SE 655(X) / SE 656(X)Toroidal Conductivity Sensors
User ManualEnglish
Latest Product Information: www.knick.de
3
The SE 655(X) and SE 656(X) toroidal conductivity sensors are particularly suited for applications in the chem-ical industry and for process analysis. Thanks to the large measuring range over six decades and the high chemical resistance of the process-wetted parts made of PEEK or PFA, resp., almost every conceivable applica-tion is possible, such as concentration determination of acids, bases and salt solutions, monitoring of product quality or phase separation of product mixtures. The SE 655X and SE 656X sensors are approved for operation in hazardous locations.
Special conditions for safe use in hazardous areas• The sensors may only be used in liquid media with a conductivity of > 10 nS/cm
(risk of electrostatic charging of the sensor).• The sensors are suitable for use in the following ambient/process temperature ranges:
Temperature class T4: -20 ... +125 °C Temperature class T6: -20 ... +75 °C
• The measuring cable must be protected from electrostatic charging when applied in Ex Zone 0 (1G).• Metallic process connection parts must be mounted at the installation site so that they are electrostatically
conductive (< 1 Mohm).
Safety Information• Application in hazardous locations: Observe the EC-Type-Examination Certificate and the provisions
of the current EN 60079-14 (Electrical installations in hazardous areas) when installing the product.• Manipulations of the product other than described in the user manual are not permitted.• Assembly/dismantling, installation, operation and maintenance shall only be carried out by qualified
personnel as defined by the automation industry in compliance with the applicable regulations and this user manual. Be sure to observe the specified ambient conditions and installation instructions.
• The sensors were developed in compliance with the applicable European guidelines and standards. The models SE 655X and SE 656X are suitable for use in hazardous locations. Compliance with the European Harmonized Standards for use in hazardous locations is confirmed by the EC Declaration of Conformity (see page 12).
• Be sure to follow the instructions given in this user manual for connecting and operating the SE 655(X) and SE 656(X) sensors.
• Observe the specified ambient conditions and temperature ranges to ensure the safe operation of the sensors.
• Take care to maintain the IP 68 protection when installing the sensor (use original gasket, pay particular attention to the cable connections).
• Make sure that the gaskets are securely in place. Otherwise, the assembly is not tight.• Do not dismantle the sensor as long as the process medium is under pressure.• For applications in aggressive media such as highly concentrated acids or bases, use the SE 656(X)
sensor with sealing sets B or C (ZU 0340N or ZU 0342N) made of particularly resistant material (PTFE).
Description
4 Table of Contents
Description ................................................................................................3
Safety Information ...................................................................................3
Table of Contents ......................................................................................4
Specifications ...........................................................................................5Chemical resistance ...................................................................................................6Temperature and pressure resistance .................................................................6Electrical connection .................................................................................................7Identification of the sensors ...................................................................................7
Assembly ...................................................................................................8Assembly, with PTFE washer (for flange protection) ....................................8Assembly, without PTFE washer (flange wetted by process medium) ..8
Dimension Drawings ..............................................................................10
Dimension Drawings of Flange .............................................................11
EC Declaration of Conformity ................................................................12
EC-Type-Examination Certificate ..........................................................13
Accessories ..............................................................................................18
5Specifications
General
Cell factor Approx. 2 (for installation conditions, see Fig. 10 Minimum extent of the free field without change of the cell factor, page 9)
Transfer ratio 120.0
Sensor coding F003
Measuring range 0 ... 2000 mS/cm
Resolution 0001 mS/cm
Accuracy (-20 ... +100 °C) ± (0.005 mS/cm + 0.5 % meas. value)
Accuracy (> 100 °C) ± (0.010 mS/cm + 0.5 % meas. value)
MaterialSE 655(X) SE 656(X)
PEEKPFA
Process temperatureSafe areaEx T4Ex T6
-20 ... +125 °C -20 ... +125 °C -20 ... +75 °C
Temperature response time t90 (DIN 746-1)
SE 655(X)SE 656(X)
Approx. 7 minApprox. 11 min(For measuring processes with rapid temperature changes, use a separate temperature detector with fast response.)
Ambient temperature -20 ... +70 °C
Storage temperature -20 ... +80 °C
Process pressureSE 655(X)SE 656(X)
0 ... 20 bar0 ... 16 bar
Temperature detector Pt 100, Class A (IEC 751)
Assembly G 3/4
CableLength
Connection
5 m (max. permissible cable length for hazardous-area applications: 55 m)Wire end ferrules
Protection (EN 60529) IP 68 (sensor mounted, with original gasket)
Explosion protectionSE 655X/SE 656X
II 1G Ex ia IIC T4/T6 Ga
Weight Approx. 1 kg
6
20
16
-20 125
16
-20 20 125
9
1,5
P/T diagram (sensor version)
P/T diagram (flange version)
Fig. 2 Pressure/temperature curves for the different flanges
Pres
sure
[bar
]
Temperature [°C]
Temperature [°C]
Pres
sure
[bar
]
SE 655(X), PEEK
SE 656(X), PFA
DN 50, stainless steel (ZU 0343)
DN 50, PVDF (ZU 0344)
Fig. 1 Pressure/temperature curves for the different sensor materials
Temperature and pressure resistance
Specifications
Chemical resistance
Medium Concentration [%]
Temperature[°C]
Resistance PFA / PEEK
Nitric acidHNO3
5 2060
++
++
up to 40 2060
++
+-
Phosphoric acidH3PO4
up to 10 2060
++
++
Sodium hydroxide solution
NaOH3
205080
+++
+++
7
Connect the sensor to you measuring device. Refer to the user manual of your measuring device for information on the electrical connection. For further information, see www.knick.de.
Electrical connection
Identification of the sensors
Safe-area model Hazardous-area model
Fig. 4 Model designation / rating plates of SE 655(X) sensors
Safe-area model Hazardous-area model
Fig. 5 Model designation / rating plates of SE 656(X) sensors
Coax RD
Coax WH
BU
BU
RD
RD
GN
WH
YE
GY
receive hi
receive lo
send lo
send hi
RTD (temp)
RTD (temp)
sense (temp)
shieldFig. 3 Wire assignment of measuring cable
Specifications
8
Assembly, with PTFE washer (for flange protection)Max. torque 20 Nm, 36-mm wrench
Fig. 6 Assembly, with PTFE washer
Fig. 7 Assembly, without PTFE washer
Nut (included with sensor) Replacement: ZU 0341 N
ZU 0343 flange
ZU 0342N: PTFE washer DN 50
O-ring (included with sensor)Replacement: ZU 0341N
Nut
Flange (stainless steel) DN 50 PN 16
Gasket
Sensor
Flange kit DN 50 PVDF PN 10
Lapped flange (UP-GF)NutFlange (PVDF)
Sensor: PFA
Assembly, without PTFE washer (flange wetted by process medium)Max. torque 20 Nm, 36-mm wrench
Assembly
9
1,40
1,20
1,00
0,800 10 30 50
am
ax 1
00
Min. distance 30 mm (no change of cell factor)With a distance of less than 30 mm to metallic tank or pipe walls, the cell factor will be reduced. With insulating walls, the cell factor will increase.
Make sure that the sensor is correctly positioned.
Minimum water level in the pipe(sensor head completely wetted)
Distance a to pipe wall [mm]
Fig. 10 Minimum extent of the free field without change of the cell factor
Fig. 8 Minimum extent of the free field without change of the cell factor
Fig. 9 Typical installation in a pipe
Insulating pipe wall
Conducting pipe wall
Installation factor
Assembly
10
Fig.11 Dimensions of the sensor [mm]
Dimension Drawings
Ø 24
Ø 39
Ø 47
Ø 15.5
27.5
G 3/4
36 A
/F 9814
8 / 1
51
11
Flange kit DN 50 PN 10, PVDF, ZU 0344
Flange kit DN 50 PN 16, 316L, ZU 0343
Fig. 12 Dimensions of the flanges [mm]
Lapped flange Flange (PVDF)
4 x 18
18
78125165
4 x 18
18
125165
27
2011
75102
27
Flange (stainless steel)
Dimension Drawings of Flange
18 Accessories
Order No.
Flange DN 50 PN 16 Material: 316 L (When using the SE 656(X) sensor or measuring in aggressive media, you require sealing set C.)
ZU 0343
Flange DN 50 PN 10 Material: PVDF ZU 0344Sealing set A spare part for SE 655(X)
Nut + FKM (Viton) O-ring (3 x) ZU 0340N
Sealing set B spare part for SE 656(X)
Nut + FFKM (Chemraz) O-ring (2 x) ZU 0341N
Sealing set C PTFE washer DN 50(protects ZU 0343 flange against aggressive media)
ZU 0342N
Conductivity standard KCl 0.1 mol/l 250 ml 12.88 mS/cm + 1.5% ZU 0348
For more accessories enabling the installation in immersion fittings or flow-through fittings, refer to the Knick “Process Analytics” catalog.
For your notes: