2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Volumetric Modulated Arc TherapyVolumetric Modulated Arc Therapy--
Patient Specific QA Patient Specific QA
Daliang Cao, PhD, DABR
Swedish Cancer Institute, Seattle, WA
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
VMAT plan QA methodsVMAT plan QA methods
Film & ion chamberFilm & ion chamber
diode array diode array
MapcheckMapcheck in in MapPhanMapPhan phantomphantom
ArcCheckArcCheck
Delta4Delta4
2D ion chamber array 2D ion chamber array
MatriXXMatriXX in MULTICube phantomin MULTICube phantom
PTW 729 in PTW 729 in OctaviusOctavius phantomphantom
Composite dose measurement Fluence measurement
2D ion chamber array2D ion chamber array
MatriXXMatriXX mounted on Gantrymounted on Gantry
Portal Portal dosimetrydosimetry
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Which system do you use/intend to use Which system do you use/intend to use in your clinic for VMAT QAin your clinic for VMAT QA
0%
0%
0%
0%
0%
10
A.A. Film & ion chamberFilm & ion chamber
B.B. Ion chamber array (Ion chamber array (MatriXXMatriXX, PTW 729), PTW 729)
C.C. Diode array (Diode array (mapcheckmapcheck, , ArcCheckArcCheck, Delta4), Delta4)
D.D. Portal Portal dosimetrydosimetry
E.E. othersothers
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Film & Ion ChamberFilm & Ion Chamber
A stack of solid water phantom with a total thickness of 15cm.A stack of solid water phantom with a total thickness of 15cm.A 0.6cc farmer chamber was inserted at the depth of 10cm.A 0.6cc farmer chamber was inserted at the depth of 10cm.A film was sandwiched 1cm above the ion chamber. A film was sandwiched 1cm above the ion chamber.
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Prostate CaseProstate Case
Plan Measurement
Ion Chamber measurement is 1.3% below the predicted value
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
H&N CaseH&N Case
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
H&N Case: absolute dose comparisonH&N Case: absolute dose comparison
The Gamma analysis passing rate in this case is 96.4%The Gamma analysis passing rate in this case is 96.4%The measured ion chamber dose was 0.2% more than the planned oneThe measured ion chamber dose was 0.2% more than the planned one
measurement plan
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
2D Diode Array2D Diode Array
A MapCheck device inserted into a A MapCheck device inserted into a MapPhanMapPhan solid water phantomsolid water phantom
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
MapCheck in MapCheck in MapPhanMapPhan QA resultQA result
With 3% and 3mm criteria, the gamma analysis passing rate is 96.With 3% and 3mm criteria, the gamma analysis passing rate is 96.5%5%Dose threshold in gamma analysis is 10%Dose threshold in gamma analysis is 10%
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Dose Rate DependencyDose Rate Dependency
0.975
0.980
0.985
0.990
0.995
1.000
1.005
0.00 50.00 100.00 150.00 200.00 250.00 300.00 350.00 400.00 450.00 500.00
dose rate (MU/minute)
rela
tive
dose
res
pons
e
Mapcheck diode
MatriXX ion chamber
The variation of dose response with the dose rate can reach up to 2.5% for MapCheck diode.
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Calibrate MapCheck with medium dose rateCalibrate MapCheck with medium dose rate
Gamma passing rate increased to 98.8%Gamma passing rate increased to 98.8%
Dose calibration was performed at 224 MU/Min
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Angular dose response of MapCheck Angular dose response of MapCheck diodediode
Significant variation of the dose response can be found when theSignificant variation of the dose response can be found when the incidence incidence beam angle is parallel to the detector plane.beam angle is parallel to the detector plane.Each diode in MapCheck may have different angular dose response Each diode in MapCheck may have different angular dose response curve. curve.
-10.0
-5.0
0.0
5.0
10.0
15.0
20.0
25.0
30.0
-180 -120 -60 0 60 120 180Gantry Angle
Per
cent
Err
or (
%)
180°°°°
90°°°°
0°°°°
-90°°°°
Couch
detector array& phantom
-180°°°°
phantom setup MapCheck Diode MatriXX ion chamber
-10.0
-5.0
0.0
5.0
10.0
15.0
20.0
25.0
30.0
-180 -120 -60 0 60 120 180
Gantry Angle
perc
ent e
rro
r (%
)
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
1386 diode detectors arranged in cylindrical geometry1386 diode detectors arranged in cylindrical geometry
Measures entrance and exit doseMeasures entrance and exit dose
The QA can be done in composite or per control point The QA can be done in composite or per control point
Arc Check Arc Check –– Sun NuclearSun Nuclear
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
ArcCHECKArcCHECK features a versatile central cavityfeatures a versatile central cavity
–– May be used to accommodate different detectors and insertsMay be used to accommodate different detectors and inserts
–– MultiPlugMultiPlug (left) also accommodates EBT film cassette and (left) also accommodates EBT film cassette and multiple detector locationsmultiple detector locations
Central Cavity OptionsCentral Cavity Options
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
SNC Patient SoftwareSNC Patient Software
Software enables composite and control point analysisSoftware enables composite and control point analysis
DICOM RT Dose is imported and DICOM RT Dose is imported and ArcCHECKArcCHECK software then extracts 3D software then extracts 3D dose corresponding to detector locations, and performs a comparidose corresponding to detector locations, and performs a comparisonson
All data files from All data files from ArcCHECKArcCHECK are open format for easy exportare open format for easy export
–– Raw data also available for exportRaw data also available for export
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
ArcCheckArcCheck QA resultQA result
Gamma analysis passing rate is 95.2% for this case.
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Virtual InclinometerVirtual Inclinometer
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
3DVH Option3DVH Option
The most advanced 3D patient dose and DVH tools availableThe most advanced 3D patient dose and DVH tools available
–– Uses existing Uses existing ArcCHECKArcCHECK measurementsmeasurements
–– Supports VMAT and Supports VMAT and RapidArcRapidArc deliveriesdeliveries
–– No secondary dose calculation algorithmNo secondary dose calculation algorithm
–– 3D dose and DVH analysis on 3D dose and DVH analysis on patientpatient geometry (not phantom geometry (not phantom geometry)geometry)
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
ScandidosScandidos DeltaDelta 44
A cylinder-shaped plastic phantom with 2 imbedded orthogonal crossing detector planes.1069 diode detectorsDose is recorded in 2 planes and a 3D dose is reconstructed for comparison with the QA plan.
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
CompositeComposite--level analysis of level analysis of SmartArc plan with DeltaSmartArc plan with Delta44
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
MatriXXMatriXX & MULTICube phantom& MULTICube phantom
MatriXX system has 1020 vented ion chambers with an active area MatriXX system has 1020 vented ion chambers with an active area of of 24.4cm x 24.4cm.24.4cm x 24.4cm.
The size of each ion chamber is 4.5cm (diameter) x 5cm (height).The size of each ion chamber is 4.5cm (diameter) x 5cm (height). The raw The raw pixel size is 7.62mm, which can be interpolated down to 1mm.pixel size is 7.62mm, which can be interpolated down to 1mm.
Laser is used to align system. Dose from both coronal and Laser is used to align system. Dose from both coronal and sagittalsagittal planes planes can be measured depending on the phantom.can be measured depending on the phantom.
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
MatriXX data conversionMatriXX data conversion
The raw image has only 1024 pixels with 7.62mm dose grid. The daThe raw image has only 1024 pixels with 7.62mm dose grid. The data set can be interpolated to ta set can be interpolated to 1mm dose grid.1mm dose grid.
linear
Cubic Spline
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Results: Prostate CaseResults: Prostate Case
With 3% and 3mm criteria, the passing rate is 99% for this coronWith 3% and 3mm criteria, the passing rate is 99% for this coronal sliceal slice
measurement plan
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
MatriXX QA: H&N CaseMatriXX QA: H&N Case
measurement plan
With 3% and 3mm criteria, the passing rate is 98.2% for this corWith 3% and 3mm criteria, the passing rate is 98.2% for this coronal sliceonal sliceNo dose threshold was applied in gamma analysis. Note the passinNo dose threshold was applied in gamma analysis. Note the passing rate reduces to 97.1% g rate reduces to 97.1% if a 10% threshold is applied. if a 10% threshold is applied.
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Angular variation of dose responseAngular variation of dose response
Up to 8% dose variation can be found when the photon beam is parUp to 8% dose variation can be found when the photon beam is parallel to the detector allel to the detector planeplane
180°°°°
90°°°°
0°°°°
-90°°°°
Couch
detector array& phantom
-180°°°°
phantom setup
-10.0
-6.0
-2.0
2.0
6.0
10.0
-180 -120 -60 0 60 120 180Gantry Angle
per
cen
t err
or
(%)
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
MatriXX EvolutionMatriXX Evolution
The gantry angle for each frame is recorded during the VMAT deliThe gantry angle for each frame is recorded during the VMAT delivery.very.
A angular dose response correction is applied to all the data baA angular dose response correction is applied to all the data based on the sed on the gantry angle information.gantry angle information.
InclinometerInput
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
With and Without Angular response With and Without Angular response correctioncorrection
The measured data without angular dose response The measured data without angular dose response correction is slightly colder than the one with correction.correction is slightly colder than the one with correction.
PlanMeasurement
Without CorrectionMeasurement
With Correction
A two-arc case: H&N
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
With and Without Angular response With and Without Angular response correctioncorrection
3% and 3mm passing criteria for Gamma analysis.3% and 3mm passing criteria for Gamma analysis.
Without Correction81.9%
With Correction96.2%
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Which of the following cannot be Which of the following cannot be measured using measured using MatriXXMatriXX 2D ion chamber 2D ion chamber
arrayarray
0%
0%
0%
0%
0%
10
A.A. Dose distribution on a Dose distribution on a sagittalsagittal planeplane
B.B. Absolute dose at detector centerAbsolute dose at detector center
C.C. Dose distribution on a coronal planeDose distribution on a coronal plane
D.D. Dose distribution on a transverse plane Dose distribution on a transverse plane
E.E. Radiation Radiation fluencefluence of a VMAT planof a VMAT plan
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
AnswerAnswer
A.A. Dose distribution on a Dose distribution on a sagittalsagittal planeplane
B.B. Absolute dose at detector centerAbsolute dose at detector center
C.C. Dose distribution on a coronal planeDose distribution on a coronal plane
D.D. Dose distribution on a transverse plane Dose distribution on a transverse plane
E.E. Radiation Radiation fluencefluence of a VMAT planof a VMAT plan
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Portal Portal DosimetryDosimetry
Portal Dosimetry Process
Created verification plan for each patient
Calibrated panel prior to collecting data
Data measured in units of CU
Definition of Calibrated Unit (CU)Field Size =10 x 10 cm2SDD = 105 cmDeliver 100 MU to panelSet reading = 1 CU1 CU is numerically approximate to 1 Gy
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Portal Portal DosimetryDosimetry –– RapidArcRapidArc (Prostate)(Prostate)
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Portal Portal DosimetryDosimetry –– RapidArcRapidArc (Prostate)(Prostate)
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Portal Portal DosimetryDosimetry –– RapidArcRapidArc (HN)(HN)
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Portal Portal DosimetryDosimetry –– RapidArcRapidArc (HN)(HN)
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Which QA system has the highest Which QA system has the highest resolutionresolution
0%
0%
0%
0%
0%
10
A.A. portal portal dosimetrydosimetry using EPIDusing EPID
B.B. filmfilm
C.C. MatriXXMatriXX 2D ion chamber array2D ion chamber array
D.D. ArcCheckArcCheck
E.E. Delta4Delta4
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
AnswerAnswer
A.A. portal portal dosimetrydosimetry using EPIDusing EPID
B.B. filmfilm
C.C. MatriXXMatriXX 2D ion chamber array2D ion chamber array
D.D. ArcCheckArcCheck
E.E. Delta4Delta4
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
SummarySummary
VMAT QA using film and ion chamber is easy to setup and perform.VMAT QA using film and ion chamber is easy to setup and perform.However, there is no instantaneous result and It is timeHowever, there is no instantaneous result and It is time--consuming consuming for data analysis.for data analysis.
DiodeDiode--array and ionarray and ion--chamberchamber--array based devices can provide array based devices can provide instantaneous data analysis and can be used for VMAT patient instantaneous data analysis and can be used for VMAT patient specific QA once the angular dose response and dose rate specific QA once the angular dose response and dose rate dependence is considered during data analysis/collection.dependence is considered during data analysis/collection.
Portal Portal dosimetrydosimetry can provide an easy and robust way for patient can provide an easy and robust way for patient specific QA. The only issue with this technique is that informatspecific QA. The only issue with this technique is that information on ion on gantry angle cannot be checked independently. gantry angle cannot be checked independently.
Users need to know limit of each QA technique/device for best Users need to know limit of each QA technique/device for best patient specific QA result.patient specific QA result.
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
AcknowledgementAcknowledgement
David M. Shepard, Ph.D.David M. Shepard, Ph.D.
Muhammad K. N. Afghan, Ph.D.Muhammad K. N. Afghan, Ph.D.
Fan Chen, Ph.D.Fan Chen, Ph.D.
Min Rao, Ph.D.Min Rao, Ph.D.
Jinsong Ye, M.S.Jinsong Ye, M.S.
Tony P. Wong, Ph.D.Tony P. Wong, Ph.D.
Vivek Mehta, M.D.Vivek Mehta, M.D.
David HousleyDavid Housley
Grace KimGrace Kim
Igor Gomola, Ph.D.Igor Gomola, Ph.D.
Gerry Gerry VantellingenVantellingen
Jie Shi, Ph.D.Jie Shi, Ph.D.
Kris LyleKris Lyle
2012 AAPM Annual Meeting, Charlotte2012 AAPM Annual Meeting, Charlotte
Thank you!Thank you!