVERTIGHTEN!Justalittleoversnugisfine.RefillreservoirwithsteamdistilledwaterONLY.Turnonchiller.Asit’spumpingthroughthesystem,addmorewatertoreservoiruntilitsfull.Leavesystempumpingfor10minutes.Shutdownchilleragain,drainwateragain,refillwateragain(withfreshdistilledwater),restartchillerandtopoffreservoiragain.Verifypumppressureat~39psi.AIRFILTERSERVICETheairfilterislocatedonthefrontofthechiller.Pullthetabsidewaysanditwillpopout.Toclean,removethefilter,vacuumoffthemajorityofdust,thenrinseitinthesink.Waituntilit’stotallydrybeforereplacinginchiller.CONDENSINGFANThecondensingfandrawsairthroughtheairfilterandoverthecondenser.Thishasaservicelifeofabout5-6years.Itcanbereplacedasnecessary.SCHEDULE
• Waterfiltershouldbeinspectedevery3monthsandchangedevery6monthsorifithasanydiscoloration.
• Watershouldbeflushedandreplacedevery3months.• Airfiltershouldbecheckedevery3monthsandcleanedasneeded.Replaceifitistorn
orcannotbethoroughlycleaned.• WaterlevelshouldbecheckedWEEKLY!• Condensingfanshouldbecheckedatabout5years.
SERVICERECORD(MaiTaiChiller)
Date H2Oflush?
H2Ofilterchanged?
Airfilterservice?
Otherdetails
1Nov17 x x x Initialinstallation(2-3gallonstofill)
Checkwaterlevelweekly.Topoffif
necessarywithsteamdistilledwaterONLY!
FilterHousingWrench Filter SteamDistilledWater ChillerRunningNormally
Top Related