Vascular & Interventional Radiology UnitDepartment of Radiological Sciences“Sapienza” – University of Rome

Fabrizio Fanelli

NEW TECHNOLOGIES: Wires, balloons, drug-eluting devices, ect.

Page 2: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Small vessels

Long lesions

Occlusions > stenoses

More vessels

CalcificationsIliac Femoral Popliteal Peroneal Post Tib Ant Tib

0

100

200

300

400

500

600

700

Stenosis

Occlusion

L. Graziani – Eur J Vasc Endovasc Surg 2007

Below The Knee Arteries

Page 3: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

55% FEM-POP

96% BTK

5% ATG

CLI : Mean Length Treated lesions

185±121 mm

116±113 mm

63±61 mm

R.Ferraresi - oral presentation CICE 2012

Page 4: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

PTA F.U. PATENCY TLR

DEBATE 1 12 mos 28.0% 24.0%

Leipzig Registry2 3 mos 31.0 % 50.0%

DEBELLUM 3 12 mos 47.1% 47.0%

BMS F.U. PATENCY TLR

ACHILLES4 12 mos 57.1% 16.0%

DESTINY5 12 mos 54.3 % 33.6%

YUKON6 12 mos 55.6% 13.0%

3) F. Fanelli - oral presentation CIRSE 2012 2) A. Schmidt – JACC 2011

1) F. Liistro - oral presentation TCT 2012

6) A. Rastan – European Heart Journal 20115) M. Bosiers – J Vasc Surg 2012

4) K. Katsanos - oral presentation CIRSE 2011

Page 5: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

…. New Technologies ….

Page 6: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Micro-puncture set

Guidewires

Micro-guidewires

Introducers

Support catheters

Low profile balloon catheters

Atherectomy devices

Drug eluting balloons (DEB)

Bare metal stents

Drug eluting stents (DES)

Page 7: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Micro-puncture set

Guidewires

Micro-guidewires

Introducers

Support catheters

Low profile balloon catheters

Atherectomy devices

Drug eluting balloons (DEB)

Bare metal stents

Drug eluting stents (DES)

Page 8: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Drug Eluting Balloons

In.Pact Amphirion – Medtronic /Invatec

Advance LP 18 PTX – Cook

Freeway II/Dior – Eurocor

Elutax – Aachen Medical

Page 9: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Study type Prospective single center, single arm, investigator initiated study

Objective Assess IN.PACT Amphirion™ efficacy for the treatment of long BTK lesions occlusions

Population Symptomatic patients with CLI or severe claudication

Eligibility At least one lesion BTK ≥ 80 mm

Prim. Endpoint 3 month restenosis rate

Nr of patients 104 / 109 limbs

IN.PACT™ Amphirion inreal world BTK complex lesions

A. Schmidt – JACC 2011

DEB BTK Leipzig Registry

Page 10: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

DEB(angio

subgroup)

PTA(historical

group)

# patients / limbs 74 / 79 58 / 62

Male gender 51 (68.9%) 38 (65.5%)

mean age (y) 73.5 ± 9.3 70.5 ± 8.08

diabetics 54 (73%) 52 (89.7%)

Renal insuff. 34 (45.9%) 30 (51.7%)

RC 3 16 (20.3%) 0 (0%)

RC 4 14 (17.7%) 16 (25.8%)

RC 5 49 (62%) 46 (74.2%)

RC 6 0 (0%) 0 (0%)

avg lesion length 173 ± 87 mm 183 ± 75 mm

Tot occlusions 61.9% 64.9%

DEB(angio subgroup)

PTA(historical group)

3m Ang. FU

Restenosis (>50%) 27.4% 69%

Full-segment Resten. 10% 56%

Restenosis Length 64 mm 155 mm

12m Clinical FU

15m Clinical FU

Deaths 16.3% 10.5%

Limb Salvage 95.6% 100%

Clinical Improvement (1) 91.2% 76.5%

Compl. wound healing 74.2% 78.6%

TLR 17.3% 50%

(1) clinical improvement = reduction in size and/or depth of ulceration or improvement of rest-pain

vs historical PTA cohort (A.Schmidt - Cath. and Cardiovasc. Interventions

2010)

DEB BTK Leipzig Registry

Page 11: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

DEB(angio

subgroup)

PTA(historical

group)

# patients / limbs 74 / 79 58 / 62

Male gender 51 (68.9%) 38 (65.5%)

mean age (y) 73.5 ± 9.3 70.5 ± 8.08

diabetics 54 (73%) 52 (89.7%)

Renal insuff. 34 (45.9%) 30 (51.7%)

RC 3 16 (20.3%) 0 (0%)

RC 4 14 (17.7%) 16 (25.8%)

RC 5 49 (62%) 46 (74.2%)

RC 6 0 (0%) 0 (0%)

avg lesion length 173 ± 87 mm 183 ± 75 mm

Tot occlusions 61.9% 64.9%

DEB(angio subgroup)

PTA(historical group)

3m Ang. FU

Restenosis (>50%) 27.4% 69%

Full-segment Resten. 10% 56%

Restenosis Length 64 mm 155 mm

12m Clinical FU

15m Clinical FU

Deaths 16.3% 10.5%

Limb Salvage 95.6% 100%

Clinical Improvement (1) 91.2% 76.5%

Compl. wound healing 74.2% 78.6%

TLR 17.3% 50%

(1) clinical improvement = reduction in size and/or depth of ulceration or improvement of rest-pain

vs historical PTA cohort (A.Schmidt - Cath. and Cardiovasc. Interventions

2010)

DEB BTK Leipzig Registry

Page 12: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Investigator initiated, single center Study

Randomized (1:1) DEB vs. standard PTA

Primary Endpoint: 12-month Angiographic

Binary (>50%) Restenosis Rate

Secondary Endpoints: TLR and Reocclusion

CLI + Diabetes150 (Tibial) Lesions

DEB(75 lesions)

Std PTA(75 lesions)

12-month Angiographic

and Clinical follow-up

Aspirin + Clopidogrel (1 month)

24-month Duplexand Clinical follow-up

Random (1:1)

• Key Inclusions• RC 4-5-6 and Diabetes• Stenosis / occlusions >40 mm in at

least 1 tibial vessel with distal run-off• Key Exclusions

• Allergy to Paclitaxel• Contraindication for combined

antiplatelet treatment• Life expectancy <1 year

DEBATE BTK

F. Liistro - oral presentation TCT 2012

Page 13: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

DEB PTA p

Patients 65 67 /

Age 74±9.4 75±9.6 0.7

Smoke 13 (20%) 7 (10.4%) 0.1

Hypertension 46 (70.8%) 52 (77.6%) 0.4

Hypercholesterolemia 23 (35.4%) 16 (23.9%) 0.1

Diabetes 65 (100%) 67 (100%) 1

Obesity 23 (35.3%) 26 (38.8%) 0.5

Dialysis 7 (10.8%) 7 (10.4%) 1

Coronary artery disease 12 (18.5%) 10 (14.9%) 0.6

Limbs 71 72 0.9

Texas wound ulcer class. I C 1 (1.4%) 3 (4.2%)

0.5

II C 3 (4.2%) 5 (6.9%)

I D 7 (9.9%) 9 (12.5%)

II D 34 (47.9%) 32 (44.4%)

III D 26 (36.6%) 23 (31.9%)

Page 14: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,
Page 15: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

• Key Inclusions– Fontaine Stage IIb, III, IV – SFA, popliteal and BTK lesions

either isolated or concomitant– De-novo lesions – Stenosis or occlusions 3 - 30 cm

Investigator initiated, single center Study

Randomized (1:1) DEB vs. standard PTA

Primary Endpoint: 6-month LLL

Secondary Endpoints: TLR, Primary Patency, Fontaine change, MAE at 6 and 12 months

DEBELLUM Drug Eluting Balloon Evaluation for Lower Limb mUltilevel treatMent

F. Fanelli - oral presentation CIRSE 2012

Page 16: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

analysis per lesion

12-mos TLR

Page 17: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

DEB PTA DEB PTA

Angio 6 mos Angio 6 mos Angio 12 mos Angio 12 mos

Restenosis rate 15.3% (2 lesions)

41.1% (7 lesions)

23.07% (3 lesions)

52.9% (9 lesions)

TLR 7.69% (1 lesion)

23.5% (4 lesions)

15.3% (2 lesions)

47.0% (8 lesions)

Occlusion rate 0 9.1% 0 9.1%

Death 0 0 0 0

Major amputation 10% (1 patient)

9.09% (1 patient)

10% (1 patient)

9.09% (1 patient)

Minor amputation 0 18.8% (2 patients)

0 18.8% (2 patients)

DEBELLUM - 12 mos F.U.

Page 18: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

• Prospective Randomized Multicenter Trial

• 50 pts.• RC: 4-6• >50% calcified stenosis• popliteal, tibial and peroneal aa.

Page 19: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

OA + PTA PTA p

Stenting 6.9 % 14.3 % 0.44

Dissection 6 pts 1 pts ns

Perforation / 1 ns

Distal Embolization / 1 ns

Max Inflation Press <0.001

2 - 4 59.3 % 11.8 %

5 - 6 14.8 % 8.8 %

7 - 8 7.4 % 26.5 %

9 - 10 7.4 % 32.4 %

11 - 12 0 8.8 %

>12 11.1 % 11.8 %

Mean 5.9 ± 4.2 9.4 ± 3.8

Page 20: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

OA + PTA PTA p

Stenting 6.9 % 14.3 % 0.44

Dissection 6 pts 1 pts ns

Perforation / 1 ns

Distal Embolization / 1 ns

Max Inflation Press <0.001

2 - 4 59.3 % 11.8 %

5 - 6 14.8 % 8.8 %

7 - 8 7.4 % 26.5 %

9 - 10 7.4 % 32.4 %

11 - 12 0 8.8 %

>12 11.1 % 11.8 %

Mean 5.9 ± 4.2 9.4 ± 3.8

Page 21: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Secondary Endpoints

Page 22: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

PTA F.U. PATENCY TLR

DEBATE 1 12 mos 28.0% 24.0%

Leipzig Registry2 3 mos 31.0 % 50.0%

DEBELLUM 3 12 mos 47.1% 47.0%

DEB F.U. PATENCY TLR

DEBATE 1 12 mos 71.0% 13.0%

Leipzig Registry2 3 mos 72.6 % 17.3%

DEBELLUM 3 12 mos 76.9% 15.3%

3) F. Fanelli - oral presentation CIRSE 2012 2) A. Schmidt – JACC 2011

1) F. Liistro - oral presentation TCT 2012

Page 23: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

What about Stent … ?

Page 24: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

DEB F.U. PATENCY TLR

DEBATE 1 12 mos 71.0% 13.0%

Leipzig Registry2 3 mos 72.6 % 17.3%

DEBELLUM 3 12 mos 76.9% 15.3%

BMS F.U. PATENCY TLR

ACHILLES4 12 mos 57.1% 16.0%

DESTINY5 12 mos 54.3 % 33.6%

YUKON6 12 mos 55.6% 13.0%

3) F. Fanelli - oral presentation CIRSE 2012 2) A. Schmidt – JACC 2011

1) F. Liistro - oral presentation TCT 2012

6) A. Rastan – European Heart Journal 20115) M. Bosiers – J Vasc Surg 2012

4) K. Katsanos - oral presentation CIRSE 2011

Page 25: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

DEB F.U. PATENCY TLR

DEBATE 1 12 mos 71.0% 13.0%

Leipzig Registry2 3 mos 72.6 % 17.3%

DEBELLUM 3 12 mos 76.9% 15.3%

DES F.U. PATENCY TLR

ACHILLES4 12 mos 75.0% 10.0%

DESTINY5 12 mos 85.2 % 8.7%

YUKON6 12 mos 80.6% 13.8%

3) F. Fanelli - oral presentation CIRSE 2012 2) A. Schmidt – JACC 2011

1) F. Liistro - oral presentation TCT 2012

6) A. Rastan – European Heart Journal 20115) M. Bosiers – J Vasc Surg 2012

4) K. Katsanos - oral presentation CIRSE 2011

Page 26: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Everolimus-eluting stent vs. bare metal stent

103 pts. CLI

Angioplasty + Bailout stenting

JVIR 2009; 20(9):1141-1150

Page 27: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

D. Siablis – JVIR 2009; 20(9):1141-1150

HR: 4.8, CI 2.9-7.9 p<0.001 HR: 2.5, CI 1.3-5.0 p=0.006

Page 28: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,
Page 29: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Xcience-Prime Xcience-Prime

Page 30: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Conclusions

DEB patency rate similar to DES but with the advantage

of

“nothing left behind”.

Efficacy of DEB in very calcified arteries still under

evaluation

Combined treatment with atherectomy must be

validated.

Page 31: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

The Future …

• Adaption to flexion, extension

•Overcome Stent fractures

•Positive Remodelling

•Might act as a transporter (Gen-

Therapy)

•Surgeon’s friend

Bioabsorbable Stent

Page 32: Vascular & Interventional Radiology Unit Department of Radiological Sciences “Sapienza” – University of Rome Fabrizio Fanelli NEW TECHNOLOGIES: Wires,

Poly-L-lactide +

Everolimus

30 pts.

ABSORB Study

5-y F.U.

- No cardiac death

- No thrombosis

- MACE: 3.4 %

- LLL 0.27 mm