Troubleshooting Failed QC
Dr Douglas ChesherClinical Biochemistry, NSW Health Pathology, Royal
North Shore HospitalNorthern Clinical, School University of Sydney
![Page 2: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/2.jpg)
![Page 3: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/3.jpg)
Outline
• Types of errors• Approach to troubleshooting• Problem fixed, now what?• Lot number changes
![Page 4: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/4.jpg)
Experimental Errors
• Systematic Error• Random Error• Blunders
![Page 5: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/5.jpg)
Systematic Errors
• Due to identified causes and can, in principle, beeliminated. Errors of this type result in measuredvalues that are consistently too high orconsistently too low.
• Types– Instrumental– Observational. For example, parallax in reading a
meter scale.– Environmental– Theoretical. Due to simplification of the model system
or approximations in the equations describing it.
http://www.physics.nmsu.edu/research/lab110g/html/ERRORS.html
![Page 6: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/6.jpg)
Random Errors
• Random errors are positive and negativefluctuations that cause about one-half of themeasurements to be too high and one-half to betoo low.– Sources of random errors cannot always be identified.
• Types of random error– Observational. For example, errors in judgment of an
observer when reading the scale of a measuringdevice to the smallest division.
– Environmental. For example, unpredictablefluctuations in line voltage, temperature, ormechanical vibrations of equipment.
http://www.physics.nmsu.edu/research/lab110g/html/ERRORS.html
![Page 7: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/7.jpg)
Blunders
• An outright mistake.• Should stick out like sore thumbs if we
make multiple measurements or if oneperson checks the work of another.
http://www.physics.nmsu.edu/research/lab110g/html/ERRORS.html
![Page 8: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/8.jpg)
“My QC Failed”
![Page 9: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/9.jpg)
![Page 10: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/10.jpg)
![Page 11: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/11.jpg)
Response to a failed QC
• Stop• Do not just rerun the QC
– Unless part of your rule (eg. 1 2s rerun ‘once’)– Regression towards the mean
![Page 12: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/12.jpg)
Response to failed QC
• Do not just recalibrate as a routine.– Calibrations add noise and increase
imprecision.
![Page 13: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/13.jpg)
120
125
130
135
140
145
150
1 3 5 7 9 11131517192123252729313335373941434547495153555759
126128130132134136138140142144146148
1 3 5 7 9 11131517192123252729313335373941434547495153555759
![Page 14: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/14.jpg)
Points of failure• Instrument Failure:
– Check for error messages (printed or displayed). Refer to thetroubleshooting section of the instrument manual.
• Reagent, Quality Control Materials, Calibrators:– In correct reconstitution, i.e. wrong or contaminated diluent or
use of the wrong volume of diluent.– In correct storage, e.g. left at room temperature for excessive
amount of time.– Prepare fresh (allow enough time to dissolve)– NOTE: Bulk liquid reagents may be contaminated.
• Human:– Reread the method - Check whether the correct sample volumes
or reagents were used. Make sure that a step in the procedurewas not missed.
– Were the correct parameters entered?
![Page 15: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/15.jpg)
Have other tests also failed?
• Blunders– Swapped QC levels
• Short sampled
![Page 16: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/16.jpg)
Are there any instrument flags?
![Page 17: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/17.jpg)
What rule has failed?
• Systematic error– 1 3s, 2 2s, 4 1s, 10x etc
• Random error– 1 3s, R4s
![Page 18: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/18.jpg)
Review your QC Chart
142
143
144
145
146
147
148
Serum X
-3SD
-2SD
-1SD
Mean
1SD
2SD
3SD
Systematicerror
Systematicerror
![Page 19: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/19.jpg)
![Page 20: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/20.jpg)
Has anything changed?
• Just calibrated• New bottle of reagent• New lot of reagent
![Page 21: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/21.jpg)
New reagent lotNew reagent lotQuarterlymaintenance
Quarterlymaintenance
![Page 22: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/22.jpg)
Review points of failure
• Reagents– Correct reagent?, sufficient volume, shelf expiry
date, on-board expiry date• Quality Control Material
– Material, lot no. & assigned value, preparation,shelf expiry date, in-use expiry date, storageconditions.
• Calibrators– Correct material, lot no. & assigned values,
preparation, storage, expiry date. Inspectcalibration trace of last calibration.
![Page 23: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/23.jpg)
Review points of failure
• Instrument– Is maintenance up to date?– Recheck flags, probes, lamps, cuvettes, water
bath.
![Page 24: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/24.jpg)
Correctiveaction
requiredimmediately
?
Correctiveaction
requiredimmediately
?
Instrumentflags?
Instrumentflags?
Maintenance up todate?
Maintenance up todate?
ReagentsOK?
ReagentsOK?
QC OKQC OK
CalibratorsOK?
CalibratorsOK?
QC FailureQC Failure
InstrumentOK?
InstrumentOK?
Monitor QC performanceand correct non-urgentfaults as appropriate
Monitor QC performanceand correct non-urgentfaults as appropriate
Correct instrument faultCorrect instrument fault
Perform all outstandingmaintenance
Perform all outstandingmaintenance
Correct reagent faultCorrect reagent fault
Make up and run correctQC
Make up and run correctQC
Make-up and run correctcalibrator
Make-up and run correctcalibrator
Fix identified instrumentfault
Fix identified instrumentfault
Rerun QC to confirmproblem fixed
Rerun QC to confirmproblem fixedCall for Help!Call for Help!
Y
N
Y
Y
N
N
N
N
Y
Y
Y
N
![Page 25: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/25.jpg)
Problem Solved!
![Page 26: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/26.jpg)
Problem solved, now what?
• Run QC for evidence that problem has beensolved.
• Document what you have done.Troubleshooting logs, QC annotation.
• Address patient results from previous goodQC to when QC failure occurred.– Repeat testing
• Exclude / Inactivate failed QC from dataanalysis if cause of the outlier is clearlyidentified.
![Page 27: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/27.jpg)
Reagent lot changes
• CLSI EP26-A User Evaluation of Between-Reagent Lot Variation; Approved Guideline.
• “The protocol attempts to balance the need toreliably detect clinically significant change inreagent performance that may affect patientresults with the recognition that reagent lotverification is a relatively frequent task thatputs demands on the laboratory’s limitedresource”
![Page 28: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/28.jpg)
QC material is not alwayscommutable• Shift in QC may not reflect a similar shift in
patient results• Just because QC does not show a shift, does
not mean the patients will not show a shift.
• Verify all new lots before they are put intouse.– Need only perform once for group of labs if using
same lot QC and reagent.
![Page 29: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/29.jpg)
Test patient samples withcurrent and candidate
reagent lots
Test patient samples withcurrent and candidate
reagent lots
Estimate averagedifference between lots
Estimate averagedifference between lots
Averagediff < CD?Average
diff < CD?
Was QCacceptable with new
lot?
Was QCacceptable with new
lot?
Investigate lot difference.Do not report patientresults with this lot
Investigate lot difference.Do not report patientresults with this lot
Update QCtargets
Update QCtargets
Candidate lot acceptablefor patient testing
Candidate lot acceptablefor patient testing
YesYes
NoNo
CD: Critical Difference
![Page 30: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/30.jpg)
Defining the critical difference
• Evaluation of comparability based on clinicaloutcomes.
• Evaluation of comparability based on clinical decisions– derived from either biological variation or– from data based on clinicians’ opinions.
• Published professional recommendations.• Performance goals set by regulatory bodies or
organizers of External Quality Assessment Schemes.• Goals based on the current state of the art as
demonstrated by data from EQA or from currentpublications on methodology.
![Page 31: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/31.jpg)
Determine the statisticalpower
Determine the statisticalpower
Determine the number ofsamples to be tested and
the rejection limit
Determine the number ofsamples to be tested and
the rejection limit
Determine the number ofreplicates per sample
Determine the number ofreplicates per sample
Determine the number oftarget concentrations
Determine the number oftarget concentrations
• Alpha (Type I error, false positive)• Probability of detecting an error
when there is no error.• Typically set at 0.05
• Beta (Type II error, false negative)• Probability of not detecting a
difference when there is adifference
• Typically set at 0.20 to 0.05• Statistical Power = 1 – Beta.
Generally measure at clinical decisionpoints, often similar to the QC levels.
Statistical tables may suggest that only onepatient is required but it is recommendedthat at least three samples are used toavoid random errors.If difficult to get sufficient samples,replicate analysis may provide thenecessary additional statistical power.However, under normal circumstancesmeasure in singlicate
![Page 32: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/32.jpg)
![Page 33: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/33.jpg)
Conclusions
• Do not automatically re-run QC or recalibratean assay.
• Type of QC failure and Quality Control Chartsmay give a clue to the nature of the problem.
• Have a structured approach totroubleshooting.
• Have a process for managing your patientresults
• Prevention is better than cure– Manage introduction of new lots to the laboratory
![Page 34: Troubleshooting Failed QC - AACB](https://reader030.fdocuments.in/reader030/viewer/2022012508/6184c75d75b8a876312f42b3/html5/thumbnails/34.jpg)