Slide 1 Chairperson: John Vis Welcome to our E-Seminar: Injection of high-concentration and high-volume samples in preparative HPLC Presenter: Ulrik Wittek

Transcript of Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables...

Page 1: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 1Chairperson: John Vis

Welcome to our E-Seminar:

Injection of high-concentrationand high-volume samples in preparative HPLC

Presenter: Ulrik Wittek

Page 2: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 2Chairperson: John Vis

Contents

• IntroductionColumn loading and overloadingConcentration overloadingVolume overloading

• High-concentration samplesSandwich injectionOrganic phase injection

• High-volume samplesSystem with injection pump

Page 3: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 3Chairperson: John Vis

Column loading and overloading

Goal:• Quantification of compounds

Important parameters:• Resolution• Peak symmetry• Peak width

mAU

Peakheight h

4σ0.13 h

0.61 h

t

amount

k’

σv

Capacity factor

Peak width σv

10

' −=ttk

Page 4: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 4Chairperson: John Vis

Column loading and overloading

mAU

t

Goal:• Separation of compounds

Important parameters:• Throughput• Yield• Purity

mAU

amount

k’

σv

k’Ana

lytica

l

Preparative

t

Page 5: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 5Chairperson: John Vis

Column loading and overloading

Concentration Overloading Volume Overloading

• Determined by solubility of compound in mobile phase

• “Preparative” area of adsorption isotherm

• Throughput determined by selectivity

• Particle size of stationary phase of low influence

• Determined by injection volume

• “Analytical” area of adsorption isotherm

• Throughput determined by column diameter

• Small particle size required

Usually a combination of concentration and volume overloading is done. Concentration overloading is preferred because higher throughput is possible.

Page 6: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 6Chairperson: John Vis

Column loading and overloading

Concentration overloading• Sample not soluble in mobile phase of gradient starting conditions• Precipitation, leads to blocking of the flow path

Volume overloading• Volume limited by max. injection volume of autosampler and

sample container

Page 7: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 7Chairperson: John Vis

Column loading and overloading

Vinj

Vinj

Vinj Vinj

Analytical run

Volume overloading

Vinj Concentration overloading

Page 8: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 8Chairperson: John Vis

Break Number 1

Please type your

question into the

Chat Box at any time

during the presentation.

Page 9: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 9Chairperson: John Vis

High-concentration samples

When injecting high-concentrated samples, precipitation in the injection valve can be a problem. This blocks the injection valve and it has to be taken apart and cleaned.

Reason:• Sample is dissolved e.g. in DMSO• Starting composition of the gradient is e.g. water/acetonitrile 90:10• As soon as the sample mixes with the mobile phase precipitation

occurs

Sample

Page 10: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 10Chairperson: John Vis

Sample precipitation

SampleSolvent Solvent

SampleSolvent Solvent

SampleSolvent Solvent

SampleSolvent Solvent

Page 11: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 11Chairperson: John Vis

Sandwich injection

Sandwich Injection:• Sample is sandwiched between two plugs of the sample solvent.• Mixing with mobile phase occurs only at the beginning and the

end of the plugs, where the sample concentration is zero.

SampleSolvent SolventPlu

g

Plu

g

SampleSolvent Solvent

Page 12: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 12Chairperson: John Vis

Injector program

• Sandwich is generated using a ChemStation injector program.• A vial containing the sample solvent must be placed into the injector.

Plug

Plug

Sample

Page 13: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 13Chairperson: John Vis

Application example

min0 2 4 6 8

mAU

100200300400

Standard injection (500 µl)

min0 2 4 6 8

mAU

100

200

300Sandwich injection (500 µl)Plugs: 50 µl

min0 2 4 6 8

mAU

100

200

300Sandwich injection (500 µl)Plugs: 200 µl

Early eluting peaks might front/tail!

Page 14: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 14Chairperson: John Vis

Advantages of sandwich injection

Advantages of sandwich injection:• Sample does not come in contact with the mobile phase until it

reaches the column. Therefore no precipitation in the critical part of the flow path can occur.

• No hardware re-configuration required. Sandwich injection can be done on a standard system using an injector program.

Page 15: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 15Chairperson: John Vis

Disadvantages of sandwich injection

Disadvantages of sandwich injection:• The sample must be retained on the column in a rather narrow

band. The strong sample and plug solvent leads to a significant band broadening of the sample on the column head.

• If the ratio of plug and sample volume compared to the column volume is too large some of the compound can elute with the solvent front as column breakthrough.

• The plug and sample solvent might disturb the equilibrium of the stationary phase.

Page 16: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 16Chairperson: John Vis

Organic phase injection

• Sample is injected onto the column using only the organic phase.• Mixing with aqueous phase occurs on the column (in the T-piece

shortly before the column).

PumpAqueous phase

PumpOrganic phase Injector

ColumnNo mixer!

Page 17: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 17Chairperson: John Vis

Organic phase injection

• Due to the design of our preparative pump, two isocratic pumps form a gradient pump, the system setup is very simple.

• Since the two physical pumps are treated as one binary pump in the software gradient programming is very easy (no flow-gradient programming)!

Page 18: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 18Chairperson: John Vis

Application example

min0 2 4 6 8

mAU

100

200

300

400Standard injection (500 µl)

min0 2 4 6

Peak broadening due to low mixing performance

mAU

100

200

300

400

Organic phase injection (500 µl)

8

Page 19: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 19Chairperson: John Vis

Advantages of organic phase injection

Advantages of organic-phase injection:• No strong sample solvent is required, which could lead to

additional peak broadening or column breakthrough.

• Since the Agilent 1100 Series preparative pump in the gradient version consists of two physical pumps anyway the gradient programming is very easy. It can be set up in the same way as for the standard configuration, no flow gradient programming is necessary.

Page 20: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 20Chairperson: John Vis

Disadvantages of organic phase injection

Disadvantages of organic-phase injection:• Sample precipitation can still occur in the mixing tee or on the

head of the pre-column. While precipitation on the column head usually leads only to an increased pressure clogging in the mixing tee is critical.

• Setting up the purification system for organic-phase injection requires re-plumbing of the system. That means it is not possible to perform organic-phase and standard injection on the same system without hardware changes.

Page 21: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 21Chairperson: John Vis

Which technique should be used?

Sandwich injection:• For highly-concentrated samples where chromatographic

performance is not critical.

Organic phase injection:• For medium-concentrated samples where good chromatographic

performance is required.

Page 22: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 22Chairperson: John Vis

Break Number 2

Please type your

question into the

Chat Box at any time

during the presentation.

Page 23: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 23Chairperson: John Vis

High-volume samples

• The usage of our autosampler for sample injection is limited to 5 ml. This is OK for most preparative work on 1 inch columns and flow rates of 20 - 35 ml/min.

• Especially for larger columns (but also for 1 inch columns) higher sample volumes have to be injected. Therefore usually an injection pump is used.

• In our system an isocratic (preparative) pump and a 2-position/6-port valve can be used as injection pump.

• Basically two approaches: Pump fills an injection loop, which is then switched into the flow or direct pumping of the sample onto the column. Second approach is favored.

Page 24: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 24Chairperson: John Vis

System configuration

Column

Waste

Injection valvePos. 1: Gradient pump Injection pump

Gradient pump

Sample valve

Page 25: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 25Chairperson: John Vis

Operation

1. At the start of the run the injection valve is in position 1, in which the flow from the gradient pump goes to the column and the flow of the injection pump goes to waste.

2. When the run is started the injection valve switches to position 2 where the flow from the injection pump goes to the column. The injection pump transfers the sample onto the column.

Column

Waste

Injection valvePos. 1: Gradient pump Injection pump

Gradient pump

Sample valve

ColumnColumnColumn

Waste

Injection valvePos. 1: Gradient pump Injection pump

Gradient pump

Sample valve

3. When the sample is applied to the column the injection valve switches back to position 1 and the gradient is started. During the run the injection pump and the injection valve are washed with an appropriate solvent and are then re-equlibrated.

Page 26: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 26Chairperson: John Vis

Sample injection system

• Pos. 1: Starting position, Solvent A gradient start composition• Pos. 2: Sample• Pos. 3: Solvent A to flush sample completely onto column• Pos. 4: Wash solvent to rinse injection valve and injection pump

• Pos. 5: Solvent A to remove wash solvent from inj. pump and inj. valve

Solvent ATo injection valve

Sample Injection pumpSolvent AWash

Sample valveSolvent A

Page 27: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 27Chairperson: John Vis

Pumps and valves timetables

Time Injection Injection pump Sample Comment(% B) Flow (mL/min) valve pos. Flow (mL/min) valve pos.

0.0 5 20 1 0 1 Starting conditions0.1 20 2 Injection valve to injection pump pos.0.2 1 0 2 Sample valve to sample pos., lower gradient pump flow0.3 10 Start injection pump

* 1.3 1 3 Sample valve to solvent A pos., transfer sample onto column** 1.8 5 20 1 Injection valve to gradient pump pos., increase gradient pump flow

1.9 4 Sample valve to wash solvent pos.** 2.4 5 Sample valve to solvent A pos., transfer sample onto column** 2.9 10

3.0 0 Decrease flow of injection pump11 95 End of gradient

Gradient pump

* timing depends on sample volume to be injected** timing depends on injection system volume (capillaries, inj. pump, valve)

Page 28: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 28Chairperson: John Vis

Sandwich injection with the injection pump system

Solvent A To injection valvePlug solvent Injection pump

SamplePlug solvent Sample valveSolvent AWashSolvent A

Page 29: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 29Chairperson: John Vis

Scope and limitations

• ChemStation only (A.10.01), system cannot be used with Purify.

• Time-, peak- and mass-based fraction collection

• Injection pump cycle controlled by injection pump and valve time-table.

• Delay volume calibration procedure not available without autosampler.

Page 30: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 30Chairperson: John Vis

Scope and limitations

• A blank run must be set up in the ChemStation since there is no autosampler in the system to give a start pulse.

• The run starts not with the injection (switching of the injection valve back to the gradient pump position) but with the start of the injection cycle. Advantage: If compound of interest brakes through the column it is still collected!

• For connecting solvent A, sample and wash solvent to the valve capillaries with rather large diameter (1.5 mm i.d., e.g. from solvent selection capillary kit G1160-68706) must be used. The capillary from the sample to the sample valve should be as short as possible.

Page 31: Welcome to our E-Seminar - Agilent · Chairperson: John Vis Slide 27 Pumps and valves timetables Time Injection Injection pump Sample Comment (% B) Flow (mL/min) valve pos. Flow (mL/min)

Slide 31Chairperson: John Vis

Injection of high-concentration and high-volume samples

• Column loading and overloadingConcentration and volume overloadingLimiting factors of column overloadingCharacteristic peak shapes

• High-concentration samplesProblems with high-concentration samplesSandwich injectionOrganic phase injectionAdvantages and disadvantages of both methods

• High-volume samplesConfiguration of a system with injection pumpSample injection systemSandwich injection with the injection pump