Troubleshooting DNA Sequences: Guidelines and Suggestions .

56
Troubleshooting DNA Sequences: Guidelines and Suggestions http://vermontcancer.org/dna

Transcript of Troubleshooting DNA Sequences: Guidelines and Suggestions .

Page 1: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Troubleshooting DNA Sequences:

Guidelines and Suggestions

http://vermontcancer.org/dna

Page 2: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Biodesktop Workshops:

• Scheduled: 300 HSRF– Jan. 29th at 10-11AM (Friday)

• Rama Kocherlakota– Basics of submitting orders– Retrieving data– Creating a team to share data– Opportunity to provide input

Page 3: Troubleshooting DNA Sequences: Guidelines and Suggestions .

SESSION OUTLINE:• Guidelines: Generic Set up and Profiles

• Impact of Template and primer ratio???

• Suggestions for different sample types and Sequence Context: Chemistry, Profile, Additives– Sample types: PCR, Plasmid, BAC, Cosmid, RNAi construct, Bisulfite

treated gDNA, gDNA– Sequence Context: GC Rich, Homopolymeric Runs, Repetitive Sequence

• Instrument and Processing Anomolies:

• Retreiving data off the biodesktop: How, What…..

• Sample Purification prior to Sequencing

• Troubleshooting Resources

Page 4: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Sequencing Instruments: AB 3100-Avant, 3130XL both

Capillary Based

• Advantages– Higher throughput– Can reinject samples– Higher separation

efficiency– Better resolution– No Plates!

• Disadvantages– Sensitive to charged

ions– Sensitive to

microparticulates or bubbles

Page 5: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Required Template Concentrations:

Template Type Required templateconcentration

Totalamount in Rxn

Double strandplasmid

50 ng/uL 375 ng

Large plasmid (>7Kb) 50 ng/uL 525 ng

Single strandplasmid

25 ng/uL 75 ng

PCR 0-200 bp

5 ng/uL 10 ng

200-500 bp 5 ng/uL 20 ng

500-1000 bp 5 ng/uL 40 ng

>1000 bp 5 ng/uL 45 ng

PCR (Exo-Sap) N/A 1-2 ul

BAC and gDNA 1ug/uL 1-2 ug

*This amount of template is needed for each primed sequencing reaction

Page 6: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Normal Conditions: Default Profile

AutoSeq1 Profile

• 96° 1min

• 96 ° 10 sec• 50 ° 5 sec 25x• 60 ° 4 min

C.S. Rxn conditions

DS plasmid-375 ng

PCR (5ng/100bp Product)

3.2 pmol Primer

1/8 dilution BDv3.1

Page 7: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Primer Titration: Plasmid

Page 8: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Template Titration: Plasmid

Page 9: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Template Titration: PCR ProductPCR Product Size=~720BP

70 ng added

30ng added

Page 10: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Template Concentration: It Can MatterSubmitting template at the

incorrect concentration

Submitted as:50 ng/ulWas:350 ng/ul

Diluted to:50 ng/ul

Page 11: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Sample Type: RNAi construct, BAC, Cosmid, gDNA, Bisulfite treated DNA

Different Sample Types May Require Different Template or primer

concentration, Chemistry, Profile, and additives

Page 12: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Sequencing RNAi Constructs:Auto Seq1 Profile, Chemistry (Default)

Page 13: Troubleshooting DNA Sequences: Guidelines and Suggestions .

RNAi Construct:GC Rich Profile, 5% DMSO

Page 14: Troubleshooting DNA Sequences: Guidelines and Suggestions .

RNAi Construct:Modified RXN Set-up, RNAi Profile

660 ng Template10 pMol Primer8ul BDT v.3.110% Betaine(Q Buffer)

98 c 5min

96 c 15 sec50 c 10 sec60 c 4 min

50X

Page 15: Troubleshooting DNA Sequences: Guidelines and Suggestions .

RNAi Construct: testing to reduce costsAutoSeq1 (Default):

Page 16: Troubleshooting DNA Sequences: Guidelines and Suggestions .

RNAi Construct:Default Chemistry, RNAi Profile:

Page 17: Troubleshooting DNA Sequences: Guidelines and Suggestions .

RNAi Construct:BDTv3.1/dGTP Chemistry, GC RichProfile:

Page 18: Troubleshooting DNA Sequences: Guidelines and Suggestions .

RNAi Construct:BDTv3.1/dGTP Chemistry, RNAi Profile:

Page 19: Troubleshooting DNA Sequences: Guidelines and Suggestions .

RNAi Construct: LOR scores for three different approaches

AutoSeq1

GC Rich

RNAi

ThermalProfiles:

Page 20: Troubleshooting DNA Sequences: Guidelines and Suggestions .

BAC’s, Cosmid’s, Genomic:

BAC DSRG Profile

• 96° 5min

• 96 ° 30 sec• 50 ° 20sec• 60 ° 4 min

C.S. Rxn conditions

DNA- 1ug

10 pmol Primer

straight BDv3.1

50X

Page 21: Troubleshooting DNA Sequences: Guidelines and Suggestions .

BAC’s: Default set-up and AutoSeq1

Page 22: Troubleshooting DNA Sequences: Guidelines and Suggestions .

BAC’s: Modified Set-up, BAC profile

Page 23: Troubleshooting DNA Sequences: Guidelines and Suggestions .

BAC Sequencing: LOR scores for two different approaches

Page 24: Troubleshooting DNA Sequences: Guidelines and Suggestions .

gDNA Sequencing: Not as easy as you may think!

1.3 Kb productgDNA = 500 ngFWD Primer: 500nMREV Primer: 100nM125 uM dNTP’sStraight Big Dye

Cycling:950C 30s

950C 15s500C 15 s 35-45600C 4 min

PCR and Cycle Sequencing combined:

Murphy, K., et.al., Clinical Chemistry 51:1 35-39 (2005)

Page 25: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Bisulfite Sequencing:Sequencing methylated gDNA

Default Set-up and Profile

Page 26: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Bisulfite Sequencing: Suggested Set-up and profile

BiSulSeq Profile: “ Vish”

• 95° 1min

• 96 ° 10 sec• 52 ° 10sec 30x• 60 ° 4 min

C.S. Rxn conditions

PCR 5ng/100 bp

3.2 pmol Primer

1/8 dilution BDv3.1

“Vish II”: 96c 1min, 25X of 96c 20s,50c 5s, 54c 4minChemistry BDT V1.1

Page 27: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Bisulfite Sequencing: Default Set-up and BiSulSeq Profile

“Vish Profile”

Page 28: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Cosmids:

BAC DSRG Profile

• 96° 5min

• 96 ° 30 sec• 50 ° 20sec• 60 ° 4 min

C.S. Rxn conditions

DNA- 1ug

10 pmol Primer

straight BDv3.1

50x

Page 29: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Sequence Context Constraints:

GC rich, Homoploymeric runs, Repetitive sequence (STR)

Page 30: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Run of G’s:Default Set-up and Profile (AutoSeq1)

Page 31: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Run of G’s:dGTP Chemistry, AutoSeq1 profile

Page 32: Troubleshooting DNA Sequences: Guidelines and Suggestions .

GC Rich Template:Generic Set up, AutoSeq1 Profile

Page 33: Troubleshooting DNA Sequences: Guidelines and Suggestions .

GC Rich Template:BDTv3.1/dGTP (3:1), GC Rich profile

Previous stop point

Page 34: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Repetititve Sequence: Template C Defaults

Stops

Page 35: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Repetititve Sequence: Template C BDT v3.1/dGTP (3:1) mix, GC Rich Profile

Page 36: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Repetititve Sequence: Template D BDT v3.1/dGTP (3:1) mix, GC Rich Profile

Lunatic!!

Page 37: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Repetititve Sequence: Template E BDT v3.1/dGTP (3:1) mix, GC Rich Profile

Page 38: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Repetititve Sequence: Template A BDT v3.1/dGTP (3:1) mix, GC Rich Profile

Not always a fix!

Page 39: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Repetititve Sequence: Template A BDT v1.1, Default Profile

Page 40: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Homopolymeric runs:

“Throw the Book”

Primer Location

Page 41: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Retrieving Data:• An email is sent notifying the user that data is ready.

The staff of the DNA facility has the ability to edit this message to include specific remarks about how your samples ran, so please look at this message!

• The format for the sequence data files: Well Location_template-primer, (A01_pGem-M13For).

In these messages, we often refer to your individual samples by the injection number (A01), not the full name.

Instructions

*

*

Page 42: Troubleshooting DNA Sequences: Guidelines and Suggestions .

What is Phred?

• Basecaller (used by genome centers)• Provides quality score for each base • Generates two files

Page 43: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Need to look at the Chromatogram!! (.AB1 File)

Trace viewers:PC: Mac:

Chromas EditView Finch TV Finch TV

Page 44: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Instrument and Processing Related Anomolies:

AND Email notifications

Page 45: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Bad Injection message:One of your samples had a bad injection. It is being re-injected at no

charge and we will send the data to you when it is ready.

Example:

Causes SolutionsGenerally, this is due to incomplete excess dye clean up of the cycle sequence reaction.

If the DNA facility cleaned the reaction, we consider this an error on our part and we will repeat the cycle sequence reaction and the sequence run for free.If the clean up was done by the user, we charge for the sample, but we will be happy to help with troubleshooting if the problem continues.

Dye blobs can sometimes occur when the cycle sequence reaction was a very low signal.

Optimize the template and primer for a successful cycle sequence reaction.

Page 46: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Loss of Resolution: In the middle

Reinjection successful!

Page 47: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Timing of Reinjections:

C fluorophore degrading

Reinjections on Monday from a Friday run may need to be set-up again

Page 48: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Dye-Blob message:• Your sequence sample, injection number (A01) has a dye leak which interferes with the basecalling in

that region. These cannot be corrected with a reinjection of the sample. It is our policy to repeat the cycle sequence reaction and sequence run once for free to eliminate this problem. If the dye leak was not a problem for you we will not repeat the sample. If it is a problem, resubmit the sample on the BioDesktop and put “re-run for free due to a dye leak” in the comment section of the order form. Please make sure that we have enough template DNA to repeat the reaction.

Example:

Causes Solutions

Generally, this is due to incomplete excess dye clean up of the cycle sequence reaction.

If the DNA facility cleaned the reaction, we consider this an error on our part and we will repeat the cycle sequence reaction and the sequence run for free.If the clean up was done by the user, we charge for the sample, but we will be happy to help with troubleshooting if the problem continues.

Dye blobs can sometimes occur when the cycle sequence reaction was a very low signal.

Optimize the template and primer for a successful cycle sequence reaction.

Page 49: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Spike message:One of your samples had spikes. It is being re-injected at no charge

and we will send the data to you when it is ready.

Causes Solutions

This is generally believed to be a sequencing instrument artifact. It is caused when a small bubble or crystals in the polymer migrate through the capillary with the sample and cause a sharp peak of multicolors. These can mask the true data peaks.

Re-injection of the sample usually gets rid on the spike.

Example:

Page 50: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Template Purification:

What’s best for sample or sequence type

Page 51: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Impact of Purification Method on Sequence Quality: Gel Purified

Run 1

Run 2

PCR Product size= ~630BP

Page 52: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Impact of Purification Method on Sequence Quality: Enzyme Treated

Same sample: Exo1/SAP treated PCR Product

Page 53: Troubleshooting DNA Sequences: Guidelines and Suggestions .

DNA Sequencing Troubleshooting Resources:

• VCC DNA Analysis Websitewww.vermontcancer.org/dna

• DSRG Sequencing Troubleshooting Web Resource www.abrf.org/index.cfm/stwr.home

• DNA Sequencing Resource: Nucleics http://www.nucleics.com/nucleics-dna-sequencing-site.html

• http://biowww.net/

• DNA Sequencing: Optimizing the Process and Analysis

• DNA Facility Staff

Page 54: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Conclusions: Successful Sequencing is Dependent on:

• Template Quality

• Template Quantity

• Upfront Identification of Sample Type

• Upfront identification of Sequence context constraints

• Read Email messages sent by facility staff

Page 55: Troubleshooting DNA Sequences: Guidelines and Suggestions .

Acknowledgements:

VCC DNA Analysis Facility:

MaryLou Shane

Meghan Brown

Scott Tighe

To all users of the VCCDNA Analysis Facility

Special thanks to all the users who have providedtemplate for research studies and those who shared their data for this presentation

Page 56: Troubleshooting DNA Sequences: Guidelines and Suggestions .

VersaDoc 4000MP:Keep in Mind these very important facts about X-Ray film vs CCD imagers:

1] X-Ray film exposures are not quantitative although many people believe they are, it is not the case. The dynamic range is so narrow, and is non-linear that to assume quantitative results can be very misleading. When a band is black, it is black! Please make note of the Gamma curve for X-Ray film.

2] The VersaDoc has a CCD imager in it. CCD’s are unique as they are true photon counters. The higher the bits, the higher the dynamic linear range. The VersaDoc uses a 16 bit CCD! That is what they were originally designed for, and that is why they are quantitative. Many reviewers also know this as well.3] Some people believe that CCD’s are not as sensitive as X-Ray film. This also is not the case. If you add the substrate to the blot and image immediately, the result is essentially the same. The difference people notice often comes from a time delay to walk to the imagers after you add substrate. That is why we suggest adding your substrate in the lab that the VersaDoc Imager is located in.