Vet ABC Troubleshooting. CHANGING THE PACK Service menu Change Pack Remove the old pack etc....

15
Vet ABC Vet ABC Troubleshooting Troubleshooting

Transcript of Vet ABC Troubleshooting. CHANGING THE PACK Service menu Change Pack Remove the old pack etc....

Vet ABC TroubleshootingVet ABC Troubleshooting

CHANGING THE PACKCHANGING THE PACK

• Service menu Change Pack Remove the old pack etc. (on-screen instructions)

• Remove the Red caps from the new pack• Install new pack ensuring arrows have disappeared• Prime the new pack into an open vessel• Re-install waste pipe onto new pipe• Write on the pack the installation date

Operational ProblemsOperational Problems

• Basic understanding of how the ABC works • Function of the main components• How to spot potential problems• How to carry out minor repairs

““STARTUP FAILED CHECK STARTUP FAILED CHECK REAGENTS”REAGENTS”

  Blank values high: PLT high, pack can be contaminated, reagent too cold

   Perform concentrated cleaning or change pack

  Electromagnetic noise (X-rays)    Isolate Vet ABC and run start-up

““STARTUP FAILED CHECK STARTUP FAILED CHECK REAGENTS”REAGENTS”

  If other parameters are OK, photometer blank failed    Check photometer (see below)

  Pack was shaken while in transport (bubbles)    After Transport by a carrier, wait 1-2 hours before

installation

  Tube on valve, needle, chamber or dispenser block is loose   Check tube connections and repair if possible

““SENSOR DEFECT OR REAGENT EMPTY”SENSOR DEFECT OR REAGENT EMPTY”

• Pack wasn’t installed properly• Re-seat the pack ensuring arrows have disappeared

• Valve on base of pack has collapsed inwards• Adjust the position of the valve

• Pack is empty• Change the pack

““SENSOR DEFECT OR REAGENT EMPTY”SENSOR DEFECT OR REAGENT EMPTY”

• Pack is distended with excess air• “Burp” the pack

• Red caps were not removed• Remove the Red caps

• Waste tube is blocked or bent• Check tube is free of obstructions

““ERROR: MOTOR PRESSURE SYRINGE”ERROR: MOTOR PRESSURE SYRINGE”• Pack is distended with excess air• “Burp” the pack

• Piston is pushed out of chamber by high pressure• Re-install the piston

• Piston is sticking• Check movement – Service menu Mechanic

Pressure Syringe.

““ERROR: LIQUID SYRINGE MOTOR” ERROR: LIQUID SYRINGE MOTOR”

• Piston or gear is damaged• Check movement – Service menu Mechanic Liquid

Syringe

• Dilution Block is damaged• Check for leaks – Call Woodley Equipment.

““START-UP FAILED HGB REFERENCE FAILURE”START-UP FAILED HGB REFERENCE FAILURE”

• WBC Chamber is contaminated• Run concentrated clean

• Reagent is contaminated• Change pack

““START-UP FAILED HGB REFERENCE FAILURE”START-UP FAILED HGB REFERENCE FAILURE”

• No reagent is available• Prime the pack – check that reagent appears in

chambers

• Photometer is out of adjustment• Run concentrated clean; Check photometer

adjustment – Service menu Technician Adjustments

IF IN ANY DOUBT CALL IF IN ANY DOUBT CALL WOODLEY!WOODLEY!