The comparison of a measuring instrument against an
accurate standard to determine any deviation.
The device with known assigned correctness is called the
Standard. The second device is the Unit Under Test (UUT).
In case of accurate thermocouple measurement, it’s a
common practice to reference the cold junction
temperature at ice point (0°C) so that copper leads may be
connected to an EMF readout
device. This procedure avoids the
compensation of cold junction
temperature at the terminal of
read out which may not be
constant and the measurement
may not be very accurate.
What is Calibration
Accurate Temperature Calibration
Temperature Calibration has many facets. It can be carried
out thermally in the case of probes or electrically in the
case of Instrument and it can be performed directly with
certified equipment, or indirectly with traceable
standards.
Thermal (Temperature) calibration is achieved by elevating
(or depressing) the temperature sensor to a known,
controlled temperature and measuring the corresponding
change in its associated electrical parameter (voltage or
resistance).
The accurately measured parameters compared with that
of a certified reference probe; the absolute difference
represents the calibration error. If the sensor is connected
to a measuring instrument, the sensor and the instrument
combination can be effectively calibrated by this
technique.
A typical general purpose system comprises of a thermal
reference (stable temperature source), a certified
reference probe with its certificate, a precision electronic
digital thermometer, bridge or digital voltmeter.
Dry Block Calibrator
Provides the most convenient, portable facilities for
checking & calibrating industrial probes and they are
usually reasonable rapid heating and cooling device. The
unit consists of a special designed heating block which has
located internal holes for the probes. Although the block
temperature is accurately controlled, any indication
provided should be used for guidance only.
As with any comparison technique a
certified sensor and indicator should be
used to measure the block temperature
and used as a reference for the test
probe. Two types of unit are available;
portable units which can be taken on to
plant for the on-site calibration and
laboratory units to which industrial
sensors are brought as required.
Temperature Source
Black Body Source
Black bodies are reference sources used for testing
infrared systems. They are required in industry for
calibration of pyrometers, infrared line scanners or
cameras. In laboratory, they are
p a r t o f b e n c h e s f o r
characterization of complex
optronic systems. Tempsens offers
a wide range of black bodies to
cover all Customer’s needs.
Cold Junction Compensation
By connecting any thermocouple to measurement device
three dissimilar metal junction are created in the circuit:
the thermocouple junction itself, or hot junction, and the
junction between each lead and the measurement device,
or cold junctions. These cold junction provide their own
thermoelectric voltages that are proportional to the
temperature at the device terminals.
A technology known as cold junction compensation is
therefore used to remove this unwanted effect.
Referernce Unit
Stirred Liquid Bath
Provide superior thermal environment
for probe immersion as no air gap exist
between the probe and the medium
Thermal coupling is therefore much
better than the alternatives described,
and the stirring results in very even heat
distribution throughout the medium.
Methanol is used for temperature
below 0°C, water from 0 to 80°C and
silicon oil for upto 250°C.
Division of contact type calibrators according to the temperature range.
Stable Reference Unit
Reference temperature could be change according to customer requirementTo be specified at time of ordering
Master Sensor
SENSOR TYPE ORANGE( C) ACCURACY
PT 25.5
PT 100
PT 200
K/N Type
R/S Type
RTD
-80 to 400
-80 to 250
-80 to 250
-80 to 250
0 to 1200
0 to 1500Thermocouple
± 1.1°C OR ± 0.4%
± 0.6°C OR ± 0.1%
Class A
1/3 DIN
1/5 DIN
1/10 DIN
SSPRT _ upto 660 long term drift ≤ ±50mk/Yr.
Bath Dry Block
-200°C
0°C
200°C
400°C
600°C
800°C
1000°C
1200°C
1400°C
1600°C
1800°C
Cal
sys
-19
6/-
80
Cal
sys
-30
/11
0
Cal
sys
-15
/11
0
Cal
sys
65
0
Cal
sys
FB
Cal
sys
12
00
Cal
sys
12
00
L
Cal
sys
12
00
3Z
Cal
sys
15
00
L
Cal
sys
17
00
L
Cal
sys
-80
/50
Cal
sys
-40
/50 Cal
sys
-35
/20
0
Cal
sys
12
0SP C
alsy
s 3
00
SP
Cal
sys
25
0
Calibrators Temperature Representation for Contact Type Calibration
*
*
TYPE CALref 0 CALref 60
Channel 20
0°C
J,K,T,E,N,R,S,B J,K,T,E,N,R,S,B
24
60°CRef. temp.
Portable Model
Wall Mounted
Model
Type of Junction
**
Division of non-contact type calibrators according to the temperature range.
Stable Reference Unit
Reference temperature could be change according to customer requirementTo be specified at time of ordering
TYPE CALref 0 CALref 60
Channel 20
0°C
J,K,T,E,N,R,S,B J,K,T,E,N,R,S,B
24
60°CRef. temp.
Portable Model
Wall Mounted
Model
Type of Junction
Master Sensor
SENSOR TYPE ORANGE( C) ACCURACY
PT 25.5
PT 100
PT 200
K/N Type
R/S Type
RTD
-80 to 400
-80 to 250
-80 to 250
-80 to 250
0 to 1200
0 to 1500Thermocouple
± 1.1°C OR ± 0.4%
± 0.6 °C OR ± 0.1%
Class A
1/3 DIN
1/5 DIN
1/10 DIN
SSPRT _ upto 660 long term drift ≤ ±50mk/Yr.
Calibrators Temperature Representation for Non-Contact Type Calibration
-200°C
0°C
200°C
400°C
600°C
800°C
1000°C
1200°C
1400°C
1600°C
1800°C
2000°C
2200°C
2400°C
2600°C
2800°C
3000°CLB
B3
3H
Cal
sys
11
0 B
B
Cal
sys
50
0B
B
Cal
sys
12
00
BB
Cal
sys
15
00
BB Cal
sys
17
00
BB
Fast
cal 3
00
0
Black Body
*
*
*
*
KEY FEATURES
ü High Stability
ü PC Interfacing
ü Simple to use & Cost effective portability
TPW in Thermal CalibrationThermal Calibration lab has facility for Fixed point calibration of SPRTs, HTPRTs and standard thermocouples probe. Fixed point calibration are used for primary temperature standard in thermal calibration to achieve the lowest possible uncertainties. Triple point of water cell may be the most commonly used type of fixed point and it is used in ITS-90 calibration. Water can exist as a solid, liquid and vapor at 0.01 deg C.Tempsens TPW CellTempsens TPW cells are manufactured with high quality borosilicate glass in two sizes mini & Standard. Special procedure for cleaning, ultra pure water with special degassing technique is used and the cell is flame sealed.
Maintenance ApparatusTempsens made maintenance dry block is best fitted for tpw cells. They are carefuly designed to give the required stability and uniformity and provide sufficient immersion depth for TPW cell. It can be also use as dry block calibrator using equalizing block for comparison calibration.Equipments Used In Realization TPW Mini CellPlatinum Resistance Thermometer (PRT)Measuring EquipmentAccessoriesŸ Master RTD Model No. TPRT - A - 300Ÿ Carry case Ÿ Operational ManualŸ Insertion Block Dia 32mm x 180mm long with 4 x 6mm
hole of 150 mm insertion depth
Maintenance Apparatus For Tpw CellDRY BLOCK FURNACE : Dry block offers medium temperature range from -15°C to 110°C. It is a highly stable standard temperature source for cal ibrat ing RTD/thermocouples. It has peltier elements to generate stable temperature in the well. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment. Self tuned PID controller used for increased or decreased temperature.
Triple Point Of Water Cell with Maintenance Apparatus
Triple point of water is the most important defining thermometric fixed point used in the calibration of thermometers. Triple point of water cell is used to create a thermal equilibrium between solid, liquid and vapor phases are independent of ambient pressure. This thermal equilibrium occurs at 273.16 Kelvin or 0.01 degree Celsius. TPW cell is cylindrical borosilicate glass filled with pure water. The water in the cell is high purity and gas free and have isotropic composition.
Uncertainty 2mK
Reproducibility <1mK
Dimensions OD-32 mm,Total Length-165mm
Material Borosilicate Glass
Immersion Profile
Depth 150 mm
Insert Construction 1 hole of 8 mm dia
Water Used 99.999% Pure distilled water
TPW Cell Type Mini
SPECIFICATIONS OF CALSYS -15/110
Ÿ Temperature Range : -15/110Ÿ Temperature Resolution : 0.1°CŸ Controlling Sensor : RTDŸ Power Requirement : 230 VACŸ Weight : 14 kgŸ Stability : 0.05°C Ÿ Time to reach min. Temp. : 30Min.Ÿ Method of control : Digital self tuned PID
controller Ÿ Dimension : 430(H) x 170(W) x
188(D) mmŸ Insert Construction : 33mm Dia x 220mm
long for dry block
CALsys -190/-80 offers ultra low temperature range from -190 to -80°C. It is a highly stable standard source for calibrating Thermocouples / RTD. It has been designed for low temperature range calibration and find application in the glass, pharma, electrical power, automotive & material processing industries. The comparison volume is a metallic block of aluminium having dimension 25mm diameter with 300 mm long which fixed into the liquid nitrogen container for calibration of RTD & thermocouple. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment.
PORTABLE RTD CALIBRATOR
Temperature Range : -190 to -80°C
Temperature Resolution : 1.0°C
Stability : ±0.1°C at-190°C
±0.1°C at -130°C
±0.1°C at -80
Uniformity : ±0.2°C at-190°C
±0.15°C at -130°C
±0.17°C at -80
Controlling Sensor : RTD PT100
Method of Control : Digital self tuned PID Controller
Insert Construction : Dia 25 x 300 mm long (2 x 6 mm & 2 x 8 mm holes) of 300 mm insertion depth
Time to Reach Max Temp : 30 Mins
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 300 W
Dimensions of Container : 580(H) x 280(Dia)
Nitrogen Container : 10Ltr.
Dimension of Control Panel : 270(H)x380(W)x270(D)
Weight : 15Kg (without packing)
SPECIFICATIONS
CALsys -190/-80
Ÿ TEMPMET 08, TEMPMET 09
Ÿ Extra Equalizing Block………Part No. EQ2
OPTIONAL ACCESSORIES
Ÿ Reference Standard RTD
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath
temperature and monitoring the PV. Graphical
representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
STANDARD ACCESSORIES
CALsys -190/-80
0-190 -130 -80
0.05
0.1
0.15
0.2
0.25
Temperature in °C
Tem
pe
ratu
re in
°C
STABILITY UNIFORMITY
-190 to -80°C
KEY FEATURES
ü Large Immersion Depths
ü Wide Operating Range (-190 to -80°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
ü Separate control box
CALsys -80/50 is a Highly Stable Stirred Liquid Bath for Calibration RTD / Thermocouple and also for compensation of cold junction. CALsys -80/50 offers the temperature range of -80 to 50°C. It has large tank which is filled with liquid Methanol. The liquid is heated or cooled to the desired temperature. It has better stability & uniformity than dry block bath and as it uses of stirred liquid. This CALsys -80/50 is built using CFC free refrigerants.
-80 to 50°CCALSYS-80/50
HIGH STABILITY BATH
Ÿ Reference Standard PRT...Part No. TPRT-A-300 NABL accredited calibration certificate - 3 point.
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual.
STANDARD ACCESSORIES
Ÿ Methanol 5 ltr. (3050)
OPTIONAL ACCESSORIES
CALsys -80/50
-80 0 50
0.04
0.06
0.08
0.02
0
0.1
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
Temp. Range : -80 to 50°C
Stability : ±0.04°C at -80°C
±0.07°C at 0°C
±0.05°C at 50°C
Uniformity : ±0.05°C at -80°C
±0.09°C at 0°C
±0.07°C at 50°C
Temperature Resolution : 0.1°C
Time to Reach Min. Temp. : 2 Hrs
Medium : Methanol
Controlling Sensor : PT-100
Volume : 9 Ltr.
Access Opening : 120 x 120 mm
Depth : 200 mm
Method of control : Digital self tuned PID Controller
Computer Interface : RS - 232
Operating Temperature : 20 to 25°C
Power Requirement : 230 VAC, 2.0 KW
Dimensions : 630 (W) x 1200(H) x 500 (D)mm
Weight : 114.7 Kg (without packing)
SPECIFICATIONS
KEY FEATURES
P Large Immersion Depths
P High Accuracy
P High Stability and Uniformity
P Wide Temperature Range
P PC Interface
P Simple to use and Cost effective
Carrying CaseInterchangeable Inserts
(as per requirement)
CALsys -40/50
-40 0 50
0.04
0.06
0.08
0.02
0
0.1
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
CALsys -40/50 is a Highly Stable Stirred Liquid Bath for
Calibration RTD / Thermocouple and also for
compensation of cold junction. CALsys -40/50 offers
the temperature range of -40 to 50°C. It has large tank
which is filled with liquid Methanol. The liquid is
heated or cooled to the desired temperature. It has
better stability & uniformity than dry block bath and
as it uses of stirred liquid. This CALsys -40/50 is built
using CFC free refrigerants.
-40 to 50°CCALSYS-40/50
HIGH STABILITY BATH
Ÿ Reference Standard PRT...Part No. TPRT-A-300 NABL accredited calibration certificate - 3 point.
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual.
STANDARD ACCESSORIES
SPECIFICATIONS
Temp. Range : -40 to 50°C
Stability : ±0.04°C at -40°C
±0.07°C at 0°C
±0.05°C at 50°C
Uniformity : ±0.05°C at -40°C
±0.09°C at 0°C
±0.07°C at 50°C
Temperature Resolution : 0.1°C
Time to Reach Min. Temp. : 1.5 Hrs
Medium : Methanol
Controlling Sensor : PT-100
Volume : 9 Ltr.
Access Opening : 120 x 120 mm
Depth : 200 mm
Method of control : Digital self tuned PID Controller
Computer Interface : RS - 232
Operating Temperature : 20 to 25°C
Power Requirement : 230 VAC, 2.0 KW
Dimensions : 630(W) x 1200(H) x 500(D) mm (Approx.)
Weight : 60 Kg (without packing)
KEY FEATURES
ü Large Immersion Depths
ü High Accuracy
ü High Stability and Uniformity
ü Wide Temperature Range
ü PC Interface
ü Simple to use and Cost effective
CALsys -35/200 is a Highly Stable Stirred Liquid Bath
for Calibration RTD / Thermocouple and also for
compensation of cold junction.
CALsys -35/200 offers the temperature range of -35 to
200°C. It has large tank which is filled with liquid
(Water+Methanol/Silicon Oil). The liquid is heated or
cooled to the desired temperature, where
Methanol+Water are used for Negative Temperature
and Silicon Oil is for Heating purpose.
It has better stability & uniformity than dry block bath
and as it uses of stirred liquid. This CALsys -35/200 is
built using CFC free refrigerants.
-35 to 200°CCALSYS-35/200
HIGH STABILITY BATH
Ÿ Reference Standard PRT…Part No.TPRT-A-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME.
Ÿ Operational Manual
STANDARD ACCESSORIES
KEY FEATURESü High Accuracy
ü High Stability and Uniformity
ü Wide Temperature Range
ü PC Interfacing
ü Simple to Use and Cost Effective
Temperature Range : -35 to 200°C
Stability : ±0.04°C at -35°C
±0.05°C at-0°C
±0.06°C at 200°C
Uniformity : ±0.03°C at -35°C
±0.06°C at 0°C
±0.07°C at 200°C
Temperature Resolution : 0.1°C
Time to Reach Max Temp : 2 Hrs
Medium Cooling : Methanol + Water
Medium Heating : Silicon Oil
Controlling Sensor : PT-100
Volume : 10 Ltr.
Access Opening : 105(L) x 105(W)
Depth : 150 mm
Method of Control : Digital self tuned PID Controller
Computer Interface : RS - 232
Operating Temperature : 20 to 25°C
Wetted Part : SS-304
Power Requirement : 230 VAC, 2.0 KW
Dimensions : 558(L) x 350(W) x 462(H)mm
Weight : Approx 45 Kg
SPECIFICATIONS
Ÿ Methanol............10 Ltr. No.(3050-M)
Ÿ Silicon Oil............10L Part No.C200S-LB SO
OPTIONAL ACCESSORIES
0.03
CALsys -35/200
-35 0 200
0.04
0.05
0.06
0.07
0.08
0.02
0.01
0
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
CALsys -35/200
CALsys -30/110
0-30 0 110
0.02
0.04
0.06
0.08
0.1
Temperature in °C
Tem
pe
ratu
re in
°C
STABILITY UNIFORMITY
Calsys -30/110 is low temperature dry block furnace.
The dry block furnace based on thermoelectric
cooling circuitry. This model provides an special
design isothermal enclosure (Aluminium block) in
which the thermocouple/RTD can be calibrated
against the temperature of the calibrator. The
calibrator controller uses a precision RTD as a
controlling sensor and controls the well temperature
with thermoelectric cooler. To obtain and maintain a
required temperature the controller varies the power
to the heater via solid-state relay. There is one
electricity driven fan which is situated under the
heating chamber for cooling the heater. The dry block
calibrator was designed for portability, moderate cost
and ease of operation.
LOW TEMPERATURE DRY BLOCK CALIBRATOR
Temperature Range : -30 to 110 °C
Temperature Resolution : 0.1°C
Stability : ±0.04°C at -30°C
±0.06°C at 0°C
±0.07°C at 110°C
Uniformity : ±0.05°C at -30°C
±0.07°C at 0°C
±0.08°C at 110°C
Controlling Sensor : Precision PRT
Method of Control : Digital self tuned
PID Controller
Fixed Insert Construction : 1 hole of 8mm and
2 hole of 6.5mm
with 120mm length
Time to Reach Max/Min Temp : 10/20 Mins
Operating Temperature : 20 to 25°C
Computer Interface : RS - 232
Power Requirement : 230 VAC, 500 W
Dimensions : 380(H) x 170(W) x
188(D) mm
Weight : 11.70Kg (without
packing)
SPECIFICATIONS
-30 to 110°CCALSYS-30/110
Carrying CaseInterchangeable Inserts
(as per requirement)
ü Wide Operating Range (-30 to110°C)
ü High Stability
ü PC Interfacing
ü Simple to use and cost effective
ü Portability
ü Low temperature calibration facility upto -30°C
ü Conveniently handle and calibration
ü Light Weight
KEY FEATURES
CALsys -15 / 110 offers medium temperature range
from -15 to 110°C. It is a highly stable standard
temperature source for ca l ibrat ing RTD
/thermocouples. It has peltier elements to generate
stable temperature in the well. The comparison
volume is a metallic block of aluminum, which has
diameter of 24mm and 120mm length. The
temperature of the calibrator is set and controlled by
a self tuned PID controller with automatic super fine
adjustment.
PELTIER BASED TEMPERATURE CALIBRATOR
Ÿ Reference standard PRT......part no.TPRT -A-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - cal soft including for setting bath
temperature and monitoring the PV. graphical
representations of PV/time with 2 hours data logging.
Ÿ Operational manual
Ÿ Carry case
STANDARD ACCESSORIES
Temperature Range : -15 to 110°C
Temperature Resolution : 0.1°C
Stability : ±0.03°C at -15°C
±0.05°C at 0°C
±0.07°C at 110°C
Uniformity : ±0.05°C at -15°C
±0.07°C at 0°C
±0.09°C at 110°C
Controlling Sensor : Precision PRT
Method of Control : Digital self tuned PID
Controller
Insert Construction : Dia 24x120 mm long,
3x6 mm hole with
115mm insertion
depth
Time to Reach Max/Min Temp : 25 Mins
Operating Temperature : 20 to 25°C
Computer Interface : RS - 232
Power Requirement : 230 VAC, 500 W
Dimensions : 380(H) x 170(W) x
188(D) mm
Weight : 11.70 Kg (without
packing)
SPECIFICATIONS
-15 to 110°CCALSYS-15/110
CALsys -15/110
0-15 0 110
0.02
0.04
0.06
0.08
0.1
Temperature in °C
Tem
pe
ratu
re in
°C
STABILITY UNIFORMITY
ü Wide Operating Range (-15 to 110°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
ü Portability
KEY FEATURES
Carrying CaseInterchangeable Inserts
(as per requirement)
CALsys 120SP offers medium temperature range from 0 to 120°C. It is a highly stable standard temperature source for calibrating RTD /thermocouples. It has peltier elements to generate stable temperature in the well. Use as a dry block & liquid bath using magnetic stirrer in one model. The comparison volume is a metallic block of aluminum, which has diameter of 24mm and 100mm length. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment.
0 to 120°CCALSYS120SP
PELTIER BASED TEMPERATURE CALIBRATOR
Temperature Range : 0 to 120°C
Temperature Resolution : 0.1°C
Stability : ± 0.03°C at 0°C
±0.05°C at 50°C
±0.07°C at 120°C
Uniformity : ±0.05°C at 0°C
±0.07°C at 50°C
±0.09°C at 120°C
Controlling Sensor : Precision PRT
Medium : Silicon Oil (for liquid bath)
Method of Control : Digital self tuned PID Controller
Insert Construction : Dia 24 x 100 mm long, 3 x 6 mm hole
Time to Reach Max/Min Temp : 25 Mins
Operating Temperature : 20 to 25°C
Computer Interface : RS - 232
Power Requirement : 230 VAC, 500 W
Dimensions : 380(H) x 170(W) x 188(D) mm
Weight : 14 Kg (without packing)
SPECIFICATIONS
Carrying CaseInterchangeable Inserts
(as per requirement)
CALsys 120SP
0 50 120
0.04
0.06
0.08
0.02
0
0.1
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
Ÿ Reference Standard PRT......Part No.TPRT -A-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
Ÿ Carry Case
STANDARD ACCESSORIES
ü Wide Operating Range (0 to120°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
ü Portability
KEY FEATURES
CALsys 300SP calibration bath is a highly stable
standard stirred liquid bath for calibrating
RTDs/Thermocouples and other temperature sensors
in the laboratory. The temperature of the bath is set
and controlled by a self tuned PID controller with
automatic super fine adjustment.
The unique features is portability with very high
stability and uniformity. The medium used in the bath
is silicon oil.
HIGH ACCURACY LIQUID CALIBRATION BATHGENERAL
50 to 250°CCALsys300 SP
Ÿ Silicon Oil......... Part No.C300S-LB-SO
Ÿ Reference Standard PRT……Part No.TPRT-A-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath
temperature and monitoring the PV. Graphical
representations of PV/TIME with 2 hours data
logging.
Ÿ Operational Manual
Ÿ Carry Case
STANDARD ACCESSORIES
Silicon Oil..........Part No C200S-LB-SO
OPTIONAL ACCESSORIES
Temperature Range : 50 to 250°C
Temperature Resolution : 0.1°C
Stability : ±0.05°C at 50°C
±0.07°C at 150°C
±0.1°C at 250°C
Uniformity : ±0.1°C at 50°C
±0.15°C at 150°C
±0.2°C at 250°C
Controlling Sensor : Precision PRT (PT-100)
Method of Control : Digital self tuned PID Controller
Medium of Heating : Silicon Oil
Time to Reach Max Temp : 45 Mins
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 2.0 KW
Dimensions : 600(H) x 350(W) x 550(D) mm
Access Opening : 100mm x 100mm
Depth : 150mm
Weight : 20 Kg(without packing)
SPECIFICATIONS
Note : Large depth and Access opening available on customer request.
ü Large Immersion Depths
ü Wide Operating Range (50 to 250°C)
ü High Stability and Uniformity
ü PC interfacing
ü Simple to use and cost effective
KEY FEATURES
CALsys 300SP
50 150 250
0.10
0.15
0.20
0.05
0
0.25
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
Provide superior thermal environment for probe immersion as no air gap exist between the probe and the medium Thermal coupling is therefore much better than the alternatives described, and the stirring results in very even heat distribution throughout the medium. Methanol is used for temperature below 0°C, water from 0 to 80°C and silicon oil for upto 250°C. We have seven types of liquid baths given below. CALsys 250 calibration bath is a highly stable standard stirred l iquid bath for calibrating RTDs / Thermocouples and other temperature sensors in the laboratory. The temperature of the bath is set and controlled by a self tuned PID controller with automatic super fine adjustment. The unique features is portability with very high stability and uniformity. The medium used in the bath is silicon oil.
50 to 250°CCALSYS250
HIGH ACCURACY LIQUID CALIBRATION BATH
SPECIFICATIONS
Temperature Range : 50 to 250°CTemperature Resolution : 0.1°CStability : ±0.015°C at 50°C ±0.025°C at 150°C ±0.05°C at 250°CUniformity : ±0.02°C at 50°C ±0.04°C at 150°C ±0.06°C at 250°CControlling Sensor : Precision PRT (PT-100)Method of Control : Digital self tuned PID
ControllerMedium of Heating : Silicon OilTime to Reach Max Temp : 30 MinsComputer Interface : RS - 232Operating Temperature : 20 to 45°CPower Requirement : 230 VAC, 1.0 KWDimensions : 360(H) x 185(W) x
285(D) mmAccess Opening : 90 mm DiaDepth : 140 mmRemovable Insert : 6 holes of 15 mm DiaConstruction Weight : 12 Kg (without
packing)Volume : 900 mlWetted Parts : Aluminium
KEY FEATURES
ü Large Immersion Depths
ü Wide Operating Range (50 to 250°C)
ü High Stability and Uniformity
ü PC interfacing
ü Simple to use and cost effective
ü Portability
0
CALsys 250
50 150 250
0.010.020.030.040.050.060.070.080.09
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
Ÿ Silicon Oil......... Part No.C300S-LB-SO
Ÿ Reference Standard PRT……Part No.TPRT-A-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath
temperature and monitoring the PV. Graphical
representations of PV/TIME with 2 hours data
logging.
Ÿ Operational Manual
Ÿ Carry Case
STANDARD ACCESSORIES
CALsys 650
050 350 650
0.02
0.04
0.06
0.08
0.1
Temperature in °C
Tem
pe
ratu
re in
°C
STABILITY UNIFORMITY
CALsys 650 offers medium temperature range from 50 to 650°C. It is a highly stable standard furnace for calibrating thermocouples / RTD.
The comparison volume is a metallic block of special material, which has diameter of 32mm and 150mm long The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment.
50 to 650°CCALSYS650
PORTABLE CALIBRATOR
Temperature Range : 50 to 650°C
Temperature Resolution : 0.1°C
Stability : ±0.025°C at 50°C
±0.035°C at 350°C
±0.07°C at 650°C
Uniformity : ±0.04°C at 50°C
±0.07°C at 350°C
±0.09°C at 650°C
Controlling Sensor : T/C “N” TYPE
Method of Control : Digital self tuned PID Controller
Insert Construction : Dia 32 x 150 mm long with 4 holes of 6.5 x 120 mm insertion depth
Time to Reach Max Temp : 25 Mins
Cooling Time 650 to 100°C : 70 Mins
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 1 KW
Dimensions : 325(H) x 185(W) x 265(D) mm
Weight : 8.80 Kg (without packing)
SPECIFICATIONS
Ÿ Customized Equalizing Block....Part No. EQ1
OPTIONAL ACCESSORIES
Carrying CaseInterchangeable Inserts
(as per requirement)
KEY FEATURESü Large Immersion Depths
ü Wide Operating Range (50 to 650°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
Ÿ Reference Standard Thermocouple ('K' Type T/C)…… Part No. TTCK-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
Ÿ Carry Case
STANDARD ACCESSORIES
CALsys FB offers medium temperature range from 50 to 650°C. It is a high stable fluidized bath for calibration thermocouple & RTD. It has been designed for medium temperature calibration with high satiability and maximum uniform temperature zone. Alumina powder is used as medium. Metallic probes are used with probe holder for calibration which is 280 mm long with 8mm diameter. Temperature of furnace is set and controlled by self tuned PID controller with automatic super fine adjustment.
SPECIFICATIONS
Temperature Range : 50 to 650°CDed Bed : 0.05°CStability : ±1.0°CDisplay Resolution : 0.1°CType of Control : Digital self Tuned
PID ControllerSensor Type : N type
thermocoupleTime to reach Max. Temp. : 105 MinutesAir pressure kPa (psi) : 420 (60)Weight of medium : 16 kg (35 lbs)Working volume : 150 x 385mmOverall Size L x W x H : 870 x 515 x
600mmWeight : 84 kg (185 lbs)
(without Packing)Serial communication : RS 232Power Requirement : 230VAC, 3.0 KWOperating temp. : 20 to 45°C
50 to 650°CCALSYSFB
FLUIDIZED BATH
Ÿ Reference Standard Thermocouple ('N' type T/C)....Part No. TTCN-300
STANDARD ACCESSORIES
Ÿ Customized Equalizing Block....Part No. EQ1
OPTIONAL ACCESSORIES
ü High Uniform zone
ü Wide Operating Range (50°C - 650°C)
ü High Stability
ü No. of Sensor Calibrated Simultaneously
KEY FEATURES
CALsys 1200 offers high temperature range from 250 to 1200°C. It is a highly stable standard furnace for calibrating Thermocouples / RTD. It has been designed for high temperature range calibration and find application in the glass, electrical power, automotive & material processing industries. The comparison volume is a metallic block of special material, which is 37 mm in diameter with 215 mm long. The temperature of the calibrator is set and controlled by a self tuned PID controller with automatic super fine adjustment. Version with 3 zone furnace is also available.
250 to 1200°CCALSYS1200
PORTABLE THERMOCOUPLE CALIBRATOR
SPECIFICATIONS
Temperature Range : 250 to 1200°CTemperature Resolution : 1.0°CStability : ±0.1°C at 250°C ±0.2°C at 700°C ±0.3°C at 1200°CUniformity : ±0.20°C at 250°C ±0.24°C at 700°C ±0.36°C at 1200°CControlling Sensor : PT/RH-PT T/CMethod of Control : Digital self tuned
PID ControllerInsert Construction : Dia 37 x 215 mm
long (2X6mm & 2X8 mm holes) of 160mm insertion depth
Time to Reach Max Temp : 1.5 HrsCooling Time 1200°C to 250°C : 180 MinsComputer Interface : RS - 232Operating Temperature : 20 to 45°CPower Requirement : 230 VAC, 1.5 KWDimensions : 405(H) x 205(W) x
285(D) mmWeight : 13.65 Kg (without
packing)
Ÿ Customized Equalizing Block....Part No. EQ1
OPTIONAL ACCESSORIES
Carrying CaseInterchangeable Inserts
(as per requirement)
0.25
0.20
0.15
0.10
0.05
0
CALsys 1200
0.30
0.35
0.40
0.45
250 700 1200
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
KEY FEATURESKEY FEATURESü Large Immersion Depths
ü Wide Operating Range (250 to 1200°C)
ü High Stability
ü PC interfacing
ü Simple to Use and Cost Effective
Ÿ Reference Standard Thermocouple ('N'Type T/C)……Part No. TTCN-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
Ÿ Carry Case
STANDARD ACCESSORIES
KEY FEATURESü Large Immersion Depths
ü Wide Operating Range (300 to 1200°C)
ü High Stability
ü PC interfacing
ü Simple to Use and Cost Effective
CALsys 1200L calibration source is a highly stable standard furnace for calibrating thermocouples in the laboratory. The temperature of the furnace is set and controlled by a self tuned PID controller with automatic super fine adjustment. The standard insert is a metallic block of special material, which is 37mm in diameter with 240 mm long and can hold up to four thermocouples.
It has been designed for high temperature range calibration and find application in the glass, electrical power, automotive, material processing industries & laboratories. Special Version with 3 zone furnaces are available.
Temperature Range : 300 to 1200°C
Temperature Resolution : 1.0°C
Stability : ±0.1°C at 300°C
±0.2°C at 700° C
±0.35°C at 1200°C
Uniformity : ±0.15°C at 300°C
±0.25°C at 700°C
±0.4°C at 1200°C
Controlling Sensor : Precision PT/RH-PT T/C
Method of Control : Digital self tuned PID Controller
Insert Construction : Dia 37 x 240 mm long (2X6mm & 2X8 mm holes) of 160mm insertion depth
Time to Reach Max Temp : 1.5 Hrs
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 2.0 KW
Dimensions : 590(H) x 450(W) x 530(D) mm
Weight : 50 Kg (without packing)
SPECIFICATIONS
300 to 1200°CCALSYS1200L
HIGH STABILITY BLOCK FURNACE
Ÿ Customized Equalizing Block....Part No. EQ1
OPTIONAL ACCESSORIES
Optional Black
0.25
0.20
0.15
0.10
0.05
0
CALsys 1200L
0.30
0.35
0.40
0.45
300 700 1200
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
Ÿ Reference Standard Thermocouple ('N' Type
T/C)……Part No. TTCN-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath
temperature and monitoring the PV. Graphical
representations of PV/TIME with 2 hours data
logging.
Ÿ Operational Manual
STANDARD ACCESSORIES
CALsys 1200 3Z calibration source is a highly stable standard furnace for calibrating thermocouples in the laboratory. The temperature of each zone of the furnace is set and controlled by a self tuned PID controller(master + Slave) with automatic super fine adjustment. The standard insert is a metallic block of special material, which is 37mm in diameter with 240 mm long and can hold up to four thermocouples. It has been designed for high temperature range calibration and find application in the glass, electrical power, automotive, material processing industries & laboratories.
Temperature Range : 300 to 1200°C
Temperature Resolution : 1.0°C
Stability : ±0.1°C at 300°C
±0.2°C at 750° C
±0.35°C at 1200°C
Radial Uniformity : ±0.15°C at 300°C
±0.25°C at 750° C
±0.40°C at 1200°C
Axial Uniformity : ±1.0°C up to 80mm at 1200°C
Controlling Sensor : Precision PT/RH-PT T/C
No. of Zone : Three
Method of Control : Digital self tuned PID Controller
Insert Construction : Dia 37 x 240 mm long
(2X6 mm & 2X8 mm holes) of 160mm insertion depth
Time to Reach Max Temp : 1.5 Hrs
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 2.0 KW
Dimensions : 500(H) x 400(W) x 490(D) mm
Weight : 50Kg (without packing)
SPECIFICATIONS
300 to 1200°CCALSYS1200 3Z
HIGH STABILITY BLOCK FURNACE
Ÿ Customized Equalizing Block....Part No. EQ1
OPTIONAL ACCESSORIES
Optional Black
KEY FEATURESü Large Immersion Depths
ü Wide Operating Range (300 to 1200°C)
ü High Stability
ü PC interfacing
ü Simple to Use and Cost Effective
0.5
0.4
0.3
0.2
0.1
0
CALsys 1200 3Z0.6
300 750 1200
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
Ÿ Reference Standard Thermocouple ('N' Type T/C)……Part No. TTCN-300
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
STANDARD ACCESSORIES
CALsys 1500L calibration source is a highly stable standard furnace for calibrating thermocouples in the laboratory. The temperature of the furnace is set and controlled by a self tuned PID controller with automatic super fine adjustment. The standard insert is a Ceramic Tube, which is 37 mm in diameter with 245 mm long and can hold up to four thermocouples. It has been designed for high temperature range calibration and find application in the glass, electrical power, automotive & material processing industries and laboratories.
Temperature Range : 500 to 1500°C
Temperature Resolution : 1.0°C
Stability : ±0.4°C at 500°C
±0.7°C at 1000°C
±1.0°C at 1500°C
Uniformity : ±0.6°C at 500°C
±0.8°C at 1000°C
±1.2°C at 1500°C
Controlling Sensor : PT-RH/PT T/C
Method of Control : Digital self tuned PID Controller
Insert Construction : Dia 37 x 245 mm long (2X6mm & 2X8 mm holes) of 140mm insertion depth
Heaters : Silicon Carbide Spiral Rod Heater
Time to Reach Max Temp : 2 Hrs
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 3.0 KW
Dimensions : 590(H) x 450(W) x 530(D) mm
Weight : 55 Kg (without packing)
SPECIFICATIONS
500 to 1500°CCALSYS1500L
HIGH STABILITY BLOCK FURNACE
Ÿ Reference Standard Thermocouple (PT-RH/PT 'R' Type T/C) ……… Part No. TTCR-300
Ÿ Extra Equalizing Block………Part No. EQ2
OPTIONAL ACCESSORIES
1.0
0.8
0.6
0.4
0.2
0
CALsys 1500L1.2
500 1000 1500
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
KEY FEATURES
ü Use of Silicon Carbide Spiral Rod Heaters
ü Large Immersion Depths
ü Wide Operating Range (500 to 1500°C)
ü High Stability
ü PC interfacing
ü Simple to Use and Cost Effective
STANDARD ACCESSORIES
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
CALsys 1700L calibration source is a highly stable standard furnace for calibrating thermocouples in the laboratory.
The temperature of the furnace is set and controlled by a self tuned advanced PID controller with automatic super fine adjustment. The standard insert is a Ceramic Tube, which is 37 mm in diameter and 240 mm long and can hold up to four thermocouples. It has been designed for high temperature range calibration and find application in the glass, electrical power, automotive & material processing industries and laboratories.
Temperature range : 500 to 1700°C
Voltage : 230 V AC
Power : Maximum
Supply Frequency : 50/60 Hz
Resolution : 0.1°C
Stability : ±0.5°C at 500°C
±1.0°C at 1200°C
±1.5°C at 1700°C
Uniformity : ±0.6°C at 500°C
±1.4°C at 1200°C
±1.9°C at 1700°C
Controlling sensor : B type T/C Duplex
Time to reach max. Temp. : 3 hours
Operating Temperature : 20 to 45°C
Controller Specifications : Eurotherm P104 series
Over temperature protection : Autonics controller with B Type T/C
Stabilization Time : 15 to 20 Min.
Dimensions : 640(H) x 500(W) x 550(D) mm
Insert Construction : 2 x 6.5, 2 x 8.5 mm with 225mm depth
Weight : 80kg (without packing)
SPECIFICATIONS
500 to 1700°CCALSYS1700L
HIGH TEMPERATURE BLOCK FURNACE
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
STANDARD ACCESSORIES
Ÿ Reference Standard Thermocouple (PT-RH/PT 'B' Type T/C) ……… Part No. TTCB-300
Ÿ Extra Equalizing Block………
Ÿ Black Body Cavity……….Part No. EQ3
OPTIONAL ACCESSORIES
1
0.5
0
CALsys 1700L
1.5
2
500 1200 1700
Temperature in °C
Tem
per
atu
re in
°C
STABILITY UNIFORMITY
Extended area black body is defined by the large emitting surface area precise temperature control with good uniformity. The blackbody is designed to provide infrared radiation as an ideal blackbody emitter. Because of the large uniform surface area the
body called extended area black body.
Tempesens offers extended area black body with two emitter size small and large have different temperature range. It desigined accordingly temperature measurement and calibration both are performed in the radiation head and transfer digitally to the controller. It is designed highly standard and stable black body for testing of infrared sensors and thermal detecting devices. We sepeartly provide high programming controller with black body source. High accuracy chamber have designed separately. The temperature of the furnace is set or change by the controller.
LBB100CH & LBB11CH
Ÿ Independent controller with calibrated area.
Ÿ Monitored controlling device.
Ÿ Portable controller.
Ÿ Serial communication and computer interfacing by port
Ÿ Rs-232.
Ÿ Quick and convenient calibration technique.
PROPERTIES
EXTENDED AREA BLOCK BODY
Temperature Range : 15 to 100° CDisplay Resolution : 0.01° CStability : ±0.5°CUniformity : ±2°CControlling Sensor : RTDEmissivity : 0.97Emissive Head : 1000x1000 mmPower Requirement : 230VACMethod of Control : Digital Self tuned PID
ControllerComputer Interface : Rs-232/USBDimensions : 1450 (H) x 1450 (W) x 600
(D) mmWeight : 35 kg
LBB100CH
KEY FEATURES
ü Large area emitter size
ü Good uniformity
ü Capable of operating at wide temperature range
ü High resolution & PC Interfacing
ü Better stability
ü High accuracy
LBB11CH
Temperature Range : 0 to 140° CDisplay Resolution : 0.01°CStability : ±0.01°CUniformity : ±0.08°CControlling Sensor : RTDEmissivity : 0.97Emissive Head : 100x100 mmPower Requirement : 220VAC, 500WMethod of Control : Digital self tuned PID
ControllerComputer Interface : Rs-232/USBDimensions : 610 (H) x 550 (W) x 610
(D) mmWeight : 30 kg
CALsys 110BB calibration source is a highly stable
standard portable black body furnace for calibrating
non contact IR thermometer for the temperature
range of 10 to 110°C.
The unique feature of this portable black body furnace
is large temperature controlled black body target with
a size of 80 mm dia which offer large view area for IR
thermometer.
The emissivity of the target is 0.95(±0.01). The
temperature of the furnace is set and controlled by a
self tuned PID controller with automatic super fine
adjustment.
10 to 110°CCALSYS110BB
PORTABLE BLACK BODY
Temperature Range : 10 to 110°C
Temperature Resolution : 0.1°C
Stability : ±0.05°C at 10°C
±0.07°C at 60°C
±0.1°C at 110°C
Controlling Sensor : Precision PRT
Method of Control : Digital self tuned PID Controller
Emissivity : 0.95 ±0.01
Time to Reach Max Temp : 25 Mins
Computer Interface : RS - 232
Operating Temperature : 20 to 25°C
Power Requirement : 230 VAC, 0.5 KW
Dimensions of cabinet : 330(H) X 355(W) X 225(D) mm
Cavity : High Emissive Painted Aluminum Plate
Aperture : 80 mm dia
Weight : 12Kg (without packing )
SPECIFICATIONS
CALsys 110BB
10 60 110
0.10
0.15
0.20
0.05
0
0.25
Temperature in °C
Tem
per
atu
re in
°C
STABILITY
Ÿ Carry Case
Ÿ Operational Manual
STANDARD ACCESSORIES
Ÿ Master Pyrometer
OPTIONAL ACCESSORIES
ü Large Aperture Area
ü Below Ambient Operating Range (10 to 110°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
ü Portability
KEY FEATURES
Ÿ Master Pyrometer
OPTIONAL ACCESSORIES
CALsys 500BB calibration source is a highly stable
standard portable black body furnace for calibrating
non contact IR thermometer for the temperature
range of 50 to 500°C.
The unique feature of this portable black body furnace
is large temperature controlled black body target with
a size of 100 mm dia which offer large view area for IR
thermometer.
The emissivity of the target is 0.95(±0.01). The
temperature of the furnace is set and controlled by a
self tuned PID controller with automatic super fine
adjustment.
50 to 500°CCALSYS500BB
PORTABLE BLACK BODY
Temperature Range : 50 to 500°C
Temperature Resolution : 0.1°C
Stability : ± 0.5°C at 50°C
±0.8°C at 250°C
±1.0°C at 500°C
Controlling Sensor : T/C “N” type
Method of Control : Digital self tuned PID Controller
Emissivity : 0.95 ±0.01
Time to Reach Max Temp : 45 Mins
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 1.0 KW
Dimensions of cabinet : 270(H) X 360(W) X 270(D) mm
Target Size : Dia - 100 mm (Ridged Aluminum Plate)
Weight : 10Kg (without packing)
SPECIFICATIONS
CALsys 500BB
50 250 500
0.4
0.6
0.8
0.2
0
1.0
Temperature in °C
Tem
per
atu
re in
°C
STABILITY
ü Large Aperture Area
ü Wide Operating Range (50 to 500°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
ü Portability
KEY FEATURES
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
Ÿ Carry Case
STANDARD ACCESSORIES
ü Wide Operating Range (300 to 1200°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
ü Sighting tube
KEY FEATURES
CALsys 1200BB calibration source is a highly stable
standard black body furnace for calibrating non
contact IR thermometer for the wide temperature
range of 300 to 1200°C.
The unique feature of this black body furnace is large
temperature controlled black body target with a
diameter of 40 mm and 85mm which offer large view
area for IR thermometer.
The emissivity of the target is 0.99(±0.01).The
temperature of the calibrator is set and controlled by
a self tuned PID controller with automatic super fine
adjustment.
HIGH STABILITY BLACK BODY FURNACE
Temperature Range : 300 to 1200°C
Temperature Resolution : 1.0°C
Stability : ±0.2°C at 300°C
±0.3°C at 800°C
±0.5°C at 1200°C
Controlling Sensor : Precision PT/RH-PT T/C
Method of Control : Digital self tuned PID Controller
Time to Reach Max Temp : 1.5 Hrs
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 2.5KW
Dimensions of cabinet : 590(H) x 450(W) x 530(D) mm
Radiation Cavity Type : Silicon Carbide
Dimension of cavity : 46 mm Dia x 85mm depth
External Aperture : 40 mm dia
Weight : 5 0 K g ( w i t h o u t packing)
SPECIFICATIONS
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
STANDARD ACCESSORIES
Ÿ Master Pyrometer
Ÿ Black Body Cavity…………Part No. EQ3
OPTIONAL ACCESSORIES
300 to 1200°CCALSYS1200BB
CALsys 1200BB
300 800 1200
0.2
0.3
0.4
0.1
0
0.5
Temperature in °C
Tem
per
atu
re in
°C
STABILITY
Ÿ Black Body Cavity…….Part No. EQ3
OPTIONAL ACCESSORIES
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
STANDARD ACCESSORIES
CALsys 1500BB calibration source is a highly stable
standard black body furnace for calibrating non
contact IR thermometer for the wide temperature
range of 500 to 1500°C.
The unique feature of this black body furnace is large
temperature controlled black body target with a
diameter of 40 mm and 85 mm depth which offer large
view area for IR thermometer.
The emissivity of the target is 0.99(±0.01). The
temperature of the calibrator is set and controlled by a
self tuned PID controller with automatic super fine
adjustment.
HIGH STABILITY BLACK BODY FURNACE
CALsys 1500BB
500 1000 1500
0.4
0.6
0.8
0.2
0
1.0
Temperature in °C
Tem
per
atu
re in
°C
STABILITY
500 to 1500°CCALSYS1500BB
ü Use of Silicon Carbide Spiral Rod Heaters
ü Wide Operating Range (500 to 1500°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
ü Sighting tube
KEY FEATURES
Temperature Range : 500 to 1500°C
Temperature Resolution : 1.0°C
Stability : ±0.4°C at 500°C
±0.7°C at 1000°C
±1.0°C at 1500°C
Controlling Sensor : PT-RH/PT T/C
Method of Control : Digital self tuned PID Controller
Heaters : Silicon Carbide Spiral Rod Heater
Time to Reach Max Temp : 2.5 Hrs
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 3.5 KW
Dimensions of cabinet : 590(H) x 450(W) x 530(D) mm
Radiation Cavity Type : Silicon Carbide,
Dimension of cavity : 46 mm Dia x 85mm depth
External Aperture : 40 mm dia
Weight : 50 Kg (without packing)
SPECIFICATIONS
Ÿ Master Pyrometer
Ÿ Black Body Cavity………. Part No. Eq3
OPTIONAL ACCESSORIES
Ÿ NABL accredited calibration certificate - 3 point
Ÿ Software - Cal Soft including for setting bath temperature and monitoring the PV. Graphical representations of PV/TIME with 2 hours data logging.
Ÿ Operational Manual
STANDARD ACCESSORIES
CALsys 1700BB calibration source is a highly stable
standard black body furnace for calibrating non
contact IR thermometer for the wide temperature
range of 500 to 1700°C.
The unique feature of this black body furnace is large
temperature controlled black body target with a
diameter of 29 mm which offer large view area for IR
thermometer.
The emissivity of the target is 0.97(±0.01). The
temperature of the calibrator is set and controlled by a
self tuned PID controller with automatic super fine
adjustment.
HIGH STABILITY BLACK BODY FURNACE
500 to 1700°CCALSYS1700BB
SPECIFICATIONSTemperature Range : 500 to 1700°C
Temperature Resolution : 1°C
Stability : ±0.7°C at 500°C
±1.4°C at 1200°C
±2.0°C at 1700°C
Controlling Sensor : Precision PT/RH-PT T/C
Heaters : MoSiO2
Method of Controlling : Digital self tuned PID Controller
Time to Reach Max Temp : 3 Hrs
Computer Interface : RS - 232
Operating Temperature : 20 to 45°C
Power Requirement : 230 VAC, 3 KW
Dimensions of cabinet : 640(H) x 500(W) x 550(D) mm
Black Body Cavity Aperture : 29 mm, End closed tube
Weight : 80Kg (without packing)
ü Use of Molybdenum Disilicide Heaters (MoSiO2)
ü Wide Operating Range (500 to 1700°C)
ü High Stability
ü PC interfacing
ü Simple to use and cost effective
KEY FEATURES
CALsys 1700BB
500 1200 1700
1.0
1.5
2.0
0.5
0
2.5
Temperature in °C
Tem
per
atu
re in
°C
STABILITY
KEY FEATURES
ü Water flow failure
ü Increase in water temperature
ü Argon gas flow failure
ü Over current
600 to 3000°CFASTCAL3000
FastCAL 3000 is a high temperature blackbody source
with a capacity to produce high temperature using
high emissivity emitter(graphite tube) and at the
same time stabilize at the required temperature
within a few minutes.
It is mainly used to calibrate non-contact temperature
sensors. Step change in temperature is achievable
within seconds. The graphite tube has an emitter
diameter of 25mm (1.00") and an emissivity of +0.995
±0.005. Target temperature is sensed by a non contact
temperature sensor/Pyrometer which gives a
feedback to PID controller to regulate the cavity
temperature precisely to the desired set points.
Remote set point is achieved via RS-232 C
communication port.
All safety features to protect equipment and operator
are easily accessible.
SPECIFICATIONS
Temperature Range 600°C to 3000°C (1112°F - 5432°F)
Wavelength 1000 nm or 1600 nm (Pyrometer)
Cavity Temp. Accuracy ± 0.25 % of set temperature ±1°C
Temperature Resolution 0.1°C
Temperature Stability ±1°C
Emissivity 0.995 ±0.005
Cavity Closed end graphite tube Length - 292mm Heated Length - 127mm
Heating Aperture/Cavity Dia Graphite tube with argon gas purge dia 25mm
CoolingWater cooled, recommended re-circulation system.All units include water cooled reflector around the blackbody to reduce the heat load to the room. A Chiller unit will be provided with the system.
Purge Gas Dry Argon gas flow with 7-10 LPM
Heating Time Approx 5 Minute from ambient to 3000°C
Safety Interlocks Cooling water over temperature, Low purge gas flow, low cooling water flow rate, over current and over temperature protection.
Controlling Sensor Pyrometer
Method of Control PID controller Eurotherm make(Option - HMI & Data Logging)
Remote Controller Set Point control and temperature monitoring by RS-232/Rs-485
Ambient Temperature 0°C to 50°C
Power 230 VAC ±10°C 50Hz (Three phase/Two Phase/Single Phase)
Dimension 1700mm(H) x 900mm(W) x 1200mm(D)
Weight 300 Kg
KEY FEATURES
ü 0°C Thermoelectric reference unit
ü Eliminates Old Fashioned “Ice Bath”
ü Versatile use in industries, Laboratories, Instrument shop
ü NABL Traceable Calibration Available
KEY FEATURES
ü 60°C Thermoelectric reference unit
ü Versatile use in factory, Laboratory, Instrument Shop
ü Rugged Outer case for safe Portability
ü NABL Traceable Calibration Available
CALREF 0
Reference Junction Temp. : 0°C(Standard)
Types of T/C : J, K, T, E, N, R, S, B
Accuracy : ±0.1°C, Errors can be compensated by adjusting controller setting
Stability : ±0.03°C
Stabilization Time : 15 Mins
Capacity : 20 No (User Defined)
Resolution : ±0.01°C
Dimension : 315(H) x 305(W) x 332(D)mm
Weight : 13Kgs
Power Supply : 230 VAC, 50 Hz
Carry Case : Aluminium modular box
SPECIFICATIONS
CALREF 60
SPECIFICATIONS
Temperature Range : +10 to 90
Types of T/C : J, K, T, E, N, R, S, B
Accuracy : ±0.5°C, Errors can be compensated by adjusting controller setting
Stability : ±0.05°C
Stabilization Time : 10 Mins
Capacity : 6-24 Channel (User Defined)
Max. Capacity : 48 Channel
Resolution : ±0.1°C
Dimension(CRCA) : 350(H) x 350(W) x 210(D) mm
Dimension(FRP) : 330(H) x 330(W) x 180(D) mm
Weight : 15Kg(CRCA), 10Kg(FRP)
Power Supply : 230 V AC, 50/60 Hz / 24 V DC
KEY FEATURES
ü 2 4 C h a n n e l C a l i b r a t i o n ( 1 2 N o . f o r thermocouples & 12 No. for RTD's)
ü Includes Easy to use Connectors
ü Internal CJC compensation
ü Data Logging of selected channel
ü Communicat ion wi th CALsys Model .& PC/Laptop
242424
The Autocal 24 is an automatic temperature calibration system for the thermocouple and RTD's. The system consist of furnaces, the autocal device and PC software, which together contribute to the whole cycle of auto calibration process.
Each thermocouple and RTD input is designated as channel for the system. The system accept 24 channels, 12 thermocouples & 12 RTD's. The connection for these channels through special type of locking connectors. The channel switching can be done with PC software/LCD display via touch screen keypad. The thermocouple microvolt & RTD ohm reading for each channel is monitored with CJC compensation. After the calibration process complete the PC software generate a report of actual calibrated values for the inputs.
ü Display LCD Display with Touch Screen
ü Resolution (°C) 0.01°C
ü Resolution (mV) 6 ½ digit
ü Display Unit °C / °F / mV / ohms
ü Mounting 19” rack / suitable for carry case
ü Calibration With accredited calibration certificates
ü Main Supply 230 VAC, 50 Hz
SPECIFICATIONS
AUTOMATIC CALIBRATION SYSTEM
AUTOCAL SYSTEM
KEY FEATURESü High stability
ü High accuracy of RTD measurement 0.01°C
ü High accuracy of T/C measurement 0.1°C
ü High Resolution
ü Measuring inputs
ü 10 Thermocouples B, C, D, E, R, S, T, N, J, K
ü 6 RTD's PT-100, PT-10, PT-50, PT-500, PT-200, PT-1000
ü Thermocouple reference junction external, internal and off
ü Units °C/ °F
ü Data Logging 4000 values
ü Suitable for 2/3/4 wire RTD.
Extended area black body is defined by the large
emitting surface area precise temperature control
with good uniformity. The blackbody is designed to
provide infrared radiation as an ideal blackbody
emitter. Because of the large uniform surface area the
body called extended area black body.
Tempesens offers extended area black body with two
emitter size small and large have different
temperature range. It desigined accordingly
temperature measurement and calibration both are
performed in the radiation head and transfer digitally
to the controller. It is designed highly standard and
stable black body for testing of infrared sensors and
thermal detecting devices. We sepeartly provide high
programming controller with black body source. High
accuracy chamber have designed separately. The
temperature of the furnace is set or change by the
controller.
CALSYSC-4004
PRECISION THERMOMETER
Display : 192x64 LCD Graphic Panel with back light
Input : 2 Channels for RTD and Thermocouple via lemo 1-S series connector
Working Temperature : Less than 95% non condensing condition
Storage Temperature : -20 to +55 °C
Mains Supply : 85-264 VAC / 120-370 VDC 47-63 Hz
Data Logging : Up to 4000 single channel values can be stored along with date and time
Dimensions : 224(H) x 318(W) x 90(D) mm
SPECIFICATIONS
Ÿ Direct Thermocouple measurement
Ÿ Direct RTD measurement
APPLICATIONS
Specifications
RTD + T/C type RTD + T/C range Accuracy
Pt100 0 0-200 C ~ +850 C ±0.05°C
S 0 00 C ~ 1768 C ±2.40°C
B 300°C ~ +1820°C ±2.40°C
T 0 0-200 C ~ +400 C ±1.20°C
J 0 0-210 C ~ +1200 C ±1.20°C
R 0°C ~ +1768°C ±2.40°C
E 0 0-200 C ~ +1000 C ±1.20°C
N -200°C ~ +1300°C ±1.80°C
K 0 0-200 C ~ +1372 C ±1.50°C
TEMPMET 08 - RTD + THERMOCOUPLE
INPUT
Thermocouple : K, S, E, T, J, R, B, N
RTD : Pt100
Unit : °C, F, K
Resolution : 0.0°C/0.0°F
Specifications
TEMPMET 09 - RTD + THERMOCOUPLE
RTD + T/C type RTD + T/C range Accuracy
Pt100 0 0-200 C ~ +850 C ±0.02% RDG +0.060°C
mA -2.0 ~ +24.0mA ±0.03% RDG +3µA
J -210 ~ +1200°C ±0.50°C
S -0.0 ~ 1768°C ±0.8°C
E -200 ~ +1000°C 0.40°C
T -200 ~ +400°C ±0.50°C
K -200°C ~ +1372°C ±0.50°C
R 0.0 ~ 1768°C ±0.8°C
mV -100.0 ~ +200.0mV ±0.015% RDG +10µV
Cu100 0 0-50 C ~ +150 C ± 0.060°C
Cu50 0 0-50 C ~ +150 C ± 0.080°C
Pt1000 0 0-140 C ~ +320 C -
Ω -0.000 ~ 2200 Ω ±0.02% RDG +50mΩ
B 300 ~ +1820°C ±0.9°C
N -200 ~ +1300°C ±0.50°C
INPUT
Thermocouple : K, S, E, T, J, R, B, N
RTD : Pt1000, Cu50, Cu10
mV : -100 to 200mV
mA : -2 to 24 mV
Resistance : 0 to 500 Ohms
Unit : Ohms, mA, mV, °C, F, K
Resolution : 0.000°C/0.000°F, 0.000
A/V/mA/mV
PRECISION THERMOMETER
TEMPMET
Operation Temperature and Humidity :
0°C ~ 50°C, <80% RH
Storage Temperature and Humidity :
0 0-10 C ~ 60 C, <70% RH
Power Requirement :(AA Size) batteries, up to 1500 hours of continuous work time
Input Protection at Thermocouple Input :
24V AC/DC Maximum
Dimension(Without holster) :
115 mm X 70 mm x 30 mm (LxWxH)
Weight :Approx.0.25 Kg (Include Battery)
Operation Temperature and Humidity
0 00 C ~ 50 C, <80% RH
Storage Temperature and Humidity
0 0-10 C ~ 60 C, <70% RH
Power Requirement(AA Size) batteries, up to 1500 hours of continuous work time
Input Protection at Thermocouple Input :
24V AC/DC Maximum
Dimension(Without holster) :
115 mm X 70 mm x 30 mm (LxWxH)
Weight :Approx.0.25 Kg (Include Battery)
THERMOGRAPHY SERVICES
Tempsens provide thermography services for various industries.
Thermography enables to monitor the thermal efficiency of
critical process systems that rely on heat transfer of retention.
This is one of the most powerful, fast and one of the most cost-
effective condition monitoring technique that has wide
application in any industry in detecting incipient faults, if left
unattended, would not only lead to loss of productivity and
quality but also increase operations and maintenance costs.
Tempsens Calibration Center is an independent unit of Tempsens instruments (I) Pvt. Ltd, having laboratories at Udaipur, Vadodara & Bangalore. It is accredited for wide range of temperature calibration services.It is the only private sector Laboratory in the country with accredited Fixed Point Temperature calibration Facilities. The lab has highly stable calibration furnaces, measuring instruments and accurate master sensors traceable to National and International Standards.
IN HOUSE CALIBRATION FACILITY
ON SITE CALIBRATION FACILITY
QualityMeasured/
Instruments
Temperature Range
Non Contact TypePyrometer
Calibration &Measurement
Capability
Contact type RTD.ThermocouplesThermometers
-25°C to 0°C>0°C to 140°C
>140°C to 250°C>250°C to 650°C
>650°C to 1200°C
0.07°C0.04°C0.09°C0.12°C1.30°C
0°C to 250°C>250°C to 500°C
>500°C to 1200°C
1.50°C2.40°C2.5°C
Multipoint Position Calibration of
Chamber, Oven, Furnaces (Thermal
Mapping(TUS))
-80°C to 200°C>200°C to 1200°C
2.8°C4.1°C
BangaloreLab
C-1226
The calibration center functions as per ISO 17025 / NABL
standards. Calibration of contact type sensors can be made in
temperature range of -196°C to 1600°C and Calibration of non
contact type sensors can be made in temperature range 0°C to
2900°C. Further the laboratory is accredited for onsite
temperature calibration.
The lab offer both at Lab & On-Site Calibration of Furnace/Bath
from -80°C to 1600°C and Black Body Calibration from 50°C to
1700°C.
Furnace/Chamber Calibration (TUS) with multiple sensors from -
80°C to 1200°C is also in the scope of the lab.
Sample Thermography Report
CALIBRATION SERVICES
FIXED POINT CALIBRATION FACILITIES
QualityMeasured/
InstrumentsTemperature Range
Calibration &
Measurement
Capability
Calibration of SPRT/PRTS/thermocouple
etc.
Triple Point of Water (0.01°C)Melting Point of Gallium (29.7646°C)
Freezing Point of Tin (231.928°C)Freezing Point of Zinc (419.527°C)
Freezing Point of Aluminum (660.323°C)
0.0038°C0.0065°C0.0065°C0.0071°C0.0075°C
Calibration ofThermocoupleat SecondaryFixed Point
Melting Point of Gold(1064.18 °C) 0.72°C
Melting Point of Palladium(1554.8 °C)
0.83°C
Contact Type RTD,ThermocouplesThermometers
Non Contact TypePyrometer
0°C to 250°C>250°C to 500°C
>500°C to 1500°C>1500°C to 1700°C>1700°C to 2900°C
1.5°C2.4°C2.5°C3.2°C4.0°C
QualityMeasured /Instruments
Temperature RangeCalibration & Measurement
Capability
-38°C to 0°C>0°C to 140°C
>140°C to 250°C>250°C to 650°C
>650°C to 1200°C>1200°C to 1600°C
-196°C -80 to -38°C
0.05°C0.03°C0.03°C0.04°C0.04°C0.12°C1.30°C2.60°C
Tempsens Instruments (l) Pvt. Ltd. U# llA-190, Road No. 5, M.I.A., Udaipur-313003 (Rajasthan) INDIAPh. : +91-294-3052900, Fax : +91-294-3052950Email : [email protected]
INDIA GERMANY
Tempsens Instruments GmbHLoehestrasse 37,53773 Hennef, GERMANYPh. : +49 2242 8703-22, Fax : +49-2242-8703-20, Email: [email protected] [email protected]
INDONESIAPt. Tempsens Asia JayaJl. Jembatan lll Komplex F.77 No. 6CJakarta Utara14450, INDONESIAHp. : (+62)877 8080 4433Telp. : (+62) 21 666 04006Fax : (+62) 21 661 6789Email : [email protected]
THERMAL & CABLE SOLUTIONS
UAE
Tempsens Gulf Oilfield Equipment LLCUnited Arab EmiratesEmail:[email protected]