Zahn_poster_4_19
-
Upload
elaine-mau -
Category
Documents
-
view
14 -
download
0
Transcript of Zahn_poster_4_19
Quantification of Particle Separation in Electrokinetically Driven
Flow through a Bifurcating MicrochannelElaine Mau, Jeffrey Zahn, Rutgers University
IntroductionTo separate various particles or cells within a
sample for diagnostic assay, a bifurcating
microchannel utilizes varying flow ratios within the
daughter channels to direct particles into a certain
channel (Figure 1).
Figure 1: Bifurcation Law using Pressure Driven Flows a) Greater pressure difference
and higher shear force causes particle to move towards higher flow branch b) Position
of centroid with respect to the critical stream line determines where the particle
flows.
The purpose of this work is to study the
properties of using electrokinetic flows to create
the different flow ratios.
Objectives� To design a means of generating electrokinetic
flow in a reproducible way
� To demonstrate the effect of voltage/flow ratio
on particle recovery counts
� To analyze the differences, if any, between
electrokinetic and pressure driven flows.
� To determine the how to apply electrokinetic
flow for particle separation using data recorded
MethodsFabrication of Device
� Standard Soft Lithography Techniques
� PDMS/Glass
� Beads used:
� 15 µm Polystyrene Divinylbenzene
(PS-DVB), green fluorescence
� 1 µm, PS-DVB, green fluorescence
� 1 µm, Melamine, red fluorescence
Well techniques
� Poked holes using needle end
� Needle as well, convenient
56
Method 1: Needle
� Drilled and poked holes
� Used various ports and fittings
� Epoxy
� PDMS
Method 2: Ports and Fittings
Figure 2: Early methods for creating wells (a) Needle (b) Ports and Fitting.
� Biopsy punch for wells in PDMS
� Oxygen Plasma Treated
� Polycarbonate lid with wells, easy
access for electrodes
� Aluminum base with viewing window
� Grease between lid and base, vigorous
tapping
Method 3: Housing
Figure 3: Housing (a) ProEngineering model (b) Actual Device
Polycarbonate Lid
Figure 4: To test the affect of Magnalube-G PTFE grease on lid, the lid was
placed on the table with and without grease.
Experimental Set-up
a) b)
Figure 5: To create the high and low flow rates using electrokinetic flow,
different voltages were applied at the end of each reservoir, such as the figure
above.
Figure 6: System used for experiment a) Schematic b) Actual.
a)
b)
Results� General Problems
� Settling of Beads
� Head pressure and back flow
� Some clogging issues
� No bubble trapped in channels
� Method 1: Syringe
� Bubble Formation
� Method 2: Ports and Fittings
� Inconvenient for electrodes
� Affected by bubble formation
� No reproducibility
� Method 3: Housing
� Convenient for electrodes
� Difficult reproducibility
� Lid’s hydrophobicity (Figure 8)
� Electrophoretic effect seen (Figure 9)
Figure 7: Bubble formation at tip of the needle.
Figure 9: When electrokinetic flow worked, beads traveled in the opposite direction as
expected. (a) 1 µm beads polystyrene beads traveled from negative to positive terminal.
Most 15 µm beads followed the direction of flow for small particles; a few went the other
direction (b) 1 µm melamine beads were seen to travel from negative to positive terminal
and vise versa.
a) b)
a) b)
Figure 8: Results of test for polycarbonate lid. (a) Without grease, water would spread on
the table; however, when assembled, there was some leaking and electrokinetic flow did
not occur. (b) With grease, hydrophobic effect was observed. When assembled, vigorous
tapping occasionally allowed connection between water from the lid and and that from
the well.
a) b)
Conclusions
Acknowledgments
I would like to acknowledge Dr. Zahn’s graduate
students, Larry Sasso and Alex Fok.
I would like to thank Rutgers University and Aresty
Research Center for providing funding for this
research.
� First two methods are difficult to implement and
produce results that are not reproducible
� The holder, although easier to assemble, is
difficult to get working because of the lid’s
properties.