Ultrasound Potential PDF Ppt
Transcript of Ultrasound Potential PDF Ppt
RSNA 2008
Ultrasound Bioeffects
Dolores H. Pretorius, MDThomas R. Nelson, Ph.D.
Professorsof Radiology
University of California. Son Diego
Ang ESBC. Gluncic V. Duque A. Schafer ME. Rakic P,Prenatal exposure to ultrasound waves impactsneuronal migration in mice, Proceedings of the NationalAcad of Sci, August 2006; 103: 12903-12910. Links
335 mice
Ultrasound for 30 - 420 min
A small number of neurons failed to acquire theirproper position
They remain scattered within inappropriate corticallayers
Consequences such as epi lepsy, schizophrenia andautism
~gcts in Ani as
Lung hemorrhage in young mice andneonatal/adult pigsIntestinal hemorrhage in adult miceBleeding near developing bone in youngmice
AlUMConsensus Report on Potential Bioeffects of DiagnosticUltrasound. JUM 2008: 27:503-515
Ultrasound Bioeffects
_~~W.arnina
Concerns have been raised in the past relatedto. Autism.Abnormal hearing, vision or language
development. Intrauterine growth retardation. Childhood cancer. Increase in non-right handedness
Abramowicz J, UOG 29:363, 2007
Approximately 250 million fetal ultrasoundexaminations performed per year in US.To date no adverse effects on fetusreportedFetal ultrasound considered to be safeUltrasound exposure can producebioeffects at diagnostic levels inlaboratory studiesValid experimental studies difficult toperformCaution advises following ALARA principle
Ultrasound Bioeffects: Reasons for Concern
Ultrasound is mechanical energy:. thermal effects. mechanical effects
Energy level depends on machinesetting
. MI - mechanical index
. TI - thermal indexBioeffects studies based on oldlimits - 94 mW/cm2New equipment limits much higher -720 mWlcm2
Use As Low As ReasonablyAchievable technique
Ultrasound Bioeffects: Background1 ,
TI is an on-screen measure of the potential fortissue heatingMI is an on-screen measure of the potential toinduce cavitation in tissues
3D ultrasound does not introduce additionalsafety considerations4D ultrasound with continuous exposure offerspotential to prolong examination times
ClinICal S<>f~fy Slote_ntfO<' Di<Ig"".tk
Ulfr<tlound, ECMUS (Europull Co",,,,ittu 01 M.d;cool Ull'NsoYnd$ofety) (2OQt.)
Ultrasound Bioeffects: Some Ph sics
Medical ultrasound basic properties:. frequency 1 - 15 MHz. pulse length 3-5 cycles. speed of sound in tissue ~ 1540 mls. attenuation ~ 1dBI (em-MHz). wavelength & resolution ~ 0.5 mm. image rates up to 150 fps (30 typical)
2.5 MHz 3.5 MHz 4.0 MHz 5.0 MHz 6.0 MHz
Ultrasound Bioeffects: Background
Ultrasound has diagnostic value
Wider range of US studies and new technologies havinghigher acoustic output levels in more patients
Current acoustic output much greater than earlierequipment
No evidence that diagnostic ultrasound produces harm
Subtle or transient effects not well understood
Diagnostic ultrasound should be used prudently
Ultrasound examinations should only be performed bytrained. competent personnel
Essential to maintain vigilance to ensure continued safety
Ultrasound Bioeffects: Some Physics
Characteristics of SoundSound is mechanical energyPropagates longitudinally through elastic media. Alternating zones of compression and
rarefactionUltrasound imaging typically uses short pulsesEnergy is reflected at interfaces
Ultrasound Bioeffects: Some Physics
Doppler.Pulses reflecting off moving interfaces, (e.g. blood cells, heart valves
or contrast) exhibit a phase shift that can be used to measure thevelocity of motion along the path of the sound beam.
.Typical Doppler shift range is 10 to 1000 Hz
Ultrasound Bioeffects: Some Ph sics
Velocity Doppler. computes the velocity in each pixel and displays acolor whose hue depends on the direction and whosesaturation depends on the velocity component measured
Ultrasound Bioeffects: Some Ph sics
Ultrasound Contrast Materials. Improved visualization of vessels and tissues is possible
using gas-filled micro-bubbles to produce a largeimpedance discontinuity. Contrast agents offer
promise in enhancingmasses, visualizing bloodflow, measuring perfusion,and delivering drugs andgenetic agents to specificsites
Ultr~sound Bioeffects: Some Ph sics
Focused Excitation
..1I3"~! Weak Focus.---.-.-.-.-.-.-.-.-.-.-.-.-.---.-.-.-.-.-
Ultrasound Bioeffects: Some Physics
Power Doppler. computes the integral of the entire velocity distribution
and displays the magnitude as a color and brightness value. has less angular dependence and better sensitivity toslow flow
Ultrasound Bioeffects: Some Physics
Beam PropertiesUltrasound propagates as longitudinal wave
Two primary zones
Near field (Fresnel) - Complex field pattern
Far field (Fraunhofer) - More coherent field pattern
Tlill\sduc 1\t'nl l.mgthQfnu,f'eld
Diar.wI"",d ':;'.. d'/4;.
BeAm <!; 'IJPf1c..
:;I1I"''''I.:2;',!<1
Ultrasound Bioeffects: Some Physics
Image formation. Scan lines sweeping through different directions are used to producea two-dimensionalimage
Ultrasound Bioeffects: Some Physics
A ~,+ I
~ I -
~
..
.,
I
!Io-_!
ij&rG.._
i
c
~'IsO'""p",,","p;l
".\
~ \,~
J'._'._.~-ISP<dlid AVI'!!;~ye(SA,I
D,,,,t;J:1c,,"-::f(.'$,,,N!<lrn
Ultrasound Bioeffects: Some Physics
ISAPA = IsATA/duty cycle. IpA = ITA/duty cycle
ISPTA = ISATA (Isp/IsA). Indicator of thermal effectsISPPA=ISATA(Isp/IsA)/duty cycle
Indicator of potential mechanical bioeffects andcavitation
ISPPA > ISPTA > ISAPA > ISATA
SA = Spatial AverageSP = Spatial Peak
PA = Pulse AverageTA = Temporal Average
TP = Temporal Peak
Ultrasound Bioeffects: Some Physics
Attenuation
1.0
.~
~ 0.5.o"'"
The lowerthe frequency,
the morethe penetration
4.5
Penelr'otion depth, em
(Dist:~nce traveled =2 pellstlation depth)
Ultrasound Bioeffects: Some Physics
Ultrasound Bioeffects: Some Physics
Acoustic Power
Rate of energy production, absorption or flow. SI unit - Watt (1 J oule/sec)
Acoustic Intensity.Rate of energy flow through a unit area
(Watts/cm2)
Power OutputOutput power
PRF = pulsed repetition frequencyFrequencyOperating mode
Ultrasound Bioeffects: Some Physics
100
10
.5
~~.~
E 01
001
Attenuation
Ultrasound Bioeffects: Some Physics
Ultrasound Bioeffects: Some Physics
SPTA =Spacial peak temporal averageSPPA =Spacial peak pulse average
Ultrasound Bioeffects: Some Physics
10 100 1000Expo$ure time, minutes
10000
Ultrasound Bioeffects: Some Physics
FDA Output Limits. ISPTA< 100 mW/cm2. ISPTAcan exceed 1000 mW/cm2 for Doppler
IndicesThermal Index (TI). Ratio of acoustic power produced to power required
to raise tissue by 1° C. TIs - soft tissue, TIb - bone, TIc - cranium
Mechanical Index (MI). Estimate of likelihood of cavitation. Related to intensity
Ultrasound Bioeffects: Recommendations=p~
"'"'' ""
Nelson TR, Fowlkes B, Abramowicz. Ultrasound Biosafety Considerations forthe Practici Son ra her/Sonol ist. JUM, 2008.
Ultrasound Bioeffects: Thermal Hazard
Thermal hazard exists with some equipment if usedimprudently
Temperature elevation < 1.5° C does not present harm toembryo
Temperature elevation> 1.5° C may cause harm
Temperature elevation> 4° C for 5 minutes potentiallyhazardous to embryo
Spectral pulsed Doppler modes can reach these levels
Color Doppler modes may reach these levels
TI provides rough guide to sonographer/sonologistregarding thermal hazard
Ultrasound Bioeffects: Non-thermal Hazard
Non-thermal hazard exists with some equipment if usedimprudently
Damage demonstrated in tissues with gas pockets at MI >0.3
Avoid unnecessary exposure to neonatal lung
Keep MI as low as possible
MI provides rough guide to sonographer/sonologistregarding non-thermal hazard
Contrast agents increase probability of cavitation
Single beam modes (A-mode, M-mode and spectral pulsedDoppler) more likely than scanned modes (B-mode, ColorDopp ler)
BMU5$<1fctyStGtc...n'(2000)
Ultrasound Bioeffects: Ultrasound Contrast Aqents
Ultrasound contrast agents typically are stable gas-filledmicrobubblesPotentially produce cavitation and micro-streamingMicrovascular damage or rupture is possible (animalstudies)Premature ventricular contractions induced in contrastenhanced echocardiography with high MI and triggeredexcitationRisk increases with increasing MI
Ultrasound Bioeffects: Literature Studies
pulsed wave Doppler TI was 1.5 ! 0.5: range 0.9-2.8
color flow Doppler TI was 0.8 ! OJ: range 0.6-1.2Significantly greater than
B-mode sonography TI was 0.3 ! OJ: range 0.1-0.7 (p<.01)
190 B-mode MI variations 1.1 ! 0.1
31 color flow Doppler variation I.O! 0.1 (p=.09)
190 pulsed wave Doppler MI variations 0.9!0.2 (p<.OOI).
MI = 1.0 and Tis = 0.2
Ultrasound Bioeffects: Preanancy Exposure
Embryo/fetus is particularly sensitive
Little information available regarding possible subtlebiological effects at diagnostic levelsTemperature rise greatest at bone surfaces andadjacent soft tissuesIncreasing mineralization raises possibility of heatingsensitive tissues such as brain and spinal cordCare should be taken to minimize eye exposureImportant to limit exposure time and the MI and TIcommensurate with acceptable diagnostic evaluationNo reason to withhold diagnostic scanning duringpregnancy provided it is medically indicated and is usedprudently by fully trained operators
Ultrasound Bioeffects: Literature Studies
Sheiner et.al. Increased TI can be achieved during Doppler OBSonography. J Ultrasound Med 2007: 26:71-76
Patients with suspected fetal growth problems undergoing Dopplerstudies of the fetal circulation in addition to B-mode sonography
Examinations took place between 21 and 40 weeks' gestation
HDI 5000 scanners (Philips Medical Systems, Bothell, WA)
transvaginal (4-8 MHz) probe. transabdominal (2-5 MHz) probe
63 examinations were evaluatedgestational age was 31.6 ! 5.1 weeks
total examination time was 17.6 ! 8.6 minutesDoppler examination time was O.9:tO.8 minutes
Ultrasound Bioeffects: Literature Studies
Increased acoustic output levels, as expressedby TI levels, are reached during obstetricDoppler studiesTI levels may reach 1.5 and higher
Doppler procedures should be performed withcautionDoppler procedures should be as brief aspossible during obstetric sonography
Ultrasound Bioeffects: Literature Studies
Color and pulsed Doppler should be avoidedin the first trimester of pregnancy due topotential for tissue heating
AILIM Cons sus Report on Potential Bioef; ts of DiagnosticLJltrasound. JLlM 2008: 27;503- 15
Ultrasound Bioeffects: Literature Studies-~
11first-trimester14 second-trimester12 third-trimester examinations were evaluatedFirst-trimester examination was 8.9 minutes. MI was 0.73; range 0.3-1.3. TI was 0.34; range 0.1-1.7
Second-trimester examination was 31.8 minutes. MI was 1.04; 0.5-1.5. TI was 0.28; 0.1-2.4
Third-trimester examination was 16.3 minutes. MI was 1.06; 0.2-1.5. TI was 0.32; 0.1-2.4
Ultrasound Bioeffects: Literature Studies
Output levels during routine obstetric ultrasoundexaminations, as expressed by the MI and TI,are generally lowHowever, higher output levels, particularly TIlevels of greater than 1.5, can be achievedTI greater than 1.5 typically account for only asmall proportion of the examination timePrudent use requires ALARA principles. Minimize study time consistent with diagnostic
goals
Ultrasound Bioeffects: Literature Studies
Sheiner et. al. Acoustic output as measured by MI and TIduring routine OB exams, J Ultrasound Med 2005;24:1665-1670
Quantify acoustic output of clinical ultrasoundinstruments during routine obstetric examinations. thermal index (TI). mechanical index (MI)
Prospective, observational studySonographers were unaware of the data being sought
HDI 5000 scanners (Philips Medical Systems, Bothell,WA). transvaginal (4-8 MHz) probe. transabdominal (2-5 MHz) probe
Ultrasound Bioeffects: Literature Studies
Statistical significance existed acrosstrimesters with regard to examination durationsand MI (P < .001)No statistical significance existed in the TIacross trimesters3.5~o of third trimester examinations had a TIof greater than 1.0. 2.4 ~owere between 1.0 and 1.49. 1.1~owere greater than 1.5
TI ~ 1 changes were brief (0.17 :t 0.08 minutes)TI changes were observed during color Dopplerimaging
er
Tricuspid valve interrogation for aneuploidy risk
Requires certification
Should be done rapidly, by experienced sonographers
Ultrasound Bioeffects: Literature Studies
Sheiner et.al., What Da Clinical Users KnowRegarding Safetyof Ultrasound During Pregnancy? J Ultrasound Med 2007;26:319-325
determine end users' knowledge regarding safety aspects ofdiagnostic ultrasound during pregnancy
A questionnaire was distributed to ultrasound end usersattending review courses and hospital grand rounds betweenApril and June 2006
130 end users completed the questionnaires (63~o responserate).63% were physicians81.7~o were obstetricians-18~o of participants routinely performed Doppler ultrasoundexaminations during the first trimester
Ultrasound Bioeffects: Literature Studies
-70/0 of users disapproved of"keepsake/entertainment" ultrasoundKnowledgeable users support limitations onthe number of ultrasound examinations inlow-risk pregnancies. Concern regarding possible adverse
effects during pregnancy
Sheiner et.ol., J Ultrasound Med 2007__26:319-325
Ultrasound Bioeffects: Guidelines for Safe Use
Doppler for fetal heart monitoring. The power levelsused for fetal heart monitoring are sufficiently low thatuse of this modality is not contra-indicated, on safetygrounds, even when it is to be used for extended periods.
Peripheral pulse monitoring. The output from CW Dopplerdevices intended for monitoring peripheral pulses issufficiently low that their use is not contra-indicated, onsafety grounds.
Transcranial ultrasound investigations Transcranialultrasound investigations may require higher acousticoutput than other applications. Safety Guidelines shouldbe applied for both imaging and stand-alone Dopplerequipment. TIC should be monitored.
Ultrasound Bioeffects: Literature Studies
32.2/0 of the participants were familiar with theterm TI. only 17.7/0 correctly described the nature of
TI~22/0 were familiar with the term MI. only 3.810 correctly described the nature of MI
8010 of end users did not know where to find theacoustic indices. Only 20.8/0 knew they are displayed on the
sonographic display
Sheiner et.ol., J Ultrasound Med 2007,' 26:319-325
Ultrasound Bioeffects: Literature Studies
Ultrasound end users are poorly informedregarding safety issues during pregnancy
Further efforts in the realm of education andtraining are needed to improve end userknowledge about the acoustic output of themachines and safety issues
That's why we are here today!
Thank 'Ou
Ultrasound Bioeffects: AIUM Guidelines
AIUM Practice Guideline for the Performance of an AntepartumObstetric Ultrasound Examination. (2003)http://www .aium.org/publicatio 051clinicall obstetri cal.pdf
Diagnostic ultrasound studies of the fetus are generally consideredto be safe during pregnancy.This diagnostic procedure should be performed only when:
there is a valid medical indicationthe lowest possible ultrasonic exposure setting should be used togain the necessary diagnostic information under the as low asreasonably achievable (ALARA) principle.
The promotion, selling, or leasing of ultrasound equipment for making'keepsake fetal videos' is considered by the US Food and DrugAdministration to be an unapproved use of a medicol device.Use of a diagnostic ultrasound system for these purposes, without aphysician's order, may be in violation of state laws or regulations.
Ultr(ls0':l.nd B.i.<>effects: Background
This Output Display Standard (ODS) consists of the thermalindex (TI) and the mechanical index (MI)
The TI expresses the potential for a rise in temperature atthe ultrasound's focal point.The MI indicates the potential for the ultrasound to induceinertial cavitation in tissues.There are 3 TIs:
the TI for soft tissues (TIS), when the ultrasound beamdoes not impinge on bone, appropriate mostly for the firsttrimesterthe TI for bones (TIB), when the beam impinges on bone ator near its focus, which should be displayed in the secondand third trimestersTI for cranial bone (TIC), when the transducer is veryclose to the bone, such as when scanning in the adult.
Ultrasound Bioeffects: Guidelines for Safe Use
GUIDELINESFOR THE SAFE USE OF DIAGNOSTIC ULTRASOUNDEQUIPMENT,Safety Group of the British MedicalUltrasoundSociety, http://www .bmus.orglultras-safe ty Ius -safety04 .asp
Medical endorsement. Ultrasound should only be used for medicaldiagnosis if endorsed by a medical practitioner
Operator training. Diagnostic ultrasound procedures should becarried out only by persons who are fully trai ned in the use of theequipment, the interpretation of its results and images and in thesafe use of ultrasound, including an appreciation of its potentialhazards.
Awareness of machine factors influencing hazard. Operators shouldunderstand the likely influence of the machine controls, theoperating mode (e.g. B-mode, color Doppler imaging or spectralDoppler) and probe frequency on the thermal and cavitation hazards.
Ultrasound Bioeffects: Backqround
National Council on Radiation Protection and Measurements(NCRP). . NCRP report 140, Exposure Criteria for MedicalDiagnostic Ultrasound, II: Criteria Based on All KnownMechanisms. Bethesda, MD: NCRP; (2002)
Before 1976, there were no limits to the permissible acousticoutput from diagnostic ultrasound equipment.
In 1976, the US Food and Drug Administration beganregulating the output levels of machines to be no more than 94mW/cm2 spatial-peak temporal-average intensity for fetaluse.
In 1992, under pressure from manufacturers and end users,the Food and Drug Administration changed this limit to 720mW/cm2 but mandated that machines capable of producinghigher outputs display to the diagnostician some indication ofthe likelihood of ultrasound-induced bioeffects (OutputDisplay Standard)
Ult~(lsoun~.<>eff~cts: Background
The TI and MI are not perfect indicators of the actualthermal and non-thermal risks
At present, they should be accepted as the most sensiblemethods of risk estimation
The implementation of the ODS puts much greaterresponsibility for patient safety on the ultrasound enduser
The ALARA (as low as reasonably achievable) principlewas also recommended.
Nevertheless, for the acoustic indices to be meaningful,the diagnostician must be familiar with safety issues andimplications
In fact, one of the major recommendations of the ODSdocument was education of end users.
Ultrasound Bioeffects: Guidelines for Safe Use
Initial power setting. Machines should be set up so that the default(switch-on) setting of the acoustic output power control is low. If alow default setting cannot be achieved, a low setting should beselected after switching on. A low setting should be selected foreach new patient. The output should only be increased during theinvestigation if this is necessary to produce Qsatisfactory result.
Exposure time. The overall examination times should be kept asshort as is necessary to produce a useful diagnostic result.
Stationary probe. The probe should not be held in a fixed positionfor any longer than is necessary, and should be removed from thepatient whenever there is no need for a real-time image or spectralDoppler acquisition.
For example, using the freeze frame or cine loop facilities allowsimages to be reviewed and discussed without continuing theexposure.
Ultrasound Bioeffects: Guidelines for Safe Use
Probe self-heating. Endo-probes (e.g. vaginal, rectal or oesophagealprobe) should not be used if there is noticeable self heati ng of theprobe when operating in air.
applies to any probeparticular care should be taken if trans-vaginal probes are to beused to investigate a pregnancy during the first eight weekspost-conception.
Pre-existing temperature elevation. Particular care should be takento reduce output and minimise exposure time of an embryo or fetuswhen the temperature of the mother is already elevated.
Sensitive tissues. Particular care should be taken to reduce the riskof thermal hazard when exposing the following to diagnosticultrasound:
an embryo less than eight weeks after conception:the head. brain or spine of any fetus or neonate:an eye (in a subject of any age).
Ultrasound Bioeffects: Guidelines for Safe Use
Thermal and Mechanical Indices.MI> 0.3 There is a possibility of minor damage to neonatal lung orintestine. If such exposure is necessary, try to reduce the exposuretime as much as passible.MI> 0.7 There is a risk of cavitation if an ultrasound contrast agentcontaining gas micro-spheres is being used.
There is a theoretical risk of cavitation without the presence ofultrasound contrast agents.The risk increases with MI values above this threshold.
TI> 0.7 The overall exposure time (including pauses) of an embryo orfetus should be restricted in accordance with Table 1.
TI> 1.0 Eye scanning is not recommended, other than as part of afetal scan.
TI 3.0 Scanning of an embryo or fetus is not recommended, howeverbriefly
Ultrasound Bioeffects: Guidelines for Safe Use
Non-diagnostic uses of diagnostic ultrasound equipment. Examplesinclude repeated scans for training. equipment demonstration usingnormal subjects. and production of fetal souvenir pictures or videos
for equipment that display safety indicesthe TI should always be less than 0.5. the MI should always be less than 0.3
for equipment that do not display safety indicesTm~ should be less than I' CMIm~ should be less than 0.3.frequent exposure of the same subject is to be avoidedfollow safe scanning guidelines and ALARA
first trimester scans should not be carried out for:the sole purpose of producing souvenir videos or photographstheir prod.uction should not increase exposure levels or extendingthe scan times beyond those needed for clinical purposes.
Ultrasound Bioeffects: Guidelines for Safe Use1-------
Pulsed Doppler. The use of spectral pulsed Doppler, or color Dopplermode with a narrowwrite-zoom box selected, is not recommendedfor the investigation of any sensitive tissues, unless an e.stimate ofthe maximum likely temperature elevation has been obtained andconsidered in relation to the anticipated exposure time.
Thermal and Mechanical Indices. for machines which display on-screen thermal index (TI) and mechanical index (MI) values,operators should continually monitor their values and use controlsettings that keep them as small as is consistent with achievingdiagnostically useful results.In obstetric investigations, TIS (soft tissue thermal index) shouldbe monitored during scans carried out in the first eight weeks afterconception, and TIS (bone thermal index) thereafter.In applications where the probe is very close to bone (e.g. trans-cranial applications), TIC (cranial thermal index) should bemonitored.for eye scanning TIS should be monitored. In other applications,TIS should be monitored.
Ultrasound Bioeffects: Guidelines for Safe Use
Thermal and Mechanical IndicesWhere an on-screen thermal index (TI) or mechanical index (MI) isnot displayed, try to obtain worst case estimates (considering allpossible combinations of control settings) of temperature elevation('Tm~) and mechanical index (MIm~) for the particular probe andmode in use.If these can be obtained, assume that the MI value is equal to MIm"and the TI value is equal to 0.5 "Tm~ and refer to Table 1.
Table 1Maximum recommended exposure times for an embryo or fetus
TI Maximum exposure time (minutes)0.7 601.0 301.5 152.0 42.5 1
Ultrasound Bioeffects: Mechanical Index---The mechanical index (MI) is intended to offer a roughguide to the likelihood of the occurrence of cavitation.
The MI is constantly updated by the machine, accordingto the control settings, using the formula
MI= maximum value of peak negative pressure / pulsecentre frequency,
NB: the maximum value of peak negative pressureanywhere in the ultrasound field measured in water isreduced by an attenuation factor equal to that whichwould be produced by a medium having an attenuationcoefficient of 0.3 dB cm-l MHz-!.
Gestation Gestatian fram Title af Majar relevant
Fram LMP Canceptlan / Canceptus eventsFertilisation
0.14 days NIL--
14-28 days 0-14 days Zygate Rapid ce II
multiplicatio.ns
29:70 days 15-56 days Embryo. ()rganagenesis
4.1-10 weeks 2.1-8 weeks Emb a
10-11 weeks 8-9 weeks Fetus Ossificatlan of
.....~pi.~~ .~.!~.~~~.
13-14 weeks 11-12 weeks Fetus Ossification of
skull and lang
bone starts
Ultrasound Bioeffects: Thermal Index
The thermal index (TI) is intended to give a rough guideto the likely maximum temperature rise that might beproduced after long exposure.
Three forms of TI may be displayed, according to theapplication.
TIS assumes that only soft tissue is insonated.TIB assumes bone is present at the depth wheretemporal intensity is greatest.TIC assumes bone is very close to the front face ofthe probe.
However, note that errors in calculating TI values, andthe limitations of the simple models on which they arebased, means that TI values can underestimate thetemperature elevation by a factor of up to two.
Ultrasound Bioeffects: Sensitive Tissues
Ultrasound Bioeffects: Conclusions
Study results at our perinatal, academic medical centersuggest that 3DUS targeted studies are shorter than general2DUS studies, especially far complex cases.
Extended scanning to watch fetal movement has the potentialto increase fetal US exposure without benefiting diagnosis ormanagement and thus caution is needed for these procedures.
-----
Ultrasound Bioeffects: Sensitive Tissues
Up to eight weeks after conception, organogenesis istaking place in the embryo.
This is a period when cell damage might lead to fetalanomalies or subtle developmental changes.
The brain and spinal chord continue to develop through tothe neonatal period.
The presence of bone within the beam greatly increasesthe likely temperature rise
due to direct absorption in the bone itself
due to conduction of heat from bone to adjacenttissues.
Ultrasound Bioeffects: Conclusions1.....---....
Technological advances enhance fetal scanningopportunities by providing improved visualization andimage quality that potentially improves fetal visualization,diagnosis and management
Notably, 3DUS equipment provides clear images of thedeveloping fetus that are recognizable to family andphysician.
4DUS equipment further facilitates observing fetalmovements similar to real-time 2DUS imaging -potentially inviting longer fetal viewing.