Submitted by:Prathibha Saseedharan
EKAHEBM046S7 BME
Guided by: Jibin Jose
Parallel Image Reconstruction in MRIUsing Wavelet Transform
-
8/3/2019 Univ Seminar
2/60
CONTENTS.Introduction.Image formation.Introduction to k-space.
Parallel image reconstructionWAVELET-REGULARIZED RECONSTRUCTION FOR
RAPID MRIAUTOCALIBRATED REGULARIZED PARALLEL MRI
RECONSTRUCTION IN THE WAVELET DOMAINConclusion
-
8/3/2019 Univ Seminar
3/60
INTRODUCTION
MRI is a non-invasive medical procedure.Nothing is inserted in a patients body, no dyes are
swallowed, and no contrast agents are injected, except under
special circumstances.Moreover, patients are not exposed to ionizing radiation,as
is the case with X-ray Computed Tomography (CT)
imaging.
-
8/3/2019 Univ Seminar
4/60
A powerful magnetic field aligns the nuclear
magnetization of hydrogen atoms in water in the body.
Radio frequency (RF) fields are used to systematicallyalter the alignment of this magnetization.
-
8/3/2019 Univ Seminar
5/60
This causes the hydrogen nuclei to produce a rotating
magnetic field detectable by the scanner.
This signal can be manipulated by additional magnetic fields
to build up enough information to construct an image of the
body.
The magnet first aligns hydrogen atoms that come within its
field, and then a radio-frequency pulse is applied to jostle
them momentarily.
-
8/3/2019 Univ Seminar
6/60
As they realign, special receivers pick up their signals and
transmit that information into computers in which special
programs convert those signals into vivid images.
The various tissues and fluids are distinguishable from
one-another largely because the concentration of hydrogen
varies within different tissues and bodily fluids
-
8/3/2019 Univ Seminar
7/60
Instrumentation.MRI scanner consists of a magnet, 3 magnetic field gradient
coils and an RF coil.
Magnet polarizes the protons in the patient, produces a
homogeneous magnetic field within patient.
Either permanent or resistive magnets are used for low
magnetic fields like 0.35T.Permanent magnet systems are made of cobalt-samarium.
Resistive magnets are created by passage of current through
copper.
-
8/3/2019 Univ Seminar
8/60
Current MRI scanners produce a field of 1.5T or more
by using superconducting magnets .
Made out of a set of 4 or 6 solenoidally woundsuperconducting wires.
Wires are made of niobium-titanium alloy.
Field is homogenised by using ferromagnetic blocks
and electrically fed resistive coils.
-
8/3/2019 Univ Seminar
9/60
Gradient coils- generation of magnetic field gradient so that
resonant frequencies within patient are spatially dependent.
Fitted directly inside the bore of cylindrical magnet.
3 separate gradients are required to encode x,y,z dimension
of image.
Maxwell pair produces a linear variation in Bo along z-axis.Production of linear gradients along x and y axes are
performed by Golay coils.(saddle coils)
-
8/3/2019 Univ Seminar
10/60
-
8/3/2019 Univ Seminar
11/60
RF Coils : produces oscillating magnetic field necessary for
creating phase coherence b/w protons.
Also receives MRI signal via faraday induction.
Volume coils irradiate the whole body or just one specific
anatomical region.
Surface coils especially phased array coils are good in
imaging organs which lie close to the surface as a good SNR
is achieved.
-
8/3/2019 Univ Seminar
12/60
-
8/3/2019 Univ Seminar
13/60
MR image is a 2-dimensional signal composed of
many distinct coded elements called pixels.
Arrangement of pixels in a planar slice of tissue gives
an array called matrix.
Volume of a pixel is called as a voxel.To accomplish MR imaging the 3 gradient coil pairs
must be timed for i.) slice selection , ii.) phase
encoding, iii.) freq encoding and k-space formalism.
-
8/3/2019 Univ Seminar
14/60
-
8/3/2019 Univ Seminar
15/60
Frequency and PhaseFrequency and Phase
== tt
The spatial information of the proton pools contributing MR signal isThe spatial information of the proton pools contributing MR signal is
determined by the spatial frequency and phase of their magnetization.determined by the spatial frequency and phase of their magnetization.
-
8/3/2019 Univ Seminar
16/60
-
8/3/2019 Univ Seminar
17/60
Slice selection. Clinical MRI studies acquire a series of slices through anatomical area
of interest.
Slice selection is accomplished using a freq selective RF pulse applied
simultaneously with one of magnetic field gradients denoted as Gslice.
Coronal, axial or sagittal slice selections corresponds to sections in y, z
or x directions.
If the selective RF pulse is applied at a freq ws with an excitationbandwidth of ws, then protons precessing at frequencies b/w
ws+ws and ws-ws are rotated into the transverse plane and those
with resonant frequencies are not affected and remain in the z-
direction.
-
8/3/2019 Univ Seminar
18/60
The thickness T of the slice corresponding to protonsthat are affected by the RF pulse is determined by T = 2ws/ Gslice.Slice thickness can be increased either by decreasing
the strength of Gslice or increasing the freq bandwidthof excitation pulse.
A longer RF pulse results in a narrower freq spectrumand therfore a thinner slice for a given value of Gslice.If the direction of Gslice is denoted by z,then
sl(z) = Gz.z. /2
where = duration of pulsez= proton position within slice
-
8/3/2019 Univ Seminar
19/60
Mathematically, signal from precessing magnetizationafter slice selection can be represented as
S slice slice p (x,y) dxdy
Where p(x,y) is the no: of protons at positions (x,y)within the body and is called the proton density.
-
8/3/2019 Univ Seminar
20/60
Phase- encodingHaving selected a slice, the other 2 dimensions must
be encoded to produce a 2-dimensional image.One of these directions is encoded by imposing a
spatially dependent phase on the signal fromprecessing protons.Other by creating a spatially dependent precessional
freq during signal acquisition.
Phase is encoded by a gradient turned on or off beforedata acquisition begins. S(Gy, pe) = slice slice p(x,y) e-j Gy pe dxdy
-
8/3/2019 Univ Seminar
21/60
Frequency encodingFreq encoding gradient Gfreq is turned on during data
acquisition.Assuming that Gfreq is applied in the x direction and
considering only the effect of this freq-encodinggradient the acquired signal is
s(Gx,t) sl sl p(x,y) e-jwxt dx dy =
slsl p(x,y) e-j Gx.xt dx dy
-
8/3/2019 Univ Seminar
22/60
-
8/3/2019 Univ Seminar
23/60
k-space is a formalism widely used in magnetic resonance
imaging independently introduced in 1983 by Ljunggren and
Twieg.
Simply speaking, k-space is the temporary image space in
which data from digitized MR signals are stored during data
acquisition.
When k-space is full (at the end of the scan), the data are
mathematically processed to produce a final image.Thus k-space holds raw data before reconstruction.
-
8/3/2019 Univ Seminar
24/60
k-space is in spatial frequency domain. Thus if wedefine kFE and kPE such that
-
8/3/2019 Univ Seminar
25/60
where FE refers tofrequency encoding, PE tophase encoding, tis the sampling time (the reciprocal of sampling frequency), is
the duration ofGPE, (gamma bar) is the gyromagnetic ratio, m is
the sample number in the FE direction and n is the sample numberin the PE direction (also known aspartition number),
the 2D-Fourier Transform of this encoded signal results in a
representation of the spin density distribution in two dimensions.
-
8/3/2019 Univ Seminar
26/60
k-space has the same number of rows and columns as the final image.
During the scan, k-space is filled with raw data one line per TR
(Repetition Time). Although a strict mathematical proof does not exist and
counterexamples can be provided, in most cases it is safe to say that
data in the middle ofk-space contain the signal to noise and contrast
information for the image, while data around the outside of the image
contain all the information about the image resolution.
-
8/3/2019 Univ Seminar
27/60
-
8/3/2019 Univ Seminar
28/60
k-space information is somewhat redundant
An image can be reconstructed using only one half of the k-
space,
Either in the PE (Phase Encode) direction saving scan time
(such a technique is known as half Fourier or half scan)
Or in the FE (Frequency Encode) direction, allowing forlower sampling frequencies and/or shorter echo times (such
a technique is known as half echo).
-
8/3/2019 Univ Seminar
29/60
Two Spaces
FTFT
IFTIFT
k-spacek-space
kkxx
kkyy
Acquired DataAcquired Data
Image spaceImage space
xx
yy
Final ImageFinal Image
-
8/3/2019 Univ Seminar
30/60
-
8/3/2019 Univ Seminar
31/60
Parallel MRI (pMRI) is a way to increase the speed of the MRI
acquisition by skipping a number of phase-encoding lines in the k-
space during the MRI acquisition.
Data received simultaneously by several receiver coils with distinct
spatial sensitivities are used to reconstruct the values in the missing k-
space lines.We focus on the minimizing of the presence of noise in the
reconstructed image and also on removing of the aliasing artifacts from
the reconstructed image (artifacts caused by skipping some phase-
encoding lines in the k-space during the acquisition).
-
8/3/2019 Univ Seminar
32/60
A k-space image is formed by measuring the retransmitted signal.
The k-space image corresponds to the image in the Fourier space.
The real image of the object is obtained by Fourier transform of the k-
space image (it resolves the correspondence of the frequency and
spatial position of the signal).
-
8/3/2019 Univ Seminar
33/60
2D FFT
====>
k-space image final image
-
8/3/2019 Univ Seminar
34/60
In MRI, signal is usually received by a single receiver coil with an
approximately homogeneous sensitivity over the whole imaged object.
In pMRI, MRI signal is received simultaneously by several receiver coils with
varying spatial sensitivity -> This brings more information about the spatial
position of the MRI signal.
-
8/3/2019 Univ Seminar
35/60
The task of pMRI is to speed up the acquisition in order to:
be able to image dynamic processes without major movement artifacts (i.e. reduce the
speed of the acquisition so the movement during the acquisition time does not cause
significant artifacts),
shorten the MRI acquisition time that could be very long.
The bottleneck of the MRI acquisition is the number of retrieved lines in k-space and the
time needed to acquire one line in k-space. In pMRI, only a fraction 1/M of k-space lines is acquired while preserving spatial
resolution.
The acquisition is M times faster.
It causes an aliasing in the images - M points from the original image overlaps over
themselves in the image with aliasing.
-
8/3/2019 Univ Seminar
36/60
Linear combination of at least M images with aliasing retrieved by
the coils with varying sensitivity is used to reconstruct the original
image
(the coil configuration is supposed to be suitable for pMRI
reconstruction - the coil sensitivities should be distinct, all parts of
the imaged slice should be covered by at least one coil with
reasonable SNR in this part of the slice).
The parameters of the reconstruction are estimated using the exact
knowledge of the coil sensitivities.
-
8/3/2019 Univ Seminar
37/60
Aliasing
=====>
+
Reconstruction
==========>
+
Aliasing
=====>
-
8/3/2019 Univ Seminar
38/60
WAVELET-REGULARIZED
RECONSTRUCTION FOR RAPID MRI
-
8/3/2019 Univ Seminar
39/60
What is wavelet analysis?A wavelet is a waveform of effectively limited duration that
has an average value of zero and is of varying freq.Fourier analysis decomposes a signal into sine waves of
various frequencies.
Similarly, wavelet analysis breaks up a signal into shiftedand scaled versions of the original wavelet.
Signals with sharp changes might be better analyzed with anirregular wavelet than with a smooth sinusoid.
Wavelet analysis can be applied to one-dimensional data(signals), two-dimensional data (images) and, in principle, tohigher dimensional data.
-
8/3/2019 Univ Seminar
40/60
Advantages of this method.
By this method artifacts are significantly reduced
compared to conventional reconstruction methods.
Capable of recovering the missing k-space regions.Employs a non-linear approach and therefore blurring,
noise propagation, undersampling,aliasing are reduced.
Speeds up the reconstruction process.
-
8/3/2019 Univ Seminar
41/60
All problems related to reconstruction can be solvedby Daubechies Tl algorithm.Potential difficulty arises when forward model is
poorly reconditioned.This paper describes a TL algorithm specially tailored
to solve this problem.
-
8/3/2019 Univ Seminar
42/60
Method Consider a single receiving coil with homogeneous sensitivity. Thecorresponding model for the complex time-varying MR signal is
where " is the unknown proton density map to be recovered and k(t)denotes the so-called k-space trajectory.
In order to perform a numerical reconstruction, we must provide a
discretized version of the forward model .Time is sampled at N instantsresulting in the k-space samples {kn}. The signal to be reconstructed is represented as a linear combination of
basis functions that are shifted versions of a generator # on a finiteCartesian 2-D grid Cs:
-
8/3/2019 Univ Seminar
43/60
The signal is thereby parametrized by a set of Mcoefficients {c[p]}, represented as a vector c.
The term bn is introduced to represent both themeasurement noise and model mismatch. This model
is linear; thus there exists a N M matrix E such that:
-
8/3/2019 Univ Seminar
44/60
Variational formulation. The solution "c is defined as the minimizer of a cost function that
involves two terms: the data fidelity F and the regularization R that
favors solutions according to given prior knowledge. This is
summarized as
where the tuning parameter balances the effects of the two terms. F is chosen
as the square of the l-norm :
#m Ex#2 , which is justified when the noise is Gaussian.
The ill-conditioning, inherent to undersampled trajectories, imposes the choice of
an adequate regularization term R.
TV reconstruction is related to the l-norm of the modulus of the gradient and is
an optimal regularization for piecewise-constant solutions.
-
8/3/2019 Univ Seminar
45/60
Wavelet regularization
The underlying idea of wavelet regularization is that naturalimages tend to be sparse in the wavelet domain.Based on the property that a small l-norm promotes sparsity,
this solution is defined as:
where W and W1 are the wavelet decomposition andsynthesis matrices, respectively.
-
8/3/2019 Univ Seminar
46/60
Principle of TL algorithmSolution of the simpler wavelet denoising problem (E is the
identity matrix and W is orthonormal) is a single-stepthresholding:
Daubechies algorithm can then be explained by
iteratively bounding the initial reconstruction problemby a simpler denoising problem.Specifically, at iteration step n, one defines the
auxiliary variable
Where the wavelet vector "cn specifies the currentestimate of the solution.
-
8/3/2019 Univ Seminar
47/60
A fundamental point for our algorithm is that the multiplicationwith the matrix EHE corresponds to a 2-D convolution (Block-Toeplitz matrix). Indeed, by defining the kernel
Note that the kernel G has a support twice as large as Cs in eachdimension.The matrix-vector multiplication with EHE correspondsto the
most computer-intensive part of Algorithm. Based on the aboveproperty, we implement this operation by a pointwisemultiplication in the frequency domain using FFTs on a gridtwice as large as Cs .
The advantage is that this computation is exact while it avoidsthe use of regridding.
-
8/3/2019 Univ Seminar
48/60
-
8/3/2019 Univ Seminar
49/60
For the linear reconstruction, a Conjugate Gradient (CG) loopis used,
with a tolerance fixed to 1e 8.
The TV reconstruction was implemented using the Iterative Re-
weighted Least Square (IRLS) method with 10 outer iterations. For the linear solver, CG with a tolerance 1e 8 is applied
For the wavelet reconstruction, we chose the Haar basis, which is the
simplest and fastest wavelet transform. 3 decomposition levels wereconsidered and cycle-spinning is used as in to avoid blocking artifacts.
-
8/3/2019 Univ Seminar
50/60
AUTOCALIBRATED REGULARIZED PARALLEL MRI
RECONSTRUCTION IN THE WAVELET DOMAIN
-
8/3/2019 Univ Seminar
51/60
IntroductionTo reduce scanning time in MRI parallel acquisition
techniques with multiple coils were used.This is usually done using the SENSE reconstruction
method.SENSE (Sensitivity Encoding) have been developed in order
to unfold the aliased registerd images in the k-space and inthe image domain, respectively.This method is supposed to achieve an exact reconstruction
in the absence of noise.
An array of multiple, simultaneously operated receiver coilsis used for signal acquisition.The array elements are usually surface coils, which exhibit
strongly inhomogeneous, mutually distinct spatial
sensitivity.
Sensitivity encoding makes it possible to reduce the density
-
8/3/2019 Univ Seminar
52/60
Sensitivity encoding makes it possible to reduce the density
and, consequently, the number of these steps.
In the widely used k-space view, reducing phase encodinginthis fashion means that the same k-space area is sampled by
fewer, more widely spaced readout lines.
The factor by which the number of readout lines is reducedis referred to as the reduction factorR.
In conventional image reconstruction, such reduced phase
encoding approach with multiple-coil acquisition, however,
permits the reconstruction of a full-FOV image without the
aliasing effect.
-
8/3/2019 Univ Seminar
53/60
-
8/3/2019 Univ Seminar
54/60
-
8/3/2019 Univ Seminar
55/60
Consequently, we propose to look for an image representation where
these localized transitions can be easily detected and hence attenuated.
The WT has been recognized as a powerful tool that enables a good
space and frequency localization. The statistics of the wavelet coefficients can also be easily modelled
allowing us to efficiently employ a Bayesian framework for the
estimation procedure.
W d fi th lti l t ffi i t fi ld f th
-
8/3/2019 Univ Seminar
56/60
We define the resulting wavelet coefficient field of thetarget image by = (a,m)m, (h,j,m, v,j,m, d,j,m)1jjmax,m
where a,m denotes an approximation coefficient atresolution level jmax and location m and o,j,m with o
{h, v, d} denotes a detail coefficient at resolution
level j, location m and orientation o which may bevertical, diagnol or horizontal.
We aim at building an estimate of from d. Then, anestimate of the objective image is easily derived by justapplying the inverse WT operatorT to .
-
8/3/2019 Univ Seminar
57/60
O ti i ti l ith
-
8/3/2019 Univ Seminar
58/60
Optimization algorithmThe goal of this algorithm is to iteratively compute a field of
coefficients that minimizes J .For doing so, we will use the concept of proximity operator which
was found to be fundamental in a number of recent works in convex
optimization.
Note that blurring effects in the Tikhonov regularized image are no
longer present in the WT regularized image.
Moreover, the aliasing artifacts in the basic-SENSE reconstructed
image are significantly smoothed with the proposed wavelet-based
approach but they are completely removed only if they were not
-
8/3/2019 Univ Seminar
59/60
CONCLUSION.Both these methods of reconstruction reduces aliasing
artifacts in data and improves the image .
-
8/3/2019 Univ Seminar
60/60