Trace Analysis of Volatile Organic Acids with the Agilent J&W DB-624UI GC Column Authors John Oostdijk, Ken Lynam, and Allen Vickers Agilent Technologies, Inc. Application Note Environmental Abstract Agilent’s new deactivation of the 6% cyanopropyl dimethylpolysiloxane (624) phase significantly improves acid performance and maintains very good performance for bases and alcohols. Similar phase selectivity makes it easy to replace existing columns. Introduction Volatile organic acids are organic compounds with acidic properties. The most common organic acids are the carboxylic acids, such as formic acid and acetic acid. Analysis of these acids has been problematic on traditional 624 phases due to reactivity within the cyanopropyl dimethylpolysiloxane phase. This resulted in loss of acidic compounds, no reproducible response, and retention times and bad peak shape that produced high detection limits. Luong et al. [1] compared different GC columns using a solventless stringent test mix and discovered very poor acid performance. Response is influenced by conditioning time/temperature and is not stable after heating the column. After heating and conditioning the column at low temperature, acid recovery improves, but this effect is immediately lost once heat is applied again. Agilent uses a new deactivation procedure for the 6% cyanopropyl dimethylpolysiloxane phase (624 phase). Using this deactivation procedure, acid performance is improved significantly, and the very good performance for bases and alcohols is maintained. The similar phase selectivity makes it easy for analysts to replace their current columns.

Transcript of Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

Page 1: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

Trace Analysis of Volatile OrganicAcids with the Agilent J&W DB-624UI GC Column

Authors

John Oostdijk, Ken Lynam, and

Allen Vickers

Agilent Technologies, Inc.

Application Note

Environmental

Abstract

Agilent’s new deactivation of the 6% cyanopropyl dimethylpolysiloxane (624) phase

significantly improves acid performance and maintains very good performance for

bases and alcohols. Similar phase selectivity makes it easy to replace existing

columns.

Introduction

Volatile organic acids are organic compounds with acidic properties. The mostcommon organic acids are the carboxylic acids, such as formic acid and acetic acid.Analysis of these acids has been problematic on traditional 624 phases due toreactivity within the cyanopropyl dimethylpolysiloxane phase. This resulted in lossof acidic compounds, no reproducible response, and retention times and bad peakshape that produced high detection limits. Luong et al. [1] compared different GCcolumns using a solventless stringent test mix and discovered very poor acidperformance.

Response is influenced by conditioning time/temperature and is not stable afterheating the column. After heating and conditioning the column at low temperature,acid recovery improves, but this effect is immediately lost once heat is appliedagain.

Agilent uses a new deactivation procedure for the 6% cyanopropyldimethylpolysiloxane phase (624 phase). Using this deactivation procedure, acidperformance is improved significantly, and the very good performance for bases andalcohols is maintained. The similar phase selectivity makes it easy for analysts toreplace their current columns.

Page 2: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

2

Materials and Methods

There are 2 different Agilent 624 Ultra Inert stationary phases.In this application note, the Agilent J&W DB-624UI phase isdescribed, but the results are also valid for Agilent J&W DB-Select 624UI for <467>. Conditions and supplies are shownwith the individual analyses.

Results and Discusion

Ultra Inert testingAgilent Ultra Inert columns are developed to deliver the bestinertness for a wide range of active compounds.

3

4

5

6

7

8

9

10

11

12

13

2 7 12 17 22 27

FID

res

pons

e (p

A)

Retention time (min)

1

2 3

4

5

6

7

8

9

10

11

12

No. Compound ng on-column1 1-propanol 12 acetic acid 23 pyridine 14 n-octane 25 1-pentanol 46 1,2-propanediol 17 n-butyric acid 58 m-xylene 19 4-picoline 210 bromoform 2511 dimethyl methylphosphonate 512 n-decane 4

Figure 1. Ultra Inert test mix for Agilent J&W 624UI columns tested after 1 hour at 260 °C and at ng levels, showing good acid performance for acetic acid at 2 ng and n-butyric acid at 5 ng.

To achieve this, every Ultra Inert column is tested using anUltra Inert test mix. For the Agilent J&W DB-624UI columnline, a new mix was developed that tests the inertness of thecolumn at a low temperature with a good separation of allcomponents (Figure 1).

ConditionsColumn: Agilent J&W DB-624UI, 30 m × 0.25 mm, 1.4 µm

(p/n 122-1334UI)Carrier: Hydrogen, 42 cm/sOven: 70 °C isothermalInlet: 1 µL, split 1:116

Page 3: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

3

Calibration curvesWhen a column is inert, linear calibration curves can beobtained. For example, Figure 2 shows 2 critical compoundson a 30 m × 0.32 mm × 1.8 µm column: n-butyric acid and 4-picoline. Curves show good linear response for acidic andbasic compounds, even at low ng levels.

Figure 2. Calibration curves for (A) n-butyric acid and (B) 4-picoline on an Agilent J&W DB-624UI GC column.

R2 = 0.9989

0

50

100

150

200

250

300

0 5 10 15 20 25

Pea

k ar

ea (p

A)

On-column amount (ng)

R2 = 0.9996

0

20

40

60

80

100

120

140

160

180

0 5 10 15 20 25 30

Pea

k ar

ea (p

A)

On-column amount (ng)

A B

Page 4: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

4

CAS No. Carbons Compound Structure

503-74-2 C5 Isopentanoic acid

109-52-4 C5 n-Pentanoic acid

123-76-2 C5 Levulinic acid

111-14-8 C7 n-Heptanoic acid

99-66-1 C8 2-Propyl pentanoicacid

149-57-5 C8 2-Ethyl hexanoicacid

502-47-6 C10 Citronellic acid

O

OH

OH

O

HO

O

CH3

O

H3COH

O

HO

O

OH

O

OH

O

Acid standardsTable 1 shows the range of organic acids used. Dilutions indichloromethane were made to obtain the low on-columnamount after split injection.

Table 1. Organic acids used in this work.

CAS No. Carbons Compound Structure

64-18-6 C1 Formic acid

64-19-7 C2 Acetic acid

79-09-4 C3 Propionic acid

79-10-7 C3 Acrylic acid

79-41-4 C4 Methacrylic acid

79-31-2 C4 Isobutyric acid

107-92-6 C4 n-Butyric acid

OHH

O

C

O

O HH C C

H

H

H C C C

H H

HH O

O H

OH

O

OH

O

HO

O

C C C C

H H

HH O

O HH

H

H

Page 5: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

5

A narrow bore DB-624UI GC column was installed in anAgilent 7890A GC with FID. After conditioning, a C1-C5 organicacid test mix (0.5-1 ng) was injected (Figure 3). Because theresponse for formic acid on an FID is very low, a highconcentration (250 ng) of this compound was added to obtaina peak. However, this has a negative influence on its peakshape.

Conditions

Column: Agilent J&W DB-624UI, 30 m × 0.25 mm, 1.4 µm (p/n 122-1334UI)

Carrier: Hydrogen, 3 mL/min constant flowOven: 70 °C (1 min), then 20 °C/min to 160 °CInlet: 250 °C, 1 µL, split 1:200Inlet liner: 4 mm, glass wool (p/n 5183-4647)FID: 260 °C

Figure 3. Organic acid mix C1-C5 on an Agilent J&W DB-624UI GC column.

2.5

3.5

4.5

5.5

6.5

7.5

8.5

9.5

0 1 2 3 4 5 6

FID

res

pons

e (p

A)

Retention time (min)

1

2

3

4

5 6 78 9

i

No. Compound [ng]1 Formic acid 2502 Acetic acid 1.03 Propionic acid 0.5i impurity 4 Acrylic acid 0.55 Isobutyric acid 0.56 n-Butyric acid 0.57 Methacrylic acid 0.58 Isopentanoic acid 0.59 n-Pentanoic acid 0.5

Page 6: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

6

To demonstrate reproducibility, the injection was repeated 6 times (Figure 4). Retention time and responses were stable.For conditions, see Figure 3.

Figure 4. Organic acid mix C1-C5 on an Agilent J&W DB-624UI GC column after 6 injections.

1.3 1.8 2.3 2.8 3.3 3.8 4.3 4.8

Retention time (min)

Inj no.1

Inj no. 3

Inj no. 6

Page 7: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

7

A test mix with C1-C8 organic acid was prepared and injectedon GC-FID (Figure 5). With this test mix, a low level amount offormic acid was used, which was below the FID detectionlimit. Higher molecular weight acids at 0.6 ng and 6 ngshowed good peak shape.

Figure 5. Organic acid mix C1-C8 on an Agilent J&W DB-624UI GC column at (A) 6 ng and (B) 0.6 ng.

2.5

12.5

22.5

32.5

42.5

52.5

62.5

72.5

82.5

92.5

0 2 4 6 8 10 12 14

FID

res

pons

e (p

A)

Retention time (min)

1

2

3

4 5 6

7

89

10

FID Bleed level at 260 °C

2.5

4.5

6.5

8.5

10.5

12.5

1.5 2.5 3.5 4.5 5.5 6.5

FID

res

pons

e (p

A)

Retention time (min)

2

3

4 56

7

8 9

10

A

B

No. Compound1 Dichloromethane2 Acetic acid3 Propionic acid4 Isobutyric acid5 n-Butyric acid6 Isopentanoic acid7 n-Pentanoic acid8 n-Heptanoic acid9 2-Propyl pentanoic acid10 2-Ethyl hexanoic acid

ConditionsColumn: Agilent J&W DB-624UI, 30 m × 0.25 mm, 1.4 µm

(p/n 122-1334UI)Carrier: Hydrogen, 3 mL/min constant flowOven: 70 °C (1 min), then 20 °C/min to 260 °CInlet: 250 °C, 1 µL and 0.1 µL, split 1:200Inlet liner: 4 mm, glass wool (p/n 5183-4647)FID: 260 °C

Page 8: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

8

Because formic acid analysis using FID is not a method ofchoice, MSD was also used for detection. A new mix wasprepared. Formic acid was injected at 17 ng on-column, aswell as some higher organic acids (Figure 6). Using MSD,good analysis of organic acids was possible, even for formicand citronellic acids.

ConditionsColumn: Agilent J&W DB-624UI, 30 m × 0.25 mm, 1.4 µm

(p/n 122-1334UI)Carrier: Helium, 3 mL/min constant flowOven: 70 °C (1 min), then 20 °C/min to 260 °CInlet: 250 °C, 1 µL, split 1:20Inlet liner: 4 mm, glass wool (p/n 5183-4647)Agilent 5973 MSD: Transfer 250 °C, EI source 230 °C, Quad 150 °C, EI

full scan m/z 10-550

Figure 6. Organic acid mix C1-C10 (6 to 17 ng) on an Agilent J&W DB-624UI GC column.

0

20000

40000

60000

80000

MS

D r

espo

nse

(cou

nts)

100000

120000

140000

0 2 4 6 8 10Retention time (min)

1

2

3 45

67

8 910

11-13

14

11

12 13

1

2

Sum m/z 46 + 60

No. Compound1 Formic acid2 Acetic acid3 Propionic acid4 Acrylic acid5 Isobutyric acid6 n-Butyric acid7 Methacrylic acid8 Isopentanoic acid9 n-Pentanoic acid10 n-Heptanoic acid11 Levulinic acid12 2-Propyl pentanoic acid13 2-Ethyl hexanoic acid14 Citronellic acid

Page 9: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

9

Comparing 0.32 mm id columnsAs well as 0.25 mm id columns, those of 0.32 mm id are alsowidely used. Using GC-FID, the C1-C10 organic acid mixturewas tested on a 0.32 mm id column and compared to a non-Agilent 624 phase with the same dimensions. Thecolumns were tested at the same time and in the same GC (2 channels) after conditioning for 1 hour at 260 °C. Asexpected, formic acid was below the detection limit, andlevulinic acid was not separated from 2-propyl pentanoic acid(fewer plates compared to a 0.25 mm id column). However,other acids showed good elution from the DB-624UI column(Figure 7). No acids were detected on the traditional phase inthe non-Agilent column.

Columns: Agilent J&W DB-624UI, 30 m × 0.32 mm, 1.8 µm (p/n 123-1334UI)Non-Agilent 624. 30 m x 0.32 mm, 1.8 µm

Carrier: Hydrogen, 4 mL/min constant flowOven: 70 °C (1 min), then 20 °C/min to 260 °CInlet: 250 °C, 1 µL, split 1:200Inlet liner: 4 mm, glass wool (p/n 5183-4647)FID: 260 °C

0

20

40

60

80

100

0 1 2 3 4 5 6 7 8

FID

res

pons

e (p

A)

Retention time (min)

2

3

4

56 7

89 10

A

R

11+12

13

14

No. Compound1 Formic acid (< DL)2 Acetic acid3 Propionic acid4 Acrylic acid5 Isobutyric acid6 n-Butyric acid7 Methacrylic acid8 Isopentanoic acid9 n-Pentanoic acid10 n-Heptanoic acid11 Levulinic acid12 2-Propyl pentanoic acid13 2-Ethyl hexanoic acid14 Citronellic acid

Figure 7. Organic acid mix C1-C10 (6 to 17 ng) on an Agilent J&W DB-624UI column (blue) and a traditional non-Agilent 624 column (red) after conditioning at 260 °C for 1 hour.

624UI

Page 10: Trace Analysis of Volatile Organic Acids with the Agilent J&W DB ...

www.agilent.com/chem

Agilent shall not be liable for errors contained herein or for incidental or consequentialdamages in connection with the furnishing, performance, or use of this material.

Information, descriptions, and specifications in this publication are subject to changewithout notice.

© Agilent Technologies, Inc., 2012Printed in the USAJuly 31, 20125991-0845EN

Conclusions

Agilent J&W DB-624UI and Agilent J&W DB-Select 624UI for<467> GC columns are well suited for the analysis of themost stringent components, including C1-C10 organic acids.These DB-624UI columns represent a major breakthrough ininertness compared to traditional 624 phases. Quality iscontrolled using an Ultra Inert test mix, which guaranteescolumn-to-column inertness for the most stringent probes.

False negatives or incorrect values for acidic compounds wereavoided. Selectivity remained similar compared to currentAgilent columns, requiring minimal revalidation for chemistswho want to replace their traditional 624 column with thelatest Agilent J&W DB-624UI column. In addition, custom-made dimensions are possible.

Reference

1. J. Luong, R. Gras, W. Jennings. J. Sep. Sci. 30, 2480(2007).

For More Information

These data represent typical results. For more information onour products and services, visit our Web site atwww.agilent.com/chem.