User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis...

21
1 User guide for Cary 5000 absorption spectrometer with external DRA 1800 attachment. (last updated 10/25/2017) This guide is for use of the Cary 5000 with the DRA only. For use without the DRA as a standard transmission spectrometer see the User guide for Cary 5000 in absorption mode. Important warnings! Do not unplug or plug in the external DRA attachment when the instrument is on. This will ruin the detectors of the integrating sphere Do not put white light into the DRA with the DRA electrically connected to the Cary. Minimize the amount of time that the integrating sphere is open. Long-term light exposure hurt the detector over time. Wear clean gloves while using the instrument Do not use liquid or powder samples in the integrating sphere. Quick Start for using DRA 1. Turn on computer 2. Click on the Cary Scan icon on the taskbar 3. Check lens and mirror are correct for your measurement. (Ware gloves to change the mirrors or lens). a. Install Small Sample Kit mirror (M3 SSK) or regular M3 mirror if needed b. If using SSK install correct lens L2. 4. Make sure cover for the DRA is fully closed and turn Cary on. 5. If you get an error turn go to the section below on Startup Errors. off 6. Wait 20 minutes for the lamps to warm up. 7. If you need to change and/or align the optics see the local user manual below and remember never put white light into the integrating sphere when the Cary is turned on. To align the instrument use 550 nm light. 8. Select Setup a. set wavelength range, %T or %R b. Under Options tab: set slit width and height, SBW to 2 nm, Detector and Grating change wavelengths c. Under Baseline tab: chose zero/baseline d. Under Storage tab: Set your filename 9. Make sure that all exits of the integrating sphere are blocked with full reflectors. When using the small spot kit you must take care that the full reflectors completely cover the exit ports with their white reflector material.

Transcript of User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis...

Page 1: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

1

UserguideforCary5000absorptionspectrometerwithexternalDRA1800attachment.

(lastupdated10/25/2017)

ThisguideisforuseoftheCary5000withtheDRAonly.ForusewithouttheDRAasastandardtransmissionspectrometerseetheUserguideforCary5000inabsorptionmode.

Importantwarnings!• DonotunplugorplugintheexternalDRAattachmentwhentheinstrumentis

on.Thiswillruinthedetectorsoftheintegratingsphere• DonotputwhitelightintotheDRAwiththeDRAelectricallyconnectedtothe

Cary.• Minimizetheamountoftimethattheintegratingsphereisopen.Long-termlight

exposurehurtthedetectorovertime.• Wearcleangloveswhileusingtheinstrument• Donotuseliquidorpowdersamplesintheintegratingsphere.

QuickStartforusingDRA1. Turnoncomputer2. ClickontheCaryScanicononthetaskbar3. Checklensandmirrorarecorrectforyourmeasurement.(Wareglovestochange

themirrorsorlens).a. InstallSmallSampleKitmirror(M3SSK)orregularM3mirrorifneededb. IfusingSSKinstallcorrectlensL2.

4. MakesurecoverfortheDRAisfullyclosedandturnCaryon.5. IfyougetanerrorturngotothesectionbelowonStartupErrors.off6. Wait20minutesforthelampstowarmup.7. Ifyouneedtochangeand/oraligntheopticsseethelocalusermanualbelowand

rememberneverputwhitelightintotheintegratingspherewhentheCaryisturnedon.Toaligntheinstrumentuse550nmlight.

8. SelectSetupa. setwavelengthrange,%Tor%Rb. UnderOptionstab:setslitwidthandheight,SBWto2nm,Detectorand

Gratingchangewavelengthsc. UnderBaselinetab:chosezero/baselined. UnderStoragetab:Setyourfilename

9. Makesurethatallexitsoftheintegratingsphereareblockedwithfullreflectors.Whenusingthesmallspotkityoumusttakecarethatthefullreflectorscompletelycovertheexitportswiththeirwhitereflectormaterial.

Page 2: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

2

10. ChooseBaselineandrun100%Tscan,thenblockthesamplebeamattheentrancetotheintegratingsphereforthe0%Tscan.Ifusinganapertureholderforreflectionmeasurements,pleaseseethe“Aperturekitforsmallsamples”sectiononpage4.

11. Mountyoursample12. ChooseStarttorunthespectra 13. Whendoneremovesampleandremountreflectors.14. TurnoffCary15. Shutdowncomputer.

IntroductionandHelpForacompleteguidetotheCaryhardwareandopentheCary“Scan”programandclickontheHelptab.

Forthebasictheoryofdiffusereflectancemeasurements,usefuldiagramsofthediffusereflectanceadaptor(DRA),alignmenttipsandprocedures,anddescriptionsofalltheDRAaccessories(VATH,SmallSpotKit,andcentermountholders)gotoHelp>Helptopics>Accessories>Cary4000/5000/6000i>SolidSample>externalDRA

TolearnmoreaboutthesoftwareandtakingmeasurementsintheScanmenu,gotoHelp>Helptopics>SoftwareApplications>Scan

StartupError1. Ifyoureceivetheastartuperror(normallyE9303)whenstartingtheCary.Dothe

following:2. ClickokandlettheCarycontinuetheinitializationprocess.3. AfteritstartsandgotoCommandsandusetheGoTomenuoptiontogoto500nm.4. Checkthatthesmallapertureinthesamplebeamlightpathisallthewayopen.5. Turnthelightsoffintheroom.Checkthatthegreenlightcomesoutofthe

monochromatorwindowandisdirectedintotheintegratingsphere.Youneedtocheckboththesampleandreferencebeams.

6. Ifyoudon’tseeanylightcomingoutofthemonochromator,youneedtogobackandusetheGoTocommandagainandwaituntilitcompletesbeforeopeningtheinstrument.

7. Checkthatthereisnothingblockingthelightfromenteringintegratingsphere.8. Checkthatthereisnosampleonanyoftheportsorinsidetheintegratingsphere.9. ClosethesamplecompartmentontheCaryandmakesureyouheartheinterlock

click.10. TurnofftheCary,wait10s,andthenturntheCarybackon.Itshouldnowstartup

correctly.

Start

Page 3: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

3

LocalUserGuidefortheCary500withDiffuseReflectanceAccessoryIntegratingSphere.

Turningtheinstrumenton:OpentheCaryScanprogramonthetaskbarorDesktopTheDRAmusttobefullyclosedinorderfortheinstrumenttoinitialize.Turnontheinstrument(0/Iswitchonthelowerleftsideofthefrontoftheinstrument).IntheScanprogram,watchthelowerleftcornertoseethestatusoftheinstrument.Whenthetrafficlightisgreen(topcenterofscreen),thenascancanberun.IfthereisanerrormessageseeStartupErrorsectionabove.

BasicsconfigurationandalignmentTherearefourpositionsinwhichthesamplecanbeplacedforusewiththeintegratingsphere(seeFigure1andFigure2).

1. Theentranceslottotheintegratingsphereisfornormalincidencediffusetransmission

2. Therearslotisfornormalincidencediffusereflection3. Thecentermountholderallowsforvariableanglediffusereflectanceand

transflectancemeasurements

4. Inaddition,theVATH(variableangletransmissionholder)accessorycanbeputinfrontoftheintegratingsphereforvariableanglespeculartransmissionmeasurements.IfyouwouldliketousetheVATHaccessory,pleasetalktooneoftheGLAsfortheinstrumentfirst.

ForthenormalorlargesamplesetuptherearethreepositionsforthelargestandardM3mirror(seeFigure4andFigure3)thatcorrespondtothethreesamplepositions(Transmission,Center,andReflection).Eachofthemirrorpositionswillfocusthelightontothepositionofthesample.ThethreedifferentmirrorpositionsaremarkedT(transmittanceslot),C(centermount),andR(rearreflectanceslot)inFigure4.Thus,themirrorneedstobemovedtotherightpositiondependingontheplacementofyoursample.ThecablefortheDRAcangetinthewayofmovingthismirror.

Figure1Integratingsphereentranceslot.

Figure2Integratingsphereexitslots.

Page 4: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

4

Wareglovestohandlethemirrorsorlens.TurnofftheCaryfirst,ifyouneedtounplugtheDRAtomovethemirror.

Forthesmallsamplesetup(seesmallsamplediscussionbelow)theM3mirrorisreplacedbythesmallsamplekitmirror(M3SSK)intheTposition.Lens(L2)mountsinfrontofthesample(seeFigure3).TherearethreelensesthatcanbeusedagainforthethreepositonsmarkedasT,CorR.

Eachtimethesampleconfigurationischanged,theopticsneedtoberealigned,suchthatthesampleandreferencebeamsarenotclippedattheentrancetotheintegratingsphereandthatthesamplebeamiscenteredonthesample.

1. Putoncleangloves.2. WiththetopdoortotheDRA

closed,underthecommandsmenuclickonGoto,andthentypein550nm(donotusethe“align”option).Whenyoudarkentheroomlights,youshouldseeagreenbeam,whichyoucanusetoalignthebeamontothesample(Alternatively,ifyouhavetroubleseeingthebeam:

Figure4DiffuseReflectanceAccessorywithbeampath.

Figure3Photoofdiffusereflectanceaccessorywithbeampath.

Page 5: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

5

Blockthesampleandreferencebeamsfromenteringtheintegratingsphere.UnderthecommandsmenugotoaligntoswitchthegratingstozeroordertosendwhitelightintotheDRA.Thewhitelightshouldnotentertheintegratingsphereoritwilldamageit!!)

3. Mostofthemirrorshavetwoknobs,whichyoucanrotatetoadjusttheverticalandhorizontalpositionsofthebeam.StartwithmirrorM2andadjustsothebeamhitsthecenterofmirrorM3thenuseM3sothebeamhitsthecenterofthesample.Someofthemirrorsaremorecomplicatedtoadjust(althoughthesecanoftenbeleftalone).Forhelpwithalignment,talktooneoftheGLAsandreadthealignmentinstructions,whichcanbefoundunderHelp.

SmallsamplesMosttypicallabsamplesaresmallerinsizethanwhattheinstrumentwasdesignedtomeasure.Forallthreetypesofmeasurements(T,C,andR),itisbesttounderfillthesamplearea.Spotsize-Therearetwowaysofreducingthebeamspotsize:

1)Slitwidth:Thesimplestwayistoreducebeamsizeisbytheslitwidth.ClickontheSetuptheoptionstab.Underslitheight,switchtoReduced.Ittakesaboutaminutefortheslittochange.

2)Smallspotkit:Therearealsoaseriesoflensesthatcanbeputintheopticalpathtofocusthebeamintoasmallerspotonthesample.AgaintherearethreelenseswithdifferentfocallengthsandmarkedT,C,andRforthethreepositionsoftheintegratingsphere.Also,thelargemirror(M3)isswitchedoutforaonewithadifferentcurvature;thismirrorisalwayskeptintheTposition,regardlessofthetypeofmeasurementbeingmade.3)Aperturekitforreflectionmeasurements

Fornormalincidencereflectionmeasurements,theopencircleintherearoftheintegratingsphereisseveralinchesindiameter.Thereareaseriesofapertures,whichhavethewhiteSpectralonmaterial(sameroughenedTeflonthatisinsidetheintegratingsphere),thatcanslideontotheslotonthebackoftheintegratingsphere.Thespring-loadedmountthatholdsthesampleagainsttheintegratingspherefirstneedstobetakenoff.Slideontheappropriateapertureandtheneithertapeyoursampleontothebackoftheaperture,orputbackonthespring-loadedsamplemount.

Ifthebeamspotoverfillstheaperture,youcancorrectforthis“straylight”byperformingaZero/baselineCorrection.Normally,the0%TBaselinescanisrunbyblockingthebeambeforeitenterstheintegratingsphere,whichaccountsforelectronicnoise.Inthiscase,donotblockthebeamsothatthe0%TBaselineaccountsforthelightreflectingontheedgesoftheaperture.

PerformingaZero/BaselineCorrection:Selectthisoptiontoapplya100%Tbaselinecorrection(100%TBaseline)andazero-linecorrection(0%TBaseline)to

Page 6: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

6

thesamplescan(S).Thiscorrectionisperformedwiththerawtransmissiondataandisthus:

𝑆 − 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒100%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛 − 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒

Withthisoption,theCarywillpromptyoutoperforma100%TBaselinefirst,followedbya0%TBaselinewhenyouperformabaselinecollection.

BasicsofmeasurementacquisitionThewavelengthrangeis1800nmto250nm.Scansnormallygofromlongtoshortwavelengths.Therearetwodetectors,aPMTandanInGaAs.Thewavelengthatwhichthesystemchangesfromonetotheother(accompaniedbyagratingchange)canbeadjusted.Usually750to800nmisagoodvalueforthischangeover.Youcanseethechangeinthespectrumwhenyoutakeabaseline(%TransmissionorReflection).Ifyouseeastepinthespectrum,itmaymeanyourbaselineisbadorthatyoursamplehasapolarizationdependency.

Setup–Thesetupbuttonlocatedintheupperleft-handcornerdefinesthescanparameters.ThefollowingusefultabscanbefoundunderSetup.

CarytabHereyoucansettheWavelengthRange,theunitsofthex-andyaxes(nm,cm-1,etc.andA,%T,%R,etc.),aswellastherangefortheaxes.Underscancontrols,theaveragetimeperdatapointandthedataintervalcanbeset.NotethattheScanRatewillbeautomaticallydeterminedfromthesevalues.

OptionstabHereyoucontroltheBeamMode,SpectralBandWidth(SBW),andSlitHeight(seeAppendixAofthisguideformoreinfo).TheDoublebeammodeneedstobeusedfortheintegratingspheremeasurementsandtheSBWistypicallysetto2nm.

TheSlitHeightcanbeusedtodecreasethebeamspotsizenormallyitissettofull.Typically,theUV-Visbuttonwillbeselected.Ifyoudonotneedtogotoshorterwavelengthsthan400nm,youcanselecttheVisbutton,whichwillturnofftheUVlamp.

Baselinetab

Thistabenablesyoutochoosethetypeofbaseline(s)thatyouwanttouseintheScanrun.ThedifferenttypesofbaselinecorrectionsareparticularlyusefulwithsomeoftheVarianaccessories.Youcancollectabaselineimmediatelybeforescanningasampleoruseastoredbaseline.

Theaccuracyofyourbaselinedeterminestheaccuracyofyourmeasurement.Youshouldperformazero/baselinecorrectionbeforerunningyoursamples

Page 7: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

7

aftertheinstrumenthaswarmedupfor~20minutesusingthesamescanparametersthatyouwilluseforyoursample.TheCaryWinUVsystemcollectsanewbaselinewhenyouselectanytypeofbaselinecorrectionintheBaselinetaboftheSetupdialogboxandthenclickBaselinebuttonintheapplicationwindow.Dependingonthetypeofcorrectionyouhaveselected,theCarywillpromptyoutoperformeitheroneortwobaselinescans.Thefirstisa100%Tbaseline(fortheDRAwiththe100%referencematerialinthesamplepositions).Thesecond,ifrequired,isa0%Tbaseline(withthesamplebeamblocked).

Oncethebaselineshavebeencollected,theywillbeautomaticallyappliedtoeachcollectedsampledatapointasitisdisplayed(indicatedbythereddisplay).Ifthereisnotacorrespondingdatapointinthecurrentbaseline,thenthefilewillbeinterpolatedtoprovidethecorrection.ThecorrectionappliedtoeachpointisdependentontheselectedbaselinetypeintheBaselinetab.

Abaselinecannotbeappliedtoasamplewherethewavelengthrangeofthesampleexceedsthatofthebaseline.Ifyouchangeyourwavelengthrangetooneoutsidethatofthebaseline,theCarywilldisplayanerrormessage.Alsotheredtextintheordinate(y)display,whichindicatesthatthebaselinecorrectionmodeisactivatedwilldisappear.

Forreflectionmeasurementsusingtheaperturekit,ifthebeamspotoverfillstheaperture,youcancorrectforthis“straylight”byperformingaZero/baselineCorrection.Normally,the0%TBaselinescanisrunbyblockingthebeambeforeitenterstheintegratingsphere,whichaccountsforelectronicnoise.Inthiscase,donotblockthebeam,butinsteadremovethereflectanceportcaptoaccountforthelightreflectingontheedgesoftheaperture(theremainderofthelightshouldpassthroughtheintegratingsphereandnotbedetected).

PerformingaZero/BaselineCorrection:Selectthisoptiontoapplya100%Tbaselinecorrection(100%TBaseline)andazero-linecorrection(0%TBaseline)tothesamplescan(S).Thiscorrectionisperformedwiththerawtransmissiondataandisthus:

(𝑆 − (0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒))(100%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒 − 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒)

Withthisoption,theCarywillpromptyoutoperforma100%TBaselinefirst,followedbya0%TBaselinewhenyouperformabaselinecollection.

Thisbaselineoptioncorrectsforanyinherentvariationsintheelectroniczerolineoftheinstrument.Youshouldusethisbaselineoptionifyoursampleshaveareasofverylowtransmission(highabsorbance)asanyvariationsintheinstrument'szerolinewillaffectyourmeasurementsiftheyyoudonotcorrectforthese.YoushouldrefertoASTMmethodE903forfurtherdetails.

SeeAppendixBformoreinfoontheacquisition,storing,andretrievalofbaselinefiles

Page 8: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

8

IndependentTab

ThisisnormallynotchangedfromdefaultAutoinMeasurementMode.InthismodetheinstrumentwillcollectdoublespectrausingafixedSBWintheUV-VisregionandafixedenergyordetectorgainintheNIRregion.Youcanexplicitlysethowtheinstrumentcontrolsthedetectorreadingsineachregion.WitheitherAutoorFixedSBWsetastheMeasurementMode,youmustsetaSBWvalueintheUV-Visgroup.Thenormalsettingis2.00nm.YouwillfindthattheEnergylevelwillchangeautomaticallytomaintaintherequiredsignallevel.Ifyouwanttheinstrumenttosaninanothermodei.e.fixeSBEintheNIRorfixedenergy(gain)intheUV-Visregionyoucansetthathere.Storagetab

TheAutoStoretaballowsyoutospecifyifyouwanttobepromptedtostorethecollecteddata.YoucanstorethedatainaBatchorDatafileatthestartorendofacollect.Thereare3differentoptionsforautomaticallystoringyourdata

StorageOff:Selectthisifyoudonotwishtobeautomaticallypromptedtosavecollecteddata.Tomanuallystoreyourunsaveddataaftertheruniscomplete,chooseSaveDataAsfromtheFileMenu.YoucansaveyourdataasaBatchorDatafile.IntheScanapplication,dataisautomaticallystoredasaDatafile.SeetheNotebelow.

StorageOn(PromptAtStart):SelectthistodisplaytheWindowsSaveAsdialogboxatthestartofthecollectwhereyoucanentertheBatchfilenameforyourdata.

StorageOn(PromptAtEnd):SelectthistodisplaytheWindowsSaveAsdialogboxattheendofacollectwhereyoucanentertheBatchfilenameforyourdata.

NoteIntheScanapplication,PromptatStartandPromptatEndsaveinformationasa*.DSWfileunlesstherearemultiorcyclescansbeingperformed.Inthesecases,informationissavedinaBatchfile.

CommandsMenuCommandsinthisapplicationcanbeaccessedbothfromthemenuitemsatthetopofthemainwindowandfromtheapplicationbuttonsontherighthandsideofthescreen.Alternatively,youcanusefunctionkeyshortcutsforsomeofthecommands.TheseappearnexttotheitemintheCommandsmenu.

Start:SelectthistostartaScanrunusingthecurrentlysetupparameters.FunctionkeyisF9.

Stop:SelectthistostopaScanrun.IfyouhaveselectedStorageOnintheAutoStoretaboftheSetupdialogbox,alldatacollectedisautomaticallystoredinthe

Page 9: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

9

filespecifiedintheBatchnamefieldoftheWindowsSaveAsdialogboxbeforeoraftertheruncommenced.FunctionkeyisF12.Connect:TheConnectcommandbecomesavailablewhenmorethanonesoftwareapplicationisrunningatanyonetime.TheapplicationthatisinusewillbeonlineandwillhavecontroloftheCary.AllotherapplicationswillbeofflineanddisplaytheConnectcommand.Ifyouwishtoswaptoadifferentapplication(andstillkeepotherapplicationsrunning)youwillneedtoselectConnectintheapplicationyouwishtoconnecttotheCary.

Zero:SelectthistozerothecurrentordinatevalueusingtheStartwavelength(orequivalent)valuesetintheCarytaboftheSetupdialogbox.TheCarywillchangetothiswavelengthifcurrentlyidlingatadifferentvalue.FunctionkeyisF5.Donotusethisifyousampleisnotzeroatthiswavelength.Goto:SelectthistodisplaytheGotoWavelengthdialogboxandchangetheinstrumenttoanewwavelength.FunctionkeyisF4.Thisoptionisonlyavailableiftheinstrumentisnotcurrentlycollectingdata.Align:ThismovestheCaryspectrophotometerto0nm(whitelight).Youcanthenusethislightforaligningaccessories.Donotuseunlessyouhaveblockedthesampleandreferenceslotstotheintegratingsphere.LampsOff:Switchesoffalllampsources.

LampsOn:Switchonselectedlamps.

GraphmenuTheGraphmenuallowsyoutoviewdatainvariousgraphicalformatsorinmultiplegraphsintheGraphicsarea.Youcandouble-clickontheGraphicsareatotogglethedisplaythroughgraphandreporttographonly.Thegraphfunctionscanalsobeaccessedby:

• clickingoneoftheToolbarbuttons.

• right-clickinginagraphboxortheGraphicsarea.Notallfunctionsareavailableoneverymenu.

NoteTheredtraceinanygraphboxistheoneforwhichallinformationappearsandtheoneusedforcursortracking.Itiscalledthefocusedtrace.Ifyouholdthecursoroveratrace,bubbletextidentifyingthetraceisshown.

UndertheGraphmenu,youcanchangeboththeTraceandGraphPreferencesTracePreferences:SelecttodisplaytheTracePreferenceswindowtochoosethetrace(s)todisplayintheselectedgraph.GraphPreferences:SelecttodisplaytheGraphPreferencesdialogboxandchangetheappearanceofyourgraph.

Page 10: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

10

SeeAppendixCformoreoptionundertheGraphmenuaswellastheTracePreferencesandGraphPreferencesmenus

StoringDataUndertheFilemenuyoucansaveyourMethod(theSetupconfigurationusedtoperformthescan),yourData,orthecombinationoftheMethodandDataasaBatchfile.

File,SaveMethodAsSelectthisoptiontodisplaytheSaveAsdialogboxandstorethecurrentmethod.YoucanusetheSaveAsdialogboxtostorethemethod,data,report,orgraphicstemplateseparatelyortogetherasaBatchfile.ThedatacanalsobesavedasaCaryGRAMSfile,acommadelimitedASCIISpreadsheetfileorinformatssuitableforotherCarysoftwaresystems(dependsonwhichapplicationyouarerunning).

File,SaveDataAs

SelectthisoptiontodisplaytheSaveAsdialogboxandstorethecollecteddata.YoucanusetheSaveAsdialogboxtostorethemethod,dataorreportseparatelyortogetherasaBatchfile.ThedatacanalsobesavedasaCaryGRAMSfile,acommadelimitedASCIISpreadsheetfileorinformatssuitableforotherCarysoftwaresystems(dependsonwhichapplicationyouarerunning).Afterabatchfilehasbeensaved,youcanusethisoptiontomanuallysavethefileasamethodfile,abatchfile,acommadelimitedASCIIspreadsheetfileoraRichTextFormatfile.

HowtoExportCollectedDataIfyouwanttoexportdatafromaCaryWinUVapplicationintoathird-partyapplication,thensaveyourdataasanASCII*.CSVfile.Todothis:

1. ClickFilemenuitem.2. ClickSaveDataAs.TheSaveAsdialogboxwillappear.3. ClickthedownarrownexttotheFilesoftypelistbox.4. SelectSpreadsheetASCII(*.CSV).5. TypeyourfilenameintotheFilenamefield.6. ClickSavetoexportyourdata.

HowtoCombineDataFilesintoaBatchFile:

Ifyouhaveanumberofdatafilesthatyouneedtocombineintoabatchfile,thenyouneedto:

1. SelectOpenDatafromFilemenu.2. Alistofpreviouslystoredbatchfileswillappear.Clickthedownarrowat

therightofthe“FilesofType”listboxandselect'Data'tolistallthedatafiles.

Page 11: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

11

3. ChecktheOverlayDatacheckbox.4. HighlightthedatafilesthatyourequireandclickOpen.Thehighlighted

datafileswillloadintotheapplicationandappearinthesamegraphbox.

5. SelectSaveDataAsfromtheFilemenu.6. Alistofpreviouslystoreddatafileswillappear.Clickthedownarrowat

therightofthe“FilesofType”listboxandselect'Batch'tolistallthebatchfiles.

7. MakesurethattheSaveonlyfocusedtracecheckboxisnotchecked.8. IntheFilenamefield,typeinthefilenameforyournewbatchfile.9. ClickSavetocreatethenewbatchfile.Thecurrentmethodwillbestored

withthebatchfiles.Allthedatafilesarenowcombinedintotheonebatchfile.

Page 12: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

12

AppendixA

HowtheCary5000/6000iControlsSBWandEnergyInaspectrometertheintensityobservedbythedetectorchangeswithwavelength.Doublebeamspectrometersnormallycontrolthereadingfromthedetectorstomakesurethattheystaywithintheiroptimumregion.Thiscanbedoneintwoways.Youcanopenandclosetheslitstoincreaseordecreasetheamountoflightthatreachesthedetectororyoucanchangethegainonthedetectortomakeithavehighergainwhenthelightenergyislowandvisaversa.NormallyonelikestohavetheSpectralBandWidth(SBW)stayconstant(i.e.constantslitwidth).ThisiseasyintheUV-Visregionwherethedetectorisaphotomultipliertube(PMT)thatallowsaverylargechangeingain.However,intheNIRregionthishistoricallywaslesspossiblesinceyouareusingaPbSdetector.ThuswhatwasdonewastolettheSBWvarytomaintainthedetectoroutput.InourCarywehaveaInGaASdetectorandwhileyoucannotchangethegainonthedetectoritisverylownoiseandyoucanuseaamplifierwithsetablegainafterthedetectortobostthesignal.InDoubleorDoubleReverseBeamModeintheUV-Visregiontheslitwidth(SBW)isnormallykeptconstantandtheEnergylevel(gain)ischanged.IntheNIRregion(InGaAsdetector)thereverseisnormallytruei.etheEnergylevelisconstantandtheSBW(slits)changes.Duetotheinherentdifferencesbetweenthetwotypesofdetector,thedifferentmethodsofcontrolnormallyensurethewidestdynamicrangeandthebestsignaltonoise.

Theinstrumentusestwodetectors(PMTandInGaAsdiode),twolightsources(WbulbandD2lamp),andtwogratingsinthemonochromator.Youcansetthewavelengthatwhichthedetectorsandgratingschange.Bothofthesearetypicallysetat800nm,ifyouhaveanimportantspectroscopicfeatureat800nm,youcanmovethechangeover±50nm.Thelampchangeoverisnormallysetto350nmbutagaincanbemoved±50nm

Notethattheinstrumentusesgratingstodispersethelightandneedsordersortingfilterstoremove2nd(and3rd)orderlightfromtheoutputbeam(i.e.at800nmyouwouldalsoget400and200nmlight).Thefilterschangeareat350,570,800,and1200nm.Youcannotmovetheirchangeoverwavelengths.IfyouarecollectingdataindoublebeammodeintheUV-Visregions,youcansethowtheinstrumentcontrolsthedetectorreadings.InSETUPundertheindependenttabifyouchoosehowthedoublebeamspectraarecollected.WitheitherAutoorFixedSBWsetastheMeasurementMode,yousetaSBWvalueintheUV-Visgroup.Thenormalsettingis2.00nm.YouwillfindthattheEnergylevelwillchangeautomaticallytomaintaintherequiredsignallevel.

IfyouarecollectingdataindoublebeammodeintheNIRregion(λ >thedetectorchangewavelengthsetting,generally800nm),witheitherAutoorFixedEnergysetastheMeasurementMode,theEnergylevelisfixedintheNIRgroup.AnEnergysettingof1.0willprovidethebestsignal-to-noiseratiobutwithalargerSBW.Inmostcases,therecommendedsettingis3.0;thisprovidesasmallerSBWovertheentireregionanda

Page 13: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

13

bettermatchoftheSBWintheNIRregionandtheVisregion.YouwillfindthattheSBWwillchangeautomaticallytomaintaintherequiredsignallevel.IfyouarecollectingdataindoublebeammodeacrosstheUV-VisandNIRregionswithFixedSBWsetastheMeasurementModethenyouwillneedtoentervaluesintheSBWfieldsinboththeUV-VisandtheNIRgroupboxes.ThevaluescanbethesameifyouwanttousethesameSBWacrossthewholescanoryoucansetdifferentvaluesforeachregion,normallytheSBWintheNIRis>>thentheSBWintheUV-Vis.NotethattheSBWisinnmandforafixedsettingoftheSBWtheresolutioninenergytermswillbemuchbetterintheNIRthenintheUVorVisregions.Aspectrabandwidthof2nmat400nmcorrespondsto124cm-1whileat1200nm124cm-1giveaSBWofabout18nm.ThisisonereasonthatmostabsorptionpeaksintheNIRaremuchbroaderinwavelengththenpeaksintheUV-Visregions.

Page 14: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

14

AppendixB:Acquiring,storing,andretrievingbaselinefilesTheCaryWinUVsystemcollectsanewbaselinewhenyouselectanytypeofbaselinecorrectionintheBaselinetaboftheSetupdialogboxandthenclicktheBaselinebuttonintheapplicationwindow.Dependingonthetypeofcorrectionyouhaveselected,theCarywillpromptyoutoperformeitheroneortwobaselinescans.Thefirstisalwaysa100%Tbaseline(withnothinginthebeam,orinthecaseofaDiffuseorSpecularReflectancemeasurements,withthe100%referencematerialinposition).Thesecond,ifrequired,isa0%Tbaseline(withthesamplebeamtotallyblockedsothattheinstrumentcanmeasuretheelectroniczerovalues).

Oncethebaselineshavebeencollected,theywillbeautomaticallyappliedtoeachcollectedsampledataaccordingtotheappropriatebaselinecorrectionequation.Ifthereisnotacorrespondingdatapointinthecurrentbaseline,thenthefilewillbeinterpolatedtoprovidethecorrection.ThecorrectionappliedtoeachpointisdependentontheselectedbaselinetypeintheBaselinetab.Abaselinecannotbeappliedtoasamplewheretheabscissarangeofthesampleexceedsthatofthebaseline.Ifyouchangeyourabscissarangetooneoutsidethatofthebaseline,theCarywilldisplayanerrormessage.Alsotheredtextintheordinatedisplaywhichindicatesthatbaselinecorrectionmodeisactivatedwilldisappear.

NotthatallvalueswithintheBaselineabscissarangearecorrected,eventhevaluedisplayedwhiletheCaryisidlingataparticularwavelength.

NoteYoushouldbecarefulapplyingcorrectionstosampleswherethefollowing

parametersaredifferentinthebaselinefileandthesamples.Energy,SBW,BeamMode,SourceChangeover,DetectorChangeoverGratingChangeover,SelectedLampforthecollection,Signal-to-NoiseMode,SlitheightandtheIndependentControl.

Baselinesaresavedasapartofabatchfile.However,ifyouwant,youcanalsostorethebaselinesbythemselvesinabaselinefilewithoutthecollecteddata.TodothisyousimplyselecttheBaseline(*.CSW)astheSaveAsTypeintheSaveAswindowaccessedfromtheFile|SaveDataAsmenuitems.WhenyouwanttousethebaselinesforanotherScanrun,yousimplyopenthefileusingeithertheFile|OpenDatacommandorclicktheBaselinebuttonintheBaselinetaboftheSetupdialogbox.

NoteIfyoudecidethatyoudonotwanttousethatstoredfile,butwanttocollectanewbaselineinstead,thentheCarywillusethenew,collectedbaselineinallcorrections.However,theBaselinetabwillstilldisplaythenameofthebaselineinthebaselinefile.Youneedtosavethecollecteddataasabatchfileifyouwantthenewbaselineincludedinthatfile.

NoteAllcorrectionsareperformedwiththerawtransmissiondatathatcomesdirectly

Page 15: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

15

fromtheinstrument.ThesetransmissionvaluesarethenconvertedintotheselectedYmodefordisplay.

Abaselinecannotbeappliedtoasamplewheretheabscissarangeofthesampleexceedsthatofthebaseline.IfyouwishtousethatbaselinethenyouwillneedtoreducetherangeofthescanbychangingtheStartandStopfieldsontheCarytaboftheSetupdialogbox.

Typesofbaselinesthatcanbeperformed.ChoosethetypeofbaselinecorrectionundertheBaselinetabintheSetupmenu

BaselineCorrection

Selectthisoptiontoapplyabaselinecorrection(Baseline)tothesamplescan(S)thatiscollectedfromtheinstrumentastransmissionvalues.Thiscorrectionisthus:

𝑆𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒

Zero/BaselineCorrectionSelectthisoptiontoapplya100%Tbaselinecorrection(100%TBaseline)andazero-linecorrection(0%TBaseline)tothesamplescan(S).Thiscorrectionisperformedwiththerawtransmissiondataandisthus:

S− 0%TBaseline100%TBaseline− 0%TBaseline

Withthisoption,theCarywillpromptyoutoperforma100%TBaselinefirst,followedbya0%TBaselinewhenyouperformabaselinecollection.

Thisbaselineoptioncorrectsforanyinherentvariationsintheelectroniczerolineoftheinstrument.Youshouldusethisbaselineoptionifyoursampleshaveareasofverylowtransmission(highabsorbance)asanyvariationsintheinstrument'szerolinewillaffectyourmeasurementsiftheyyoudonotcorrectforthese.YoushouldrefertoASTMmethodE903forfurtherdetails.

ZeroSRAcorrectionSelectthisoptiontoapplya100%Tbaselinecorrection(100%TBaseline)andazero-linecorrection(0%TBaseline)toasample(S)measuredwithanAbsoluteVWSpecularReflectanceAccessory(SRA).

NoteTheVNSRAandthesinglebouncerelativeSRAaccessoriesshouldnotusethiscorrection.SelectaBaseline/Zerocorrectioninstead.

Thiscorrectionisperformedwiththerawtransmissiondataand,duetothedoublebounceonthesample,isthus:

Page 16: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

16

S− 0%TBaseline100%TBaseline− 0%TBaseline

Withthisoption,theCarywillpromptyoutoperforma100%TBaselinefirst,followedbya0%TBaselinewhenyouperformabaselinecollection.YoushouldrefertoASTMmethodE903forfurtherdetails.Zero*Std.Ref.Correction

Selectthisoptiontoapplya100%Tbaseline(100%TBaseline)correctionandazero-linecorrection(0%TBaseline)toasample(S)measuredwithareflectanceaccessorywhichisnotabsolute(e.g.,aDiffuseReflectanceAccessoryortheVariableAngleSpecularReflectanceaccessory).TheCarythenmultipliestheresultbytheappropriatevalueinthe'BaselineStdref'filewhicharethetransmissionvaluesassociatedwiththeStandardReferencematerialyouusedtoestablishthebaseline.RefertothehelpfortheDiffuseRelectanceaccessoriesforfurtherdetailsaboutStandardReferencematerials.

Thiscorrectionisperformedwiththerawtransmissiondataandisthus:S− 0%TBaseline

100%TBaseline− 0%TBaseline Baseline Stdref

Withthisoption,theCarywillpromptyoutoperforma100%TBaselinefirst,followedbya0%TBaselinewhenyouperformabaselinecollection.

NoteIfyouselectthisoptiontheBaselinebuttoninthemainwindowwillnotbeenableduntilyouhaveretrievedaStdReffile.ClickthisbuttontobrowseforandselectaStdReffile.Thisfilemustbeinthecorrectformat.

KnownMirrorSelectthisoptionwhenyouareusingaVWSpecularReflectanceAccessory(SRA)withasmallsampleorasampleoflowreflectance.RefertoyourSRAmanualformoreinformation.KnownMirrorCorrection

Thecorrectionequationisasfollows:S− 0%TBaseline

100%TBaseline− 0%TBaselineKnown mirror corrected

where

Known mirror corrected =known mirror− 0%TBaseline100%TBaseline− 0%TBaseline

TipIfbaselinecorrectionisactivated,thetypeofcorrectionthattheCarywillapplywillappearintheinred.

Page 17: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

17

However,sometimesyoumaynoticethatthetexthasturnedtograyandisinitalics.ThisindicatesthattheCaryiscurrentlyidlingoutsidetheabscissarangeofthebaselinefile,butthatthecurrentmethodStart/Stopparametersdonotexceedtheabscissarangeofthebaseline.

RetrieveBaselineFile

ThisbuttondisplaystheWindowsOpendialogboxwhereyoucanselectastoredbaselinefile.Thetrace(s)inthisfilewillbeusedfortheselectedbaselinecorrection.ThenameoftheselectedfileappearsnexttotheBaselinebutton.

TipIfyouwanttousethebaselineagainwithothersamples,savethemethodusingFile|SaveMethodAs.Then,whenyoure-openthemethod,thebaselinewillalsoopenandbereadytouse.Itispreferabletosavethebaselinewiththemethod,ratherthanabaselinefile,asthatwayyoucanbesurethatallyourcollectionparametersareexactlythesameforthenewScanruns.However,GoodLaboratoryPracticerecommendsthatyoucollectanewbaselineforeachlaboratorysession.

ViewBaselineFileThisdisplaysthetracesintheselectedbaselinefileintheGraphicsarea.

RetrieveStdRefFile

ThisbuttondisplaystheWindowsOpendialogboxwhereyoucanselectastoredreferencefile.ThetraceinthisfilewillbeusedintheZeroxStdRefCorrection.ThenameoftheselectedfileappearsnexttotheStdRefbutton.

TipTheStdReffilemustbeinthevalidformat.Youcanobtainthesefilesfromthemanufacturerofthestandardthatyouareusing.

• Howtocreateanduseastandardreferencefile

Q.WhyismyRetrieveStdRefbuttonunavailable?A.YouneedtoselecttheCorrectiontypeas'Zero*Std.Ref.Correction'.

ViewStdRefFileThisdisplaysthetracesintheselectedstandardreferencefileintheGraphicsarea.

ShowStatusDisplaySelectthischeckboxtodisplaytheShowStatuswindow.Thiswindowallowsyoutoviewthestatusofvariousparametersduringthedatacollection.

Page 18: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

18

AppendixCGraphmenu,TracePreferencesandGraphPreferencesoptions

1) GraphmenuoptionsPrintXYpoint:Selectthistoprintthecursor'scurrentXandYcoordinatesinthereportarea.Thisoptionisonlyavailablewhenright-clickinginagraphbox.

TracePreferences:SelectthistodisplaytheTracePreferenceswindowtochoosethetrace(s)todisplayintheselectedgraph(TracePreferencesarelistedbelow).GraphPreferences:SelectthistodisplaytheGraphPreferencesdialogboxandchangetheappearanceofyourgraph.Anychangesyoumaketotheappearanceofthegraphwillapplytoallgraphsuntilthesettingsarechangedagain.(Graphpreferencesoptionsarelistedbelow).

CursorMode:SelectthistodisplaytheCursorModesdialogboxtoenableyoutousedifferentcursorsonthescreen.Youcanselecttouseafreecursorthatcanmoveanywhereinthegraphorexactlytracethefocusedscan(redscan).

AxesScales:SelectthistodisplaytheAxesScalesdialogboxandchangetherangeoftheaxes.

Autoscale[XY]:SelectAutoscale[XY]toautomaticallyscalecollecteddatabyscalingtheordinate(Y)heightandabscissa(X)widthtofullscreen.

Autoscale[X]:SelectAutoscale[X]toautomaticallyscalecollecteddatabyscalingtheabscissa(X)widthtofullscreen.Autoscale[Y]:SelectAutoscale[Y]toautomaticallyscalecollecteddatabyscalingtheordinate(Y)heighttofullscreen.

ZoomOut:SelectZoomOuttoZoombacktotheoriginalXandYaxisscalingsetintheCarySetuptab.Thiswillundoallzooming.

AddLabel:SelectAddLabeltodisplaytheAddLabeldialogboxtoenableyoutoplaceatextlabelonyourgraph.

AddPicture:SelectAddPicturetodisplaytheAddPicturedialogboxtoinsertapicture,bitmaporchemicalstructureintoyourgraph.EditAnnotation:SelectEditAnnotationtoalterthecontentorlookofanannotation(suchasalabelorapicture)intheGraphicsarea.DeleteAnnotation:SelectDeleteAnnotationtoremoveanannotation(suchasalabelorapicture)fromtheGraphicsarea.

Single/MultiGraph:SelectSingle/MultiGraphtotogglebetweendisplayingoneorallgraphs.AutoArrangeGraphs:SelectAutoArrangeGraphstoviewandarrangeallgraphssimultaneously.

Page 19: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

19

CopyGraph:SelectthistocopytheGraphicsareaontotheWindowsClipboard.

PastetoGraph:SelectthistopastethecontentsoftheWindowsClipboardintotheGraphicsarea.Textispastedasalabel;graphicsarepastedaspictures.

AddGraph:ClickAddGraphtoaddanewgraphboxtotheGraphicsarea.RemoveGraph:ClickRemoveGraphtodeletetheselectedgraphboxfromtheGraphicsarea.

ClearAllTraces:SelectClearAllTracestodeletealltracesfromthegraph/s.UserDataForm:SelectthistodisplaytheUserDataForm.

Maths:SelectthistoenabletheMathswindow.Thisoptionisonlyavailablewhenright-clickinginagraphbox.HistoryList:Thisdisplaysalistofthecurrentgraphs.

2) TracePreferencesAsinglegraphboxcandisplaymanytraces.Thiswindowenablesyoutoselectthetrace(s)todisplay(makevisible)intheselectedgraphboxorcolorspace.TracePreferencesisaccessiblefromtheGraphmenu,byright-clickingonatraceorbyclickingonthetoolbar.

TheTracePreferencesmenuisalsoavailablebyright-clickinganywhereinthetracelisting.

Visible:Thisfieldallowsyoutodeterminewhichtraceswillappearontheselectedgraph.Clickinthiscolumntoplaceaticknexttothetraceyouwanttoappearonthegraph.Clickagaintoremovetheticknexttothetrace.

Ifyouareinsinglegraphview,allselectedtraceswillappearontheonegraph.IfyouareinMultiGraphview,thetraceswillappearonthegraphthatiscurrentlyselected.

Color:Thisshowsthecolorinwhichthetracewillbedisplayedinthegraphbox.Theredtraceinanygraphboxistheoneonwhichallinformationappearsandtheoneusedforthecursortracking.Itiscalledthefocusedtraceanditisalwaysred.

ClickingonatraceintheColorcolumnwillcausethattracetobecomethefocusedtraceintheselectedgraphbox.Youcanalsochangethecolorofanytrace(whenitisnotthefocusedtrace),fromtheTracePreferencesmenu.

WhenyouchangethetracelinewidthintheGraphPreferencedialogbox,italsowillchangethewidthofthecolortracelineinthiswindow.

NameofTrace:Thisfieldcontainsadescriptionofthetracescurrentlyavailablefordisplayintheselectedgraphbox.Ifatraceisnotalreadytickedasvisible,youcanclicktoactivatethisproperty.Right-clickinginthiscolumndisplaystheTracePreferencesmenu.Thismenuallowsyoutochangesomeofthepropertiesofthetraces,aswellashiding,displayingordeletingcertaintraces.

Youcanresizethe“NameofTrace”and“File”namefieldsbyselectinganddragging

Page 20: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

20

thesplitterbarbetweenthefields.

FileName:Thisfielddisplaysthefilenameandpathofthetrace.Ifthenameistoolongtoappearinthefield,youcanusethearrowkeystoscrolltotheendofthefilenameoryoucanmaximizetheTracePreferenceswindow.Ifthefilepathforatraceshows'notsaved',thetraceisunsaveddataandisonlystoredinmemory.TraceAuditLog:Thisfielddisplaysinformationaboutthetracethatishighlighted.Itincludesinstrumentparametersanddataforminformationrelatingtothetracethathasbeenselected.Whenmorethanonetraceishighlighted,thisdisplaysinformationonthetracethatwaslastselected.

SelectTraces:Thedrop-downarrowrevealsanycolumnsyouhavecreatedintheUserDataForm.Youcanselecttracesbasedonanyofthecolumnsyouhavecreated.

3) GraphPreferencesThisdialogboxallowsyoutochangetheappearanceofyourgraph.GraphPreferencesisaccessiblefromtheGraphmenu,byright-clickingonatraceorbyclickingGraphpreferencesicononthetoolbar.AllgraphswillappearlabeledwiththeXaxislabelbeinghorizontalandtheYaxisvertical.Youcannotchangetheorientationoftheselabels.However,youcanalterthefontandsizeoftheaxesandtheirlabels.Anychangesyoumaketotheappearanceofthegraphwillapplytoallgraphsuntilthesettingsarechangedagain.

StandardTab:Thistaballowsyoualterthewayyourgraphappearsonscreen.YoucanviewthechangesyoumakebeforeacceptingtheminthePreviewbox.PreviewBox:Apreviewofeachoftheaxesvariablesisshowntotherightofthisdialogbox.Tochangetheview,alterthesettingsinthevariablesontheleft.AxesColor:Fromthedrop-downlist,selectthecolorthatyouwanttoassigntotheaxesofthegraph(s).

AxesWidth:Fromthedrop-downlist,selectthewidthoftheaxes.Thewidthisgiveninpixelsandrangesbetweenoneandfivepixels.

GridStyle:Thisdrop-downlistallowsyoutospecifywhetherornotyouwantagridonyourgraph.Youcanalsodefinethetypeofgrid.PlotAreaColor:Thisdrop-downlistenablesyoutospecifywhat(ifany)coloryouwouldlikeasthebackgroundofthegraph.TraceWidth:Fromthisdrop-downlistyoucanselectthewidthofthetrace.Thewidthisgiveninpixelsandrangesbetweenoneandfivepixels.

Font:ThisbuttondisplaystheWindowsFontdialogboxthatallowsyoutodefinethefontstylefortheaxesofthegraph(s).Thestyleandsizeofthefontwillbeappliedtoallgraphsuntilthefontischangedagain.

TraceStyle:Thisfieldallowsyoutodefinethewaythatyourtracewillappearonscreen.

Page 21: User guide for Cary 5000 absorption spectrometer with ...mmrc.caltech.edu/Cary UV-Vis Int.Sphere/manuals/Cary5000...User guide for Cary 5000 absorption spectrometer with external DRA

21

Points EachindividualdatapointcollectedbytheCaryisshownonthegraph.

Pointsandsolidlines

EachindividualdatapointcollectedbytheCaryisshownonthegraph.Aninterpolatedlineisdrawnconnectingthepoints.

Solidlines TheCarysysteminterpolatesthedatabetweentheactual,collecteddatapointsandshowsthedataasacontinuousline.

Stylecyclesautomatically

TheCarysystemwillgiveeachtraceoneoftheabovestylesmakingiteasiertodifferentiatebetweentraces,especiallyonblack-and-whiteprinters.StylecyclingwillonlyoccurwhentheTraceWidthhasbeenselectedat1pixel.

AdvancedTab:Thistaballowsyoutofurtherrefinethewayyourgraphlooks.ForeachoptionyoucanmakeindividualchangestotheX,Yorbothaxes.

Autotickmarks:Usethesliderbarstoalterthenumberoftickmarks(graduations)thatyouwanttoappearonyourgraph.IfyouhaveselectedtheMajorTickscheckbox,thisoptionisunavailable.Majorticks:Selectthistodeterminehowmanytickmarks(graduations)youwanttoappearoneachaxisofyourgraph.

Decimals:Selectthistodeterminehowmanydecimalplacesyouwanttoassigntoeachtickmark(graduation)onyourgraph.

ShowLegend:Selectthistodisplayalegendontherightsideofeachgraphbox.Asingleclickonasamplenameorexamplelinewithinthelegendwillmakeitthefocusedorredscan.InformationaboutthetracewillbedisplayedbelowtheXaxis.

Double-clickingthelegendwillautomaticallydisplaytheTracePreferenceswindow,allowingyoutoviewmoredetailedinformationaboutyourtrace(e.g.,auditlog,filename,etc.).