7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
1/26
CARESTREAMSPT1 Radiographic Table
Users Guide
2013-04-08
PN AA3426
Version 4.0
H225_1001HC
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
2/26
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
3/26
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
4/26
ii
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
5/26
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
6/26
2 AA3426
Users Guide
CAUTION:
Cautions point out procedures that you must follow
precisely to avoid injury to yourself, others, damage to the
system or any of its components, loss of data, or
corruption of files in software applications.
In the European Union, this symbol indicates that when thelast user wishes to discard this product, it must be sent to
appropriate facilities for recovery and recycling.
Contact your local authorized representative for additionalinformation.
Safety and Regulatory
Safety CautionsCAUTION:
Follow all safety labels on the equipment.
CAUTION:
Do not roll the patient table over toes or other extremities.
CAUTION:
Keep fingers and other body parts away from the table
pinch points when the table is in motion.
CAUTION:
The weight limit of the table is 227 kg (500 lb). Exceeding
this limit may result in injury to the patient or
malfunction of the table controls.
CAUTION:
Do not use the table to move patients.
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
7/26
Users Guide
AA3426 3
CAUTION:
Avoid damaging the table top. Keep surface clear of
sharp-edged objects.
CAUTION:
Thoroughly check that there is no interference or
possibility of collision between the patient and other
equipment.
CAUTION:
Do not use the table on stairs or steep ramps.
Conforming Standards -Safety
This medical device meets the provisions of the Medical Directive93/42 EEC, and the basic requirements according to Annex II. If
you have further questions regarding the applicable national or
international standards, please contact your Carestream Health
representative.
IEC 606011: 1988 + A1 + A2/EN 606011: 1990 + A1 + A2
IEC/EN 6060113: 1994
IEC/EN 60601232: 1994
UL 606011: 2003 R6.03
CAN/CSAC22.2 No. 601.1M90 IEC 606011:2005
IEC/EN 6060113: 2008
IEC/EN 60601254: 2009
EN 606011:2006
This product complies with 21 CFR 1020.30/.31 Performance
Standards for Radiation SafetyRadiographic Equipment.
Carestream Health France
1, rue Galile
93192 NOISY-LE-GRANDE CEDEX
FRANCE
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
8/26
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
9/26
Users Guide
AA3426 5
Labels CARESTREAM Data Plate Sample
1 Official product name
2 Service code and serial number
3 Manufactured date
4 Model type as required/consistent with product safety test reports
5 Voltage range, rated frequency in hertz and amps/consistent withproduct safety test reports: V = volts, Hz = hertz, A = amperes
6 Symbols for product safety, EMC, and CE marking
7 SFDA Registration number 8 SFDA Product Standard number
9 Compliance statement according to FDA requirements for radiationemitting products
10 Label manufacturer registration number and material specification
11 Made in China by (signature)
Made in China byCarestream Health, Inc.150 Verona Street, Rochester, NY 14608 USA1510
SNSC
MODEL
Manufactured:
P/N XXXXXX REV A
Official Product Name (Engl ish)
Official Product Name (Japanese, i f required)Official Product Name (Chinese, i f required)
THIS PRODUCT COMPLIES WITH 21 CFR CHAPTER 1, SUBCHAPTER J
10
9
8
7
6
5
4
3
2
1
11
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
10/26
6 AA3426
Users Guide
Float Label
Up-Down Label
H225_1076HC
Data Plate
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
11/26
Users Guide
AA3426 7
Table Top Weight Limit
Use caution to prevent pinching fingers and other body parts
while moving the table.
Warning Label: Hold The Table Top Firmly While Pressing
the Float Control Pedal.
2G7598 REV. B
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
12/26
8 AA3426
Users Guide
Mobile Patient TableOverview
Indication for Use The SPT1 Radiographic Table is intended to be used to supportpatients during radiographic examinations. It is a single pedestal,
moveable table designed for maximum maneuverability when
used with a DIRECTVIEW system. It is not intended for moving
patients.
All wheel locking at any position allows for quicker patient set up.
Red pedals lock the wheels and green pedals release them.
You can float the table top 2.0 cm (0.8 in.) transversely and
52.0 cm (20.5 in.) longitudinally for precise placement of the
patient. When using the longitudinal float, you can X-ray the
complete length of the table top (212.0 cm) (83.5 in.) withoutmoving the patient.
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
13/26
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
14/26
10 AA3426
Users Guide
System Operation
Moving and Locking thePatient Table
CAUTION:
Check the wheel locks before moving a patient. For
maximum efficiency of movement, floors should be flat,
clean, and dry.
The hand grip allows for easy movement in any direction.
The maximum height of an obstacle or threshold is 15.0 mm (0.6
in.).
H225_1065AC
hand grip
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
15/26
Users Guide
AA3426 11
There are six brake pedals, three red and three green. Press a red
pedal to lock all four wheels:
Press a green pedal to release all four wheels:
H225_1071AC
red pedal
H225_1072AC
green pedal
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
16/26
12 AA3426
Users Guide
The wheels at the end of the table have only one lock pedal and
one release pedal:
Press the red pedal to lock all wheels. Press the green pedal to
release all wheels.
H225_1073AC
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
17/26
Users Guide
AA3426 13
Operating the Table TopFloat
The standard table top position is locked. To allow the table top to
float, press the float control pedal down. This allows you to move
the table top in all four directions by simply pushing it.
Longitudinal Float
H225_1069BC
float control pedal
H225_1070BC
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
18/26
14 AA3426
Users Guide
Transversal Float
To stabilize and lock the table in the desired position, release the
pedal.
H225_1067AC
H225_1068AC
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
19/26
Users Guide
AA3426 15
Adjusting the Table Height Use the table top lift and lower pedals to adjust the table height.Press the lift pedal, multiple times if necessary, to raise the table to
the desired height.
NOTE: Maximum table height is 86.0 cm (33.9 in.).
Press and hold the left pedal to lower the table.
H225_1075AC
lift pedal
H225_1074AC
lower pedal
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
20/26
16 AA3426
Users Guide
Highest and lowest table positions are shown below:
H225_1066AC
H225_1068AC
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
21/26
Users Guide
AA3426 17
Imaging Area Areas outside the lines cannot be imaged.
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
22/26
18 AA3426
Users Guide
Maintenance
Cleaning Follow your hospitals approved cleaning and disinfectingprocedure daily.
Lubricating Once a year lubricate the slides.
Disposal The patient table does not contain hazardous materials. Sort outequipment or parts separately and dispose of them in accordance
with national and regional laws and regulations for waste.
CAUTION:
The elevating system contains minimum quantities of
hydraulic oil.
Service Contact your authorized Service provider if the table does notperform to your satisfaction.
Free Space/BreakdownOutline for the Bucky
218.0 cm (85.5 in.)
86.0 cm (33.9 in.)
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
23/26
Users Guide
AA3426 19
Technical Data
Table Dimensions Length (including handle): 218.0 cm (85.5 in.)
Width: 73.0 cm (28.8 in.)Maximum height: 86.0 cm (33.9 in.)
Table Top Dimensions Length: 212.0 cm (83.5 in.)
Width: 65.0 cm (25.6 in.)
Thickness: 5.4 cm (2.2 in.)
Threshold Height Maximum height: 15.0 mm (0.6 in.)
Table Weight Weight: 155.00 kg (341.70 lb)
Table Deflection Lighter than Average Weight Patient: the head of the table will beelevated 20.0 mm (0.9 in.) or less
Average Weight Patient 90.7 kg (200.0 lb): the table is horizontal
Above Average Weight Patient up to 227 kg (500 lb): the table top
will be below horizontal by 40.0 mm (1.6 in.) or less
X-ray Transparent Surface Length: 212.0 cm (83.5 in.)
Width: 65.0 cm (25.6 in.)
X-ray Absorption Factor Factor: less than or equal to 0.8 mm Aluminum Equivalents
Travel Position of theTabletop
Elevating range: 25.0 cm (9.8 in.)
Longitudinal float: 52.0 cm (20.5 in.)
Transversal float: 24.0 cm (9.5 in.)
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
24/26
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
25/26
7/25/2019 AA3426_SPT1 User Guide_2013-04-08_en
26/26
CARESTREAM HEALTH, INC.
150 Verona Street
Rochester, NY 14608
United States
CARESTREAM is a trademark of Carestream Health, Inc.
Carestream Health, Inc., 2013
PN AA3426
Printed in China
Top Related