Model TP700/760
Software version: V5.0x OS: Windows XP / Vista / 7
Instrument Manual
TAKARA BIO INC.
Attention: Thermal Cycler DiceTM
Real Time System Software Operation Manual
should be stored for reference in a secure and accessible location.
TM
NOTICE
P10 PCR Notice
Purchase of this instrument conveys a limited non-transferable immunity from suit for the
purchaser's own internal research and development and applied fields other than human in vitro
diagnostics under one or more of U.S. Patents No. 5,475,610 or corresponding claims in their
non-U.S. counterparts, owned by Applera Corporation.
No right is conveyed expressly, by implication or by estoppel under any other patent claim.
Further information on purchasing licenses may be obtained by contacting the Director of
Licensing, Applied Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404, USA.
Applied Biosystems does not guarantee the performance of this instrument.
L45 Heated Cover
This product is covered by the claims of U.S. Patent 5,552,580, and their corresponding foreign
counterpart patent claims.
L47 Real-Time PCR Quantification Method
The purchase of this product includes a limited, non-transferable license for all fields other than
human or veterinary in vitro diagnostics under specific claims of U.S. Patent Nos. 6,174,670,
6,569,627, 6,303,305, and 6,503,720, owned by the University of Utah Research Foundation
and licensed to Idaho Technology, Inc. and Roche Diagnostics GmbH.
- Thermal Cycler Dice™
is a trademark of TAKARA BIO INC.
- Microsoft® Windows
® XP , Windows Vista
® , Windows
® 7 , Excel
® , PowerPoint
®
is a trademark of Microsoft Corporation.
- Adobe® Reader
® is a trademark of Adobe Systems Incorporated.
- All marks are the property of their respective owners. Certain trademarks may not be
registered in all jurisdictions.
Copyright© 2012 by TAKARA BIO Inc., Otsu, Shiga, Japan
TP700/760 Instrument Manual
Rev. A
= Contents =
page
Precautions for Using TaKaRa Scientific Instruments --- 1
Chapter 1: Outline --- 6
Chapter 2: Installation Requirements and Maintenance --- 8
1. Installation Conditions --- 8
2. Opening the Packaging --- 9
3. Installation -- 11
4. Moving the System -- 12
Chapter 3: Part Names and Functions -- 13
Chapter 4: Explanation of Screens -- 17
1. System Startup and Shut Down -- 17
2. Initial Screen Menu -- 19
Chapter 5: Maintenance -- 22
1. Self-diagnostic Function -- 22
2. Optical Calibration -- 25
3. Daily Inspections -- 36
Chapter 6: Troubleshooting -- 37
Chapter 7: Specifications -- 42
Chapter 8: Consumption and Options List -- 45
Appendix : Thermal Cycler Dice Real Time System TP700/760 -- 46
1. Computer Driver Setup -- 47
2. Software Install / Uninstall -- 52
3. Computer Basic Setup -- 56
1
Precautions for Using TaKaRa Scientific Instruments
●Users are recommended to carefully read the content of this manual before operating the TaKaRa PCR
Thermal Cycler Dice system. ●After familiarizing themselves with the information in this manual, users are recommended to store it in
secure and convenient location from which it is always available for reference
Introduction
All Takara scientific instruments are sold strictly for use in research applications and should not be
used by those who are not experts in this field.
Due to ongoing improvements, product specifications and the external appearance are subject to
change without prior notification.
Warranty and Responsibilities
Refer to the Warranty card included with this product for detailed description of the Warranty
conditions and period.
This product is designed to ultimate safety standards. Nevertheless, should any safety-related issues
arise, users are requested to immediately contact the nearest sales agent.
Precautions (safety and accident prevention) for System Use
(1)Danger Warnings
The various messages and labels provided on the product itself and in the various manuals are
designed to ensure the system is operated safely to prevent personal injury or damage to property.
Users are strongly recommended to read and observe the warnings included with the symbols
shown below.
Warnings indicate incorrect system operation and handling procedures that can lead to serious accidents.
Cautions indicate incorrect system operation and
handling procedures that can lead to accidents. ! CAUTION
! WARNING
2
(2)General Precautions Regarding Takara Scientific Instruments
It is important to observe the various precautions described below to ensure optimum system
performance and prevent accidents before they happen. Ensure safe and correct system operation
by carefully reading the precautions.
【Precautions for Installation Location】
Observe the following precautions in order to prevent fires and electrical shocks.
● Do not install the system on an unstable or sloped foundation or any
other location subject unstable conditions.
● If the system tips over or falls, immediately turn off power to the
system and all connected devices and disconnect the power supply
cables.
● Install the system away from direct sunlight or rainwater.
● Avoid blocking the paths to any air vents included on the product.
Blocked air vents can lead to internal overheating that may result in
physical burns or fire.
● Avoid locations subject to corrosive or flammable gases.
● Avoid installing the system in locations subject to dust or strong
vibrations.
● Avoid installing in locations subject to direct sunlight and wide
temperature fluctuations as this may result in internal overheating and
may cause physical burns or fire.
● Do not place objects on top of the system.
● Do not install the system in locations other than research institutions or
related facilities.
【Electrical Precautions】
Observe the following precautions when using electrically powered scientific instruments in order to
prevent electric shocks and fires.
● Avoid operating system with power sources that do not conform to
system power supply specifications.
● Avoid sharply twisting, altering, marking or otherwise damaging the
power supply cable. Avoid applying stress to, heating or placing
objects on the power cable.
● Never use a multiple outlet extension plug.
● Always ground the system to prevent electric shocks due to leakage
current.
! WARNING
! CAUTION
! WARNING
3
● Be sure to compare system power supply requirements with the rated
power capacity of the power source to ensure system current
requirements do not exceed power socket capacity. Confirm that the
total power requirements of all connected equipment do not exceed
the rated power supply capacity of the common power source.
● The system body and power supply cable are subject to high voltage
levels(The maximum voltage is a line voltage AC100V~240V). Avoid
inserting metal or flammable objects inside the system or handling it
with wet hands, as this can result in electric shock and/or other
dangers.
● When moving the system, be sure to first disconnect the power cable
as well as any peripheral devices.
● Be sure to disconnect the power cable prior to cleaning or servicing
the system.
【Other precautions】
Observe the following precautions in order to prevent electric shocks and fires.
● Avoid inserting metal objects, flammable materials or other foreign
objects into the various openings on the system case.
● Never remove the system cover or attempt to alter it in anyway.
● Avoid disassembling or making any direct contact with internal parts of
the system.
● Should noxious fumes smoke or other abnormal signs appear,
immediately terminate operation, turn off system power and
disconnect the power cable.
Wait until conditions stabilize before contacting the nearest sales
agent or the company technical support line.
● Should water or other foreign matter enter the system, immediately
terminate operation, turn off system power and disconnect the power
cable.
Wait until conditions stabilize before contacting the nearest sales
agent or the company technical support line.
● Do not sit on the system.
● Do not disassemble the system
! WARNING
! CAUTION
! WARNING
! CAUTION
4
(3)Warning Safety Labels Applied to System Case
Users are advised to avoid accidents by carefully reading the warnings and cautions contained on
warning stickers at key locations on the interior and exterior of the system.
Possible
Danger
Warning Type
Location of Danger
Type d’Avertissement
Localisation du Danger
Warning Label Description of
Danger
Electric
shocks and
fires
Disassembly
Prohibited
System Interior
Démontage Interdit
Intérieur de l’Appareil
Never disassemble
system, as this may
lead to electrical
short-circuits,
leakage current
and/or fire.
Electric
shocks
Warning! High
Voltage
Power cable
Attention! Haute
tension
Cordon d’alimentation
The case and power
cable are subject to
high voltage levels,
so take care to avoid
electric shocks.
Electrical
shocks and
fires
Grounding Mandatory
Power cable
Mise à la Terre
Obligatoire
Cordon d’alimentation
Always ground the
system to prevent
electric shock from
leakage current.
Burns High Temperatures
Heat cover
& Heat block
Haute Température
Couvercle Chauffant
Avoid touching the
heat cover, which
reaches very high
temperatures and
may cause burns.
Burns and
fires
Installation
Precaution
Air vents
Précautions
d’Installation
Prises d’air
Be sure not to block
air vents when
installing the system,
as this can cause
internal overheating.
Personal
injury
Do not Open When in
Operation
System cover
Ne Pas Ouvrir en
cours de
fonctionnement
Couvercle
Be sure to close the
cover while the
system is in
operation. Failure to
do so can lead to
burns and/or
damage to the eyes.
5
This symbol means caution or warning.
This symbol means an action is prohibited.
This symbol means an instruction must be followed.
(4) Electrical Symbols
The symbols shown on this page are adapted to the TaKaRa PCR Thermal Cycler Dice. Read the
explanations and make sure you understand what the symbols mean before you interact with the
instrument in any way.
This symbol indicates the on position of the main power switch.
This symbol indicates the off position of the main power switch.
Precautions against Transferring or Copying Content of This Document
The content of this document cannot be transferred or copied without the express permission of the
copyright holder.
Copyright holder: TAKARA BIO INC.
Printed in Japan
Contact TAKARA BIO INC.
TAKARA BIO INC.
SETA3-4-1, OTSU, SHIGA, 520-2193, JAPAN
http://www.takara-bio.com
Technical Inquiries (Contact):
URL: http://www.takara-bio.co.jp/english/index.htm
TP700/760 Instrument Manual
Rev. A
6
● Chapter 1: Outline
Thank you for purchasing the Thermal Cycler Dice™
Real Time System Lite.
This system consists of the following components:
<TP700> Thermal Cycler Dice™
Real Time System Lite
1. Thermal Cycler Dice™
Real Time System Lite Main Unit (Code: TP710)
2. Thermal Cycler Dice™
Real Time System Control Computer (Code: TP820)
3. Thermal Cycler Dice™
Real Time System Software for TP700 (Code: TP730)
<TP760> Thermal Cycler Dice™
Real Time System Lite MRQ
1. Thermal Cycler Dice™
Real Time System Lite Main Unit (Code: TP710)
2. Thermal Cycler Dice™
Real Time System Control Computer (Code: TP820)
3. Thermal Cycler Dice™
Real Time System Software for TP700 (Code: TP730)
4. Multiplate RQ Software (Code: TP840)
System construction table
System Lite code TP700 TP760
TP710 Dice Real Time System Lite Main Unit ○ ○
TP820 Control Computer ○ ○
TP730 Dice Real Time System Software for TP700 ○ ○
TP840 「Multiplate RQ」 Software - ○
TP700/760 Instrument Manual
Rev. A
7
The main features of the Thermal Cycler Dice™
Real Time System Lite are as follows:
【 TP700/760 】
This model has the 48-well Sample Block Assembly. Use a 48 well plate, 24well
plates, 8 strips tubes and individual tubes for your situation.
The reaction block of copper with a high thermal conductivity is use. Temperature
uniformity has improved better than the reaction block of aluminum.
Original calibration processing eliminates the need for calibration using secondary
fluorescent dye. This reduces running costs and reagent preparation time.
The miniature design saves space by using the smallest width and height possible.
With an internal mirror and built-in tilt and sliding lid mechanism, the device can be
stored on a shelf, but is still easy to operate.
( Main Unit Dimensions: Width 294 x Depth 480 x Height 304 [mm] )
The CCD camera system gives no difference in measuring time between wells.
The system includes automatic compensation for stable detection, maintenance
functions and easy optical calibration functions.
Expandability is provided for by allowing installation of a maximum of four (4) types
of filter unit. Two fluorescent filter units are installed as standard (FAM and ROX).
The application software is user-friendly. Basic settings are easy, and application
settings provide a high degree of freedom. It is possible to set the fluorescent
measurement point and to view data even with other computers. The software
uses a unique data format that prevents alteration of the data by general-purpose
software.
The application software have the Absolute Quantification method, the Relative
Quantification mode and added the Plus/Minus Assay program and the SNP
Genotyping Assay program.
The application software has the List manager of Target and Sample, making text
files for RDML(Real-time PCR Data Markup Language) and changeable graphic
colors patterns.
【 TP760 only 】
TP760 system includes Multiplate RQ software ( Code: TP840).
TP700/760 Instrument Manual
Rev. A
8
● Chapter 2: Installation Requirements and Maintenance 1. Installation Conditions
To ensure optimum performance and long and stable product life, observe the
following precautions when selecting an installation location.
Note that installing the system in a location that does not meet the requirements
listed below can result in malfunctions and/or system breakdown. Replacement
of components and system repairs for any deterioration in performance or
damage caused by failure to adhere to the following conditions will not be
covered by warranty.
1. Install system in a comparatively dust-free location.
2. Avoid locations subject to corrosive or flammable gases.
3. Avoid locations subject to either strong shocks or continuing low levels of
vibration.
4. Install the system away from direct sunlight or rainwater.
5. Install the system as far away as possible from electrical equipment that
generates strong magnetic fields, electrical fields and high frequencies. Avoid
locations near equipment that generate noise caused by electrical motors and
transformers.
6. Choose a flat location capable of withstanding the design weight of the
system.
7. Do not put it on the slippery seat. Avoid to slip down when the earthquake.
8. The installation environment must be within 15 to 30ºC temperature range and
a 20 to 80% relative humidity range.
9. Avoid locations subject to sharp temperature fluctuations, exhaust from
air-conditioning systems and condensation caused by sharp humidity
fluctuations.
10. Avoid blocking air vents located on the right and left sides, front and rear of the
system.
11. Choose an operating location with good ventilation.
12. Do not operate the system in a cold room.
13. Select a room that can provide sufficient power (frequency, voltage and
current) for system operation.
14. Never use a multiple outlet extension plug.
15. Do not attempt to move the system while holding it by the lid handle or the
control panel, as this can damage to key components. Doing so can also
result in personal injury due to part breakage resulting in the system falling
down.
TP700/760 Instrument Manual
Rev. A
9
2. Opening the Packaging
This equipment comes packaged in carton, as shown below. The power cord is bundled at
the bottom of the carton box. The gross weight with packaging is 21.5 kg, and the equipment
itself weighs 18 kg. Take sufficient care when removing the equipment from the packaging.
■ Main Unit Package
【Packed Items】
Main Unit
Power Cord
Instruction Manual
Warranty card
Consumable Accessory Pack
1. 48 well snap plate (NJ700) 4 plates
2. Flat cap for snap plate (NJ720) 24 strips
3. 0.2ml Hi-8-Tube (NJ300) 25 strips
4. 0.2ml Hi-8-Flat Cap (NJ302) 25 strips
5. 0.2ml 8-strip tube, individual Flat caps (NJ600) 24 strips
Power Cord
Consumable
Accessory Pack
Inside the upper
tension plate
Instruction Manual
and Warranty card
TP700/760 Instrument Manual
Rev. A
10
■ Control Computer Package
Notebook PC
AC adapter for PC
USB Cable (supporting USB 2.0)
Thermal Cycler Dice™
Real Time System Software CD for TP700
Thermal Cycler Dice™
Real Time System Driver Software CD
Microsoft® Office Home & Business 2010
Accessory Pack
Note: The contents of the accessory pack may vary.
Check the sama serial Number, TP710 main body and PC label.
same number
TP700/760 Instrument Manual
Rev. A
11
3. Installation
Use the following procedures when installing system.
1. Confirm that the system power switch is set to off.
2. Connect the system side of the power cord to the power connector on the rear
of the system.
3. Plug the AC power plug of the cord into a power outlet.
The system is designed to operate normally with a 100 to 240 V AC, 50/60Hz
power line.
Make sure that the power outlet is designed for a 3-pronged plug with ground
line.
4. Install the Control Computer in a location near the main unit.
5. Connect the USB cable provided to the connector
on the back of the main unit, and connect the other
end to the USB port on the computer.
Note: Connect the cable to one of all computer USB ports of which the drivers
are installed.
6. After turning the main unit power ON, turn the computer power ON and start
the software. “Instrument: Connected” should be displayed at the bottom right
of the initial screen.
Caution: Connect the AC adapter as a power source for the computer. If
battery power is used alone, experiments may be cut off when
battery power runs out.
7. Please refer to the Software Manual for information about real time PCR and
analysis.
8. When experiments are complete, close the software by File -> Exit.
9. Shut down the Control Computer.
10. Turn the power source for the main unit to the OFF position.
In the interests of safety, and avoiding electric shocks, be sure to connect
the system to a properly grounded power outlet. ! CAUTION
USB port
TP700/760 Instrument Manual
Rev. A
12
4. Moving the System
Follow the procedures described below when moving the system to a new location after it has been installed. ● New Location Within the Same Site Follow the precautions listed below when moving the system to a new location
within the same site.
1. Thoroughly clean the system and make sure that there are no tubes or plates remaining in the sample block.
2. Avoid any shock during movement that can affect the operation of this highly sophisticated system.
3. When lifting and carrying the system, be sure to place both hands on each side of the system and lift and carry in a balanced way.
4. Make sure that the new location conforms to the installation requirements described in section “1. Installation Conditions” in Chapter 2.
5. Follow the installation procedures outlined in section “2. Installation” in Chapter 2.
● Moving to a New Site
When moving the system to a new site, clean it carefully and repack it in the packaging in which it was delivered. Note that damage sustained by the system during transport due to improper packaging will not be covered by Warranty even if the Warranty period is still valid. Follow the instructions provided in this chapter when reinstalling the system at the new location.
TP700/760 Instrument Manual
Rev. A
13
● Chapter 3: Part Names and Functions
Exterior (Lid Closed)
POWER [ON] when the power switch turn ON ( | ).
[OFF] when the power switch turn OFF (○).
(always ON when power is being supplied).
Standby [Flashing] when the main unit is getting ready;
[ON] when ready for operation (a run can be started
immediately using the Run command).
RUN [ON] during program (run) operation;
[OFF] after the operation is complete.
Power ON
Preparing for Operation
POWER: ON, Standby: Flashing, RUN: OFF
Ready for Operation
(Lid temperature stable)
POWER: ON, Standby: ON, RUN: OFF
Run Start
(Program start)
POWER: ON, Standby: OFF, RUN: ON
Returning to Standby
condition
POWER: ON, Standby: Flashing, RUN: OFF
LED display
TP700/760 Instrument Manual
Rev. A
14
Front (Lid Open)
Explanation of Parts ――――――――――――――――――――――――――――――
① Lid Handle To open the Lid, lift the Lid Handle, slide the Lid towards the
back and push it into the main unit. To close the Lid, pull the Lid
Handle forward and push it down.
② Lid The Lid prevents evaporation by heating the top of the Reaction
Plate. The Lid also allows through path of excitation light and
reflected light.
[Caution]: The equipment becomes very hot inside (108-110°C).
Do not insert fingers.
③ Sample Block Place the Reaction Plate or Tube on the Reaction Block.
[Caution]: Be careful to avoid injuries from burns. The block
heats up to a maximum of 100°C.
④ Lid Lock While the program is in operation, the Lid becomes locked so
that it cannot be opened.
⑤ Ventilation Intake (Front Side)
Openings in the main unit for providing internal cooling by air
ventilation.
[Caution]: Do not cover the Ventilation Intakes.
① Lid Handle
② Lid
④ Lid Lock
⑤ Ventilation Intake
(Front Side)
⑥ Power Switch
ON (|), OFF (○)
⑦ Ventilation Intake
(Left and Right Sides)
⑧ Lid Lock Catch
③ Sample Block
TP700/760 Instrument Manual
Rev. A
15
⑥ Power Switch Switch for turning the main unit power to ON (|) or OFF (○).
⑦ Ventilation Intake (Left and Right Sides)
Openings in the main unit for providing internal cooling by air
ventilation.
[Caution]: Do not cover the Ventilation Intakes.
⑧ Lid Lock Catch Catch for the Lid Lock cylinder.
TP700/760 Instrument Manual
Rev. A
16
Back
Explanation of Parts ――――――――――――――――――――――――――――――
⑨ Power Cord Connector Connector for Power Cord.
[Caution]: Ensure that you use the Power Cord that
came with the equipment.
⑩ Ventilation Exhaust (Back Side) Openings in the main unit for providing internal cooling.
[Caution]: Do not cover.
⑪ USB Cable Port Connection for the USB cable mini-connector plug.
⑫ Top Cover Cover for the equipment.
[Caution]: Do not open. For service..
⑨ Power Cord Connector
⑩ Ventilation
Exhaust (Back)
⑪ USB Cable Port
⑫ Top Cover
TP700/760 Instrument Manual
Rev. A
17
● Chapter 4: Explanation of Screens
1. System Startup and Shut Down
1) Startup
① Verify that the Thermal Cycler Dice™
Real Time System Lite main unit is connected
to the computer via the USB cable.
② Turn the Power Switch located at the bottom right of the Thermal Cycler Dice™
Real
Time System Lite main unit to the ON (|) position.
When the power is turned on in the main unit, The Lid begins to warm up. It takes
about 5 minutes for the Lid plate until its temperature is greater than about 100°C.
The Standby LED blinks until Lid have warmed up sufficiently. When warm-up is
complete and the equipment is ready for use, the Standby LED changes to ON.
③ After the Thermal Cycler Dice™
Real Time System Lite main unit is powered, turn
on the computer.
④ At the Login Screen, input User Name and Password if necessary.
⑤ When the computer has finished starting up, double-click the Thermal Cycler Dice™
Real Time System Software icon on the desktop.
Alternatively, the user may select [All Programs]→[TaKaRa Thermal Cycler Dice]→
[Thermal Cycler Dice Real Time System ] from the Start menu at the bottom left of
the computer screen.
⑥ The splash screen is displayed, and the Thermal Cycler Dice™
Real Time System
Software launches.
TP700/760 Instrument Manual
Rev. A
18
⑦ The Initial Screen for the Thermal Cycler Dice™
Real Time System Software is
displayed.
Refer to the section on software operation in the Operation Manual for a
description of the software and its method of use.
2) Shut Down
① Select Exit from the File Menu in the Thermal Cycler Dice™
Real Time System
Software.
Caution: If there are experimental files
that have not been saved, a
dialog box will be displayed
asking whether or not the files
should be saved before shutting
down. Save the experiment files
if necessary.
② Select “Turn Off Computer” from the Start menu at the bottom left of the computer
screen.
③ The Shut Down Screen is displayed.
④ Click “Turn Off”. The computer shuts down.
⑤ Turn the Power Switch on the Thermal Cycler Dice™
Real Time System main unit to
the OFF (○) position.
TP700/760 Instrument Manual
Rev. A
19
2. Initial Screen Menu
The connection status can be verified using the Instrument Status and the Camera Status displays in the Status Bar (bottom of screen). “Connected” is displayed when the computer and any units are connected, and “Disconnected” is displayed if they are not connected, or if power to the main unit is OFF.
Displays connection to computer and main unit.
① ② ③ ④
Temperature:
the current Lid
Temperature.
Lid: Close/Open
Lid Status (Open/Closed)
TP700/760 Instrument Manual
Rev. A
20
① File Menu
New: Create a new Experiment File
Open/Close: Open experiment files / Close the opened file
Data Manager: Open the Data Manager
Save: Overwrite as the same file name
Save As: Save the experiment file to another file name
Load Template: Load the only format of selected stored experiment
files
Make Full Report: Make a report of experiment file as Word/Power point
data form.
Print: Print out a report of experiment file
Exit: Close the software (program)
② Instrument menu
Run: Start Run/Add Cycle/Pause/Resume/Force Quit
(forcibly terminate)
Filter Set: Settings for Excitation and Detection Filter /
Information for Fluorescent dye / Settings for Display,
Addition, Deletion and Measurement Order
Calibrate: Optical System Calibration Screen
Display Thermal Service Mode for service only
Diagnosis: Self-diagnostics for equipment heating and cooling
systems
TP700/760 Instrument Manual
Rev. A
21
③ User Menu
Change: Change a user
Setting: Display the User Setting screen
Management: Display the User Management screen
④ Help Menu
Help: Display User’s manual by selected Japanese or
English ( need to install “Adobe Reader”)
About: Display Software version
TP700/760 Instrument Manual
Rev. A
22
● Chapter 5: Maintenance
1. Self-diagnostic Function
If the PCR reaction does not proceed well, it possible to determine whether the
temperature control for the main equipment is operating normally using the self-diagnostic
function for Sample Block temperature efficiency.
In addition, the status of the temperature control function can be tracked by performing
periodic self-diagnostic tests. There are 3 items in the diagnostic test:
① Rate Test
Tests the heating and cooling rates.
② Cycle Test
Tests the cycle time, overshoot and undershoot .
③ Up-Down Test
Measures the heating and cooling time required to go up 98°C and then down to
10°C.
1) Select [Instrument]→[Diagnosis] from the Initial screen.
Self-diagnostic function
from software Initial Screen
Self-diagnostic function
from after File→Open operation
TP700/760 Instrument Manual
Rev. A
23
2) The Self-Diagnosis Screen is displayed.
Select the diagnostic items, and click the Start button. The first step of the
diagnostic program starts.
[Note]: If the first phase of the diagnostic program does not change to orange in
the diagram, click the Stop button once and click the Start button again.
① Rate Test Program Execution Screen (Refer to the previous figure.)
Program Contents: 25°C/30 seconds, 94°C/30 seconds, 40°C/30 seconds
Specification:
Maximum heating rate greater than or equal to 3.0°C per second
Maximum cooling rate greater than or equal to 2.4°C per second
1. Check diagnostic items.
2. Click Start
button. 3. It is progressive time.
Results are displayed when self-diagnostics are complete.
TP700/760 Instrument Manual
Rev. A
24
② Cycle Test Program Execution Screen
Program Contents: 94°C/30 seconds, 55°C/30 seconds, 72°C/30 seconds
Specification:
Overshoot less than or equal to 0.5°C
Undershoot less than or equal to 0.5°C
Length of run time : Within 4 minutes
③ Up-Down Test Program Execution Screen
Program Contents: 98°C/1 minute, 10°C/1 minute
Specification:
Length of run time within 6 minutes
TP700/760 Instrument Manual
Rev. A
25
2. Optical Calibration
Optical calibration is necessary for calculation of baseline values to ensure that the
obtained 48-well fluorescent display quantities are uniform. When the two
standard-specification wavelength filters are used, use the FAM and ROX calibration plates
which were specially designed for use with the Thermal Cycler Dice™
Real Time System
Lite. It is recommended that calibration be carried out once a year. Check for any soiling or
contamination of the Block before performing optical calibration, and clean if necessary.
Prepare the Test Reagent and the following tools and equipment:
One (1) FAM calibration plate for TP700
One (1) ROX calibration plate for TP700
Cotton swab (for cleaning the Block)
Ethanol (for cleaning the Block)
Plate Centrifuge
Powder-free gloves
1) Cleaning the Sample Block
If the Sample Block is dirty, it may affect the strength of the fluorescence. In order to
check whether the Sample Block is soiled/dirty, tighten the Lid without putting anything
on the Sample Block, and then photograph in FAM and ROX modes and verify that
fluorescence is not visible. If dirty, clean using alcohol and a cotton swab.
Verification of Sample Block Soiling
① Verify that there is no plate on the Sample Block, and close the Lid.
② Open [Instrument]→[Calibrate]→[Optical] from the Initial Screen Menu.
③ Using operation system Windows® 7,
the Use Account Control message appears.
Click “Yes” on the screen.
TP700/760 Instrument Manual
Rev. A
26
④ Close the initial screen menu and appear again.
In this condition, allow to access the program.
Open again [Instrument]→[Calibrate]→[Optical] from the Initial Screen
Menu.
⑤ Click “Take Picture” in the Optical Calibration Screen.
Set the fluorescent dye to FAM with Expo=200, Gain = 1
in the “Camera _Setting” items.
Set to
Expo [200]
Gain [ 1 ]
Set to “FAM”.
Photo Shoot button
Click
TP700/760 Instrument Manual
Rev. A
27
⑥ Click the TakePicture button, and a Block Image photograph is taken.
A confirmation message is displayed when photographing is complete.
Click OK.
[Note] Take Picture is started on 103 degree C. of Lid temperature.
⑦ Verify the image.
⑧ If the Sample Block is soiled, soak the cotton swab ethanol and wipe off
any accumulated dirt.
TP700/760 Instrument Manual
Rev. A
28
⑨ Change to ROX, set Gain to [6] and photograph the Block in the same
way .
⑩ If the Sample Block is soiled, soak the cotton swab in ethanol and wipe
off the accumulated dirt in the same way.
Caution: Be sure to verify the image with both FAM and ROX. Sometimes
soiling is not visible with FAM but is visible with ROX, and vice versa.
Expo [300]
Gain [ 3 ]
ROX
Photo shoot button
TP700/760 Instrument Manual
Rev. A
29
2) Optical Calibration
Preparation ① The FAM and ROX Calibration Plates each perform 3 cycles of PCR
in the Thermal Cycler Dice™
Real Time System. Melt the calibration
reagent and stir by convection.
PCR program: (95°C/5 seconds, 60°C/30 seconds) x 3 cycles.
② For the 48-well plate, perform the centrifugal process for 3 minutes at
1500 to 1800 rpm to remove Reagent and air bubbles adhering to the
well wall surface. Verify by visual inspection.
Caution: After the centrifuging process, verify that there is no liquid adhering to the
plate wall surface and that no bubbles are present.
If there is still liquid adhering to the walls or bubbles remaining, continue
the centrifugal process until the liquid drops solidly into the tip.
Caution: Wear powder-free gloves for experiments when conducting Optical
Calibration, and be careful not to touch the bottom surface of the plate or
the top surface of the seal.
When the Calibration Plate becomes dirty, Optical Calibration cannot be
performed accurately, and this may affect all test data for experiments
performed afterwards.
Procedure
① Verify that the main unit has completed warming up and the Standby LED is ON.
② Select [Instrument]→[Calibrate]→[Optical] from the Software Screen.
If appear the User Account Control message,
click “Yes”. The initial screen close and open again.
Select [Instrument]→[Calibrate]→[Optical].
○
Liquid level is even
X
Liquid coming into contact with wall surface
X
Contains bubbles
TP700/760 Instrument Manual
Rev. A
30
③ The Optical Calibration Screen appears.
Verify that only "SearchWellCenter" is at the top
and select FAM from the drop-down menu.
④ Slide the Lid into the main unit and insert the FAM
Plate into the Sample Block.
Caution: Adjust the direction of the plate and insert
securely into the Sample Block.
⑤ Close the Lid securely and wait for 1 minute.
Caution: When the Calibration Plate sheet becomes
cloudy, this influences calibration.
Therefore, when the sheet is heated,
wait approximately 1 minute until the
cloudiness disappears.
Caution: During calibration, the Lid does not lock.
Therefore, be careful not to open the Lid
while calibration is being performed.
⑥ Click the SearchWellCenter button.
⑦ A message is displayed confirming that the Calibration Plate (FAM) is set on the
Sample Block.
Verify that the FAM plate is in place, and click OK.
The right screen is displayed when the
center of each well is detected.
Click OK. The center points of the detected wells are displayed.
③
⑥
TP700/760 Instrument Manual
Rev. A
31
⑧ A red spot is displayed in the center of each well. Verify by visual inspection that
this is the center of the Well.
Click the Save button.
Caution: If the red spot is not displayed in the center,
click the SearchWellCenter button again.
Caution: Data is not retained if Save is not selected.
Always be sure to click the Save button.
⑨ When the Save operation is complete, the screen below appears. Click the OK
button.
The diameter of the red circle becomes large, and a confirmation message
is displayed in the center.
⑩ Click the Calibration tab to proceed to Calibration.
Perform Calibration.
Verify that the fluorescent
pigment is FAM, and click
the Calibrate button.
TP700/760 Instrument Manual
Rev. A
32
⑪ The Calibration Verification Screen is displayed.
Verify that the Calibration Plate (FAM) is on the Sample Block and click OK.
⑫ FAM calibration starts.
⑬ When FAM calibration ends, a verification screen is displayed. Click OK.
FAM calibration is now complete.
After leaving to cool, open the Lid and remove the FAM Calibration Plates.
Caution: The Calibration Plate becomes hot due to proximity to the Lid heater.
Remove only after it has cooled off.
Next, perform ROX calibration in the same way as FAM calibration only installed
ROX filter.
TP700/760 Instrument Manual
Rev. A
33
⑭ Insert the ROX Calibration Plate into the Sample Block and close the Lid.
Click the SearchWellCenter tab A in the area on the left and designate the
fluorescent pigment as ROX in the drop-down menu B . Click the
SearchWellCenter button C , verify that it is the ROX plate, and click OK D
when the confirmation message is displayed.
Caution: Verify that the pigment is ROX in the Fluorescent Pigment Selection
Screen.
Caution: The Lid does not lock during calibration. Therefore, be careful not to
0pen the Lid while calibration is being performed.
⑮ The task of detecting the central position of the wells is complete. Click OK.
A B C D
TP700/760 Instrument Manual
Rev. A
34
⑯ When the central points of the wells are displayed, verify by visual inspection that
they represent the well centers, and click the Save button.
Caution: If the Save button is not selected, data is not retained.
Always be sure to click Save .
⑰ When saving is complete, the screen below appears. Click OK.
Then click the Calibration tab to proceed to Calibration.
⑱ Select ROX as the fluorescent pigment type from the drop-down menu and click
the Calibrate button.
When the confirmation screen appears, verify that the ROX Plate is set,
and click OK.
TP700/760 Instrument Manual
Rev. A
35
⑲ ROX calibration begins.
⑳ When calibration ends, a confirmation screen is displayed. Click OK.
Lastly, remove the ROX Calibration Plate; be sure not to forget this step.
This completes the calibration operation.
TP700/760 Instrument Manual
Rev. A
36
3. Daily Inspections
Perform the following inspections and cleaning periodically in order to maintain the
operational performance of the equipment,
Recommended Once-Weekly Maintenance
Clean the Ventilation Intakes and exhaust (front/back/left/right).
Wipe clean with a cloth, or if there is significant accumulation of dust, clean with a
toothbrush.
* The main equipment, including the Sample Block, LED’s light source and
Electronic circuits, need to be cooled by air. Dirty ventilation intakes may cause
decreased functional capability and shortening of operating lifetime.
Clean the Lid and the main unit.
Wipe the outside of the equipment with a cloth soaked a little in a neutral detergent.
Then wipe with a dry cloth.
Verify the sliding action of the Lid. Wipe off any dirt or debris around the rails.
* Wiping the Lid rails with a dry cloth is sufficient. It is not necessary to lubricate.
Disinfect by wiping with a 10% bleach solution. Allow to leave for 10 minutes, then
wipe with a 70% alcohol solution in the same place. Lastly, wipe with a dry cloth to
dry the surface.
Caution: If bleach solution remains on the equipment, it may lead to corrosion.
Therefore, always wipe with alcohol after using bleach.
Recommended Once-Monthly Maintenance (or occasionally based on frequency of
use or soiling conditions):
Clean the Sample Block.
* For the cleaning method, refer to 3. Optical Calibration 1) Sample Block Cleaning
in this chapter.
Once Yearly
Optical Calibration is recommended.
TP700/760 Instrument Manual
Rev. A
37
● Chapter 6: Troubleshooting
Symptom Cause Correspondence
Nothing happens when the
Power Switch is turned ON.
Power is not being supplied. Verify the Power Cord connection on
the main unit side and the power
outlet side.
The Power Cord is not securely
plugged in to the power outlet or
the main unit connector.
The power outlet is not providing
power.
Check the power source equipment.
Thermal protector worked on
overheat of the sample block.
Wait the moment that the sample
block cooled.
Thermal fuse on the cooling fin
was melt down.
Need to replace it. Make a request
for repair.
Trouble occurred on electronic
circuits.
The Power Switch remains
OFF even when it is turned
on.
Malfunction of internal electrical
circuits in the main unit.
(The Power Switch works as the
over-current breaker.)
Immediately unplug the Power
Cable and make a request for repair.
Power Switch malfunction. Replace Power Switch.
It occur unusually when the
power switch turned on.
Optical filter changer is not
stopped on home position.
You hear a strange noise.
Turn off the power as soon as
possible and make a request for
repair.
The Lid does not close. There is an obstruction between
the Lid Heater and Sample Block.
Remove the obstruction.
Sample plate or tubes do not
seated correctly into the well.
Re-seat them correctly.
The Lock function of the Lid is
broken.
Disassemble the Lid and make a
request for repair.
Software do not launch. Software is not installed correctly. Delete the software and re-install it.
Not installed .NET Framework 2.0
or later version.
Install .NET Framework 2.0 or later
version.
TP700/760 Instrument Manual
Rev. A
38
Symptom Cause Correspondence
The program does not
start.
The main unit is not connected to
the PC by USB cable.
Connect the USB cable.
No communication between the
main unit and the PC.
Refer to the last symptom.
[ Instrument “Disconnected” and/or
Camera “Disconnected”].
Open the experiment file that is
not a new experiment file.
Create a new experiment file.
Warm-up for the Lid Heater has
not yet finished to warm up.
Try again after allowing 5 minutes
for the equipment to warm up.
The heater or a safety fuse has
burned out.
A part needs to be replaced. Make a
request for repair.
When starting the program,
the message “Please put
the lid on” is displayed.
The Lid is not completely closed. Close the Lid completely.
The Lid switch is broken. The switch must be replaced or
adjusted. Make a request for repair.
The program cycle stops
before completion.
The stepping motor has
malfunctioned.
The part needs to be replaced and
calibrated. Make a request for
repair.
The position-detecting sensor has
malfunctioned.
Make a request for repair.
Under PC Power Settings, any
settings are OFF.
Monitor/Hard Disk power is always
ON. Change to “OFF”.
Sleep setting is some hours.
Change to “Never” under “sleep
after”.
Under Screen Saver Settings on
PC, its setting is not OFF.
Change to “OFF” .
Thermal protector switch on the
Sample block heat sink works for
over temperature.
The cycler is too hot. Set the
environmental operating
temperature to 20-23℃. Make
sure the Ventilation. If the protector
or a fan is broken, need to replace.
Make s request for repair.
TP700/760 Instrument Manual
Rev. A
39
Symptoms Cause Correspondence
The program cycle stops
before completion.
When the PC was started up, the
camera or/and the USB-serial
bridge driver was not recognized.
On the start screen, display the
message under bar.
[Camera: Disconnected]
Refer to 1,2,3
[Instrument: Disconnected]
Refer to 1,2,4
Refer to the section on Computer
settings in Appendix .
1. Shut down the system. After
turning the power ON for the main
unit, turn the power ON for the PC.
2. Alternatively, leaving the power
ON for both the PC and the main
unit, unplug the USB cable, plug it in
again, and wait 1 minute.
3. The method for verifying the
camera drivers is to select [Imaging
device] under [Start]→[Control
Panel]→[System]→[Device
Manager]. Check to see if the device
is recognized.
4. Under the device manager on the
PC, make sure “COM1” of Ports /
ATEN USB to Serial Bridge (COMx).
On the control PC, work on
another operation except unit
control.
Operate the system only.
PC had too loaded and then
stopped.
The fluorescence
measurement value
(graph) is abnormal.
< Graph is too low around
baseline.>
The Test Reagent Plate is not in
place.
Set the Test Reagent Plate.
LED can not light. LED or a power supply is broken.
Make a request for repair.
The camera has malfunctioned. The part needs to be replaced.
Make a request for repair.
The fluorescence
measurement value
(graph) is abnormal.
< Graph is wavy.>
The Filter Unit is no correctly set. Correctly set the Filter Unit.
Filter unit exchanger is no move
smoothly.
Need it to replace or repair.
Make a request for repair.
LED or power supply for LED is
malfunction.
make a request for repair.
TP700/760 Instrument Manual
Rev. A
40
Symptom Cause Correspondence
The rate of temperature
heating or cooling is too
low.
The environmental operating
temperature being used is not
appropriate.
Set the environmental operating
temperature to 20-23°C and perform
the self-diagnostic test. If out of
range, make a request for repair.
The Peltier element is broken. The part needs to be replaced.
Make a request for repair.
Intake and exhaust are close to
the walls.
Make sure that ventilation is good.
Cooling fan or Exhaust fan is
broken.
Need to replace it. Make a request
for service.
The computer froze. CPU confused. Re-start the computer. Running data
is saved until PC froze.
PC can not communicate to
the main unit.
On the start screen, display
the followings under status
bar.
Instrument “Disconnected”
and/or
Camera ”Disconnected”
Power switch is OFF position. Turn the switch to ON.
The camera driver or/and the
USB-serial bridge driver was not
recognized.
1. Shut down the system. After
turning the power ON for the main
unit, turn the power ON for the PC.
2. Alternatively, leaving the power
ON for both the PC and the main
unit, unplug the USB cable, plug it in
again, and wait 1 minute.
Connected another USB terminal
on the PC for moving the system.
Connect the same USB port of start
setting system.
Under the device manager on the
PC ,verify “COM1” of Ports/ ATEN
USB to Serial Bridge(COM x).,
Change to “COM1” if not.
The camera driver was not
recognized.
The method for verifying the camera
drivers is to select [Imaging device]
under [Start]→[Control Panel]→
[System]→[Device Manager].
Check to see if the device is
recognized.
Port setting of using PC changed
automatically. for example case,
Insert another USB port or
connect a USB device to
recognized USB port.
Verify the driver settings as follows;
Start>Control Panel>Hard>Device
Manager/ Port ATEN USB=COM1,
and Imaging devices =Lumenera
(refer to Appendix 1)
TP700/760 Instrument Manual
Rev. A
41
Error code
Display the following error when the system catches error. For recover the system shut
down and restart. If it will not clear, please make a request for repair.
Error code Cause
11 Indicates disconnected center sensor on the sample block.
12 Indicates short in center sensor on the sample block.
13 Indicates disconnected left sensor on the sample block.
14 Indicates short in right sensor on the sample block.
15 Indicates disconnected right sensor on the sample block.
16 Indicates short in right sensor on the sample block.
17 Indicates disconnected lid sensor.
18 Indicates short in lid sensor.
19 Indicates disconnected cooling plate sensor on the sample block.
20 Indicates short in cooling plate sensor on the sample block.
21 Indicates disconnected circuit of the cooling fan.
22 Indicates short in circuit of the cooling fan.
23 Indicates lid temperature control.
24 Indicates block temperature control error.
26 Indicates home position error.
TP700/760 Instrument Manual
Rev. A
42
● Chapter 7: Specifications
Main Unit Basic Specifications
Product Name Thermal Cycler Dice™ Real Time System Litet
Part Number TP710
Objects Processed 48-Well Plate Cap Seal Type
8 Series Tube Cap Seal Type
Optical Detection
System
CCD CAMERA (1280 x 960 resolution).
Computed for each well based on the image data.
Optical Filter Filter Unit (Include excitation/emission filter and dichroic mirror)
2 Filter Units as standard:
FAM (SYBR Green I ), ROX (Texas Red)
Maximum of 4 can be installed: FAM, ROX, Cy5, HEX/VIC
Light Source LED x 3 (Blue LED x 1, White LED x2 )
recommended total usage time: 50,000 hours
Temperature Control
Method
PID Control (Temperature Sensor Pt 1000Ω)
Heating and Cooling
System
Peltier Element
Cooling Capability 10°C (reaction block temperature when the environment
temperature is 30°C)
Temperature Range 10.0 - 99.9°C (increments of 0.1°C)
Temperature Display
Information Display
Sample temperature calculated from block temperature; display
sent to control PC (increments of 0.1°C)
Main Unit Display LED display: POWER / Standby / RUN
Rate of Temperature
Increase/Decrease
Maximum 3.0°C / second temperature heating rate
Maximum 2.4°C / second temperature cooling rate
(with an environment temperature of 20°C).
Run Modes Fast mode, Normal mode
Lid Holding
Temperature
When set program temperature to 20°C or more:
Lid temperature at about 108°C.
When set program temperature to less than 20°C:
Lid temperature at about 40°C.
Lid Lock Function Locks the Lid with an electromagnetic lock while running,
and releases the lock when all processes are complete.
Operating the Pause or Stop functions during a run releases
the lock. When a paused run is restarted, the lid is locked again
and the run continues.
TP700/760 Instrument Manual
Rev. A
43
Optical Calibration
Function
For use with Takara Thermal Cycler Dice Real Time System Lite
FAM/ROX Calibration Plate. (FAM/ROX).
Self-diagnostic
Function
Various sensors for broken wires, short circuits, fan lock, and
Block Temperature, Lid Temperature, or Heating/Cooling Rate
abnormalities
Safety Devises Lid Heater : Thermo Fuse ( 169 degree C)
Heat Sink of the Sample Block : Thermal Guard (85 degree C)
+ Thermo Fuse (113 degree C)
Main Switch : Over Current Switch (10 A)
Timer Battery Lithium Cell, Type CR2032 3V
Connection with PC USB 1 port (COM1 Port use)
Supply Voltage AC 100-240 V 50/60 Hz
Maximum Current 3.7 A (at 100 V)
Power Cord 125 V 15 A cable (3 m) including a 3-pin plug connection with a
ground pin (commonly used in Japan and North America)
Exterior Dimensions
/ Weight
Width 294 × Depth 480 × Height 304mm
18 kg for main unit only. (21.5 kg including packaging.)
TP700/760 Instrument Manual
Rev. A
44
Basic Specifications of Computer
Product Name Computer to be used with TaKaRa Thermal Cycler Dice™ Real Time
System
Part Number TP820
PC Model (ex.) Fujitsu LIFE BOOK A561/D
Dimensions /
Weight
374 (Width) × 245 (Depth) × 35 (Height) mm
2.4 kg
Specifications OS: Microsoft® Windows
® 7 Professional Service Pack 1
32bit (English / Japanese)
Intel® Core™ i 5-2520M Processor (2.50GHz)
Memory: 2 GB
Hard Disk: 250 GB
LCD Size: 15.6 inch HD(1366x768) LED back light LCD display
Optical Drive: DVD super multi drive
USB port x 4 USB version 2.0
Standard
Accessory
USB mouse, USB Memory key
Microsoft® Office Home & Business 2010 (English or Japanese):
include Word 2010/ Excel® 2010/ PowerPoint
® 2010/ Outlook
® 2010
/ OneNote®2010
Note: The computer specifications may change according to market
conditions. This specification is at the time of June, 2012.
Application Software Basic Specifications
Product Name Thermal Cycler Dice™ Real Time Software
Part Number TP730
Media One (1) Program CD and One(1) Driver software CD
For OS Windows® XP / Vista / 7 32bit / 64bit *
* This system driver are installed 32bit version for the Camera and the USB
Converter. The software can use on another computer that is installed the 64 bit
OS for analysis only.
TP700/760 Instrument Manual
Rev. A
45
● Chapter 8: Consumption
■ 48 well plate
Code Product Quantity Comments
NJ700 48 well Snap Plate 20 plates
NJ720 Flat Cap for Snap Plate 120 seats Cap for NJ700
■ Strip Tubes
Code Product Quantity Comments
NJ300 0.2ml Hi-8-Tube 125 strips Strip of 8 attached caps for use
with 0.2 mL strip tubes
NJ302 0.2ml Hi-8-Flat Cap 125 strips Flat caps for easy closure and
sample labeling
■ Strip Tube with individual Flat Cap
Code Product Quantity Comments
NJ600 0.2ml 8-Strip Tube, individual Flat
Caps
120 strips Strip of 8 tubes with each cap.
Can use one tube by separation.
■ Option Filter Units(TP900/960, TP800/860 Multi Wave Model)
Code Product Quantity Comments
TP702 Filter Unit HEX for TP700
(For HEX / VIC)
1 unit Only Factory setting
TP703 Filter Unit Cy5 for TP700
(For Cy5)
1 unit Only Factory setting
TP700/760 Instrument Manual
Rev. A
46
Appendix
Thermal Cycler DiceTM
Real Time System TP700/760
Computer settings
☆☆ Windows 7 ☆☆
A. Computer Driver Setup - - - - - 47 B. Software Install / Uninstall - - - - - 52 C. Computer Basic Setup - - - - - 57
TP700/760 Instrument Manual Windows7
Rev. A
47
☆☆ Windows 7 ☆☆
A.Computer Driver Setup
Computer Driver Settings
① Search the “flycapture 2.2.3.14_x86.exe” from the Thermal Cycler Dice™ Real Time
System Driver Software CD and double-click to launch it.
On the way, when the User Account
Control dialog appear, click on Yes .
② Appear the Setup Wizard, Click the
Next button.
③ Check the accept the terms in the
License Agreement to agree to the
License Agreement and
click on Next .
④ Click on Next .
FlyCapture2 Installer
CD/DVD drive
TP700/760 Instrument Manual Windows7
Rev. A
48
⑤ Input your name and organization if
need, select “Anyone who uses this
computer” and release the check of
“internet”.
Click on Next .
⑥ Click on Next again.
⑦ In the next screen, select “Complete” .
Then click on Next .
⑧ Check two items of USB and release
the check mark of “Install UsbPRO....”.
Click on Next .
Check on
□Install UsbPRO.. remove the check
check
In the next screen, check “Click
confirm” and then click on Next .
Remove the check.
TP700/760 Instrument Manual Windows7
Rev. A
49
⑨ Confirm the check “The installer will disable ....” , and click on Next .
Then click on Install .
⑩ Start to install and finish the wizard of
camera driver. Click on Finish .
⑪ The restarts windows appear, click on Yes for restart.
⑫ Open the device manager.
Show the “PGR Chameleon CMLN-13S2M”.
TP700/760 Instrument Manual Windows7
Rev. A
50
⑬ Install a USB – Serial Bridge driver.
Double-click on the ATEN “cvrs9_1.0.075.exe” file inside the Driver CD.
⑭ Right-click on “Computer” icon on the
desktop ( or Start > Computer ) and select
Properties from the menu. Open the
“Device Manager”.
⑮ Under Ports (COM & LPT), set “ATEN
USB to Serial Bridge” (COMx) to COM1.
If OK, go to the item ⑲.
⑯ To change the settings, right-click on
“ATEN USB to Serial Bridge (COM x) ”
and open “Properties” from the menu.
⑰ In the Properties box, under Port
Settings, click on Advanced .
Caution: If “ATEN USB to Serial Bridge”
is not set to COM1, then change to “COM1”
as the following procedures.
TP700/760 Instrument Manual Windows7
Rev. A
51
⑱ Set COM Port Number to COM1
and click OK.
⑲ Close the Device Manager, and open it again to verify that the settings are correct.
Under Ports (COM & LPT) Settings, verify that the “ATEN USB to Serial Bridge
(COM1)” setting is in effect. For other devices, verify that “?” is not displayed. If a
question mark is displayed, double-click on that device, open Properties, and click
Reinstall Driver. The driver is then reinstalled.
⑳ If the computer has multiple USB ports, set up the driver according to the
procedure described above after changing the USB cable connection. Also, make
sure that the port setting is COM1.
Caution: Make sure to connect the main unit to the USB port on the computer that
has been set. If there are multiple ports and all of them are set in the same
way, then every time the computer is used, the connected USB cable will
be automatically detected. If this setting has not been performed, the Install
Shield Wizard appears and the driver settings are requested. Please set
the driver settings if they have not already been carried out.
21. Eject the CD and reboot the computer.
TP700・760 Instrument Manual Windows 7
Rev. A
52
☆☆ Windows 7 ☆☆
B.Software Install / Uninstall
If the Thermal Cycler Dice Real Time System Software is being installed for the
first time, perform the steps described in section 1 (Install).
If the software version is being upgraded, first perform the steps in section 2
(Uninstall), and then start the procedure from section 1.
1. Install
① Insert the Thermal Cycler Dice Real Time System Software CD in to the CD drive.
② Make sure the OS is Windows 7
English version. Double-click on
the English folder in the CD drive.
③ Double-click on the setup.exe file
inside the open folder to launch the
installation program.
④ In the initial screen, click the Next
button.
TP700・760 Instrument Manual Windows 7
Rev. A
53
⑤ Click the Next button in the screen
that is displayed below.
○ The installation folder can be
changed, but please use the
default folder unless there are
special circumstances.
○ Use default (“Just me”) for the
user settings.
⑥ Click the Next button in the screen
below. The installation process
begins.
⑦ The Installation Execution screen
is displayed.
On the way, when
“User Account Control”
appear ,click on Yes .
⑧ When installation is complete, the right screen is displayed. Click “Close” button.
And then The restarts window appear, click on “Yes”.
CD/DVD drive
RP1_RealTimePCR_Res.exe
TP700・760 Instrument Manual Windows 7
Rev. A
54
2. Uninstall
① From the Start menu, go to Control Panel and double-click on the Programs icon.
② When the “Uninstall or Change a Program” window is displayed, select “Thermal
Cycler Dice Real Time System Software ” from the Currently Installed Programs list.
③ Click the Uninstall button.
When the confirmation dialog box is displayed, click Yes.
View by Category View by Small icons
TP700・760 Instrument Manual Windows 7
Rev. A
55
④ The uninstall operation starts.
On the way, when the User Account
Control dialog
appear, click on “Yes”.
⑤ When complete, a dialog box inquiring whether to restart now or later is displayed.
If additional software is to be installed, select “No”. The following screen is
displayed. After verifying that the Thermal Cycler Dice Real Time System program
has been removed, click Close. Then proceed to the installation steps detailed in
section 1.Alternatively, click Yes to restart the computer and continue settings.
RP1_RealTimePCR_Res.exe
TP700/760 Instrument Manual Windows 7
Rev. A
56
☆☆ Windows 7 ☆☆
C. Computer Basic Setup
Computer Settings
① In the Start menu, right-click on
Computer.
Click “Show on Desktop” to display
the Computer icon on the desktop.
This icon can then be selected when
needed for setting purposes.
② Right-click on the screen and open
“Personalize”. Under the Screen
Saver tab, set “None” for Screen
Saver and click on “Apply” .
Then, click on the change power
settings
TP700/760 Instrument Manual Windows 7
Rev. A
57
③ Under Power Options, select “High performance” and “Change plan settings”.
If “High performance” is not appeared, click on the arrow (Show additional plans).
In the Edit Plan Settings, set as shown below and click on the Save changes.
④ In the same way, Power Options >
High performance > Change plan
settings > Change advanced
power settings. Set as follows.
Dim the display
Turn off the display all “Never”
Put the computer to sleep
Click on the Save changes.
TP700/760 Instrument Manual Windows 7
Rev. A
58
⑤ Right-click on the Computer icon
and click Properties to enter the
System Properties window.
Click on “Advanced system
settings” to enter the system
properties, click Advanced tab,
then “Settings...”.
⑥ Under the Visual Effects tab,
remove the checks from the
following two boxes and click OK.
→ □ Save taskbar thumbnail
previews
→ □ Show windows contents
while dragging
⑦ Ensure that file extensions are
displayed. Under [Start]→[Control
Panel (view by Large icon or Small
icon)] →[Folder Options], click the
View tab. Under Advanced Settings,
remove the check mark beside
“Hide extensions for known file
types”. Click on “OK” button.
TP700/760 Instrument Manual Windows 7
Rev. A
59
⑧ Hardware Verification
Right-click on the Computer
icon on the desktop and select
[Properties]→[Device Manager].
Verify that the following devices
are installed normally. If a
question mark is shown for an
item, right-click on that icon and
select [Properties]→ [Re-install]
to re-install the drivers.
・ Imaging devices: Lumenera
USB 2.0 Camera is normal.
・ Ports (COM & LPT) is normal
and COM 1 is selected.
・ All the USB drivers are
normal.