Vicat device callibration.xlsx
-
Upload
ratkovic-predrag -
Category
Documents
-
view
251 -
download
0
Transcript of Vicat device callibration.xlsx
-
7/28/2019 Vicat device callibration.xlsx
1/5
ID*:
ID*:
ID*:
s
30 1 15 60 120 180 240
30 1 15 60 120 181 240
0 0 0 0 0 1 0
OK OK OK OK OK OK OK
Table 2. Needle inspection: d1 d2 d3
1.11 1.11 1.11
max
min
OK OK OK
min
299.2 297.1 298.3
max
min
OK OK OK OK
295
305
Measured weight (g)
Required (g)
Range test:
Table 4. Needle inspection:
Table 5. Needle inspection:
Yes, needle is in longitudinal center.
No, needle is bended.
Test passed: OK
YES
Test pass
Test passed
Date of calibration:
Instrument name:
ID:
VICAMATIC, Automatic setting time device
Referent norm:
Final report:
Table 1. Timer inspection:
Table 3. Needle inspection:
Needle diameter
(mm):
Weight (needle +
holder)(g):
Needle position (central longitudinal) visual
description
Required (mm)
Range test:
OK
1.09
1.21
Test pas
OK
Calibration tools:
Timer device:
Length device:
Weight device:
EN196-3
APPROVED
Units:
Time (Vicat):
Time device:
Difference:
Range test:
Measured diameter (mm)
OK
Needle length
(45mm)Required (mm)
Measured length (mm) 47
45
Range test:
Test pass
OK
minute
R011/1CALIBRATION REPORT, VICAT DEVICE
1
-
7/28/2019 Vicat device callibration.xlsx
2/5
L1 L2 L3
69.0 70.0 40.0
max 65 65 39.5
min 85 85 40.5
OK OK OK
YES
NO
YES
YES
Table 6. Mold inspection:
Table 7. Software inspection:
Table 8. Software inspection:
Table 9. Base plate dimension:
Table 10 Visual inspection:
YESYes, all base plates are same thickens
No, If not, remove all which are different OK
All base plates used for setting time
test has to be same thickness
Test
passed:
Test
passed:
Test
passed:
Software parameters
Check in the last version of LI001
Needle ZERO position, auto
calibration
OK
Final Conclusion:
Signature:
Team for calibration:
Yes, parameters are set good
No, parameters are not set good
Device is clean
Any visible damage (if YES, give description in annex)
Does the room condition suitable for this device
Electricity connector is proper and safe
OK
OK
Required
(mm)
Yes, needle ZERO position is set
No, needle ZERO position is not set OK
YES
Range test:
YES
1)
2)
Approved by:
R011/1
Dimension (mm)
Mold
Dimension
(mm)
Test pass
If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this
document)
This device is properly calibrated and is approved to use in laboratory testing.
APPROVED
2
-
7/28/2019 Vicat device callibration.xlsx
3/5
ID*:
ID*:
ID*:
s
30 1 15 60 120 180 240
-- -- -- -- -- -- --
-- -- -- -- -- -- --
Table 2. Needle inspection: d1 d2 d3
max
min
-- -- --
min
max
min
-- -- --
Instrument name:
Calibration tools:
Timer device:
ID:
Referent norm: EN196-3
Date of calibration:
Weight device:
Length device:
Time device:
Difference:
Table 1. Timer inspection:
Units:
Test passedTime (Vicat):
Range test: --
Needle diameter
(mm):
Measured diameter (mm)
Test pasRequired (mm)
1.09
Range test: -- --
1.21
Range test: --
Table 3. Needle inspection:
Needle length
(45mm)
Measured length (mm)
minute
Range test: --
Weight (needle +
holder)(g):
Measured weight (g)
Test pass
Required (g)295
305
Test passed: --
--
VICAMATIC, Automatic setting time device
Final report:
R011/1CALIBRATION REPORT, VICAT DEVICE
Table 5. Needle inspection:
Table 4. Needle inspection:
Needle position (central longitudinal) visual
description
Yes, needle is in longitudinal center.
No, needle is bended.
Test passRequired (mm) 45
3
-
7/28/2019 Vicat device callibration.xlsx
4/5
L1 L2 L3
max 65 65 39.5
min 85 85 40.5
-- -- --
Table 6. Mold inspection: R011/1
Mold
Dimension
(mm)
Dimension (mm)
Test pass
Range test: OK
Table 7. Software inspection: Test
passed:
Required
(mm)
--
Table 8. Software inspection: Test
passed:Software parameters
Check in the last version of LI001
Yes, parameters are set good
Needle ZERO position, auto
calibration
Yes, needle ZERO position is set
No, needle ZERO position is not set
Yes, all base plates are same thickens
No, If not, remove all which are different --
Table 10 Visual inspection:
No, parameters are not set good --
Table 9. Base plate dimension: Test
passed:All base plates used for setting time
test has to be same thickness
Device is clean
Any visible damage (if YES, give description in annex)
Does the room condition suitable for this device
--
This device is properly calibrated and is approved to use in laboratory testing.
If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this
document)
Electricity connector is proper and safe
Final Conclusion: --------
Approved by:
Signature:
2)
Team for calibration:
1)
4
-
7/28/2019 Vicat device callibration.xlsx
5/5
VICAMATIC, Automatic setting time deviceR011/1
Calibration description (EN 196-3)
Table 3
L = 45mm
Table 5
WeightNeedle and holder
together
295-305(g)
Ta
ble2
d3
d2
d1
L3 = 40.0 2mm
Table 6
L1 = 65-85mm
L2 = 65-85mm
Use L key
for needlereplacement onlyUse L key
for needle
ZERO calibration
Glass base plate
Table 7
Table 9
OKNOT
Table 4
5