Toxic and Harmful Compounds by ICP-MS final ESAC 2008.pptamcham.dk/dl/esac/ESAC08-1.pdf ·...
-
Upload
nguyennhan -
Category
Documents
-
view
224 -
download
0
Transcript of Toxic and Harmful Compounds by ICP-MS final ESAC 2008.pptamcham.dk/dl/esac/ESAC08-1.pdf ·...
ESAC Copenhagen15th April 200815th April 2008
Determination of Essential, Therapeutic and Toxic Elements and Their Compoundsand Toxic Elements and Their Compounds
in Biological Materials; Applications of ICP MS i th Cli i l L b tICP-MS in the Clinical Laboratory
Ed McCurdy, ICP-MS Specialist, Agilent Technologies Ltd
ESAC April 2008
Scope
Introduction to ICP-MS for clinical sample analysis• Requirement for elemental analysis in clinical samples• Routine trace element monitoring
f f• Analysis of toxic and harmful elements– Elemental screening, poisoning, radionuclides
T i t lli d– Toxic organometallic compoundsICP-MS analysis for the measurement of organic compounds• Identification and quantification of pesticide residues• Quantification of Chemical Warfare Agent degradation productsF i A li ti f ICP MSForensic Applications of ICP-MS• Laser Ablation ICP-MS• Direct Analysis of Elemental Distribution in Tissues
Page 2
ESAC April 2008ESAC April 2008
What is ICP-MS?
ICP Inductively Coupled Plasma MS Mass Spectrometer
An inorganic (elemental) analysis techniqueICP - Inductively Coupled Plasma• high temperature electrical discharge,
which decomposes, atomizes and ionizes
MS - Mass Spectrometer• quadrupole (“quad”) mass analyzer• mass range from 5 to 260 amu (Li toc deco poses, ato es a d o es
samples– forms ions, so compounds not measured
directly
• mass range from 5 to 260 amu (Li to U...)– separates all elements in rapid
sequential scandirectly sequential scan– isotopic information available
• ions measured using dual mode detector– ppt level LODs for most elements– Calibration range up to 1000’s ppmCalibration range up to 1000 s ppm
• Spectral interferences removed using collision/reaction cell
Page 3
ESAC April 2008
Agilent 7500cx ICP-MS System withCollision/Reaction Cell (CRC)Collision/Reaction Cell (CRC)
Multi-element interference removal by on-axis octopole reaction cell
Fast simultaneous dual mode detector (9 orders dynamic range)
Reaction Gas InletHigh temperature 27MHz plasma
Reaction Gas Inlet
27MHz plasma generator
Off-axis L
OctopolePlasma
Low flow sample introduction system
High frequency hyperbolic quadrupole
Lens
Page 4
ESAC April 2008ESAC April 2008
Elements of Interest in Easily Extractable Fluids (Urine Blood Serum Plasma)Fluids (Urine, Blood, Serum, Plasma)
Highly toxic heavy metalsHighly toxic heavy metals• As, Cd, Pb and Hg
Potentially toxic elements• Al, Sb, Ba, Be, Bi, Li, Ni, Sr and Tl
Essential elementsC C C M M S V d Z• Cr, Co, Cu, Mg. Mn, Se, V and Zn
Wid f l t b d i l diWide range of elements may be measured, including many, such as Be, As, Se, Hg, U, which are considered
“difficult” by other techniques
Page 5
ESAC April 2008ESAC April 2008
y q
Which Elements can be Measured Using ICP MS?ICP-MS?
All elements in colour can be measured only those elements present in the– only those elements present in the
plasma gas, <5amu, >260amu, and those which are not ionized, are inaccessible
Page 6
ESAC April 2008
Analytical Needs for Inorganic Measurements in Clinical Laboratoriesin Clinical Laboratories
The technique must possess robust sample introductionThe technique must possess robust sample introduction
• To handle large sample numbers and differing matrices routinely monitored
Measurement of many elements in the same fast acquisitionMeasurement of many elements in the same fast acquisition
• Removal of interferences• Screening applications Quadrupole ICP MS can• Screening applications• Sample turnaround/productivity• Reduced cost of analysis
Quadrupole ICP-MS can meet all of these criteria
Reduced cost of analysis
Make measurements at trace levels (low detection limits) and at high concentrations in the same acquisitionq
• To reduce reruns, improve productivity and lower costs
The ability to measure element species is also provided by ICP-MSThe ability to measure element species is also provided by ICP MS
• Expands range of applications, and provides useful background research info
Page 7
ESAC April 2008ESAC April 2008
Collision/Reaction Cell to Remove M t i B d S t l I t fMatrix-Based Spectral Interferences
20
Ad t f C lli i /
15
(ppb
)
Cr52 Advantage of Collision/ Reaction Cell ICP-MS• Provides removal of
A C (
ArC and ClOoverlaps in15
tion
in B
lank
spectral overlaps –allows accurate trace element analysis in variable high matrix
ArC (mass 52) overlap in acetic acid matrix
overlaps in combined matrix
10
52 C
once
ntra
t
H2
He
NoGas
variable high-matrix samples
• Example shows the removal of ClO ArC
matrix
5
App
aren
t Cr5 NoGas removal of ClO, ArC
interferences in various matrix blanks
• He collision mode gives
ClO/ClOH(mass 52) overlap in HCl t i
00 1% 5% 5% HCl 1% 1% 200 200 500 Mi d
A He collision mode gives reliable removal of all polyatomic interferences regardless of the sample
HCl matrix
0.1% HNO3
5% HNO3
5% HCl 1% H2SO4
1% AcOH
200ppm Na
200ppm Ca
500ppm P
Mixed Matrix
Matrix Blank
matrix composition
Page 8
ESAC April 2008
Routine Analysis of Urine – Sample Preparation and CalibrationPreparation and Calibration
Di t l i f i l f ll i 1/5 1/10 ( / )Direct analysis of urine samples following a 1/5 or 1/10 (v/v) dilution with deionized water and nitric acid• No clogging of the nebulizer• No particle deposition in the injector tube over 12 hours of analyses
Calibration using Method of Standard Additions (MSA), so matrix-matched standards• Standard addition calib is then converted to external calib and applied to
subsequent samplessubsequent samples• 5 µg/L Tb internal standard added to all sample and calibration solutions
Agilent Application Note: Rapid and reliable routine analysis of urine by Octopole Reaction Cell ICP-MS, 5989-2482EN, by Peter Heitland, Medical Laboratory Bremen, Germany
Page 9
ESAC April 2008ESAC April 2008
MS, 5989 2482EN, by Peter Heitland, Medical Laboratory Bremen, Germany
Measured and Certified Concentrations in Urine Reference Material Lyphochek®Urine Reference Material Lyphochek®
Element Concentration (µg/L - ppb)Lyphochek, level 1 Lyphochek, level 2yp , yp ,
Measured(n=10, external)
certified Measured(n=10, external)
certified(n 10, external) (n 10, external)
Cr 1.7 ± 0.2 1.2 ± 0.2 18.6 ± 2.6 20.2 ± 4.1Co 6.6 ± 0.7 6.9 ± 1.4 18.9 ± 1.4 19.1 ± 4.2C 24 ± 2 1 26 7 ± 5 4 45 ± 5 5 50 ± 10Cu 24 ± 2.1 26.7 ± 5.4 45 ± 5.5 50 ± 10Se 56 ± 5.3 49 ± 10 192 ± 17 187 ± 37As 65 ± 6 67 ± 14 162 ± 15 163 ± 33Cd 8.4 ± 1.1 8.6 ± 1.7 14.9 ± 1.9 15.6 ± 3.1Sb 6.9 ± 1.1 9 ± 1.8 34.8 ± 4.4 36.9 ± 7Tl 9 6 ± 0 8 9 7 ± 2 0 185 ± 17 198 ± 40Tl 9.6 ± 0.8 9.7 ± 2.0 185 ± 17 198 ± 40Pb 13.5±1.1 14.3 ± 2.9 68 ± 5 69 ± 14
Data: Medical Laboratory Bremen, Germany
Page 10
ESAC April 2008
High Throughput Analysis of Blood Samples using 7500ce ICP MSusing 7500ce ICP-MS
Biomonitoring of trace elements in human blood samples - an important tool g p pfor occupational and environmental health
Goals of this study y
– Determine a high number of trace metals in blood of 130 unexposed subjects– Develop a rapid routine method for the multi-element analyses of blood usingDevelop a rapid routine method for the multi element analyses of blood using
collision/reaction cell-ICP–MS
Blood samples were collected in lithium heparin monovettes
• 500 uL of the sample was diluted with 100 uL 0.1% (v/v) Triton-X-100 solution and 500 uL of the internal standard solution
• This solution was made up to 5 mL with a 0.5% (v/v) NH4OH solution in a 10 mL polypropylene autosampler tube
More than 100 samples can be prepared in less than 1hour by one personMore than 100 samples can be prepared in less than 1hour by one person
Biomonitoring of 37 trace elements in blood samples from inhabitants of northern Germany by ICP–MS, Peter Heitland, Helmut D. Koster, Journal of Trace Elements in Medicine and Biology 20 (2006) 253–262
Page 11
ESAC April 2008ESAC April 2008
Heitland, Helmut D. Koster, Journal of Trace Elements in Medicine and Biology 20 (2006) 253 262
Analytical Figures of Merit for Blood Analysis
Limits of quantification q(LOQs), calculated in undiluted blood, range from 0 003 ug/L for 238U to 0 1 ug/L0.003 ug/L for 238U to 0.1 ug/L for 69Ga
Spike recoveries of 1 ug/L (10 ug/L for B, Mn and Sr; 200 ug/L for Cu and Rb) fromug/L for Cu and Rb) from single element calibration solutions are in the range g94 -111%
Page 12
ESAC April 2008ESAC April 2008Data: Peter Heitland, Helmut D. Koster, Medical Laboratory Bremen, Germany
Effects of Sample Matrix on the Sample Introduction SystemIntroduction System
Photos of the interface and l i t d ti tsample introduction system
after a 90-sample run (whole blood).blood).
Both the sampler and skimmer cones show only
Sample Cone Skimmer Cone
skimmer cones show only minor matrix deposits – none at the cone tips
The standard 2.5mm injector torch used was virtually deposit-free. The blood deposits on spray chamber and the nebulizer blockchamber and the nebulizer block were removed using a sodium hypochlorite solution
Page 13
ESAC April 2008Images: R. Wahlen et al., LGC Limited, UK
Screening for Toxic and Harmful Elements
Elemental Screening of 1:10 diluted Urine (with interference removal in He mode)
• Unique capability of ICP-MS to acquire a scan across the entire mass range in Screening to identify poisons (1:10 diluted urine scan)
• “Unknown” element spiked into urine sample
1 . 0 E 5
[ 1 ] S p e c t r u m N o . 1 [ 1 8 1 . 5 2 5 s e c ] : 0 0 2 S M P L . D # / T u n e # 1 [ C P S ] [ L i n e a r ] [
about 2 minutes, screening elements from 1000’s ppm to sub-ppb levels
1 . 0 E 5
[ 1 ] S p e c t r u m N o . 1 [ 1 8 1 . 5 3 1 s e c ] : 0 0 4 S M P L . D # / T u n e # 1 [ C P S ] [ L i n e a r ]
Unknown element spiked into urine sample
RbC NaMg
RbC NaMg
5 . 0 E 4
Zn
5 . 0 E 4
Zn
“Unknown” element
Ca
Li MoISr
Zn
CuLi MoISr
Zn
Cu
Ca
1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 0 1 0 0 1 1 0 1 2 0 1 3 0 1 4 0 1 5 0 1 6 0 1 7 0 1 8 0 1 9 0 2 0 0 2 1 0 2 2 0 2 3 0 2 4 0 2 5 0 2 6 0/ 0 6 0
PbFeBr Ba
ICs
SrSb
Sn1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 0 1 0 0 1 1 0 1 2 0 1 3 0 1 4 0 1 5 0 1 6 0 1 7 0 1 8 0 1 9 0 2 0 0 2 1 0 2 2 0 2 3 0 2 4 0 2 5 0 2 6 0/ > 0 0 2
PbFeBr Ba
ICs
SrSb
SnAsAs
Page 14
ESAC April 2008ESAC April 2008
1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 0 1 0 0 1 1 0 1 2 0 1 3 0 1 4 0 1 5 0 1 6 0 1 7 0 1 8 0 1 9 0 2 0 0 2 1 0 2 2 0 2 3 0 2 4 0 2 5 0 2 6 0m / z - > 0 6 01 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 0 1 0 0 1 1 0 1 2 0 1 3 0 1 4 0 1 5 0 1 6 0 1 7 0 1 8 0 1 9 0 2 0 0 2 1 0 2 2 0 2 3 0 2 4 0 2 5 0 2 6 0m / z - > 0 0 2
Analysis of Toxic and Harmful ElementsScreening to identify poisons (1:10 diluted urine scan) • Confirmation (from isotopic template) of presence of Thallium (2ppb spike)
[1 ] S p e c t r u m N o .1 [ 1 7 5 .1 1 8 s e c ] : 0 0 3 S M P L .D / T u n e # 1 [C P S ] [L in e a r ]0 5 S M P L .D # /
– Can be quantified (semiquant) by reference to known concentration element– Note 210Po would also be seen in this mass region of the screening acquisition
1 .0 E 5
Alexander Litvinenko - poisoning
5 .0 E 4
T l
205Tl
210Po203Tl
1 9 4 1 9 6 1 9 8 2 0 0 2 0 2 2 0 4 2 0 6 2 0 8 2 1 0 2 1 2 2 1 4 2 1 6m /z >
P b
2 0 4
208Pb
Page 15
ESAC April 2008ESAC April 2008
1 9 4 1 9 6 1 9 8 2 0 0 2 0 2 2 0 4 2 0 6 2 0 8 2 1 0 2 1 2 2 1 4 2 1 6m /z - > 2 0 4
Measurement of Radionuclides
Left: Calibration for radionuclides is easily achieved at sub ng/L (ppt) levels, even with standard sample introduction (U p (used for illustration)
Scan of 1ppt 237Np standard, showing high sensitivity and low random backgroundg
Radionuclides are typically easily ionized and the spectrum is free from overlaps andfrom overlaps and backgrounds, so LOD’s in the pg/L (ppq) range are achieved
Page 16
ESAC April 2008
Radionuclide Analysis – Quantification and Isotope Ratio Measurement
2.0E4
[1] Spectrum No.1 [ 115.786 sec]:10Urine.d / Tune #1 [CPS] [Linear]Isotope Ratio MeasurementVery high sensitivity and low background for 10ppt U in 1:10
238U (99.27%)
gdiluted urine
~1.2 million cps/ppb U
U
( )Natural U spike, so 235U (0.72% abundance) concentration was 72ppq
1.0E4ICP-MS also provides isotopic information, so U isotopic patterninformation, so U isotopic pattern (isotope ratio) can be used to identify source of contamination
N t l 0 72% 235U
235U (0.72%)
• Natural = 0.72% 235U • Waste depleted = 0.2 - 0.4% 235U• Pile depleted 0 6% 235U
228 229 230 231 232 233 234 235 236 237 238 239 240 241 242m/z->
• Pile depleted ~0.6% 235U• Enriched >0.72% 235U
Page 17
ESAC April 2008ESAC April 2008
Toxic Elemental Forms or “Species”
For many elements, the level of toxicity is highly dependent on the chemical form of the element, so separation (chromatography) is required
GC Laser Ablation
Optional
ICP-MSOptional
ConventionalDetector(s)
ESI MS
LC
– e.g. ESI-MS
1 0 0CE6 0
8 0
1 0 0
%
1 3 9 .0
1 4 4 .1
1 6 0 .11 7 4 .1 2 5 8 .9
2 7 6 .9
20000
25000
30000
35000
onse
(CP
S)
31P
123
4
56
78
910
0
2 0
4 0
72 04 36 68 00 32 64 96 28 60
2 0 3 .1
2 8 2 .1
0 5 10 15 20 25
0
5000
10000
15000
Res
po 10
Page 18
ESAC April 2008ESAC April 2008
7 10 13 16 20 23 26 29 32 36
m /z0 5 10 15 20 25
Time (min)
Example: LC-ICP-MS for As Speciation
Toxic! Many As species exist – theexist the inorganic As species are known to be toxic andLess-Toxic to be toxic and most organic species are
l i l h l
Less Toxic
Non-Toxic relatively harmless to humans.
The potentialThe potential toxicity of some species, such as th h i t f
?the huge variety of arsenosugars, has not yet been established.
Page 19
ESAC April 2008ESAC April 2008Courtesy Ute Kohlmeyer GALAB, Germany
Chromatogram of As Standard (1.0 µg/L each)
Agilent Application Note: Routine
A new column has been developed to provide routine separation of the 5 most common As species in urine:Column G3288-80000 (4.6 x 250 mm) g pp
Analysis of Toxic Arsenic Species in Urine Using HPLC with ICP-MS, 5989-5505EN, by Tetsushi Sakai and Steven
( )Guard Column G3154-65002
Mobile Phase (Basic): Wilbur, Agilent TechnologiesMobile Phase (Basic):2 mM phosphate buffer solution (PBS)pH 11.0 adjusted with NaOH0 2 mM EDTA0.2 mM EDTA10 mM, CH3COONa3.0 mM NaNO31% ethanol
Page 20
ESAC April 2008
Determination of Organo-As SpeciesUsing HPLC with ESI-MS & ICP-MSUsing HPLC with ESI-MS & ICP-MS
ESIESI--MSMSAgilent 1100Agilent 1100
ICPICP--MSMSAgilent 7500Agilent 7500 HPLCHPLC Agilent 1100Agilent 1100Agilent 7500Agilent 7500 Agilent Agilent
11001100
Elemental specific detectionElemental specific detection Molecular specific detectionMolecular specific detection
15 %15 % 85 %85 %
Single HPLC System with controlled split to provide sample flow to ESI-MS and ICP-MS. Provides simultaneous measurement of As-containing compounds and As concentration (ICP-MS) and concentration/structural information on the organic part of the As compounds (ESI MS)
Elemental specific detectionElemental specific detection Molecular specific detectionMolecular specific detection
Page 21
ESAC April 2008
Page 21
concentration/structural information on the organic part of the As compounds (ESI-MS)
Courtesy Jörg Feldmann et al, Aberdeen Univ.
Separation and Identification of Organo-Arsenic SpeciesArsenic Species
100
120
167.0Peak 1 (B)[M+H]+DMAE
Peak 2
P k 1pH 5.3
60
80
%
155.1
191 1y
Peak 3
Peak 1
ICP-MS m/z 75
ICP-MS
0
20
4094.9
116.1 173.1
191.1213.1
229.0
Inte
nsity
m/z 139
m/z 167
m/z 277/ 2 9 84 10
8
132
156
180
204
228
252
276
300
m/z
100 120DMA DMAA?m/z 361
m/z 181
m/z 259ESI-MS
60
80139.0
144.1
160.1 258 9
276.980
100
120
181.0
361.0
Peak 2 (C ) Peak 3 (D)[M+H]+[M+H]+
[2M+H]+[2M+H]+0 500 1000
Retention time (s)
m/z 91
Chromatograms (above) for ICP MS
40
60
%
174.1 258.9
282.1 40
60%
144.1
Chromatograms (above) for ICP-MS measurement of As (mass 75) and ESI-MS of various indicator masses. Mass spectra (ESI-MS) for the 3 peaks show
[2M-H2O]+
0
20
72 104
136
168
200
232
264
296
328
360
203.1
0
20
72 105
138
171
204
237
270
303
336
369
163.1155.1
214.2 302.0
characteristic fragmentation patterns for the identified species – DMAE, DMA and (maybe) DMAA
Page 22
ESAC April 2008
Page 22
m/z m/z
Hansen et al. J Anal At. Spectrom, 18, 474
Which Elements can be Measured Using ICP-MS?MS?
ICP-MS can also measure non-metals used in highly toxic compounds suchused in highly toxic compounds, such as pesticides and chemical warfare agents, provided the backgrounds can be controlled – e.g. using GC-ICP-MS
Page 23
ESAC April 2008
Agilent GC-ICP-MS Interface
GC-ICP-MS System used:ICP-MS: Agilent 7500
Fully heated and insulated GC transfer line
Modified torch with heated injector replacesGC: Agilent 6890Interface: Agilent G3158A
Modified torch with heated injector replaces standard demountable torch
“Silicosteel” transfer line and injector liner for inertness
GC effluent injected directly into base of lplasma
Very high transport efficiency, high plasma temperature (no water vapour/aerosol) andtemperature (no water vapour/aerosol) and no solvent-based interferences
High plasma temperature means elemental g p presponse is high, even for poorly ionized elements.
Also, elemental response is independent of compound, so compound independent calibration (CIC) is possible
Page 24
ESAC April 2008
( ) p
Courtesy Raimund Wahlen, LGC Teddington
I 12 00 (11 70 t 12 70) CICCAL3 DCarbon
Pesticide Analysis by GC-ICP-MS
0
2000000
4000000
6000000
8000000 Ion 12.00 (11.70 to 12.70): CICCAL3.DSingle ion chromatograms for C, P and S (right) and Cl, Br and I (below), extracted from multi-element GC-ICP-MS acquisition
3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.000
200000300000400000
Ion 31.00 (30.70 to 31.70): CICCAL3.DPhosphorusLow backgrounds (due to absence of
solvent) and good ionization (due to high temperature of “dry” plasma) leads to
3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.000100000
80000100000
Ion 34.00 (33.70 to 34.70): CICCAL3.DSulphur
temperature of dry plasma), leads to excellent signal to background and low LOD
Chl i
3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00020000400006000080000
3 00 4 00 5 00 6 00 7 00 8 00 9 00 10 00 11 00 12 00 13 000100000200000300000400000 Ion 35.00 (34.70 to 35.70): CICCAL3.DChlorine
Compound Conc (pg/uL - ppb) Calib Elements Elemental %3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00
200000300000400000
Ion 79.00 (78.70 to 79.70): CICCAL3.DBromine
p (pg pp )
Dichlobenil 610 Cl 41.3
2,4,6-TBA 287 Br 72.5
Ethoprop 39 P, S 12.8, 26.4
DBOB 100 Br 35 1
3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.000100000
200000300000400000 Ion 127.00 (126.70 to 127.70): CICCAL3.DIodine
DBOB 100 Br 35.1
Phorate 210 P, S 11.9, 36.9
PCNB 169 Cl 60.1
Terbufos 745 P, S 10.8, 33.3
Diazinon 976 P S 10 2 10 5
3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.000100000200000 Diazinon 976 P, S 10.2, 10.5
Malathion 107 P, S 9.37, 19.4
Dursban 569 Cl, P, S 30.3, 8.82, 9.15
Ioxynil (methyl ester) 50 I 66
TPP 158 P 50 3Table (right) shows components, concentrations and l t l i ht % i 1/10 dil t d CIC ti id i
Page 25
ESAC April 2008
TPP 158 P 50.3elemental weight % in 1/10 diluted CIC pesticide mix.
CIC - Sulphur in Pesticide MixSulphur elemental response is independent of the compound
Sulphur Response
Sulphur elemental response is independent of the compound
R2 = 0.9996800000
900000
TerbufosR 0.9996
600000
700000
Terbufos
Compound Concentration RT S conc S
400000
500000
Are
a
Di i
CompoundConcentration
pg/uLRT
(min)S conc (ppb)
S response
Ethoprop 385 6.4 101.64 23544
200000
300000
DursbanPhorate
Diazinon Ethoprop 385 6.4 101.64 23544Phorate 2102 7.01 775.64 261462Terbufos 7454 7.87 2280.92 785089Diazinon 9755 8.1 1024.28 360585
0
100000
0 500 1000 1500 2000 2500
EthopropMalathion
DursbanMalathion 1072 9.75 207.97 62313Dursban 5690 9.94 597.45 197738
0 500 1000 1500 2000 2500
Concentration (ppb) Pesticide compound LoD’s typically single ppb or sub-ppb
Page 26
ESAC April 2008
Chemical Warfare Agent (CWA) Regulation
• Chemical Weapons Convention of January 1993
• Enforcement began April 1997
• August 24th, 2006; Meeting of 180 countries (representing 98% of World Population) which are members of the OPCW.
O i i f h P hibi i f Ch i l W• Organization for the Prohibition of Chemical Weapons
• ~70,000 Metric Tons of Chemical Weapons Declared (24/8/2006)
• ~14,000 Metric Tons Destroyed (24/8/2006)
• 2006 Budget
•$96 Million
"Determined for the sake of all mankind, to exclude completely the possibility of the use of chemical weapons..."
Page 27
ESAC April 2008From Doug Richardson, Univ Cincinnati
Chemical Warfare Agent Analysis by ICP-MS
N er ve Agent s
G-Type V-Type
N er ve Agent s
G-Type V-Type
P
O
CH 3
F
O CH
CH
CH 3
CH 3
CH
P
O
CH 3
F
O CH
CH 3
CH 3P
O
CH 3 O CH 2 CH 3 P
O
CH 3 O CH 3P
O
CH 3
F
O CH
CH
CH 3
CH 3
CH
P
O
CH 3
F
O CH
CH 3
CH 3P
O
CH 3 O CH 2 CH 3 P
O
CH 3 O CH 3
F CH 3 CH 3
S oman (GD)F
S ar in (GB)S
NO
S
N
CH 3
F CH 3 CH 3
S oman (GD)F
S ar in (GB)S
NO
S
N
CH 3
VX
P
O
O
N
C NP
O
CH3
F
O
Russian VX (RVX)VX
P
O
O
N
C NP
O
CH3
F
O
Russian VX (RVX)VXTAbun (GA) Cycl osar in (GF) VXTAbun (GA) Cycl osar in (GF)
31P Selective Detection
All these agents contain a P atom, so ICP-MS can be used to identify and quantify the concentration of agent, based on the g ,consistent (compound independent) response for 31P. Nerve agents mostly decompose in the environment to MPA (via EMPA, IMPA, CMPA…)
Page 28
ESAC April 2008From Doug Richardson, Univ Cincinnati
CWA Analysis in Natural Samples by LC-ICP-MS
Right: Standards
Below: Unspiked and spiked 30000
35000
31P
1
2 4 78
Apple Juice
Column: Hamilton PRP-X100 Anion Exchange
20000
25000
onse
(CP
S)
3
56 9
10g
30000 2 10000
15000
Res
po 10
20000S)
31P 13 4 7 8 9 0 5 10 15 20 25
0
5000
Elution Order1. MPA2. H2PO4
-
20000
pons
e (C
PS
Apple Juice+ Spike (3ppm)
35 6
910
0 5 10 15 20 25
Time (min)
2 43. EPA4. DMHP5. PPA6. EMPA
10000
Res
p
Apple Juice
7. IMPA8. DEHP9. IPHEP10. IBHMP0 5 10 15 20 25
0
Time (min)
Page 29
ESAC April 2008
Time (min)
From Doug Richardson, Univ Cincinnati
Forensic Applications of ICP-MS –Glass Fragment AnalysisGlass Fragment Analysis
Almost any solid fragment collected from a suspect individual or location may be suitable for analysis using laser ablation ICP-MS – sample size as small as 50um diameter can be measured routinely
Glass samples can be analysed using simple screening scan (qualitative or semi-quantitative) or
lib t d i t ll h t i d fcalibrated against well-characterized reference glasses. NIST 600 series Trace Elements in Glass – eg NIST 612 ~ 50ppm
ESAC April 2008Images courtesy of New Wave Research
Trace Element Distribution Patterns (sumt 100%) S th ti Gl d U k S lto 100%) – Synthetic Glass and Unknown Samples
Page 31
ESAC April 2008Data courtesy of New Wave Research
Analysis of Bic Black Pen Inks
15 black Bic pen inksdifferent sources
Non-ablated Inkdifferent sources
analyzed in triplicate
24 elements / 26 isotopes
105 comparisons taken as pairs
86/105 (82%) pairs were discriminated by Pb
Of 19 remaining, 17 pairs were discriminated by CoAblated Ink
Of 2 remaining, 1 pair was discriminated by Ba
Both Zn and Cu have similar discriminating power to Co
Only 1 pair (<1%) could not be discriminated
Most useful discriminators:Pb, Co, Ba, Cu, Zn
Page 32
ESAC April 2008Images courtesy of FBI Academy
Eight Bic Black Pen Inks25.000
20.000
Unknown #2b is 02 Bic
15.000Al/Cu
Unknown #2b is 02 Bic(high W and low Mo)
10 000
Al/CuPbW/MoMo/CoZn/BaUnknown #8b is 08 Bic
10.000
(High Pb and low W)
5.000
0.00001 BicMean
02 BicMean
03 BicMean
04 BicMean
05 BicMean
06 BicMean
07 BicMean
08 BicMean
Unk (#2b) Unk (#8b)
Page 33
ESAC April 2008Data courtesy of FBI Academy
Metal Imaging Mass Spectrometry ( S)(MIMS)
10 µm tissue sections
LA-ICP-MS
Slice frozen tissue, e.g. brain
D t i itiImage reconstruction Raster Laser
Page 34
ESAC April 2008
Data acquisition
Images courtesy of Dominic Hare, UTS
MIMS maps of various elements in brain section of Parkinson’s disease rat model systemof Parkinson s disease rat model system
31P
Lesion side of brain(Parkinson-like)
Intact side of brain
Intensity increases from blue – green – yellow – red
EM 31P
56Fe 57Fe
Page 35
ESAC April 2008ESAC April 2008
Page 35
Images courtesy of Dominic Hare, UTS
Conclusions
ICP-MS offers a unique combination of:Wide elemental coverage (almost all elements can be measured)
Low limits of detection (typically 10’s ppq for easily ionized elements)Low limits of detection (typically 10 s ppq for easily ionized elements)
Wide dynamic range (from sub-ppt to 1000’s ppm)
Very rapid analysis (<3 minutes for full elemental screening, much less for consistent matrices and fewer elements)
Freedom from spectral interferences (using collision/reaction cell)
Tolerance of a wide range of matrices (due to high plasma temperature)g ( g p p )
Facility to link to chromatographic and direct solid sampling accessories, to further extend the range of applicationsg pp
Page 36
ESAC April 2008ESAC April 2008