TROUBLE SHOOTING GUIDE FOR FRESENIUS 4008 HEMODIALYSIS UNITS
General Guidelines:
1) Make sure that the following errors / alarms are not present before running the T1 test.
a) Power Failure
b) Dialines not conn
c) Shunt cover open
d) Connect conc.Line
e) Blood sensed by OD
f) Flow alarm
g) Water alarm
2) When T1 test fails, press “Test” key again and note the displayed error code before proceeding to technical manual for further clues.
3) If more than 1 test fails select “Test Service” ON (PCB LP 632 DIP switch array 1, SW 4 ON) & do the test individually to get the error codes of the concerned test fail.
Test Equipments:
a) Hemodialysis tester
b) Pressure gauge
c) Graduated Measuring jar (100ml , 1000ml)
d) Flashlight.
e) Digital multimeter
Pressure Units:
1bar = 14.5 psi =749.94mmHg
The Order of Troubleshooting :
1) Water Alarms
2) Flow Alarms
3) Upper Flow Alarm
4) Temperature
5) Conductivity
6) Blood Leak
7) Pressure Holding Test Failures.
8) TMP Continuous Alarms
9) Hydraulic Test Fails
10) Blood Pump Errors / Level Detector Alarms /Arterial Alarms/ Venous Alarms /Heparin Pump Errors/ Diasafe Errors
11) Ultra Filtration Errors
12) Battery Fail
S NO. Error Possible Causes Tips Remarks 1
Water Alarm
a) No Water Supply/ Low Pressure.
b) Pressure Regulator (61)
c) Check the Inlet Filter
d) Water Inlet Valve (41)
a) Water Inlet Pressure(0.95 to 1.05 Bar)
c) Check Valve 41
Use Diagnostics Program to check the water Flow Path.
2 Flow Alarm a) Blocked filter
b) Defective valve/pump (valve resistance 50 to 80 Ohms)
c) Calibration
a) Verify flow 300, 500, 800
b) Calibrate if required.
a) Create flow from diagnostics.
b)Verify Loading pressure (1.2 – 1.45 Bar)
3 Upper Flow Alarm
a) Degassing filter
b) Degassing pressure
c) Flow calibration
4 Temperature
a) High temperature
b) Low temperature
a) Incoming water temperature
b) Temp Calib
a) Temp Calib
b) Heater Rod.
a) Incomplete hot disinfection can be a reason!
a) Flow errors can lead to Low Temp Alarm!
b) Heater rod resistance around 30 Ohms
S NO. Error Possible Causes
Tips Remarks
5
Conductivity
i) Low Conductivity
a) Dilution Ratio.
b) Concentrate or Bicarbonate
c) Conc. &
a) Verify Dilution Ratio in Set Up Menu
b) Wrong or Expired Solution??
a)Normally 1+35.83(Nacl26);depends on the A & B part
b) Provided the concentrate and BIC strokes and suction is OK
c) Ref: Calibration-Mixing System
ii) High Conductivity
BIC pump strokes and output??
d) Conc and BIC pump spring set broken??
e) Conductivity (Display ) Calibration
f) Faulty Conc or BIC Pump
a)Flow Errors
b)Dilution
c) Concentrate or Bicarbonate
d) Conductivity (Display ) Calibration
c) Calibrate if required
d) Replace if required
e) Temperature and Conductivity (display) Calibration
a) Verify flow ,Calibrate if required
b) Verify Dilution
c) Wrong or Expired Solution??
d) Temperature and Conductivity (display) Calibration
d) Calibrate the Conc and BIC pumps after replacing the springs. Use the graduated measuring Jar to measure the stroke volume.
c) Ref: Set up- Dilution
S NO. Error Possible Causes
Tips Remarks
6 TMP / Dial Pressure
a) Dialyser membrane defects
b) Dial Pressure Calibration Required.
a) Replace Dialyser.
b)Continuous TMP Alarm / Pressure fail indicates Dial Pressure Calibration
a)Adjust 0 mmHg ( LP 633 Pot P3) and do dial pressure calibration.
7 Blood Leak Blood Leak and Dimness Calibration
Adjust both to 5 Volts.
8 Pressure Holding Test
i) Positive Pressure Errors
ii) Negative Pressure Errors
a) Leakage in Hydraulic system.
b) Dial Pressure Calibration
a) UF filter blockage
b) UF Pump Calibration
c) Broken UF Pump Spring
a) Observe for leaks during Positive pressure test or Induced pressure test.
c) Observe/ Replace Spring set
a) Use a flash light.
b) Do Dial Pressure calibration only if the error code indicates dial pressure calibration.
b) Use the graduated measuring Jar.
d) Use Diagnostics program
set
d) Minute Leakage in Hydraulic system
d) Check tightness of hoses
S NO. Error Possible Causes
Tips Remarks
9 Hydraulic Test Fail
i)Rinse Failure Alarm
ii) V39 Failure.
a) Rinse Filters 148 & 149
b)Rinse Valves 91,99,100
c) Bibag Pressure switch 134
a) Clogged
Filter 210
a) Replace Filters if needed.
b) Most cases pressure leak due to loose hose at these valves
OR Leak from Conc and BIC pick up line rubber seals.
c) Check PS 134 from Diagnostics.
a) 3 minutes before the end of cleaning program.
a) 1 minute before the end of cleaning program.
10 i) Blood Pump Errors
ii)Level detector Alarm
iii) Arterial Pressure Alarms
a) BP Power cord
b) Jammed Motor
c) Control Board
a) Air Detector
a) Arterial Pressure calibration
a) Replace power cord
b) Replace bearing
a) Calibrate Air detector
a) Power cord is the one connecting control board to motor
c) Do complete arterial calibration and BP speed calibration after replacing the control board
S NO. Error Possible Causes
Tips Remarks
10 Contd… iv) Venous Pressure Alarms
v) Heparin Pump Errors
vi) Diasafe Errors
a) Transducer Protector ( P Ven)
b) Venous Pressure calibration
c) Venous Line set.
a) Mechanical problem of gear
a)F 03/ F02
b)F 34/ 43
c) Diasafe Pressure Transducer
Partially clogged female connection.
a) Manually rotate the gear to Top or bottom end and retest the pump
a) Dial Pressure calibration required
b) Damaged Diasafe Filter
a)
c) Replace Diasafe filter.
11 Ultrafiltration Errors.
i) Less fluid removal
a)UF Pump filter (74)
b) UF Pump Calibration
b) Verify
b) Verify the stroke volume (1ml/ stroke)
c) UF pump spring set
d) UF Pump
Calibration
c)Replace if required
d) Replace if defective.
12 Battery Fail.
a) Battery
b) Battery Fuse (3.15 A)
c) LP 647
a) Battery Vol. 18 V .
c) LP 647 is in the Power supply module.
Diagnostics Program
1) On the service switch
2) Enter Diagnostics- Write outputs – Write Digital outputs – CPU 1: Digital outputs.
Rinsing V 84 ( Disinfection Valve)
1) On V 87, 30, 32, 34, 36, 38, 84 (On - 1111)
2) Start Flow Pump (FP On - 0000)
Creating the Flow
1) Switch On V 24, 24 B, 30,31,32,33,34,35,36,37,38,87
2) Start Flow Pump and Degassing pump
3) Set On FL_ SW_ EN (Float Switch Enable)
Note: Balancing Chamber inlet /outlet valve combination can be switched off one group at a time to locate the faulty BC valve, if any.