VAPORIZER Lakshmanraj

198
VAPORIZER Dr.K.M.LAKSHMANARAJAN . MD.

Transcript of VAPORIZER Lakshmanraj

Page 1: VAPORIZER Lakshmanraj

VAPORIZER

Dr.K.M.LAKSHMANARAJAN . MD.

Page 2: VAPORIZER Lakshmanraj

REFERENCES

• DORSCH AND DORSCH• WARD’S• MILLER

Page 3: VAPORIZER Lakshmanraj

• WHAT IS VAPOR ? • Vapor is the gaseous phase of a substance which is

normally a liquid at room temp. and atm. pressure. Eg. water vapor is the vapor form of water

• Gas is substance which exists only in the gaseous state at room temp. and amt. pressure.

3

Page 4: VAPORIZER Lakshmanraj

WHAT IS A VAPORIZER ?

• A vaporizer is an instrument designed to facilitate the change of a liquid anaesthetic agent into a vapor

And

Add a controlled amount of this vapor to the FGF• The SVP of most inhalation agents is MUCH more that

is required to produce anesthesia i.e. 32% vs 0.75 or 243 mm Hg vs 5.7 mm Hg for halothane

• Need to dilute this vapor with the carrier gas and deliver a controlled amount of this vapor to the patient

Page 5: VAPORIZER Lakshmanraj

DISCUSSION

• EVALUATION OF VAPORIZER• PHYSICS APPLIED TO VAPORIZER• HOW IT WORKS?• CLASSIFICATION• INDIVIDUAL VAPORIZER• HAZARDS & SAFETY FEATURES • IDEAL VAPORIZER

5

Page 6: VAPORIZER Lakshmanraj

EVOLUTION….EVOLUTION….

Wilson and Pinson's Ether

"Bomb

Schimmelbusch's Inhaler

Foregger Model of Morris' Copper Kettle

Vaporizer

EMO VAPORIZER

MORTON INHALER

Page 7: VAPORIZER Lakshmanraj

7

Page 8: VAPORIZER Lakshmanraj

FLAGG’S CAN

8

Page 9: VAPORIZER Lakshmanraj

EVOLUTION OF VAPORIZERS

• MORTONS ETHER INHALER

1846, OCT 16

Page 10: VAPORIZER Lakshmanraj

SNOW – MARCH 1847

10

Page 11: VAPORIZER Lakshmanraj

CLOVER PORTABLE REGULATING ETHER INHALER 1877

Page 12: VAPORIZER Lakshmanraj

VERNON HARCOURT CHLOROFORM INHALER 1903

Page 13: VAPORIZER Lakshmanraj

OMBREDANNE ETHER INHALER 1908

Page 14: VAPORIZER Lakshmanraj

14

Page 15: VAPORIZER Lakshmanraj

SOMNOFORM INHALER 1908

• Mixture of ethyl chloride, methyl chloride & ethyl bromide

Free Template from www.brainybetty.com 15

Page 16: VAPORIZER Lakshmanraj

OPEN DROP MASK

Page 17: VAPORIZER Lakshmanraj

OPEN DROP MASK

Page 18: VAPORIZER Lakshmanraj

OGSTON INHALER

Page 19: VAPORIZER Lakshmanraj

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 20: VAPORIZER Lakshmanraj

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 21: VAPORIZER Lakshmanraj

GOLDMAN VAPORIZER

Page 22: VAPORIZER Lakshmanraj

ROWBOTHOM VAPORIZER

• Modification of Goldman vaporizer

Page 23: VAPORIZER Lakshmanraj

POCKET TRILENE INHALER

23

Page 24: VAPORIZER Lakshmanraj

PHYSICSAPPLIED TO VAPORIZER

24

Page 25: VAPORIZER Lakshmanraj

WHAT IS CRITICAL TEMPERATURE ?

• For any substance there is a max. temp. at which it can be compressed so as to convert it from a gas to a liquid. This is known as the critical temp. and above this temp. no amount of compression will liquefy it.

• Under this condition the substance is a gas. • Below that critical temp it is a vapor.

Page 26: VAPORIZER Lakshmanraj

• A gas that is currently below its critical

temperature is called a “vapour”.• If compressed with enough pressure, it will condense into a

liquid.

26

Page 27: VAPORIZER Lakshmanraj

Critical temperature of Critical temperature of

• Oxygen –118Oxygen –11800CC

• Nitrogen –147Nitrogen –147ooCC

• Air –141Air –141ooC C

• Carbon dioxide 31Carbon dioxide 31ooCC

• Nitrous oxide 36.4Nitrous oxide 36.4ooCC

• Entonox - 5 to -7 (psuedocrit-temp)Entonox - 5 to -7 (psuedocrit-temp)

Page 28: VAPORIZER Lakshmanraj

ISOFLURANE ON VENUS

• The critical temperature of Isoflurane is about 200 C

• The temperature on the planet Venus is about 500 C

So if you lived on Venus, Isoflurane would exist as a gas.

• On earth, since The temperature where you live is hopefully below the critical temperature of Isoflurane, it exists as a vapour / liquid.

28

Page 29: VAPORIZER Lakshmanraj

VP & SVP

• Vapor exerts a pressure on its surroundings – Vapor pressure .

• Vapor pressure depends only on the temp. and liquid.

• Vapor pressure of an agent determines how much of vapor will be formed from 1 ml of the liquid.

• Since different anaesthetic agents have different vapor pressures the need separate vaporizers.

Page 30: VAPORIZER Lakshmanraj

SVP ?

Page 31: VAPORIZER Lakshmanraj

• For a particular liquid at a particular temp. there occurs an equilibrium at which the number of molecules leaving the liquid equals the number reentering

• It is the maximum VP at a particular temp. Increases rapidly as boiling point approaches.

Page 32: VAPORIZER Lakshmanraj

HOW DO YOU DEFINE BOILING POINT…?HOW DO YOU DEFINE BOILING POINT…?

-Is the temperature at which the vapor pressure reaches the atmp - decrease the atmp – dec the BP -increase the atmp – incre the BP

IF THE B.P OF ETHER IS 36.5 C , WILL IT BE BOILING IF THE B.P OF ETHER IS 36.5 C , WILL IT BE BOILING ALWAYS IN THE ROOM TEMP IN THE SUMMER IN ALWAYS IN THE ROOM TEMP IN THE SUMMER IN MADURAI WHEN THE TEMP IS 40 C…?MADURAI WHEN THE TEMP IS 40 C…?

• SVP MAY BE > ATMP SVP MAY BE > ATMP • ATMP MAY BE > SVP ---- its not BP ATMP MAY BE > SVP ---- its not BP •To get boiling point SVP to reach ATMP To get boiling point SVP to reach ATMP

Page 33: VAPORIZER Lakshmanraj

BOILING POINT ?

Page 34: VAPORIZER Lakshmanraj

S.NOS.NO LIQUID ANESTHETICLIQUID ANESTHETIC B.P in CB.P in C

11 ETHYL CHLORIDEETHYL CHLORIDE 12.412.4

22 DESFLURANEDESFLURANE 22.822.8

33 ISOFLURANEISOFLURANE 48.548.5

44 HALTHANEHALTHANE 50.250.2

55 ENFLURANEENFLURANE 56.556.5

77 SEVOFLURANESEVOFLURANE 58.658.6

88 ETHER ETHER 34.634.6

99 METHOXY FLURANEMETHOXY FLURANE 105105

BOILING POINT BOILING POINT

Page 35: VAPORIZER Lakshmanraj

GAS CONCENTRATION:GAS CONCENTRATION:

Partial pressure:Partial pressure:

•Molecular agitationMolecular agitation

•Bombardment on the walls of containerBombardment on the walls of container

•Creation of a pressureCreation of a pressure

•Dalton’s law of partial pressure: Dalton’s law of partial pressure: total pressure of the mixture is the sum of thetotal pressure of the mixture is the sum of thepartial pressures of the constituent gasespartial pressures of the constituent gases

•Volume percent: partial pressure/ total pressure x 100Volume percent: partial pressure/ total pressure x 100

Page 36: VAPORIZER Lakshmanraj

HEAT of VAPORIZATION

• In a liquid, molecules are in a state of continuous motion because of mutual attraction by Van-der Waal’s forces.

• Some molecules may develop velocities sufficient to escape from these forces and may enter the vapour phase.

• The amount of heat required to convert a unit mass of liquid into a vapour without a change in temperature of the liquid is termed

the heat of vaporization.

36

Page 37: VAPORIZER Lakshmanraj

HEAT OF VAPORISATION:HEAT OF VAPORISATION:

•No of calories necessary to convert 1 G of liquid into vapourNo of calories necessary to convert 1 G of liquid into vapour

•Effect of heat loss on continuous vaporisationEffect of heat loss on continuous vaporisation

•Gradient of temp. between liquid and surroundingGradient of temp. between liquid and surrounding

•Ways of Supplied heatWays of Supplied heat

Page 38: VAPORIZER Lakshmanraj

SPECIFIC HEAT CAPACITYSPECIFIC HEAT CAPACITY

Quantitiy of heat required to raise the temp of 1 G of the Substance by 1 C

Material used to construct the vaporiser should have High sp.heat capacity

THERMAL CONDUCTIVITY:THERMAL CONDUCTIVITY:

Measure of the speed with which heat flows throughMeasure of the speed with which heat flows througha substancea substanceCu > Al > brass > steel >> glass

Page 39: VAPORIZER Lakshmanraj

39

Agent

Boiling Point (°C,

760 mm Hg)

Vapor Pressure

(torr, 20°C)

Density of Liquid (g/mL)

Heat of Vaporization Specific Heat of Liquid

MACa in O2 (%)cal/g cal/mL cal/mL cal/g

Halothane

50.2 243 1.86 (20°C)

35 (20°C) 65 (20°C) 0.35 0.19 0.75

Enflurane 56.5 175 1.517 (25°C)

42 (25°C) 63 (25°C)     1.68

Isoflurane 48.5 238 1.496 (25°C)

41 (25°C) 62 (25°C)     1.15

Desflurane

22.8 669 1.45 (20°C)

        6.4

Sevoflurane

58.6 157           2.0

Page 40: VAPORIZER Lakshmanraj

VOLATILE AGENTSBoiling point

SVP at 20ºc MAC BG coefficient

Desflurane 22.8º c 669 mmhg 6.4 0.45

Ether 34.6º c 425 mmhg 1.92

Isoflurane 48.5º c 175 mmhg 1.15 1.4

Halothane 50.2º c 243 mmhg 0.75 2.5

Enflurane 56.5º c 175 mmhg 1.68 1.8

Sevoflurane 58.5º c 157 mmhg 2.0 0.65

Methoxyflurane 105 º c 20 mmhg 0.2 12-15

Nitrous oxide 21.2º c 4000 mmhg 104 0.47

Xenon -108.1ºc 60-70 0.115

Page 41: VAPORIZER Lakshmanraj

HOW IT WORKS ?

41

Page 42: VAPORIZER Lakshmanraj

VAPORIZATION

42

Page 43: VAPORIZER Lakshmanraj

VAPORIZATION

43

Page 44: VAPORIZER Lakshmanraj

VAPORIZATION AT HIGH FLOW

• If one uses a high fresh gas flow, the vaporisation process can't keep up with so much gas arriving into the vaporisation chamber.

• The result is that, relative to the high flow of fresh gas flow, the amount of anaesthetic vaporised is inadequate.

• So,, the vaporiser delivers less anaesthetic concentration than is set on the dial

44

Page 45: VAPORIZER Lakshmanraj

VAPORIZER IN HIGH FLOW

45

Page 46: VAPORIZER Lakshmanraj

VAPORIZATION AT HIGH FLOW

• The solution employed by modern vaporisers to solve this problem is to increase the surface area of contact between the fresh gas and anesthetic agent.

• The output concentration remains accurate to the setting on the dial over a wide range of flows.

• One method that vaporisers use to increase the efficiency of vaporisation is to dip wicks into the anaesthetic agent

46

Page 47: VAPORIZER Lakshmanraj

WICKS

47

Page 48: VAPORIZER Lakshmanraj

BUBBLE THROUGH

48

Page 49: VAPORIZER Lakshmanraj

BUBBLE THROUGH

49

Page 50: VAPORIZER Lakshmanraj

SPECIFIC HEAT

50

Page 51: VAPORIZER Lakshmanraj

51

Page 52: VAPORIZER Lakshmanraj

This property is used in the design of temperature compensating valves in vaporizers.

52

Page 53: VAPORIZER Lakshmanraj

THERMAL STRIP

53

Page 54: VAPORIZER Lakshmanraj

54

Page 55: VAPORIZER Lakshmanraj

Another method uses a "bi metallic" strip. Different metals expand and contract to differing extents when exposed to temperature changes

55

Page 56: VAPORIZER Lakshmanraj

56

Page 57: VAPORIZER Lakshmanraj

EFFECT OF TEMPERATURE ON THE EFFECT OF TEMPERATURE ON THE PERFORMANCE OF VAPORISERSPERFORMANCE OF VAPORISERS

•As liquid is vaporised, energy is lost in the form of As liquid is vaporised, energy is lost in the form of heatheat

•As the temp. decreases, so does the vapour pressureAs the temp. decreases, so does the vapour pressure

• vaporisation decreases, so does the vaporiser outputvaporisation decreases, so does the vaporiser output

Page 58: VAPORIZER Lakshmanraj

Effect of heat loss on the performance of Effect of heat loss on the performance of Boyle’s bottle:Boyle’s bottle:

Page 59: VAPORIZER Lakshmanraj

METHODS OFMETHODS OF TEMPERATURE COMPENSATION:TEMPERATURE COMPENSATION:

1.1. Supplied heatSupplied heatLarge mass of copper – high specific heat and Large mass of copper – high specific heat and thermal conductivitythermal conductivitywater jacketwater jacketelectric heaterelectric heater

2. Flow compensation2. Flow compensationIncrease the % of carrier gas flow that is directedIncrease the % of carrier gas flow that is directedthrough the vaporizing chamberthrough the vaporizing chamber

Page 60: VAPORIZER Lakshmanraj

EFFECT OF EFFECT OF INTERMITTENT BACK PRESSURE:INTERMITTENT BACK PRESSURE:

Positive pressure from breathing system, O2 flushPositive pressure from breathing system, O2 flush

1.PUMPING EFFECT1.PUMPING EFFECT

2. PRESSURISING EFFECT2. PRESSURISING EFFECT

Page 61: VAPORIZER Lakshmanraj

PUMPING EFFECT:PUMPING EFFECT:

Increased pressure fluctuation - increased gas flow Increased pressure fluctuation - increased gas flow into vaporising chamber – picking up more vapour – into vaporising chamber – picking up more vapour – which enters into the bypass chamber – diluent which enters into the bypass chamber – diluent containing vapour – increased vaporiser outputcontaining vapour – increased vaporiser output

Page 62: VAPORIZER Lakshmanraj

Pumping effect

• Common in plenum ( high resistant vaporizers)• Seen in low flow anesthesia, low dial settings, low

level of liquid agent• Seen in when high RR used, high PIP, rapid drops

in airway pressure during expiration • When use O2 flush also ( during intraop)

Page 63: VAPORIZER Lakshmanraj

Pumping effect (Think traffic signal, starting from signal )

During IPPV – gas flow to pt & to back bar also

Pressure rises in the back bar

Pressure transmitted to bypass channel (only FGF coming ) & to vaporizer chamber

Page 64: VAPORIZER Lakshmanraj

Vapor molecules compressed inside the vapor chamber

Compressed FGF again will diverted more into vaporizer

Now SVP raised

More molecules of concentrated vapor produced

Page 65: VAPORIZER Lakshmanraj

On expiration – pressure drops at common gas outlet

High concentrated vapor than the dial setting delivered to the pt

Pumping effect

Page 66: VAPORIZER Lakshmanraj

PUMPING EFFECT:PUMPING EFFECT:

Page 67: VAPORIZER Lakshmanraj

67

Page 68: VAPORIZER Lakshmanraj

Modifications to minimise the pumping effect:Modifications to minimise the pumping effect:

1. Increasing the size of bypass chamber

2. Decreasing the size of the vaporising chamber

3. Long inlet tube

4. Avoiding wicks in the inlet area

5. Increase resistance to gas flow through vaporiser

6. Unidirectional / pressurising valve at the outlet of the vaporiser

Page 69: VAPORIZER Lakshmanraj

69

Page 70: VAPORIZER Lakshmanraj

70

Page 71: VAPORIZER Lakshmanraj

71

Page 72: VAPORIZER Lakshmanraj

72

Page 73: VAPORIZER Lakshmanraj

PRESSURISING EFFECT:PRESSURISING EFFECT:

Will vaporisation increase as the pressure in the vaporisingchamber increase?

SVP depends on temperature , not on pressure

This effect leads to a decrease in the vaporiser output

Page 74: VAPORIZER Lakshmanraj

Pressurizing effect• According to the second concept

• Common in high flow anesthesia

• Common in VIC vaporizer• SVP of vapor not increased• Less common effect and less danger than

pumping effect

Page 75: VAPORIZER Lakshmanraj

Ippv – back flow/pressure transmitted back to vaporizer / bypass

Vaporized molecules are compressed without altering SVP

Same no molecules but diluted (density reduced)

Vapor concentration decreased & delivered on expiration

Pressurizing effect

Page 76: VAPORIZER Lakshmanraj

PRESSURISING EFFECT:PRESSURISING EFFECT:

Page 77: VAPORIZER Lakshmanraj

Which effect is important…?Which effect is important…?

•Pumping effect is more dangerous…

•Pressurising effect occurs in high gas flows

•Pumping effect in low flows..

Page 78: VAPORIZER Lakshmanraj

EFFECT OF ALTERED BAROMETRIC EFFECT OF ALTERED BAROMETRIC PRESSURE PRESSURE ON THE ON THE PERFORMANCE OF VAPORISERS:PERFORMANCE OF VAPORISERS:

•Vaporisers in high altitude and hyperbaric chambers

•Vaporiser output concentration differs

•Partial pressure remains the same, so does depth of anesthesia

C’ = C x P/ P’

Page 79: VAPORIZER Lakshmanraj

PLACING THE VAPORISER PLACING THE VAPORISER IN THE BREATHING CIRCUITIN THE BREATHING CIRCUIT

1.1. VAPORISER OUTSIDE THE CIRCLE ( VOC )VAPORISER OUTSIDE THE CIRCLE ( VOC )

2.2. VAPORISER INSIDE THE CIRCLE ( VIC )VAPORISER INSIDE THE CIRCLE ( VIC )

Page 80: VAPORIZER Lakshmanraj

VOCVOC

IN THE BACK BAR….

•Between the flow meter and common gas outletBetween the flow meter and common gas outlet

•Or between common gas outlet and breathing systemOr between common gas outlet and breathing system

•Vaporisers in a row..Vaporisers in a row..

•Selectatec mechanismSelectatec mechanism

•Difficult to equalise the vaporiser concentration withDifficult to equalise the vaporiser concentration with alveolar concentrationalveolar concentration

Page 81: VAPORIZER Lakshmanraj

VOC – VAPORISER IN A ROW IN THE BACK BARVOC – VAPORISER IN A ROW IN THE BACK BAR

Two consideratons:Two consideratons:

•Parallel / serials Parallel / serials •Cross contamination should be minimumCross contamination should be minimum•If it occurs, it should be least dangerousIf it occurs, it should be least dangerous

With agents equal potency the agent with the lower With agents equal potency the agent with the lower vapor pressurevapor pressureshould be upstream…should be upstream…

With agents of equal vapour pressure, the less With agents of equal vapour pressure, the less potent should be upstreampotent should be upstream

Explosive agents, trilene should be downstream..Explosive agents, trilene should be downstream..

Page 82: VAPORIZER Lakshmanraj

VAPORIZER ARRANGEMENT

• In the order of SVP and potency( less SVP , less potent –first)

• Thymol added vaporizer – in the downstream

Page 83: VAPORIZER Lakshmanraj

ORDER OF VAPORIZERS

• Less potent - upstream• More potent - downstream• If equipotent

Low VP - upstreamHigh VP - downstream

If explosive - downstreamTrilene - downstreamEasy to clean - downstream

Page 84: VAPORIZER Lakshmanraj

UP STREAM → → → DOWNSTREAM

SEVOFLURANE

ENFLURANE ISOFLURANE

HALOTHANE

DESFLURANE

VP- 157 175 238 243 669

Page 85: VAPORIZER Lakshmanraj

VAPORISERS IN A VAPORISERS IN A ROW…ROW…

Page 86: VAPORIZER Lakshmanraj

VAPORISER INSIDE THE CIRCLE:VAPORISER INSIDE THE CIRCLE:

Low resistance vaporiser incorporated in the circle systemLow resistance vaporiser incorporated in the circle system•Economical

•Faster induction

•Vaporiser output and concentration are not same

•Easy to achieve alveolar concentration similar to vaporiser concentration

•Dangerous concentration may build up•2 positions:

Expiratory side, Inspiratory limb

Page 87: VAPORIZER Lakshmanraj

VIC - IN THE EXPIRATORY LIMB:VIC - IN THE EXPIRATORY LIMB:

•SafeSafe•Moisture condensationMoisture condensation•Wastage through relief valveWastage through relief valve

VIC - IN THE INSPIRATORY LIMB:VIC - IN THE INSPIRATORY LIMB:

•Faster inductionFaster induction•Danger of liquid anesthetic aspirationDanger of liquid anesthetic aspiration

Page 88: VAPORIZER Lakshmanraj

VAPORIZER

• Vapour output depends on

Temperature of liquid

Volatility

Surface area of liquid

FGF rate

Ippv

Carrier gas composition

Barometric temperature ( eg – high altitude)

Page 89: VAPORIZER Lakshmanraj

FACTORS AFFECTING VAPORIZER OUTPUT

• 1. FLOW THROUGH THE VAPORIZING CHAMBER: Varying the proportion of gas passing through the vaporizing chamber and bypass chamber

• 2. SURFACE AREA OF THE LIQUID GAS INTERFACE: Greater the surface area, more will be the vaporization.

• Bubble through > flow over

Page 90: VAPORIZER Lakshmanraj

• 3. TEMPERATURE: as temperature increase, output increase

• 4. TIME: Output concentration tend to fall over time

• 5. GAS FLOW RATE: (A) At high flowrates, the gas leaving vaporization chamber is less saturated

(B) Alters the total flow that passes through the vaporization chamber

Page 91: VAPORIZER Lakshmanraj

• 6. CARRIER GAS COMPOSITION:

(a) Changes in viscosity & density may affect the proportion of the total flow passing through the vaporization chamber

(b) N2O dissolves in the flow, thus altering the effective volume passing through the vaporization chamber

• 7. BOILING POINT: Higher the boiling point, less will be the vapor output

Page 92: VAPORIZER Lakshmanraj

• 8. AMBIENT PRESSURE:

- SVP is solely a function of temp. so if ambient pressure is reduced, the constant SVP becomes a greater proportion of the total pressure

-> output increases.

- Agents with low boiling points are more susceptible to the influence of ambient pressure

Page 93: VAPORIZER Lakshmanraj

EFFECT OF LOW ATMOSPHERIC PRESSURE

CONCENTRATION CALIBRATED • High resistance pathway through the vaporizing

chamber offers less resistance, under hypobaric conditions and so a slight increase in vapor output occurs.

• Deliver higher conc. if measured in vol. % but deliver same PP so clinicall effect unchanged

CP =CP or C=CP/P At 0.5 atm. C=Cx1/.5=2%

Page 94: VAPORIZER Lakshmanraj

EFFECT OF LOW ATMOSPHERIC PRESSURE

MEASURED FLOW• Here the delivered partial pressure & volume

percent increases.• Amount of increase depends on the barometric

pressure & the vapor pressure of the agent.• Closer the vapor pressure is to barometric

pressure, greater the effect.

Page 95: VAPORIZER Lakshmanraj

CONCENTRATION CALIBRATED• ATM PRESSURE INCREASES -> Density of gas

CHANGES -> More resistance to flow of gas through the vaporizing chamber

-> Decreased vapor O/P (Partial Pressure & Volume Percent)

Effect on partial pressure is less dramatic

EFFECT OF HIGH ATMOSPHERIC PRESSURE

Page 96: VAPORIZER Lakshmanraj

MEASURED FLOW

• Lower concentration in terms of PP/Volume percent

EFFECT OF HIGH ATMOSPHERIC PRESSURE

Page 97: VAPORIZER Lakshmanraj

1988 ASTM MACHINE STANDARDS FOR VAPORIZERS

1. The effects of variations in ambient temperature and pressure, tilting, back pressure, and input flow rate and gas mixture composition on vaporizer performance must be stated in the accompanying documents.

2. The average delivered concentration from the vaporizer shall not deviated from the set value by more than ± 20% or ± 5% of the maximum setting whichever is greater without back pressure.

Page 98: VAPORIZER Lakshmanraj

3. The average delivered concentration from the vaporizer shall not deviate from the set value by more that + 30% or – 20% or by more than + 7.5% or – 5% of the max. setting whichever is greater, with pressure fluctuations at the common gas outlet of 2 kPa with a total gas flow of 2 L/minute or 5 kPa with a total gas flow of 8 L/minute

4. A system that prevents gas from passing through the vaporizing chamber or reservoir of one vaporizer and the through that of another must be provided

5. The output of the vaporizer shall be less than 0.05% in the “OFF” or “zero” position if the zero position is also the “OFF” position.

Page 99: VAPORIZER Lakshmanraj

6. All vaporizer control knobs must open counterclockwise

7. Either the max. and min. filling levels or the actual usable volume and capacity shall be displayed

8. The vaporizer must be designed so that it cannot be overfilled when in the normal operating position.

9. Vaporizer unsuitable for use in the breathing system must have noninterchangeable proprietary or 23 mm fittings. Conical fittings of 15 mm and 22 mm cannot be used. When 23 mm fittings are used the inlet must be male and the outlet female. The direction of gas flow must be marked.

Page 100: VAPORIZER Lakshmanraj

10. Vaporizers suitable for use in the breathing system must have standard 22-mm fittings or screwthreaded, weight-bearing fittings with the inlet female and the outlet male. The direction of gas flow must be indicated by arrows and the vaporizer marked “for use in the breathing system.

Page 101: VAPORIZER Lakshmanraj

CLASSIFICATION

1-Regulation of output concentration

A. measured outflow

B. Concentrations calibrated

2-Method of vaporization

A. Flow over ( over surface of liquid)

B. Bubble thro ( ether bottle)

C. Draw over ( cyprane inhaler,EMO )

D. Injector ( TEC6)

3- Location

A. VOC ( in back bar)

B. VIC ( in closed circuit )

Page 102: VAPORIZER Lakshmanraj

4.Plenum or not

A occupied / closed room, in which the pressure of the air is greater than that of the outside atmosphere

Here the air is forced in to the chamber

Eg – ether , copper kettle, tec

5.Compensations

A. Temperature (bimetallic strip,heater)

B. Back pressure (tec 3- 7)

C. Tilt

D. Ambient pressure ( high altitude )

6. Agent specific or not

7.Electronically controlled or not ( TEC 6, aladin casette)

8. Measured output flow rate compensation ( TEC 6 plus)

Page 103: VAPORIZER Lakshmanraj

• Plenum types – have high resistance

• Draw over – low resistance

Page 104: VAPORIZER Lakshmanraj

VARIABLE BYPASS VAPORISERSVARIABLE BYPASS VAPORISERS

Vaporiser has to dilute the saturated vapor as vapourpressure of most volatile anesthetics are much greaterthan the partial pressure required to produce anesthesia

Accomplished by splitting the gas flow that passes through the vaporiser

Some of the gas flows through the vaporising chamberand remainder gas through a bypass to the vaporiser outlet

Both flows join downstream to give the desired concentration

Splitting ratio

Page 105: VAPORIZER Lakshmanraj

Flowmeter

Patient

VARIABLE BYPASS

Page 106: VAPORIZER Lakshmanraj

Flowmeter

Patient

VARIABLE BYPASS

Page 107: VAPORIZER Lakshmanraj

FLOW OVER

• Carrier gas flows OVER liquid picking up vapor.• Efficiency improved by increasing area that carrier

gas flows over gas-liquid interface. ie. Baffles or wicks

Page 108: VAPORIZER Lakshmanraj

FLOW OVER VAPORISERSFLOW OVER VAPORISERS

Performance depends onPerformance depends on

1.1. Vapour pressure of the liquidVapour pressure of the liquid

2.2. Efficiency of vaporisationEfficiency of vaporisation

3.3. Area of the gas-liquid interfaceArea of the gas-liquid interface

4.4. Velocity of carrier gas flowVelocity of carrier gas flow

5.5. Height of gas flow above the liquidHeight of gas flow above the liquid

Page 109: VAPORIZER Lakshmanraj

BUBBLE THROUGH VAPORISERS:BUBBLE THROUGH VAPORISERS:

Performance depends on:Performance depends on:

•Size of the bubblesSize of the bubbles

•Depth of the liquidDepth of the liquid

•Velocity of gas flowVelocity of gas flow

Page 110: VAPORIZER Lakshmanraj

Principle of the draw-over vaporizer

DRAW-OVER VAPORIZER

Page 111: VAPORIZER Lakshmanraj

PLENUM VAPORIZER

Page 112: VAPORIZER Lakshmanraj

INDIVIDUAL VAPORISERSINDIVIDUAL VAPORISERS

BOYLE’S ETHER BOTTLE:BOYLE’S ETHER BOTTLE:

Variable bypass – flow over or bubble through – out of systemVariable bypass – flow over or bubble through – out of system-No temperature compensationNo temperature compensation

-U inlet tube ,plunger with hood ( cowl )U inlet tube ,plunger with hood ( cowl )

-Cork with an attached chainCork with an attached chain

Page 113: VAPORIZER Lakshmanraj

BOYLES BOTTLE

• Parts: (1) vaporizing bottle 300 mL (2) Metal top incorporating controls (3) Lever, plunger which is chrome plated (4) Stopper & Retaining chain

• Concentration calibrated• Flowover or bubble through• Not temperature compensated• Multiple agents• Vaporizer outside circuit

Page 114: VAPORIZER Lakshmanraj

BOYLE BOTTLE

Page 115: VAPORIZER Lakshmanraj

BOYLE’S ETHER BOTTLE:BOYLE’S ETHER BOTTLE:

Page 116: VAPORIZER Lakshmanraj

EVALUATION OF ETHER BOTTLE:EVALUATION OF ETHER BOTTLE:

Factors influencing the vaporiser output:Factors influencing the vaporiser output:

•Temperature of the liquid in the vaporiserTemperature of the liquid in the vaporiser

•Plunger levelPlunger level

•Position of the control leverPosition of the control lever

•Volume flow of the gas through the vaporiserVolume flow of the gas through the vaporiser

•Level of liquid in the vaporiserLevel of liquid in the vaporiser

•Agitation of the vaporiserAgitation of the vaporiser

Page 117: VAPORIZER Lakshmanraj

BOYLE’S BOTTLE:BOYLE’S BOTTLE:

ETHER BOTTLE:ETHER BOTTLE:

Vaporising chamber is largeCopper inlet tube and hood

HALOTHANE BOTTLE:HALOTHANE BOTTLE:No hood or plungerChromium plated inlet tubeInlet tube is plugged at the tip and a side hole 1 cm aboveControl lever scale is largerVaporising chamber is smaller than ether’s

Page 118: VAPORIZER Lakshmanraj

OXFORD MINIATURE VAPORIZER (OMV)

• Variables bypass• Flowover with steel wicks• Temperature compensated• For use by multiple agents• Vaporizer outside circuit• Halothane trilene methoxyflurane ether isoflurane

Page 119: VAPORIZER Lakshmanraj

OMV

119

Page 120: VAPORIZER Lakshmanraj

MEASURED FLOW (COPPER KETTLE & VERNITROL)

• Also known as bubble through or Vernitrol. • Flow meter-measured. Manually calculated

bypassed carrier flow.• Temperature compensated by construction

materials with high specific heat and thermal conductivity to offset cooling from vaporization induced heat loss. ie. Copper

• Amount of carrier gas (CG) O2 bubbled through is determined by a dedicated Thorpe tube flowmeter

• A valve separates the vaporizer circuit from the standard O2 & N2O flowmeters

Page 121: VAPORIZER Lakshmanraj
Page 122: VAPORIZER Lakshmanraj

METAMATIC VAPORIZER

122

Page 123: VAPORIZER Lakshmanraj

FOREGGER COPPER KETTLE 1952

Page 124: VAPORIZER Lakshmanraj

FOREGGER COPPER KETTLE 1952

Page 125: VAPORIZER Lakshmanraj

GOLDMAN VAPORISERGOLDMAN VAPORISER

Variable bypass – flow over without wick – in or out of system – No temperature compensation

Designed for intermittent flow machine

When vaporising chamber is full, 2% halothane concentrationcan be achieved at a lever position of 3.

VIC – concentration may reach 3.5 % - 4.5%

Problem of the preservative - Thymol

Page 126: VAPORIZER Lakshmanraj

GOLDMAN VAPORIZER

Page 127: VAPORIZER Lakshmanraj

• Young’s modification of goldman

• Placing wicks

• 3—4%

127

Page 128: VAPORIZER Lakshmanraj

EPSTEIN-MACINTOSH-OXFORD VAPORISEREPSTEIN-MACINTOSH-OXFORD VAPORISER

Variable bypass – flow over with wick – in system – temperatureCompensation by supplied heat and flow alteration

Designed for field condition

Can be used for both spontaneous and controlled ventilation

A ventilating bellows has to be added in the circuit forcontrolled ventilation

Page 129: VAPORIZER Lakshmanraj

EMO vaporizer 1953• Variable bypass• Incomplete vapourizatio • Low resistance• Agent-specific for ether or may not be• Temperature compensated by bellows, • Temperature stabilized by water jacket, • Transportable but heavy (10kg). .

Page 130: VAPORIZER Lakshmanraj

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 131: VAPORIZER Lakshmanraj

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 132: VAPORIZER Lakshmanraj

E.M.O VAPORISER

Page 133: VAPORIZER Lakshmanraj

FLUTEC

• All TEC

Variable bypass

Plenum

Temperature compensated

Concentrations calibrated

Flow compensated – by using wicks ( except TEC 2 )

Page 134: VAPORIZER Lakshmanraj

• TEC 2- 4 – not agent specific• Backpressure (pumping effect) compensation starts

from TEC 3• Anti spill proof starts from TEC 4• Agent specific starts from TEC 4• Selectatec starts from TEC 4• Color coding starts from TEC 5 ( but our TEC 3

having color coding)• Tilt compensation from TEC 6• Barometric pressure compensation TEC 7

But TEC 2 used only for HALOTHANE

Page 135: VAPORIZER Lakshmanraj

• Bimetallic strip(temp compensating device )

may be in inside the vaporizer chamber

or in bypass channel

• TEC 5 PUMPING EFFECT> TEC 4• INTERNAL BAFFLES TEC5

Page 136: VAPORIZER Lakshmanraj

FLUTEC Mk-IIIFLUTEC Mk-III

Variable bypass – flow over with wick – out of system – Variable bypass – flow over with wick – out of system – Temp. compensation by flow alteration – Temp. compensation by flow alteration – agent specific ( Halothane )agent specific ( Halothane )

Bypass is located concentrically within the vaporising chamberFlow alteration is by a bimettalic stripVolume of vaporising chamber is 270 ml, wick retains 35 mlMaximum 5%Accuracy is good up to 3% with all flows

Page 137: VAPORIZER Lakshmanraj

FLUTEC Mk - IIIFLUTEC Mk - III

Page 138: VAPORIZER Lakshmanraj

TEC – 5 VAPORISERSTEC – 5 VAPORISERS

•Meant for isoflurane, sevofluraneMeant for isoflurane, sevoflurane

•Improvised to check the pumping effectImprovised to check the pumping effect

•Accuracy is assured in clinically used flow ratesAccuracy is assured in clinically used flow rates

•Liquid anesthetic can’t enter the bypass channelLiquid anesthetic can’t enter the bypass channel

•Fitted with coded filling systemFitted with coded filling system

Page 139: VAPORIZER Lakshmanraj

TEC Mk 5 VAPORISERS:TEC Mk 5 VAPORISERS:

Page 140: VAPORIZER Lakshmanraj

TEC Mk 5 VAPORISER

Page 141: VAPORIZER Lakshmanraj

TEC 6- why • Desflurane BP 22.8 ºc• So at room temperature – will be in vapor or as gas form ( u

wont able to see even drops of desflurane)• So we may not know how much it is delivered to the patient

if u give it by scimmelbusch mask like ether• Suppose if u use desflurane by other Tec - high output will

be delivered • Suppose if u are reducing desflurane output by diluting its

vapor molecules in other Tec – u need to keep FGF of 73 L/min to have 1% vapor

Page 142: VAPORIZER Lakshmanraj

• If u supply desflurane in ordinary Tec – all desflurane vapoured in room tem

• This leads to excessive cooling of vaporizer-lead on to low output

• So Desflurane needs special (controlled vaporization) vaporizer

Page 143: VAPORIZER Lakshmanraj

TEC 6

• Parts

1. It has 2 independent gas circuits in parallel

1.FGF inlet circuit – passes thro R1(fixed restrictor)

2.Vapor circuit – passes thro R2 (variable restrictor, ie the dial setting)

2.Differential pressure transducers

3.Control electronics

4.Pressure regulating valve

5.Desflurane resorvoir(heating sump)

6.Shut off valve

Page 144: VAPORIZER Lakshmanraj

TEC 6

Page 145: VAPORIZER Lakshmanraj

145

Page 146: VAPORIZER Lakshmanraj

146

Page 147: VAPORIZER Lakshmanraj

147

Page 148: VAPORIZER Lakshmanraj

148

Page 149: VAPORIZER Lakshmanraj

149

Page 150: VAPORIZER Lakshmanraj

TEC 6

Page 151: VAPORIZER Lakshmanraj

Tec 6 – how it works

Sump (electrically heated) is heated up 39ºc

Vapor pressure in sump – reaches 2atm absolute

Shutt of valve opened ( dial setting on)

Pass thro R2

R2 reduces pressure to 1.1atm absolute

Page 152: VAPORIZER Lakshmanraj

After R2 vapor joins with FGF

When FGF rate – constant

Back pressure to the diaphragm of differential pressure transducer(dpt)

DPT conveys pressure diff between fgf circuit & vapor circuit

Now control electronics – act on pressure regulative valve

Page 153: VAPORIZER Lakshmanraj

Pressure in vapor circuit = pressure in FGF circuit

Now this is the working pressure of vaporizer

• If FGF rate constant – working pressure – constant• If u increase the FGF rate – working pressure is also

increased• But change in FGF rate or dial setting – will not

change vapor pressure ( vapor output is constant)

Page 154: VAPORIZER Lakshmanraj

Special operating features of TEC 6

• Once vaporizer – connected to power supply – the vaporizer starts to heat the sump automatically

• No separate switch to on heaters

• Warm up time – 5-10 mins• Front panel has 5 LED lights

Page 155: VAPORIZER Lakshmanraj

Tec 6 LEDs – top to bottom

• Operational LED (ready to use) - green

• No output LED - red

• Low agent LED(refilling is required) – amber

• Warm up LED - amber

• Alarm Low battery LED – amber

Page 156: VAPORIZER Lakshmanraj

Amber LED

• Amber color LED - Emits light during

Warm up

Agent <50ml

Muted the alarm for >120 secs

Low battery

If press the mute button continuosly – vaporizer will go for self testing

Page 157: VAPORIZER Lakshmanraj

RED LED• If fluid level < 20ml• Tilting >105 º ( > 15 deg from vertical axis)• Power failure• Internal malfunction

There is 10 seconds delay from the time detected malfunction to time alarm activated

Page 158: VAPORIZER Lakshmanraj

TEC 6 - LCD

• Battery – 9 volt comes for 1 year• LCD bars (20 small bars, instead of fluid column

in others)- receives signal from sump• If heaters switched off or Low agent – LCD bars

also will drop• Upper most LCD bar = 390 ml in sump• Middle bar = 240 ml • Lowest = only 60 ml

Page 159: VAPORIZER Lakshmanraj

TEC 6 • Beginning of warm up – all LED s flash for 1 sec with

alarm• Tec 6 tilt proff - when tilt > 105 º - shutts of vaporizer• Up to 90 º u can use the vaporizer• Filling speciality

SAF –T-FIT – specific filler nozzle in desflurane bottle

Nozzle is attached with filler port in vaporizer

then turned upward and invert the bottle

Page 160: VAPORIZER Lakshmanraj

• Now the vaporiser is filled

• If u fill the vaporiser when LCD bar showing < 240 ml ( below middle bar)- port will accept the whole bottle

• Once bottle emptied – bottle turned to starting (upward) position

• Now the filler port ejecting nozzle of bottle

• You can fill when vaporizer on flow

• Overfilling prevented

• Dial setting – 1 – 18%

Page 161: VAPORIZER Lakshmanraj

TEC 6 CHECK OUT

• 1-Press and hold the mute button until all lights and alarms activated.• 2-Turn on to at least 1% and unplug the electrical connection. A "No Output"alarm should ring within seconds. This tests battery power for the alarms.

This step is crucial in relation to the quick emergence characteristics of this agentany interruption in its supply must be noted and responded to at once

161

Page 162: VAPORIZER Lakshmanraj

TEC 6

• Tell me one disadvantage of Tec 6 ?

Not barometric pressure compensated so dial settings to be adjusted I tell u why ?

Page 163: VAPORIZER Lakshmanraj

High altitude & vaporizer

• For classical plenum vaporizers,

the percentage output increases roughly in proportion to the fall in barometric pressure,

But a smaller partial pressure increase.• Depth of anesthesia depends on partial pressure• The changes in partial pressure are relatively small, • So normal vaporizer settings work as expected at

altitude

Page 164: VAPORIZER Lakshmanraj

TEC & barometric compensation

• At 20º c SVP of isoflurane 238 mmhg• At high altitude temperature rises to 35º c – SVP rised

to 450mmhg• Increase pressure in vaporizing chamber bend the

bimetallic strip• So the bimetallic strip decrease FGF to the vaporizing

chamber – reduce the vapor output – so same concentration is delivered = same anesthetic potency

Page 165: VAPORIZER Lakshmanraj

High altitude & Tec 6 • The percentage delivered is essentially held constant• so partial pressure FALLS in proportion to the fall in

atmospheric pressure. • The dial setting should be turned up to compensate

• Required dial setting = normal dial setting * 760 mmhg/ ambient pressure (mmhg)

• I mean the vapor pressure of deslurane at outlet to the pt is always constant becoz 2 pressure resistors

Page 166: VAPORIZER Lakshmanraj

SAFETY FEATURESVAPORIZER

166

Page 167: VAPORIZER Lakshmanraj

INTERLOCK SYSTEM

167

Page 168: VAPORIZER Lakshmanraj

INTERLOCK DEVICES

• Selector valve

• Selectatec system

• Back bar devices

• Ohio switch

• Drager lock

Page 169: VAPORIZER Lakshmanraj

169

Page 170: VAPORIZER Lakshmanraj

BACK BAR

• Safety Selectatec’ system

Opening more than 2 vaporizer not possitbe at a time

Page 171: VAPORIZER Lakshmanraj

SELECTA TEC

171

Page 172: VAPORIZER Lakshmanraj

Fillers • Types

Funnel fillers – simple funnel there (most of our vaporisers)

Keyed filler – specific attached to vapor bottle and vaporiser ( not in our institute)

Quick filler – sevoflurane vaporizer is (eg our theatre sevo)

SAF-T-FIT – special for desflurane ( we r having)

Easy filler – Tec 7 ( bottle adapter)

Page 173: VAPORIZER Lakshmanraj

FILLING – FUNNEL

173

Page 174: VAPORIZER Lakshmanraj

KEYED FILLING

174

Page 175: VAPORIZER Lakshmanraj

175

Page 176: VAPORIZER Lakshmanraj

176

Page 177: VAPORIZER Lakshmanraj

177

Page 178: VAPORIZER Lakshmanraj

EASY FILL

178

Page 179: VAPORIZER Lakshmanraj

SAFE –T-FIT

179

Page 180: VAPORIZER Lakshmanraj

1. Difficulty in filling

2. Misalignment of adaptor in filler receptacle

3. Adaptor not sealing at the bottle end

4. Leak in the bottle adaptor

5. Air bubbles.

6. Lost bottle adaptor

7. Failure of keyed system

8. Liquid leaks

AGENT SPECIFIC FILLING SYSTEMS PROBLEMS

Page 181: VAPORIZER Lakshmanraj

9. VAPORIZER TIPPING

If filler receptacle on vaporizer extends beyond the base → it cannot be set upright on a flat surface → this can cause tipping prevented by setting the vaporizer receptacle at the edge of the surface or a ring can be fitted at the base that extends below the projection of the filler reeptacle

10.POOR DRAINAGE

11.BROKEN INNER TUBE

AGENT SPECIFIC FILLING SYSTEMS PROBLEMS

Page 182: VAPORIZER Lakshmanraj

ADU ALADIN CASSETTE

182

Page 183: VAPORIZER Lakshmanraj

ALADIN CASSETTE

• Cassettes – color & magnetic coded • Handle • Filler• 10% fluid only – messages in machine& alarm • Vaporizer oultet flow-controlled by CPU• Cassette temp<18C – fan activated• Over fill protection ( air – deactivate it)• Temperature maintenance- brass shell around the

cassette

183

Page 184: VAPORIZER Lakshmanraj

184

Page 185: VAPORIZER Lakshmanraj

DRAGER VAPORIZER

• VAPOR 19.1• Calibration with air• Over fill protection even the vaporizer in on

position• VAPOR 2000 ( DRAGER)• Air calibration• DRAGER D• For desflurane

185

Page 186: VAPORIZER Lakshmanraj

PENLAN DELTA

• O2 – CALIBRATION

• TEMP COMPENSATION- 1- 2HRS

• PENLON ALPHA

• FOR DES

• FAST WARM UP SLOWER POWER CONSUMPTION

• CALCULATE THE AMOUNT OF DES USED FOR EACH CASE

• CPU

• CAUTERY MAY AFFECT

186

Page 187: VAPORIZER Lakshmanraj

HAZARDS

• Wrong agent• Tipping• Overfilling • Reversed flow• Control dial in wrong position• Leaks• Vapor leak • Cross contamination

187

Page 188: VAPORIZER Lakshmanraj

• TIPPING• Liquid from vaporizing

Chamber→bypass/outlet→high output• Drained before moving• OVERFILLING• In majority, the design of the filling port and agent

specific filling systems prevent this• During filling do not:

Turn the dail ON/Unscrew the bottle adaptor

HAZARDS

Page 189: VAPORIZER Lakshmanraj

• Reversed flow• Inlet male & outlet female• Increased output• Physical damage• Obstruction to fresh gas flow• Interlock malfunction

HAZARDS

Page 190: VAPORIZER Lakshmanraj

VAPOURISER check

• Check for the required vaporizer• Check for the fluid level• Filler caps and draining port should be tight• Check for the correct seating and locking of the a

detachable vaporizer• Ensure power supply if electrically operated.

Page 191: VAPORIZER Lakshmanraj

• VO: Vapor Output (mL/min)

• CG: Carrier Gas Flow (mL/min)

• VP: Vapor Pressure (mmHg)

• BP: Barometric Pressure (760mmHg)

• VAA%: Volatile Anesthetic Agent Concentration

VO = CG x VP VO = Total Gas Flow

BP – VP VAA%

Vapor output

Page 192: VAPORIZER Lakshmanraj

Examples

• Halothane: VP 243 mmHg• CG: 100 mL/min

VO = (100 mL/min)(243mL/min) = 24300 = 47mL/min

(760mmHg – 243mmHg) 517

If 1% of Halothane is desired:

47mL/min = 4700mL/min Total gas flow needed

0.01

Page 193: VAPORIZER Lakshmanraj

HALOTHANE

• If 100mL/min of O2 enters , 147ml of gas will exit (100mL of O2 picked up 47mL of Halothane)

• % Halothane exiting is 32% (243/760=0.32) at 1 atm

• If Only 1% Halothane needed for anesthesia, the 32% needs to be diluted with 4553mL/min

• 4700mL/min – 147mL/min = 4553mL/min

Page 194: VAPORIZER Lakshmanraj

194

Page 195: VAPORIZER Lakshmanraj

Isoflurane • Avagadro law• 1.5% iso 5 litre/min flow• Total /hr = 300l• 1.5% = 300* 1.5% = 4.5l• 184.5 mol wt iso – 22.4l• 4.5 – 37 gms• Sp gravity 1.496• 37/1.496 = 24.7 ml

195

Page 196: VAPORIZER Lakshmanraj

IDEAL VAPORISER CHARACTERISTICS:IDEAL VAPORISER CHARACTERISTICS:

•Performance should not get affected by changes in fresh gas flow, volume of the liquid,ambient temp and pressure, decrease in liquid temp,pressure fluctuation due to mode of respiration

•Low resistance to flow

•Lightweight with small liquid requirement

•Economy and safety in use with minimal servicing requirements

•Corrosion and solvent resistant construction

Page 197: VAPORIZER Lakshmanraj

Are we having such type of a vaporiser…?Are we having such type of a vaporiser…?

Page 198: VAPORIZER Lakshmanraj

Concluding…

Inhalational anesthesia is the most important part of modern balanced anesthesia

To efficiently and safely conduct an inhalational anesthesia,a proper knowledge of a vaporiser is a must….

THANK YOU