1
Test on RPC Veto Detector ModelTest on RPC Veto Detector Model—— Anticoincidence Detector —— Anticoincidence Detector
for Daya Bay Neutrino Exp.for Daya Bay Neutrino Exp.
Test on RPC Veto Detector ModelTest on RPC Veto Detector Model—— Anticoincidence Detector —— Anticoincidence Detector
for Daya Bay Neutrino Exp.for Daya Bay Neutrino Exp.
Speaker:Speaker: Jiawen Zhang Jiawen Zhang
5 June 20065 June 2006
2
OutlineRPC IntroductionRPC R&D in IHEP Of CAS RPC Performance Study RPC Mass Production for BESIIIRPC for Daya bay Neutrino Exp. Precondition and requirement Test Design Test ResultSummary
3
RPC IntroductionWhat is RPC?
RPC is composed of two resistive plates with gas flowing between them. High voltage is applied on the plates to produce a strong electric field in the gas. When a cosmic-ray passes through the gas between the two plates, a signal will produce, which is then picked up by the pickup strip and sent to the DAQ system.
Advantages:Simple structure
Cheap to make a larger area detector
4
RPC R&D in IHEP Of CAS
The RPCs for the BESIII Muon tracker detector were constructed by using a new type of phenolic rosin laminates developed in IHEP of CAS.
The methods of improving surface quality is similar to other bakelite plates, and have been used to construct RPCs elsewhere.
Oscilloscope traces of 100 triggered cosmic ray registered in a RPC prototype at 8 kV. The average signal amplitude from a pickup electrode is about 400 mV with a 50 Ω termination. No secondary streamers were recorded
5
Bakelite surface
7
R&D —— prototype performance
1.18 0.044
96-98%
8
R&D ---- long-term stability
neutron irradiationother
experiments
Beam test
neutron irradiation
Max: 98.8%
Min: 95.3%Average: 97.2%
9
R&D ---- Humidity effect• During the R&D test, we added water
vapor into the gas for about one month. No effects were observed.
• All the HV connectors to the RPC are covered with insulation glue. Last summer, the humidity was about 80-90% lasting about 2 week in Beijing. We tested the RPC bare chambers and the assembled modules, no problem was found, and we will do more test soon.
10
R&D---Flammability gas
• The flammability Iso-butane of the mixture gas maybe catch fire, if its proportion is more than 15%.
• We used the gas mixture of argon:F134a: Iso-butane = 50:42:8, Therefore the problem won’t be avoided with the ventilation requirement for Radon removal.
11
RPC mass production for BESIII
• Single layer RPC (bare chamber) efficiency ε>95%
Barrel 7.5KvMin. 85.6%Max. 99.02%Aver.95.39%
Barrel 8.0KvMin. 90.38%Max. 99.2%Aver.96.4%
12
• A new bare chamber single counting rate is below 1000Hz/m2( only training 1-3days), if training for a long time, the single counting rate will be below 500Hz/m2
Barrel 7.5KvMin. 0.016Max. 0.599Aver.0.095
Barrel 8.0KvMin. 0.022Max. 0.872Aver.0.130
RPC mass production for BESIII
13
• 2 layers of RPCs form a super layer , ε >98%
Mean 0.98
Average eff: 0.99
RPC mass production for BESIII
14
The requirements of Daya bay Neutrino Exp.
The anticoincidence detector is used to decrease the cosmic ray background to improve measurement precision
Near Site Far Site
Neutrino rate (/day) 560 80
Muon Flux (Hz/m2) ~1 0.045
Table 1. Neutrino instance and Cosmic ray flux
According to the design, we know the radius of central detector is 1.6m. And by calculation, the cosmic ray through the detector is 8Hz near site, and 0.36Hz far site.
Taking example of far site, the cosmic ray through 4 modules every day is 4×0.36Hz×3600s×24h=125000 >>80
15
• Anticoincidence detector ε~99.9%, 125(>80) cosmic ray won’t be removed
• Anticoincidence detector ε~99.99%, 12(~15%) cosmic ray won’t be removed
Most of cosmic ray can be removed by spectrum & time relation of later neutron signal
So the result is acceptable!
16
The design of detectorRequirement Higher efficiency Less noise
But these two factors are contradictions, so the key problem is how to balance the two factors.
A fact is that a water Cherenkov may be used. (efficiency is ~ 95%[1] ,noise <0.1Hz. [2] )
17
The design of detector(2)Outer detector scheme• Adopt 2 dimension readout• RPC operate in streamer mode• The gas mixture used as Ar:C2H2F4:C4H10 =50:42:8
• HV:+4000V , -4000VSingle gap, 3 layers in one
moduleEach layer overlapping
assembly, no dead space.Each module overlapping
assembly too, so between modules no dead space.
18
The design of detector(3)
Efficiency and noise Efficiency of each layer is ε ~95%, and adopt choose 2 out of 3 as a hit,
their coincidence efficiency is εeff =ε3+C3
2 ε2(1- ε)=0.953+3×0.952×(1-0.95)=99.3% The efficiency of the module with the water Cherenkov is
1-(1- εeff )(1- ε)=1-(1-0.99)(1-0.95)=99.95% The RPC bare chamber noise rate ~800Hz/ m2 , the shaped signal width
is τ=100ns=10-7s, so the module noise rate is 3C3
2r2 τ=3 × 3 X(800)2×10-7=0.576Hz/m2
since the module has 3 layers RPC, so we can do the track for itself. By using the track information, the noise can be reduced to <0.05Hz/m2. If we reduce RPC noise rate to 300Hz/m2, the noise can be reduce to more lower.
In addition, Because the noise of the water Cherenkov is very small, the total noise is
not more than 1Hz !
19
• Electronics and readout
The design of detector(4)
the same as BESIII Muon detector.
Each FEC can handle 16 channels , and a total of 16 FEC composed a data chain.
All the data chains are connected to the VME readout system.
A fast-OR signal from each FEC is sent to the trigger system.
The primary bitmap signal are transferred from parallel to serial, hence reduce significantly the cables.
The width of the shaped signal is 100ns.
20
Test Result(1)• Group 1 (have been used as a telescope sys.)
0
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 9000
0
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 90000
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 9000
0
0. 01
0. 02
0. 03
0. 04
0. 05
0. 06
6000 7000 8000 9000
00. 010. 020. 030. 04
0. 050. 060. 070. 080. 09
6000 7000 8000 9000
0
0. 01
0. 02
0. 03
0. 04
0. 05
0. 06
0. 07
0. 08
6000 7000 8000 9000
0
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 90000
0. 005
0. 01
0. 015
0. 02
0. 025
0. 03
0. 035
0. 04
6000 7000 8000 9000
99.5±0.25%
21
Test Result(2)• Group 2(haven’t used)
0
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 90000
0. 005
0. 01
0. 015
0. 02
0. 025
0. 03
6000 7000 8000 9000
0
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 9000
0
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 90000
0. 2
0. 4
0. 6
0. 8
1
1. 2
6000 7000 8000 9000
00. 020. 040. 060. 08
0. 10. 120. 140. 160. 18
6000 7000 8000 9000
0
0. 02
0. 04
0. 06
0. 08
0. 1
0. 12
6000 7000 8000 90000
0. 02
0. 04
0. 06
0. 08
0. 1
0. 12
6000 7000 8000 9000
99.3±0.4%
23
SummaryRPC is economical for a larger area detectorThe RPC performance developed by IHEP is excellent,
single gap RPC efficiency>95%, noise rate <800Hz/m2, dark current <2μA/m2
Adopt 3 layers, 2 dimension readout, RPC and module overlapping assembly, no dead space. Adopt choose 2 out of 3 as a hit, their coincidence efficiency >99%, the noises <0.05Hz/m2
Electronics and readout are same to BESIII Muon detector.
24
Reference
1. Determination of Neutrino Mixing-Angle θ13 Using the Daya Bay Nuclear Power Facilities , version 3.1
2. Preliminary study of Daya Bay reactor neutrino experiment, Yaxuan Sun, Ph.D thesis
The end
Thanks!