TPS modeling and Beam commissioning at UIHC Radiation Oncology

26
ROTATION 5: TPS MODELING AND BEAM COMMISSIONING at UIHC Radia9on Oncology Mentors: John Bayouth, Ph.D. (04/23/09 – 06/11/09) By Vibha Chaswal

description

 

Transcript of TPS modeling and Beam commissioning at UIHC Radiation Oncology

Page 1: TPS modeling and Beam commissioning at UIHC Radiation Oncology

ROTATION  5:  TPS  MODELING  AND  BEAM  COMMISSIONING  at  UIHC  Radia9on  

Oncology  Mentors:  John  Bayouth,  Ph.D.    

(04/23/09  –  06/11/09)  

By    Vibha  Chaswal  

Page 2: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Goals  

•  Model  1  photon  beam    -­‐  Know  parameters  and  their  effects  

•  Model  1  electron  beam  

-­‐  Know  parameters  and  their  effect  

•  Model  Wedges  

•  Validate  the  beams  modeled  

•  Understand  the  nuances  of  TPS  modeling    

Page 3: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Photon  beams:  data  requirements  

•  Pdds  •  Profiles  •  Calibra9on  output  factor  •  Rela9ve  output  factors  Data  For  Valida9on!  -­‐  A  subset  of  above  -­‐  Sc  and  Sp  -­‐  Dosimetric  tests,  match  computed  and  measured  dose  distribu9ons  

Page 4: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Data  collected  for  assumed  clinical  applica9on  

Page 5: TPS modeling and Beam commissioning at UIHC Radiation Oncology

TPS  modeling:  The  Machine  

Page 6: TPS modeling and Beam commissioning at UIHC Radiation Oncology

TPS  Modeling:  The  machine  

Page 7: TPS modeling and Beam commissioning at UIHC Radiation Oncology

TPS  Modeling:  The  Radia9on  Beam  

•  Fine  tuning  Profiles    -­‐  Effec9ve  source  size  for  Penumbra  matching  

-­‐  Gaussian  width  and  height  of  FF  scacer  source  for  tails    

-­‐  Transmission  factors  of  Jaw  and  MLC  for  tails  region    

Page 8: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Energy  Spectrum  

Page 9: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Energy  Spectrum:  Build-­‐up  region  

Page 10: TPS modeling and Beam commissioning at UIHC Radiation Oncology

In-­‐field  Region:  FF  acenua9on  using  cone  model  

With  Off-­‐axis  Soeening  factor  

Page 11: TPS modeling and Beam commissioning at UIHC Radiation Oncology

In-­‐field  Region:  FF  acenua9on  using  arbitrary  profile  

With  Off-­‐axis  Soeening  factor  

Page 12: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Off-­‐axis  soeening  

Page 13: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Out-­‐of  field  regions:    Penumbra  and  tails  

Page 14: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Out-­‐of  field  modeling  parameters  

Page 15: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Beam  Valida9on:  PDDs  

Result:  All  Pdds  pass  within  2%  or  2mm  

Page 16: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Beam  Valida9on:  Profiles  

Result:  Within  3%  or  3mm  except  outliers  in  tail  regions  

Page 17: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Beam  Valida9on  tests:  MU  verifica9on  

Page 18: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Beam  valida9on:  R  and  V  import  

Page 19: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Rela9ve  Decision  Maps:  error  Limit  =  3  %,  dta  Limit  =  3  mm  

Page 20: TPS modeling and Beam commissioning at UIHC Radiation Oncology

TPS  Modeling:  The  machine,  Collimator  rota9on  incorrect  

Page 21: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Electron  beams:  Data  required    

Page 22: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Electron  Beam  Parameters  

Page 23: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Two  adjustable  parameters  

- Off-axis ratios, and

- The scatter correction factor (FMCS), dose spread out in media relative to air.

Page 24: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Electron  Beam  modeling  

Page 25: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Electron  PDD  

Page 26: TPS modeling and Beam commissioning at UIHC Radiation Oncology

Thank  you!!!!!