Vicat device callibration.xlsx

download Vicat device callibration.xlsx

of 5

Transcript of Vicat device callibration.xlsx

  • 7/28/2019 Vicat device callibration.xlsx

    1/5

    ID*:

    ID*:

    ID*:

    s

    30 1 15 60 120 180 240

    30 1 15 60 120 181 240

    0 0 0 0 0 1 0

    OK OK OK OK OK OK OK

    Table 2. Needle inspection: d1 d2 d3

    1.11 1.11 1.11

    max

    min

    OK OK OK

    min

    299.2 297.1 298.3

    max

    min

    OK OK OK OK

    295

    305

    Measured weight (g)

    Required (g)

    Range test:

    Table 4. Needle inspection:

    Table 5. Needle inspection:

    Yes, needle is in longitudinal center.

    No, needle is bended.

    Test passed: OK

    YES

    Test pass

    Test passed

    Date of calibration:

    Instrument name:

    ID:

    VICAMATIC, Automatic setting time device

    Referent norm:

    Final report:

    Table 1. Timer inspection:

    Table 3. Needle inspection:

    Needle diameter

    (mm):

    Weight (needle +

    holder)(g):

    Needle position (central longitudinal) visual

    description

    Required (mm)

    Range test:

    OK

    1.09

    1.21

    Test pas

    OK

    Calibration tools:

    Timer device:

    Length device:

    Weight device:

    EN196-3

    APPROVED

    Units:

    Time (Vicat):

    Time device:

    Difference:

    Range test:

    Measured diameter (mm)

    OK

    Needle length

    (45mm)Required (mm)

    Measured length (mm) 47

    45

    Range test:

    Test pass

    OK

    minute

    R011/1CALIBRATION REPORT, VICAT DEVICE

    1

  • 7/28/2019 Vicat device callibration.xlsx

    2/5

    L1 L2 L3

    69.0 70.0 40.0

    max 65 65 39.5

    min 85 85 40.5

    OK OK OK

    YES

    NO

    YES

    YES

    Table 6. Mold inspection:

    Table 7. Software inspection:

    Table 8. Software inspection:

    Table 9. Base plate dimension:

    Table 10 Visual inspection:

    YESYes, all base plates are same thickens

    No, If not, remove all which are different OK

    All base plates used for setting time

    test has to be same thickness

    Test

    passed:

    Test

    passed:

    Test

    passed:

    Software parameters

    Check in the last version of LI001

    Needle ZERO position, auto

    calibration

    OK

    Final Conclusion:

    Signature:

    Team for calibration:

    Yes, parameters are set good

    No, parameters are not set good

    Device is clean

    Any visible damage (if YES, give description in annex)

    Does the room condition suitable for this device

    Electricity connector is proper and safe

    OK

    OK

    Required

    (mm)

    Yes, needle ZERO position is set

    No, needle ZERO position is not set OK

    YES

    Range test:

    YES

    1)

    2)

    Approved by:

    R011/1

    Dimension (mm)

    Mold

    Dimension

    (mm)

    Test pass

    If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this

    document)

    This device is properly calibrated and is approved to use in laboratory testing.

    APPROVED

    2

  • 7/28/2019 Vicat device callibration.xlsx

    3/5

    ID*:

    ID*:

    ID*:

    s

    30 1 15 60 120 180 240

    -- -- -- -- -- -- --

    -- -- -- -- -- -- --

    Table 2. Needle inspection: d1 d2 d3

    max

    min

    -- -- --

    min

    max

    min

    -- -- --

    Instrument name:

    Calibration tools:

    Timer device:

    ID:

    Referent norm: EN196-3

    Date of calibration:

    Weight device:

    Length device:

    Time device:

    Difference:

    Table 1. Timer inspection:

    Units:

    Test passedTime (Vicat):

    Range test: --

    Needle diameter

    (mm):

    Measured diameter (mm)

    Test pasRequired (mm)

    1.09

    Range test: -- --

    1.21

    Range test: --

    Table 3. Needle inspection:

    Needle length

    (45mm)

    Measured length (mm)

    minute

    Range test: --

    Weight (needle +

    holder)(g):

    Measured weight (g)

    Test pass

    Required (g)295

    305

    Test passed: --

    --

    VICAMATIC, Automatic setting time device

    Final report:

    R011/1CALIBRATION REPORT, VICAT DEVICE

    Table 5. Needle inspection:

    Table 4. Needle inspection:

    Needle position (central longitudinal) visual

    description

    Yes, needle is in longitudinal center.

    No, needle is bended.

    Test passRequired (mm) 45

    3

  • 7/28/2019 Vicat device callibration.xlsx

    4/5

    L1 L2 L3

    max 65 65 39.5

    min 85 85 40.5

    -- -- --

    Table 6. Mold inspection: R011/1

    Mold

    Dimension

    (mm)

    Dimension (mm)

    Test pass

    Range test: OK

    Table 7. Software inspection: Test

    passed:

    Required

    (mm)

    --

    Table 8. Software inspection: Test

    passed:Software parameters

    Check in the last version of LI001

    Yes, parameters are set good

    Needle ZERO position, auto

    calibration

    Yes, needle ZERO position is set

    No, needle ZERO position is not set

    Yes, all base plates are same thickens

    No, If not, remove all which are different --

    Table 10 Visual inspection:

    No, parameters are not set good --

    Table 9. Base plate dimension: Test

    passed:All base plates used for setting time

    test has to be same thickness

    Device is clean

    Any visible damage (if YES, give description in annex)

    Does the room condition suitable for this device

    --

    This device is properly calibrated and is approved to use in laboratory testing.

    If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this

    document)

    Electricity connector is proper and safe

    Final Conclusion: --------

    Approved by:

    Signature:

    2)

    Team for calibration:

    1)

    4

  • 7/28/2019 Vicat device callibration.xlsx

    5/5

    VICAMATIC, Automatic setting time deviceR011/1

    Calibration description (EN 196-3)

    Table 3

    L = 45mm

    Table 5

    WeightNeedle and holder

    together

    295-305(g)

    Ta

    ble2

    d3

    d2

    d1

    L3 = 40.0 2mm

    Table 6

    L1 = 65-85mm

    L2 = 65-85mm

    Use L key

    for needlereplacement onlyUse L key

    for needle

    ZERO calibration

    Glass base plate

    Table 7

    Table 9

    OKNOT

    Table 4

    5