VISIONIST X4 CRT-P and VISIONIST CRT-P Models U225, U226, U228 Provides an ImageReady MR Conditional system* at 3T and 1.5T, full body, with no time limitations, with automatic MRI timeout feature to optimize workflow in the MR environment Extended Longevity device projected to last up to 13.8 years** The X4 CRT-P and ACUITY X4 lead portfolio features: – electrodes on a 3D spiral designed to allow pacing from a basal location without sacrificing fixation or thresholds – 17 pacing options to manage phrenic nerve stimulation and high thresholds – multiple lead shapes with a 2.6F (0.86 mm) tapered lead tip to improve predictability of accessing target vessels Advanced AF and HF diagnostic reports provide a comprehensive and proactive approach for comorbidity management RF telemetry for wireless transmission of information and efficiency in the operating room and follow-up setting Remote patient monitoring with the LATITUDE NXT Patient Management System RightRate MV sensor is the only MV sensor clinically proven to restore chronotropic competence 1 Enhanced features and diagnostics, including AP Scan and Respiratory RateTrend, designed to provide you with greater insight into your patient’s disease progression based on the patient’s own respiration POST function to facilitate patient follow up with a fully automatic device and lead check EASYVIEW header with port identifiers designed to make the implant experience more efficient *Model U228 only. When conditions of use are met, please refer to MRI Technical Guide. **Assumes 70 min-1 LRL, DDDR mode; 100% biventricular pacing; 15% atrium pacing and 0.4 ms pacing. Pulse Width (RA, RV, LV); RA 500 Ω; RV/LV 700Ω; sensors Off, EGM Onset On. Mechanical Specifications Size (cm) Connector Type Model Type (W x H x D) Mass (g) Volume (cc) (RA RV LV) U225 (non-MRI model) CRT-P 4.45 x 6.13 x 0.75 30.6 15.2 RA/RV/LV: IS1 U226 (non-MRI model) CRT-P 4.45 x 6.13 x 0.75 31.1 15.2 RA/RV: IS1 – LV:LV1 U228 X4 CRT-P 4.45 x 6.17 x 0.75 33.0 16.6 RA/RV: IS1 – LV:IS4 Assumes 70 min-1 LRL, DDDR mode; 100% biventricular pacing; 15% atrium pacing and 0.4 ms pacing. PulseWidth (RA, RV, LV); RA Impedance 500 Ω; sensors On, EGM Onset On. These calculations also assume that the pulse generator spends 6 months in Storage mode during shipping and storage, the ZIP telemetry use for 3 hours at implant time and for 40 minutes annually for in-clinic follow-up checks. Power Supply: lithium-carbon monofluoride cell; Boston Scientific; 40229. Additional Longevity Information Longevity Projected Longevity RA/RV LV 500Ω 700Ω 700Ω, no MV/RS Typical programmed setting 2.5V 3.0V 10.3 11.2 11.7 Maximum labeled longevity 2.0V 2.0V 12.4 13.1 13.8

Transcript of VISIONIST X4 CRT-P and VISIONIST CRT-P - Boston Scientific › content › dam › bostonscientific...

Page 1: VISIONIST X4 CRT-P and VISIONIST CRT-P - Boston Scientific › content › dam › bostonscientific › … · VISIONIST™ X4 CRT-P and VISIONIST ™ CRT-P Models U225, U226, U228

VISIONIST™ X4 CRT-P and VISIONIST™ CRT-PModels U225, U226, U228

•ProvidesanImageReady™MRConditionalsystem*at3Tand1.5T,fullbody,withnotimelimitations,withautomaticMRItimeoutfeaturetooptimizeworkflowintheMRenvironment

•ExtendedLongevitydeviceprojectedtolastupto13.8years**•TheX4CRT-PandACUITY™X4leadportfoliofeatures: –electrodesona3Dspiraldesignedtoallowpacingfroma

basallocationwithoutsacrificingfixationorthresholds –17pacingoptionstomanagephrenicnervestimulationand

highthresholds –multipleleadshapeswitha2.6F(0.86mm)taperedlead

tiptoimprovepredictabilityofaccessingtargetvessels•AdvancedAFandHFdiagnosticreportsprovidea

comprehensiveandproactiveapproachforcomorbiditymanagement

•RFtelemetryforwirelesstransmissionofinformationandefficiencyintheoperatingroomandfollow-upsetting

•RemotepatientmonitoringwiththeLATITUDE™NXTPatientManagementSystem

•RightRate™MVsensoristheonlyMVsensorclinicallyproventorestorechronotropiccompetence1

•Enhancedfeaturesanddiagnostics,includingAPScan™andRespiratoryRateTrend,designedtoprovideyouwithgreaterinsightintoyourpatient’sdiseaseprogressionbasedonthepatient’sownrespiration

•POSTfunctiontofacilitatepatientfollowupwithafullyautomaticdeviceandleadcheck•EASYVIEW™headerwithportidentifiersdesignedtomaketheimplantexperiencemoreefficient*Model U228 only. When conditions of use are met, please refer to MRI Technical Guide.

**Assumes 70 min-1 LRL, DDDR mode; 100% biventricular pacing; 15% atrium pacing and 0.4 ms pacing. Pulse Width (RA, RV, LV); RA 500 Ω; RV/LV 700Ω; sensors Off, EGM Onset On.

Mechanical Specifi cations Size (cm) Connector Type Model Type (W x H x D) Mass (g) Volume (cc) (RA RV LV)

U225 (non-MRI model) CRT-P 4.45x6.13x0.75 30.6 15.2 RA/RV/LV:IS1

U226 (non-MRI model) CRT-P 4.45x6.13x0.75 31.1 15.2 RA/RV:IS1–LV:LV1

U228 X4CRT-P 4.45x6.17x0.75 33.0 16.6 RA/RV:IS1–LV:IS4

• Assumes70min-1LRL,DDDRmode;100%biventricularpacing;15%atriumpacingand0.4mspacing.PulseWidth(RA,RV,LV);RAImpedance500Ω;sensorsOn,EGMOnsetOn.Thesecalculationsalsoassumethatthepulsegeneratorspends6monthsinStoragemodeduringshippingandstorage,theZIP™telemetryusefor3hoursatimplanttimeandfor40minutesannuallyforin-clinicfollow-upchecks.

• PowerSupply:lithium-carbonmonofluoridecell;BostonScientific;40229.

Additional Longevity Information

Longevity

Projected Longevity RA/RV LV 500Ω 700Ω700Ω,

no MV/RS

Typicalprogrammedsetting 2.5V 3.0V 10.3 11.2 11.7

Maximumlabeledlongevity 2.0V 2.0V 12.4 13.1 13.8

Page 2: VISIONIST X4 CRT-P and VISIONIST CRT-P - Boston Scientific › content › dam › bostonscientific › … · VISIONIST™ X4 CRT-P and VISIONIST ™ CRT-P Models U225, U226, U228

VISIONIST™ X4 CRT-P and VISIONIST™ CRT-PModels U225, U226, U228

1. ChronotropiccompetenceisdefinedbytheModeloftheCardiacChronotropicResponsetoExercise.WilkoffB,CoreyJ,BlackburnG.Amathematicalmodelofthecardiacchronotropicresponsetoexercise.JournalofElectrophysiology.1989;3:176–180.

All cited trademarks are the property of their respective owners. CAUTION:Thelawrestrictsthesedevicestosalebyorontheorderofaphysician.Indications,contraindications,warningsandinstructionsforusecanbefoundintheproductlabelingsuppliedwitheachdevice.Informationfortheuseonlyincountrieswithapplicablehealthauthorityproductregistrations.GuidantCorporationandGuidantEuropenv/saareBostonScientificcompanies.InformationcontainedhereinisfordistributionoutsidetheU.S.only.InformationnotintendedfordistributioninFrance.

CRM-251102-ACProducedbyGosling.

©2018BostonScientificCorporationoritsaffiliates.Allrightsreserved.DINCRM0851EB

Pacing TherapyBrady Modes Normal:DDD(R)-DDI(R)-VDD(R)-VVI(R)-AAI(R)-DOO-VOO-

AOO-Off

Temporary:DDD-DDI-VDD-VVI-AAI-DOO-VOO-AOO-Off

AT/AF Management ATRModeSwitch,VentricularRateRegulation(VRR),AtrialFlutterResponse(AFR),AtrialPacingPreference(APP),ProACt,RateSmoothing

Automaticity AutomaticGainControl(AGC)forsensitivity

RightAtrialAutomaticThreshold(RAAT)

RightVentricularAutomaticThreshold(RVAT)

LeftVentricularAutomaticThreshold(LVAT)*

Rate Adaptive Pacing Accelerometer,RightRate™(MinuteVentilation)orblendedsensorswithsensortrendingfunction

Rate Management SuddenBradyResponse(SBR),PMTTermination,PVARPafterPVC,DynamicPVARP

RA/RV Pace-Sense Configuration

Unipolar,Bipolar,Bipolar/Unipolar,Unipolar/Bipolar,Unipolar/Off,Bipolar/Off,LeadSafetySwitch,AutomaticLeadRecognition

*Models U225 and U226 only

Implant/In Clinic Follow UpImplant Communication Mode

Programmablevalues:EnableuseofZIP™telemetry(MICS)(RequiresinitialuseofwandfordeviceID)orusewandforalltelemetry

Nominal:EnableuseofZIP™telemetry(RequiresinitialuseofwandfordeviceID)

In Clinic Follow Up SnapshotFunctionupto12secondstraceofECG/EGMdisplaystored

POST(Post-OperativeSystemTest):providesanautomaticdevice/leadcheckatapre-determinedtimepost-implanttohelpdocumentpropersystemfunctionalitywithoutrequiringmanualsystemtesting

Indications-Based Programming (IBP)

Toolthatprovidesspecificprogrammingrecommendationsbasedonthepatient’sclinicalneedsandprimaryindications

Patient DiagnosticsArrhythmia Logbook EventSummary,StoredElectrogramswithAnnotation

Markers(Intervalsandapproximately14minutesallmultichannelEGM,alwayswith10secondsOnsetandeventstorageprioritization).ImplantactivationofallavailableEGMs.Onscreenmeasurementsofallstoredsignal,amplitudesandtiming.SnapshotFunction(upto12secondstraceofECG/EGMdisplaystored)

Histograms & Counters

VentricularTachyCounter,Brady/CRTCounter,Histograms,IntrinsicPromotion(RateHysteresis%successful)

HF Therapy / Diagnostics

HeartRateVariability(HRV)withSDANNandABM,RespiratoryRateTrend,APScan,AT/AFBurden,ActivityLevel,A&VArrhythmias,WeightandBloodPressure**

Atrial Arrythmia Report

AT/AF%andTotalTimeinAT/AF,AT/AFBurdenTrend,RVRateduringAT/AFTrend,PacingPercentTrend,HeartRateTrend,ActivityLevelandRespiratoryRateTrends,RVRateduringAT/AFHistogram.Timelinehistoryofinterrogations,programming,andcounterresetsforoneyear.LongestAT/AF,FastestRVSrateinAT/AF,andmostrecentepisode.

DAILY TREND for last 365 Days

Events,ActivityLevel,AT/AFBurden,PacingPercent,RespiratoryRate,APScan,HeartRate,SDANN,HRVFootprint,ABM,LeadImpedanceandAmplitude,RAATTrend,RVATTrend,LVATTrend

**Weight and Blood Pressure are only available via Latitude.

Remote Follow UpRemote Monitoring ThisdeviceisdesignedtobeLATITUDE™NXTenabled;

LATITUDE™NXTavailabilityvariesbyregion

Thresholds AutomaticstorageoflastsuccessfuldailyPaceSafethresholdtestforallactivechambers

Wireless Remotefollow-upforalldevices(MICS)

Patient Triggered Monitor (PTM)

Triggersthestorageoftwominutesonsetandoneminutepost–EGMs,intervals,andannotatedmarkerdataduringasymptomaticepisodebyplacingamagnetoverthedevice

Heart Failure TherapyHeart Failure Therapy Optimization

SmartDelay™(AVOptimization),BiVTrigger,LVOffset,LVSensing

LV Lead Options QuadripolarLVlead(VISIONISTX4CRT-P)

BipolarandUnipolarLVlead(VISIONISTCRT-P)

LV Pacing Vector Options

17options(VISIONISTX4CRT-P)

6options(VISIONISTCRT-P)

LV Sensing Vector Options

8optionsplusOFF(VISIONISTX4CRT-P)

5optionsplusOFF(VISIONISTCRT-P)

Safety Functions***

Safety Core Isintendedtoprovidelife-sustainingtherapyifcertainnon-recoverableorrepeatfaultconditionsoccur.SafetyCoreoperatesindependentlyandactsasabackuptothesecomponents

Electrocautery Protection Mode

ProvidesasynchronouspacingattheprogrammedoutputsandLRLwhencommandedbytheprogrammer

***The Safety Functions do not have programmable parameters.

www.bostonscientific.eu

ImageReady™ MR-Conditional Pacing SystemMRI Lead Selection MRConditionalPulseGeneratorU128andU228

combinedwith:

PacingLead:FINELINE™II,INGEVITY™

LVLead:ACUITY™X4

MRI Conditions 1.5Tand3T

Fullbody

MRI Mode PacingMode:AOO,VOO,DOO,Off

VentricularPacingChamber:BiV,RV-Only

ProtectionModeTimeOut:Off,3,6,9,12,24,48hours