ViscoTron VP-FT5500 180529 - viscometer.onlineFactory calibration with Newtonian ASTM traceable...
Transcript of ViscoTron VP-FT5500 180529 - viscometer.onlineFactory calibration with Newtonian ASTM traceable...
Torsionally vibrating sensor, no wearing parts Straight flow path, no internal restrictions Eliminates sample cells and pockets Rugged, virtually no maintenance In-line viscosity measurement Laser welded construction Proven operating principle
Your process viscosity measurement solution
Viscotronics Co., Ltd.
ViscoTron VP-FT5502/3ViscoTron VP-FT5502/3Flow through viscosity transducer for process installations
VP-FT5503LL-LTE-A33-4C-551
The VP-FT5500 series of flow through viscosity sensors has been developed employing experience gained over many years with our process viscometers. Using the same torsional vibration principle, the flow through transducer is capable of measuring viscosities accurately over a very wide range in real time.
The sensor has a straight flow through Path. There are no bends or restrictions and therefore can be easily cleaned with foam pigs commonly used in some applications, should this be required. The straight flow through tube makes the sensor rugged and virtually maintenance free.
Laser welded construction and innovative assembly allows the series VP-FT5500 sensors to be manufactured to exacting standards at competitive prices. Laser welding replaces alternative mounting methods of internal components and provides a stable measurement environment.
The series VP-FT5500 is available in versions to measure very low, low, medium and high viscosities. Sensors can be constructed for pressures up to 10 MPa (1,440 psi) and temperatures up to 175˚C (350˚F). Factory calibration with Newtonian ASTM traceable fluids over two decades is standard with three decades optional.
The viscometer creates a shear wave by torsionally resonating the tube. The drag of the fluid on the twisting motion is a measure of the viscosity. The system is factory calibrated using ASTM traceable standards. Speed, direction and turbulence of flow have no influence on the measurement.
The resonant frequency is dependent on the mechanical construction of the sensor and typically between 200 and 600 Hz. The microscopic motion is fast enough not to be influenced by
outside mechanical vibrations, yet low enough to provide excellent sensitivity and resolution for newtonian and non-newtonian fluids.
For fluids which easily adhere to walls, the housing can be heated or cooled with a fluid or electrical heating tape. A PT100 is optionally provided to measure the internal
housing temperature. Low friction coatings for the flow through tube are available as an option.
The VP-FT5500 series works with ViscoTron VT-IRFT induced resonance transmitters. ViscoTron VT-IRFT transmitters induce the natural resonance response in the measuring tube. Applying a fast Fourier Transform (FFT) algorithm provides accurate frequency measurement of the natural resonance. Measuring the power of the Fourier Transform at the resonant frequency is a measure for the viscosity.
VP-FT5502/3
VP-FT5503M-LTI-V73-0N-101
© 2018Viscotronics Co., Ltd.- Aavos International BVBA
[email protected] www.aavos.be www.viscometer.online
© 2018
VP-FT5502/3 SpecificationsDescription VP-FT5502/3LL VP-FT5502/3L VP-FT5502/3M VP-FT5502/3H
Measurement range (mPa·s x g/cm³)
2 decade calibration3 decade calibration
0.00 to 500.00
5.00 to 500.000.5 to 500.00
0.0 to 2,500.0
25.0 to 2,500.02.5 to 2,500.0
0 to 25,000
250 to 25,00025.0 to 25,000
0 to 250,000
2,500 to 250,000250 to 250,000
The measurement and calibration ranges indicated above are a guide only, the final range capability is dependent on pressure requirements, total sensor length and other design factors
Optionally larger calibrated ranges are available
Sensor length 500 to 1000 mm, dependent on range, pressure requirements andprocess connections
Maximum process temperature with external transmitter
LTESTE
< 130°C / 265°F<175°C / 350°F
Maximum process temperature with internal transmitter
LTI < 50°C / 120°F
Resonant frequency Shear rate
200 to 600 Hz (dependent on range and design requirements)Typically between 1,250 to 3,800 sec-1 (dependent on range and design requirements)
Calibration Cable length independent calibration with NIST traceable standards2 decade calibration standard, 3 decade calibration optional
Reproducibility 0.5% or ±1 digit, whichever is greater
Accuracy 1% or ±1 digit (factory calibrated with NIST standards)
Housing IP65
Wetted partsSUS316 / SUS304 standard
Policoat low friction, corrosion resistant coatings up to 175°COther coatings or materials are available on request
Process connection1” NPT and 1” 600# ASME flange are standard
Other process connections optional(maximum pressure capability of sensor: 10 MPa / 1,440 psig)
Cable length 500 meters maximum between sensor and transmitter (50 Ohm max.)
Hazardous area approval (optional)
ATEX: II 1/2 G EEx ia IIC T3 – T6CSA International Class I, Div I, Group C and D
IECEx Ex ia IIB T3-T6 Ga
Process temperature measurement (optional) Internal PT100 or External PT100 provided by customer
Housing Powder coated steel
Housing temperature
A tubular ring, entering and exiting at one end of the housing, can be used for heating or cooling using electrical heating tape or a flowing liquid
Maximum housing temperature 93°C / 200°F with external transmitterMaximum housing temperature 50°C / 120°F with internal transmitter
Internal PT100 optional
Viscotronics Co., Ltd.- Aavos International BVBA [email protected] www.aavos.be
www.viscometer.online
VP-FT5500 ViscoTron sensor Standard range examplesVP-FT5502 Flow through viscometer with 2 decade calibration within range capability (example: 25 to 2,500 mPa.s x gr/cm³)VP-FT5503 Flow through viscometer with 3 decade calibration within range capability (example: 2.5 to 2,500 mPa.s x gr/cm³)
Code Sensor type mPa·s x g/cm³ Other ranges on requestLL Extra Low viscosity 0 to 500 Required maximum viscosity:L Low viscosity 0 to 2,500 Required maximum viscosity:M Medium viscosity 0 to 25,000 Required maximum viscosity:H Highe Viscosity 0 to 250,000 Required maximum viscosity:S Built to customer requirements Required maximum viscosity:
Code Process temperatureLTE < 130°C / 265°F for use with external transmitterSTE < 175°C / 350°F for use with external transmitterLTI < 50°C / 120°F for use with internal transmitter (see VT-IRFT for details)
Code Process connectionN1 1" NPTA2 1" 300# ASME FlangeA3 1” 600# ASME FlangeD4 DN25 PN40 DIN FlangeD5 DN25 PN100 DIN FlangeT6 1“ Tri-ClampV7 25 mm Varivent flangeN8 1/2” NPTS9 Others on request
Code Tube diameter 1 Ø 6 mm2 Ø 12 mm3 Ø 25 mm9 Other diameters on request
Code Installation type (Approvals include safety barriers)0 General purpose area1 Safety barriers for viscosity only4 Explosion proof sensor for viscosity only
Code Agency ApprovalsN None, for general purpose areaA ATEX II 1/2G EEx ia IIC T3 – T6
Safety barriers for viscosity onlyC CSA Class I, Div I, Group C and DI IECEx Ex ia IIB T3-T6 Ga
Code Coating and materials for wetted parts1 5
SUS3165 Policoat (request separate data sheet for more information)9 Special sensor coating or material on request
Code Process temperature measurement0 No temperature measurement5 PT100 for housing temperature
Code Housing (Nema 4)1 Powder coated9 Others on request
^ ^ ^ ^ ^ ^ ^ ^ ^ ^VP-FT5503 L STE 1 3 2 C 5 2 1 EXAMPLE ORDER INFORMATION
VP-FT5502/3 Configuration
Not all combinations are available for all configurations, some can only be combined with other options Technical improvements reserved
VP-FT5502H-LTE-N82-0N-101
© 2018Viscotronics Co., Ltd.- Aavos International BVBA
[email protected] www.aavos.be www.viscometer.online
V18/05
/29
Compatible TransmittersFor more info see the transmitter brochure
Emulsions Crude Oil Blending
Black Liquor Ceramics
Filling Operations Additives Slurries
Coatings Spray Painting Food Industry
Creams Sauces
Approvals and Application Examples
Viscotronics Co., Ltd. Engineering Office
#22- 2475 Dobbin Rd.West Kelowna, BC V4T 2E9
Canada+1-250-801 8089
Calibration and Service 32 Stradwick Way
Calgary, AB T3H 1G9Canada
+1-403-249 [email protected]
ViscoTron VT-IRFTi-TFT(Wall mount enclosure)
ViscoTron VT-IRFTx-TFTTP(DIN Rail / Wall mount enclosure)
D
E
C
B
A
321 5
C
D
4 6 7 8
A
B
321 54 6 7 8
Drawn
Last RevisionMFR Appr.
CAD generated, do not manually update. Distribution format is PDF in A3.
Wetted Materials:
DMB 2018-03-08
Do not scale from drawing
Size: A3
Drawing #:
Sheet: 1 of 1
VP-FT5500-A1/600
Dimensions:
ViscoTronDesigned in British Columbia, [email protected]
www.viscotronics.com
West Kelowna, BC, CanadaViscotronics Co., Ltd. Qty.:
1Copyright:This drawing is the proprietary property of Viscotronics Co., Ltd. and its its subsidiaries or affiliates. It is not to be divulged to any third party. This drawing shall not be copied mechanically or otherwise without the written consent of Viscotronics Co., Ltd.
Tolerance: No decimal: 0.5 mmIn millimeters unless otherwise noted
Customer:
Project:
Inquiry / PO #:
Approval signature and date:
Notes:
CustomerProject
email inquiry
Repr
esen
tativ
e:
Policoat
Flow through viscometer for low viscosities
1000.0 750
312.5 367.6
Distributor for Europe:
AAVOS International bvba Sparkevaardekenstraat 3
8600 Diksmuide Belgium
Tel : +32 (0)51 69 78 15 Fax: +32 (0)51 69 78 17
[email protected] www.aavos.be