AQT90 FLEX Service TrainingHANDS ON Task list
RADIOMETER PRESENTATION2 19/04/23
Task – Wet Section
Service Program “Wet Section”
Test Pumps P1 and P2 Test Pumps P4 and P5 Test Pump P3 Test Valve V1 Test Valve V2 Calibrate Liquid sensor LS1 Calibrate Liquid sensor LS2 Calibrate Liquid sensor LS3
Refer to the service manual chapter 3-6 and chapter 7-19
RADIOMETER PRESENTATION3 19/04/23
Task – Wet Section
Service Program “CAVRO”
Examine the different menus Run the self test
Refer to the service manual chapter 3-6 and chapter 7-19
Service Program “Reagent Pack” Release RP Lock - unlock Tag – see the information
RADIOMETER PRESENTATION4 19/04/23
Task – Wet Section
Service Program “Reagent Pack”
Release RP Lock - unlock Tag – read the information Write the values:
• Remaining buffer• Cups left• Other
Refer to the service manual chapter 3-6 and chapter 7-19
RADIOMETER PRESENTATION5 19/04/23
Tasks – Cup transport
Service Program “Cup transport”
perform stepper movement actions. Remove the RP and feel the cup sensor – calibrate the cup
sensor. Run a “cup transport” Exchange needle – use torque screwdriver 0.95 Nm Needle calibration – important to follow the guidelines in SM
RADIOMETER PRESENTATION6 19/04/23
Task – Inlet
Service Program “Inlet Wheel”
Mix and check sensor positions - try to adjust inlet wheel position on stepper motor rod
Lock and unlock. See the function
Service Program “Inlet Door” Open – close. See the functionality
Service Program “Storage Door” Open – close. See the movement to the sensors
Look eventually into Colour definitions and Vision Control
Refer to the service manual chapter 3
RADIOMETER PRESENTATION7 19/04/23
Task – Inlet
Service Program “Inlet Door”
Open – close. examine the functionality
Service Program “Storage Door” Open – close. See the movement to the sensors
Refer to the service manual chapter 3
RADIOMETER PRESENTATION8 19/04/23
Task – Pipette
Service Program “Pipette”
Sensor Check Park Waste Sample Presample Preassay Wash Pierce Service Needle movements Disassemble Wash Block – examine!
Refer to the service manual chapter 3
RADIOMETER PRESENTATION9 19/04/23
Task – Process Unit
Pls. perform the following tests and adjustments:
Service Program “Process Unit”
Raise Lower Sensor and movement check
Service Program “Cup Dryer” Find target temperature Find blower flow Dryer flow test Check the change of values in the special config
Refer to the service manual chapter 3
RADIOMETER PRESENTATION10 19/04/23
Task – Optical Unit
Pls. perform the following test and adjustments:
Service Program “Measuring”
Go to reference Dark Meas. Measure 7 times Calibrate Factory Reference Compare value and save.
Refer to the service manual chapter 3 and chapter 6-19
RADIOMETER PRESENTATION11 19/04/23
Electronics – Power Supply - Status
Pls. perform the following tests and adjustments:
Service Program “Status”
Check the values for the voltages:
12V ________
24Va ________
24Vb ________
24Vc ________
RADIOMETER PRESENTATION12 19/04/23
Software Tasks
Simulate exchange of Main Board PCB (Service Manual) Write down the steps you will do: Files saved on a USB stick: ____________________ ____________________ ____________________
Simulate exchange of CF card What will you do, step by step: Files that must be saved: ____________________ ____________________
Software upgrade What will you do, step by step:
____________________ ____________________ ____________________ ____________________
Steps:
Steps:
RADIOMETER PRESENTATION13 19/04/23
Software UpgradeDirect approach of accessing a USB device and upgrade DMS software
No sample tests must be running!
Insert the USB device. Press Ctl+Shift+Esc to access the Task Manager. Click on New Task. A “Create New Task” dialog box comes up. Enter "explorer" in the dialog text box; a Windows Start icon appears at the
bottom of the screen. Click Start->My Computer. Double click on file D:\PISA\Temp\ResetAU.bat Current instance of AQT software will be terminated Access the root of the USB and click on Setup.cmd Choose to upgrade – tick the box Prompted “Calibrate Display” pls. perform precisely Prompted “Choose Language” pls. choose English After the installation the analyser will be restarted. Once Analyser is restarted then Windows start-up sequence comes up and the
user is logged on as ABLUser. Use Task manager to “End Process” explore
RADIOMETER PRESENTATION14 19/04/23
No sample tests must be running! All data logs will be cleared All parameter calibrations will be cleared
Insert the USB device. Press Ctl+Shift+Esc to access the Task Manager. Click on New Task. A “Create New Task” dialog box comes up. Enter "explorer" in the dialog text box; a Windows Start icon appears at the
bottom of the screen. Click Start->My Computer. Double click on file D:\PISA\Temp\ResetAU.bat Current instance of AQT software will be terminated Access the root of the USB and click on Setup.cmd Choose to “clean install” – tick the box if not already Prompted “Calibrate Display” pls. perform precisely Prompted “Choose Language” pls. choose English After the installation the analyser will be restarted. Once Analyser is restarted then Windows start-up sequence comes up and the
user is logged on as ABLUser. Use Task Manager to “End Process” explore
Software InstallDirect approach of accessing a USB device and installing DMS software
Hands-on tasks
Refer to the Service Manual
1. Replace Process Wheel – easy way2. Disassemble Bottom plate complete3. Disassemble Pipette unit complete4. Adjust BCR for cartridge5. Needle calibration – save values6. Cup transport7. Adjust Inlet wheel8. Clean BCR mirror under LCD screen
USB Stick
Contents of the pen drive:
Agenda Documents Manuals OSK Photos Pre-learning Service training Software Video
Additional
1. Run osk.exe2. Calibrate touch screen3. Setup AQT90 for video and show it
Analyze
Service Dump:
1.Always make a service dump when a failure is observed2.If you are not there ask the lab tech to give as many info as
possible.
19/04/23
End of HANDS ONTask list