Without reference, identify principles about volume/pressure and high frequency ventilators with at...
-
Upload
bernadette-neal -
Category
Documents
-
view
223 -
download
1
Transcript of Without reference, identify principles about volume/pressure and high frequency ventilators with at...
Without reference, identify principles about volume/pressure and high frequency ventilators with at least 70 percent
accuracy.
Purpose of Volume/Pressure Ventilators ◦ Device used to move gas into the lungs ◦ Required when
Acute respiratory failure No absolute rule
Volume/Pressure and High Frequency Ventilators
Categories ◦ Based on the type of pressure used to move gas
into the lungs ◦ Negative pressure ventilators
Apply a sub-atmospheric pressure around the chest
Air at atmospheric pressure is drawn in An example is an iron lung Major disadvantages
Cannot provide adequate ventilation in all patients Problems with fit and comfort during long-term use
Volume/Pressure and High Frequency Ventilators
◦ Pressure ventilators Apply a positive pressure directly to the airway Similar to mouth-to-mouth artificial ventilation Provides better control over ventilation Most common Disadvantage is that the increased pressure in the
thoracic cavity decreases venous blood return to the heart
Volume/Pressure and High Frequency Ventilators
Patient Connection Methods ◦ Face mask
Disadvantages Difficult to get a good seal Airway blockage may occur
Acceptable for short term emergency use only◦ Intubation
A rigid tube inserted into the trachea Types
Endotracheal tubes may be inserted through the mouth or nose
Tracheostomy tube is surgically inserted into the trachea o Use a balloon-like cuff which seals against the walls of the
trachea All gas flow is through the intubation tube
Volume/Pressure and High Frequency Ventilators
Limiting Devices - Required to prevent barotraumas (lung damage)o Pressure limiter
Pressure developed by the drive mechanism may be high enough to injure the lungs
Methods Vent excess pressure End the inspiration completely
Volume/Pressure and High Frequency Ventilators
◦ Flowrate control When the generated pressure is high there
must be some means of controlling flow rate Methods Increasing resistance With injectors, decreasing the driving pressure
Volume/Pressure and High Frequency Ventilators
Phases of the Ventilation Cycle ◦ Inspiratory phase - moves gas into the lungs ◦ Changeover from inspiration to expiration
Cycling mechanisms determine when to end an inspiration and begin an expiration
There are four ways of cycling the ventilator into an expiratory phase Pressure cycled ventilators o Inspiration is ended when a predetermined pressure is
reached o Volume, flow rate, and inspiratory time may all vary from
one inspiration to another
Volume/Pressure and High Frequency Ventilators
Flow cycled ventilators o Inspiration is ended when the flow of gas
into the patient falls below a specific point oVolume, pressure, and inspiratory time
may vary from one inspiration to another Volume cycled ventilators o Inspiration is ended when a predetermined
volume of gas has been delivered to the patient
oPressure, flow rate, and inspiratory time may vary from one inspiration to another
Volume/Pressure and High Frequency Ventilators
Time cycled ventilators o Inspiration is ended after a predetermined time o Volume, pressure, and flow rate may vary from one
inspiration to another
Inspiratory supplemental functions Sigh- a periodic deep breath, usually delivered 6 to
10 times an hour Inflation hold o It is a period of time at the end of an inspiration that the
lungs are kept inflated o Also known as inspiratory hold, inspiratory pause, or
inspiratory plateau
Volume/Pressure and High Frequency Ventilators
◦ Expiratory phase Allows the lungs to empty
Normally not restricted by the ventilator
Adjuncts of the expiratory phase Expiratory resistanceo Application of mechanical resistance to the exhalation
port of the breathing circuit o Increases the mean airway pressure
Volume/Pressure and High Frequency Ventilators
◦ Positive end-expiratory pressure (PEEP) Maintains a positive pressure in the lungs
throughout the respiration cycle Used during mechanical breathing Purposes
Increases the mean airway pressure and functional residual capacity
May improve lung compliance
Volume/Pressure and High Frequency Ventilators
◦ Continuous positive airway pressure (CPAP) Maintains a positive pressure in the lungs
throughout the respiration cycle Used during spontaneous breathing Purposes:
Keeps airways open Increases the mean airway pressure and functional
residual capacity
Volume/Pressure and High Frequency Ventilators
◦ Changeover from expiration to inspiration Cycling mechanisms determine when to end
an expiration and begin an inspiration Three ways of cycling the ventilator into an
inspiratory phase Manual cycling - expiration is ended and inspiration is
begun by operator intervention Patient cycled o Started when the patient attempts to take a breath o Once initiated the ventilator delivers a controlled breath o Breath is known as an "assisted" breath o The sensitivity or trigger sensitivity parameter must be
set to the patient's efforts
Volume/Pressure and High Frequency Ventilators
Timed cycled o Inspiration begins after a predetermined time has passed o Timer is independent of the patient o Controlled by the "breaths/min" setting o Classified as a "strict controller" if no assist mechanism is
used o Often used as a backup to patient cycling
Volume/Pressure and High Frequency Ventilators
Special Ventilator Modes ◦ Developed to make it less tedious to wean patient from
the ventilator ◦ Intermittent mandatory ventilation (IMV)
Allows the patient to breathe spontaneously between controlled breaths
May be divided into two phases IMV phase Spontaneous phase
Controlled IMV breath is delivered at the beginning of the IMV phase
Disadvantage is that the ventilator can inspire against the patient Uncommon
Volume/Pressure and High Frequency Ventilators
◦ Synchronized intermittent mandatory ventilation (SIMV) Allows the patient to breathe spontaneously
between assisted breaths, with a mandatory breath as backup
May be divided into two phases SIMV phase Spontaneous phase Assisted breath in SIMV phase is synchronized to
patient efforts If no assisted breath is taken during the SIMV phase o A mandatory SIMV breath will be given o At the beginning of the next spontaneous phase
Volume/Pressure and High Frequency Ventilators
◦ Mandatory minute ventilation (MMV) • Additional option to SIMV • Number of mandatory breaths is increased or decreased
dynamically • Used to ensure a specific minute ventilation
High Frequency Ventilation (HFV) ◦ Uses breath rates of 1 HZ to 20 HZ◦ Purpose
Low diffusionPatient requires higher pO2 (partial pressure of oxygen) gradient to force
O2 to diffuse Allows the Paw to be reduced but keeps pO2 high
InfantsRequire a higher breathing rate Have a lower compliance
Volume/Pressure and High Frequency Ventilators
GENERAL CALIBRATION PROCEEDURES FOR VOLUME/PRESSURE VENTILATORS◦ Ensure annual manufacturer’s calibration ◦ Ensure equipment is designed to accurately measure the
delivery from the unit under test High flow adult Low flow pediatric High frequency
◦ Obtain ventilator information What settings must the ventilator be placed in to measure
volumes and flows Test device must be set accordingly to ensure proper
measurements Failure to do so may result in the maladjustment of the
ventilator
Volume/Pressure and High Frequency Ventilators
Ventilator Measurement Standards ◦ Flow and volume measurements are affected by
Type of gas (air, oxygen or mixed) Temperature Pressure Water vapor
◦ BTPS – Body Temperature Pressure Saturated Body temperature – 37° C Pressure – ambient barometric pressure Saturated – water vapor at 100% relative humidity Usually used when the flow of gases has been
changed to patient conditions (exhalation)
Volume/Pressure and High Frequency Ventilators
◦ STPD – Standard Temperature Pressure Dry Standard temperature – usually 0 or 21° C Pressure – ambient barometric pressure Dry – water vapor at 0% relative humidity Usually used when the flow of gases has been
changed to standard conditions (wall supply/inhalation)
Volume/Pressure and High Frequency Ventilators