VIDAS BLUE
VIDAS BLUEVIDAS BLUE PRESENTATIONPRESENTATION
VIDAS BLUE
VIDAS BLUE:
New Vidas and mini Vidas look
VIDAS BLUE
The main changes
VIDAS BLUE
VIDAS BLUE
Strip preparation tray
Blue tray covers
VIDAS BLUE
Removingstrip preparation tray
Mini VIDAS BLUE
Mini VIDAS BLUE
New screen / keyboard Holder
New printer paperholder
Internal changes
VIDAS BLUE
VIDAS BLUE
On mini VIDAS:Addition of a long screw to fix the right trim
VIDAS BLUE
On mini VIDAS:
Screen/keyboard holder 3 fixing points:
VIDAS BLUE
New Power Supply Assembly: - Only 4 external points of fixation (4 screw)- Central chimney (V30)- No railing(bar)
VIDAS BLUE
New Power Supply Assembly:- Only 1 Power Supply Board (V12/V30)- 1 air fan- Self switching Power Supply (110/220V)
VIDAS BLUE
New Power Supply Board:-Dedicated mini Vidas and Vidas connectors.
VIDAS
Mini VIDAS
Be carefullRespect the correct connector!
VIDAS BLUE
New Power Supply Board:
-Air Fan•Fixed by 4 plastic retainersFixed by 4 plastic retainers•Spare partSpare part•If Air Fan out of service:If Air Fan out of service: Software T° error will be Software T° error will be generatedgenerated Automatic Power SupplyAutomatic Power Supplycut offcut off
VIDAS BLUE
New tray ball bearings:
Excentric ball b
earing
replaced by a normal
one fixed on a sprin
g
mechanism.
No more lateral play to adjustNo more lateral play to adjust
VIDAS BLUE
Tray motor assembly:
Addition of a pretensionning spring Addition of a pretensionning spring
CAUTION!CAUTION!
Lock the motor Assy
Lock the motor Assy
when it iswhen it is
in the good position
in the good position
VIDAS BLUE
Tower screw:
Addition of a spacerAddition of a spacer
To reduce the tower misalignmentsTo reduce the tower misalignments
VIDAS BLUE
Incubator door:
- New door plunger ball nut- New door plunger ball nut
- Addition of a safety latch- Addition of a safety latch
VIDAS BLUE
New electronic boards:
Scanner BoardScanner Board
Incubator BoardsIncubator Boards
VIDAS BLUE
New Scanner Board:
Flash ROM U5 ( V2.51)Flash ROM U5 ( V2.51)(replace the EEPROM)(replace the EEPROM)
DIP Switch SW3DIP Switch SW3(replace JP7/JP8/JP9) (replace JP7/JP8/JP9)
Power Supply LEDsPower Supply LEDs(+5V/+10V/+12V/-(+5V/+10V/+12V/-12V)12V)
New Power Supply 3.3VNew Power Supply 3.3VLEDLED
Sensors LEDsSensors LEDs
Scanner cable Scanner cable breakdown detectionbreakdown detectionLEDsLEDs
Communication LEDsCommunication LEDs
Mini Vidas Blue BacklightMini Vidas Blue BacklightPower Supply connectorPower Supply connector
J15: Scanner Cable new J15: Scanner Cable new Connector for failure Connector for failure detectiondetection(replace coax. cables ) (replace coax. cables )
VIDAS BLUE
New Scanner Board:
•100 % compatible with the old boards100 % compatible with the old boards
•One Flash ROM One Flash ROM V 2.51V 2.51 instead of 3 EPROM V2.41 instead of 3 EPROM V2.41
•SW3 replaces Jumpers JP7 to JP9SW3 replaces Jumpers JP7 to JP9
•New power supply 3,3VNew power supply 3,3V
•New connector (J18) dedicated to the mini New connector (J18) dedicated to the mini Vidas screen backlight power supply.Vidas screen backlight power supply.
•New connector (J15) for scanner failure New connector (J15) for scanner failure detection ( replaces coax cables) detection ( replaces coax cables)
REF. SCANNER BOARDS : VD8201-1REF. SCANNER BOARDS : VD8201-1
VIDAS BLUE
VIDAS BLUE
VIDAS BLUE
New Incubator Board:New Incubator Board:Double support Flash ROM U5Double support Flash ROM U5or EPROM V 3.10or EPROM V 3.10
DIP switch SW2DIP switch SW2(replace adress/(replace adress/reset jumpers)reset jumpers)
Power Supply LEDsPower Supply LEDs
Heaters LEDsHeaters LEDs
Serial communicationSerial communicationwith scanner boardwith scanner boardLEDsLEDs
Sensor LEDsSensor LEDs
VIDAS BLUE
New Incubator Board:New Incubator Board:
• Available with EPROM 3.10 Available with EPROM 3.10 oror Flash ROM Flash ROM (same version)(same version)
• Dip Switch SW2 replace all Adress and Dip Switch SW2 replace all Adress and Reset jumpers ( Reset jumpers ( silkscreensilkscreen directly on the directly on the board).board).• LEDS for Power Supply, communication, LEDS for Power Supply, communication, sensors and heaters.sensors and heaters.
• Fuses (SMD) F1 to F4Fuses (SMD) F1 to F4
REF. INCUB.Board : VD8202-2REF. INCUB.Board : VD8202-2
• 100% compatible100% compatible with the old boards with the old boards
VIDAS BLUE
VIDAS BLUE
VIDAS BLUE
SCANNERSCANNER: new parts
• New mini VIDAS/ VIDAS scanner cablesNew mini VIDAS/ VIDAS scanner cables
New scanner cables with failure detectionNew scanner cables with failure detection(replace coaxial cables)(replace coaxial cables)
The blue flat cable is still the The blue flat cable is still the samesame
VIDAS BLUE
SCANNERSCANNER: new parts• New Scanner PRE-AMP boardNew Scanner PRE-AMP board
• J1 connector unchangedJ1 connector unchanged, use with , use with both old scanner câble kits and newboth old scanner câble kits and newScanner cables.Scanner cables.
• New J2 connectorNew J2 connector, is only used with , is only used with new scanner cables and new scannernew scanner cables and new scannerBoard.Board.(Allows scanner cable failure detection)(Allows scanner cable failure detection)
VIDAS BLUE : reminder
The scanner cable failure detection system is available onlyThe scanner cable failure detection system is available only with a new Scanner Cable, new Scanner Board, with a new Scanner Cable, new Scanner Board, and new Scanner Head.and new Scanner Head.
The LEDs (DS9 to DS12 from the new scanner board) are onThe LEDs (DS9 to DS12 from the new scanner board) are onWhen a cable is broken.When a cable is broken.( ( off = cables OKoff = cables OK))
If a scanner cable failure is detected, a blocking error 161 If a scanner cable failure is detected, a blocking error 161 will be generated.will be generated.
The new Scanner head is compatible with the old Scanner headThe new Scanner head is compatible with the old Scanner headBut But don’t forget to replace the new scanner cable by an old don’t forget to replace the new scanner cable by an old scanner cable kit in order to be compatible with the old scanner scanner cable kit in order to be compatible with the old scanner boardboard
VIDAS BLUE
Changes impact
Decontamination: Decontamination: a new procedure has been defined to take a new procedure has been defined to take Into account the « Strip preparation tray ».Into account the « Strip preparation tray ».
Error 161: Error 161: If err. 161, check the integrity of the scanner If err. 161, check the integrity of the scanner cables. ( LEDs off on scanner board = good)cables. ( LEDs off on scanner board = good)
PM : Scanner cable replacement every 2 yearsPM : Scanner cable replacement every 2 years Non ApplicableNon Applicable
PM : PM : tray lateral play not necessary to set tray lateral play not necessary to set
PM : Power Supply PM : Power Supply not necessary to open it during a PM not necessary to open it during a PM(If dead, replace the complete block.)(If dead, replace the complete block.)
VIDAS BLUE
Compatibility table
(*) Scanner Head Blue : put an old scanner cable kit(*) Scanner Head Blue : put an old scanner cable kit
(**) Scanner Cable Blue: only used with a new (**) Scanner Cable Blue: only used with a new Scanner Head and new Scanner Board.Scanner Head and new Scanner Board.
VIDAS BLUE
You will receive a specific Mandatory Action Required (MAR)You will receive a specific Mandatory Action Required (MAR) from GCS soon.from GCS soon.
Feel free to contact us for many information :Feel free to contact us for many information :
E-mail : E-mail : [email protected]
More details will be added in this MAR .More details will be added in this MAR .