XBridge ost C Method guidelines 18 - Waters Corporation · 2008. 4. 15. · [method guidelines ]...

8
[ method guidelines ] XBridge OST C 18 columns are based on Waters second generation of hybrid-silica BEH Technology particles and can be effectively used for the lab scale purification and analysis of detritylated synthetic oligonucleotides using ion-pair, reversed-phase chromatography. This document provides useful method guidelines for the effective use of this column chemistry for this group of compounds. CONTENTS I. PRINCIPLES OF OLIGONUCLEOTIDE SEPARATIONS II. SAMPLE PREPARATION III. RECOMMENDED MOBILE PHASES IV. RECOMMENDED INJECTOR WASH SOLVENT V. GENERAL CONSIDERATIONS IN DEVELOPING SEPARATIONS VI. ANALYSIS OF MODIFIED OLIGONUCLEOTIDES VII. PURIFICATION CONSIDERATIONS XBRIDGE OST C 18 METHOD GUIDELINES *OST: OLIGONUCLEOTIDE SEPARATION TECHNOLOGY

Transcript of XBridge ost C Method guidelines 18 - Waters Corporation · 2008. 4. 15. · [method guidelines ]...

  • [ method guidelines ]

    XBridge™ OST C18 columns are based on Waters second generation of

    hybrid-silica BEH Technology™ particles and can be effectively used

    for the lab scale purification and analysis of detritylated synthetic

    oligonucleotides using ion-pair, reversed-phase chromatography. This

    document provides useful method guidelines for the effective use of

    this column chemistry for this group of compounds.

    Cont ents

    I. PRINCIPLES OF OLIGONUCLEOTIDE SEPARATIONS

    II. SAMPLE PREPARATION

    III. RECOMMENDED MOBILE PHASES

    IV. RECOMMENDED INJECTOR WASH SOLVENT

    V. GENERAL CONSIDERATIONS IN DEVELOPING SEPARATIONS

    VI. ANALYSIS OF MODIFIED OLIGONUCLEOTIDES

    VII. PURIFICATION CONSIDERATIONS

    XBridge™ ost C18 Method guidelines

    *ost: oligonuCleot ide separat ion t eChnology

  • 2

    [ method guidelines ]

    i . p rinCiples oF oligonuCleot ide separat ions Separations of detritylated synthetic oligonucleotides on an XBridge™

    OST C18 column are based on ion-pair, reversed-phase chromatographic

    principles (IP-RP-LC). As shown in Figure 1, the ion-pairing additive in

    the mobile phase is adsorbed on a hydrophobic sorbent and provides

    for charge-to-charge interactions with negative charges contained on

    the oligonucleotide backbone (e.g., phosphate groups).

    Figure 1: Proposed Mechanism of IP-RP-LC for Synthetic Oligonucleotide Separations

    As a result, an efficient charge-based (length-based) oligonucleotide

    separation is achieved (Figure 2). Gradient elution using an acetonitrile

    or methanol eluent displaces both ion-pairing agent and the oligo-

    nucleotides from the sorbent surface.

    Separation selectivity and resolution decreases with increasing

    oligonucleotide length (Figure 2) making the separation of long

    oligonucleotides challenging. Modified oligonucleotides such as phos-

    phorothioates and 2-O alkyl modified species are also more difficult

    to analyze. Special mobile phase may be required (see Section III,

    Recommended Mobile Phases).

    Figure 2: Separation of a 15 - 60mer Deoxythymidine Ladder on XBridge™ OST C

    18

    HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample Injected: Approximately 100 pmoles of a detritylated 15 – 60mer oligonucleotide ladder diluted in 0.1 M TEAAColumn: Waters XBridge™ OST C18, 2.5 µm (2.1 x 50 mm)Mobile Phases: A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80, v/v Flow rate: 0.2 mL/min Column Temp.: 60 ˚C Gradient delay: 0.45 mLGradient: 40 to 62.5% B in 30 minutes (8-12.5% acetonitrile, 0.15% acetonitrile per minute) Detection: 260 nm, 5 scans per second

    Two commonly used ion-pairing agents for oligonucleotide applications

    are triethyl ammonium and dimethylbutyl ammonium ions. The final pH of

    these mobile phases containing either of these ion-pairing reagents is

    adjusted by the addition of Acetic Acid, or in some cases, Hexafluo-

    roisopropanol (HFIP). These mobile phases are volatile making them

    suitable for LC-MS applications.

    The ability to adequately resolve synthetic oligonucleotide mixtures

    by ion-pair, reversed-phase chromatography is significantly affected

    by the particle size of the material contained in an efficiently packed

    column (see Figure 3). Consequently, XBridge™ OST C18 columns are

    efficiently packed with 2.5 micron material to maximize detritylated

    oligonucleotide component resolution. In order to improve oligonucleotide

    separation efficiency and speed, elevated separation temperature (e.g.

    60 ˚C) is recommended. Elevated temperature will also reduce operating

    LC System back pressure.

    ++

    + + - -

    - -

    - - -

    - +

    T EA

    XBridge™ OST C18

    chain PO group on Oligo chain

  • 3

    [ method guidelines ]

    Figure 3: Effectiveness of Waters BEH Technology™ Hybrid-Silica C18

    Particle Size on Deoxythymidine Ladder Separations

    HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample Injected: Approximately 100 pmoles of detritylated 15 – 60mer crude oligonucleotide ladder diluted in 0.1 M TEAAColumn: Waters BEH Hybrid-Silica C18 particles (2.1 x 50 mm)Mobile Phases: A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80, v/v Flow rate: 0.2 mL/minColumn Temp.: 60 ˚C Gradient delay: 0.45 mLGradient: 40 to 62.5% B in 30 minutes (8-12.5% acetonitrile, 0.15% acetonitrile per minute) Detection: 260 nm, 5 scans per second

    In addition to ion-pairing, a hydrophobic reversed-phase mechanism

    also takes place in the oligonucleotide separation. The residual interaction

    of nucleobases has an impact on overall retention and separation

    selectivity, especially when using Triethylammonium Acetate (TEAA)

    ion-pairing mobile phases. Separation of N and N-1mers may be either

    enhanced or suppressed by the sequence contribution. More potent

    ion-pairing systems such as Triethylammonium ion with Hexafluo-

    roisopropanol counter ion provide for more regular “charge-based”

    separations (Figure 4).

    Figure 4: Impact of Ion-pairing System on Separation of a 10-30mer Heterooligonucleotide Ladder

    HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample: 20 mer: TCC CTA GCG TTG AAT TGT CC 25 mer: TCC CTA GCG TTG AAT TGT CCC TTA G 30 mer: TCC CTA GCG TTG AAT TGT CCC TTA GCG GGT Ladder was prepared by hydrolyzing detritylated 20, 25, and 30mer oligonucleotides with a 3’-exonucleaseColumn: Waters XBridge™ OST C18, 2.5 µm (4.6 x 50 mm)Mobile phases: Upper chromatogram: 0.1 M TEAA with acetonitrile gradient; Lower chromatogram: 16.3 mM TEA - 400 mM HFIP with methanol gradientFlow rate: 1.0 mL/min Column Temp.: 60 ˚C Gradient delay: 0.45 mLDetection: 260 nm, 5 scans per second

    i i . saMple p reparat ion

    1. Dissolve the detritylated synthetic oligonucleotide sample in Mobile

    Phase A (e.g., 0.1 M TEAA). For example, a 0.05 - 0.2 µmole scale

    synthesis can be prepared in 0.1 mL of 0.1 M TEAA. Proportionately

    larger or smaller volumes of 0.1M TEAA are required when dissolving

    samples from different scale syntheses. Due to the nature of gradient

    separations, relatively large volumes of sample (in low organic

    strength eluent) can be injected and concentrated onto the head of

    the column before beginning the gradient elution program.

    2. Samples must be completely in solution and free of particulates

    before injecting onto the column. Remove all particles from the

    sample (Controlled Pore Glass Synthesis Support, etc.), which may

    block the inlet column frit, increase the operating pressure, and

    shorten the column life time. Sample contamination with high con-

    centration of salts and/or detergents may also interfere with analysis.

  • 4

    [ method guidelines ]

    3. To remove particulates the sample may be filtered with a 0.2 μm

    membrane. Be sure that the selected membrane is compatible and

    does not dissolve with the selected Mobile Phase diluent. Contact

    the membrane manufacturer with solvent compatibility questions.

    An alternative method of particulate removal involves centrifugation

    for 20 minutes at 8,000 rpm, followed by the transfer of the supernatant

    liquid to an appropriate vial.

    i i i . reCoMMended MoBile phases The most common ion-pair mobile phase for synthetic oligonucleotide

    separations is based on Triethylammonium Acetate (TEAA). This

    mobile phase can be prepared by titrating Glacial Acetic Acid aqueous

    solution with Triethylamine (TEA).

    Note: To maximize column life, it is ESSENTIAL that all prepared OST Mobile Phases be filtered through a solvent compatible, 0.45 µm mem-brane and contained in bottles that are clean and particulate free.

    TEAA1L of 0.1 M TEAA may be prepared as follows:

    1) Perform work in a hood.

    2) Add 5.6 mL of glacial Acetic Acid into 950 mL of water and mix well.

    3) Slowly add 13.86 mL of TEA.

    4) The pH should be adjusted to pH 7 +/- 0.5 by careful addition

    of Acetic Acid.

    5) Adjust final volume to 1 L with water.

    Alternatively, premixed TEAA can be used [(e.g., Sigma 1 M TEAA

    (part no. 90357)]. Mix 100 mL with 900 mL of water to prepare 1 L of

    0.1 M TEAA mobile phase.

    Alternative ion-pairing reagents are recommended for improved sepa-

    ration of phosphorothioates or when performing LC-MS analyses. An

    ion-pairing mobile phase based on Triethylamine (TEA) and Hexafluo-

    roisopropanol (HFIP) as the buffering acid produces an efficient eluent

    system for improved separations involving these application types.

    As indicated below, two ion-pairing systems are useful.

    For routine detritylated oligonucleotide applications, aqueous buffer

    consisting of 8.6 mM TEA and 100 mM HFIP is effective. For applications

    such as those involving the separation of G-rich oligonucleotides, it is

    advisable to use aqueous buffer consisting of 15 mM TEA and 400

    mM HFIP (pH 7.9).

    TEA-HFIP System 11L of 8.6 mM TEA / 100 mM HFIP is prepared as follows:

    1) Perform work in a hood

    2) Add 10.4 mL of HFIP (16.8 g) into 988.4 g of water and mix well.

    3) Slowly add 1.2 mL of TEA.

    4) The pH is approximately 8.3 +/- 0.1.

    TEA-HFIP System 21 L of 15 mM TEA / 400 mM HFIP is prepared as follows:

    1) Perform work in a hood

    2) Add 41.56 mL (67.17 g) of HFIP into 956.36 g of water and mix well.

    3) Slowly add 2.08 mL (1.52 g) of TEA.

    4) The pH of final buffer is approximately 7.9 +/- 0.1.

    iV. reCoMMended inJeCtor Wash solV entsBetween analyses, the HPLC system injector seals should be washed. A

    90% Water / 10% Acetonitrile injector wash solvent is recommended.

    V. general Consideration in deVeloping separationsSeparation of detritylated synthetic oligonucleotides by ion-pair,

    reversed-phase chromatography uses very shallow gradients. With

    both TEAA and TEA-HFIP ion-pairing systems, a rate of strong eluent

    change between 0.1-0.25 % Acetonitrile (or Methanol) per minute is

    recommended. However, the formation of shallow gradients can place

    performance demands on LC pumps and mixers that can compromise

    the quality of the separation. Consequently, it is strongly advised

    that Mobile Phase B formulation contain a premix blend of aqueous

    and organic solvents (e.g., Mobile Phase A= 0.1 M TEAA and Mobile

    Phase B = Acetonitrile / 0.1M TEAA, 20/80, v/v) to minimize poten-

    tially inadequate solvent mixing that can compromise component

    resolution.

  • 5

    [ method guidelines ]

    As illustrated in Figures 5 through 7, these analyses were performed

    with the following mobile phases:

    Mobile Phase A: 0.1 M TEAA

    Mobile Phase B: Acetonitrile (ACN) containing 0.1 M TEAA, 20:80 (v:v)

    The 0.1% ACN / min gradient change from an initial 5 to 10% Acetonitrile

    concentration over 50 minutes was programmed as specified in Table 1:

    Table 1

    Time % A % B Actual Acetonitrile (ACN) Concentration

    0 min 75 25 5%

    50 min 50 50 10%

    Example:

    For the initial 5% Acetonitrile concentration:

    Initial %B = desired ACN % / Volume Fraction of ACN in Mobile Phase

    B. So, initial %B = 5% / 0.2 = 25%

    For the final 10% Acetonitrile concentration:

    Final %B = desired ACN % / Volume Fraction of ACN in Mobile Phase

    B. So, final %B = 10% / 0.2 = 50%

    With TEAA mobile phases, the unmodified oligonucleotides elute

    within a 7-10 % ACN gradient window. However, C and G rich oligo-

    nucleotide sequences are generally less retained (i.e., elute within a

    5-8% ACN gradient window) than A and T rich sequences (i.e., elute

    within a 8-11% ACN gradient span). When using a shallow gradient,

    the total length of analysis for an unknown sample sequence may be

    excessive. Use of a fast scouting gradient with a 1% ACN per minute

    change is recommended in such cases. Information gathered from this

    scouting analysis can then be used to create a more appropriate and

    time efficient set of gradient conditions for the particular sample.

    Gradient slope has a direct impact on the achievable oligonucleotide

    component resolution (along with the type of ion-pairing agent,

    sequence, and oligonucleotide modification). Steeper gradients (e.g.,

    1% ACN change per minute on a 4.6 x 50mm column at a 1.0 mL/min

    flow) are recommended for labeled oligonucleotides or for short, 5-15

    mer sequences. Separation of longer sequences are typically performed

    using more shallow gradient slopes (e.g. 0.15% ACN change per minute

    on a 4.6 x 50mm column at a 1.0 mL/min flow).

    The organic solvent concentration at initial sample loading conditions

    has to be well chosen. If the initial organic solvent strength is too

    high, then some desired oligonucleotide sequences may be unretained.

    In the other extreme, when the gradient starts with too low an organic

    concentration, the analysis is excessively long without the benefit

    of enhanced component resolution. A suitable gradient separation

    method can be approximated from the oligonucleotide base (C, G, A,

    and T) composition. The initial gradient is typically adjusted while

    keeping the gradient slope constant.

    Table 2: Suggested Gradient Conditions for Non-Standard Detritylated Synthetic Oligonucleotide Sequences

    Gradient 1 [Standard

    oligonucleotides (1)]

    Gradient 2 [High GC content or short

    oligonucleotides (2)]

    Gradient 3 [High AT content or long

    oligonucleotides (3)]

    Initial % ACN 7.00% 5.25% 7.50%

    Final % ACN 10.75% 9.00% 12.50%

    Gradient Length(4) 15 min 15min 20min

    1: Standard oligonucleotides: 10 – 30mers

    2: Short oligonucleotides: Less than 10mer

    3: Long oligonucleotides: 30 – 60mers

    4: Assuming use of a 2.1 x 50mm XBridge™ OST C18 column at a flow

    of 0.2 mL/min and a separation temperature of 60 ˚C.

    The retention of single and dual dye-labeled oligonucleotides is dictated

    by the nature of label. For example, the retention of 25 mer oligonu-

    cleotide increases according to the type of label attached as follows:

    no label

  • 6

    [ method guidelines ]

    Figure 5: Analysis of a Digested 25mer Phosphorothioate Oligonucleotide

    HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample: Detritylated 25mer phosphorothioate oligonucleotide mix (CTC TCG CAC CCA TCT CTC TCC TTC T) digested with 3’ exonucleaseColumn: XBridge™ OST C18, 2.5 µm (2.1 x 50 mm)Mobile phase: A: 15 mM TEA with 400 mM HFIP B: methanolFlow rate: 0.2 mL/minColumn Temp.: 60 ˚CGradient delay: 0 mL Gradient: 15 to 20% B in 20 minutes (0.25% methanol per minute)Detection: 260 nm, 2 scans per second

    Peptide nucleic acids (PNA) can also be analyzed using XBridge™ OST C18

    columns. The ion-pairing system recommended for analysis of PNA is

    similar to those used for peptide analysis (0.1% Trifluoroacetic Acid

    or Formic Acid).

    UV detection of eluted oligonucleotide peaks is often performed at

    260 nm. Injection of 50 pmol of detritylated oligonucleotide sample

    on a 2.1 x 50mm XBridge™ OST C18 column yields relatively abundant

    peaks. Limits of quantitation (LOQ) vary with the type of oligonucleotide,

    LC system and detector; LOQ generated on Waters 2996 PDA detector

    equipped with micro UV cell is approximately 1 pmol (2.1 x 50mm

    XBridge™ OST C18). The Limit of detection (LOD) estimate is shown in

    Figure 6.

    Figure 6: Analysis of a 20mer Oligodeoxythymidine Crude Synthesis Mixture

    HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cellSample: ~600 pmol of a detritylated 20mer, ~18 pmol of 19mer, ~4.5 pmol of 17mer was injected on column.Column: XBridge™ OST C18, 2.5 µm (2.1 x 50mm)Mobile phase: A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80 (v/v) Flow rate: 0.2 mL/min Column Temp: 60 ˚CGradient Delay: 0 mLGradient: 35 to 50% B in 30 minutes (7-10% acetonitrile)Detection: 260 nm, 2 scans per second

    V ii. puriF iCat ion Considerat ionsXBridge™ OST C18 columns are designed for laboratory scale oligo-

    nucleotide purifications and analyses. Sufficient amount of isolated

    material suitable for molecular biology and other experiments can

    be prepared in a single injection. For example, a 4.6 X 50 mm XBridge™

    OST C18 column can suitably purify approximately 20-200 nmoles of

    sample in a single injection. It is important to understand that column

    overloading results in a peak broadening and that some earlier eluting

    impurities may co-elute with the component of interest. With a proper

    heart-cutting technique, a good purity of the target oligonucleotide

    can be obtained without significant yield sacrifice (Figure 7).

    Chromatographers frequently develop a separation on the analytical

    scale before moving to preparative work. The steps required to opti-

    mize the analytical separation involve:

    1) Selecting the appropriate column packing material and mobile phase.

    2) Determining the optimal flow rate, gradient during and separation

    temperature.

    3) Determining the amount of material that can be satisfactorily

    loaded and separated on the analytical scale column.

  • 7

    [ method guidelines ]

    Once the separation has been optimized, one begins preparing for the

    preparative separation. The steps to successfully scale a separation

    from an analytical to a preparative column, containing the same packing

    material composition, are detailed below

    Step A: Calculate the flow rate for use on the Preparative column.

    Preparative Column Flow Rate =

    Analytical Column Flow Rate x (Diameter of Prep Column)2

    (Diameter of Anal Column)2

    Step B: To get similar chromatography, the gradient elution profile

    should be created on both columns using the same number of column

    volumes. When the analytical and preparative columns are of the

    same length, as is recommended for this application, then the gradient

    duration should be the same.

    Note: This assumes use of the same flow rate linear velocity for both runs as calculated above. For preparative runs, it is also important to note that an initial gradient delay is required to allow the entire sample to load onto the head of the column prior to beginning chromatography.

    Step C: The last calculation involves determining how much sample

    can be loaded on the preparative column. This calculation compares

    the relative volumes of the two columns assuming that both columns

    are the same length as recommended for this application.

    Preparative Column Sample Load =

    Previously Determined Analytical Column Sample Load x (Diameter of Prep Column)2

    (Diameter of Anal Column)2

    Table 3: XBridge™ OST C18 Column Selection Guide for Detritylated

    Oligonucleotide Purification

    Column (mm) Approx Mass Load (µmoles)** Flow Rate (mL/min)

    2.1 x 50 0.04 0.2

    4.6 x 50 0.20 1.0

    10.0 x 50 1.00 4.5

    19.0 x 50* 4.00 16.0

    30.0 x 50* 9.00 40.0

    50.0 x 50* 25.00 110.0

    * XBridge™ OST C18 Custom Column

    ** Values are only approximate and vary depending on detritylated oligonucleotide length, base composition, and “heart-cutting” fraction collection method used

    Figure 7 shows the separation of 90 nmoles of a detritylated 30 mer deoxythymidine crude reaction mixture on a 4.6 x 50mm XBridge™

    OST C18 column. The collection interval is suggested by the lines. Due to partial column overloading, the N -1, N -2... impurities are partially displaced and elute earlier than expected. With the proper hearth-cutting technique, 95-98% purity is typically achieved for 15-35 mer oligo-nucleotides at this purification scale.

    Figure 7: Purification of a Detritylated 30mer Deoxythymidine Sample

    HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample: Crude detritylated 30mer oligothymidine, 200 nmole dissolved in 100 µl of mobile phase A, 45 µl was injected on columnColumn: XBridge™ OST C18, 2.5 µm (4.6 x 50mm)Mobile phase: A: 0.1M TEAA with 400 mM HFIP B: Acetonitrile/0.1M TEAA, 20/80 (v/v)Flow rate: 1.0 mL/minColumn Temp.: 60 ˚CGradient delay: 0 mL (compensated)Gradient: 35 to 65% B in 24 minutes (7-13% ACN, 0.25% ACN per minute)Detection: 260 nm, 2 scans per second

    Table 4: Ordering Information

    Description Particle Size Pore Size Dimension Part No.

    XBridge™ OST C18 2.5 μm 135Å 2.1 x 50 mm 186003952

    XBridge™ OST C18 2.5 μm 135Å 4.6 x 50 mm 186003953

    XBridge™ OST C18 2.5 μm 135Å 10.0 x 50 mm 186003954

    Custom XBridge™ OST C18

    -- -- -- 186003955

  • Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com

    [ method guidelines ]

    Waters, XBridge, BEH Technology and BioAlliance are trademarks of Waters Corporation.

    ©2008 Waters Corporation. April 2008 715001476 Rev B SC

    Australia Waters Australia Pty. Limited Tel: 2 9933 1777 Fax: 2 9898 1455

    Austria and European Export (Central South Eastern Europe, CIS and Middle East) Tel: 431 877 18 07 Fax: 431 877 18 08

    Belgium Waters S.A.-N.V. Tel: 32 2 726 1000 Fax: 32 2 726 1100

    Brazil Waters Technology Tel: 55 11 5094 3788 Fax: 55 11 5093 6413

    Canada Waters Limited Tel: 800 252 4752 Fax: 905 678 9237

    China Waters Technologies (Shanghai) Limited Tel: 8621 6495 6999 Fax: 8621 6495 1999

    CIS/Russia Tel: +7 495 727 4490 Fax: +7 495 336 7000

    Czech Republic Tel: 42 02 617 11384 Fax: 42 02 617 11386

    Denmark Waters A/S Tel: 45 46 59 8080 Fax: 45 46 59 8585

    Finland Waters Finland Tel: (09) 5659 6288 Fax: (09) 5659 6282

    France Waters S.A Tel: (33) 1 30 48 72 00 Fax: (33) 1 30 48 72 01

    Germany Waters GmbH Tel: 49 6196 400600 Fax: 49 6196 4006010

    Hungary Tel: 36 1 350 5086 Fax: 36 1 350 5087

    Hong Kong Waters China Ltd. Tel: 852 29 64 1800 Fax: 852 25 49 6802

    India Waters India Pvt. Ltd. Tel: 91 80 2 837 1900 Fax: 91 80 2 839 2157

    Ireland Waters Chromatography Ireland Ltd. Tel: 353 1 448 1500 Fax: 353 1 448 1510

    Italy Waters S.P.A. Tel: 39 02 274 211 Fax: 39 02 250 1827

    Japan Nihon Waters K.K. Tel: (81) 3 3471 7191 Fax: (81) 3 3471 7118

    Korea Waters Korea Limited, Tel: (82) 2 820 2700 Fax: (82) 2 820 2730

    Mexico Waters S.A. DE C.V. Tel: 5255 5200 1860 Fax: 5255 5524 9376

    The Netherlands Waters Chromatography B.V. Tel: +31 (0)76-50 87 200 Fax: +31 (0)76-50 87 280

    Norway Waters AS Tel: 47 63 84 60 50 Fax: 47 63 84 60 51

    Poland Tel: (48) 22 833 4400 Fax: (48) 22 833 0987

    Puerto Rico Waters Technologies Corporation Tel: 787 747 8445 Fax: 787 747 8448

    Singapore Waters Asia Ltd. Tel: 65 6273 1221 Fax: 65 6273 6116

    Spain Waters Cromatografía, S.A. Tel: 34 93 600 93 00 Fax: 34 93 600 93 60

    Sweden Waters Sverige AB Tel: 46 8 555 11500 Fax: 46 8 555 11550

    Switzerland Waters AG Tel: 41 62 889 2030 Fax: 41 62 889 2059

    Taiwan Tel: 886 2 2543 1898 Fax: 886 2 2543 1918

    United Kingdom Waters U.K. Ltd. Tel: 44 208 238 6100 Fax: 44 208 207 7070

    All Other Countries

    Waters Corporation 34 Maple Street Milford, MA 01757 USA Tel: 508 478 2000 800 252 4752 Fax: 508 872 1990

    http://www.waters.com

    sales oFF iC es