Page 1: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020

Analyzer

Service Manual

July 2009

Page 2: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Windows is a registered trademark of Microsoft Corporation.

© Micromeritics Instrument Corporation, 2009. All rights reserved.

Page 3: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TABLE OF CONTENTS

1. GENERAL INFORMATIONOrganization of the manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1Using This Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

2. FUNCTIONAL DESCRIPTIONOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1New Product Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

Cabinetry & Sample Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2Interior Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

Instrument Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3Cabinetry/Sample Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

Top Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3Instrument Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4Side Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

Interior Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6Gas entrance manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6Dosing volume manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6Sample analysis manifold. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7

Schematic Board & Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8Printed Circuit Boards (PCBs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9

Analysis Transducer Interface PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9Card Cage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9

DSP-ISA CPU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10Digital Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10Backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10

Power Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11

3. THE ANALYSIS SEQUENCEAnalysis Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2Analysis Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3

4. PREINSTALLATION

5. INSTALLATION

Jul 09 i

Page 4: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Table of Contents TriStar II 3020 Service Manual

6. OPERATOR TRAINING

7. MAINTENANCE DOCUMENTS

8. CALIBRATION PROCEDURE

9. MECHANICAL AND ELECTRICAL DRAWINGSMechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

10. SERVICE PARTS

11. SOFTWARE

12. OPERATOR’S MANUAL

13. SERVICE TRAINING

ii Jul 09

Page 5: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Organization of the manual

1. GENERAL INFORMATION

This service manual contains information and instructions for providing instrument service on the Micromeritics TriStar II 3020 Analyzer and is intended for use by Micromeritics' factory-trained and supported service personnel.

Organization of the manual

The service manual is organized as follows:

Chapter 1 GENERAL INFORMATION

Provides information about the contents of the service manual.

Chapter 2 FUNCTIONAL DESCRIPTION

Provides a functional description of analyzer components.

Chapter 3 THE ANALYSIS SEQUENCE

Provides a description of the analysis sequence.

Chapter 4 PREINSTALLATION

Provides Preinstallation requirements and checklist.

Chapter 5 INSTALLATION

Provides Installation Instructions.

Chapter 6 OPERATOR TRAINING

Provides a checklist for training operators on the use of the analyzer.

Chapter 7 MAINTENANCE DOCUMENTS

Provides scheduled maintenance documents.

Chapter 8 CALIBRATION PROCEDURE

Provides a copy of the Factory Calibration Procedure.

Jul 09 1-1

Page 6: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Using This Manual TriStar II 3020 Service

Using This Manual

This manual is most efficiently used from the Micromeritics web site. However, it can be downloaded to a portable media device, such as a CD or a USB stick.

Chapter 9 MECHANICAL AND ELECTRICAL DRAWINGS

Contains mechanical and electrical drawings.

Chapter 10 SERVICE PARTS

Contains a list of service replacement parts.

Chapter 11 SOFTWAREContains the demo software for the TriStar II 3020 Analyzer.

Chapter 12 OPERATOR’S MANUAL

Provides links to current versions of the operator manuals.

Chapter 13 SERVICE TRAINING

Provides a syllabus for training service personnel.

1-2 Jul 09

Page 7: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Overview

2. FUNCTIONAL DESCRIPTION

This chapter describes the systems and components that comprise the TriStar II 3020 Analyzer. It includes graphics showing the location of major systems, modules, and components, as well as functional descriptions.

Overview

As the latest addition to the TriStar Analyzer family, the TriStar II 3020 is a fully automated, three-station physisorption instrument that provides high-quality surface area and porosimetry measurements on solid materials by using the gas adsorption technique.

The TriStar II 3020 has three ports, allowing analysis of up to three samples simultaneously and independently. A user can operate up to four TriStar analyzers with one Windows-based computer.

Jul 09 2-1

Page 8: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

New Product Features TriStar II 3020Service

New Product Features

Cabinetry & Sample Chamber

• The TriStar II 3020 is smaller than the TriStar 3000 due to elimination of the vacuum pump cavity.

• The vacuum pump is now located external to the cabinet.

• The rotating door allows easier access to the sample ports for mounting and removing sample tubes.

Interior Modules

• Solid-state transducers with stainless steel diaphragms replace the exposed-silicon transducers of the 3000 model.The TriStar II 3020 adds a continuous saturation pressure measurement feature.

• An additional port has been added to the gas inlet manifold to bring the total number of ports, for readily available gases, to four.

Software

• The use of any of the IUPAC standard measurement units is available on the TriStar II 3020.

• Data from the TriStar 3000 and Gemini can be imported to the 3020 to allow overlay plots and other comparisons to be generated.

• Information can be scanned directly into input fields using a bar code reader. This elimi-nates human error on data entry.

• The TriStar II 3020 software allows opening of multiple sample files simultaneously for editing.

• Diagnostics may be performed remotely by way of the Internet. This allows MIC engi-neers to access the instrument and run diagnostics programs.

2-2 Jul 09

Page 9: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Instrument Components

Instrument Components

The TriStar 3020 Analyzer is comprised of these major components and modules:

• Cabinetry/Sample Chamber• Interior• Printed Circuit Boards (PCBs)• Power Module

Cabinetry/Sample Chamber

Due to the elimination of the vacuum pump cavity, the redesigned TriStar II 3020 analyzer is significantly smaller than its predecessor.

Top Panel

The top panel of the TriStar provides access to the interior analysis module and its subassemblies. The top panel is molded from plastic and internally coated with a special conductive coating to prevent radio frequency interference.

Front

The redesigned front of the TriStar II 3020 consists of a tinted and transparent rounded sample chamber door. The front panel is intended to shield the operator from any flying debris resulting from a sample tube explosion or liquid nitrogen splatter. When the door is opened, it allows for easy, 270-degree access to the three sample ports.

Top Panel

Elevator

Dewar

Sample Tubes

Door

Communications status indicator

Jul 09 2-3

Page 10: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Cabinetry/Sample Chamber TriStar II 3020Service

Elevator

The TriStar II 3020 elevator raises the dewar to the sample tubes. During analysis, it is automatically raised and lowered. The elevator can be manipulated by the user when the system is in manual mode.

Dewar

The Dewar for the TriStar 3000 carried over to the TriStar 3020 has a capacity of 2.75-L.

Sample Tubes

The TriStar has the ability to test up to three samples at a time The tubes are made out of borosilicate glass to withstand the temperatures reached inside the Dewar or during degassing.

Communications status indicator

A green LED indicator is located on the front of the TriStar’s top panel to indicate the instrument has power.

Instrument Back

The back area of the TriStar is where the gas entrance manifold, the vacuum pump connection and the power connection are located.

Gas entrance manifold

Vacuum pump connector

Power switch

Power cord connector

2-4 Jul 09

Page 11: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Cabinetry/Sample Chamber

Vacuum Pump

A new feature of the TriStar II 3020 is that the vacuum system is outside the instrument cabinet. Wet and dry pumps are available for nitrogen systems, while the krypton option must use a dry pump.

Side Panels

The left panel of the TriStar is plain. The right panel, however, provides access to the unit’s card cage, ethernet, and RS-232 connection.

Oil/ Wet pump Oil-free / Dry Pump

Jul 09 2-5

Page 12: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Interior Modules TriStar II 3020Service

Interior Modules

The analysis module consists of three subassemblies discussed below.

Gas entrance manifold

The Gas Entrance Manifold provides four gas inlet ports, that face the rear of the instrument. The inlets have dust filters (frits) to prevent dirt from gas lines getting into the system. This manifold also has the vacuum connection.

The servo valve sets dosing pressures in the manifold and regulates the rate of pressure decrease during gas evacuation.The manifold transducer is also located on the gas entrance manifold. All other valves are solenoid valves.

Dosing volume manifold

The Dosing Chamber or Volume Chamber is an aluminum chamber that has a capacity of 18 mL of volume. The volume chamber does not have any valves attached to it in order to prevent

TransducerServo valve

Gas inlet ports

Vacuum pump connector

Gas Inlet Photo Gas Inlet CAD

Solenoid valves

Dosing Volume CAD

2-6 Jul 09

Page 13: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Interior Modules

local fluctuations of temperature, which would reduce the precision of gas measurements. There is a temperature sensing device (RTD) mounted to the chamber.

Sample analysis manifold

The Sample Manifold contains the three sample ports and a P0 port for the instrument. Each of the TriStar’s three ports has a 20-µm frit to prevent sample from entering and clogging the system, a small-orifice solenoid valve for dosing, a large-orifice solenoid valve for evacuation, 1000-Torr transducer, and a pressure relief valve to protect against excess pressure build up caused by too rapid desorption. The relief valves vent at 10 to 20 psig, even if there is no power applied to the instrument and protect the sample transducers, sample tubes and dewar. The P0 port is used for continuous saturation pressure measurement.

Sample ports

Transducer

Pressure relief valve

Solenoid valve

Dosing volume manifold

10-Torr Transducer (Krypton units only)

Sample Analysis Photo Sample Analysis CAD

Jul 09 2-7

Page 14: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Schematic Board & Software TriStar II 3020Service

Schematic Board & Software

The schematic diagram of the 3020 is displayed on the computer screen by the TriStar software. The schematic shows:

• The current state of each valve.• The pressure reading from the transducers.• The temperature reading in the manifold.• Location and status of the elevator.

The software is designed to work on computers running Windows XP or Windows Vista. The software also supports importing data from the TriStar 3000 and Gemini for analysis and comparison with TriStar II 3020 plots. For more instructions on how to use the software and run the TriStar analyzer, refer to Chapter 3 - Operational Procedures in the Operator’s Manual.

1

2

3

45

6789

10

11

12

1 - Gas Inlet Manifold2 - Vacuum Pump3 - Servo Valve4 - Manifold Transducer

5 - Manifold Thermocouple6 - Port 1 Transducer7 - Port 3 Transducer8 - Port 2 Transducer

9 - P0 Transducer10 - Elevator11 - Status Bar12 - 10-Torr Transducer (Krypton only)

All transducers are 1000-Torr unless otherwise stated.

2-8 Jul 09

Page 15: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Printed Circuit Boards (PCBs)

Printed Circuit Boards (PCBs)

Analysis Transducer Interface PCB

The Analysis Transducer Interface PCB is responsible for controlling most functions throughout the analysis module. This includes actions performed by the gas inlet, dosing, and sample manifolds. The analysis transducer interface board is partially enclosed by a metal barrier to protect its components from radio frequency interference.

Card Cage

Analysis Transducer Interface PBC

Transducer Interface Photo Transducer Interface CAD

1 - Backplane PB

2 - DSP-ISA CPU PCB

3 - Digital Interface PCB

Card Cage Photo

1

23

Card Cage CAD

Jul 09 2-9

Page 16: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Printed Circuit Boards (PCBs) TriStar II 3020Service

The card cage is the main assembly of electrical systems and the connection between those systems and other components inside and outside of the instrument. The card cage is comprised of three boards: Backplane, CPU, and Digital Interface PCBs.

DSP-ISA CPU

The DSP-ISA CPU PCB provides control of the instrument. Application software is loaded into this PCB. This board is in the first slot from right to left.

Digital Interface

The digital interface board generates control signals for valve drivers and the elevator interface. This board also stores all PCB IDs, and instrument calibration information. Finally, it has the necessary RS-232 channels for serial communications and connections for ethernet. The digital interface board is in the second slot from right to left.

Backplane

The backplane carries the necessary signals and DC power to and from the boards in the card cage.

2-10 Jul 09

Page 17: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Power Module

Power Module

The TriStar’s power supply assembly is the same model as used in the SediGraph 5120. The power supply is mounted at the rear of the instrument providing access to the power switch.

Power supply (from interior of instrument)

Power switch (from rear of instrument)Power supply (from rear of instrument)

Jul 09 2-11

Page 18: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Power Module TriStar II 3020Service

2-12 Jul 09

Page 19: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Analysis Sequence

3. THE ANALYSIS SEQUENCE

The analysis sequence describes what is taking place throughout the instrument during an analysis.The instrument schematic, shown below, will be used to show the state of the system components during the various phases of an analysis.

The state of a valve is indicated by the icon color.

10 mmHg transducer (Displays only if the Krypton option is installed.)

Sample

Sample tubes

Elevator

Gasinletports

1000 mmHg transducer Temperature

Po tube

transducers

Green = open

White = closed (when in Manual Mode, yellow = closed)

Jul 09 3-1

Page 20: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Analysis Sequence TriStar II 3020 Service

The valves in the schematic are described below.

Analysis Technique

The TriStar II 3020 Analyzer uses physical adsorption and capillary condensation principles to obtain information about the surface area and porosity of a solid material. The analytical technique is simple: a sample contained in an evacuated sample tube is cooled (typically) to cryogenic temperature, then is exposed to analysis gas at a series of precisely controlled pressures. With each incremental pressure increase, the number of gas molecules adsorbed on the surface increases. The equilibrated pressure (P) is compared to the saturation pressure (Po) and their relative pressure ratio (P/Po) is recorded along with the quantity of gas adsorbed by the sample at each equilibrated pressure.

As adsorption proceeds, the thickness of the adsorbed film increases. Any micropores in the surface are filled first, then the free surface becomes completely covered, and finally the larger pores are filled by capillary condensation. The process may continue to the point of bulk condensation of the analysis gas. Then, the desorption process may begin in which pressure systematically is reduced resulting in liberation of the adsorbed molecules. As with the adsorption process, the changing quantity of gas on the solid surface at each decreasing equilibrium pressure is quantified. These two sets of data describe the adsorption and desorption isotherms. Analysis of the shape of the isotherms yields information about the surface and internal pore characteristics of the material.

Valve(s) Description1 through 3 Sample ports4 Po port5 Isolation valve6 Isolation valve7 Helium port8 and 9 Analysis gas ports10 Vacuum11 through 13 Unrestricted vacuum for sample ports14 Analysis gas port15 10-mmHg isolation valve (displays

only if the Krypton option is installed).Unmarked Servo

3-2 Jul 09

Page 21: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Analysis Sequence

Analysis Sequence

1. Create and save up to three sample files for use during an analysis.

2. Place a plug in any port that is not being used for analysis.

3. Prepare samples to be analyzed. This may include weighing the sample and degassing the sample.

4. Load sample tubes in their respective ports and fill the Dewar.

5. Select Unit n > Sample Analysis > Browse to load the newly created sample file for each port.

6. Select Start to begin the analysis.

Jul 09 3-3

Page 22: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Analysis Sequence TriStar II 3020 Service

7. The system begins to evacuate the manifold at a rate and time specified in the sample file.

8. All valves close, then the N2 valve opens and the Po tube is filled to the target pressure.

9. The manifold is evacuated and helium is purged from the gas line.

3-4 Jul 09

Page 23: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Analysis Sequence

10. A series of manifold dosings and port equilibrations begin to take place on each port to prepare for analysis under cryogenic conditions. Helium warm free space is measured. Then the elevator is raised to measure cold freespace.

Jul 09 3-5

Page 24: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Analysis Sequence TriStar II 3020 Service

11. Helium is evacuated from the samples and the Dewar is lowered if specified in the sample file.

12. Samples are evacuated for the amount of time specified in the sample file.

3-6 Jul 09

Page 25: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Analysis Sequence

13. The Po tube is dosed and allowed to equilibrate.

14. If the pressure does not change measurably, the Psat measurement is taken. If the pressure changes measurably, another dose may be required.

1000.00

Jul 09 3-7

Page 26: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Analysis Sequence TriStar II 3020 Service

15. After evacuation, the analysis phase begins. The instrument starts dosing to the set point specified in the sample file and continues until the specified number of data points have been collected. Data points are displayed in a graph as they are collected.

16. You can view the instrument schematic by selecting Instrument Schematic from the View drop-down list.

3-8 Jul 09

Page 27: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Preinstallation

4. PREINSTALLATION

This chapter contains the Preinstallation Instructions and Checklist for the TriStar II 3020. This document helps to ensure that the customer has prepared the laboratory for installation of the instrument and that the supplies, equipment, and personnel are available for installation.

Double-click on the following link to access the preinstallation instructions and checklist.

• Preinstallation Instructions and Checklist, P/N: 302-42870-01.

Jul 09 4-1

Page 28: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Preinstallation TriStar II 3020 Service

4-2 Jul 09

Page 29: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Installation

5. INSTALLATION

Typically, the TriStar analyzers are customer-installed, except for the Confirm version which usually is installed by a qualified Micromeritics service representative.

Installation instructions for the TriStar analyzers are located as follows:

• TriStar II 3020 Standard Windows version: This is a user-installed product; therefore it contains an installation video on the setup CD.

• TriStar II 3020 Confirm Windows version: TriStar II 3020 Confirm Installation Instructions and Checklist, part number: 302-42870-04

Jul 09 5-1

Page 30: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Installation TriStar II 3020 Service

5-2 Jul 09

Page 31: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Operator Training

6. OPERATOR TRAINING

The Operator Training Checklist is used in conjunction with the operator’s manual and soft-ware to train the person(s) who will be operating the analyzer. Click below to access the Operator Training Checklist.

• Operator Training Checklist, P/N: 302-42870-03.

Jul 09 6-1

Page 32: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Operator Training TriStar II 3020 Service

6-2 Jul 09

Page 33: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Maintenance Documents

7. MAINTENANCE DOCUMENTS

The documents listed below are included in Service PM kits, which may be purchased by the user for proper maintenance of his analyzer. Use of these PM plans also help to prolong the life of the analyzer.

• Maintenance Schedule, P/N: 302-42890-00

• Scheduled Maintenance Checklist, P/N: 302-42891-00

• Scheduled Maintenance Procedures, P/N: 302-42892-00

• As-Found Datasheet, P/N: 302-42893-00

• Calibration Certificate, P/N: 302-42894-00

Jul 09 7-1

Page 34: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Maintenance Documents TriStar II 3020 Service

7-2 Jul 09

Page 35: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Calibration Procedure

8. CALIBRATION PROCEDURE

This chapter contains the calibration and test procedure used in Micromeritics’ Final Assem-bly Department. It is provided as a reference so that you will be familiar with the calibration procedures and instrument tests that are performed on the instrument during manufacture.

The methods used at the factory may differ from the methods documented in Chapter 7, Main-tenance Documents. Use the established service calibration and test procedure unless otherwise directed by a Micromeritics Service Managers or Service Specialist.

• Product Calibration Procedure, P/N: 302-34000-76.

Jul 09 8-1

Page 36: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Calibration Procedure TriStar II 3020 Service

8-2 Jul 09

Page 37: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Mechanical

9. MECHANICAL AND ELECTRICAL DRAWINGS

This section contains mechanical and electrical drawings for the TriStar II 3020. These draw-ings show assemblies in great detail. Some parts listed in these prints may be available only as an assembly or in a kit. Do not provide copies of these drawings to a customer.

Mechanical

Refer to the following table for a list of the mechanical prints. Double-click on the part number to view the print.

Electrical

Refer to the following table for a list of the electrical drawings. Double-click on the part num-ber to view the print.

Print Number Description

302-00000-20 Assembly, Core Unit

302-34000-00 Assembly, Semi-Final

302-34001-00 Assembly, Main Chassis

302-34002-00 Assembly, Card Cage

302-34003-00 Assembly, Power

302-34004-00 Assembly, Analysis Module, Nitrogen

302-34004-01 Assembly, Analysis Module, Krypton

Print Number Description

302-17701-011 Backplane PCB

302-17701-511 Backplane Schematic

302-17702-011 Digital Interface PCB

302-17702-511 Digital Interface Schematic

302-17703-011 Analysis Module PCB

302-17703-511 Analysis Module Schematic

302-17704-011 Vacuum Pump Reset PCB

302-17704-511 Vacuum Pump Reset Schematic

302-63803-00 Cable, Inlet Manifold Valves

Jul 09 9-1

Page 38: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Electrical TriStar II 3020 Service

9-2 Jul 09

Page 39: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Service Parts

10. SERVICE PARTS

This section contains a list of approved repair parts for this instrument. The items with an asterisk (*) are included in the Service Parts Kit, P/N: 302-33680-00.

Part Number Description

003-17642-011* CPU PCB Assembly

003-22617-00* Solenoid Valve 24 VDC, Clippard ET-2M-24

003-22626-00* Valve, Solenoid 24VDC 10-32 MT, .060 dia. orifice (25 psig)

003-40051-00* Power Supply, Quad

003-51188-00 Circuit Breaker

003-55617-01 Transducer, 10-torr

003-55635-00* Transducer, 0-30 psia

003-60046-01 Cable, 18 GA 3 Cond Power Belden

003-60805-00 Cable, RTD, 1000 Ohm

003-60820-00 Cable, LED Panel Mnt, Grn, 42 in.

003-60822-00 Cable IEC, Pwr

003-63801-00 Cable, Cat5, 8 Cond, RJ45, RJ45, N, 120

003-63811-07 Cable, Ribbon, 9 Cond, DF, DF, N, 055 in.

003-63811-36 Cable, Ribbon, 9 Cond, DF, DF, N, 036 in.

003-63813-26 Cable, Ribbon 26 Cond, Twst Pr, N, 201

003-63833-04 Cable, Cable, 9 Cond, UML9 UML9, NS, 571

004-24001-00 Handle, 1.50 x .75 x 5/32 dia.

004-25006-00* O-ring, -006 70 Duro Buna-N

004-25022-00* O-ring, -0012 70 Duro Buna-N

004-25044-00* O-ring, -0013 70 Duro Buna-N

004-25116-00 Ferrule, Front, 1/2 in., Teflon

004-25117-00 Ferrule, Front, 1/2 in., Nylon

004-25131-00* Ferrule, Front, 1/8 in., Teflon

004-25132-00* Ferrule, Front, 1/8 in., Nylon

004-25138-00* Fitting, Comp, 1/8 Tube

004-25336-00 Dowel Pin, 1/16 dia x 1/2 Lg SS

Jul 09 10-1

Page 40: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Service Parts TriStar II 3020 Service

Part Number Description

004-25466-00* O-ring, -010 70 Dura Buna-N

004-25468-01* O-ring, -007 70 Dura Buna-N

004-25471-00* O-ring, -018 70 Dura Buna-N

004-25583-01* Nut, Swaging 1/8 Brass

004-25583-10 Plug, 1/8 Gas Port

004-25626-07 Flex Tube, 3/4 OD x 20 Lg, NW-16, NW-40

004-25630-00 Ring, Centering, NW 16

004-25860-00 Fitting, Elbow, NW-16

004-25913-00* Ferrule Set, 1/8 PEEK

004-27046-00* Frit, 40 µm SS 1/4 D x 1/16T

004-28179-00 Dowel Pin 1/8 x 3/8

004-28400-00* Bumper, Screw 6-32

004-28404-00* Foot, .63 OD x .28 High x .13

004-28428-00* Foot, Rubber Cabinet

004-28512-00 Clamp, Vacuum, Flange

004-28850-02* Dowel Pin 3/8 x 1-1/2 in. SS

004-32150-04 Tape, UHMWPE, Acrylic Adhesive, 4 ft.

004-62611-00* Valve, Check, 10 psi Cracking

202-25846-00 Nut, Transducer

202-60807-00 Cable, P.S. DC Outlet

300-17705-011* Assy, Elevator Ctrl

300-25813-00 Clamp, Tube

300-25816-00 Clamp, Tube, 1/4

300-25842-00 Nut, Sample Tube, 1/2 in.

300-25845-10 Tray, Elevator

300-25847-10 Bar, Travel Limit Mounting

300-25848-10 Disk, Travel Limit Trip

300-25861-20 Dewar

300-34005-00 Assembly, Elevator

10-2 Jul 09

Page 41: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Service Parts

* Included in Service Parts Kit, P/N: 302-33680-00

Part Number Description

300-60800-00 Cable, Servo Valve

300-63805-00 Cable, BP Power to Elevator

302-17701-011 Backplane

302-17702-011 Digital Interface

302-17703-011 Analysis Module

302-33680-00 3020 Repair Parts Kit

302-46000-04 TriStar 3020 Service Agreement 04

302-46000-05 TriStar 3020 Service Agreement 05

302-46000-1A TriStar 3020 Service Agreement 1A

302-46000-1B TriStar 3020 Service Agreement 1B

302-46000-2A TriStar 3020 Service Agreement 2A

302-46000-2B TriStar 3020 Service Agreement 2B

302-46000-3A TriStar 3020 Service Agreement 3A

302-46000-3B TriStar 3020 Service Agreement 3B

302-63803-00 Cable, Inlet Manifold Valves

302-63803-01 Cable, TriStar Manifold Valves

302-63803-02 Cable, AC Breaker to Quad PS

512-34003-01 Assy, Power Supply, Quad

Jul 09 10-3

Page 42: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Service Parts TriStar II 3020 Service

10-4 Jul 09

Page 43: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Software

11. SOFTWARE

A demo version of the software for is available for either the standard version or the Confirm version of the TriStar II 3020 system. Double-click the demo software version you wish to view.

• TriStar II 3020 standard software

• TriStar II 3020 Confirm software

Jul 09 11-1

Page 44: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the
Page 45: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Operator’s Manual

12. OPERATOR’S MANUAL

A printed operator’s manual is shipped with every Gemini analyzer. Analyzers operated in a Windows environment also have an electronic copy accessible from the Help menu.

Listed below are the operator manuals available for the Gemini II 2390 series analyzers. Click the document you wish to view:

• TriStar II 3020 standard version, P/N 302-42801-01

• TriStar II 3020 Confirm version, P/N 302-42811-01

Jul 09 12-1

Page 46: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the
Page 47: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

TriStar II 3020 Service Service Training

13. SERVICE TRAINING

Service Training is provided to service personnel so that they can become familiar with the TriStar system. Click Service Training below to view a syllabus of the training class.

• Service Training

Jul 09 13-1

Page 48: TriStar II 3020 Analyzer · 2016-04-01 · TriStar II 3020 Service Overview Jul 09 2-1 2. FUNCTIONAL DESCRIPTION This chapter describes the systems an d components that comprise the

Service Training TriStar II 3020 Service

13-2 Jul 09