The immunogenicity of therapeutic proteins-what you don't know can ...
Ultra-high-performance liquid chromatography for the characterization of therapeutic proteins
Transcript of Ultra-high-performance liquid chromatography for the characterization of therapeutic proteins
Accepted Manuscript
Title: Ultra-high-performance liquid chromatography for the characterization
of therapeutic proteins
Author: Szabolcs Fekete, Davy Guillarme
PII: S0165-9936(14)00183-6
DOI: http://dx.doi.org/doi: 10.1016/j.trac.2014.05.012
Reference: TRAC 14303
To appear in: Trends in Analytical Chemistry
Please cite this article as: Szabolcs Fekete, Davy Guillarme, Ultra-high-performance liquid
chromatography for the characterization of therapeutic proteins, Trends in Analytical Chemistry
(2014), http://dx.doi.org/doi: 10.1016/j.trac.2014.05.012.
This is a PDF file of an unedited manuscript that has been accepted for publication. As a service
to our customers we are providing this early version of the manuscript. The manuscript will
undergo copyediting, typesetting, and review of the resulting proof before it is published in its
final form. Please note that during the production process errors may be discovered which could
affect the content, and all legal disclaimers that apply to the journal pertain.
1
Ultra-high-performance liquid chromatography for the
characterization of therapeutic proteins
Szabolcs Fekete *, Davy Guillarme School of Pharmaceutical Sciences, University of Geneva, University of Lausanne, Boulevard d’Yvoy 20, 1211
Geneva 4, Switzerland
HIGHLIGHTS
UHPLC conditions were successfully applied to characterize therapeutic proteins
High-resolution separations are of prime interest in biopharmaceuticals analysis
IEX and SEC columns packed with sub-2-µm particles are commercially available
HILIC columns packed with small particles are useful for glycan profiling
The coupling of UHPLC with MS is highly informative
ABSTRACT
The pharmaceutical market has markedly changed over the past few years, and there are today
an increasing number of therapeutic drugs produced from biological sources. These
biopharmaceuticals include recombinant peptides, proteins, and monoclonal antibodies
(mAbs). Their detailed characterization could be difficult and time consuming, so it requires
powerful chromatographic and spectroscopic methods. In this context, the use of columns
packed with sub-2-µm particles at very high pressure, also known as ultra-high performance
(or pressure) liquid chromatography (UHPLC) has been reported as successful. Various
modes of chromatography are compatible with UHPLC columns and conditions, including
reversed-phase liquid chromatography (RPLC), size-exclusion chromatography (SEC), ion-
exchange chromatography (IEX) and hydrophilic interaction chromatography (HILIC).
Keywords:
Biopharmaceutical
Characterization
Hydrophilic interaction chromatography (HILIC)
Ion-exchange chromatography (IEX)
Mass spectrometry (MS)
Monoclonal antibody (mAb)
Reversed-phase liquid chromatography (RPLC)
Size-exclusion chromatography (SEC)
Therapeutic protein
Ultra-high-performance liquid chromatography (UHPLC)
* Tel.: +41 22 37 963 34; Fax: +41 22 379 68 08.
E-mail address: [email protected] (S. Fekete)
1. Introduction
The large majority of traditional pharmaceuticals are chemically-synthesized small-
molecule compounds. In addition to these “chemical” substances, there are a number of
substances that are produced from biological sources (i.e., biological systems or biological
molecules). These “biopharmaceuticals” include recombinant peptides, proteins or
glycoproteins [1,2]. The pharmaceutical potential of numerous proteins (e.g., interferons,
interleukins, and growth factors) that are naturally produced in the body was originally
Page 1 of 15
2
demonstrated more than 40 years ago. These molecules have obvious advantages, including
high efficacy, high specificity, wide therapeutic range, limited side effects, and exceptional
chemical and biological diversity. The clinical use of therapeutic proteins has enabled the
treatment of a wide range of life-threatening diseases, which were considered incurable or
untreatable only a few decades ago. Dozens of new drugs for the treatment of cancer, AIDS
and arthritis are on the market or are very close to regulatory approval [3]. Today, the global
protein therapeutics market is worth over $100b, thereby evolving towards a total
pharmaceutical market share of 20% [4,5]. We expect that, within the current decade, more
than 50% of the new drug approvals will be biologics [4,6,7].
Therapeutic proteins are large, heterogeneous and subject to a variety of enzymatic and
chemical modifications during expression, purification and long-term storage [5]. These
changes include several possible modifications, such as oxidation, deamidation,
glycosylation, aggregation, misfolding, or adsorption, leading to a potential loss of therapeutic
efficacy or unwanted immune reactions.
Because the development of biopharmaceuticals and biosimilars is especially complex,
regulatory agencies, such as the United States Food and Drug Administration (FDA) and
European Medicines Agency (EMA), require a demonstration of detailed characterization
(e.g., verifying primary structure and appropriate post-translational modifications, secondary
and tertiary structure), lot-to-lot and batch-to-batch comparisons, stability studies, impurity
profiling, glycoprofiling, determination of related proteins, excipients and protein aggregates
[8]. For this purpose, a single analytical technique is generally not sufficient, and a variety of
orthogonal methods are required to characterize such a complex sample fully, as summarized
elsewhere [9].
The primary structure of proteins can be identified with two reference techniques, namely
mass spectrometry (MS) and Edman degradation/sequencing [10]. The use of liquid
chromatography (LC) with tandem MS (LC/MS/MS) in de nova sequencing dominates the
process of sequencing proteins, and with peptide-mass fingerprinting combined with MS/MS
became the preferred techniques. Numerous spectroscopic techniques are available to assess
protein secondary and tertiary structure, such as X-ray crystallography, nuclear magnetic
resonance (NMR), UV/Vis spectrophotometry, fluorescence, circular dichroism (CD),
dynamic light scattering (DLS), static light scattering (SLS), differential scanning calorimetry
(DSC) and infrared spectrophotometry (IR) [11,12].
In addition to these methodologies, electrophoresis is also a key technique for protein
analysis, and different modes can be employed in slab gel or capillary format. Isoelectric
focusing (IEF) is able to distinguish charge differences (isoelectric point) among proteins
through the use of a pH gradient [13]. Capillary zone electrophoresis (CZE) has several well-
established attractive features for the characterization of such complex samples, including its
high resolving power and throughput. High-performance LC (HPLC) is another option for the
detailed characterization of proteins [14].
The three most common modes of chromatography are size-exclusion chromatography
(SEC), which separates proteins based on their size or hydrodynamic volume, ion-exchange
chromatography (IEX), able to separate proteins based on their charge, and reversed-phase
chromatography (RPLC), where separation occurs on the basis of the hydrophobicity of the
proteins. This last strategy offers a higher resolving power than SEC and IEX.
Today, one of the most widely-used analytical methods for therapeutic protein
characterization is LC, probably due to the impressive developments of the past few years,
enabling a new level of chromatographic performance. Recent developments in LC columns,
such as ultra-high-performance LC (UHPLC), packed with wide-pore superficially porous
particles (SPPs) and organic monolith columns allow a dramatic increase in the efficiency and
the resolution of protein separations, even with large, intact molecules.
Page 2 of 15
3
The aim of this article is to review the current trends in UHPLC and the potential for
UHPLC strategy for the characterization of therapeutic proteins. In this review, we focus
solely on unidimensional separations and particle-based, stationary-phase formats for routine
UHPLC applications.
2. Need for high kinetic performance in protein analysis
Higher separation efficiencies and throughput have always been of great interest in LC.
The pharmaceutical industry is interested in using rapid, efficient procedures for qualitative
and quantitative analyses to cope with the large number of samples and to reduce the time
required to deliver results [14]. When dealing with protein analysis, selectivity is often limited
so the only way to improve separation is to increase chromatographic efficiency. Moreover,
high-molecular-weight compounds, such as intact proteins, may have numerous
conformations, post-translational modifications, or multiple isoforms that can cause
broadened peaks and shifted retention times in the chromatograms. Another reason for
broadened peaks is the slow molecular diffusion of these compounds due to their large size.
Typical routine tasks are the separation of oxidized, deamidated or reduced forms of intact
proteins. Because the differences in molecular structure are small, similar retention behaviors
of the different forms are expected. In many cases, the selectivity cannot be improved and, as
a result, enhancement of separation efficiency must be considered. The stationary-phase
dimensions and morphology, and the mobile-phase temperature are the two most relevant
parameters for improving the efficiency of protein separations. Thanks to the latest stationary-
phase technologies, such as sub-2-μm porous particles, superficially porous particles (SPPs),
or wide-pore monolithic columns, the separation power was significantly increased in recent
years. In addition, it is also possible to extend the column length (e.g., coupling columns in
series) to achieve the required plate count or peak capacity. The use of elevated mobile-phase
temperature in the range 60–90°C further improves performance. At higher temperature, the
viscosity of the mobile phase decreases and the diffusivity of large proteins increases, leading
to sharp chromatographic peaks. The other benefit of elevated mobile-phase temperature is
reduction in adsorption of undesired proteins at the surface of the stationary phase.
3. Reversed-phase liquid chromatography (RPLC)
In RPLC, solute retention is predominantly mediated through the hydrophobic interactions
between the non-polar amino-acid residues of the proteins and the bonded n-alkyl ligands of
the stationary phase. Proteins are thus eluted based on their hydrophobicity. Large molecules
possess a so called “on-off” retention mechanism. Their retention strongly depends on minor
variations in the solvent strength, and a small change (<1%) in the organic modifier content
can lead to a significant retention change. For this reason, isocratic conditions are impractical,
and gradient elution mode is recommended. The efficiency of RPLC is superior to other
chromatographic modes and its robustness makes it well suited for use in a routine analysis
environment [5]. Mobile phases typically consist of water, acetonitrile or methanol and 0.1%
trifluoroacetic acid or formic acid (TFA, FA). The separation mechanism is based on a
combination of solvophobic and electrostatic interactions, the latter governed by the ion-
pairing interaction of TFA with basic side chains of a few amino acids (i.e., arginine, lysine,
histidine) and the N-terminus amino groups [14].
It was recognized very early that one of the best approaches to improving intact protein
RPLC separations would be to use small-particle packing material [15]. It came into practice
in the late 1990s, when very fine particles (sub-2-μm or sub-μm) could strongly improve the
separation performance, but required a huge increase in system pressure. To overcome the
Page 3 of 15
4
pressure limitations of conventional HPLC with a pressure limit of 400 bar, the research
groups of Jorgenson [16,17] and Lee [18] constructed prototype instrumentation and
experimental columns packed with very small non-porous material and performed analyses at
very high pressures (up to 7200 bar). New nomenclatures have appeared to describe this
higher back-pressure requirement, including ultra-high-pressure liquid chromatography, ultra-
high-performance liquid chromatography or very high-pressure liquid chromatography (i.e.,
UHPLC, UPLC, VHPLC or vHPLC).
The first commercial system for ultra-high pressure operation was released in 2004. It was
able to operate at pressures as high as 1000 bar, and the system was known as ultra-
performance liquid chromatography (UPLC). Since then, UHPLC systems compatible with
pressures up to 1200–1400 bar have become available from most HPLC manufacturers.
The use of such UHPLC systems and columns packed with highly-efficient sub-2-μm
particles became popular for the separation of small molecules very soon after their
introduction. At the beginning, only columns packed with standard pore-size particles (e.g.,
80–130 Å) were available, so the UHPLC approach was not amenable to separation of larger
proteins. In 2006, columns packed with wide-pore (300 Å) 1.7-µm particles were introduced
and opened up new avenues for protein analysis. These columns packed with sub-2-µm, fully
porous, hybrid particles (Acquity BEH300) were used with great success in protein and
peptide separations. Due to the intrinsic chemical stability of this hybrid-particle technology,
a wide pH range (pH 1–12) and high-temperature operation (up to 80°C) can be employed,
enabling a versatile, robust separation technology for method development. Since then,
several new generations of superficially porous, wide-pore materials and organic monoliths
have also been commercialized.
3.1. Non-porous versus fully porous particles
Non-porous and porous particles are the two main types of spherical packing materials
used for fast HPLC separation of biomolecules. The major difference between these two types
of particles is that the porous particles show higher resistance to the transparticle mass-
transfer process in the stagnant mobile phase inside the pores. Issaeva et al. demonstrated an
extremely high-speed separation of proteins and peptides using 1.5-µm non-porous silica
particles [19]. Non-porous particles can provide lower mass-transfer resistance and higher
efficiency than porous particles, but they have significantly lower surface area and sample-
loading capacities.
According to Wu et al., the loading capacity of 1.7-µm Acquity C18 porous particles is
~16.5 times larger than that of Micra C18 non-porous 1.5-µm particles [20].
Another issue is the very low retention achieved on non-porous particles compared to
totally porous particles. In spite of the early success of non-porous particles in protein
analysis, most of the applications are currently performed with sub-2 µm, fully porous
particles or using the new generation of superficially porous particles (SPPs). Today, two
providers (Waters and Agilent) offer sub-2-µm, fully porous, wide-pore (300 Å) materials for
reversed-phase protein separations. Those packings are available with C3, C4, C8, C18 and
diphenyl chemistries, offering a broad range of selectivity. There is another stationary phase
packed with 1.9-µm particles and possessing pore size of 175 Å (ThermoFisher Scientific)
that may also be of interest for the separation of proteins possessing molecular weights below
20 kDa [21].
A recent study systematically compared the peak capacity achieved on conventional 3–5-
µm wide-pore packings against 1.7-µm, fully porous, and a new generation of 3.6-µm, SPP,
wide-pore particles [21]. The last two types of column provided higher separation power,
especially for large proteins (e.g., myoglobin, bovine serum albumin, and filgrastim).
Page 4 of 15
5
Another study showed the possibilities of performing very fast separation (1.5–2 min) of
therapeutic protein variants using short columns of 50 mm packed with sub-2-µm porous
particles [22]. UHPLC was also used to separate a mixture of 10 protein standards. The
optimized method yielded improved chromatographic resolution, enhanced sensitivity, and a
three-fold increase in throughput. Subsequent cell-lysate analysis demonstrated no
compromise in chromatographic or mass spectral data quality at one-third of the original run
time [23].
A possible complication of UHPLC relates to the effect of pressure and mobile-phase
velocity (frictional heating) on the retention properties of large molecules. Dramatic changes
of retention with pressure were reported in a recent study [24]. An increase of retention factor
of around 150% for peptides (~ 1.3 kDa), 800% for insulin (~ 6 kDa) and up to > 3000% for
myoglobin (~ 17 kDa) was observed when increasing the pressure from 100 bar to 1100 bar.
The important effect observed for the isocratic elution of proteins probably relates to
conformational changes of the protein in addition to the effect of pressure on molecular size.
In gradient-elution mode, the pressure-related effects on protein retention were found to be
less pronounced, but still readily observable (an increase of apparent retention factor between
0.2–2.5 was observed).
3.1.1. Intact protein analysis (“top-down” approach)
In RPLC, it is possible to analyze therapeutic proteins in their intact forms. This approach
is often considered for impurity profiling or heterogeneity evaluation. This can be useful in
the biopharmaceutical field, to evaluate the stability and the purity of therapeutic protein
samples, as a release test, or to compare the chromatographic profile of biosimilar products.
Typically oxidized, reduced and deamidated forms of the native protein can be separated and
quantified in RPLC conditions.
Fig. 1 illustrates the analysis of native filgrastim (18.8 kDa) and its related
impurities/degradants (oxidized and reduced forms). As shown in Fig. 1, the baseline
resolution of these different forms can be obtained in less than 1.4 min using a 50-mm
column, at 80°C, and a reasonable flow rate. The resolution of these peaks was excellent,
demonstrating that high-throughput analysis can be carried out effectively for these
biomolecules.
In another study, the separation of the related oxidized and reduced forms of interferon α-
2A (19.2 kDa) were successfully performed in 5 min on a 150-mm long column packed with
1.7-µm fully porous particles [25].
UHPLC is useful for not only fast separations, but also performing high-resolution
separations by using long columns. Fig. 2 shows the gain in resolution upon conversion of a
conventional HPLC method on a column packed with 5-µm particles to a UHPLC method
using columns packed with sub-2 µm particles [5]. All the pre-eluting and post-eluting
variants of this 30-kDa protein were better resolved in the UHPLC method with a comparable
analysis time.
Coupling columns packed with sub-2-µm particles in series was recently applied to
improve peak capacity for the characterization of a 150-kDa intact monoclonal antibody
(mAb) [26]. In this work, a 150-mm-long column generated a peak-capacity value of nc = 62,
while the 30-cm and 45-cm-long columns provided nc = 95 and nc = 117, respectively.
Despite the resolution power being doubled with the 45-cm-long column, the baseline
resolution of the mAb from a few variants eluted prior to the main isoform could not be
achieved. Nevertheless, the separation cannot be further improved by simply increasing the
column length or gradient time because the experiments were performed at 80°C to limit
adsorption at the surface of the stationary phase. A gradient time longer than 20 min may be
detrimental for the stability of intact mAbs and should be employed with caution as
Page 5 of 15
6
documented elsewhere [27]. Finally, this improved analysis of intact mAbs could be useful
for comparing the chromatographic profile of between batches or with biosimilar products.
3.1.2. Analysis of protein fragments (limited proteolysis, “middle-down” approach)
Complete characterization of a large protein cannot be accomplished by the analysis of the
intact protein alone, so various enzymes, such as pepsin, papain, Lys-C and IdeS, are often
used to obtain relatively large fragments to facilitate the investigation of their micro-
heterogeneity. This limited proteolysis approach – or “middle down” – was used recently for
the characterization of mAbs [13]. Papain is primarily used to cleave mAbs into three
fragments, namely one crystallizable fraction (Fc) and two identical antigen-binding
fragments (Fab) of ~50 kDa each However, pepsin and IdeS generate F(ab′)2 fragments of
~100 kDa.
The reduction of disulfide bonds is also commonly used to produce two light-chain (LC)
and two heavy-chain (HC) fragments of 25 kDa and 50 kDa, respectively. To prepare only
25-kDa fragments, the reduced mAb (HC and LC) can be further digested with papain. This
process yields three main fragments of ~25 kDa, namely the native LC, the single-chain Fc
(sFc), and the Fab portion of the HC (Fd). This approach (generating ~ 25-kDa domains) is
generally known as the combinatory strategy (combination of intact-mAb reduction and
papain digestion) and is useful for the characterization and comparison of mAbs by RPLC
[28]. In most cases, the limited proteolysis approach can determine the heterogeneity of mAb
related to the different parts of the molecule.
Fig. 3 shows an example of a fast analysis of 25-kDa mAb (IgG1) fragments obtained after
limited proteolytic digestion approach. Several variants of the LC, sFc and Fd fragments can
be separated on a 150-mm-long column, applying a 6-min gradient at elevated temperature.
The inset on Fig. 3 displays the variability of the mAb related to the Fd domain. A generic
computer-assisted method-development approach was recently proposed for the RPLC
analysis of mAb fragments on sub-2-µm fully or superficially porous particles [29].
Outstanding separations within relatively short analysis times (typically 6–10 min) were
demonstrated on rituximab, bevacuzimab and panitumumab employing gradients of 30–40%
acetonitrile, temperatures above 70°C and 0.1% TFA as ion-pairing additive [29].
Fast, efficient separation of mAb-fragment variants in UHPLC conditions were reported in
several studies [5,8,25–29], suggesting that UHPLC is indeed a very powerful, promising tool
for characterization of mAbs. A recent study showed the advantages of the middle-down
approach versus the top-down approach to mAb analysis [30]. To improve structural analysis
of mAbs, a middle-down UHPLC-MS protocol was suggested using fast sample preparation
(about 1 h) followed by subsequent analysis of 25-kDa fragments. The reported results
showed that the application of this middle-down approach doubled the mAb-sequence
coverage previously obtained with top-down MS [30].
3.1.3. Peptide mapping (“bottom-up” approach)
Peptide mapping is commonly undertaken to determine the primary structure of proteins
after cleavage into proteolytic peptides. The power of peptide mapping lies in the large
number of site-specific molecular features that can be detected. When using one digestion
enzyme (e.g., trypsin), peptide mapping is typically carried out for protein identification.
When using multiple enzymes, peptide mapping is applied for the confirmation of a complete
amino-acid sequence, e.g., when confirming the amino-acid sequence of a biosimilar and
comparing it with the originating molecule. Historically, this approach is certainly the most
important one in RPLC analysis of proteins. However, peptide mapping is highly demanding
in terms of chromatographic performance, since samples are very complex, containing
Page 6 of 15
7
hundreds of peaks [5,31]. So far, many papers have been published on UHPLC peptide
mapping, but we discuss only a few key references in this review.
As an example, the determination of methionine oxidation in mAb – by peptide mapping –
was performed using BEH300 C18 1100 x 2.1 mm, 1.7-µm column [32].
In a similar study, peptide fragments generated after tryptic digestion were analyzed to
characterize the oxidation of mAb samples [33]. UHPLC coupled with electrospray ionization
time-of-flight MS (ESI-TOF/MS) and multivariate analysis were used to detect alterations in
the conformational states of the protein. For this study, a (100 x 1 mm) C18 column packed
with 1.7-µm particles was used.
Quantification and characterization of mAb deamidation was performed by Wang et al.
using UHPLC-MS after tryptic digestion [34].
An alanine replaced with a serine sequence variants was separated and identified in an
IgG4 monoclonal antibody by UHPLC-MS [35].
A new UHPLC peptide-mapping approach using ammonia as an mobile-phase additive
was reported by Liu et al. [36]. Compared to traditional methods using acidic mobile phase
additives (trifluoroacetic acid or formic acid), this method exhibited excellent overall
performance. It was particularly advantageous over the traditional approaches in terms of the
UV and MS sensitivities, and sequence coverage. Due to significant differences in the
chromatographic selectivity, supplemental peptide-mapping analysis could potentially be
carried out using basic mobile-phase conditions yielding “orthogonal” separations.
The most popular and widespread stationary-phase chemistries in peptide-mapping
analysis are C18, C8 and C4. The new CSH (Charged Surface Hybrid) phenyl-hexyl phase
was studied as an alternative bonded phase in reversed-phase peptide separation [37]. Peak
capacities of 150–200 were achieved on a 100-mm-long column in 8 min. Moreover, a
possible mathematical treatment of peptide maps was evaluated to improve resolution.
Applying power functions on raw chromatographic data can be a neat tool to improve peak
recognition and resolution [37].
3.2. Superficially porous particles as an alternative to fully porous and non-porous particles
Columns packed with fully porous particles have constraints in separation speed and
efficiency because of limitations in the stationary-phase mass transfer, resulting from the
relatively long diffusion times required for proteins to cross the porous structure [38]. The
concept of SPPs (often named shell, core-shell, fused-core, or partially porous) was first
applied by Horváth in the late 1960s [39,40]. They were initially intended for the analysis of
macromolecules, such as peptides and proteins. SPPs are made of a solid, non-porous silica
core surrounded by a porous shell layer having similar properties to that of the fully porous
materials conventionally used in HPLC. Later, Kirkland showed that 30–40-μm diameter
SPPs provided much faster separations compared with the large, fully porous particles that
were used earlier in LC [41]. The rationale behind this concept was to improve the column
efficiency by shortening the diffusion path that the analyte molecules must travel and to
improve their mass-transfer kinetics.
Nowadays, columns packed with the latest generation of sub-3-µm SPPs rival the
efficiency of columns with sub-2-µm fully porous particles, but at only half the operating
pressure. To improve the separation efficiency further, columns packed with sub-2-µm SPPs
(1.3–1.7 µm) were recently commercialized and showed impressive efficiency [42–45]. SPP
materials have been successfully applied in many fields of application, including the
characterization of large biomolecules under their intact, fragmented or digested form.
Columns packed with wide-pore (300 Å) 3.6-µm SPPs (Aeris Widepore) showed
significant gains in analysis time and peak capacity compared to fully porous materials for
Page 7 of 15
8
intact protein analysis [21]. Fig. 4 shows the separation of intact Interferon α-2A-related
impurities (oxidized and reduced forms) within 5 min applying a 150-mm-long column
packed with 3.6-µm SPPs [25]. In another recent study, the same column was used to separate
N-methionylated variants of Interferon α-2A in a drug substance and finished product, and for
the analysis of variants in untreated, oxidized and slightly degraded samples [46].
A column packed with 400-Å SPPs of 3.4-µm (Halo Fused-core) was also used for the
efficient separation of large intact proteins. This study reported the fast separation of intact-
protein mixtures, and very high-resolution separations of mAbs and associated variants [47].
In another work, the efficiency and the possible analysis time of 1.7-µm SPPs and fully
porous materials were compared for peptides and large intact proteins [48]. This study
suggests a two-fold increase in terms of achievable peak capacity and analysis speed for large
proteins when using SPPs, compared to those achieved with fully porous materials of the
same particle size.
For the separation of peptides and moderate size proteins, a 160-Å packing was introduced
in 2010 by Advanced Material Technology and Supelco [49,50]. An average pore size of 160
Å allowed unrestricted access of molecules up to ~15 kDa, depending on their molecular
conformation [51].
Recently introduced 1.3-µm and 1.6-µm SPPs were applied to peptide mapping of mAb
samples [44,45]. By combining long columns (200–300 mm) with long analysis time, peak
capacity around 1000 could be reached using 1.3-µm SPPs for 0.5–2-kDa peptides. When
using 50-mm-long columns and analysis time of 10 min, peak capacities of 200–300 can be
readily achieved.
4. Size exclusion chromatography (SEC)
SEC is a traditional technique widely employed for the detailed characterization of
therapeutic proteins. It can be considered a reference technique for the qualitative and
quantitative evaluation of aggregates. The main advantage of this approach is the mild
mobile-phase conditions that permit the characterization of proteins with minimal impact on
the conformational structure and local environment.
There is now a wide variety of porous packing materials with different particle sizes
available for the separation and characterization of macromolecules by SEC [52]. As an
example, UHPLC technology, originally developed for RPLC applications, is available for
SEC operation using 1.7-µm BEH silica particles with varying pore sizes of 125–900 Å [53].
Using the recently introduced sub-2-µm particles in SEC (aqueous mobile phase, 150-mm
column length), very high pressures (up to 500–600 bar) can be generated. Thermal effects
and shear forces might therefore become critical for temperature- or pressure-sensitive
proteins [54,55]. These effects might also cause some changes in protein structure, and there
is an additional risk of on-column protein denaturation and aggregation.
Nevertheless, very fast separations of peptides, myoglobulin and insulin aggregates were
demonstrated with 1.7-µm SEC columns [56]. These very efficient columns were also applied
to characterize recombinant mAbs and aggregates within 3–4 min [57].
5. Ion-exchange chromatography (IEX)
IEX is widely used for profiling the charge heterogeneity of proteins resulting from
enzymatic or chemical modifications [5]. The technology is used for characterization and
release testing. Currently, cation exchange (CEX) is considered the gold standard for the
analysis of mAbs, given their basic nature. In IEX, the sample is loaded onto a column under
Page 8 of 15
9
mobile-phase pH where the net charge of the protein is opposite to the charged ligand of the
stationary phase. Elution is then achieved using a salt gradient and/or a pH gradient [5].
Since electrostatic/ionic interactions are energetically strong, the IEX mechanism is
considered a slower mass-transfer process, so non-porous materials are typically used for IEX
separations of proteins to eliminate the contribution of trans-particle mass-transfer resistance
that strongly contributes to band broadening of large biomolecules. Usually 5–10-µm non-
porous particles are used in routine IEX applications. However, particle size also plays a role
in band broadening with non-porous material, but its impact is certainly less important than
for fully porous or SPP particles. Indeed, particle size mostly affects the Eddy dispersion of
columns packed with non-porous particles. This is probably the reason why UHPLC has not
yet been considered for IEX separations. Currently, only two manufacturers offer IEX
columns of 1.7-µm non-porous particles (Agilent BioMab and Sepax Technologies’
Antibodix).
6. Hydrophilic interaction chromatography (HILIC)
HILIC combined with fluorescence detection is a key technique to analyze N-glycans
originating from biopharmaceuticals [58–62].
This mode of chromatography is often performed on columns packed with sub-2-µm
particles. This approach enables the very efficient separation of glycans and allows structural
isomers to be resolved. The method has also been applied successfully to the measurement of
sialic acid and phosphate-containing glycans that provide increased retention over their
neutral counterparts [58–62].
Ahn et al. separated N-linked glycans by HILIC utilizing a 1.7-µm glycan column and 2-
aminobenzamide (2-AB) to label glycans [60]. Similar to the RPLC mode, the HILIC column
packed with 1.7-µm amide sorbent improved significantly the peak capacity compared to 3-
µm packing. The optimal peak capacity was achieved at a flow rate of 0.2–0.5 mL/min.
Excellent resolutions were achieved for the high-throughput separation of 2-AB-labeled
glycans released from fetuin, RNase B, and human IgG.
Bones et al. applied UHP-HILIC as a second dimension of a multidimensional separation
[61]. The combined IEX x HILIC allowed the rapid, comprehensive analysis of complex N-
glycosylation profiles of therapeutic glycoproteins.
The combination of UHP-HILIC and capillary electrophoresis with laser induction
fluorescence detection (CE-LIF) was applied in another study for the identification of
oligosaccharides on recombinant β-glucuronidase [62].
7. Combination of UHPLC and MS
MS has become increasingly popular for the characterization of peptides and proteins. The
success of this technique is related to the introduction of two highly sensitive and “soft”
ionization techniques [i.e., electrospray ionization (ESI) and matrix-assisted laser desorption/
ionization (MALDI)], enabling the transfer of intact proteins into the gas phase without
fragmentation, and to the continuous improvements of MS instruments that can detect high
m/z ratios at elevated sensitivity [63]. In addition, modern MS instruments can discriminate
proteins with close m/z ratios because of their very high spectral resolution {e.g., maximal
resolution of 50,000 for time-of-flight (TOF), up to 500,000 for Orbitrap [64] and up to
several millions with Fourier-transform ion-cyclotron resonance (FT-ICR) [65]}, in
comparison with quadrupole-based instruments (resolution 1000).
ESI and MALDI can be readily combined with RPLC and HILIC, while the other
chromatographic approaches (SEC or IEX) are compatible with MALDI (more tolerant to
Page 9 of 15
10
non-volatile salts) and have limited on-line compatibility with ESI. The combination of RPLC
and MALDI is primarily off-line and consists in the collection of spots originating from the
RPLC column. In contrast, ESI is a flow-based on-line approach, where the liquid sample is
transformed into an aerosol via a nebulization process and ionized to produce gas-phase ions.
The main application of LC-MS in the analysis of therapeutic proteins is identification of
the protein primary structure and impurities or degradation products. Nowadays, de novo
sequencing for the rapid determination of protein sequence can be achieved by peptide
mapping using RPLC-ESI/MS/MS. Alternatively, MALDI/MS of collected RPLC peaks is
also feasible, albeit slower and more labor intensive. LC-MS-based peptide mapping of mAbs
typically covers over 98% of the sequence determination and the confirming protein
identification [5]. Implementing tandem MS systems obtains fragmentation data on the
peptides and further enhances confidence in peptide identity, sequence and modification sites
[5].
Enzymatically-derived peptide fragments of a mAb were generated using a dedicated
sample-preparation protocol, then separated by RP-UHPLC and detected by ESI-MS/MS
using a bottom-up approach [66]. The MS/MS was able to sequence co-eluting peptides.
UHPLC was also combined with QqTOF/MS in a detailed study of mAb deamidation
under extreme conditions to achieve chromatographic resolution of the deamidated products,
while maintaining relatively short analysis time [67].
The on-line LC-MS approach can also be applied to separation and identification of HC
and LC species in SEC by adding organic modifiers (20% acetonitrile) and a mixture of TFA
and FA into the mobile phase, albeit at the cost of separation power [68].
The feasibility of direct coupling of IEX methods applying low ionic strength pH gradient
with MS was recently studied for the separation of mAb variants [69]. By applying an
ammonium-hydroxide buffer system and methanol in the mobile phase, an improvement in
the ionization efficiency was observed.
Using an additional make-up pump (switching valve) between the column and the MS inlet
to increase the organic solvent content of the mobile phase can also improve ionization
efficiency and sensitivity. This approach is sometimes applied to the analysis of small
molecules in RPLC and supercritical fluid chromatography (SFC). Such make-up pumps,
splitters and switching valves are available from different vendors. However, applications
using SEC and IEX have not been found in the literature for protein analysis.
8. Conclusion
As shown in this review paper, there are some significant advantages in using UHPLC
columns packed with sub-2-µm particles to enhance the throughput or resolving power of
regular HPLC. The advantages observed in UHPLC conditions with small molecules are still
valid in most cases with large biomolecules, at least in the RPLC mode, though a few
modifications of the mobile and stationary phases may be required.
For a successful analysis of intact proteins, mAbs (150 kDa) or mAb fragments of 25–100
kDa (top-down or middle-down approaches), the RPLC mobile-phase conditions have to be
adjusted. It is advisable to add an ion-pairing reagent (0.1% TFA) and to thermostat the
column at T > 70°C. In terms of stationary phases, sub-2-µm C18 or C4 bonded phases on
fully porous or SPP wide-pore particles appear to be ideal. Under these conditions, a peak
capacity of more than 100 can be achieved in about 10–20 min, with relatively long UHPLC
columns of 150–450 mm. In the case of peptide mapping (bottom-up approach), there is no
need for a very high temperature (50°C can be sufficient) and the column pore size can be in
the range 130–200 Å. Under such conditions, a peak capacity of 300–600 can be obtained in 1
h, using a 150-mm column packed with sub-2-µm particles.
Page 10 of 15
11
Other than their use for RPLC, columns packed with sub-2-µm particles can also be employed
with other chromatographic modes, such as HILIC, SEC or IEX. HILIC is generally
employed for profiling glycans. In SEC, caution should be exercised to minimize the
artifactual aggregates under UHPLC conditions. For IEX, the interest in columns packed with
sub-2-µm particles is less obvious, and non-porous particles are generally used, as the kinetic
advantage of UHPLC-type columns is much more limited in IEX.
Page 11 of 15
12
References [1] G. Walsh, Biopharmaceuticals, an overview [In] Biopharmaceuticals, an industrial perspective, Edited by G. Walsh, B. Murphy, Kluwer Academic Publishers, Dordrecht, The Netherlands, 1999, pp. 1-34. [2] G. Walsh, Pharmaceuticals, biologics and biopharmaceuticals [In] Pharmaceutical biotechnology: concepts and applications, Wiley, West Sussex, England, 2007, pp. 1-11. [3] S. Tandon, S. Sharma, R. Rajput, B. Semwal, P.K. Yadav, K. Singh, Biotech Drugs: The next boom in pharmaceutical market, J. Pharm. Res. Op. 2 (2011) 76-79. [4] G. Walsh, Nat. Biotechnol. Biopharmaceutical benchmarks 28 (2010) 917-924. [5] K. Sandra, I. Vandenheede, P. Sandra, Modern chromatographic and mass spectrometric techniques for protein biopharmaceutical characterization, J. Chromatogr. A 1335 (2013), 81-103. http://dx.doi.org/10.1016/j.chroma.2013.11.057. [6] S.A. Berkovitz, J.R. Engen, J.R. Mazzeo, G.B. Jones, Analytical tools for characterizing biopharmaceuticals and the implications for biosimilars, Nat. Rev. Nat. Discov. 11 (2012) 527-540. [7] J.G. Elvin, R.G. Couston, C.F. van der Walle, Int. Therapeutic antibodies: market considerations, disease targets and bioprocessing, J. Pharm. 440 (2013) 83-98. [8] I.S. Krull, A. Rathore, T.E. Wheat, Current applications of UHPLC in biotechnology, Part I: peptide mapping and amino acid analysis, LCGC North. Am. 29 (2011) 838-848. [9] A. Staub, D. Guillarme, J. Schappler, J.L. Veuthey, S. Rudaz, Intact protein analysis in the biopharmaceutical field, J. Pharm. Biomed. Anal. 55 (2011) 810-822. [10] Zs.K. Szekeres, M. Olajos, K. Ganzler, J. Pharm. Biomed. Anal. 69 (2012) 185-195. [11] R.L. Lundblad, Approaches to the Conformational Analysis of Biopharmaceuticals Chapman and Hall/CRC (2009) [12] A. Bax, Two-dimensional NMR and protein structure Annu. Rev. Biochem. 58 (1989), 223–256. [13] S. Fekete, A.L. Gassner, S. Rudaz, J. Schappler, D. Guillarme, Analytical strategies for the characterization of therapeutic monoclonal antibodies, Trends in Anal. Chem. 42 (2013) 74-83. [14] S. Fekete, J.L. Veuthey, D. Guillarme, New trends in reversed-phase liquid chromatographic separation of therapeutic peptides and proteins: Theory and applications, J. Pharm. Biomed. Anal., 69 (2012) 9-27. [15] A.J.P. Martin, R.L.M. Synge, A new form of chromatogram employing two liquid phases, Biochem. J. 35 (1941) 1358-1368. [16] J.E. MacNair, K.C. Lewis, J.W. Jorgenson, Ultra high pressure reversed phase liquid chromatography in packed capillary columns, Anal. Chem. 69 (1997) 983-989. [17] J.E. MacNair, K.D. Patel, J.W. Jorgenson, Ultra high pressure reversed-phase liquid chromatography: Isocratic and gradient elution using columns packed with 1.0 μm particles, Anal. Chem. 71 (1999) 700-708. [18] N. Wu, J.A. Lippert, M.L. Lee, Practical aspects of ultrahigh pressure capillary liquid chromatography, J. Chromatogr. A 911 (2001) 1-12. [19] T. Issaeva, A. Kourganov, K. Unger, Super-high-speed liquid chromatography of proteins and peptides on non-porous Micra NPS-RP packings, J. Chromatogr. A 846 (1999) 13-23. [20] N. Wu, Y. Liu, M.L. Lee, Sub-2 μm porous and nonporous particles for fast separation in reversed-phase high performance liquid chromatography, J. Chromatogr. A 1131 (2006) 142-150. [21] S. Fekete, R. Berky, J. Fekete, J.L. Veuthey, D. Guillarme, Evaluation of a new wide pore core-shell material (Aeris TM WIDEPORE) and comparison with other existing stationary phases for the analysis of intact proteins, J. Chromatogr. A 1236 (2012) 177-188. [22] S. Fekete, D. Guillarme, Reversed-phase liquid chromatography for the analysis of therapeutic proteins and recombinant monoclonal antibodies, LCGC Eur. 25 (2012) 540-550. [23] R.A. Everley, T.R. Croley, Ultra-performance liquid chromatography/mass spectrometry of intact proteins, J. Chromatogr. A 1192 (2008) 239–247.
Page 12 of 15
13
[24] S. Fekete, J.L. Veuthey, D.V. McCalley, D. Guillarme, The effect of pressure and mobile phase velocity on the retention properties of small analytes and large biomolecules in ultra-high pressure liquid chromatography, J. Chromatogr. A 1270 (2012) 127-138. [25] S. Fekete, R. Berky, J. Fekete, J.L. Veuthey, D. Guillarme, Evaluation of recent very efficient wide-pore stationary phases for the reversed-phase separation of proteins, J. Chromatogr. A 1252 (2012) 90-103. [26] S. Fekete, M. Dong, T. Zhang, D. Guillarme, High resolution reversed phase analysis of recombinant monoclonal antibodies by ultra-high pressure liquid chromatography column coupling, J. Pharm. Biomed. Anal. 83 (2013) 273-278. [27] S. Fekete, S. Rudaz, J.L. Veuthey, D. Guillarme, Impact of mobile phase temperature on recovery and stability of monoclonal antibodies using recent reversed-phase stationary phases, J. Sep. Sci. 35 (2012) 3113–3123. [28] S. Fekete, J.L. Veuthey, S. Eeltink, D. Guillarme, Comparative study of recent wide-pore materials of different stationary phase morphology, applied for the reversed-phase analysis of recombinant monoclonal antibodies, Anal. Bioanal. Chem. 405 (2013) 3137-3151. [29] S. Fekete, S. Rudaz, J. Fekete, D. Guillarme, Analysis of recombinant monoclonal antibodies by RPLC: Toward a generic method development approach, J. Pharm. Biomed. Anal. 70 (2012) 158-168. [30] L. Fornelli, D. Ayoub, K. Aizikov, A. Beck, Y.O. Tsybin, Middle-Down Analysis of Monoclonal Antibodies with Electron Transfer Dissociation Orbitrap Fourier Transform Mass SpectrometryAnal. Chem. 86 (2014) 3005-3012. dx.doi.org/10.1021/ac4036857. [31] K. Sandra, M. Moshir, F. D’hondt, K. Verleysen, K. Kas, P. Sandra, Highly efficient peptide separations in proteomics Part 1. Unidimensional high performance liquid chromatography, J. Chromatogr. B 866 (2008) 48-63. [32] S. Wang, R. Ionescu, N. Peekhaus, J. Leung, S. Ha, J. Vlasak, Separation of post-translational modifications in monoclonal antibodies by exploiting subtle conformational changes under mildly acidic conditions, J. Chromatogr. A 1217 (2010) 6496–6502. [33] L. Zamani, F.O. Andersson, P. Edebrink, Y. Yang, S.P. Jacobsson, Conformational studies of a monoclonal antibody, IgG1, by chemical oxidation: structural analysis by ultrahigh-pressure LC-electrospray ionization time-of-flight MS and multivariate data analysis, Anal. Biochem., 380 (2008) 155-163. [34] W. Wang, A.R. Meeler, L.T. Bergerud, M. Hesselberg, M. Byrne, Z. Wu, Quantification and characterization of antibody deamidation by peptide mapping with mass spectrometry, Inter. J. Mass. Spec. 312 (2012) 107-113. [35] J. Fu, J. Bongers, L. Tao, D. Huang, R. Ludwig, Y. Huang, Y. Qian, J. Basch, J. Goldstein, R. Krishnan, L. You, Z. J. Li, R. J. Russell, Characterization and identification of alanine to serine sequence variants in an IgG4 monoclonal antibody produced in mammalian cell lines, J. Chromatogr. B 908 (2012) 1-8. [36] H. Liu, B. Xu, M.K. Ray, Z. Shahrokh, Peptide mapping with liquid chromatography using a basic mobile phase, J. Chromatogr. A 1210 (2008) 76-83. [37] R. Berky, S. Fekete, J. Fekete, Enhancing the quality of separation in one-dimensional peptide mapping using mathematical transformation, Chromatographia, 75 (2012) 305-312. [38] S. Fekete, D. Guillarme, M.W. Dong, Superficially porous particles: perspectives, practices and trends, LCGC 2014 in press [39] C. Horvath, B.A. Preiss, S.R. Lipsky, Fast liquid chromatography: An investigation of operating parameters and the separation of nucleotides on pellicular ion exchangers, Anal. Chem. 39 (1967) 1422-1428. [40] C. Horvath, S.R. Lipsky, Column design in high pressure liquid chromatography, J. Chromatogr. Sci. 7 (1969) 109-116. [41] J.J. Kirkland, Controlled surface porosity supports for high speed gas and liquid chromatography, Anal. Chem. 41 (1969) 218-220. [42] A. C. Sanchez, G. Friedlander, S. Fekete, J. Anspach, D. Guillarme, M. Chitty, T. Farkas, Pushing the performance limits of reversed-phase ultra high performance liquid chromatography with 1.3 µm core–shell particles, J. Chromatogr. A 1311 (2013) 90-97.
Page 13 of 15
14
[43] S. Fekete, D. Guillarme, Kinetic evaluation of new generation of column packed with 1.3 µm core–shell particles, J. Chromatogr. A, 1308 (2013) 104-113. [44] S. Fekete, D. Guillarme, Possibilities of new generation columns packed with 1.3 µm core–shell particles in gradient elution mode, J. Chromatogr. A, 1320 (2013) 86-95. [45] B. Bobály, D. Guillarme, S. Fekete, Systematic comparison of a new generation of columns packed with sub-2 µm superficially porous particles, J. Sep. Sci. 37 (2014) 189-197. [46] Y. Lia, C. Raoa, L. Taoa, J. Wanga, B. Lorbetskieb, M. Girard, Improved detection of variants in recombinant human interferon alpha-2a products by reverse-phase high-performance liquid chromatography on a core–shell stationary phase, J. Pharm. Biomed. Anal. 88 (2014) 123. [47] S.A. Schuster, B.M. Wagner, B.E. Boyes, J.J. Kirkland, Optimized superficially porous particles for protein separations, J. Chromatogr. A, 1315 (2013) 118-126. [48] S. Fekete, K. Ganzler, J. Fekete, Efficiency of the new sub-2 µm core–shell (KinetexTM) column in practice, applied for small and large molecule separation, J. Pharm. Biomed. Anal., 54 (2011) 482-490. [49] F. Gritti, G. Guiochon, The mass transfer kinetics in columns packed with Halo-ES shell particles, J. Chromatogr. A 1218 (2011) 907-921. [50] S.A. Schuster, B.M. Wagner, B.E. Boyes, J.J. Kirkland, Wider pore superficially porous particles for peptide separations by HPLC, J. Chromatogr. Sci. 48 (2010) 566-571. [51] S.A. Schuster, B.E. Boyes, B.M. Wagner, J.J. Kirkland, Fast high performance liquid chromatography separations for proteomic applications using Fused-Core® silica particles, J. Chromatogr. A 1228 (2012) 232-241. [52] T. Hofe, G. Reinbold, J. McConville, The challenge of using small particle packing materials in SEC/GPC, risks and possibilities, Chrom. Today, 4 (2011) 18-23. [53] E. Gazal, Can size exclusion chromatography (SEC) be done on sub-3 um particles?, presented at the 17th annual meeting of the Israel Analytical Chemistry Society, 2014, Tel Aviv, Israel [54] J. Vajda, R. Römling, New approaches to hplc analysis of antibody aggregates and fragments, Chromatogr. Today 5 (2012) 44-47. [55] Y. Liu, W. Radke, H. Pasch, Coil-stretch transition of high molar mass polymers in packed-column hydrodynamic chromatography, Macromolecules 38 (2005) 7476-7484. [56] S. M. Koza, P. Hong, K.J. Fountain, Advantages of ultra performance liquid chromatography using 125 Å pore size, sub-2 µm particles for the analysis of peptides and small proteins, poster presented at Medimmune 2012, Rockville, MD. [57] S. Fekete, K. Ganzler, D. Guillarme, Critical evaluation of fast size exclusion chromatographic separations of protein aggregates, applying sub-2 µm particles, J. Pharm. Biomed. Anal, 78-79 (2013) 141-149. [58] M. Melmer, T. Stangler, A. Premstaller, W. Lindner, Comparison of hydrophilic-interaction, reversed-phase and porous graphitic carbon chromatography for glycan analysis, J. Chromatogr. A 1218 (2011) 118-123. [59] M. Melmer, T. Stangler, M. Schiefermeier, W. Brunner, H. Toll, A. Rupprechter,W. Lindner, A. Premestaller , HILIC analysis of fluorescence-labeled N-glycans from recombinant biopharmaceuticals, Anal. Bioanal. Chem. 398 (2010) 905-914. [60] J. Ahn, J. Bones, Y.Q. Yu, P.M. Rudd, M. Gilar, Separation of 2-aminobenzamide labeled glycans using hydrophilic interaction chromatography columns packed with 1.7 um sorbent, J. Chromatogr. B 878 (2010) 403-408. [61] J. Bones, N. McLoughlin, M. Hilliard, K. Wynne, B.L. Karger, P.M. Rudd, 2D-LC analysis of BRP 3 erythropoietin N-glycosylation using anion exchange fractionation and hydrophilic interaction UPLC reveals long poly-N-acetyl lactosamine extensions, Anal. Chem. 83 (2011) 4154-4162. [62] J. Bones, S. Mittermayr, N. McLoughlin, M. Hilliard, K. Wynne, G.R. Johnson,J.H. Grubb, W.S. Sly, P.M. Rudd, Identification of N-glycans displaying mannose-6-phosphate and their site of attachment on therapeutic enzymes for lysosomal storage disorder treatment, Anal. Chem. 83 (2011) 5344-5352.
Page 14 of 15
15
[63] G. L. Glish, R. W. Vachet, The basics of mass spectrometry in the twenty-first century, Nature Rev. Drug Discov. 2 (2003) 140-150. [64] A. Makarov, Electrostatic axially harmonic orbital trapping: a high-performance technique of mass analysis, Anal. Chem. 72 (2000) 1156-1162. [65] A. G. Marshall, C. L. Hendrickson, Fourier transform ion cyclotron resonance detection: principles and experimental configurations, Int. J. Mass Spectrom. 215 (2002) 59-75. [66] C. Ji, N. Sadagopan, Y. Zhang, C. Lepsy, A universal strategy for development of a method for absolute quantification of therapeutic monoclonal antibodies in biological matrices using differential dimethyl labeling coupled with ultra performance liquid chromatography-tandem mass spectrometry, Anal. Chem. 81 (2009) 9321-9328. [67] S. Sinha, L. Zhang, S. Duan, T.D. Williams, J. Vlasak, R. Ionescu, E.M. Topp, Effect of protein structure on deamidation rate in the Fc fragment of an IgG1 monoclonal antibody, Protein Sci. 18 (2009) 1573-1584. [68] H. Liu, G.G. Bulseco, C. Chumsae, Analysis of reduced monoclonal antibodies using size exclusion chromatography coupled with mass spectrometry, J. Am. Soc. Mass Spectrom. 20 (2009) 2258-2264. [69] M. Talebi, A. Nordborg, A. Gaspar, N.A. Lacher, Q. Wang, X.Z. He, P.R. Haddad, E.F. Hilder, Charge heterogeneity profiling of monoclonal antibodies using low ionic strength ion-exchange chromatography and well-controlled pH gradients on monolithic columns, J. Chromatogr. A 1317 (2013) 148-154. Captions
Fig. 1. Fast separation of filgrastim-related proteins. Column: Acquity BEH300 C18 50x2.1 mm, 1.7
µm. Mobile phase A: 0.1% TFA in water, mobile phase B: 0.1% TFA in acetonitrile. Gradient: 47–62
% B, Temperature: 80°C, Detection FL: 280–350 nm. Injected volume: 1 µL. Peaks: 1,2: oxidized
filgrastim, 3: filgrastim, 4: reduced filgrastim. (Adapted from [22], with permission).
Fig. 2. Separation of a 30-kDa therapeutic protein on a reversed-phase HPLC column packed with 5-
µm (a) and sub-2-µm particles (b). Both columns were operated on the same UHPLC system under
identical gradient conditions at 60°C using mobile phases containing water, acetonitrile and 0.1%
TFA. The 280-nm UV traces are displayed. (Adapted from [5], with permission).
Fig. 3. Fast analysis of 25-kDa mAb (IgG1) fragments obtained after limited proteolytic digestion.
Column: Acquity BEH300 C18 150x2.1 mm, 1.7 µm. Mobile phase A: 0.1% TFA in water, mobile
phase B: 0.1% TFA in acetonitrile. Gradient: 31–40 %B in 6 min, Temperature: 80°C, Detection FL:
280–360 nm. (Unpublished data from the authors’ laboratory).
Fig. 4. Representative chromatograms of Interferon α-2A related proteins. Conditions: mobile-phase
A: 0.1% TFA in water, mobile phase B: 0.1% TFA in ACN, gradient profile: 35–60% B in 5 min,
followed by 2 min re-equilibrating, flow rate: 0.3 mL/min, T = 60°C, injected volume = 5 µL, λ = 210
nm. Column: Aeris WP C18, 150 x 2.1 mm, 3.6 µm. (Adapted from [25], with permission).
Page 15 of 15