DENTAL X-RAY INSTALLATION INSTRUCTIONS (for USA & Canada) MODEL 096 REV.4 ·Wall Mount Type .................... WK ·Room Mount Type .................. RK

Transcript of ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate...

Page 1: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

DENTAL X-RAY

INSTALLATIONINSTRUCTIONS (for USA & Canada)

MODEL 096

REV.4

·Wall Mount Type ....................WK·Room Mount Type ..................RK

Page 2: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

INDEX PAGE

SECTION ONE : TECHNICAL DATA [1]ELECTRICALANDRADIATIONDATA......................................................... 3 [2]PHYSICALDIMENSIONS............................................................................... 4 [3]TUBEHEADTHERMALCHARACTERISTICS............................................ 5

SECTION TWO : PRE-INSTALLATION INSTRUCTIONS [1]SUPPORTREQUIREMENTS........................................................................... 6 [2]ELECTRICALREQUIREMENTS.................................................................... 6 [3]LOCATIONOFCOMPONENTS...................................................................... 7

SECTION THREE : INSTALLATION INSTRUCTIONS [1]INSTALLATIONREQUIREMENTS................................................................ 8 [2]UNPACKING..................................................................................................... 8 [3]ARMASSEMBLYINSTALLATION................................................................. 9 [4]HEADASSEMBLYINSTALLATION............................................................ 12 [5]CONTROLBOXINSTALLATION................................................................. 12

SECTION FOUR : POST INSTALLATION INSPECTION [1]ARMASSEMBLY............................................................................................ 14 [2]BALANCEARMASSEMBLY........................................................................ 15 [3]HEADPOSITIONING..................................................................................... 15

SECTION FIVE : CONTROL IDENTIFICATION AND OPERATION [1]CONTROLIDENTIFICATION....................................................................... 16 [2]FUNCTIONOFCONTROLS.......................................................................... 16 [3]OPERATINGPROCEDURES......................................................................... 18 [4]ERRORCODES............................................................................................... 19

SECTION SIX : POST INSTALLATION CONFIRMATION [1]CONFIRMATIONOFPOWERSUPPLYVOLTAGE..................................... 20 [2]CONFIRMATIONOFTUBECURRENT....................................................... 20 [3]CONFIRMATIONOFEXPOSUREWARNINGLAMP&BUZZER............ 20 [4]CONFIRMATIONOFLINEVOLTAGEREGULATION.............................. 20

SECTION SEVEN : INITIAL SETTING [1]FILMSPEED.................................................................................................... 21 [2]PRIORITYOFSELECTIONS......................................................................... 23 [3]ELECTRONICCHIMEON/OFF.................................................................... 23

APPENDIX ONE : CIRCUIT DIAGRAM........................................................... 23

APPENDIX TWO : PARTS IDENTIFICATION [1],[2]ARMANDHEAD/CONTROLBOXASSEMBLY................................ 24 [3]SLIDINGSECTION/ARMANDHEADASSEMBLYOFRKTYPE.......... 26

APPENDIX THREE : CERTIFICATION............................................................ 28

CAUTIONThismanualprovidesinformationandinstructionsfortheinstallation,assemblycalibrationand certification procedures for BELMONT BELRAY MODEL 096dentalx-ray.Theinstructionscontainedinthisbookshouldbethoroughlyreadandunderstoodbeforeattempting to install the X-ray unit. After the installation is complete, file this manual andreferbacktoitwhenperformingperiodicmaintenance.

Page 3: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

SECTIONONE:TECHNICALDATA

[1] ELECTRICAL AND RADIATION DATA

1.Focalpointmeasurement.............................................. 0.8mmx0.8mm2.Ratedpeaktubepotential.............................................. 70kVp3.Ratedtubecurrent......................................................... 10mA4.Maximumratedpeaktubepotential............................. 70kVp5.Ratedlinevoltage......................................................... 120VAC6.Linevoltagerange......................................................... 108VAC~132VAC7.Rangeoflinevoltageregulation................................... 2~5%8.Ratedlinecurrent.......................................................... 10.8Aat70kVp,10mA9.Maximumlinecurrent................................................... 11.9Aat70kVp,10mA10.Exposuretime............................................................. 0.02~3sec. (ONandOFFarezerocrossed.)11.Timeraccuracy............................................................ ±1pulse(1/60sec.)12. Inherent filtration........................................................ 1.3mmAlEquivalent13. Added filtration........................................................... 0.8mmAl14. Minimum filtration permanently in useful beam........ 2.1mmAlEquivalentat70kVp15.Nominalroentgenoutputa.Distalendofregularcone........................................... 8.2mGy/sec.+30%,-40%b.Distalendoflongcone............................................... 3.7mGy/sec.+30%,-40%(Dataobtainedbydirectmeasurementintheusefulbeam)16.Sourcetoskindistancea.Regularcone................................................................ 204mmb.Longcone.................................................................... 305mm17.Leakagetechniquefactor............................................ 70kVp/0.16mA 0.16mAismaximumratedcontinuouscurrent for10mAwithadutycycle1:6018.Dutycycle................................................................... 1:60(0.5sec.exposurewith30sec.interval)19.Maximumdeviationoftubepotentialandtubecurrent Pulse TubePotential TubeCurrent +8 1st,2nd&3rd 70kVp 10±2mA -11

+7 4th&Up 70kVp 10±1mA -1020.Measurementbaseoftechniquefactorsa.peaktubepotential...................................................... Peaktubepotentialofconductinghalfcycleb.tubecurrent................................................................. Averageoftubecurrentduringonecycleof linefrequencyc.exposuretime.............................................................. Impulsesofpowerlinefrequency21.Halfvaluelayer........................................................... 1.5mmAlover22.Sourcetothebaseofconedistance............................ 81mm

Page 4: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 4 -

[2] PHYSICAL DIMENSIONS( ) = mm

096-WK

096-RK

From2004July

600˚

4-3/

16"

(106

)

300˚

Max

. 25-

9/16

"(6

50)

Max

. 58-

1/16

"(1

475)

19-11/16"(500)

19-1

1/16

"(5

00)

Str

oke

27-9

/16"

(700

)

17-1

/16"

(450

)

Str

oke

7-7/

8"(2

00)

30-1

/2"

(775

)

3-9/16"(90)

7-1/2"(190)

20"(507)

4-3/16"(106)

Max

. 84-

1/4"

(214

0)

15-3/4"(400)

13-3/4"(350)

180˚

260˚

220˚

600˚

40-5

/8"

(103

2)

4-1/8"(104)

31-1/2"(800)

3-9/16"(90)Max. 46-7/8" (1190)

Stroke : 43 5/16" (1100)OPTION: 20" (500) or 12" (300)Horizontal Armis available.

300˚Max.75"(1904)

Min

. 19-

1/4"

(48

9)

Max

. 62-

9/16

" (1

589)

Min

. 34-

5/8"

(87

9)

M

ax. 7

7-15

/16"

(19

79)

11"(280)

3-15

/16"

(100

)

17-1

1/16

"(32

2)

3-11

/16"

(93)

47-1

/16"

(119

5)

40-5

/8"(

1032

)

42-1

/2"(

1080

)

48"(

1219

)

39"(

990.

6)

SH

OR

T C

ON

E :

15-

15/1

6"(4

04)

LON

G C

ON

E :

19-

7/8"

(505

)

46-11/16"(1185)

40-1/16"(1017)6-1/2"(165)

9-1/8"(232)

10-5/16"(261)

CENTER OFELECTRIC

Page 5: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

[3] TUBE HEAD THERMAL CHARACTERISTICS A. Interval between each exposure

Thetemperatureinsideofthetubeheadrises,whenanexposureismade.ThevalueoftheheatgeneratedismeasuredinHeatUnit(HU),whichistheproductoftubepotential,tubecurrentandexposuretime.Excessiveheatwillbeaccumulatedinsideofthetubehead,ifthex-rayisusedwithoutapropercooldownintervalbetweeneachexposure.Theexcessiveheatmaydamagethex-raytube,highvoltagegeneratororboth.

B. Duty cycleToavoidtheaccumulationofexcessheatinanefforttoprolongthetubeheadlife,acooldownintervalof60secondsormoremustbeallowedbetweeneach1secondexposure.ora30secondcooldownmustbeallowedbetweeneach0.5second(30impulses)exposure.

C. Tube head cooling curve 1. Tube Housing cooling curve 2. Anode thermal characteristics

3. Maximum rating chart

1H

U/S

HU/S

1HU/S

TUBE MODEL D-0

ANODE THERMAL CHARACTERISTICS

Hea

t Sto

rang

e (H

.U.)

Time in Minutes0 1 4 6

0000

1000

14000

10000

6000

000

FOCAL SPOT 0.8mm

0 kV

60 kV

0 kV0 kV

90 kV

Tube

Cur

rent

(mA)

Load Time (sec.)0.1 0. 0. 0. 0. 1.0 .0 .0 .0

0

0

10

Tube Housing Cooling Curve

Heat

Sto

rang

e (K

H.U.

)

Time in Minutes0 0 40 60 0 100 10 140

0

00

10

100

0

Page 6: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 6 -

SECTIONTWO:PRE-INSTALLATIONINSTRUCTIONS[1] SUPPORT REQUIREMENTS

Control box:When mounting the model 096 control box, the wall and mounting hardware must be sufficient towithstanda25poundshearload.

Arm and head:TheMODEL096wallplateisdesignedformountingontwo2x4woodstudswith16inchescenters. For other types of construction, the wall and mounting hardware must be sufficient to withstanda100poundshearloadanda450poundwithdrawalforceateachofthefourmountingbolts.Thearmmountingbracket,partNo.39,mustbemountedtothewallplateusingtheboltssuppliedwiththex-ray.

CAUTION:If the MODEL 096 is to be mounted in a manner other than what is specified in this manual or ifthe hardware to be used is other than what is supplied, the support capability of the wall andthe strength of the hardware must be checked and verified to be adequate.

[2] ELECTRICAL REQUIREMENTS Power supply:

TheMODEL096x-raysystemwilloperateonapowersupplyof120VAC,+12VAC,-12VAC(108VACto132VAC).A3wireGROUNDEDcircuit,separatelyconnectedtothecentraldistributionpanelwithanover-currentprotectiondeviceratedfor15amperes.Recommendedwiresizeis12AWGBUTifthewirerundistanceistoexceed50feet10AWGisrequired.Forwirerundistancesinexcessof75feet8AWGisrequired.Linevoltageregulationmustbewithin2~5%at10.8amps.

Interconnecting wiring, control box to arm and head assemblies:6conductor12AWGisrecommendedforwirerundistancesupto50feet.Forwirerunsbetween50and75feet10AWGisrequiredandforwirerundistancesinexcessof75feet8AWGisrequired.

Concealed wiring for WK type:Concealed wiring is accomplished by bringing conduit and wires into (2) flush mounted junction boxeslocated(1)behindthecontrolboxand(1)behindthearmmountingplate.Recommendedheightsfortheflush junction boxes are : 51 5/8" for behind control box and 44 1/2" for behind wall plate. Wiring done inthis manner should extend 12 inches beyond the wall surface to allow sufficient wire for connections.

NOTE:Allconnections,workmanshipandmaterialsusedmustcomplywiththelocalcodes.

Junction Boxes

FROMPOWERSOURCE

Terminal Blockfor Control Box

Pigtail for ArmMounting Bracket

(1) ()

WALLMin 1AWG

6 4 L N 4 6

Page 7: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

[3] LOCATION OF COMPONENTSA. Arm and head assemblies for WK type:

UsingtheinformationprovidedinFIGURE1,determinethecorrectlocationfortheinstallationofthearmandheadassemblies.NOTE:Stateandlocalrequirementssupersedeguidelinesindicatedbelow.

B. Control box :Whendeterminingthelocationforthecontrolboxthefollowingradiationrequirementsconcerningoperatorpositioningmustbeconsidered.Theoperatormust;

1.havefullviewofthepatient.2.havefullviewofkVp,mA,timerselectionsandexposurewarninglight.3.beaminimumof6feetawayfromthepatient.4.beoutoflineoftheusefulbeamofradiationorbepositionedbehindaprotectivedevicewithX-ray protectionequivalentof1mmoflead.

FIGURE 1

4-1/

"

1-1

/16"

(10

4)

"(1

904)

1

"(

0)

1

"(4

0)

(66-1/16")0-/"

4-1/"

(46-/16")

220˚

( ) : WITH LONG CONE

From2004July

Page 8: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

SECTIONTHREE:INSTALLATIONINSTRUCTIONSWithin the installation and confirmation procedures are inspection/test steps which the installer must perform to insure that the installation meets the manufacturer’s specifications. These steps require the installer to record thenecessaryinformationontothe“ASSEMBLER’SINSTALLATIONSECTIONOFTHELIMITEDWARRANTYREPORTFORM”supplied,whichmustbereturnedtoBELMONTalongwiththewarrantycard.

[1] INSTALLATION REQUIREMENTSTools:Standardtoolkitincludingwirecrimpingpliers(AMP,“SuperChamp”orequivalent).1.5mm,2mm,3mmand5mmallenkeys.Instruments:Digitalmultimeterwithanaccuracyof1%,capableofmeasuring150VACand10mADC,andcapableofindicatingtrueRMSvaluewithin1second.Impulsecounter,capableofregistering1to999half-waveprimaryvoltageimpulsesatarateof60persecondoverarangeof90to100Vrms,60Hz.Standardcalculator.

POWER SUPPLY:Prior to starting the installation inspect the power supply and confirm that the power supply is 120V AC,+12VAC,-12VAC,andthatthesupplyisa3wireGROUNDEDcircuit,separatelyconnectedtothecentraldistributionpanelwithanovercurrentprotectiondeviceratedfor15amperes(RefertoPage6,[2]ELECTRICALREQUIREMENTS).Recordthevoltagereadingofpowersupplyon“Assembler’sInstallationReport”.

[2] UNPACKINGUnpacktheentirecontentsoftheshippingcarton.Includedwithintheshippingcartonare:

Identification QuantityHead............................................................................................................. 1/WK,RKRegularCone............................................................................................... 1/WK,RKLongCone................................................................................................... (1)/WK(OPTION)ControlBox................................................................................................. 1/WK,RKHeadkey...................................................................................................... 1/WK,RKCollar........................................................................................................... 1/WK,RKBalanceArm................................................................................................ 1/WKBalanceArmWrench................................................................................... 1/WKHorizontalArmW/2/ScrewCover............................................................. 1/WKArmMountingBracketW/3MachineBolt(M8x20mm),3/Boltcap....... 1/WKWallPlateW/4CoachBolts,3bolts,Washers,7/boltcapandtemplate...... (1)/WKBrakeScrew(M6x6mm)........................................................................... 2/WK,RKBrakePlug(BrassPlug).............................................................................. 2/WK,RKRetainingBolt(M6x35mm)...................................................................... 2/WKStopperScrew(M6x15mm)...................................................................... 1/WK,RKControlBoxMountingScrew(ø5.8x32mm)............................................. 4/WK,RKBrakeSpring(ø5)........................................................................................ 1/RKBase............................................................................................................. 1/RKColumn........................................................................................................ 1/RKSlidingPost.................................................................................................. 1/RKSwingArm................................................................................................... 1/RKGasPump..................................................................................................... 1/RKSeat.............................................................................................................. 1/RKLagBolt(ø8x45mm)................................................................................. 5/RKManual......................................................................................................... 1/WK,RK

Inspectcontentsofshippingcartonfordamageormissingcomponents.

Page 9: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 9 -

[3] ARM ASSEMBLY INSTALLATION[3-1] WK TYPE

TheINSTRUCTIONSgivenbelowareformountingthearmandheadassemblyontwo2X4woodstudswith 16-inch centers. Should the MODEL-096 be mounted in a manner other than what is specified here,the wall and the strength of the hardware used must be checked and verified as being adequate to withstanda100poundshearloadanda450-poundwithdrawalforceateachofthefourmountingbolts.When using concealed wiring, a flush mounted junction box with the necessary conduit and wiring must bepre-installed at 40 5/8 inches from the floor and centered between the two studs (refer to template).

A. WALL PLATE (FIGURE 2):

1.Tapethewallmountplatetemplatetothewall,positioningit so that the holes are aligned with the vertical 2 X 4" stud.

NOTE:Innoinstanceisthewallmountplate,orthearmmountingbracket,tobeasinglestud.

2.Marktheholelocations.3. Using a 3/16" drill, drill a pilot hole approximately 2" deep for eachplatebolt.CAUTION:Donotuselargerdia.drill. Theelectricalwireaccessholeshouldbepredrilledusingthe height from the floor specified on the mounting template.4.Placethewirethroughtheholeintothestudmountwallplate andmounttheplatetothestudwith4coachboltssupplied. DO NOT FULLY TIGHTEN BOLTS.5. Holding plate firmly against the wall, place a level(a)upright onthesurface.Ifthewallisnotplumb,useshimsbehind theplatetolevel.6.Placealevel(b)acrosstopedgeofwallplate,levelandtightenboltssecurely. NOTE:Finallevelingofwallplateisdescribedonpage13.7.Putboltcaptoeachheadofbolt.

B. ARM MOUNTING BRACKET (FIGURE 3):

1.Removeaccesscoverfrombottomofarmmountingbracket.Snakeelectricalinterconnectingwires throughbracketandoutaccesshole.2.UsingM8X20mmboltwithwashersinlowermountingholeandtwoM8X20mmboltsintop mountingholes,mountarmmountingbrackettowallplate.DONOTFULLYTIGHTEN.3.Placinglevelacrosstopedgeofarmmountingbracket,levelbracketthentightenboltssecurely.4.Puttheboltcaptoeachheadofbolt. NOTE:FinallevelingofwallmountingbracketisdescribedonPage13.

FIGURE 3

Access Hole

Coach Bolt

Level(b)

Level(a)

Bolt Cap

FIGURE 2

Access HoleAccess Cover

M x 0Bolt

Level(b)

WasherBolt Cap

C. If wall plate is not used, other optional, special, wall plate for "pas thru" (4X4") installation must be installed.

Page 10: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 10 -

D. HORIZONTAL ARM (FIGURE 4):

1.Cutpullstringonhorizontalarm.DONOTREMOVESTRING. ALLOWONEENDTOEXTENDBEYONDMALEBARB ANDTHEOTHERENDTOEXTENDBEYONDTHE FEMALEMOUNT.2.Placeathrustwasherovertheholeofarmmountingbracket, andinsertmalebarbintoarmmountingbracket,allowingpull stringtoextendthroughaccessopeningonbottomofthearm mountingbracket.3.Inserttworetainingboltssecurelyintoupperthreadedholesofarm mountingbracketandtightensecurely.

IMPORTANT:Theretainingboltsmustsecurelyengagetheannulargroovehorizontalarm,theremovaloftheretainingboltswillallowthehorizontalarmtorisevertically,andoutof,thearmmountingbracket. DO NOT FULLY TIGHTEN.4.Insertbrakeplugthenbrakescrew(M6x6mm)intothelowerthreadedholeofthearmmountingbracket.5. Place a level on the horizontal arm and confirm that the arm is level in its left and right swing positions. NOTE:FinallevelingofhorizontalarmisdescribedonPage13.

E. BALANCE ARM ASSEMBLY (FIGURE 4a):

WARNING:DONOTRELEASEARMHOLDINGBANDUNTILTHEX-RAYHEADHASBEENINSTALLED.BALANCEARMASSEMBLYISSPRINGLOADEDANDCANCAUSEEQUIPMENTDAMAGEANDINJURYIFNOTHANDLEDINTHEPROPERMANNER.

1.DONOTREMOVEARMHOLDINGBAND.2.Securepullstringtocableandpullingtheopposite end,snakecablethroughhorizontalarmandarm mountingbracket.3.Insertbrakeplugthenbrakescrew(M6X6mm) intothehorizontalarmcollar. DONOTFULLYTIGHTEN.4.Removeendcapscrewandopenendcap.5.Insertstopperscrewintoupperthreadedholeinside horizontalarmandtightensecurely. CAUTION:Ifstopperscrewisnottightened securely,thescissorsarmcanmoveverticallyup andoutofthehorizontalarm6.Cutcableandinterconnectingwirestoaworkable length. Strip 3/8" of wire insulation from each lead. Withwirecrimpingpliersusesuppliedwirenuts tomakewireconnections.7.Insertconnectedwiresintothearmmounting bracketandsecuretheaccesscovertothebottom ofthearmmountingbracket.8.Secureendcapwithendcapscrew,andplacea screwcover.

Brake Plug

Brake Screw

Thrust WasherRetaining Bolts

Pull String

FIGURE 4

FIGURE 4a

Arm Holding Band

Brake Plug

Brake ScrewEnd Cap

Screw Cover

Stopper ScrewAccess Cover

Access Cover Screw

End Cap Screw

From2004July

Page 11: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 11 -

[3-2] RK TYPE (FIGURE5)

1. Fix the base on the floor with lag bolts (supplied) or withappropriatemeans.

CAUTION :MAKE SURE THE BASE IS

FIXED ON THE FLOOR FIRMLY.

2. Insert the slidingpostwith columncover into thecolumn.

3.Installthecolumnonthebasewithmountingbolts.Make it vertical with adjusting bolts.

4.Putthethrustwashertotopoftheslidingpostandinstalltheswingarmassemblytoslidingpost.

5.Set the stopper screw into lower threaded hole ofswingarm2.

6.Set thebrake plug then brake spring and brakescrewintotheupperholeofswingarm2.

TightenthebrakescrewIFARMDRIFTS. DO NOT FULLY TIGHTEN.

7.Run the cable from swing arm1 through a cableguide.

8.Slideupthebackrestcushiontothetopofcolumn.

9. Insert the gas pump into the gas pump bracket.Mounttheseatonthegaspump.

10.Refertopage12forHeadassemblyinstallation.

11.Refertopage13forControlboxinstallation.

12.Performthepostinstallationinspection.(page14)

Brake Screw

Stopper Screw

Thrust Washer

Brake SpringBrake Plug

Swing Arm

Swing Arm 1

Sliding Post

Column Cover

Column

Cable Guide

Cable

Mounting Bolt

Adjusting Bolt

Lag Bolt

Base

Gas Pump

Gas Pump Bracket

SeatBackrest Cushion

FIGURE 5

Page 12: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 1 -

1mm

Head Key

Arm Collar

Balance Arm

Balance ArmWiring

Head ShaftStopper Ring

Head Yoke

Terminal

mm

Yoke Inside Cover

Arm Holding Band

(a)

(b)

3 Blue4 Brown5 Red6 Yellow7 Gray8 Green/Yellow

[4] HEAD ASSEMBLY INSTALLATION (FIGURE 6):

FIGURE 6

1.Removeascrew(a)onarmcollar,andplacethearmcollaroverthestopperringontheheadshaft.

(DirectionofarmcollarisshowninFIGURE6.)

2.Removetheyokeinsidecoverbylooseningscrew(b).

3.Makingsurethestopperringandarmcollarisplacedontheheadshaft, insert thewiringsofbalancearmthroughtheheadshafttotheheadyoke.

4.Inserttheheadshaftintobalancearmassembly,andwhileholding thehead inposition, installheadkeysecurelyintoretaininggroove.

5.Slide the armcollar upward to proper position andsecureitinplacewithascrew(a).

6.Loosen5 screwson the terminal using a small dia.screw driver, and insert thewirings of balance armassembly into the terminal according to the color.(FIGURE6-1)

7. Retighten the screws on the terminal and confirm that the wirings are fixed on the terminal.

8.RemoveM5PhillipsheadscrewfromearthterminalandsecureNo.8wiringtogethertoearthterminal.

9.Reattachtheyokeinsidecoverwithascrew.

10.Removearmholdingband.

4

6

Small Diameter Screw Driver

Earth Terminal

Wirings of Balance

Arm Assembly

FIGURE 6-1

[5] CONTROL BOX INSTALLATION

The wall and the strength of the hardware used must be checked and verified as being adequate towithstanda25poundshearload.

A flush mounted junction box with the necessary conduit and wiring must be pre-installed at 51 5/8" fromthe floor.

NOTE:Refertothecontrolboxtemplate,and: 1. Be certain the electrical wire entry hole is aligned with the junction box. 2.Thatthereisadequatesupportinthewalltosecurethecontrolbox.

From2004July

Page 13: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 1 -

A. CONTROL BOX - MOUNTING (FIGURE 7):

1.Tapethecontrolboxtemplatetothewallattherecommendedheight.1a.Confirm relationship of access hole in for the electric wires with the entry hole for the wires in the back ofthecontrolbox.2. Using a 3/32" drill, drill a pilot hole 2" deep for each mounting screw.2a.Themethodofdrillingthepilotholeand,thehardwareusetosecurethecontrolboxdependuponthestructure.3. Drive two wood screws into upper two holes remaining about 3/4" undriven. (Fig.7a)4.RemovetwoM3Phillipsheadscrewsfromtopofthecontrolboxandopenfrontpanel.(Fig.7b)5.Removearestrictionplate.(Fig.7c)6.Snakepowersupplycablelinesandinterconnectingwiresthroughaccessholeinbackpanel.7.Hookthecontrolboxchassistowoodscrewsdriveninstep3abovethroughtwomountingholes onuppersideofchassis.Tightenscrewsslightly.8.Attachtwowoodscrewstomountingholesonlowersideofbackpanel.(Donotfullytighten.)(Fig.7d)9.Placinglevelacrosstopedgeofthecontrolbox,level,thentightenfourscrewessecurely.

a

b

c

d

FIGURE 7

B. CONTROL BOX, WIRING (FIGURE 8-a & 8-b)

WARNING:MAKESURETHEPOWERSUPPLYISTURNEDOFFATTHECENTRAL DISTRIBUTIONPANEL.1. Strip approximately 3/8" of insulation off the power supply leads and interconnecting cables.2.Followingwiringdiagram,connectthosewirestotheterminalblockofcontrolbox.3.Settherestrictionplatetotheoriginalplace.

Junction Boxes

FromPower

Source

Terminal Blockfor Control Box

Pigtail for ArmMounting Bracket

WALL

L N 4 6 6 4

GR

EEN

/YEL

LOW

BLU

E(W

HIT

E)

BR

OW

N(B

LAC

K)

BLUE

RED

GRAYGREEN/YELLOW

YELLOW

BROWN

FIGURE 8-a46

LN

FIGURE 8-b

Page 14: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 14 -

Bolt CapTop Mounting Bolt

Bottom Mounting Bolt

Brake Screw

LEVEL

#

#1

FIGURE 10-b

LEVEL

Wall error adjustment;installshimsbehindthewallplate

FIGURE 10-a

C. CLOSING FRONT PANEL (FIGURE 9).

CAUTION: BEFORECLOSINGTHE FRONTPANEL, PERFORMPOST INSTALLATIONCON-FIRMATION(PAGE20).

1. Confirm all the post installation confirmation are performed.2.ClosethefrontpanelandsecuretwoM3phillipsheadscrewsonthetoppanel.

FIGURE 9

SECTIONFOUR:POSTINSTALLATIONINSPECTION

[1] ARM ASSEMBLY1.Incorrectlevellingofthewallbracketcancausearmdrift.First,checklevelwitharmsinposition#l.If not correct, bracket must be adjusted by placing shims behind the wall plate (FIGURE 10-a).IMPORTANT: Iftheendofthehorizontalarm#1ispitchedbelowlevel,thenthetubeheadwilldriftawayfromthewall.Iftheendofthehorizontalarm#1ispitchedabovelevel,thenthelevelarmwillrequireonlyminimumadjustment of brake (friction) screw.2. Check level in position #2. If not correct, adjust as follows: (FIGURE 10-b) a)Removeboltcapsonmountingbolts. b)SLIGHTLYloosentwotopmountingbolts. c)Shiftthebracketleftorrightuptothearmsarcaccuratelylevelled. d)Movethehorizontalarmtoposition#1. e)Fullytightentwotopmountingbolts. f)Fullytightenbottommountingbolt. g)Puttheboltcaptoeachheadofmountingbolt.

NOTE:SLIGHTTENDENCIESTODRIFTCANBECORRECTEDBYTIGHTENING BRAKESCREWSINHORIZONTALARMAND/ORWALLBRACKET. DONOTTIGHTENBEYONDWHATISREQUIREDTOPREVENTDRIFT.

Page 15: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 1 -

[2] BALANCE ARM ASSEMBLY

1.Placethebalancearmassemblyintoposition.2.If either balance arm drifts either higher or lower from the set position, remove the spring adjuster covers and with the supplied wrench adjust the balance arm springs (FIGURE 11).

Adjustment Nut

less tensionLOOSEN

more tension TIGHTEN

Balance Spring

Cover

FIGURE 11

[3] HEAD POSITIONING

A.Placeheadintoposition.

B. If head drifts from the set position, adjust the brake screws according to the following procedures.(FIGURE12)

1.Removetheyokeoutsidecoverbylooseningcoverscrew. 2. Adjust 6 brake screws using phillips screw driver. 3. After adjustment, reattach the yoke outside cover with the cover screw.

Phillips Screw Driver

Yoke Outside Cover

Cover ScrewBrake Screw

FIGURE 12

Page 16: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 16 -

SECTIONFIVE:CONTROLIDENTIFICATIONANDOPERATION[1] CONTROL IDENTIFICATION

W A R N I N G THIS X-RAY UNIT MAY BE DANGEROUSTO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORSAND OPERATING INSTRUCTIONS ARE OBSERVED.

MODEL 096 0kVp 10mA

P C F

T1 T2 T3 T4 T5

a b c

1514

2

3

4

98

76

121110

16

5

1

13 (1) Main Power switch(2) Ready Lamp(3) Exposure Time Adjusting Sw. (Up)(4) Exposure Time Adjusting Sw. (Down)(5) Tooth Selection Switch (T1)(6) Tooth Selection Switch (T2)(7) Tooth Selection Switch (T3)(8) Tooth Selection Switch (T4)(9) Tooth Selection Switch (T5)(10) Patient Size Selection Switch(11) Cone Type Selection Switch(12) Film Speed Selection Switch(13) Exposure Time Display Window(14) Exposure Warning Light(15) Exposure Switch(16) Technical Switch

[2] FUNCTION OF CONTROLS

Main Power switch Pushingrightsideofthisswitchenergizesthex-rayunit. (Ready lamp and pre-selected lamps for patient size, cone type and film speed illuminates.) ItisrecommendedtokeepthisswitchOFFwhentheunitisnotinuseinordertopreventanacci-

dentalexposure.

Ready Lamp Thislamplightswhenthelinevoltageiswithinoperablerange. Whenthislampisnoton,exposurecannotbemade.

Exposure Time Adjusting Switches Bymomentarilypushing (or )switch,exposuretimedisplayedincreases(ordecreases)byone

step.Bykeepingtheswitchdepressedmorethan2sec.,exposuretimedisplayedincreases(orde-creases)continuouslyuntiltheswitchisreleased.

Tooth Selection Switch (T1 ~ T5) Pushingoneoftheseswitchessettheexposuretimeautomaticallyincombinationwithfollowing ~ . T1:IncisorofMandible T2:IncisorofMaxilla,Cuspid&PremolarofMandible T3:Cuspid&PremolarofMaxilla,MolarsofMandible,Bitewing T4:MolarsofMaxilla,BitewingMolars T5:Occlusal

1

3

2

4

5 9

10 1256789

~

Page 17: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 1 -

Patient Size Selection Switch Pushingthisswitchalterstheselectionofpatientsize(small->medium->large->small)andsets

theexposuretimeaccordingly.

Cone Type Selection Switch Theexposuretimecorrespondingtotheconetypebeingused(StandardRegularConeorOptional

LongCone)canbeselectedbythisswitch.

Film Speed Selection Switch Three types of film speed can be registered. Pushing this switch momentarily indicates the film

speednumberbeingselectedinexposuretimedisplaywindow . Depressing the switch for more than 2 seconds alters the film type being selected.

NOTE : Setting or adjusting the exposure time manually (with or switch) supersedes ~ functions.

Exposure Time Display Window Normallytheexposuretimeselectedisdisplayed.

1.E.00~E.12 :Errorcode[Seepage19ofthismanual] 2.F.00~F.15 :Filmtype[Seepage21&22ofthismanual] 3.TubeCurrent:Thetubecurrentofthelastpulseoftheexposurecanbedisplayediftheexposure switchiskeptdepressedafterexposureisoverandtechnicalswitch isdepressed. 4.bu.0,bu.1 :BuzzerON/OFFwhenaswitchisactivated.[Seepage23ofthismanual] 5. FIn :Confirmation of tube current [See page 20 of this manual] 6. PH.0 ~ PH.F :To adjust tube current at beginning of exposure [Refer to service manual] 7. EP.0 ~ EP.F :To adjust tube current when stabilized [Refer to service manual] 8. Pt.0 ~ Pt.F :To adjust preheat time [Refer to service manual]

Exposure Warning Light Illuminationofthislightindicatestheunitisproducingx-radiation.

Exposure Switch DeadmanTypeexposureswitch.Whenmakinganexposure,depress thisswitchandkeepitde-

presseduntiltheexposurewarninglight andtheaudiblewarningterminate. Failuretokeepthisswitchdepressedwillresultinprematureterminationoftheexposure.

Technical Switch Thisswitchisexclusivelyfortheinstallerandservicepersonnel.Itisusedforfollowingpurposes: 1.Thetubecurrent(mA)ofthelastpulseoftheexposurecanbedisplayedinexposuretimedisplay

window iftheexposureswitchiskeptdepressedevenaftertheexposureisoverandthistechni-calswitchisdepressed.Displaywillreturntoexposuretimewhentheexposureswitchisreleased.

2.Priorityofselection(PatientSize,ConeType,FilmSpeed)canbechangedwhenthemainpowerswitchisturnedonwhilethisswitchisdepressed.[SeesectionSevenofthismanual]

3. Memorizing : After setting film speed, priority of selection and/or buzzer ON/OFF, this switch isusedformemorizethesesettings.[SeesectionSevenofthismanual.]

10

11

12

13

14

15

16

5 12

13

16

14

13

Page 18: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 1 -

[3] OPERATING PROCEDURES

1. TurnONthemainpowerswitch .

2. Confirm that ready lamp isilluminated. NOTE: The ready lamp will not illuminate unless the incoming line voltage is correct and within the x-ray’s operable range.

3. Selecttheappropriatetoothtype( ~ ), and confirm if the pre-selected conditions(patientsize ,conetype and film speed )aresuitable forradiographing.

NOTE: To manually set the exposure time, depress either manual exposure time adjust switch( or ) until the desired exposure time is displayed in exposure time display window . While the unit is in manual mode, other selection switches( ~ ) do not affect exposure time. (All the tooth selection lamps are off.)

Toreturntotheautomaticexposuretimeselectionmode,depressany oneoftoothselectionswitches.

4. Depresstheexposureswitch .Whentheexposureswitchisdepressed, theexposurewarninglamp illuminatesandtheaudiblewarningsounds. Donotreleasetheexposureswitchuntiltheaudiblewarningandthe warninglampterminate. Failuretokeeptheswitchdepressedwillresult intheexposurebeingterminatedprematurely.

5. To continue to radiograph other teeth, just select appropriate tooth selectionswitch.

IMPORTANT : To protect x-ray tubehead from heat accumulation, wait for 60 times of exposure time between exposures. [Ex. 30 second wait interval for each 0.5 sec.(30 impulses) exposure]

6.AfteruseturnOFFthemainpowerswitch inordertopreventaccidental exposures.

NOTE : If the unit is left over 8 minutes without being operated and the main power switch is kept on, figure 1 runs through the exposure time display window. This does not mean that a malfunction of the unit has occurred, but saves energy. The unit returns to normal condition by pressing any one of the switches except the exposure switch.

1

2

5 9

10 1211

3 4

14

15

13

5 12

1

Page 19: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 19 -

[4] ERROR CODES

Whenabnormalconditionexistsintheunit,ormalfunctionoccurs,errorcodeisdisplayedinexposuretimedisplaywindow.Pleaserefertothetablebelow.

ErrorcodeE.00E.01E.02E.03E.04E.05E.06E.07E.08E.09E.10E.11E.12

ConditionExposureswitchwasreleasedbeforethe exposureterminates.Exposureswitchwasdepressedwithin10sec.ofpreviousexposure.Exposureswitchwasdepressedwithin3sec.afterthemainpower switchhasbeenturnedon.Linevoltagewaslessthan 90%ofratedvoltage.Linevoltagewasmorethan 110%ofratedvoltage.Excesscurrentduringexposure.Tubecurrentofthelastpulsewaslessthan7.5mA.Tubecurrentofthelastpulsewasmorethan12.5mA.Tubecurrentduringex-posurewaslessthan5mA.Tubecurrentduringex-posurewasmorethan15mA.Malfunctionofthemicrocomputer. Exposureswitchorex-posurecircuithadbeenON,whenmainpowerswitchisturnedon.Tubecurrentisdetectedduringpre-heatingperiod.Tubecurrentisdetectedwhenmainpowerswitchisturnedon.

Steptobetaken Allthetoothselection switchesblink. Depressoneoftheswitch. A10seconddelayisbuilt inbetweeneachexposure. Releaseexposureswitch. Turnoffthemainpowerswitchandwaitfora while.Turnonthemainpowerswitchagain.

PossiblesolutionReleaseexposureswitch afterexposurelamp turnsoff.There is to be an "wait"intervalof60times ofexp.timebetween successiveexposures.Exposureswitchshould bedepressedaftertheready lampcomesON. Iflinevoltageisless than90%ofratedline voltage,correctitby usingstep-uptransformer.Iflinevoltageismore than110%ofratedline voltage,correctitby usingstep-downtransformer.Conducttheconfirmation oftubecurrentdescribedinsectionsix,item[2]on page20.Refertotheservicemanual.

Page 20: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 0 -

SECTIONSIX:POSTINSTALLATIONCONFIRMATION

[1] CONFIRMATION OF POWER SUPPLY VOLTAGEAs specified in Electrical Requirements (Page 6), power supply voltage must be within the operablerange of 108~132V AC. Confirm the power supply voltage again before turning on the unit.1.Openthefrontpanelofcontrolboxbylooseningtwoscrewsontopofthecontrolbox.2.Settherangeofdigitalmultimeterat200VAC,connectprobesofmultimetertoLandNof terminalblockofcontrolbox.3. Confirm that the reading is 120V ±10%(108~132VAC). NOTE :096X-raycannotbeoperatedunlessthepowersupplyvoltageiswithinthisrange.

[2] CONFIRMATION OF TUBE CURRENTModel 096 x-ray incorporates self diagnose system to check if the tube current is within specifiedrangebothatthebeginningofexposureandduringstabilizedperiod.1.WhiledepressingtoothselectionswitchesT1,T4&T5together,turnonthemainpowerswitch.2.Exposuretimeof0.50isdisplayedandreadylightison. [Ifnot,turnoffmainpowerswitchandrepeat1.] ThenreleaseT1,T4&T5switches.3.Makeanexposurebydepressinghandexposureswitch. WARNING:X-RADIATIONISGENERATEDFOR0.5SEC.4. Confirm that "F I n" is displayed at exposure time display window. Then,turnoffthemainpowerswitch.5. If "PH. " and "EP. " are displayed alternately, follow Step 6 through 8 below. 6. When "PH. " and "EP. " are displayed alternately, leave the unit for about 30 seconds until display returns to "0.50". Then make an exposure again and confirm that "F I n" is displayed. 7. Repeat Step 5 & 6 until "F I n" is displayed.8.Record final PH & EP values in "Assemblers Installation Section of Limited Warranty Report Form" if  proceduresinStep5through7areperformed.

[3] CONFIRMATION OF EXPOSURE WARNING LAMP & BUZZERA.EXPOSUREWARNINGBUZZER 1. Make an exposure and confirm that the exposure warning buzzer located within the control box isactivatedduringtheentireexposure.B.EXPOSUREWARNINGLAMP Exposurewarninglampislocatedonthefrontpanelofthecontrolbox, 1. Make an exposure and confirm that the warning lamp illuminates during the exposure.

[4] CONFIRMATION OF LINE VOLTAGE REGULATION1. Make sure that main power switch is "OFF". 2.Openthefrontpanelofcontrolboxbylooseningtwoscrewsontopofthecontrolbox.3.Settherangeofdigitalmultimeterat200VAC,connectprobesofmultimetertoLandNof terminalblockofcontrolbox.4.Turnthemainpowerswitchon,andsettheexposuretimeat2.00sec.withmanualswitch .5.Recordtheno-loadlinevoltage(VN)indicatedbythemultimeterbeforeexposure.6.Makeanexposureandrecordtheloadvoltage(VL)indicatedbythemultimeterduringexposure. WARNING :X-RADIATIONISGENERATEDFOR2SECONDS. NOTE:Readthemultimeterwhenthevalueisstabilized(aboutonesecondafterexposure).7.CalculatelinevoltageregulationR(%)intheformulabelow:

VN-VL R = X 100 Record this value in "Assemblers Installation Report". VL

NOTE: LINEVOLTAGEREGULATIONMUSTNOTEXCEEDTHERANGEOF2~5%. IFITISGREATERTHAN5%,THESIZEOFTHEPOWERSUPPLYWIRESMUSTBE INCREASED.REFERTOTHEPOWERSUPPLYREQUIREMENTSOUTLINEDON PAGE5TODETERMINETHECORRECTWIRESIZENECESSARY.

Page 21: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 1 -

SECTIONSEVEN:INITIALSETTING[1] FILM SPEED

As factory installation, following three kinds of film speed are registered to be selected byFilmSpeedSelectionSwitch:

a = Film speed No. F.09 (equivalent to ISO speed group "D", or Kodak Ultra-Speed film) b = Film speed No. F.04 (equivalent to ISO speed group "F/E", or Kodak InSight film)c = Film speed No. F.02 (equivalent to ISO speed group "F")

Including these three, Model 096 can provide 16 different types of film speed and any three of themcanberegisteredforeasyselection.If the doctor uses different speed of film, or prefers darker (or lighter) radiograph, substitute speedcanberegisteredasfollows:

1.Whiledepressingtechnicalswitch ,turnonthemainpowerswitch . Film type lamp "a" is lit, andF.09isdisplayedinexposuretimedisplaywindow . Then,releasetechnicalswitch.

2.Bydepressing switch(or switch),increase (decrease) film speed number until desired numberisdisplayed.[RefertoExposureTimeTableonnextpage.]

3.Depresstechnicalswitch , an electronic chime sounds and the selected film speed number is registered at film type "a".

4.Turnoffthemainpowerswitch.

5. If different film speeds are to be registered at "b" and "c", depress "F"switch afterstep1above to light the appropriate film type lamp, and repeat steps 2 & 3.

12

13

16 1

16

Page 22: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

TABLE 1 : FILM SPEED AND EXPOSURE TIME (REGULAR CONE)

TABLE 2 : FILM SPEED AND EXPOSURE TIME (LONG CONE)

T10.00.00.00.00.040.00.060.00.00.100.10.10.10.00.40.9

T20.00.040.00.060.00.00.100.10.140.10.190.40.90.0.410.0

T30.040.00.060.060.00.100.10.140.10.00.40.90.0.410.00.

Patient SizeToothF. 00F. 01F. 02F. 03F. 04F. 05F. 06F. 07F. 08F. 09F. 10F. 11F. 12F. 13F. 14F. 15

SMALL MEDIUM LARGE(UNIT : SEC.)

(UNIT : SEC.)Patient Size

ToothF. 00F. 01F. 02F. 03F. 04F. 05F. 06F. 07F. 08F. 09F. 10F. 11F. 12F. 13F. 14F. 15

SMALL MEDIUM LARGE

T40.00.060.00.090.100.10.160.10.0.0.10.0.40.40.660.6

T50.00.090.110.10.160.190.0.0.0.0.40.40.660.10.91.1

T10.00.040.00.00.060.00.090.110.140.160.190.40.0.0.0.4

T20.00.060.00.090.110.10.160.190.0.0.10.0.40.40.660.1

T30.060.00.090.100.10.160.190.0.0.0.0.40.40.660.10.9

T40.00.100.10.10.10.00.0.90.60.440.00.60.60.1.01.

T50.10.10.10.00.0.10.60.440.40.60.60.1.01.1.11.6

T10.040.00.060.060.00.100.110.140.10.190.40.90.0.410.40.

T20.060.00.100.110.140.160.190.40.0.0.0.40.0.660.11.00

T30.00.090.110.10.160.190.40.0.0.410.40.0.660.11.001.1

T40.100.10.10.10.00.0.10.60.440.40.60.60.91.01.1.1

T50.10.10.0.0.10.0.440.40.660.60.91.01.1.61.6.

T10.00.00.060.00.00.100.10.10.10.00.0.10.60.440.00.6

T20.00.00.100.10.10.10.00.0.90.60.410.00.60.10.1.0

T30.00.100.10.140.10.00.0.90.60.440.00.60.10.1.01.

T40.110.140.160.190.0.0.0.0.40.0.660.11.001.11.411.6

T50.160.190.40.0.0.410.40.0.10.11.001.11.411..00.44

T10.00.00.100.110.140.10.190.40.90.0.410.00.0.10.11.00

T20.110.140.10.190.40.0.0.410.40.0.660.11.001.11.411.

T30.140.160.190.0.0.0.410.40.0.10.11.001.11.411..00

T40.10.0.0.10.60.440.40.60.60.91.01.1.61.6..6

T50.0.10.0.440.40.660.60.91.11.1.61.6..0.00.00

T10.00.100.10.140.10.00.40.90.60.410.00.60.10.1.001.

T20.140.10.00.40.90.0.410.00.0.10.11.001.1.411..1

T30.10.190.40.0.0.410.00.0.10.1.001.1.411..1.44

T40.0.0.10.0.440.40.660.60.91.11.1.6.00..0.00

T50.10.0.40.40.660.10.91.11.411.6.00..0.00.00.00

Page 23: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

L

N

Control Box Head

X-RayTube[D-0]

6

4

6

4AVR 1

1

1

1

H.V. Transformer

Fil. Trans.

PowerSwitch

Exposure Switch

Fuse

Blue

Yellow Yellow

Gray Gray

Blue Blue

BrownBrown

Red

Green/Yellow Green/Yellow

1 CN1 16

1 CN

CN4

CN

CN

N.C.

T.B.T.B.

T.B. T.B.

16

500Ω

Brown

NC

CN1

Fuse

Control Board096P-CN00

Power Generator Board096P-PW00

RY

Green/Yellow

Arm

APPENDIXONE:CIRCUITDIAGRAM

[2] PRIORITY OF SELECTION

Asfactoryinstallation,followingselectionlampslightwhenthemainpowerswitchisturnedon:

PatientSize :MediumCone :RegularFilm Speed : "a"

Ifnecessary,thisprioritycanbechangedasfollows;[For example, at pedodontistry, patient size of "small" should be preferentially selected.]

1.Whiledepressingtechnicalswitch ,turnonthemainpowerswitch.2. Select the patient size "small" by depressing patient size selection "P"switch .3.Depresstechnicalswitch , an electronic chime sounds and the patient size of "small" will be stored as primaryselection.4. Priority of selection for cone type and film speed can be changed by same procedures.

[3] ELECTRONIC CHIME ON/OFF

Asfactoryinstallation,electronicchimesoundswheneachswitchisdepressed.Ifpreferred,thissoundcanbeeliminated.

1.WhiledepressingtoothselectionswitchesT1&T2together,turnonthemainpowerswitch.2. "bu.1" will be displayed in exposure time display window .3.Bydepressingeither or ( or ), display changes to "bu.0".4.Thendepresstechnicalswitch ,andturnoffthemainswitch.

NOTE :ExposureWarningBuzzerandalarmsoundoferrorcodecannotbeeliminated.

16

1610

163 4

13

Page 24: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 4 -

APPENDIXTWO:PARTSIDENTIFICATION[1] ARM AND HEAD ASSEMBLY OF WK TYPE

[2] CONTROL BOX ASSEMBLY (60)

347

48

4

67

5

2

1

49

279

1146

12

1413

15

26

23

8

221718

14

1317

2425

1420

2119

10

16

14

2932

3031

33

28

3435

3738

393031

4142

4443

40

36

45

50

From2004July

Page 25: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -From2004July

ID.No. PartsNo. Description QTY USA Code

1 E04-EHLL0A0 X-Ray Head Assembly (RAL-900) 1 096-1010

E04-ECPE0G0 Yoke (RAL-900) 1 096-100

E04-EHLL1A0 Housing Cover Set (RAL-900) 1 096-100

4 E04-ECPRB0 Lock Ring (R1-100) 1 096-1040

E04-ECPR6A0 X-Ray Exposure Sleeve 1 096-100

6 E04-ECPR1D0 Regular Cone (RAL-900) 1 096-1060

E04-EHLL1A0 Long Cone (RAL-900) (Option) (1) 096-100

E04-EHLLA0 Balance Arm Assembly (RAL-900) (1) 096EHLLA0

9 E04-ECLS01A0 Stopper Ring 1 096-1090

10 E04-ECPE1B0 Joint No. 1 096ECPE1B0

11 E04--------------- Collar Screw (SUS M4-10) 1 096SRWM0410

1 E04-ECQR4B0 Collar (R1-100) 1 096ECQR4B0

1 E04-ECQR0C0 Spring Adjuster Cover 096ECQR0C0

14 E04-ECQRB0 Crevice Cover 4 096ECQRB0

1 E04-ECPE19F0 Left Cover for Joint No. 1 096ECPE19F0

16 E04-ECPE1F0 Right Cover for Joint No. 1 096ECPE1F0

1 E04-ECPJ64C0 Cover for Joint No. 096ECPJ64C0

1 E04-ECPJB0 Joint No. 1 096ECPJB0

19 E04-ECPJ6F0 Left Cover for Joint No.1 1 096ECPJ6F0

0 E04-ECPJ6F0 Right Cover for Joint No.1 1 096ECPJ6F0

1 E04-ECPE1B0 Joint No.1 1 096ECPE1B0

E04-ECPJ60A0 Arm Cover No. 1 096ECPJ60A0

E04-ECPE0B0 Balance Arm No. 1 096ECPE0B0

4 E04-ECPJ9A0 Arm Cover No.1 1 096ECPJ9A0

E04-ECPE1B0 Balance Arm No.1 1 096ECPE1B0

6 E04-ECQRA0 Cushion Absorber 1 096ECQRA0

E04-ECLR9B0 Head Key 1 096-10

E04-EHLK90A0 Horizontal Arm Ass'y (00mm) (RAL-900) (1) 096-10

E04-EHLKA0 Horizontal Arm Ass'y (00mm) (RAL-900) (1) 096-11

E04-EHLKA0 Horizontal Arm Ass'y (00mm) (RAL-900) (1) 096-1

E04-EHLK9A0 Horizontal Arm Ass'y (60mm) (RAL-900) (1) 096-1

E04-EHLK91A0 Horizontal Arm Ass'y (1000mm) (RAL-900) (1) 096-14

9 E04--------------- Horizontal Arm 1 096-190

0 E04-ECLS06A0 Brake Plug 096-100

1 E04--------------- Brake Screw (M6-6) 096-110

E04-ECLS09A0 Stopper Screw 1 096-10

E04-ECLJ6A0 End Cap (R1-100) 096-10

4 E04--------------- End Cap Screw (M6-1) 096-140

E04-ECNR4A0 Hole Plug for End Cap (R1-100) 096-141

6 E04-EHLL0A0 Arm Mounting Bracket Ass'y (RAL-900) (1) 096-10

E04--------------- Machine Bolt (M-0) 096-160

E04-ECPRA0 Bolt Cap (RAL-900) 096-10

9 E04-ECPR44A0 Retaining Bolt 096-10

40 E04-EHLL1A0 Arm Mounting Bracket Ass'y (RAL-900) (1) 096-190

41 E04-ECPR4B0 Bottom Cover 1 096-1400

4 E04--------------- Bottom Cover Screw (M-6) 096-1410

4 E04-ECPRA0 Bolt Cap (RAL-900) 4 096-140

44 E04--------------- Coach Bolt (ø9-) (RAL-900) 4 096-140

4 E04-ECPJ19A0 Wall Mount Plate (RAL-900) 1 096-160

E04-ECPJ1A0 Wall Mount Plate for Cover Type (1) 096-161

46 E04-EHLL11A0 Wire Harness in Balance Arm 1 096-101

4 E04-ECNR1A0 Yoke Side Cap (R1-100) 1 096-160

4 E04-ECPJ1A0 Yoke Inside Cover (RAL-900) 1 096-101

49 E04-ECLJA0 Adjust Wrench 1 096-10

0 E04-EHLL14A0 Cover Set for Wall Mount Plate (RAL-900) (Option) (1) 096-16

60 E04-EHLL0A0 Control Box Assembly (for 10V) (RAL-900) (1) 096-1000

61 E04-ECPE01A0 Chassis (R1-100) 1 096-1440

6 E04--------------- Fuse (F1/F - 10A) 096-140

6 E04--------------- Terminal Block (4P) 1 096-1460

64 E04--------------- Power Switch 1 096-140

6 E04-ECPJ0D0 Covering Plate 1 096-140

66 E04--------------- Timer Board 1 096-1490

6 E04-ECPB01G0 Front Cover 1 096-100

6 E04--------------- Top Screw (M-) 096-110

69 E04--------------- Fuse (F - 0.A) 1 096-10

0 E04--------------- Wood Screw for Chassis (ø.-) 4 096-10

1 E04--------------- Fuse (F4 - 1A) 1 096-140

E04--------------- Terminal Block (P) 1 096-10

E04-ECNJ4A0 Hook for Hand Exposure Switch 1 096-160

4 E04-EHLL1A0 Hand Exposure Switch Ass'y 1 096-10

E04--------------- Screw for Hook (ø-1) 1 096-10

6 E04-ECPR0A0 Restriction Plate 1 096-190

E04--------------- Power Board for 10V 1 096-1600

E04-ECPJ0D0 Front Sheet (RAL-900) 1 096-1610

Page 26: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- 6 -

34

65

2 1

1011

13

33

14

15

78

2019181817

21

2524

23

22

3321

3223 31

3029

2726

28

35

36

37

38

39

41

42

43

44

4546474849

50

51

52

16

9

3440

[3] SLIDING SECTION / ARM AND HEAD ASSEMBLY OF RK TYPE

Page 27: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

- -

ID.No. PartsNo. Description QTY USA Code

1 E04-EHLL0A0 X-Ray Head Assembly (RAL-900) 1 096-1010

E04-ECPR1D0 Regular Cone (RAL-900) 1 096-1060

E04-ECPR6A0 X-Ray Exposure Sleeve 1 096-100

4 E04-ECPRB0 Lock Ring (R1-100) 1 096-1040

E04-EHLL1A0 Housing Cover Set (RAL-900) 1 096-100

6 E04-ECNR1A0 Yoke Side Cap (R1-100) 1 096-160

E04-ECPJ1A0 Yoke Inside Cover (RAL-900) 1 096-101

E04-ECPE0G0 Yoke (RAL-900) 1 096-100

9 E04-EHLL0A0 Swing Arm Assembly (RAL-900) 1

10 E04-ECLR9B0 Head Key 1 096-10

11 E04-ECLS01A0 Stopper Ring 1 096-1090

1 E04--------------- Collar Screw (SUS M4-) 1 096-1110

14 E04-ECLSA0 Collar (R1-100) 1 096-110

1 E04--------------- Stopper Screw(M4-) 1 096-110

16 E04-ECLF0B0 Swing Arm 1(RAL-900) 1

1 E04--------------- Swing Arm Pin Screw(M6-16) 1

1 E04--------------- Stopper Screw(M4-)

19 E04-ECLT4B0 Swing Arm Sleeve 1

0 E04-ECLT46C0 Swing Arm Pin 1

1 E04--------------- Brake Screw(M6-6)

E04-ECLS11B0 Brake Spring(ø-11) 1

E04-ECLS06A0 Brake Plug

4 E04-ECLS1A0 Brake Sleeve 1

E04--------------- Swing Arm Sleeve Screw(M-)

6 E04-ECLT44A0 Swing Arm Joint 1

E04-ECLF19B0 Swing Arm W/Belmont Seal(RAl-900) 1

E04--------------- Tapping Screw(ø4.-40) 4

9 E04-ECLT4A0 Thrust Washer(ø4-t) 1

0 E04-EEMU19A0 Bushing 1

1 E04-ECLS09A0 Stopper Bolt(M6-1) 1

E04-ECLS4A0 Brake Spring(ø4.-) 1

E04-EHLZA0 Cable W/Connector 1

4 E04-EHLL19A0 Column Assembly (RAL-900) 1

E04-ECLK6A0 Seat Cushion (NL-) 1

6 E04-F0 Knob(black) 1

E04-P000 Gas Cylinder 1

E04-ECLKA0 Backrest Cushion (NL-) 1

9 E04-ECLT0A0 Cushion Plate 4

40 E04-ECLK69A0 Sliding Post 1

41 E04-EALJ14A0 Column Cover 1

4 E04-EHLZA0 Guide Roller Assembly 4

4 E04-EFPR0A0 Cable Guide W/Nut 1

44 E04-EHLL1A0 Constant Tension Spring Assembly 1

4 E04-EHLL1A0 Column W/Sub Plate(RAL-900) 1

46 E04--------------- Mounting Bolt(M6-0) 4

4 E04--------------- Spring Washer(M6) 4

4 E04--------------- Washer(M6) 4

49 E04--------------- Set Screw(M-1) 4

0 E04--------------- Lag Bolt(ø-4)

1 E04-ECLTA0 Gas Cylinder Sleeve(R1-100) 1

E04-ECLKB0 Base Plate(R1-100) 1

Page 28: ·Wall Mount Type WK ·Room Mount Type · 2020. 6. 11. · The MODEL 096 x-ray system will operate on a power supply of 120V AC,+12V AC,-12V AC (108V AC to 132V AC). A 3 wire GROUNDED

FOR FDA USE ONLY

a. NAME OF HOUSPITAL, DOCTOR OR OFFICF WHERE INSTALLED

a. THIS REPORT IS FOR ASSEMBLY OF CERTIFIED COMPONETS WHICH ARE(Check appropriate box(es))

b. INTENDED USE(S)(Check applicble box(es))

c. THE X-RAY SYSTEM IS (Check one) d. THE MASTER CONTROL IS IN ROOM e. DATE OF ASSEMBLY

(mo.) (day) (yr.)

STATIONARY

MOBILE

a. THE MASTER CONTROL IS

f. SELECTED COMPONENTSf. (Enter number of each installed in appropriate blocks)

b. CONTROL MANUFACTURER

c. CONTROL MODEL NUMBER

a. PRINTED NAME b. SIGNATURE c. DATE

d. COMMENTS

Complete the following infomation for the certified components listed below which you installed. For beam limited device, tables and CT gantries the manufacturer and Model number in theindicated spaces. For other certified components, enter in the appropriat block how many of each you installed in this system.

b. CONTROL SERIAL NUMBER

OTHER CERTIFIED COMPONENTS

X-RAY CONTROL

HIGH VOLTAGE GENERATOR

VERTICAL CASSETTE HOLDER

TUBE HOUSING ASSEMBLY(Medical)

DENTAL TUBE HEAD

DATE MANUFACTURED

DATE MANUFACTURED

DATE MANUFACTURED

DATE MANUFACTURED

DATE MANUFACTURED

MODEL NUMBER

MODEL NUMBER

MODEL NUMBER

MODEL NUMBER

MODEL NUMBER

MANUFACTURER

MANUFACTURERBEAM

LIMIT

ING

DEVI

CETA

BLES

CTGA

NTRY

MANUFACTURER

MANUFACTURER

MANUFACTURER

CRADLE

FILM CHANGER

IMAGE INTENSIFIER

SPOT FILM DEVICE

OTHER (Specify)

c. SYSTEM MODEL NAME (CT Systems Only)

A NEW INSTALLATIONEXISTING (Certified)

EXISTING (Non-certified)

ANGIOGRAPY

TOMOGRAPHY (Other than CT)

GENERAL PURPOSE FLUOROSCOPY

GENERAL PURPOSE RADIOGRAPHY

CHEST

CHIROPRACTIC

MAMMOGRAPHY

UROLOGY

PODIATRY

DENTAL - INTRAORAL

DENTAL - CEPHALOMETRIC

HEAD - NECK (Medical)

REASSEMBLY - MIXED SYSTEM (Both certified and uncertified components)

REASSEMBLY - FULLY CHRTIFIED SYSTEM

NEW ASSEMBLY - FULLY CHRTIFIED SYSTEM

AN ADDITION TO AN EXISTING SYSTEM

REP;ACEMENT COMPONETS IN AN EXISTING SYSTEM

CT WHOLE BODY SCANNER

CT HEAD SCANNER

DIGITAL

OTHER (Specify in comments)

C - ARM FLUOROSCOPIC

RADIATION THERAPY SIMULATOR

DENTAL PANRAMIC

b. STREET ADDRESS

c. CITY d. STATE

e. ZIP CORD

a. COMPANY NAME

b. STREET ADDRESS

c. CITY

e. ZIP CORDf. TELEPHONE NUMBER

d. STATE

From Approved: OMB No. 0910-0213Expiration Date: December 31, 1992See reverse for OMB statement.

D XXXXXX

f. TELEPHONE NUMBER

1. EQUIPMENT LOCATION

3. GENERAL INFORMATION

4. COMPONENT INFORMATION (If additional space is needed for this section use another form, replacing the preprinted number with form Number and complete items 1,4, and 5 only)

5. ASSEMBLER CERTIFICATION

FORM FDA 2579(7/92) PREVIOUS EDITION IS OBSOLETE

2. ASSEMBLER INFORMATION

DEPARTMENT OF HEALTH AND HUMAN SERVICESPUBLIC HEALTH SERVICE

REPORT OF ASSEMBLYOF A DIAGNOSTIC X-RAY SYSTEM

FOOD AND DRUG ADMINISTRATION

I affirm that all certified components assembled or installed by me for which this report is being made, were adjusted and tested by me according to the instructions provided by the manufacturer(s), were of the type required by the diagnostic x-ray performance standard (1 CFR Part 100), were not modified to adversely affect performance, and were insralled in accordance with provisions of 1 CFR Part 100. I also affirm that all instruction manuals and other infomation required by 1 CFR Part 100 for this assembly have been furnished to the purchaser and within 1 days from the date of assembly, each copy of this report will be distributed as indicated at the bottom of each copy.

BELMONT EQUIPMENT, Division of Takara Belmont, USA, Inc.101 Belmont Drive Somerset, New Jersey 0 U.S.A. TEL.:() 469-000 / (00) -119 Fax.:()6-6 / (00) 0-04

TAKARA CO, CANADA LTD.06 S. Sheridan Way, Mississauga, Ont., LJM4, Can. TEL.:(90) - Fax.:(90)-60

Book No. EDAP Printed in Japan 004 MA

APPENDIXTHREE:CERTIFICATIONFORMFDA2579FROMTHEDEPARTMENTOFHEALTHANDHUMANSERVICESMUSTBEFILLEDAND MAILEDTOTHE RESPECTIVEAGENCIES FORTHIS INSTALLATIONTOBECONSIDEREDCOMPLETE.ALSOCOMPLETETHEWARRANTYCARDANDTHEASSEMBLER'SINSTALLATIONREPORTANDRETURNTOBELMONTEQUIPMENTCORP.

REFER TO THE SAMPLE FORM: