Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

17
Why routine dosimetry doesnt work Why routine dosimetry doesn t work Scott McFeeters Industrial Hygienist Naval Medical Center San Diego

Transcript of Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

Page 1: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

Why routine dosimetry doesn’t work Why routine dosimetry doesn t work

Scott McFeeters

Industrial Hygienist

Naval Medical Center San Diego

Page 2: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

Is our job only to document hearing loss???

Page 3: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` Objective ` Background ` The Team ` Types of Noise ` Characteristics of Weapon Fire ` D t CData Compariison ` Future Studies `̀ ReferencesReferences ` Questions

Page 4: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` ` Raise awareness to the limitations of personal Raise awareness to the limitations of personal

noise dosimeters

Page 5: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

`

` NIOSH and other health researchers have documented the inadequacy of using personal noise dosimeters to measure impulse noise during weapons firingimpulse noise during weapons firing.

` Our study is in the process of characterizingOur study is in the process of characterizingimpulse noise from a variety of military issuefirearms at indoor and outdoor firing ranges.

Page 6: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

`

`

` Scott McFeeters, Industrial HygienistScott McFeeters, Industrial Hygienist ` Jane Nowell, MS, CIH

` Leif Olsen, MPH, CIH

` Tom Hartsog Industrial Hygiene Technician Tom Hartsog, Industrial Hygiene Technician

` Michelle Dewitt, Industrial Hygiene Technician

Page 7: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` Impulse, impact, peak are often used interchangeably ◦ Short ◦ High frequency ◦ High frequency

` Less than 1 sec in duration ◦ Mayy re ppeat after delayy of 1 sec

Page 8: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` Continuous noise:

industrial machinery ` Impulse noise: door

slamming automobileindustrial machinery, aircraft and auto engines, landscaping

slamming, automobile backfire, weaponfiring, riveting, drop

equipment, etc. forging, etc.

Page 9: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` Rappid

` Short duration

` Large change in instantaneous sound

pressure level

◦ Rapid expansion of gas

`̀ Impulsive Noise Impulsive Noise

Page 10: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` i. Measure impulse noise levels using calibrated SLMs that: ◦ (1) Meet or exceed specifications in Reference (i).

◦ (2) Have a peak hold circuit.

◦ (3) Have a rise time not exceeding 35 microseconds.(3) Have a rise time not exceeding 35 microseconds.

◦ (4) Are capable of measuring peak SPLs in excess of

140 dB peak SPL140 dB peak SPL.

Page 11: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` 130-175 dB ` Duration of few usec

` Decay of 10 msec or more

Instrument Maximum SPL Peak Response Time

Personal Dosimeter

140 dBP 50 µsec

Type I SLMType I SLM Quest Sound Pro ¼” mic >165 dBP 12-18 µsec

Larson Davis LXT ¼” mic >165 dBP 30 µsec Larson Davis 824 ¼” mic >165 dBP 50 µsec

Page 12: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

141.20dBP Personal Noise DosimeterPersonal Noise Dosimeter

M16-A4 Rifle Single Shot Peak Pressure Clipping

Page 13: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

164dBP SoundPro Type 1 SLMSoundPro Type 1 SLM164dBP

M16-A4 Rifle Single Shot Peaks

Page 14: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

`

` Collect peak data ` Multiple shooters ` Sample in personalSample in personal

hearing zone ` Control booth noise

attenuationattenuation ` Document noise

reduction from the apppplication of acoustical material

Page 15: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

` Berger, E.H., L.H. Royster, and J.D. Royster. The Noise Manual. Fairfax: American Industrial Hygiene Association 2003Fairfax: American Industrial Hygiene Association, 2003.

` Kardous, Chucri. "Limitations of Using Dosimeters in Impulse Noise Environments". Journal of Occupational and Environmental Hygiene July 2004: 456-462.

` Kardous, Chucri. "Noise Exposure Assessment and AbatementStrategies at an Indoor Firing Range". Applied Occupational and Environmental Hygiene 2003: 629-636.

` M h Willi "A t f N i E f IMurphy, William. "Assessment of Noise Exposure for Inddoor and Outdoor Firing Ranges". Journal of Occupational and Environmental Hygiene September 2007: 688-697.

Page 16: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work

Is our job only to document hearing loss???

Page 17: Weapon Noise Assessment: Why Routine Dosimetry Doesn't Work